Erschienen in: Journal of Children's Orthopaedics 2/2015

Open Access 01.04.2015 | Technical Note

A functional electrical stimulation system improves knee control in crouch gait

verfasst von: Sam Khamis, Raz Martikaro, Shlomo Wientroub, Yoram Hemo, Shlomo Hayek

Erschienen in: Journal of Children's Orthopaedics | Ausgabe 2/2015

Abstract

Background

Crouch gait is a major sagittal plane deviation in children diagnosed with cerebral palsy (CP). It is defined as a combination of excessive ankle dorsiflexion and knee and hip flexion throughout the stance phase. To the best of our knowledge, functional electrical stimulation (FES) has not been used to decrease the severity of crouch gait in CP subjects and assist in achieving lower limb extension.

Purpose

To evaluate the short- and long-term effects of FES to the quadriceps muscles in preventing crouch gait and achieving ankle plantar flexion, knee and hip extension at the stance phase.

Methods

An 18-year-old boy diagnosed with CP diplegia [Gross Motor Function Classification System (GMFCS) level II] was evaluated. The NESS L300® Plus neuroprosthesis system provided electrical stimulation of the quadriceps muscle. A three-dimensional gait analysis was performed using an eight-camera system measuring gait kinematics and spatiotemporal parameters while the subject walked shod only, with ground reaction ankle foot orthotics (GRAFOs) and using an FES device.

Results

Walking with the FES device showed an increase in the patient’s knee extension at midstance and increased knee maximal extension at the stance phase. In addition, the patient was able to ascend and descend stairs with a “step-through” pattern immediately after adjusting the FES device.

Conclusions

This report suggests that FES to the quadriceps muscles may affect knee extension at stance and decrease crouch gait, depending on the adequate passive range of motion of the hip, knee extension, and plantar flexion. Further studies are needed in order to validate these results.

Introduction

Cerebral palsy (CP) is the most common cause of upper motor neuron lesions in children, causing spasticity and muscle tendon contractures, leading to bony deformation, weakness, and loss of selective motor control. Rodda et al. and Kerr Graham classified sagittal gait patterns in spastic diplegia by providing a management algorithm based on these patterns [1, 2]. Crouch gait is a major sagittal plane deviation defined as a combination of excessive ankle dorsiflexion and knee and hip flexion throughout the stance phase. This gait pattern is commonly found in children afflicted with severe CP diplegia or quadriplegia. Crouch gait often rapidly progresses during the adolescent growth spurt. Walking aids are essential in crouch gait [3].
Several hypothesized causes for crouch gait include bony malformations causing lever arm dysfunction [4, 5], tight hip flexors [69], and weak muscle support [5, 1013]. One of the most common causes of crouch gait is the overlengthening of the gastrocsoleus, producing significant weakness of the gastrocsoleus extensor mechanism of the lower limbs [3]. Crouch gait can also be a natural history of the disability [1, 3].
Multilevel orthopedic surgery has been performed to correct severe crouch gait [14, 15]. The outcomes of surgical interventions vary [13, 16]. Treatment options are very limited and dependent on conservative intervention. Strengthening the extensor moment, particularly the gluteus maximus, may help improve both hip and knee extension [13].
The use of orthotics is essential in preventing crouch gait. Orthotics, such as ground reaction ankle foot orthotics (GRAFOs), are used to enhance knee extension ability. This type of orthotic frequently leads to discomfort, painful areas, and functional disability. For the GRAFOs to be effective, the lever arm should be within normal values [17].
Functional electrical stimulation (FES), a well-known intervention, has been used for many years to facilitate muscle groups during walking. The aim of the system is to facilitate the proper muscle group at the proper timing. So an accurate synchronization between the stimulator and gait cycle is essential. FES has been mainly used to control dorsiflexors muscles and prevent drop foot at the swing phase [18, 19]. Peroneal FES is widely used due to its feasibility, simplicity, and synchronization with gait.
The aim of the present report was to evaluate the short- and long-term effects of muscle stimulation in preventing crouch gait and achieving ankle plantar flexion and knee and hip extension at the stance phase by applying FES to the quadriceps muscles. Immediate and long-term effects were assessed on a subject diagnosed with CP diplegia with a crouch gait pattern.

