Erschienen in: Clinical Orthopaedics and Related Research® 2/2014

01.02.2014 | Symposium: 2013 Hip Society Proceedings

Stability and Trunnion Wear Potential in Large-diameter Metal-on-Metal Total Hips: A Finite Element Analysis

verfasst von: Jacob M. Elkins, MD, PhD, John J. Callaghan, MD, Thomas D. Brown, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 2/2014

Einloggen, um Zugang zu erhalten

Abstract

Background

Large-diameter femoral heads for metal-on-metal THA hold theoretical advantages of joint stability and low bearing surface wear. However, recent reports have indicated an unacceptably high rate of wear-associated failure with large-diameter bearings, possibly due in part to increased wear at the trunnion interface. Thus, the deleterious consequences of using large heads may outweigh their theoretical advantages.

Questions/purposes

We investigated (1) to what extent femoral head size influenced stability in THA for several dislocation-prone motions; and the biomechanics of wear at the trunnion interface by considering the relationship between (2) wear potential and head size and (3) wear potential and other factors, including cup orientation, type of hip motion, and assembly/impaction load.

Methods

Computational simulations were executed using a previously validated nonlinear contact finite element model. Stability was determined at 36 cup orientations for five distinct dislocation challenges. Wear at the trunnion interface was calculated for three separate cup orientations subjected to gait, stooping, and sit-to-stand motions. Seven head diameters were investigated: 32 to 56 mm, in 4-mm increments.

Results

Stability improved with increased diameter, although diminishing benefit was seen for sizes of greater than 40 mm. By contrast, contact stress and computed wear at the trunnion interface all increased unabatedly with increasing head size. Increased impaction forces resulted in only small decreases in trunnion wear generation.

Conclusions

These data suggest that the theoretical advantages of large-diameter femoral heads have a limit. Diameters of greater than 40 mm demonstrated only modest improvement in terms of joint stability yet incurred substantial increase in wear potential at the trunnion.

