Background

Ceramic bearings were introduced to reduce wear and increase long-term survivorship of total hip arthroplasty. In a previous study comparing ceramic with metal-on-polyethylene at 5 to 8 years, we found higher survivorship and no osteolysis for the ceramic bearings.

Questions/Purposes

We asked whether ceramic bearings have equal or superior survivorship compared with that for metal-on-polyethylene at longer followup; we also determined survivorship of the implant systems, the presence or absence of radiographic osteolysis, and incidence of device squeaking.

Methods

Five surgeons at five sites have followed 189 patients (216 hips) for a minimum of 10 years and average of 10.3 years (range, 10–12.4 years) comparing alumina ceramic bearings (144 hips) with cobalt chrome-on-polyethylene bearings (72 hips). We determined Kaplan-Meier survivorship of the bearing surface and implant systems and collected radiographic and clinical data.

Results

We observed no difference between the control metal-on-polyethylene and the alumina-bearing couple cohorts with regard to bearing-related failures (98.9% versus 99.1%). Revisions for any reason occurred in 10.5% of the control patients and 3.1% of the patients with alumina bearings. All femoral implants remain well fixed (100%), whereas one acetabular component (1%) is unstable in the control group. Osteolysis occurred in 26% of the control patients and in none of the patients with alumina bearings. Squeaking occurred in two of 144 hips (1.4%) of the patients with ceramic bearings.

Conclusions

Patients receiving the ceramic-on-ceramic bearings had fewer revisions for any reason and less osteolysis than the control metal-on-polyethylene at 10 years. Our data suggest ceramic bearings continue to provide an option for the young and more active patient and provide for a measure to compare other new alternative bearings that are currently available.