Methods

Subject and experimental protocol

We present an 18-year-old boy diagnosed with CP diplegia (Gross Motor Function Classification System (GMFCS) level II] [20], without cognitive impairment. He was considered independent according to the activities of daily living. When walking outdoors, he wore GRAFOs bilaterally for 9 years with two quadripods. Indoors, he was able to ambulate without any orthotics. For long distances, he used a wheelchair or electric scooter. His main complaints were weakness, difficulty in prolonged walking and standing, ascending and descending stairs, and dependency on the orthotics. Clinical examination exhibited bilateral hip flexion contracture (Thomas test 25°), bilateral limited knee extension of 5°, bilateral extension lag of 30°, and severe hyperpronated midfoot break, leading to excessive external foot progression. Manual muscle testing revealed severe weakness: bilateral weak hip extensors (2/5), knee extensors (4/5), gastrocsoleus (2/5). No evidence of patella alta was found.

FES

We used the NESS L300® Plus neuroprosthesis system (Bioness Inc., Valencia, CA, USA), which delivers electrical stimulations to the common peroneal nerve, hamstrings, and quadriceps muscles (Fig. 1). For this case study, we used the NESS system for electrical stimulation of the quadriceps muscle. As described in detail by Hausdorff and Ring [21], this system has three main components that communicate via radio frequency signals: (1) a remote unit giving the patient control over the system; (2) an integrated stimulation unit with electrodes placed around the thigh (the electrodes of the thigh cuff—two oval cloth electrodes, proximal: 130 × 75 mm; distal: 120 × 63 mm—positioned over the quadriceps to extend the knee) [22]; and (3) a force-sensitive gait sensor placed beneath the foot to detect heel strike and toe off. Computed algorithms analyze the gait sensor’s data and information is then transmitted to the stimulation unit to induce knee extension at the appropriate time and exact duration [10]. Necessary adjustments may be performed by a clinician using a handheld computer (PDA) to set the stimulation intensity, pulse frequency, and gait parameters [9]. The thigh stimulation (quadriceps) can begin and end at any segment in the gait cycle, as defined by the clinician [22].

Gait analysis system

A three-dimensional gait analysis was performed using an eight-camera system (Vicon MX Giganet motion analysis system, Oxford Metrics, UK) at 100 Hz and a capturing volume of 3.5 m–3.5 m–2.5 m. Retro-reflective markers were applied on anatomical landmarks to capture gait performance according to the biomechanical model Plug-In-Gait developed by Vicon (based on the work of Murali Kadaba and Helen Hayes Hospital) [23, 24].

Evaluation protocol

The initial evaluation (T1) at the gait laboratory included a three-dimensional evaluation of walking at a self-selected normal speed on a 12-m walkway barefoot and wearing his own GRAFOs.
Two months after this initial evaluation (T2), he was fitted with an FES device placed on the quadriceps. During the fitting process, the stimulation parameters and timing of stimulation were set. Quadriceps stimulation was configured to start immediately after heel strike to the pre-swing phase. A symmetrical waveform was used with a phase duration of 300 μs, pulse rate 40 Hz, and intensity 40 ma. After adjusting the FES device, the subject practiced walking for 20 min prior to initiating the data capturing. After completing the fitting procedure, the subject used the system for training sessions at home on a treadmill. Every day, the subject walked 25 min and then trained for 20 min sitting. After 2 months, he walked with the FES device daily for 30 min and climbed two flights of stairs. After 6 months of conditioning (T3), he was evaluated at the gait laboratory.
At T2, he was evaluated while performing with and without the FES device (shod only). At T3, his gait was assessed in all four conditions: barefoot, GRAFOs, and with and without the FES device (shod only). Our goal was to compare the different conditions and any carryover effect on the maximum knee and hip extension and ankle plantar flexion throughout the stance phase and at midstance. Spatiotemporal parameters were compared to assess the functional effect. In addition to his walking evaluation, his ability to ascend and descend stairs was evaluated at T2 and T3.

