Prospective diagnostic performance of semiconductor SPECT myocardial perfusion imaging: wall thickening analysis reduces the need for an additional prone acquisition

verfasst von: Loïc Djaileb, Benjamin Dubois, Nicolas de Leiris, Julien Leenhardt, Marjorie Canu, Olivier Phan Sy, Adrien Carabelli, Bastien Boussat, Laurent Dumas, Alexis Broisat, Gérald Vanzetto, Daniel Fagret, Catherine Ghezzi, Gilles Barone-Rochette, Laurent M. Riou

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 10/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To determine whether the assessment of regional wall thickening (WT) in addition to myocardial perfusion from stress supine acquisitions could compensate for the lack of prone acquisition and the corresponding decrease in the diagnostic performance of SPECT myocardial perfusion imaging (MPI) in patients with known or suspected coronary artery disease (CAD).

Methods

The study group comprised 41 patients (123 vessels) with known or suspected CAD prospectively recruited for systematic prone and supine 201Tl stress SPECT MPI. The diagnostic performance of SPECT MPI was determined for various image sets including nongated supine images (supine NG), nongated combined prone and supine images (prone and supine NG) and gated supine images, allowing WT evaluation from NG images in addition to perfusion (supine NG + WT) using invasive coronary angiography and fractional flow reserve as the gold standards.

Results

The rate of false positives was significantly higher among the supine NG images (20.8%) than among either the prone and supine NG or the supine NG + WT images (3.3% and 2.7%, respectively, P < 0.05 vs. supine NG). Consequently, specificity was higher for the prone and supine NG images than for the supine NG images (96.1% vs. 76.1%, P < 0.01) and was highest for the supine NG + WT images (96.8%, P not significant vs. prone and supine NG), without significant differences in sensitivity (80.0%, 86.6% and 73.3%, respectively, P not significant for all comparisons).

