Clinical Comparison of Polyethylene Wear with Zirconia or Cobalt-Chromium Femoral Heads

verfasst von: Maiken Stilling, MD, Kjeld Anton Nielsen, MD, Kjeld Søballe, MD, DMSc, Ole Rahbek, MD, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 10/2009

Einloggen, um Zugang zu erhalten

Abstract

Ceramic femoral heads were developed to reduce the wear of conventional ultrahigh-molecular-weight polyethylene (UHMWPE) bearing surfaces in THA. We compared the wear rates of PE acetabular components bearing against femoral heads of zirconia (Zr) or cobalt-chromium (CoCr) in young patients. One surgeon inserted CoCr femoral heads in all 33 patients (33 hips) having THA for primary osteoarthritis from 1996 to 1997 and then Zr femoral heads in all 34 patients (36 hips) from 1998 to 1999. The mean age of the entire cohort was 52.5 years (range, 29–64 years). The shells were solid and hydroxyapatite (HA) -coated, liners were argon-sterilized UHMWPE, and head size was 28 mm. The minimum clinical followup was 56 months (mean, 65 months; range, 56–77 months); minimum 5-year radiographs were available for 62 of the 68 patients. Wear analysis of digitized anteroposterior (AP) radiographs was performed with a computerized method. Demographic data were comparable in the two groups. Mean femoral head penetration rate was similar in the two types of heads (CoCr: 0.25 mm/year; range, 0.21–0.33 mm/year; Zr: 0.23 mm/year; range, 0.20–0.29 mm/year), as was mean linear wear (CoCr: 1.22 mm; range, 0.28–3.78 mm; Zr: 1.11 mm; range, 0.15–2.05 mm). There were no revisions. These data support skepticism regarding the clinical wear advantage of Zr compared with CoCr femoral heads. The explanation for the clinical similarity of wear, despite the theoretical advantages of ceramic heads, needs further investigation.
Level of Evidence: Level III, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Allain J, Le Mouel S, Goutallier D, Voisin MC. Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br. 1999;81:835–842.PubMedCrossRef Allain J, Le Mouel S, Goutallier D, Voisin MC. Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br. 1999;81:835–842.PubMedCrossRef
2.
Zurück zum Zitat Altman DG. Practical Statistics for Medical Research. 1995 ed. London, UK: Chapman & Hall; 2009. Altman DG. Practical Statistics for Medical Research. 1995 ed. London, UK: Chapman & Hall; 2009.
3.
Zurück zum Zitat Bauer TW. Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating. J Bone Joint Surg Br. 1998;80:745.PubMed Bauer TW. Severe osteolysis after third-body wear due to hydroxyapatite particles from acetabular cup coating. J Bone Joint Surg Br. 1998;80:745.PubMed
4.
Zurück zum Zitat Bauer TW, Hirakawa K. Osteolysis: a generic problem. Orthopedics. 1995;18:791–794.PubMed Bauer TW, Hirakawa K. Osteolysis: a generic problem. Orthopedics. 1995;18:791–794.PubMed
5.
Zurück zum Zitat Cales B. Zirconia as a sliding material: histologic, laboratory, and clinical data. Clin Orthop Relat Res. 2000;379:94–112.PubMedCrossRef Cales B. Zirconia as a sliding material: histologic, laboratory, and clinical data. Clin Orthop Relat Res. 2000;379:94–112.PubMedCrossRef
6.
Zurück zum Zitat Clarke IC, Gustafson A. Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear. Clin Orthop Relat Res. 2000;379:34–40.PubMedCrossRef Clarke IC, Gustafson A. Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear. Clin Orthop Relat Res. 2000;379:34–40.PubMedCrossRef
7.
Zurück zum Zitat Cohn RM, Gonzalez Della Valle A, Peterson M, Cornell CN. Similar wear in total hip arthroplasties with metallic or zirconia femoral heads. HSS J. 2008;4:107–111.PubMedCrossRef Cohn RM, Gonzalez Della Valle A, Peterson M, Cornell CN. Similar wear in total hip arthroplasties with metallic or zirconia femoral heads. HSS J. 2008;4:107–111.PubMedCrossRef