Level of Evidence

Level I, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2005;87:2456–2463.PubMedCrossRef Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2005;87:2456–2463.PubMedCrossRef
2.
Zurück zum Zitat Boehler M, Knahr K, Plenk HJ Jr, Walter A, Salzer M, Schreiber V. Long-term results of uncemented alumina acetabular implants. J Bone Joint Surg Br. 1994;76:53–59.PubMed Boehler M, Knahr K, Plenk HJ Jr, Walter A, Salzer M, Schreiber V. Long-term results of uncemented alumina acetabular implants. J Bone Joint Surg Br. 1994;76:53–59.PubMed
3.
Zurück zum Zitat Capello WN, D’Antonio JA, Feinberg JR, Manley MT, Naughton M. Ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2008;23(Suppl 1):39–43.PubMedCrossRef Capello WN, D’Antonio JA, Feinberg JR, Manley MT, Naughton M. Ceramic-on-ceramic total hip arthroplasty. J Arthroplasty. 2008;23(Suppl 1):39–43.PubMedCrossRef
4.
Zurück zum Zitat Capello WN, D’Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res. 2011;469:825–830.PubMedCrossRef Capello WN, D’Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res. 2011;469:825–830.PubMedCrossRef
5.
Zurück zum Zitat Chevillotte C, Pibarot V, Carret JP, Bejui-Hugues J, Guyen O. Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty. Int Orthop. 2010 Dec 21 [Epub ahead of print]. Chevillotte C, Pibarot V, Carret JP, Bejui-Hugues J, Guyen O. Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty. Int Orthop. 2010 Dec 21 [Epub ahead of print].
6.
Zurück zum Zitat Chevillotte C, Trousdale RT, Chen Q, Guyen O, An KN. The 2009 Frank Stinchfield Award: ‘Hip squeaking’: a biomechanical study of ceramic-on-ceramic bearing surfaces. Clin Orthop Relat Res. 2010;468:345–350.PubMedCrossRef Chevillotte C, Trousdale RT, Chen Q, Guyen O, An KN. The 2009 Frank Stinchfield Award: ‘Hip squeaking’: a biomechanical study of ceramic-on-ceramic bearing surfaces. Clin Orthop Relat Res. 2010;468:345–350.PubMedCrossRef
7.
Zurück zum Zitat Clarke IC. Role of ceramic implants: design and clinical success with total hip prosthetic ceramic-to-ceramic bearings. Clin Orthop Relat Res. 1992;282:19–30.PubMed Clarke IC. Role of ceramic implants: design and clinical success with total hip prosthetic ceramic-to-ceramic bearings. Clin Orthop Relat Res. 1992;282:19–30.PubMed
8.
Zurück zum Zitat D’Antonio JA, Capello WN, Manley MT, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.PubMedCrossRef D’Antonio JA, Capello WN, Manley MT, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171.PubMedCrossRef
9.
Zurück zum Zitat D’Antonio JA, Manley MT, Capello WN, Bierbaum B, Ramakrishnan R, Naughton M, Sutton K. Five-year experience with Crossfire highly crosslinked polyethylene. Clin Orthop Relat Res. 2005;441:143–150.PubMedCrossRef D’Antonio JA, Manley MT, Capello WN, Bierbaum B, Ramakrishnan R, Naughton M, Sutton K. Five-year experience with Crossfire highly crosslinked polyethylene. Clin Orthop Relat Res. 2005;441:143–150.PubMedCrossRef
10.
Zurück zum Zitat DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.PubMed
11.
Zurück zum Zitat Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.PubMedCrossRef Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.PubMedCrossRef
12.
Zurück zum Zitat Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–661.PubMedCrossRef Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–661.PubMedCrossRef
13.
Zurück zum Zitat Engh CA, Massin P, Suthers KE. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res. 1990;257:107–128.PubMed Engh CA, Massin P, Suthers KE. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res. 1990;257:107–128.PubMed
14.
Zurück zum Zitat Girard J, Bocquet D, Autissier G, Fouilleron N, Fron D, Migaud H. Metal-on-metal hip arthroplasty in patients thirty years of age or younger. J Bone Joint Surg Am. 2010;92:2419–2426.PubMedCrossRef Girard J, Bocquet D, Autissier G, Fouilleron N, Fron D, Migaud H. Metal-on-metal hip arthroplasty in patients thirty years of age or younger. J Bone Joint Surg Am. 2010;92:2419–2426.PubMedCrossRef
15.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC. Modes of failure of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. Modes of failure of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
16.
Zurück zum Zitat Hannouche D, Nich C, Bizot P, Meunier A, Nizard R, Sedel L. Fractures of ceramic bearings. Clin Orthop Relat Res. 2003;417:19–26.PubMed Hannouche D, Nich C, Bizot P, Meunier A, Nizard R, Sedel L. Fractures of ceramic bearings. Clin Orthop Relat Res. 2003;417:19–26.PubMed
17.
Zurück zum Zitat Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fracture: treatment by mold arthroplasty: an end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fracture: treatment by mold arthroplasty: an end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755.PubMed
18.
19.
Zurück zum Zitat Hellman EJ, Capello WN, Feinberg JR. Omnifit cementless total hip arthroplasty: a 10 year average follow-up. Clin Orthop Relat Res. 1999;364:164–174.PubMedCrossRef Hellman EJ, Capello WN, Feinberg JR. Omnifit cementless total hip arthroplasty: a 10 year average follow-up. Clin Orthop Relat Res. 1999;364:164–174.PubMedCrossRef
20.
Zurück zum Zitat Heros RJ, Willmann G. Ceramics in total hip arthroplasty: history, mechanical properties, clinical results, and current manufacturing state of the art. Semin Arthroplasty. 1998;9:114–122. Heros RJ, Willmann G. Ceramics in total hip arthroplasty: history, mechanical properties, clinical results, and current manufacturing state of the art. Semin Arthroplasty. 1998;9:114–122.
21.
Zurück zum Zitat Holmer P, Nielsen PT. Fracture of ceramic femoral heads in total hip arthroplasty. J Arthroplasty. 1993;8:567–571.PubMedCrossRef Holmer P, Nielsen PT. Fracture of ceramic femoral heads in total hip arthroplasty. J Arthroplasty. 1993;8:567–571.PubMedCrossRef
22.
