Minimally invasive dynamic screw stabilization using cortical bone trajectory

verfasst von: Chih-Chang Chang, Chao-Hung Kuo, Hsuan-Kan Chang, Tsung-Hsi Tu, Li-Yu Fay, Jau-Ching Wu, Henrich Cheng, Wen-Cheng Huang

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2020

Abstract

Background

The conventional pedicle-screw-based dynamic stabilization process involves dissection of the Wiltse plane to cannulate the pedicles, which cannot be undertaken with minimal surgical invasion. Despite some reports having demonstrated satisfactory outcomes of dynamic stabilization in the management of low-grade spondylolisthesis, the extensive soft tissue dissection involved during pedicle screw insertion substantially compromises the designed rationale of motion (muscular) preservation. The authors report on a novel method for minimally invasive insertion of dynamic screws and a mini case series.

Methods

The authors describe innovations for inserting dynamic screws via the cortical bone trajectory (CBT) under spinal navigation. All the detailed surgical procedures and clinical data are demonstrated.

Results

A total of four (2 females) patients (mean age 64.75 years) with spinal stenosis at L4–5 were included. By a combination of microscopic decompression and image-guided CBT screw insertion, laminectomy and dynamic screw stabilization were achieved via one small skin incision (less than 3 cm). These patients’ back and leg pain improved significantly after the surgery.

Conclusion

This innovative dynamic screw stabilization via the CBT involved no discectomy (or removal of sequestrated fragment only), no interbody fusion, and little muscle dissection (not even of the Wiltse plane). As a minimally invasive surgery, CBT appeared to be a viable alternative to the conventional pedicle-screw-based dynamic stabilization approach.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ALIF
Anterior lumbar interbody fusion
ASD
Adjacent segment disease
CBT
Cortical bone trajectory
PLIF
Posterior lumbar interbody fusion
TLIF
Transforaminal lumbar interbody fusion
LLIF
Lateral lumbar interbody fusion

Background

Various fusion techniques have been accepted as the surgical management for disc degenerative disease or spondylolisthesis, including anterior, posterior, transforaminal, and extreme lateral lumbar interbody fusion (ALIF, PLIF, TLIF and LLIF) [14]. In the past decade there has been an emerging option of dynamic stabilization to treat low-grade degenerative disease. The design concept of dynamic stabilization is to ameliorate the instability while maintaining segmental motility, thus yielding the potential for the prevention of adjacent segment disease (ASD). Although the actual benefit in the reduction of ASD remains elusive, more and more reports have demonstrated that dynamic stabilization is a viable option for degenerative disease or spondylolisthesis [57].
Pedicle-screw based dynamic stabilization systems use elastic materials, for example, a synthetic cord-and-spacer or metallic spring, instead of a rigid metallic rod, to connect between individual pedicle screws [8]. Placement of these pedicle screws requires either a midline open approach or dissection of the Wiltse plane bilaterally. Therefore, these dynamic pedicle screw stabilization systems inevitably involve extensive muscular dissection and therefore the procedure is not minimally invasive.
The authors report in this technical note an innovative method to insert four dynamic screws via a 3-cm skin incision to neutralize a degenerative spine or to stabilize low-grade spondylolisthesis. Under image guidance, the screws were placed along the cortical bone trajectory (CBT) and assembled with dynamic cords. The surgery involved microscopic decompression of the bilateral neuroforamen and dynamic screw insertion, but no discectomy (removal of ruptured part only) or interbody fusion procedures. Compared to the standard minimally invasive TLIF, the novel technique preserved the intervertebral disc and required less muscle dissection. Although the theoretical benefit of reducing ASD requires longer-term follow up to confirm, this technique spared arthrodesis and was done with minimal soft tissue or facet destruction. This is the first report to describe dynamic screw placement via the CBT.