8.
Zurück zum Zitat Collier MB, Kraay MJ, Rimnac CM, Goldberg VM. Evaluation of contemporary software methods used to quantify polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2410–2418.PubMed Collier MB, Kraay MJ, Rimnac CM, Goldberg VM. Evaluation of contemporary software methods used to quantify polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2410–2418.PubMed
9.
Zurück zum Zitat Cooper RA, McAllister CM, Borden LS, Bauer TW. Polyethylene debris-induced osteolysis and loosening in uncemented total hip arthroplasty: a cause of late failure. J Arthroplasty. 1992;7:285–290.PubMedCrossRef Cooper RA, McAllister CM, Borden LS, Bauer TW. Polyethylene debris-induced osteolysis and loosening in uncemented total hip arthroplasty: a cause of late failure. J Arthroplasty. 1992;7:285–290.PubMedCrossRef
10.
Zurück zum Zitat Cuckler JM, Bearcroft J, Asgian CM. Femoral head technologies to reduce polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res. 1995;317:57–63.PubMed Cuckler JM, Bearcroft J, Asgian CM. Femoral head technologies to reduce polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res. 1995;317:57–63.PubMed
11.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
12.
Zurück zum Zitat Derbyshire B, Fisher J, Dowson D, Hardaker C, Brummitt K. Comparative study of the wear of UHMWPE with zirconia ceramic and stainless steel femoral heads in artificial hip joints. Med Eng Phys. 1994;16:229–236.PubMedCrossRef Derbyshire B, Fisher J, Dowson D, Hardaker C, Brummitt K. Comparative study of the wear of UHMWPE with zirconia ceramic and stainless steel femoral heads in artificial hip joints. Med Eng Phys. 1994;16:229–236.PubMedCrossRef
13.
Zurück zum Zitat Devane PA, Bourne RB, Rorabeck CH, Hardie RM, Horne JG. Measurement of polyethylene wear in metal-backed acetabular cups: I. Three-dimensional technique. Clin Orthop Relat Res. 1995;319:303–316.PubMed Devane PA, Bourne RB, Rorabeck CH, Hardie RM, Horne JG. Measurement of polyethylene wear in metal-backed acetabular cups: I. Three-dimensional technique. Clin Orthop Relat Res. 1995;319:303–316.PubMed
14.
Zurück zum Zitat Devane PA, Bourne RB, Rorabeck CH, MacDonald S, Robinson EJ. Measurement of polyethylene wear in metal-backed acetabular cups: II. Clinical application. Clin Orthop Relat Res. 1995;319:317–326.PubMed Devane PA, Bourne RB, Rorabeck CH, MacDonald S, Robinson EJ. Measurement of polyethylene wear in metal-backed acetabular cups: II. Clinical application. Clin Orthop Relat Res. 1995;319:317–326.PubMed
15.
Zurück zum Zitat Dowd JE, Sychterz CJ, Young AM, Engh CA. Characterization of long-term femoral-head-penetration rates: association with and prediction of osteolysis. J Bone Joint Surg Am. 2000;82:1102–1107.PubMed Dowd JE, Sychterz CJ, Young AM, Engh CA. Characterization of long-term femoral-head-penetration rates: association with and prediction of osteolysis. J Bone Joint Surg Am. 2000;82:1102–1107.PubMed
16.
Zurück zum Zitat Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD. Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003;85:2378–2384.PubMed Ebramzadeh E, Sangiorgio SN, Lattuada F, Kang JS, Chiesa R, McKellop HA, Dorr LD. Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements. J Bone Joint Surg Am. 2003;85:2378–2384.PubMed
17.
Zurück zum Zitat Hernigou P, Bahrami T. Zirconia and alumina ceramics in comparison with stainless-steel heads: polyethylene wear after a minimum ten-year follow-up. J Bone Joint Surg Br. 2003;85:504–509.PubMedCrossRef Hernigou P, Bahrami T. Zirconia and alumina ceramics in comparison with stainless-steel heads: polyethylene wear after a minimum ten-year follow-up. J Bone Joint Surg Br. 2003;85:504–509.PubMedCrossRef
18.
