Point-like radioactive source with multiple absorber capsules for evaluating PET scanners

verfasst von: Tomoyuki Hasegawa, Eiji Yoshida, Yasushi Sato, Keiichi Oda, Yasuhiro Wada, Takahiro Yamada, Taiga Yamaya, Hideo Murayama, Kyoko Saito

Erschienen in: Annals of Nuclear Medicine | Ausgabe 5/2010

Einloggen, um Zugang zu erhalten

Abstract

Objective

Radioactive sources for evaluating sensitivity and uncertainty in the radioactivity measurements performed using PET scanners must be equipped with absorber materials that ensure the annihilation of positrons. Attenuation and scattering owing to the absorber materials produce uncertainty in the performance evaluation. The aim of this study is to propose a point-like radioactive source with multiple absorber capsules, for which evaluation can be independent of scatter and attenuation owing to the source absorbers.

Methods

The point-like source consists of a small spherical radioactive part and a set of successively sized cylindrical aluminum absorber capsules. Data were collected for different total absorber thicknesses. By an extrapolation technique, the effects of the source absorbers were eliminated. Sensitivity and uncertainty in the radioactivity measurements of PET scanners were evaluated with this technique.

Results

Sensitivity and uncertainty of radioactivity measurement to the point-like radioactive source were evaluated successfully with this method.