Results

Tables 1, 2, and 3 summarize the kinematic results compared to typically developed (TD) subjects’ data recorded at the gait laboratory. Sagittal plane movement left and right knee at T3 are illustrated in Fig. 2.
Table 1
Sagittal plane kinematic measures of the knee at T1 [mean ± standard deviation (SD)] compared to typically developed (TD) subjects (mean ± SD)
 
T1
TD
Barefoot
GRAFO
Left
Right
Left
Right
Maximal knee extension at midstance (°)
36.35 (1.29)
32.08 (1.98)
39.64 (2.40)
34.01 (1.78)
8.95 (1.89)
Maximal knee extension at the stance phase (°)
33.01 (0.08)
30.74 (0.98)
36.24 (1.83)
32.15 (1.55)
8.86 (2.15)
Table 2
Sagittal plane kinematic measures of the ankle and knee at T2 compared to TD subjects (mean ± SD)
 
T2
TD
Shod
FES
Left
Right
Left
Right
Maximal dorsiflexion at midstance (°)
21.50 (0.95)
16.18 (0.37)
19.17 (1.95)
15.60 (0.72)
17.61 (1.22)
Maximal dorsiflexion at the stance phase (°)
31.44 (0.29)
24.43 (1.11)
28.80 (0.55)
21.74 (0.933)
19.19 (1.83)
Maximal knee extension at midstance (°)
39.16 (0.96)
37.36 (1.33)
33.23 (0.35)
36.26 (0.85)
8.95 (1.89)
Maximal knee extension at the stance phase (°)
38.63 (1.09)
37.23 (1.22)
32.36 (0.67)
34.44 (0.16)
8.86 (2.15)
Table 3
Sagittal plane kinematic measures of the ankle and knee at T3 compared to TD subjects (mean ± SD)
 
T3
TD
GRAFO
Shod
FES
Left
Right
Left
Right
Left
Right
Maximal dorsiflexion at midstance (°)
18.55 (1.55)
15.53 (1.54)
14.05 (1.59)
16.25 (1.19)
11.18 (0.84)
15.78 (0.29)
17.61 (1.22)
Maximal dorsiflexion at the stance phase (°)
25.55 (0.38)
21.37 (0.71)
31.44 (1.58)
30.75 (0.55)
30.32 (1.20)
30.44 (0.25)
19.19 (1.83)
Maximal knee extension at midstance (°)
38.03 (1.12)
34.99 (0.67)
34.99 (0.86)
35.22 (1.44)
33.42 (1.58)
33.45 (1.50)
8.95 (1.89)
Maximal knee extension at the stance phase (°)
36.69 (1.38)
34.26 (0.45)
33.71 (0.64)
34.29 (1.47)
32.65 (1.09)
31.04 (0.75)
8.86 (2.15)

T1 comparing barefoot to GRAFOs (Table 1)

At T1, the subject demonstrated a severe crouch gait pattern, with 36.35° left knee flexion at midstance and 32.08° on the right. Walking with GRAFOs at T1 did not improve knee kinematics. A decrease in knee extension was demonstrated at midstance and in maximum knee extension at the stance phase. Knee flexion increased by 3.29° on the left and 1.93° on the right at midstance, while maximum knee extension at the stance phase deteriorated by 3.23° on the left and 1.41° on the right. Similar results were observed at T3.

T2 comparing shod only to FES (Table 2)

At T2, walking with the FES device showed an increase in the patient’s ankle plantar flexion and knee extension at midstance (left ankle 2.33°, right ankle 0.58°, left knee 5.93°, right knee 1.1°) and increased knee maximal extension, and ankle plantar flexion at the stance phase (left ankle 2.64°, right ankle 2.69°, left knee 6.27°, right knee 2.79°). No change occurred at the hip level during the stance phase.