Conclusion

The diagnostic performance of supine stress SPECT MPI is improved when WT assessment of ischaemic segments is used as an additional diagnostic criterion to values not significantly different from those with combined prone and supine acquisitions.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Arsanjani R, Hayes SW, Fish M, Shalev A, Nakanishi R, Thomson LE, et al. Two-position supine/prone myocardial perfusion SPECT (MPS) imaging improves visual inter-observer correlation and agreement. J Nucl Cardiol. 2014;21:703–11.CrossRefPubMedPubMedCentral Arsanjani R, Hayes SW, Fish M, Shalev A, Nakanishi R, Thomson LE, et al. Two-position supine/prone myocardial perfusion SPECT (MPS) imaging improves visual inter-observer correlation and agreement. J Nucl Cardiol. 2014;21:703–11.CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Taasan V, Wokhlu A, Taasan MV, Dusaj RS, Mehta A, Kraft S, et al. Comparative accuracy of supine-only and combined supine-prone myocardial perfusion imaging in men. J Nucl Cardiol. 2016;23:1470–6.CrossRefPubMed Taasan V, Wokhlu A, Taasan MV, Dusaj RS, Mehta A, Kraft S, et al. Comparative accuracy of supine-only and combined supine-prone myocardial perfusion imaging in men. J Nucl Cardiol. 2016;23:1470–6.CrossRefPubMed
3.
Zurück zum Zitat Malkerneker D, Brenner R, Martin W, Sampson U, Feurer I, Kronenberg M, et al. CT-based attenuation correction versus prone imaging to decrease equivocal interpretations of rest/stress Tc-99m tetrofosmin SPECT MPI. J Nucl Cardiol. 2007;14:314–23.CrossRefPubMed Malkerneker D, Brenner R, Martin W, Sampson U, Feurer I, Kronenberg M, et al. CT-based attenuation correction versus prone imaging to decrease equivocal interpretations of rest/stress Tc-99m tetrofosmin SPECT MPI. J Nucl Cardiol. 2007;14:314–23.CrossRefPubMed
4.
Zurück zum Zitat Segall GM, Davis MJ, Goris ML. Improved specificity of prone versus supine thallium SPECT imaging. Clin Nucl Med. 1988;13:915–6.CrossRefPubMed Segall GM, Davis MJ, Goris ML. Improved specificity of prone versus supine thallium SPECT imaging. Clin Nucl Med. 1988;13:915–6.CrossRefPubMed
5.
Zurück zum Zitat Slomka P, Nishina H, Abidov A, Hayes S, Friedman J, Berman D, et al. Combined quantitative supine-prone myocardial perfusion SPECT improves detection of coronary artery disease and normalcy rates in women. J Nucl Cardiol. 2007;14:44–52.CrossRefPubMed Slomka P, Nishina H, Abidov A, Hayes S, Friedman J, Berman D, et al. Combined quantitative supine-prone myocardial perfusion SPECT improves detection of coronary artery disease and normalcy rates in women. J Nucl Cardiol. 2007;14:44–52.CrossRefPubMed
6.
Zurück zum Zitat Berman D, Kang X, Nishina H, Slomka P, Shaw L, Hayes S, et al. Diagnostic accuracy of gated Tc-99m sestamibi stress myocardial perfusion SPECT with combined supine and prone acquisitions to detect coronary artery disease in obese and nonobese patients. J Nucl Cardiol. 2006;13:191–201.CrossRef Berman D, Kang X, Nishina H, Slomka P, Shaw L, Hayes S, et al. Diagnostic accuracy of gated Tc-99m sestamibi stress myocardial perfusion SPECT with combined supine and prone acquisitions to detect coronary artery disease in obese and nonobese patients. J Nucl Cardiol. 2006;13:191–201.CrossRef
7.
Zurück zum Zitat Hayes SW, De Lorenzo A, Hachamovitch R, Dhar SC, Hsu P, Cohen I, et al. Prognostic implications of combined prone and supine acquisitions in patients with equivocal or abnormal supine myocardial perfusion SPECT. J Nucl Med. 2003;44:1633–40.PubMed Hayes SW, De Lorenzo A, Hachamovitch R, Dhar SC, Hsu P, Cohen I, et al. Prognostic implications of combined prone and supine acquisitions in patients with equivocal or abnormal supine myocardial perfusion SPECT. J Nucl Med. 2003;44:1633–40.PubMed
8.
Zurück zum Zitat Goto K, Takebayashi H, Kihara Y, Yamane H, Hagikura A, Morimoto Y, et al. Impact of combined supine and prone myocardial perfusion imaging using an ultrafast cardiac gamma camera for detection of inferolateral coronary artery disease. Int J Cardiol. 2014;174:313–7.CrossRefPubMed Goto K, Takebayashi H, Kihara Y, Yamane H, Hagikura A, Morimoto Y, et al. Impact of combined supine and prone myocardial perfusion imaging using an ultrafast cardiac gamma camera for detection of inferolateral coronary artery disease. Int J Cardiol. 2014;174:313–7.CrossRefPubMed