Methods

Operative techniques

Four consecutive patients (2 females), with an average age of 64.75 years, who underwent minimally invasive dynamic screw stabilization with the Dynesys Dynamic Stabilization (Zimmer Biomed, Warshaw, IN, USA) system were retrospectively analyzed.
After general anesthesia, the patient who had a L4–5 herniated disc with bilateral foraminostenosis (Fig. 1) was put in a prone position on a radiolucent table. The index level was identified by fluoroscopy. A midline incision, approximately 3-cm long, was made and deepened for sub-periosteum muscular dissection. Using a self-retaining blade retractor, the L4 spinous process and lamina were exposed without violation of the nearby facet joints, either L3–4 or L4–5 facets bilaterally. Under microscopes, the L4 laminectomy was carried out for resection of the hypertrophic ligamentum flavum and decompression of lateral recesses. The neuroforamen were enlarged with curved-tip Kerrison’s rongeurs and probed through with microsurgical instruments for confirmation of adequate decompression. Typically, the surgery required no discectomy. However, removal of sequestrated disc fragments was performed on those patient who had ruptured intervertebral discs.
After decompression, the O-arm (Medtronic Spinal and Biologics, Memphis, TN) was brought into the room to acquire images for navigated screw placement. The reference pin was inserted into the left posterior superior iliac crest. The acquired images were transferred to the Stealth Station (Medtronic Spinal and Biologics, Memphis, TN) for spinal navigation. Prior to the beginning of instrumentation, we used an image-guided ball tip probe to confirm the accuracy of navigation and locate the entry points. The entry points of CBT screws are generally over the cephalad lateral part of the pars interarticularis, slightly caudal to the sulcus of the facet complex. An image-guided high-speed drill was used to break through the cortex. The trajectory was chosen, using the navigation system, to allow the screw to course through the dense cortical bone with the screw tip barely penetrating the cortex of the vertebral body laterally. A 5.2 mm image-guided cannulated tap was used to create the screw tract. The Dynesys screws (5.2x35mm, top loading with hydroxyapatite coating, Zimmer Biomed, Warshaw, IN) were subsequently placed under navigation. A total of four CBT screws were inserted into the L4 and L5 bodies with bi-cortical purchase (Fig. 2). Compared to traditional screw-and-rod fixation, the dynamic fixation system utilized elastic spacers (polycarbonate-urethan) and cords (polyethylene-terephthalate) to replace the rigid titanium rod in conventional constructs of instrumentation. Thus, after assembling the elastic cords and spacers and tightening of the nuts, the dynamic fixation construct remained elastic and could allow segmental motility.

Results

Case presentation (Table 1)

Case 1

A 68-year-old female presented with chronic low back pain and left sciatica for more than 2 years. The pain would be exacerbated after prolonged walking, thus limiting her activity. Her radiological evaluations demonstrated Meyerding grade one degenerative spondylolisthesis at L4–5. Prior to the visit at our institution, she had tried conservative treatment for more than 6 months. However, the pain persisted and greatly influenced her daily activity. Since she failed conservative management and her symptoms could be highly associated with the L4–5 spondylolisthesis that demonstrated instability on dynamic lateral radiographs, surgical decompression and dynamic stabilization were recommended. After surgery, her VAS of back pain was greatly improved from 5 to 3 (out of 10) and she was discharged from the hospital on post-operation day 5.
Table 1
Case presentation
 
Age
Gender
Symptoms
Radiological diagnosis
Surgical time (mins)
Blood loss
(ml)
Hospital stay (days)
Follow-up (months)
Post-operative narcotics a
Pre-operative ROM
last follow up
ROM b
VAS improvement
Case 1
68
Female
Back pain
Left sciatica
L4/5 grade I spondylolisthesis
225
250
5
28
Tramadol
75 mg/day
6.1
3.1
Back: 5- > 3
Leg: 3- > 2
Case 2
58
Male
Back pain
Left sciatica
L4/5 ruptured disc
225
30
3
14
Tramadol
150 mg/day
15.4
3.1
Back:3- > 0
Leg: 10- > 0
Case 3
62
Female
Back pain
L4/5 grade I spondylolisthesis
265
100
9
15
Tramadol 37.5 mg/day
9.6
1.2
Back:9- > 2
Leg:0- > 0
Case 4
71
Male
Left Leg radiation numbness
L4/5 ruptured disc
260
50
3
31
Tramadol
150 mg/day
10.3
2.4
Back:2- > 0
Leg:1- > 0
VAS visual analog scale
a All patients were weaned off narcotics within 3 months after operation
b After dynamic stabilization, the range of motion was significantly reduced