Clinical Relevance

Our model has potential to help investigators and designers of hip implants to better understand the optimization of trunnion design for long-term durability.
Literatur
1.
Zurück zum Zitat Archard J. Contact and rubbing of flat surfaces. J Appl Phys. 1953;24:981–988.CrossRef Archard J. Contact and rubbing of flat surfaces. J Appl Phys. 1953;24:981–988.CrossRef
3.
Zurück zum Zitat Bartz RL, Noble PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.PubMed Bartz RL, Noble PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.PubMed
4.
Zurück zum Zitat Berton C, Girard J, Krantz N, Migaud H. The Durom large diameter head acetabular component: early results with a large-diameter metal-on-metal bearing. J Bone Joint Surg Br. 2010;92:202–208.PubMedCrossRef Berton C, Girard J, Krantz N, Migaud H. The Durom large diameter head acetabular component: early results with a large-diameter metal-on-metal bearing. J Bone Joint Surg Br. 2010;92:202–208.PubMedCrossRef
5.
Zurück zum Zitat Bolland B, Culliford D, Langton D, Millington J, Arden N, Latham J. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93:608–615.PubMedCrossRef Bolland B, Culliford D, Langton D, Millington J, Arden N, Latham J. High failure rates with a large-diameter hybrid metal-on-metal total hip replacement: clinical, radiological and retrieval analysis. J Bone Joint Surg Br. 2011;93:608–615.PubMedCrossRef
6.
Zurück zum Zitat Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:128–133.PubMedCrossRef Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:128–133.PubMedCrossRef
7.
Zurück zum Zitat Brockett C, Williams S, Jin Z, Isaac G, Fisher J. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater. 2007;81:508–515.PubMedCrossRef Brockett C, Williams S, Jin Z, Isaac G, Fisher J. Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater. 2007;81:508–515.PubMedCrossRef
8.
Zurück zum Zitat Brown S, Flemming C, Kawalec J, Placko H, Vassaux C, Merritt K, Payer J, Kraay M. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995;6:19–26.PubMedCrossRef Brown S, Flemming C, Kawalec J, Placko H, Vassaux C, Merritt K, Payer J, Kraay M. Fretting corrosion accelerates crevice corrosion of modular hip tapers. J Appl Biomater. 1995;6:19–26.PubMedCrossRef
9.
Zurück zum Zitat Browne JA, Bechtold CD, Berry DJ, Hanssen AD, Lewallen DG. Failed metal-on-metal hip arthroplasties: a spectrum of clinical presentations and operative findings. Clin Orthop Relat Res. 2010;468:2313–2320.PubMedCrossRef Browne JA, Bechtold CD, Berry DJ, Hanssen AD, Lewallen DG. Failed metal-on-metal hip arthroplasties: a spectrum of clinical presentations and operative findings. Clin Orthop Relat Res. 2010;468:2313–2320.PubMedCrossRef
10.
Zurück zum Zitat Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes: an in vitro study. J Arthroplasty. 2005;20:11–19.PubMedCrossRef Burroughs BR, Hallstrom B, Golladay GJ, Hoeffel D, Harris WH. Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes: an in vitro study. J Arthroplasty. 2005;20:11–19.PubMedCrossRef
11.
Zurück zum Zitat Cinotti G, Lucioli N, Malagoli A, Calderoli C, Cassese F. Do large femoral heads reduce the risks of impingement in total hip arthroplasty with optimal and non-optimal cup positioning? Int Orthop. 2011;3:317–323.CrossRef Cinotti G, Lucioli N, Malagoli A, Calderoli C, Cassese F. Do large femoral heads reduce the risks of impingement in total hip arthroplasty with optimal and non-optimal cup positioning? Int Orthop. 2011;3:317–323.CrossRef
12.
Zurück zum Zitat Collier JP, Mayor MB, Williams IR, Surprenant VA, Surprenant HP, Currier BH. The tradeoffs associated with modular hip prostheses. Clin Orthop Relat Res. 1995;311:91–101.PubMed Collier JP, Mayor MB, Williams IR, Surprenant VA, Surprenant HP, Currier BH. The tradeoffs associated with modular hip prostheses. Clin Orthop Relat Res. 1995;311:91–101.PubMed
13.
Zurück zum Zitat Cook SD, Barrack RL, Clemow A. Corrosion and wear at the modular interface of uncemented femoral stems. J Bone Joint Surg Br. 1994;76:68–72.PubMed Cook SD, Barrack RL, Clemow A. Corrosion and wear at the modular interface of uncemented femoral stems. J Bone Joint Surg Br. 1994;76:68–72.PubMed
