Ceramic-Ceramic Bearing Decreases Osteolysis: A 20-year Study versus Ceramic-Polyethylene on the Contralateral Hip

verfasst von: Philippe Hernigou, MD, Sebastien Zilber, MD, Paolo Filippini, MD, Alexandre Poignard, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 9/2009

Einloggen, um Zugang zu erhalten

Abstract

Although ceramic implants have been in use for many years and they are intended to minimize wear debris it is unknown whether alumina-on-alumina or alumina-on-polyethylene produce less wear and osteolysis. We therefore investigated wear and osteolysis on 28 bilateral arthroplasties (one ceramic-ceramic and the contralateral ceramic-polyethylene) of patients who had survived 20 years without revision and without loosening of either hip. Osteolysis was identified on anteroposterior pelvic radiographs and 3-D volume from CT scans. The number of osteolytic lesions detected with CT scan was higher than with radiographs. The number of lesions was higher on the side with the alumina-PE couple. With a similar length of followup on each side, the surface and the volume of osteolysis were consistently higher on the side with the alumina-PE couple. We found no correlation between the volume of osteolysis and the volume of estimated wear in each couple of friction. Hips with osteolysis had a lower Harris score.
Level of Evidence: Level II, prognostic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Amstutz HC, Campbell P, Kossovsky N, Clarke IC. Mechanism and clinical significance of wear debris-induced osteolysis. Clin Orthop Relat Res. 1992;276:7–18.PubMed Amstutz HC, Campbell P, Kossovsky N, Clarke IC. Mechanism and clinical significance of wear debris-induced osteolysis. Clin Orthop Relat Res. 1992;276:7–18.PubMed
2.
3.
Zurück zum Zitat Bizot P, Banallec L, Sedel L, Nizard R. Alumina-on-alumina total hip protheses in patients 40 years of age or younger. Clin Orthop Relat Res. 2000;379:68–76.PubMedCrossRef Bizot P, Banallec L, Sedel L, Nizard R. Alumina-on-alumina total hip protheses in patients 40 years of age or younger. Clin Orthop Relat Res. 2000;379:68–76.PubMedCrossRef
4.
Zurück zum Zitat Boehler M, Knahr K, Plenk H, Long term results of uncemented alumina acetabular implants. J Bone Joint Surg Br. 1994;76:53–59.PubMed Boehler M, Knahr K, Plenk H, Long term results of uncemented alumina acetabular implants. J Bone Joint Surg Br. 1994;76:53–59.PubMed
5.
Zurück zum Zitat Borssen B, Karrholm J, Snorrason F. Osteolysis after ceramic-on-ceramic hip arthroplasty. A case report. Acta Orthop Scandinavic. 1991;62:73–75.CrossRef Borssen B, Karrholm J, Snorrason F. Osteolysis after ceramic-on-ceramic hip arthroplasty. A case report. Acta Orthop Scandinavic. 1991;62:73–75.CrossRef
6.
Zurück zum Zitat Bos I, Henssge EJ, Willmann G. Morphological characterization of joint capsules around hip prostheses with alumina or alumina combinations. In: Puhl W, ed. Die Keramikpaarung Biolox in der Hueftendoprothetik. Stuttgart, Germany: Enke Verlag; 1996:24–30. Bos I, Henssge EJ, Willmann G. Morphological characterization of joint capsules around hip prostheses with alumina or alumina combinations. In: Puhl W, ed. Die Keramikpaarung Biolox in der Hueftendoprothetik. Stuttgart, Germany: Enke Verlag; 1996:24–30.
7.
Zurück zum Zitat Boutin P. Arthroplastie totale de la hanche par prothèse en alumine frittee. Etude expérimentale et premières applications cliniques. Rev Chir Orthop. 1972;58:229–246.PubMed Boutin P. Arthroplastie totale de la hanche par prothèse en alumine frittee. Etude expérimentale et premières applications cliniques. Rev Chir Orthop. 1972;58:229–246.PubMed