Case 2

A 58-year-old male presented with low back pain and left sciatica for 6 months. The symptoms were on and off several times and it got worse prior to a visit to our clinic.
His radiographic evaluations demonstrated a huge L4–5 ruptured disc with caudal migration and left L4–5 foramen compression. Since his symptoms were compatible with nerve root compression by the L4–5 ruptured disc, surgical decompression, removal of sequestrated disc fragments and dynamic stabilization were recommended. The post-operative fluoroscopy demonstrated that 3.1 degree motility was preserved at post-op 12 months (Fig. 3). Post-operation, his sciatica improved immediately. He was discharged on post-operation day 3.

Case 3

A 62-year-old female had a history of a traffic accident about 6 years prior to her admission and had experienced on and off low back pain since then. A second falling-down accident happened 5 years later and she experienced left sciatica, which was refractory to rehabilitation for more than 6 months. The radiographic evaluations demonstrated a herniated intervertebral disc over L4–5, grade I spondylolisthesis with severe stenosis and bilateral foraminostenosis. The patient underwent L4 laminectomy, partial removal of the ruptured disc and L4–5 dynamic stabilization. In the post-operative first few days, she still complained about back pain while it has been relieved substantially. She stayed in the hospital for another 7 days for rehabilitation.

Case 4

A 71-year-old male patient experienced progressive left leg radiation numbness for 1 year and his symptom was exacerbated during the 2 months prior to admission. He also had the accompanying symptom of neurologic claudication. The radiographic image revealed a L4–5 ruptured disc with caudal migration. Since his symptom was quite compatible with the L4–5 ruptured disc, the patient underwent L4 laminectomy, removal of the ruptured disc and L4–5 dynamic stabilization. Post-operatively, his radiation numbness improved gradually within the following 6 months.

Discussion

Several clinical series have demonstrated satisfactory clinical outcomes of Dynesys dynamic stabilization (DDS) for lumbar degenerative disease [6, 9, 10]. It is still uncertain whether the claimed effects of DDS, preservation of segmental mobility and thus reduction of ASD, can be achieved and maintained in the long term [11, 12]. There also have been concerns about screw loosening, due to a lack of bone grafts that would likely fuse eventually in such dynamic constructs [7, 13, 14]. In this report the authors have described, for the first time, the innovative technique of inserting dynamic screws via a biomechanically stronger and denser corridor, the CBT. The CBT insertion provided at least two advantages for dynamic screw stabilization, solid bone-screw interface and less muscle dissection, via a mid-line to lateral insertion angle that could be done with minimal invasive surgery [1518].
The CBT used for screw insertion was first published by Santoni et al. in 2009 for posterior midline lumbar fusion surgery. Compared with the traditional pedicle-screw trajectory used, several biomechanical studies have demonstrated the superiority of CBT screws, which could yield a 30% increase in pullout strength and a 71% increase in insertional torque [17, 18]. Moreover, using the CBT to place dynamic screws allowed avoidance of wide exposure of the facet joint and minimized muscle detachment at both the index and the cephalad facet joints. This minimally invasive approach itself might help reduce future ASD. Therefore, the CBT might be a reasonable alternative for placing dynamic screws through minimally invasive surgery. Dynamic screw stabilization via the CBT requires no discectomy or violation of any facet joints, and spares the need for muscle dissection along the Wiltse plane. Therefore, the CBT screw-based DDS is a minimally invasive option for low grade degenerative spondylolisthesis. Studies with longer term follow-up and larger sample sizes are warranted to investigate the actual clinical outcome of this novel strategy.