T3 comparing FES to GRAFOs and shod only conditions (Table 3)

The left knee was more extended at midstance with shoes only when compared to GRAFOs at T3 (38.03° and 34.99°, respectively). No change was found on the right. However, maximal ankle dorsiflexion increased with shoes only at the stance phase (by 5.89° on the left and 9.38° on the right), demonstrating that the GRAFOs’ maximal effect occurs at terminal stance.
Walking with the FES device at T3 revealed improvement in knee extension ability at midstance and maximal knee extension at the stance phase bilaterally compared to shoes only and GRAFOs conditions (maximum knee extension at stance, left and right): FES 32.65°, 31.04°, shod only 33.71°, 34.29°, and GRAFOs 36.69°, 34.26°. Similar results were found at T2; however, there was no effect on hip kinematics.

Spatiotemporal parameters (Table 4)

Spatiotemporal parameters at T3 revealed an increase in walking speed with the FES device and GRAFOs compared to walking barefoot (left: 0.73, 0.67, and 0.57 m/s, respectively; right: 0.74, 0.67, and 0.55 m/s, respectively). When walking with the FES device, the highest speed was attained due to an increase in step length and cadence. When comparing the patient’s walking speed with FES in T3 versus T2, a higher walking speed, cadence, and bigger step length occurred at T3.
Table 4
Spatiotemporal parameters measures at T3 compared to TD subjects (mean ± SD)
 
GRAFO 3
NESS off 3
NESS on 3
TD
Right
Left
Right
Left
Right
Left
Walking speed (m/s)
0.67 (0.03)
0.67 (0.05)
0.59 (0.03)
0.61 (0.03)
0.74 (0.02)
0.73 (0.02)
1.39
Step length (m)
0.52 (0.02)
0.52 (0.03)
0.52 (0.03)
0.5 (0.01)
0.55 (0.03)
0.56 (0.00)
0.74
Cadence (steps/min)
76.2 (2.24)
76.2 (3.16)
69.79 (1.47)
72.29 (0.44)
77.77 (1.47)
78.36 (3.46)
111

Stairs

At T2, the patient was able to ascend and descend stairs with a “step-to” pattern, whereas he was able to ascend and descend the stairs with a “step-through” pattern immediately after adjusting the FES. This effect was also seen at T3. In addition, at T3, he was able to “step through” stairs in all conditions.