9.
Zurück zum Zitat Nishiyama Y, Miyagawa M, Kawaguchi N, Nakamura M, Kido T, Kurata A, et al. Combined supine and prone myocardial perfusion single-photon emission computed tomography with a cadmium zinc telluride camera for detection of coronary artery disease. Circ J. 2014;78:1169–75.CrossRefPubMed Nishiyama Y, Miyagawa M, Kawaguchi N, Nakamura M, Kido T, Kurata A, et al. Combined supine and prone myocardial perfusion single-photon emission computed tomography with a cadmium zinc telluride camera for detection of coronary artery disease. Circ J. 2014;78:1169–75.CrossRefPubMed
10.
Zurück zum Zitat Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol. 2014;21:695–702.CrossRefPubMed Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol. 2014;21:695–702.CrossRefPubMed
11.
Zurück zum Zitat Redgate S, Barber DC, Fenner JW, Al-Mohammad A, Taylor JC, Hanney MB, et al. A study to quantify the effect of patient motion and develop methods to detect and correct for motion during myocardial perfusion imaging on a CZT solid-state dedicated cardiac camera. J Nucl Cardiol. 2016;23:514–26.CrossRefPubMed Redgate S, Barber DC, Fenner JW, Al-Mohammad A, Taylor JC, Hanney MB, et al. A study to quantify the effect of patient motion and develop methods to detect and correct for motion during myocardial perfusion imaging on a CZT solid-state dedicated cardiac camera. J Nucl Cardiol. 2016;23:514–26.CrossRefPubMed
12.
Zurück zum Zitat Daou D, Sabbah R, Coaguila C, Boulahdour H. Impact of data-driven cardiac respiratory motion correction on the extent and severity of myocardial perfusion defects with free-breathing CZT SPECT. J Nucl Cardiol. 2018;25:1299–309.CrossRefPubMed Daou D, Sabbah R, Coaguila C, Boulahdour H. Impact of data-driven cardiac respiratory motion correction on the extent and severity of myocardial perfusion defects with free-breathing CZT SPECT. J Nucl Cardiol. 2018;25:1299–309.CrossRefPubMed
13.
Zurück zum Zitat Reyes E, Wiener S, Underwood SR, European Council of Nuclear Cardiology. Myocardial perfusion scintigraphy in Europe 2007: a survey of the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2012;39:160–4.CrossRefPubMed Reyes E, Wiener S, Underwood SR, European Council of Nuclear Cardiology. Myocardial perfusion scintigraphy in Europe 2007: a survey of the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2012;39:160–4.CrossRefPubMed
14.
Zurück zum Zitat Agostini D, Marie P-Y, Ben-Haim S, Rouzet F, Songy B, Cardiovascular Committee of the European Association of Nuclear Medicine (EANM), et al. Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the Cardiovascular Committee of the European Association of Nuclear Medicine (EANM). Eur J Nucl Med Mol Imaging. 2016;43:2423–32.CrossRef Agostini D, Marie P-Y, Ben-Haim S, Rouzet F, Songy B, Cardiovascular Committee of the European Association of Nuclear Medicine (EANM), et al. Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the Cardiovascular Committee of the European Association of Nuclear Medicine (EANM). Eur J Nucl Med Mol Imaging. 2016;43:2423–32.CrossRef
15.
Zurück zum Zitat Mannting F, Morgan-Mannting MG. Gated SPECT with technetium-99m-sestamibi for assessment of myocardial perfusion abnormalities. J Nucl Med. 1993;34:601–8.PubMed Mannting F, Morgan-Mannting MG. Gated SPECT with technetium-99m-sestamibi for assessment of myocardial perfusion abnormalities. J Nucl Med. 1993;34:601–8.PubMed
16.
Zurück zum Zitat Germano G, Erel J, Kiat H, Kavanagh PB, Berman DS. Quantitative LVEF and qualitative regional function from gated thallium-201 perfusion SPECT. J Nucl Med. 1997;38:749–54.PubMed Germano G, Erel J, Kiat H, Kavanagh PB, Berman DS. Quantitative LVEF and qualitative regional function from gated thallium-201 perfusion SPECT. J Nucl Med. 1997;38:749–54.PubMed
17.
Zurück zum Zitat Smanio PE, Watson DD, Segalla DL, Vinson EL, Smith WH, Beller GA. Value of gating of technetium-99m sestamibi single-photon emission computed tomographic imaging. J Am Coll Cardiol. 1997;30:1687–92.CrossRefPubMed Smanio PE, Watson DD, Segalla DL, Vinson EL, Smith WH, Beller GA. Value of gating of technetium-99m sestamibi single-photon emission computed tomographic imaging. J Am Coll Cardiol. 1997;30:1687–92.CrossRefPubMed