14.
Zurück zum Zitat Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: what they do and don’t do. Clin Orthop Relat Res. 2004;429:102–107.PubMedCrossRef Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: what they do and don’t do. Clin Orthop Relat Res. 2004;429:102–107.PubMedCrossRef
15.
Zurück zum Zitat Cuckler JM, Moore KD, Lombardi AV, McPherson E, Emerson R. Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplasty. 2004;19:41–44.PubMedCrossRef Cuckler JM, Moore KD, Lombardi AV, McPherson E, Emerson R. Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplasty. 2004;19:41–44.PubMedCrossRef
16.
Zurück zum Zitat De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.PubMedCrossRef De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.PubMedCrossRef
17.
Zurück zum Zitat D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82:315–321.PubMed D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82:315–321.PubMed
18.
Zurück zum Zitat Dobbs H, Minski M. Metal ion release after total hip replacement. Biomaterials. 1980;1:193–198.PubMedCrossRef Dobbs H, Minski M. Metal ion release after total hip replacement. Biomaterials. 1980;1:193–198.PubMedCrossRef
19.
Zurück zum Zitat Donell S, Darrah C, Nolan J, Wimhurst J, Toms A, Barker T, Case C, Tucker J. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs. J Bone Joint Surg Br. 2010;92:1501–1508.PubMedCrossRef Donell S, Darrah C, Nolan J, Wimhurst J, Toms A, Barker T, Case C, Tucker J. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs. J Bone Joint Surg Br. 2010;92:1501–1508.PubMedCrossRef
20.
Zurück zum Zitat Dorr LD, Wolf AW, Chandler R, Conaty JP. Classification and treatment of dislocations of total hip arthroplasty. Clin Orthop Relat Res. 1983;173:151–158.PubMed Dorr LD, Wolf AW, Chandler R, Conaty JP. Classification and treatment of dislocations of total hip arthroplasty. Clin Orthop Relat Res. 1983;173:151–158.PubMed
21.
Zurück zum Zitat Dowson D, Hardaker C, Flett M, Isaac GH. A hip joint simulator study of the performance of metal-on-metal joints. Part II: Design. J Arthroplasty. 2004;19:124–130. Dowson D, Hardaker C, Flett M, Isaac GH. A hip joint simulator study of the performance of metal-on-metal joints. Part II: Design. J Arthroplasty. 2004;19:124–130.
22.
Zurück zum Zitat Elias JJ, Nagao M, Chu YH, Carbone JJ, Lennox DW, Chao E. Medial cortex strain distribution during noncemented total hip arthroplasty. Clin Orthop Relat Res. 2000;370:250–258.PubMedCrossRef Elias JJ, Nagao M, Chu YH, Carbone JJ, Lennox DW, Chao E. Medial cortex strain distribution during noncemented total hip arthroplasty. Clin Orthop Relat Res. 2000;370:250–258.PubMedCrossRef
23.
Zurück zum Zitat Elkins JM, Kruger KM, Pedersen DR, Callaghan JJ, Brown TD. Edge-loading severity as a function of cup lip radius in metal-on-metal total hips—a finite element analysis. J Orthop Res. 2012;30:169–177.PubMedCrossRef Elkins JM, Kruger KM, Pedersen DR, Callaghan JJ, Brown TD. Edge-loading severity as a function of cup lip radius in metal-on-metal total hips—a finite element analysis. J Orthop Res. 2012;30:169–177.PubMedCrossRef
24.
Zurück zum Zitat Elkins JM, O’Brien MK, Stroud NJ, Pedersen DR, Callaghan JJ, Brown TD. Hard-on-hard total hip impingement causes extreme contact stress concentrations. Clin Orthop Relat Res. 2011;469:454–463.PubMedCrossRef Elkins JM, O’Brien MK, Stroud NJ, Pedersen DR, Callaghan JJ, Brown TD. Hard-on-hard total hip impingement causes extreme contact stress concentrations. Clin Orthop Relat Res. 2011;469:454–463.PubMedCrossRef
25.
Zurück zum Zitat Elkins JM, Stroud NJ, Rudert MJ, Tochigi Y, Pedersen DR, Ellis BJ, Callaghan JJ, Weiss JA, Brown TD. The capsule’s contribution to total hip construct stability—a finite element analysis. J Orthop Res. 2011;29:1642–1648.PubMedCentralPubMedCrossRef Elkins JM, Stroud NJ, Rudert MJ, Tochigi Y, Pedersen DR, Ellis BJ, Callaghan JJ, Weiss JA, Brown TD. The capsule’s contribution to total hip construct stability—a finite element analysis. J Orthop Res. 2011;29:1642–1648.PubMedCentralPubMedCrossRef