Conclusion

Dynamic screw stabilization could be achieved with minimally invasive surgery via the CBT by a combination of microscopic decompression and spinal navigation. This innovative dynamic screw stabilization involves no discectomy, no interbody fusion, no facet violation, and little muscle dissection (not even of the Wiltse plane). As a minimally invasive surgery, CBT appeared to be a viable alternative to the conventional pedicle-screw-based dynamic stabilization. However, long term studies are needed to investigate such a minimally invasive surgical strategy.

Acknowledgements

Not applicable.
All procedures performed were in accordance with the ethical standards of Taipei Veterans General Hospital and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. All patients provided written consent before being admitted for surgery.
All patients provided written consent for publication.

Competing interests

All authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Mummaneni PV, Haid RW, Rodts GE. Lumbar interbody fusion: state-of-the-art technical advances. Invited submission from the joint section meeting on disorders of the spine and peripheral nerves, March 2004. J Neurosurg Spine. 2004;1(1):24–30.CrossRef Mummaneni PV, Haid RW, Rodts GE. Lumbar interbody fusion: state-of-the-art technical advances. Invited submission from the joint section meeting on disorders of the spine and peripheral nerves, March 2004. J Neurosurg Spine. 2004;1(1):24–30.CrossRef
2.
Zurück zum Zitat Mummaneni PV, Dhall SS, Eck JC, Groff MW, Ghogawala Z, Watters WC 3rd, Dailey AT, Resnick DK, Choudhri TF, Sharan A, et al. Guideline update for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 11: interbody techniques for lumbar fusion. J Neurosurg Spine. 2014;21(1):67–74.CrossRef Mummaneni PV, Dhall SS, Eck JC, Groff MW, Ghogawala Z, Watters WC 3rd, Dailey AT, Resnick DK, Choudhri TF, Sharan A, et al. Guideline update for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 11: interbody techniques for lumbar fusion. J Neurosurg Spine. 2014;21(1):67–74.CrossRef
3.
Zurück zum Zitat Woods KR, Billys JB, Hynes RA. Technical description of oblique lateral interbody fusion at L1-L5 (OLIF25) and at L5-S1 (OLIF51) and evaluation of complication and fusion rates. Spine J. 2017;17(4):545–53.CrossRef Woods KR, Billys JB, Hynes RA. Technical description of oblique lateral interbody fusion at L1-L5 (OLIF25) and at L5-S1 (OLIF51) and evaluation of complication and fusion rates. Spine J. 2017;17(4):545–53.CrossRef
4.
Zurück zum Zitat Lu G, Kuang L, Wang B. Transforaminal lumbar interbody fusion versus mini-open anterior lumbar interbody fusion with oblique self-anchored stand-alone cages for the treatment of lumbar disc herniation: a retrospective study with 2-year follow up. Spine (Phila Pa 1976). 2017;42(21):E1259–65. https://doi.org/10.1097/BRS.0000000000002145. Lu G, Kuang L, Wang B. Transforaminal lumbar interbody fusion versus mini-open anterior lumbar interbody fusion with oblique self-anchored stand-alone cages for the treatment of lumbar disc herniation: a retrospective study with 2-year follow up. Spine (Phila Pa 1976). 2017;42(21):E1259–65. https://​doi.​org/​10.​1097/​BRS.​0000000000002145​.
5.
Zurück zum Zitat Fay LY, Wu JC, Tsai TY, Wu CL, Huang WC, Cheng H. Dynamic stabilization for degenerative spondylolisthesis: evaluation of radiographic and clinical outcomes. Clin Neurol Neurosurg. 2013;115(5):535–41.CrossRef Fay LY, Wu JC, Tsai TY, Wu CL, Huang WC, Cheng H. Dynamic stabilization for degenerative spondylolisthesis: evaluation of radiographic and clinical outcomes. Clin Neurol Neurosurg. 2013;115(5):535–41.CrossRef