18.
Zurück zum Zitat DePuey EG, Rozanski A. Using gated technetium-99m-sestamibi SPECT to characterize fixed myocardial defects as infarct or artifact. J Nucl Med. 1995;36:952–5. DePuey EG, Rozanski A. Using gated technetium-99m-sestamibi SPECT to characterize fixed myocardial defects as infarct or artifact. J Nucl Med. 1995;36:952–5.
19.
Zurück zum Zitat Danias PG, Papaioannou GI, Ahlberg AW, O’Sullivan DM, Mann A, Boden WE, et al. Usefulness of electrocardiographic-gated stress technetium-99m sestamibi single-photon emission computed tomography to differentiate ischemic from nonischemic cardiomyopathy. Am J Cardiol. 2004;94:14–9.CrossRefPubMed Danias PG, Papaioannou GI, Ahlberg AW, O’Sullivan DM, Mann A, Boden WE, et al. Usefulness of electrocardiographic-gated stress technetium-99m sestamibi single-photon emission computed tomography to differentiate ischemic from nonischemic cardiomyopathy. Am J Cardiol. 2004;94:14–9.CrossRefPubMed
20.
Zurück zum Zitat Yoda S, Sato Y, Matsumoto N, Tani S, Takayama T, Nishina H, et al. Incremental value of regional wall motion analysis immediately after exercise for the detection of single-vessel coronary artery disease: study by separate acquisition, dual-isotope ECG-gated single-photon emission computed tomography. Circ J. 2005;69:301–5.CrossRefPubMed Yoda S, Sato Y, Matsumoto N, Tani S, Takayama T, Nishina H, et al. Incremental value of regional wall motion analysis immediately after exercise for the detection of single-vessel coronary artery disease: study by separate acquisition, dual-isotope ECG-gated single-photon emission computed tomography. Circ J. 2005;69:301–5.CrossRefPubMed
21.
Zurück zum Zitat Fleischmann S, Koepfli P, Namdar M, Wyss CA, Jenni R, Kaufmann PA. Gated (99m)Tc-tetrofosmin SPECT for discriminating infarct from artifact in fixed myocardial perfusion defects. J Nucl Med. 2004;45:754–9.PubMed Fleischmann S, Koepfli P, Namdar M, Wyss CA, Jenni R, Kaufmann PA. Gated (99m)Tc-tetrofosmin SPECT for discriminating infarct from artifact in fixed myocardial perfusion defects. J Nucl Med. 2004;45:754–9.PubMed
22.
Zurück zum Zitat Bailliez A, Lairez O, Merlin C, Piriou N, Legallois D, Blaire T, et al. Left ventricular function assessment using 2 different cadmium-zinc-telluride cameras compared with a γ-camera with cardiofocal collimators: dynamic cardiac phantom study and clinical validation. J Nucl Med. 2016;57:1370–5.CrossRefPubMed Bailliez A, Lairez O, Merlin C, Piriou N, Legallois D, Blaire T, et al. Left ventricular function assessment using 2 different cadmium-zinc-telluride cameras compared with a γ-camera with cardiofocal collimators: dynamic cardiac phantom study and clinical validation. J Nucl Med. 2016;57:1370–5.CrossRefPubMed
23.
Zurück zum Zitat Coupez E, Merlin C, Tuyisenge V, Sarry L, Pereira B, Lusson JR, et al. Validation of cadmium-zinc-telluride camera for measurement of left ventricular systolic performance. J Nucl Cardiol. 2018;25:1029–36.CrossRefPubMed Coupez E, Merlin C, Tuyisenge V, Sarry L, Pereira B, Lusson JR, et al. Validation of cadmium-zinc-telluride camera for measurement of left ventricular systolic performance. J Nucl Cardiol. 2018;25:1029–36.CrossRefPubMed
24.
Zurück zum Zitat Barone-Rochette G, Leclere M, Calizzano A, Vautrin E, Céline G-C, Broisat A, et al. Stress thallium-201/rest technetium-99m sequential dual-isotope high-speed myocardial perfusion imaging validation versus invasive coronary angiography. J Nucl Cardiol. 2015;22:513–22.CrossRefPubMed Barone-Rochette G, Leclere M, Calizzano A, Vautrin E, Céline G-C, Broisat A, et al. Stress thallium-201/rest technetium-99m sequential dual-isotope high-speed myocardial perfusion imaging validation versus invasive coronary angiography. J Nucl Cardiol. 2015;22:513–22.CrossRefPubMed
25.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–42.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–42.CrossRefPubMed
26.
Zurück zum Zitat Konno M, Morita K, Adachi I, Ito Y, Kohya T, Kitabatake A, et al. Quantitative analysis of regional wall motion and thickening by quantitative gated SPECT: comparison with visual analysis. Clin Nucl Med. 2001;26:202–7.CrossRefPubMed Konno M, Morita K, Adachi I, Ito Y, Kohya T, Kitabatake A, et al. Quantitative analysis of regional wall motion and thickening by quantitative gated SPECT: comparison with visual analysis. Clin Nucl Med. 2001;26:202–7.CrossRefPubMed