Discussion

To the best of our knowledge, this is the first report evaluating the short- and long-term effects of an FES device on the quadriceps, compared to GRAFO, aimed to decrease the severity of crouch gait in a CP subject and assist in achieving lower limb extension. It is the first step to use rapidly developing technology to improve gait and functional ability in CP children by FES. Our patient’s gait was a typical crouch gait with excessive hip and knee flexion and dorsiflexion throughout the stance phase. He has been wearing GRAFOs for 9 years (for most of the day), together with two quadripods. However, with GRAFOs, he still complained of muscle fatigue when standing for prolonged periods of time and walking.
The subject preferred GRAFOs over barefoot and shoes only, in spite of the fact that his kinematic results did not reveal significant improvement in his lower limb extension. It was surprising to find that his knee kinematics still revealed a crouch gait pattern when comparing GRAFOs and shoes only conditions. It appears that the orthotics in this case prevented lower limb flexion rather than facilitated extension. The subject wore standard sports shoes with a higher heel than the forefoot, which might have caused forward tilting of the tibia, leading to flexion at all levels. The GRAFOs also supported him and, with proprioceptive feedback, facilitated leaning forward against the orthotics rather than knee extension.
These results concur with previous studies, where knee flexion was not reduced as a result of solid ankle foot orthosis (AFO) or GRAFOs at the stance phase. Hayek et al. [25] reported no change in ankle dorsiflexion and knee flexion angle at single limb support in children with diplegic CP using solid and articulated AFOs. Abel et al. [26] reported no significant change in knee position at stance as a result of fixed AFOs in the CP diplegia group with equinus or pes planovalgus deformities. Rethlefsen et al. [27] also reported no change in knee position during stance in subjects with spastic diplegia walking with fixed or articulated AFOs. The limited influence of orthotics on our subject with crouch gait can be explained by his severe external foot progression. Excessive external foot progression causes lever arm dysfunction, thus decreasing the efficiency of the orthotics [17]. Our results are consistent with the results of Postans and Granat [28]. They found an immediate increase in knee extension at mid terminal stance. The exact data were not reported and, thus, cannot be compared. However, they did not evaluate the long-term effects after daily use and training with an FES device nor did they compare the results to walking with shoes only or GRAFO.
Comparing all shod conditions from a kinematic point of view favored FES by improving the patient’s ability to extend the knees at the stance phase. The stimulation mainly affected the knee joint and with a small effect on the ankle. However, the hips were unaffected. The subject favored FES over GRAFOs due to achieving a more extended support on the lower limb without excessive orthotic support. In addition, FES gave him the feeling of actively strengthening and extending the knee as opposed to the passive effect of the orthotics. However, the effect of “attaining stronger muscles” as described by the subject should be further evaluated. In addition, the most impressive effect was that FES provided much more support, increased the amount of time the patient was able to walk or stand, and achieved a more upright position. This was consistent for 6 months. The positive effect can be attributed to the trend seen in Fig. 2, which shows his ability to achieve a more extended knee position throughout the stance phase. Some difficulty with the use of the FES device was reported due to the discomfort caused by sitting while the device was connected to his thigh.
It should be emphasized that FES will assist in achieving knee extension by supporting the muscle force against restraints preventing knee extension. Resistance to extension can be a short muscle–tendon or joint contractures, such as knee or hip joint flexion contracture, as is common in crouch gait. Our patient had a knee flexion contracture of 5° and tight hip flexors (Thomas test 15°). In cases of severe contractures, the FES effect might be limited.
Walking speed increased at T3, indicating that the patient acclimated to the FES device over time. The walking speed increased due to the increased step length. A carryover effect might have occurred over time on the left side only.
The effect of the stimulation on the patient’s functional ability to ascend and descend stairs was immediate, with no need for reeducation. At T2, he was able to ascend and descend stairs, “step through” versus “step to” (which had been his strategy for years), immediately after adjusting to the FES device, which was achieved by receiving more support from the stance limb and its extension, rather than leaning on the device. However, the kinematics were not captured in the stairs condition. A carryover effect with stairs occurred at T3. The subject was able to ascend and descend stairs in all conditions in a “step-through” pattern. It seems that the effect is not only due to stronger quadriceps but also due to a motor learning effect.

Conclusions

This report suggests that functional electrical stimulation (FES) of the quadriceps muscles may have an important effect on achieving knee extension at stance and decreasing crouch gait, depending on an adequate passive range of motion of the hip, knee extension, and plantar flexion. Patient satisfaction was high due to improved functional ability and the feeling of active strengthening. The positive effect observed demonstrates the potential of FES as a means of achieving improved gait kinematics and functional ability in a patient with crouch gait. Further studies are needed in order to validate these results.

Conflict of interest

The authors declare that they have no conflict of interest.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Pädiatrie