26.
Zurück zum Zitat Fessler H, Fricker D. Friction in femoral prosthesis and photoelastic model cone taper joints. Proc Inst Mech Eng H. 1989;203:1–14.PubMedCrossRef Fessler H, Fricker D. Friction in femoral prosthesis and photoelastic model cone taper joints. Proc Inst Mech Eng H. 1989;203:1–14.PubMedCrossRef
27.
Zurück zum Zitat Fisher J. Bioengineering reasons for the failure of metal-on-metal hip prostheses: an engineer’s perspective. J Bone Joint Surg Br. 2011;93:1001–1004.PubMedCrossRef Fisher J. Bioengineering reasons for the failure of metal-on-metal hip prostheses: an engineer’s perspective. J Bone Joint Surg Br. 2011;93:1001–1004.PubMedCrossRef
28.
Zurück zum Zitat Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award. Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2010;468:318–325.PubMedCrossRef Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award. Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2010;468:318–325.PubMedCrossRef
29.
Zurück zum Zitat Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.PubMedCrossRef Gilbert JL, Buckley CA, Jacobs JJ. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations: the effect of crevice, stress, motion, and alloy coupling. J Biomed Mater Res. 1993;27:1533–1544.PubMedCrossRef
30.
Zurück zum Zitat Girard J, Bocquet D, Autissier G, Fouilleron N, Fron D, Migaud H. Metal-on-metal hip arthroplasty in patients thirty years of age or younger. J Bone Joint Surg Am. 2010;92:2419–2426.PubMedCrossRef Girard J, Bocquet D, Autissier G, Fouilleron N, Fron D, Migaud H. Metal-on-metal hip arthroplasty in patients thirty years of age or younger. J Bone Joint Surg Am. 2010;92:2419–2426.PubMedCrossRef
31.
Zurück zum Zitat Goldberg JR, Gilbert JL. In vitro corrosion testing of modular hip tapers. J Biomed Mater Res B Appl Biomater. 2003;64:78–93.PubMedCrossRef Goldberg JR, Gilbert JL. In vitro corrosion testing of modular hip tapers. J Biomed Mater Res B Appl Biomater. 2003;64:78–93.PubMedCrossRef
32.
Zurück zum Zitat Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.PubMedCrossRef Goldberg JR, Gilbert JL, Jacobs JJ, Bauer TW, Paprosky W, Leurgans S. A multicenter retrieval study of the taper interfaces of modular hip prostheses. Clin Orthop Relat Res. 2002;401:149–161.PubMedCrossRef
33.
Zurück zum Zitat Heiney JP, Battula S, Vrabec GA, Parikh A, Blice R, Schoenfeld AJ, Njus GO. Impact magnitudes applied by surgeons and their importance when applying the femoral head onto the Morse taper for total hip arthroplasty. Arch Orthop Trauma Surg. 2009;129:793–796.PubMedCrossRef Heiney JP, Battula S, Vrabec GA, Parikh A, Blice R, Schoenfeld AJ, Njus GO. Impact magnitudes applied by surgeons and their importance when applying the femoral head onto the Morse taper for total hip arthroplasty. Arch Orthop Trauma Surg. 2009;129:793–796.PubMedCrossRef
34.
Zurück zum Zitat Insall JN. Presidential address to The Knee Society: choices and compromises in total knee arthroplasty. Clin Orthop Relat Res. 1988;226:43–48.PubMed Insall JN. Presidential address to The Knee Society: choices and compromises in total knee arthroplasty. Clin Orthop Relat Res. 1988;226:43–48.PubMed
35.
Zurück zum Zitat Jacobs JJ, Gilbert JL, Urban RM. Current concepts review: corrosion of metal orthopaedic implants. J Bone Joint Surg Am. 1998;80:268–282.PubMed Jacobs JJ, Gilbert JL, Urban RM. Current concepts review: corrosion of metal orthopaedic implants. J Bone Joint Surg Am. 1998;80:268–282.PubMed
36.
Zurück zum Zitat Khan M, Kuiper JH, Richardson J. The exercise-related rise in plasma cobalt levels after metal-on-metal hip resurfacing arthroplasty. J Bone Joint Surg Br. 2008;90:1152–1157.PubMedCrossRef Khan M, Kuiper JH, Richardson J. The exercise-related rise in plasma cobalt levels after metal-on-metal hip resurfacing arthroplasty. J Bone Joint Surg Br. 2008;90:1152–1157.PubMedCrossRef
37.