27.
Zurück zum Zitat Pereztol-Valdés O, Candell-Riera J, Santana-Boado C, Angel J, Aguadé-Bruix S, Castell-Conesa J, et al. Correspondence between left ventricular 17 myocardial segments and coronary arteries. Eur Heart J. 2005;26:2637–43.CrossRefPubMed Pereztol-Valdés O, Candell-Riera J, Santana-Boado C, Angel J, Aguadé-Bruix S, Castell-Conesa J, et al. Correspondence between left ventricular 17 myocardial segments and coronary arteries. Eur Heart J. 2005;26:2637–43.CrossRefPubMed
28.
Zurück zum Zitat Bailliez A, Blaire T, Mouquet F, Legghe R, Etienne B, Legallois D, et al. Segmental and global left ventricular function assessment using gated SPECT with a semiconductor cadmium zinc telluride (CZT) camera: phantom study and clinical validation vs cardiac magnetic resonance. J Nucl Cardiol. 2014;21:712–22.CrossRefPubMed Bailliez A, Blaire T, Mouquet F, Legghe R, Etienne B, Legallois D, et al. Segmental and global left ventricular function assessment using gated SPECT with a semiconductor cadmium zinc telluride (CZT) camera: phantom study and clinical validation vs cardiac magnetic resonance. J Nucl Cardiol. 2014;21:712–22.CrossRefPubMed
29.
Zurück zum Zitat Cochet H, Bullier E, Gerbaud E, Durieux M, Godbert Y, Lederlin M, et al. Absolute quantification of left ventricular global and regional function at nuclear MPI using ultrafast CZT SPECT: initial validation versus cardiac MR. J Nucl Med. 2013;54:556–63.CrossRefPubMed Cochet H, Bullier E, Gerbaud E, Durieux M, Godbert Y, Lederlin M, et al. Absolute quantification of left ventricular global and regional function at nuclear MPI using ultrafast CZT SPECT: initial validation versus cardiac MR. J Nucl Med. 2013;54:556–63.CrossRefPubMed
30.
Zurück zum Zitat Giorgetti A, Masci PG, Marras G, Rustamova YK, Gimelli A, Genovesi D, et al. Gated SPECT evaluation of left ventricular function using a CZT camera and a fast low-dose clinical protocol: comparison to cardiac magnetic resonance imaging. Eur J Nucl Med Mol Imaging. 2013;40:1869–75.CrossRefPubMed Giorgetti A, Masci PG, Marras G, Rustamova YK, Gimelli A, Genovesi D, et al. Gated SPECT evaluation of left ventricular function using a CZT camera and a fast low-dose clinical protocol: comparison to cardiac magnetic resonance imaging. Eur J Nucl Med Mol Imaging. 2013;40:1869–75.CrossRefPubMed
31.
Zurück zum Zitat Rozanski A, Diamond GA, Berman D, Forrester JS, Morris D, Swan HJC. The declining specificity of exercise radionuclide ventriculography. N Engl J Med. 1983;309:518–22.CrossRefPubMed Rozanski A, Diamond GA, Berman D, Forrester JS, Morris D, Swan HJC. The declining specificity of exercise radionuclide ventriculography. N Engl J Med. 1983;309:518–22.CrossRefPubMed
32.
Zurück zum Zitat Rozanski A, Berman DS. The efficacy of cardiovascular nuclear medicine exercise studies. Semin Nucl Med. 1987;17:104–20.CrossRefPubMed Rozanski A, Berman DS. The efficacy of cardiovascular nuclear medicine exercise studies. Semin Nucl Med. 1987;17:104–20.CrossRefPubMed
33.
Zurück zum Zitat Berman DS, Kang X, Tamarappoo B, Wolak A, Hayes SW, Nakazato R, et al. Stress thallium-201/rest technetium-99m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging. 2009;2:273–82.CrossRefPubMed Berman DS, Kang X, Tamarappoo B, Wolak A, Hayes SW, Nakazato R, et al. Stress thallium-201/rest technetium-99m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging. 2009;2:273–82.CrossRefPubMed
Metadaten
Titel
Prospective diagnostic performance of semiconductor SPECT myocardial perfusion imaging: wall thickening analysis reduces the need for an additional prone acquisition
verfasst von
Loïc Djaileb
Benjamin Dubois
Nicolas de Leiris
Julien Leenhardt
Marjorie Canu
Olivier Phan Sy
Adrien Carabelli
Bastien Boussat
Laurent Dumas
Alexis Broisat
Gérald Vanzetto
Daniel Fagret
Catherine Ghezzi
Gilles Barone-Rochette
Laurent M. Riou
Publikationsdatum
18.07.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 10/2019
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-019-04415-3

Weitere Artikel der Ausgabe 10/2019

European Journal of Nuclear Medicine and Molecular Imaging 10/2019 Zur Ausgabe