Zurück zum Zitat Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed
19.
Zurück zum Zitat Hummer CD 3rd, Rothman RH, Hozack WJ. Catastrophic failure of modular zirconia-ceramic femoral head components after total hip arthroplasty. J Arthroplasty. 1995;10:848–850.PubMedCrossRef Hummer CD 3rd, Rothman RH, Hozack WJ. Catastrophic failure of modular zirconia-ceramic femoral head components after total hip arthroplasty. J Arthroplasty. 1995;10:848–850.PubMedCrossRef
20.
Zurück zum Zitat Kim YH. Comparison of polyethylene wear associated with cobalt-chromium and zirconia heads after total hip replacement: a prospective, randomized study. J Bone Joint Surg Am. 2005;87:1769–1776.PubMedCrossRef Kim YH. Comparison of polyethylene wear associated with cobalt-chromium and zirconia heads after total hip replacement: a prospective, randomized study. J Bone Joint Surg Am. 2005;87:1769–1776.PubMedCrossRef
21.
Zurück zum Zitat Kim YH, Kim JS. Tribological and material analyses of retrieved alumina and zirconia ceramic heads correlated with polyethylene wear after total hip replacement. J Bone Joint Surg Br. 2008;90:731–737.PubMed Kim YH, Kim JS. Tribological and material analyses of retrieved alumina and zirconia ceramic heads correlated with polyethylene wear after total hip replacement. J Bone Joint Surg Br. 2008;90:731–737.PubMed
22.
Zurück zum Zitat Kim YH, Kim JS, Cho SH. A comparison of polyethylene wear in hips with cobalt-chrome or zirconia heads: a prospective, randomised study. J Bone Joint Surg Br. 2001;83:742–750.PubMedCrossRef Kim YH, Kim JS, Cho SH. A comparison of polyethylene wear in hips with cobalt-chrome or zirconia heads: a prospective, randomised study. J Bone Joint Surg Br. 2001;83:742–750.PubMedCrossRef
23.
Zurück zum Zitat Kraay MJ, Thomas RD, Rimnac CM, Fitzgerald SJ, Goldberg VM. Zirconia versus Co-Cr femoral heads in total hip arthroplasty: early assessment of wear. Clin Orthop Relat Res. 2006;453:86–90.PubMedCrossRef Kraay MJ, Thomas RD, Rimnac CM, Fitzgerald SJ, Goldberg VM. Zirconia versus Co-Cr femoral heads in total hip arthroplasty: early assessment of wear. Clin Orthop Relat Res. 2006;453:86–90.PubMedCrossRef
24.
Zurück zum Zitat Kumar P, Oka M, Ikeuchi K, Shimizu K, Yamamuro T, Okumura H, Kotoura Y. Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy. J Biomed Mater Res. 1991;25:813–828.PubMedCrossRef Kumar P, Oka M, Ikeuchi K, Shimizu K, Yamamuro T, Okumura H, Kotoura Y. Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy. J Biomed Mater Res. 1991;25:813–828.PubMedCrossRef
25.
Zurück zum Zitat Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–528.PubMed
26.
Zurück zum Zitat Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am. 1997;79:1635–1641.PubMed
27.
Zurück zum Zitat Oldenburg M, Wegner R, Baur X. Severe cobalt intoxication due to prosthesis wear in repeated total hip arthroplasty. J Arthroplasty. 2008 Oct 1. [Epub ahead of print]. Oldenburg M, Wegner R, Baur X. Severe cobalt intoxication due to prosthesis wear in repeated total hip arthroplasty. J Arthroplasty. 2008 Oct 1. [Epub ahead of print].
28.
Zurück zum Zitat Piconi C, Maccauro G, Angeloni M, Rossi B, Learmonth ID. Zirconia heads in perspective: a survey of zirconia outcomes in total hip replacement. Hip Int. 2007;17:119–130.PubMed Piconi C, Maccauro G, Angeloni M, Rossi B, Learmonth ID. Zirconia heads in perspective: a survey of zirconia outcomes in total hip replacement. Hip Int. 2007;17:119–130.PubMed
29.
Zurück zum Zitat Saikko V. Wear of polyethylene acetabular cups against zirconia femoral heads studied with a hip-joint simulator. Wear. 1994;176:207–212.CrossRef Saikko V. Wear of polyethylene acetabular cups against zirconia femoral heads studied with a hip-joint simulator. Wear. 1994;176:207–212.CrossRef
30.