8.
Zurück zum Zitat Boutin P, Blanquaert D. Le frottement alumine-alumine en chirurgie de la hanche. 1205 arthroplasties totales : avril 1970-juin 1980. Rev Chir Orthop. 1981;67:279–287.PubMed Boutin P, Blanquaert D. Le frottement alumine-alumine en chirurgie de la hanche. 1205 arthroplasties totales : avril 1970-juin 1980. Rev Chir Orthop. 1981;67:279–287.PubMed
9.
Zurück zum Zitat D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.PubMedCrossRef D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.PubMedCrossRef
10.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.
11.
Zurück zum Zitat Fenollosa J, Seminario P, Montijano C. Ceramic hip protheses in young patients. Clin Orthop Relat Res. 2000;379:55–67.PubMedCrossRef Fenollosa J, Seminario P, Montijano C. Ceramic hip protheses in young patients. Clin Orthop Relat Res. 2000;379:55–67.PubMedCrossRef
12.
Zurück zum Zitat Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5 year follow-up study. J Bone Joint Surg Am. 2002;84:69–77.PubMed Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5 year follow-up study. J Bone Joint Surg Am. 2002;84:69–77.PubMed
13.
Zurück zum Zitat Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed
14.
Zurück zum Zitat Hernigou P, Bahramy T. Zirconia and alumina ceramics in comparison with metal heads: polyethylene wear after a minimum ten year follow-up. J Bone Joint Surg Br. 2003;85:504–509.PubMedCrossRef Hernigou P, Bahramy T. Zirconia and alumina ceramics in comparison with metal heads: polyethylene wear after a minimum ten year follow-up. J Bone Joint Surg Br. 2003;85:504–509.PubMedCrossRef
15.
Zurück zum Zitat Hernigou P, Intrator L, Bahrami T, Bensoussan A, Farcet JP. Interleukin 6 in the blood of patients with total hip arthroplasty without loosening. Clin Orthop Relat Res. 1999;366:147–154.PubMedCrossRef Hernigou P, Intrator L, Bahrami T, Bensoussan A, Farcet JP. Interleukin 6 in the blood of patients with total hip arthroplasty without loosening. Clin Orthop Relat Res. 1999;366:147–154.PubMedCrossRef
16.
Zurück zum Zitat Kabo JM, Gebhard JS, Loren G, Amstutz H. In vivo wear of polyethylene acetabular component. J Bone Joint Surg Br. 1993;75:254–258.PubMed Kabo JM, Gebhard JS, Loren G, Amstutz H. In vivo wear of polyethylene acetabular component. J Bone Joint Surg Br. 1993;75:254–258.PubMed
17.
Zurück zum Zitat Kitamura N, Leung S, Engh CA. Characteristics of pelvic osteolysis on computed tomography after total hips arthroplasty. Clin Orthop Relat Res. 2005;441:291–297.PubMedCrossRef Kitamura N, Leung S, Engh CA. Characteristics of pelvic osteolysis on computed tomography after total hips arthroplasty. Clin Orthop Relat Res. 2005;441:291–297.PubMedCrossRef
18.
Zurück zum Zitat Knahr K, Bohler M, Frank P, Plenk H, Salzer M. Survival analysis of an uncemented ceramic acetabular component in total hip replacement. Arch Orthop Trauma Surg. 1987;106:297–300.PubMedCrossRef Knahr K, Bohler M, Frank P, Plenk H, Salzer M. Survival analysis of an uncemented ceramic acetabular component in total hip replacement. Arch Orthop Trauma Surg. 1987;106:297–300.PubMedCrossRef
19.
Zurück zum Zitat Kummer FJ, Stuchin SA, Frankel VH. Analysis of removed Autophor ceramic-on-ceramic components. J Arthroplasty. 1990;5:28–33.PubMedCrossRef Kummer FJ, Stuchin SA, Frankel VH. Analysis of removed Autophor ceramic-on-ceramic components. J Arthroplasty. 1990;5:28–33.PubMedCrossRef
20.
