Are Quadrilateral Surface Buttress Plates Comparable to Traditional Forms of Transverse Acetabular Fracture Fixation?

verfasst von: Brian J. Kistler, MD, Ian R. Smithson, MD, Seth A. Cooper, MD, Jacob L. Cox, MD, Aniruddh N. Nayak, MS, Brandon G. Santoni, PhD, H. Claude Sagi, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 11/2014

Einloggen, um Zugang zu erhalten

Abstract

Background

Several construct options exist for transverse acetabular fracture fixation. Accepted techniques use a combination of column plates and lag screws. Quadrilateral surface buttress plates have been introduced as potential fixation options, but as a result of their novelty, biomechanical data regarding their stabilizing effects are nonexistent. Therefore, we aimed to determine if this fixation method confers similar stability to traditional forms of fixation.

Questions/purposes

We biomechanically compared two acetabular fixation plates with quadrilateral surface buttressing with traditional forms of fixation using lag screws and column plates.

Methods

Thirty-five synthetic hemipelves with a transverse transtectal acetabular fracture were allocated to one of five groups: anterior column plate + posterior column lag screw, posterior column plate + anterior column lag screw, anterior and posterior column lag screws only, infrapectineal plate + anterior column plate, and suprapectineal plate alone. Specimens were loaded for 1500 cycles up to 2.5x body weight and stiffness was calculated. Thereafter, constructs were destructively loaded and failure loads were recorded.

Results

After 1500 cycles, final stiffness was not different with the numbers available between the infrapectineal (568 ± 43 N/mm) and suprapectineal groups (602 ± 87 N/mm, p = 0.988). Both quadrilateral plates were significantly stiffer than the posterior column buttress plate with supplemental lag screw fixation group (311 ± 99 N/mm, p < 0.006). No difference in stiffness was identified with the numbers available between the quadrilateral surface plating groups and the lag screw group (423 ± 219 N/mm, p > 0.223). The infrapectineal group failed at the highest loads (5.4 ± 0.6 kN) and this was significant relative to the suprapectineal (4.4 ± 0.3 kN; p = 0.023), lag screw (2.9 ± 0.8 kN; p < 0.001), and anterior buttress plate with posterior column lag screw (4.0 ± 0.6 kN; p = 0.001) groups.

Conclusions

Quadrilateral surface buttress plates spanning the posterior and anterior columns are biomechanically comparable and, in some cases, superior to traditional forms of fixation in this synthetic hemipelvis model.

