Skip to main content
Erschienen in: HSS Journal ® 1/2018

12.09.2017 | Original Article

Scaling Marker Position Determines the Accuracy of Digital Templating for Total Hip Arthroplasty

verfasst von: Austin J. Ramme, MD, PhD, Nina D. Fisher, BS, Jonathan Egol, Gregory Chang, MD, Jonathan M. Vigdorchik, MD

Erschienen in: HSS Journal ® | Ausgabe 1/2018

Einloggen, um Zugang zu erhalten

Abstract

Background

Digital templating systems foster patient-specific measurements for preoperative planning.

Questions/Purposes

We aim (1) to verify the accuracy of a templating system, (2) to describe the effects of scaling marker position on the accuracy of digital templating of the hip, and (3) to provide a practical guide for scaling marker position using patient body mass index (BMI).

Methods

A scaling sphere was placed in five positions along the anterior-posterior axis of an acetabular implant and pelvis phantom, and x-rays were obtained. Each radiograph was templated for the acetabular component and recorded. A retrospective review identified CT scans of preoperative hip arthroplasty cases. The center of the greater trochanter was calculated from these CT scans as the percent distance from the anterior thigh and recorded with the patient’s BMI.

Results

By centering the scaling sphere on the acetabular component, an accurate cup size was achieved. A difference of 3.5 cm in sphere placement resulted in a full cup size magnification error. Positioning the scaling sphere at the level of the pubic symphysis resulted in a difference of four cup sizes. This patient population had an average BMI of 28.72 kg/m2 (standard deviation 6.26 kg/m2) and an average position of the center of the greater trochanter of 51% (standard deviation of 6%) from the anterior surface of thigh.

Conclusions

Digital templating relies on scaling marker position to accurately estimate implant size. Based on the findings in this study, scaling markers for hip imaging should be placed laterally, mid-thigh in the anterior-posterior direction for patients with a BMI between 25 and 40 kg/m2. If abnormal hip anatomy or extremes of BMI are discovered, then scaling sphere positioning should be optimized on a case-by-case basis. Digital templating systems for total hip arthroplasty must use precisely placed scaling markers at the level of the hip joint to allow for accurate implant size estimation.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
2.
Zurück zum Zitat Brew CJ, Simpson PM, Whitehouse SL, et al. Scaling digital radiographs for templating in total hip arthroplasty using conventional acetate templates independent of calibration markers. J Arthroplasty. 2012; 27:643–647.CrossRefPubMed Brew CJ, Simpson PM, Whitehouse SL, et al. Scaling digital radiographs for templating in total hip arthroplasty using conventional acetate templates independent of calibration markers. J Arthroplasty. 2012; 27:643–647.CrossRefPubMed
3.
Zurück zum Zitat Hsu AR, Gross CE, Bhatia S, et al. Template-directed instrumentation in total knee arthroplasty: cost savings analysis. Orthopedics. 2012; 35:e1596–1600.CrossRefPubMed Hsu AR, Gross CE, Bhatia S, et al. Template-directed instrumentation in total knee arthroplasty: cost savings analysis. Orthopedics. 2012; 35:e1596–1600.CrossRefPubMed
4.
Zurück zum Zitat Iorio R, Siegel J, Specht LM, et al. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe? J Arthroplasty. 2009; 24:175–179.CrossRefPubMed Iorio R, Siegel J, Specht LM, et al. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe? J Arthroplasty. 2009; 24:175–179.CrossRefPubMed
5.
Zurück zum Zitat Issa K, Pivec R, Boyd B, et al. Comparing the accuracy of radiographic preoperative digital templating for a second- versus a first-generation THA stem. Orthopedics. 2012; 35:1028–1034.CrossRefPubMed Issa K, Pivec R, Boyd B, et al. Comparing the accuracy of radiographic preoperative digital templating for a second- versus a first-generation THA stem. Orthopedics. 2012; 35:1028–1034.CrossRefPubMed
6.
Zurück zum Zitat Ramme AJ, Iorio R, Smiaronksi J, et al. Organizational and Technical Considerations for the Implementation of a Digital Orthopaedic Templating System. Bull Hosp Jt Dis (2013); 74:254–261. Ramme AJ, Iorio R, Smiaronksi J, et al. Organizational and Technical Considerations for the Implementation of a Digital Orthopaedic Templating System. Bull Hosp Jt Dis (2013); 74:254–261.
7.
Zurück zum Zitat Sinclair VF, Wilson J, Jain NP, et al. Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty. J Arthroplasty. 2014; 29:1658–1660.CrossRefPubMed Sinclair VF, Wilson J, Jain NP, et al. Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty. J Arthroplasty. 2014; 29:1658–1660.CrossRefPubMed
8.
Zurück zum Zitat White SP and Shardlow DL. Effect of introduction of digital radiographic techniques on pre-operative templating in orthopaedic practice. Ann R Coll Surg Engl. 2005; 87:53–54.CrossRefPubMedPubMedCentral White SP and Shardlow DL. Effect of introduction of digital radiographic techniques on pre-operative templating in orthopaedic practice. Ann R Coll Surg Engl. 2005; 87:53–54.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Wimsey S, Pickard R and Shaw G. Accurate scaling of digital radiographs of the pelvis. A prospective trial of two methods. J Bone Joint Surg Br. 2006; 88:1508–1512.CrossRefPubMed Wimsey S, Pickard R and Shaw G. Accurate scaling of digital radiographs of the pelvis. A prospective trial of two methods. J Bone Joint Surg Br. 2006; 88:1508–1512.CrossRefPubMed
Metadaten
Titel
Scaling Marker Position Determines the Accuracy of Digital Templating for Total Hip Arthroplasty
verfasst von
Austin J. Ramme, MD, PhD
Nina D. Fisher, BS
Jonathan Egol
Gregory Chang, MD
Jonathan M. Vigdorchik, MD
Publikationsdatum
12.09.2017
Verlag
Springer US
Erschienen in
HSS Journal ® / Ausgabe 1/2018
Print ISSN: 1556-3316
Elektronische ISSN: 1556-3324
DOI
https://doi.org/10.1007/s11420-017-9578-0

Weitere Artikel der Ausgabe 1/2018

HSS Journal ® 1/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.