Zurück zum Zitat Kop AM, Swarts E. Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty. 2009;24:1019–1023.PubMedCrossRef Kop AM, Swarts E. Corrosion of a hip stem with a modular neck taper junction: a retrieval study of 16 cases. J Arthroplasty. 2009;24:1019–1023.PubMedCrossRef
38.
Zurück zum Zitat Langton D, Jameson S, Joyce T, Gandhi J, Sidaginamale R, Mereddy P, Lord J, Nargol A. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93:1011–1016.PubMedCrossRef Langton D, Jameson S, Joyce T, Gandhi J, Sidaginamale R, Mereddy P, Lord J, Nargol A. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93:1011–1016.PubMedCrossRef
39.
Zurück zum Zitat Langton D, Jameson S, Joyce T, Hallab N, Natu S, Nargol A. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.PubMedCrossRef Langton D, Jameson S, Joyce T, Hallab N, Natu S, Nargol A. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.PubMedCrossRef
40.
Zurück zum Zitat Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J. High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res. 2009;467:2259–2265.PubMedCrossRef Leslie IJ, Williams S, Isaac G, Ingham E, Fisher J. High cup angle and microseparation increase the wear of hip surface replacements. Clin Orthop Relat Res. 2009;467:2259–2265.PubMedCrossRef
41.
Zurück zum Zitat Liu F, Leslie I, Williams S, Fisher J, Jin Z. Development of computational wear simulation of metal-on-metal hip resurfacing replacements. J Biomech. 2008;41:686–694.PubMedCrossRef Liu F, Leslie I, Williams S, Fisher J, Jin Z. Development of computational wear simulation of metal-on-metal hip resurfacing replacements. J Biomech. 2008;41:686–694.PubMedCrossRef
42.
Zurück zum Zitat Lombardi AV, Skeels MD, Berend KR, Adams JB, Franchi OJ. Do large heads enhance stability and restore native anatomy in primary total hip arthroplasty? Clin Orthop Relat Res. 2011;469:1547–1553.PubMedCrossRef Lombardi AV, Skeels MD, Berend KR, Adams JB, Franchi OJ. Do large heads enhance stability and restore native anatomy in primary total hip arthroplasty? Clin Orthop Relat Res. 2011;469:1547–1553.PubMedCrossRef
43.
Zurück zum Zitat Mao X, Tay GH, Godbolt DB, Crawford RW. Pseudotumor in a well-fixed metal-on-polyethylene uncemented hip arthroplasty. J Arthroplasty. 2012;27:493e13–493e16.CrossRef Mao X, Tay GH, Godbolt DB, Crawford RW. Pseudotumor in a well-fixed metal-on-polyethylene uncemented hip arthroplasty. J Arthroplasty. 2012;27:493e13–493e16.CrossRef
44.
Zurück zum Zitat McKellop HA, Sarmiento A, Brien W, Park SH. Interface corrosion of a modular head total hip prosthesis. J Arthroplasty. 1992;7:291–294.PubMedCrossRef McKellop HA, Sarmiento A, Brien W, Park SH. Interface corrosion of a modular head total hip prosthesis. J Arthroplasty. 1992;7:291–294.PubMedCrossRef
45.
Zurück zum Zitat Morlock M, Schneider E, Bluhm A, Vollmer M, Bergmann G, Muller V, Honl M. Duration and frequency of everyday activities in total hip patients. J Biomech. 2001;34:873–881.PubMedCrossRef Morlock M, Schneider E, Bluhm A, Vollmer M, Bergmann G, Muller V, Honl M. Duration and frequency of everyday activities in total hip patients. J Biomech. 2001;34:873–881.PubMedCrossRef
46.
Zurück zum Zitat Murray D. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.PubMed Murray D. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.PubMed
47.
Zurück zum Zitat Nadzadi ME, Pedersen DR, Yack HJ, Callaghan JJ, Brown TD. Kinematics, kinetics, and finite element analysis of commonplace maneuvers at risk for total hip dislocation. J Biomech. 2003;36:577–591.PubMedCrossRef Nadzadi ME, Pedersen DR, Yack HJ, Callaghan JJ, Brown TD. Kinematics, kinetics, and finite element analysis of commonplace maneuvers at risk for total hip dislocation. J Biomech. 2003;36:577–591.PubMedCrossRef
48.
Zurück zum Zitat Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.PubMedCrossRef Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.PubMedCrossRef
49.