Clinical Relevance

Quadrilateral surface buttress plates may present a viable alternative for the treatment of transtectal transverse acetabular fractures. Clinical studies are required to fully define the use of this new form of fixation for such fractures when accessed through the anterior intrapelvic approach.
Literatur
1.
Zurück zum Zitat Chang JK, Gill SS, Zura RD, Krause WR, Wang GJ. Comparative strength of three methods of fixation of transverse acetabular fractures. Clin Orthop Relat Res. 2001;392:433–441.PubMedCrossRef Chang JK, Gill SS, Zura RD, Krause WR, Wang GJ. Comparative strength of three methods of fixation of transverse acetabular fractures. Clin Orthop Relat Res. 2001;392:433–441.PubMedCrossRef
2.
Zurück zum Zitat Cole JD, Bolhofner BR. Acetabular fracture fixation via a modified Stoppa limited intrapelvic approach. Description of operative technique and preliminary treatment results. Clin Orthop Relat Res. 1994;305:112–123.PubMedCrossRef Cole JD, Bolhofner BR. Acetabular fracture fixation via a modified Stoppa limited intrapelvic approach. Description of operative technique and preliminary treatment results. Clin Orthop Relat Res. 1994;305:112–123.PubMedCrossRef
3.
Zurück zum Zitat Cristofolini L, Viceconti M, Cappello A, Toni A. Mechanical validation of whole bone composite femur models. J Biomech. 1996;29:525–535.PubMedCrossRef Cristofolini L, Viceconti M, Cappello A, Toni A. Mechanical validation of whole bone composite femur models. J Biomech. 1996;29:525–535.PubMedCrossRef
4.
Zurück zum Zitat Gardner MP, Chong AC, Pollock AG, Wooley PH. Mechanical evaluation of large-size fourth-generation composite femur and tibia models. Ann Biomed Eng. 2010;38:613–620.PubMedCrossRef Gardner MP, Chong AC, Pollock AG, Wooley PH. Mechanical evaluation of large-size fourth-generation composite femur and tibia models. Ann Biomed Eng. 2010;38:613–620.PubMedCrossRef
5.
Zurück zum Zitat Giordano V, do Amaral NP, Pallottino A, Pires e Albuquerque R, Franklin CE, Labronici PJ. Operative treatment of transverse acetabular fractures: is it really necessary to fix both columns? Int J Med Sci. 2009;6:192–199.PubMedCrossRefPubMedCentral Giordano V, do Amaral NP, Pallottino A, Pires e Albuquerque R, Franklin CE, Labronici PJ. Operative treatment of transverse acetabular fractures: is it really necessary to fix both columns? Int J Med Sci. 2009;6:192–199.PubMedCrossRefPubMedCentral
6.
Zurück zum Zitat Grover P, Albert C, Wang M, Harris GF. Mechanical characterization of fourth generation composite humerus. Proc Inst Mech Eng H. 2011;225:1169–1176.PubMedCrossRef Grover P, Albert C, Wang M, Harris GF. Mechanical characterization of fourth generation composite humerus. Proc Inst Mech Eng H. 2011;225:1169–1176.PubMedCrossRef
7.
Zurück zum Zitat Heiner AD. Structural properties of fourth-generation composite femurs and tibias. J Biomech. 2008;41:3282–3284.PubMedCrossRef Heiner AD. Structural properties of fourth-generation composite femurs and tibias. J Biomech. 2008;41:3282–3284.PubMedCrossRef
8.
Zurück zum Zitat Hirvensalo E, Lindahl J, Bostman O. A new approach to the internal fixation of unstable pelvic fractures. Clin Orthop Relat Res. 1993;297:28–32.PubMed Hirvensalo E, Lindahl J, Bostman O. A new approach to the internal fixation of unstable pelvic fractures. Clin Orthop Relat Res. 1993;297:28–32.PubMed
9.
Zurück zum Zitat Khajavi K, Lee AT, Lindsey DP, Leucht P, Bellino MJ, Giori NJ. Single column locking plate fixation is inadequate in two column acetabular fractures. A biomechanical analysis. J Orthop Surg Res. 2010;5:30. Khajavi K, Lee AT, Lindsey DP, Leucht P, Bellino MJ, Giori NJ. Single column locking plate fixation is inadequate in two column acetabular fractures. A biomechanical analysis. J Orthop Surg Res. 2010;5:30.
10.
Zurück zum Zitat Letournel E, Judet R. Fractures of the Acetabulum. 2nd ed. Berlin, Germany: Springer Verlag; 1993.CrossRef Letournel E, Judet R. Fractures of the Acetabulum. 2nd ed. Berlin, Germany: Springer Verlag; 1993.CrossRef
11.
Zurück zum Zitat Olson SA, Kadrmas MW, Hernandez JD, Glisson RR, West JL. Augmentation of posterior wall acetabular fracture fixation using calcium-phosphate cement: a biomechanical analysis. J Orthop Trauma. 2007;21:608–616.PubMedCrossRef Olson SA, Kadrmas MW, Hernandez JD, Glisson RR, West JL. Augmentation of posterior wall acetabular fracture fixation using calcium-phosphate cement: a biomechanical analysis. J Orthop Trauma. 2007;21:608–616.PubMedCrossRef
12.
Zurück zum Zitat Qureshi AA, Archdeacon MT, Jenkins MA, Infante A, DiPasquale T, Bolhofner BR. Infrapectineal plating for acetabular fractures: a technical adjunct to internal fixation. J Orthop Trauma. 2004;18:175–178.PubMedCrossRef Qureshi AA, Archdeacon MT, Jenkins MA, Infante A, DiPasquale T, Bolhofner BR. Infrapectineal plating for acetabular fractures: a technical adjunct to internal fixation. J Orthop Trauma. 2004;18:175–178.PubMedCrossRef
13.
Zurück zum Zitat Sagi HC, Afsari A, Dziadosz D. The anterior intra-pelvic (modified Rives-Stoppa) approach for fixation of acetabular fractures. J Orthop Trauma. 2010;24:263–270.PubMedCrossRef Sagi HC, Afsari A, Dziadosz D. The anterior intra-pelvic (modified Rives-Stoppa) approach for fixation of acetabular fractures. J Orthop Trauma. 2010;24:263–270.PubMedCrossRef
14.
Zurück zum Zitat Sawaguchi T, Brown TD, Rubash HE, Mears DC. Stability of acetabular fractures after internal fixation. A cadaveric study. Acta Orthop Scand. 1984;55:601–605.PubMedCrossRef Sawaguchi T, Brown TD, Rubash HE, Mears DC. Stability of acetabular fractures after internal fixation. A cadaveric study. Acta Orthop Scand. 1984;55:601–605.PubMedCrossRef
15.
Zurück zum Zitat Shazar N, Brumback RJ, Novak VP, Belkoff SM. Biomechanical evaluation of transverse acetabular fracture fixation. Clin Orthop Relat Res. 1998;352:215–222.PubMedCrossRef Shazar N, Brumback RJ, Novak VP, Belkoff SM. Biomechanical evaluation of transverse acetabular fracture fixation. Clin Orthop Relat Res. 1998;352:215–222.PubMedCrossRef
16.
Zurück zum Zitat Zdero R, Elfallah K, Olsen M, Schemitsch EH. Cortical screw purchase in synthetic and human femurs. J Biomech Eng. 2009;131:094503.PubMedCrossRef Zdero R, Elfallah K, Olsen M, Schemitsch EH. Cortical screw purchase in synthetic and human femurs. J Biomech Eng. 2009;131:094503.PubMedCrossRef
17.
Zurück zum Zitat Zdero R, Olsen M, Bougherara H, Schemitsch EH. Cancellous bone screw purchase: a comparison of synthetic femurs, human femurs, and finite element analysis. Proc Inst Mech Eng H. 2008;222:1175–1183.PubMedCrossRef Zdero R, Olsen M, Bougherara H, Schemitsch EH. Cancellous bone screw purchase: a comparison of synthetic femurs, human femurs, and finite element analysis. Proc Inst Mech Eng H. 2008;222:1175–1183.PubMedCrossRef
Metadaten
Titel
Are Quadrilateral Surface Buttress Plates Comparable to Traditional Forms of Transverse Acetabular Fracture Fixation?
verfasst von
Brian J. Kistler, MD
Ian R. Smithson, MD
Seth A. Cooper, MD
Jacob L. Cox, MD
Aniruddh N. Nayak, MS
Brandon G. Santoni, PhD
H. Claude Sagi, MD
Publikationsdatum
01.11.2014
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 11/2014
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-014-3800-x

Weitere Artikel der Ausgabe 11/2014

Clinical Orthopaedics and Related Research® 11/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.