Zurück zum Zitat Jarrett CA, Ranawat AS, Bruzzone M, Blum YC, Rodriguez JA, Ranawat CS. The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. J Bone Joint Surg Am. 2009;91:1344–1349.PubMedCrossRef Jarrett CA, Ranawat AS, Bruzzone M, Blum YC, Rodriguez JA, Ranawat CS. The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. J Bone Joint Surg Am. 2009;91:1344–1349.PubMedCrossRef
23.
Zurück zum Zitat Johnston RC, Fitxgerald JR, Harris WH, Muller ME, Sledge CB. Clinical and radiographic evaluation of cementless THA. J Bone Joint Surg Am. 1990;72:162–168. Johnston RC, Fitxgerald JR, Harris WH, Muller ME, Sledge CB. Clinical and radiographic evaluation of cementless THA. J Bone Joint Surg Am. 1990;72:162–168.
24.
Zurück zum Zitat Kabo JM, Gebhard JS, Loren G, Amstutz HC. In vivo wear of polyethylene acetabular components. J Bone Joint Surg Br. 1993;75:254–258.PubMed Kabo JM, Gebhard JS, Loren G, Amstutz HC. In vivo wear of polyethylene acetabular components. J Bone Joint Surg Br. 1993;75:254–258.PubMed
25.
Zurück zum Zitat Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Stat Assoc. 1958;53:457–481.CrossRef Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Stat Assoc. 1958;53:457–481.CrossRef
26.
Zurück zum Zitat Kress AM, Schmidt R, Holzwarth U, Forst R, Mueller LA. Excellent results with cementless total hip arthroplasty and alumina-on-alumina pairing: minimum ten-year follow-up. Int Orthop. 2011;35:195–200.PubMedCrossRef Kress AM, Schmidt R, Holzwarth U, Forst R, Mueller LA. Excellent results with cementless total hip arthroplasty and alumina-on-alumina pairing: minimum ten-year follow-up. Int Orthop. 2011;35:195–200.PubMedCrossRef
27.
Zurück zum Zitat Lee YK, Ha YC, Yoo JJ, Koo KH, Toon KS, Kim HJ. Alumina-on-alumina total hip arthroplasty at minimum 10 year follow-up. J Bone Joint Surg Am. 2010;92:1715–1718.PubMedCrossRef Lee YK, Ha YC, Yoo JJ, Koo KH, Toon KS, Kim HJ. Alumina-on-alumina total hip arthroplasty at minimum 10 year follow-up. J Bone Joint Surg Am. 2010;92:1715–1718.PubMedCrossRef
28.
Zurück zum Zitat Lewis PM, Al-Belooshi A, Olsen M, Schemitch EH, Waddell JP. Prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene bearings in total hip arthroplasty. J Arthroplasty. 2010;25:392–397.PubMedCrossRef Lewis PM, Al-Belooshi A, Olsen M, Schemitch EH, Waddell JP. Prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene bearings in total hip arthroplasty. J Arthroplasty. 2010;25:392–397.PubMedCrossRef
29.
Zurück zum Zitat Lusty PK, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676.PubMedCrossRef Lusty PK, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676.PubMedCrossRef
30.
Zurück zum Zitat Mahoney OM, Dimon JH. Unsatisfactory results with a ceramic total hip prosthesis. J Bone Joint Surg Am. 1990;72:663–671.PubMed Mahoney OM, Dimon JH. Unsatisfactory results with a ceramic total hip prosthesis. J Bone Joint Surg Am. 1990;72:663–671.PubMed
31.
Zurück zum Zitat McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2010;91:773–782.CrossRef McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2010;91:773–782.CrossRef
32.
Zurück zum Zitat Murphy SB, Ecker TM, Tannast M. Two- to 9-year clinical results of alumina ceramic-on-ceramic THA. Clin Orthop Relat Res. 2006;453:97–102.PubMedCrossRef Murphy SB, Ecker TM, Tannast M. Two- to 9-year clinical results of alumina ceramic-on-ceramic THA. Clin Orthop Relat Res. 2006;453:97–102.PubMedCrossRef
33.
Zurück zum Zitat Park YS, Park SJ, Lim SJ. Ten-year results after cementless THA with sandwich-type alumina ceramic bearing. Orthopedics. 2010;33:796.PubMed Park YS, Park SJ, Lim SJ. Ten-year results after cementless THA with sandwich-type alumina ceramic bearing. Orthopedics. 2010;33:796.PubMed
34.
Zurück zum Zitat Restrepo C, Matar WY, Parvizi J, Rothman RH, Hozack WJ. Natural history of squeaking after total hip arthroplasty. Clin Orthop Relat Res. 2010;468:2340–2345.PubMedCrossRef Restrepo C, Matar WY, Parvizi J, Rothman RH, Hozack WJ. Natural history of squeaking after total hip arthroplasty. Clin Orthop Relat Res. 2010;468:2340–2345.PubMedCrossRef
35.
Zurück zum Zitat Restrepo C, Post ZD, Kai B, Hozack WJ. The effect of stem design on the prevalence of squeaking following ceramic-on-ceramic bearing total hip arthroplasty. J Bone Joint Surg Am. 2010;92:550–557.PubMedCrossRef Restrepo C, Post ZD, Kai B, Hozack WJ. The effect of stem design on the prevalence of squeaking following ceramic-on-ceramic bearing total hip arthroplasty. J Bone Joint Surg Am. 2010;92:550–557.PubMedCrossRef
36.
Zurück zum Zitat Schmalzried TP, Huk OL. Patient factors and wear in total hip arthroplasty. Clin Orthop Relat Res. 2004;418:94–97.PubMedCrossRef Schmalzried TP, Huk OL. Patient factors and wear in total hip arthroplasty. Clin Orthop Relat Res. 2004;418:94–97.PubMedCrossRef
37.
38.
Zurück zum Zitat Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.PubMedCrossRef Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503.PubMedCrossRef
Metadaten
Titel
Ceramic Bearings for Total Hip Arthroplasty Have High Survivorship at 10 Years
verfasst von
James A. D’Antonio, MD
William N. Capello, MD
Marybeth Naughton, BS
Publikationsdatum
01.02.2012
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 2/2012
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-011-2076-7

Weitere Artikel der Ausgabe 2/2012

Clinical Orthopaedics and Related Research® 2/2012 Zur Ausgabe

Symposium: Papers Presented at the Annual Meetings of The Hip Society

Bony Impingement Limits Design-related Increases in Hip Range of Motion

Symposium: Papers Presented at the Annual Meetings of The Hip Society

A Short Tapered Stem Reduces Intraoperative Complications in Primary Total Hip Arthroplasty

Symposium: Papers Presented at the Annual Meetings of The Hip Society

Incidence of Contralateral THA After Index THA for Osteoarthritis

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.