6.
Zurück zum Zitat Kuo CH, Chang PY, Wu JC, Chang HK, Fay LY, Tu TH, Cheng H, Huang WC. Dynamic stabilization for L4-5 spondylolisthesis: comparison with minimally invasive transforaminal lumbar interbody fusion with more than 2 years of follow-up. Neurosurg Focus. 2016;40(1):E3.CrossRef Kuo CH, Chang PY, Wu JC, Chang HK, Fay LY, Tu TH, Cheng H, Huang WC. Dynamic stabilization for L4-5 spondylolisthesis: comparison with minimally invasive transforaminal lumbar interbody fusion with more than 2 years of follow-up. Neurosurg Focus. 2016;40(1):E3.CrossRef
7.
Zurück zum Zitat Wu JC, Huang WC, Tsai HW, Ko CC, Wu CL, Tu TH, Cheng H. Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients. Neurosurg Focus. 2011;31(4):E9.CrossRef Wu JC, Huang WC, Tsai HW, Ko CC, Wu CL, Tu TH, Cheng H. Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients. Neurosurg Focus. 2011;31(4):E9.CrossRef
8.
Zurück zum Zitat Greiner-Perth R, Sellhast N, Perler G, Dietrich D, Staub LP, Roder C. Dynamic posterior stabilization for degenerative lumbar spine disease: a large consecutive case series with long-term follow-up by additional postal survey. Eur Spine J. 2016;25(8):2563–70.CrossRef Greiner-Perth R, Sellhast N, Perler G, Dietrich D, Staub LP, Roder C. Dynamic posterior stabilization for degenerative lumbar spine disease: a large consecutive case series with long-term follow-up by additional postal survey. Eur Spine J. 2016;25(8):2563–70.CrossRef
9.
Zurück zum Zitat Welch WC, Cheng BC, Awad TE, Davis R, Maxwell JH, Delamarter R, Wingate JK, Sherman J, Macenski MM. Clinical outcomes of the Dynesys dynamic neutralization system: 1-year preliminary results. Neurosurg Focus. 2007;22(1):E8.CrossRef Welch WC, Cheng BC, Awad TE, Davis R, Maxwell JH, Delamarter R, Wingate JK, Sherman J, Macenski MM. Clinical outcomes of the Dynesys dynamic neutralization system: 1-year preliminary results. Neurosurg Focus. 2007;22(1):E8.CrossRef
10.
Zurück zum Zitat Beastall J, Karadimas E, Siddiqui M, Nicol M, Hughes J, Smith F, Wardlaw D. The Dynesys lumbar spinal stabilization system: a preliminary report on positional magnetic resonance imaging findings. Spine (Phila Pa 1976). 2007;32(6):685–90.CrossRef Beastall J, Karadimas E, Siddiqui M, Nicol M, Hughes J, Smith F, Wardlaw D. The Dynesys lumbar spinal stabilization system: a preliminary report on positional magnetic resonance imaging findings. Spine (Phila Pa 1976). 2007;32(6):685–90.CrossRef
11.
Zurück zum Zitat Fay LY, Chang PY, Wu JC, Huang WC, Wang CH, Tsai TY, Tu TH, Chang HK, Wu CL, Cheng H. Dynesys dynamic stabilization-related facet arthrodesis. Neurosurg Focus. 2016;40(1):E4.CrossRef Fay LY, Chang PY, Wu JC, Huang WC, Wang CH, Tsai TY, Tu TH, Chang HK, Wu CL, Cheng H. Dynesys dynamic stabilization-related facet arthrodesis. Neurosurg Focus. 2016;40(1):E4.CrossRef
12.
Zurück zum Zitat St-Pierre GH, Jack A, Siddiqui MM, Henderson RL, Nataraj A. Nonfusion does not prevent adjacent segment disease: Dynesys long-term outcomes with minimum five-year follow-up. Spine (Phila Pa 1976). 2016;41(3):265–73.CrossRef St-Pierre GH, Jack A, Siddiqui MM, Henderson RL, Nataraj A. Nonfusion does not prevent adjacent segment disease: Dynesys long-term outcomes with minimum five-year follow-up. Spine (Phila Pa 1976). 2016;41(3):265–73.CrossRef