Conclusion

The proposed point-like radioactive source is useful for evaluating performance characteristics of PET scanners in a way that is independent of the effects of the source absorbers.
Literatur
1.
Zurück zum Zitat National Electrical Manufacturers Association (NEMA). Performance measurements of positron emission tomographs. NEMA Standards Publication 1994; NU2-1994:1–28. National Electrical Manufacturers Association (NEMA). Performance measurements of positron emission tomographs. NEMA Standards Publication 1994; NU2-1994:1–28.
2.
Zurück zum Zitat Industries Association of Radiation Apparatus (JESRA). A method of PET performance evaluation (in Japanese). JESRA 1993; X-73. Industries Association of Radiation Apparatus (JESRA). A method of PET performance evaluation (in Japanese). JESRA 1993; X-73.
3.
Zurück zum Zitat Japan Radioisotope Association (JRIA). A guideline of PET performance evaluation (in Japanese). Radioisotopes 1994;43:115–35. Japan Radioisotope Association (JRIA). A guideline of PET performance evaluation (in Japanese). Radioisotopes 1994;43:115–35.
4.
Zurück zum Zitat International Electrotechnical Commission (IEC). Radionuclide imaging devices—characteristics and test conditions. Part 1. Positron Emission Tomographs. IEC Standard 1998;61675–1. International Electrotechnical Commission (IEC). Radionuclide imaging devices—characteristics and test conditions. Part 1. Positron Emission Tomographs. IEC Standard 1998;61675–1.
5.
Zurück zum Zitat National Electrical Manufacturers Association (NEMA). Performance measurements of positron emission tomographs. NEMA Standards Publication 2001; NU2-2001:1–40. National Electrical Manufacturers Association (NEMA). Performance measurements of positron emission tomographs. NEMA Standards Publication 2001; NU2-2001:1–40.
6.
Zurück zum Zitat Bailey DL, Jones T, Spinks TJ. A method for measuring the absolute sensitivity of positron emission tomographic scanners. Eur J Nucl Med. 1991;18:374–9.CrossRefPubMed Bailey DL, Jones T, Spinks TJ. A method for measuring the absolute sensitivity of positron emission tomographic scanners. Eur J Nucl Med. 1991;18:374–9.CrossRefPubMed
7.
Zurück zum Zitat Bailey DL, Jones T. A method for calibrating three-dimensional positron emission tomography without scatter correction. Eur J Nucl Med. 1997;24:660–4.PubMed Bailey DL, Jones T. A method for calibrating three-dimensional positron emission tomography without scatter correction. Eur J Nucl Med. 1997;24:660–4.PubMed
8.
Zurück zum Zitat Bailey DL, Snowdon G, Cooper RG, Roach PJ. The use of molecular sieves to produce point sources of radioactivity. Phys Med Biol. 2004;49:N21–9.CrossRefPubMed Bailey DL, Snowdon G, Cooper RG, Roach PJ. The use of molecular sieves to produce point sources of radioactivity. Phys Med Biol. 2004;49:N21–9.CrossRefPubMed
9.
Zurück zum Zitat Hirayama H, Namito Y, Bielajew AF, Wilderman SJ, Nelson WR. The EGS5 code system, SLAC-R-730 and KEK Report 2005–8, 2006. Hirayama H, Namito Y, Bielajew AF, Wilderman SJ, Nelson WR. The EGS5 code system, SLAC-R-730 and KEK Report 2005–8, 2006.
10.
Zurück zum Zitat Toramatsu C, Suzuki T, Fukumura T, Suzuki K. Evaluation of dose calibrator as a working standard (in Japanese, with an abstract in English). Ioniz Radiat. 2007;33:19–28. Toramatsu C, Suzuki T, Fukumura T, Suzuki K. Evaluation of dose calibrator as a working standard (in Japanese, with an abstract in English). Ioniz Radiat. 2007;33:19–28.
11.
Zurück zum Zitat Yoshida E, Yamaya T, Watanabe M, Kitamura K, Kobayashi A, Hasegawa T, et al. Design and initial evaluation of a 4-layer DOI-PET system: the jPET-D4. Jpn J Med Phys. 2006;26:131–40. Yoshida E, Yamaya T, Watanabe M, Kitamura K, Kobayashi A, Hasegawa T, et al. Design and initial evaluation of a 4-layer DOI-PET system: the jPET-D4. Jpn J Med Phys. 2006;26:131–40.
12.
Zurück zum Zitat Hasegawa T, Yoshida E, Kobayashi A, Shibuya K, Nishikido F, Kobayashi T, et al. Evaluation of static physics performance of jPET-D4 with Monte Carlo simulation. Phys Med Biol. 2007;52:213–30.CrossRefPubMed Hasegawa T, Yoshida E, Kobayashi A, Shibuya K, Nishikido F, Kobayashi T, et al. Evaluation of static physics performance of jPET-D4 with Monte Carlo simulation. Phys Med Biol. 2007;52:213–30.CrossRefPubMed
13.
Zurück zum Zitat National Electrical Manufacturers Association (NEMA). Performance measurements of small animal positron emission tomographs. NEMA Standards Publication 2008; NU4-2008:1–23. National Electrical Manufacturers Association (NEMA). Performance measurements of small animal positron emission tomographs. NEMA Standards Publication 2008; NU4-2008:1–23.
14.
Zurück zum Zitat Sato Y, Murayama H, Yamada Y, Hasegawa T, Oda K, Unno, et al. Monte Carlo simulation of standardization of 22Na using multi channel scintillation detectors. In: 16th International conference on radionuclide metrology and its applications (Cape Town, South Africa) 2009. Sato Y, Murayama H, Yamada Y, Hasegawa T, Oda K, Unno, et al. Monte Carlo simulation of standardization of 22Na using multi channel scintillation detectors. In: 16th International conference on radionuclide metrology and its applications (Cape Town, South Africa) 2009.
15.
Zurück zum Zitat Geant4 Collaboration. Geant4—a simulation toolkit. Nucl Instrum Method 2003;A506:250–303. Geant4 Collaboration. Geant4—a simulation toolkit. Nucl Instrum Method 2003;A506:250–303.
Metadaten
Titel
Point-like radioactive source with multiple absorber capsules for evaluating PET scanners
verfasst von
Tomoyuki Hasegawa
Eiji Yoshida
Yasushi Sato
Keiichi Oda
Yasuhiro Wada
Takahiro Yamada
Taiga Yamaya
Hideo Murayama
Kyoko Saito
Publikationsdatum
01.06.2010
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 5/2010
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-010-0350-5

Weitere Artikel der Ausgabe 5/2010

Annals of Nuclear Medicine 5/2010 Zur Ausgabe