Zurück zum Zitat Lerouge S, Huk O, Yahia L. Ceramic-ceramic or metal-PE. A comparison of periprosthetic tissues from loosened total hip arthroplasty. J Bone Joint Surg Br. 1997;79:135–139.PubMedCrossRef Lerouge S, Huk O, Yahia L. Ceramic-ceramic or metal-PE. A comparison of periprosthetic tissues from loosened total hip arthroplasty. J Bone Joint Surg Br. 1997;79:135–139.PubMedCrossRef
21.
Zurück zum Zitat Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed
22.
Zurück zum Zitat Lusty PJ, Watson A, Tuke MA, Walter WL, Walter WK, Zicat B. Orientation and wear of the acetabular component in third generation alumina-on-alumina ceramic bearings. J Bone Joint Surg Br. 2007;89:1158–1164.PubMedCrossRef Lusty PJ, Watson A, Tuke MA, Walter WL, Walter WK, Zicat B. Orientation and wear of the acetabular component in third generation alumina-on-alumina ceramic bearings. J Bone Joint Surg Br. 2007;89:1158–1164.PubMedCrossRef
23.
Zurück zum Zitat Maloney WJ, Jasty M, Harris WH, Galante JO, Callaghan JJ. Endosteal erosion in association with stable uncemented femoral components. J Bone Joint Surg Am. 1990;72:1025–1034.PubMed Maloney WJ, Jasty M, Harris WH, Galante JO, Callaghan JJ. Endosteal erosion in association with stable uncemented femoral components. J Bone Joint Surg Am. 1990;72:1025–1034.PubMed
24.
Zurück zum Zitat Maloney WJ, Jasty M, Rosenberg A, Harris WH. Bone lysis in well-fixed cemented femoral components. J Bone Joint Surg Br. 1990;72:966–970.PubMed Maloney WJ, Jasty M, Rosenberg A, Harris WH. Bone lysis in well-fixed cemented femoral components. J Bone Joint Surg Br. 1990;72:966–970.PubMed
25.
Zurück zum Zitat Mittelmeier H, Heisel J. Sixteen-years’ experience with ceramic hip prostheses. Clin Orthop Relat Res. 1992;282:64–72.PubMed Mittelmeier H, Heisel J. Sixteen-years’ experience with ceramic hip prostheses. Clin Orthop Relat Res. 1992;282:64–72.PubMed
26.
Zurück zum Zitat Nevelos JE, Prudhommeaux F, Hamadouche M, Doyle C, Ingham E, Meunier A, Nevelos AB, Sedel L, Fisher J. Comparative analysis of two different types of alumina-alumina hip prosthesis retrieved for aseptic loosening. J Bone Joint Surg Br. 2001;83:598–603.PubMed Nevelos JE, Prudhommeaux F, Hamadouche M, Doyle C, Ingham E, Meunier A, Nevelos AB, Sedel L, Fisher J. Comparative analysis of two different types of alumina-alumina hip prosthesis retrieved for aseptic loosening. J Bone Joint Surg Br. 2001;83:598–603.PubMed
27.
Zurück zum Zitat Puri L, Wixson RL, Stern SH, Kohli JBA, Hendrix RW, Stulberg SD. Use of helical computed tomography for the assessment of acetabular osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 2002;84:609–614.PubMed Puri L, Wixson RL, Stern SH, Kohli JBA, Hendrix RW, Stulberg SD. Use of helical computed tomography for the assessment of acetabular osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 2002;84:609–614.PubMed
28.
Zurück zum Zitat Refior JJ, Plitz W, Walter A. Ex vivo and in vitro analysis of the alumina/alumina bearing system for hip joint prostheses. Bioceramics. 1997;10:127–130.CrossRef Refior JJ, Plitz W, Walter A. Ex vivo and in vitro analysis of the alumina/alumina bearing system for hip joint prostheses. Bioceramics. 1997;10:127–130.CrossRef
29.
Zurück zum Zitat Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone and Joint Surg Am. 1992;74:849–863. Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone and Joint Surg Am. 1992;74:849–863.
30.
Zurück zum Zitat Schuller HM, Marti RK. Ten-year socket wear in 66 hip arthroplasties. Ceramic versus metal heads. Acta Orthop Scand. 1990;61:240–243.PubMedCrossRef Schuller HM, Marti RK. Ten-year socket wear in 66 hip arthroplasties. Ceramic versus metal heads. Acta Orthop Scand. 1990;61:240–243.PubMedCrossRef