Kombi-Abonnement

Mit e.Med Pädiatrie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Pädiatrie, den Premium-Inhalten der pädiatrischen Fachzeitschriften, inklusive einer gedruckten Pädiatrie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Kerr Graham H, Selber P (2003) Musculoskeletal aspects of cerebral palsy. J Bone Joint Surg Br 85(2):157–166CrossRefPubMed Kerr Graham H, Selber P (2003) Musculoskeletal aspects of cerebral palsy. J Bone Joint Surg Br 85(2):157–166CrossRefPubMed
2.
Zurück zum Zitat Vuillermin C, Rodda J, Rutz E, Shore BJ, Smith K, Graham HK (2011) Severe crouch gait in spastic diplegia can be prevented: a population-based study. J Bone Joint Surg Br 93(12):1670–1675CrossRefPubMed Vuillermin C, Rodda J, Rutz E, Shore BJ, Smith K, Graham HK (2011) Severe crouch gait in spastic diplegia can be prevented: a population-based study. J Bone Joint Surg Br 93(12):1670–1675CrossRefPubMed
3.
Zurück zum Zitat Schwartz M, Lakin G (2003) The effect of tibial torsion on the dynamic function of the soleus during gait. Gait Posture 17(2):113–118CrossRefPubMed Schwartz M, Lakin G (2003) The effect of tibial torsion on the dynamic function of the soleus during gait. Gait Posture 17(2):113–118CrossRefPubMed
4.
Zurück zum Zitat Gage JR (2004) The treatment of gait problems in cerebral palsy. Mac Keith Press, London Gage JR (2004) The treatment of gait problems in cerebral palsy. Mac Keith Press, London
5.
Zurück zum Zitat Roosth HP (1971) Flexion deformity of the hip and knee in spastic cerebral palsy: treatment by early release of spastic hip-flexor muscles. J Bone Joint Surg Am 53(8):1489–1510PubMed Roosth HP (1971) Flexion deformity of the hip and knee in spastic cerebral palsy: treatment by early release of spastic hip-flexor muscles. J Bone Joint Surg Am 53(8):1489–1510PubMed
6.
Zurück zum Zitat Reimers J (1973) Static and dynamic problems in spastic cerebral palsy. J Bone Joint Surg Br 55(4):822–827PubMed Reimers J (1973) Static and dynamic problems in spastic cerebral palsy. J Bone Joint Surg Br 55(4):822–827PubMed
7.
Zurück zum Zitat Bleck EE (1987) Orthopaedic management in cerebral palsy. Mac Keith Press, London Bleck EE (1987) Orthopaedic management in cerebral palsy. Mac Keith Press, London
8.
Zurück zum Zitat Novacheck TF, Trost JP, Schwartz MH (2002) Intramuscular psoas lengthening improves dynamic hip function in children with cerebral palsy. J Pediatr Orthop 22(2):158–164PubMed Novacheck TF, Trost JP, Schwartz MH (2002) Intramuscular psoas lengthening improves dynamic hip function in children with cerebral palsy. J Pediatr Orthop 22(2):158–164PubMed
9.
Zurück zum Zitat Damiano DL, Vaughan CL, Abel MF (1995) Muscle response to heavy resistance exercise in children with spastic cerebral palsy. Dev Med Child Neurol 37(8):731–739CrossRefPubMed Damiano DL, Vaughan CL, Abel MF (1995) Muscle response to heavy resistance exercise in children with spastic cerebral palsy. Dev Med Child Neurol 37(8):731–739CrossRefPubMed
10.
Zurück zum Zitat Wiley ME, Damiano DL (1998) Lower-extremity strength profiles in spastic cerebral palsy. Dev Med Child Neurol 40(2):100–107CrossRefPubMed Wiley ME, Damiano DL (1998) Lower-extremity strength profiles in spastic cerebral palsy. Dev Med Child Neurol 40(2):100–107CrossRefPubMed
11.
Zurück zum Zitat Beals RK (2001) Treatment of knee contracture in cerebral palsy by hamstring lengthening, posterior capsulotomy, and quadriceps mechanism shortening. Dev Med Child Neurol 43(12):802–805CrossRefPubMed Beals RK (2001) Treatment of knee contracture in cerebral palsy by hamstring lengthening, posterior capsulotomy, and quadriceps mechanism shortening. Dev Med Child Neurol 43(12):802–805CrossRefPubMed
12.