Zurück zum Zitat Peters CL, McPherson E, Jackson JD, Erickson JA. Reduction in early dislocation rate with large-diameter femoral heads in primary total hip arthroplasty. J Arthroplasty. 2007;22:140–144.PubMedCrossRef Peters CL, McPherson E, Jackson JD, Erickson JA. Reduction in early dislocation rate with large-diameter femoral heads in primary total hip arthroplasty. J Arthroplasty. 2007;22:140–144.PubMedCrossRef
50.
Zurück zum Zitat Sanders AP, Brannon RM. Assessment of the applicability of the Hertzian contact theory to edge-loaded prosthetic hip bearings. J Biomech. 2011;44:2802–2808.PubMedCentralPubMedCrossRef Sanders AP, Brannon RM. Assessment of the applicability of the Hertzian contact theory to edge-loaded prosthetic hip bearings. J Biomech. 2011;44:2802–2808.PubMedCentralPubMedCrossRef
51.
Zurück zum Zitat Scifert CF, Brown TD, Pedersen DR, Callaghan JJ. A finite element analysis of factors influencing total hip dislocation. Clin Orthop Relat Res. 1998;355:152–162.PubMedCrossRef Scifert CF, Brown TD, Pedersen DR, Callaghan JJ. A finite element analysis of factors influencing total hip dislocation. Clin Orthop Relat Res. 1998;355:152–162.PubMedCrossRef
52.
Zurück zum Zitat Silva M, Heisel C, Schmalzried TP. Metal-on-metal total hip replacement. Clin Orthop Relat Res. 2005;430:53–61.PubMedCrossRef Silva M, Heisel C, Schmalzried TP. Metal-on-metal total hip replacement. Clin Orthop Relat Res. 2005;430:53–61.PubMedCrossRef
53.
Zurück zum Zitat Stuchin SA. Anatomic diameter femoral heads in total hip arthroplasty: a preliminary report. J Bone Joint Surg Am. 2008;90(suppl 3):52–56.PubMedCrossRef Stuchin SA. Anatomic diameter femoral heads in total hip arthroplasty: a preliminary report. J Bone Joint Surg Am. 2008;90(suppl 3):52–56.PubMedCrossRef
54.
Zurück zum Zitat Svensson O, Mathiesen EB, Reinholt F, Blomgren G. Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty: a case report. J Bone Joint Surg Am. 1988;70:1238–1242.PubMed Svensson O, Mathiesen EB, Reinholt F, Blomgren G. Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty: a case report. J Bone Joint Surg Am. 1988;70:1238–1242.PubMed
55.
Zurück zum Zitat Walsh AJ, Nikolaou VS, Antoniou J. Inflammatory pseudotumor complicating metal-on-highly cross-linked polyethylene total hip arthroplasty. J Arthroplasty. 2012;27:324.e5–324.e8. Walsh AJ, Nikolaou VS, Antoniou J. Inflammatory pseudotumor complicating metal-on-highly cross-linked polyethylene total hip arthroplasty. J Arthroplasty. 2012;27:324.e5–324.e8.
56.
Zurück zum Zitat Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints: a clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.PubMedCrossRef Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints: a clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.PubMedCrossRef
57.
Zurück zum Zitat Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J. Tribology and wear of metal-on-metal hip prostheses: influence of cup angle and head position. J Bone Joint Surg Am. 2008;90:111–117.PubMedCrossRef Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J. Tribology and wear of metal-on-metal hip prostheses: influence of cup angle and head position. J Bone Joint Surg Am. 2008;90:111–117.PubMedCrossRef
58.
Zurück zum Zitat Zywiel MG, Sayeed SA, Johnson AJ, Schmalzried TP, Mont MA. Survival of hard-on-hard bearings in total hip arthroplasty. Clin Orthop Relat Res. 2011;469:1536–1546.PubMedCrossRef Zywiel MG, Sayeed SA, Johnson AJ, Schmalzried TP, Mont MA. Survival of hard-on-hard bearings in total hip arthroplasty. Clin Orthop Relat Res. 2011;469:1536–1546.PubMedCrossRef
Metadaten
Titel
Stability and Trunnion Wear Potential in Large-diameter Metal-on-Metal Total Hips: A Finite Element Analysis
verfasst von
Jacob M. Elkins, MD, PhD
John J. Callaghan, MD
Thomas D. Brown, PhD
Publikationsdatum
01.02.2014
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 2/2014
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-013-3244-8

Weitere Artikel der Ausgabe 2/2014

Clinical Orthopaedics and Related Research® 2/2014 Zur Ausgabe

Symposium: 2013 Hip Society Proceedings

Editorial Comment: Symposium: 2013 Hip Society Proceedings

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.