Zurück zum Zitat Schmalzried TP, Huk OL. Patient factors and wear in total hip arthroplasty. Clin Orthop Relat Res. 2004;418:94–97.PubMedCrossRef Schmalzried TP, Huk OL. Patient factors and wear in total hip arthroplasty. Clin Orthop Relat Res. 2004;418:94–97.PubMedCrossRef
31.
Zurück zum Zitat Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone Joint Surg Am. 1992;74:849–863.PubMed Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone Joint Surg Am. 1992;74:849–863.PubMed
32.
Zurück zum Zitat Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO, dela Rosa M, Fa’vae F, McKellop HA, McClung CD, Martell J, Moreland JR, Amstutz HC. The John Charnley Award. Wear is a function of use, not time. Clin Orthop Relat Res. 2000;381:36–46.PubMedCrossRef Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO, dela Rosa M, Fa’vae F, McKellop HA, McClung CD, Martell J, Moreland JR, Amstutz HC. The John Charnley Award. Wear is a function of use, not time. Clin Orthop Relat Res. 2000;381:36–46.PubMedCrossRef
33.
Zurück zum Zitat Sychterz CJ, Engh CA Jr, Yang A, Engh CA. Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called bedding-in. J Bone Joint Surg Am. 1999;81:821–830.PubMedCrossRef Sychterz CJ, Engh CA Jr, Yang A, Engh CA. Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called bedding-in. J Bone Joint Surg Am. 1999;81:821–830.PubMedCrossRef
34.
Zurück zum Zitat von Schewelov T, Sanzen L, Borlin N, Markusson P, Onsten I. Accuracy of radiographic and radiostereometric wear measurement of different hip prostheses: an experimental study. Acta Orthop Scand. 2004;75:691–700.CrossRef von Schewelov T, Sanzen L, Borlin N, Markusson P, Onsten I. Accuracy of radiographic and radiostereometric wear measurement of different hip prostheses: an experimental study. Acta Orthop Scand. 2004;75:691–700.CrossRef
35.
Zurück zum Zitat von Schewelov T, Sanzen L, Onsten I, Carlsson A, Besjakov J. Total hip replacement with a zirconium oxide ceramic femoral head: a randomised roentgen stereophotogrammetric study. J Bone Joint Surg Br. 2005;87:1631–1635. von Schewelov T, Sanzen L, Onsten I, Carlsson A, Besjakov J. Total hip replacement with a zirconium oxide ceramic femoral head: a randomised roentgen stereophotogrammetric study. J Bone Joint Surg Br. 2005;87:1631–1635.
36.
Zurück zum Zitat Wan Z, Dorr LD. Natural history of femoral focal osteolysis with proximal ingrowth smooth stem implant. J Arthroplasty 1996;11:718–725.PubMedCrossRef Wan Z, Dorr LD. Natural history of femoral focal osteolysis with proximal ingrowth smooth stem implant. J Arthroplasty 1996;11:718–725.PubMedCrossRef
Metadaten
Titel
Clinical Comparison of Polyethylene Wear with Zirconia or Cobalt-Chromium Femoral Heads
verfasst von
Maiken Stilling, MD
Kjeld Anton Nielsen, MD
Kjeld Søballe, MD, DMSc
Ole Rahbek, MD, PhD
Publikationsdatum
01.10.2009
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 10/2009
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-0799-5

Weitere Artikel der Ausgabe 10/2009

Clinical Orthopaedics and Related Research® 10/2009 Zur Ausgabe

Symposium: ABJS Carl T. Brighton Workshop on Health Policy Issues in Orthopaedic Surgery

The Impact of Disruptive Innovations in Orthopaedics

Symposium: ABJS Carl T. Brighton Workshop on Health Policy Issues in Orthopaedic Surgery

Executive Summary: Aligning Stakeholder Incentives in Orthopaedics

Symposium: ABJS Carl T. Brighton Workshop on Health Policy Issues in Orthopaedic Surgery

Executive Summary: Value-based Purchasing and Technology Assessment in Orthopaedics

Symposium: ABJS Carl T. Brighton Workshop on Health Policy Issues in Orthopaedic Surgery

Professionalism in 21st Century Professional Practice: Autonomy and Accountability in Orthopaedic Surgery

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.