Zurück zum Zitat Arnold AS, Anderson FC, Pandy MG, Delp SL (2005) Muscular contributions to hip and knee extension during the single limb stance phase of normal gait: a framework for investigating the causes of crouch gait. J Biomech 38(11):2181–2189CrossRefPubMed Arnold AS, Anderson FC, Pandy MG, Delp SL (2005) Muscular contributions to hip and knee extension during the single limb stance phase of normal gait: a framework for investigating the causes of crouch gait. J Biomech 38(11):2181–2189CrossRefPubMed
13.
Zurück zum Zitat Rodda JM, Graham HK, Carson L, Galea MP, Wolfe R (2004) Sagittal gait patterns in spastic diplegia. J Bone Joint Surg Br 86(2):251–258CrossRefPubMed Rodda JM, Graham HK, Carson L, Galea MP, Wolfe R (2004) Sagittal gait patterns in spastic diplegia. J Bone Joint Surg Br 86(2):251–258CrossRefPubMed
14.
Zurück zum Zitat Rodda JM, Graham HK, Nattrass GR, Galea MP, Baker R, Wolfe R (2006) Correction of severe crouch gait in patients with spastic diplegia with use of multilevel orthopaedic surgery. J Bone Joint Surg Am 88(12):2653–2664CrossRefPubMed Rodda JM, Graham HK, Nattrass GR, Galea MP, Baker R, Wolfe R (2006) Correction of severe crouch gait in patients with spastic diplegia with use of multilevel orthopaedic surgery. J Bone Joint Surg Am 88(12):2653–2664CrossRefPubMed
15.
Zurück zum Zitat Stout JL, Gage JR, Schwartz MH, Novacheck TF (2008) Distal femoral extension osteotomy and patellar tendon advancement to treat persistent crouch gait in cerebral palsy. J Bone Joint Surg Am 90(11):2470–2484CrossRefPubMed Stout JL, Gage JR, Schwartz MH, Novacheck TF (2008) Distal femoral extension osteotomy and patellar tendon advancement to treat persistent crouch gait in cerebral palsy. J Bone Joint Surg Am 90(11):2470–2484CrossRefPubMed
16.
Zurück zum Zitat Arnold AS, Blemker SS, Delp SL (2001) Evaluation of a deformable musculoskeletal model for estimating muscle-tendon lengths during crouch gait. Ann Biomed Eng 29(3):263–274CrossRefPubMed Arnold AS, Blemker SS, Delp SL (2001) Evaluation of a deformable musculoskeletal model for estimating muscle-tendon lengths during crouch gait. Ann Biomed Eng 29(3):263–274CrossRefPubMed
18.
Zurück zum Zitat Laufer Y, Ring H, Sprecher E, Hausdorff JM (2009) Gait in individuals with chronic hemiparesis: one-year follow-up of the effects of a neuroprosthesis that ameliorates foot drop. J Neurol Phys Ther 33(2):104–110CrossRefPubMed Laufer Y, Ring H, Sprecher E, Hausdorff JM (2009) Gait in individuals with chronic hemiparesis: one-year follow-up of the effects of a neuroprosthesis that ameliorates foot drop. J Neurol Phys Ther 33(2):104–110CrossRefPubMed
19.
Zurück zum Zitat Danino B, Khamis S, Hemo Y, Batt R, Snir E, Wientroub S, Hayek S (2013) The efficacy of neuroprosthesis in young hemiplegic patients, measured by three different gait indices: early results. J Child Orthop 7(6):537–542CrossRefPubMedCentralPubMed Danino B, Khamis S, Hemo Y, Batt R, Snir E, Wientroub S, Hayek S (2013) The efficacy of neuroprosthesis in young hemiplegic patients, measured by three different gait indices: early results. J Child Orthop 7(6):537–542CrossRefPubMedCentralPubMed
20.
Zurück zum Zitat Palisano R, Rosenbaum P, Walter S, Russell D, Wood E, Galuppi B (1997) Development and reliability of a system to classify gross motor function in children with cerebral palsy. Dev Med Child Neurol 39(4):214–223CrossRefPubMed Palisano R, Rosenbaum P, Walter S, Russell D, Wood E, Galuppi B (1997) Development and reliability of a system to classify gross motor function in children with cerebral palsy. Dev Med Child Neurol 39(4):214–223CrossRefPubMed
21.