13.
Zurück zum Zitat Kuo CH, Chang PY, Tu TH, Fay LY, Chang HK, Wu JC, Huang WC, Cheng H. The effect of lumbar Lordosis on screw loosening in Dynesys dynamic stabilization: four-year follow-up with computed tomography. Biomed Res Int. 2015;2015:152435.PubMedPubMedCentral Kuo CH, Chang PY, Tu TH, Fay LY, Chang HK, Wu JC, Huang WC, Cheng H. The effect of lumbar Lordosis on screw loosening in Dynesys dynamic stabilization: four-year follow-up with computed tomography. Biomed Res Int. 2015;2015:152435.PubMedPubMedCentral
14.
Zurück zum Zitat Ko CC, Tsai HW, Huang WC, Wu JC, Chen YC, Shih YH, Chen HC, Wu CL, Cheng H. Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomes. Neurosurg Focus. 2010;28(6):E10.CrossRef Ko CC, Tsai HW, Huang WC, Wu JC, Chen YC, Shih YH, Chen HC, Wu CL, Cheng H. Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomes. Neurosurg Focus. 2010;28(6):E10.CrossRef
15.
Zurück zum Zitat Dayani F, Chen YR, Johnson E, Deb S, Wu Y, Pham L, Singh H. Minimally invasive lumbar pedicle screw fixation using cortical bone trajectory - screw accuracy, complications, and learning curve in 100 screw placements. J Clin Neurosci. 2019;61:106–11.CrossRef Dayani F, Chen YR, Johnson E, Deb S, Wu Y, Pham L, Singh H. Minimally invasive lumbar pedicle screw fixation using cortical bone trajectory - screw accuracy, complications, and learning curve in 100 screw placements. J Clin Neurosci. 2019;61:106–11.CrossRef
16.
Zurück zum Zitat Marengo N, Ajello M, Pecoraro MF, Pilloni G, Vercelli G, Cofano F, Zenga F, Ducati A, Garbossa D. Cortical bone trajectory screws in posterior lumbar Interbody fusion: minimally invasive surgery for maximal muscle sparing-a prospective comparative study with the traditional open technique. Biomed Res Int. 2018;2018:7424568.CrossRef Marengo N, Ajello M, Pecoraro MF, Pilloni G, Vercelli G, Cofano F, Zenga F, Ducati A, Garbossa D. Cortical bone trajectory screws in posterior lumbar Interbody fusion: minimally invasive surgery for maximal muscle sparing-a prospective comparative study with the traditional open technique. Biomed Res Int. 2018;2018:7424568.CrossRef
17.
Zurück zum Zitat Matsukawa K, Yato Y, Kato T, Imabayashi H, Asazuma T, Nemoto K. In vivo analysis of insertional torque during pedicle screwing using cortical bone trajectory technique. Spine (Phila Pa 1976). 2014;39(4):E240–5.CrossRef Matsukawa K, Yato Y, Kato T, Imabayashi H, Asazuma T, Nemoto K. In vivo analysis of insertional torque during pedicle screwing using cortical bone trajectory technique. Spine (Phila Pa 1976). 2014;39(4):E240–5.CrossRef
18.
Zurück zum Zitat Santoni BG, Hynes RA, McGilvray KC, Rodriguez-Canessa G, Lyons AS, Henson MA, Womack WJ, Puttlitz CM. Cortical bone trajectory for lumbar pedicle screws. Spine J. 2009;9(5):366–73.CrossRef Santoni BG, Hynes RA, McGilvray KC, Rodriguez-Canessa G, Lyons AS, Henson MA, Womack WJ, Puttlitz CM. Cortical bone trajectory for lumbar pedicle screws. Spine J. 2009;9(5):366–73.CrossRef
Metadaten
Titel
Minimally invasive dynamic screw stabilization using cortical bone trajectory
verfasst von
Chih-Chang Chang
Chao-Hung Kuo
Hsuan-Kan Chang
Tsung-Hsi Tu
Li-Yu Fay
Jau-Ching Wu
Henrich Cheng
Wen-Cheng Huang
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2020
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-020-03629-z

Weitere Artikel der Ausgabe 1/2020

BMC Musculoskeletal Disorders 1/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.