31.
Zurück zum Zitat Sedel L, Kerboull L, Christel. Alumina-on-alumina hip replacement. Results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.PubMed Sedel L, Kerboull L, Christel. Alumina-on-alumina hip replacement. Results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.PubMed
32.
Zurück zum Zitat Sedel L, Nizard R, Kerboull L, Witwoet J. Alumina-alumina hip replacement in patients younger than 50 years old. Clin Orthop Relat Res. 1994;298:175–183.PubMed Sedel L, Nizard R, Kerboull L, Witwoet J. Alumina-alumina hip replacement in patients younger than 50 years old. Clin Orthop Relat Res. 1994;298:175–183.PubMed
33.
Zurück zum Zitat Shih CH, Wu CC, Lee ZL, Yang WE. Localized femoral osteolysis in cementless ceramic total hip arthroplasty. Orthop Rev. 1994;23:325–328.PubMed Shih CH, Wu CC, Lee ZL, Yang WE. Localized femoral osteolysis in cementless ceramic total hip arthroplasty. Orthop Rev. 1994;23:325–328.PubMed
34.
Zurück zum Zitat Sugano N, Nishii T, Nakata K, Masuhara K, Takaoka K. Polyethylene sockets and alumina ceramic heads in cemented total hip arthroplasty: a ten-year study. J Bone and Joint Surg Br. 1995;77:548–556. Sugano N, Nishii T, Nakata K, Masuhara K, Takaoka K. Polyethylene sockets and alumina ceramic heads in cemented total hip arthroplasty: a ten-year study. J Bone and Joint Surg Br. 1995;77:548–556.
35.
Zurück zum Zitat Urban JA, Garvin KL, Boese CK, Bryson L, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one-year results. J Bone Joint Surg Am. 2001;83:1688–1694.PubMed Urban JA, Garvin KL, Boese CK, Bryson L, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-polyethylene bearing surfaces in total hip arthroplasty: seventeen to twenty-one-year results. J Bone Joint Surg Am. 2001;83:1688–1694.PubMed
36.
Zurück zum Zitat Walter A. On the material and the tribology of alumina-alumina coupling for hip joint prostheses. Clin Orthop Relat Res. 1992;282:31–46.PubMed Walter A. On the material and the tribology of alumina-alumina coupling for hip joint prostheses. Clin Orthop Relat Res. 1992;282:31–46.PubMed
37.
Zurück zum Zitat William S, Isaac G, Portier N, Fischer J, Older J. Long term radiographic assessment of cemented polethylene acetabular cups. Clin Orthop Relat Res. 2008;466:366–372.CrossRef William S, Isaac G, Portier N, Fischer J, Older J. Long term radiographic assessment of cemented polethylene acetabular cups. Clin Orthop Relat Res. 2008;466:366–372.CrossRef
38.
Metadaten
Titel
Ceramic-Ceramic Bearing Decreases Osteolysis: A 20-year Study versus Ceramic-Polyethylene on the Contralateral Hip
verfasst von
Philippe Hernigou, MD
Sebastien Zilber, MD
Paolo Filippini, MD
Alexandre Poignard, MD
Publikationsdatum
01.09.2009
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 9/2009
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-0773-2

Weitere Artikel der Ausgabe 9/2009

Clinical Orthopaedics and Related Research® 9/2009 Zur Ausgabe

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Factors Influencing the Stability of Stems Fixed with Impaction Graft in Vitro

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

The Classic: On Rest and Pain: Lecture XIV

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Proximal Femoral Reconstructions with Bone Impaction Grafting and Metal Mesh

Symposium: Papers Presented at the Annual Closed Meeting of the International Hip Society

Accuracy of Computer-assisted Navigation for Femoral Head Resurfacing Decreases in Hips with Abnormal Anatomy

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.