Zurück zum Zitat Hausdorff JM, Ring H (2008) Effects of a new radio frequency-controlled neuroprosthesis on gait symmetry and rhythmicity in patients with chronic hemiparesis. Am J Phys Med Rehabil 87(1):4–13CrossRefPubMed Hausdorff JM, Ring H (2008) Effects of a new radio frequency-controlled neuroprosthesis on gait symmetry and rhythmicity in patients with chronic hemiparesis. Am J Phys Med Rehabil 87(1):4–13CrossRefPubMed
22.
Zurück zum Zitat Springer S, Vatine JJ, Lipson R, Wolf A, Laufer Y (2012) Effects of dual-channel functional electrical stimulation on gait performance in patients with hemiparesis. Sci World J 2012:530906CrossRef Springer S, Vatine JJ, Lipson R, Wolf A, Laufer Y (2012) Effects of dual-channel functional electrical stimulation on gait performance in patients with hemiparesis. Sci World J 2012:530906CrossRef
23.
Zurück zum Zitat Kadaba MP, Ramakrishnan HK, Wootten ME, Gainey J, Gorton G, Cochran GV (1989) Repeatability of kinematic, kinetic, and electromyographic data in normal adult gait. J Orthop Res 7(6):849–860CrossRefPubMed Kadaba MP, Ramakrishnan HK, Wootten ME, Gainey J, Gorton G, Cochran GV (1989) Repeatability of kinematic, kinetic, and electromyographic data in normal adult gait. J Orthop Res 7(6):849–860CrossRefPubMed
24.
Zurück zum Zitat Kadaba MP, Ramakrishnan HK, Wootten ME (1990) Measurement of lower extremity kinematics during level walking. J Orthop Res 8(3):383–392CrossRefPubMed Kadaba MP, Ramakrishnan HK, Wootten ME (1990) Measurement of lower extremity kinematics during level walking. J Orthop Res 8(3):383–392CrossRefPubMed
25.
Zurück zum Zitat Hayek S, Hemo Y, Chamis S, Bat R, Segev E, Wientroub S, Yzhar Z (2007) The effect of community-prescribed ankle-foot orthoses on gait parameters in children with spastic cerebral palsy. J Child Orthop 1(6):325–332CrossRefPubMedCentralPubMed Hayek S, Hemo Y, Chamis S, Bat R, Segev E, Wientroub S, Yzhar Z (2007) The effect of community-prescribed ankle-foot orthoses on gait parameters in children with spastic cerebral palsy. J Child Orthop 1(6):325–332CrossRefPubMedCentralPubMed
26.
Zurück zum Zitat Abel MF, Juhl GA, Vaughan CL, Damiano DL (1998) Gait assessment of fixed ankle-foot orthoses in children with spastic diplegia. Arch Phys Med Rehabil 79(2):126–133CrossRefPubMed Abel MF, Juhl GA, Vaughan CL, Damiano DL (1998) Gait assessment of fixed ankle-foot orthoses in children with spastic diplegia. Arch Phys Med Rehabil 79(2):126–133CrossRefPubMed
27.
Zurück zum Zitat Rethlefsen S, Kay R, Dennis S, Forstein M, Tolo V (1999) The effects of fixed and articulated ankle-foot orthoses on gait patterns in subjects with cerebral palsy. J Pediatr Orthop 19(4):470–474CrossRefPubMed Rethlefsen S, Kay R, Dennis S, Forstein M, Tolo V (1999) The effects of fixed and articulated ankle-foot orthoses on gait patterns in subjects with cerebral palsy. J Pediatr Orthop 19(4):470–474CrossRefPubMed
28.
Zurück zum Zitat Postans NJ, Granat MH (2005) Effect of functional electrical stimulation, applied during walking, on gait in spastic cerebral palsy. Dev Med Child Neurol 47(1):46–52CrossRefPubMed Postans NJ, Granat MH (2005) Effect of functional electrical stimulation, applied during walking, on gait in spastic cerebral palsy. Dev Med Child Neurol 47(1):46–52CrossRefPubMed
Metadaten
Titel
A functional electrical stimulation system improves knee control in crouch gait
verfasst von
Sam Khamis
Raz Martikaro
Shlomo Wientroub
Yoram Hemo
Shlomo Hayek
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Journal of Children's Orthopaedics / Ausgabe 2/2015
Print ISSN: 1863-2521
Elektronische ISSN: 1863-2548
DOI
https://doi.org/10.1007/s11832-015-0651-2

Weitere Artikel der Ausgabe 2/2015

Journal of Children's Orthopaedics 2/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.