Skip to main content
Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine 5/2022

Open Access 18.04.2022 | Research Article

T2* assessment of the three coronary artery territories of the left ventricular wall by different monoexponential truncation methods

verfasst von: Pandji Triadyaksa, Jelle Overbosch, Matthijs Oudkerk, Paul Eduard Sijens

Erschienen in: Magnetic Resonance Materials in Physics, Biology and Medicine | Ausgabe 5/2022

Abstract

Objectives

This study aimed at evaluating left ventricular myocardial pixel-wise T2* using two truncation methods for different iron deposition T2* ranges and comparison of segmental T2* in different coronary artery territories.

Material and methods

Bright blood multi-gradient echo data of 30 patients were quantified by pixel-wise monoexponential T2* fitting with its R2 and SNR truncation. T2* was analyzed at different iron classifications. At low iron classification, T2* values were also analyzed by coronary artery territories.

Results

The right coronary artery has a significantly higher T2* value than the other coronary artery territories. No significant difference was found in classifying severe iron by the two truncation methods in any myocardial region, whereas in moderate iron, it is only apparent at septal segments. The R2 truncation produces a significantly higher T2* value than the SNR method when low iron is indicated.

Conclusion

Clear T2* differentiation between the three coronary territories by the two truncation methods is demonstrated. The two truncation methods can be used interchangeably in classifying severe and moderate iron deposition at the recommended septal region. However, in patients with low iron indication, different results by the two truncation methods can mislead the investigation of early iron level progression.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Cardiac magnetic resonance imaging (MRI) by multi-gradient echo (MGE) has become a widespread method for the non-invasive assessment of myocardial iron deposition [13]. Myocardial iron deposition assessment is essential for treatment decisions in patients with thalassemia, hemochromatosis, cardiomyopathy, and sickle cell disease [36]. The deposition quantification is conducted by applying T2* evaluation on bright blood mode MGE image series [1, 2, 4, 713] at mid-ventricular septal [2, 10, 14] or global myocardium [1, 4, 8]. Several post-mortem studies have verified the relation between cardiac iron deposition with T2* value [7, 9, 11, 15, 16]. On its evaluation, iron distribution is found at different circumferential myocardial regions [4, 11, 17]. Therefore, the segmental approach is suggested to be more sensitive in detecting the early progression of iron distribution than the evaluation of the septum only.
The established myocardial T2* is assessed by either pixel-wise quantification [7, 11, 12, 18, 19] or per region of interest [1, 2, 4, 810]. In their methods, pixel-wise quantification has the advantage of identifying the heterogeneity of iron distribution [7, 12] but comes with the drawback of it being prone to susceptibility artifact [13, 19] as located in the MGE image series in Fig. 1. Therefore, in pixel-wise T2* quantification, a study suggests using the median rather than mean T2* values to characterize myocardial segments [12].
In the bright blood mode, signal plateau or offset can occur at longer TEs. This offset is due to noise bias, heterogeneous iron distribution in myocytes, signal offset, myocardial blood volume, motion artifact, or iron-poor tissue contributions [20]. Therefore, one might suggest adding an offset constant C to the classic monoexponential equation [20]. However, in the presence of artifacts, an increase by a constant value will consequently reduce the initial signal intensity estimation of the monoexponential equation and, therefore, lower the T2* estimation. This estimation leads to a biased interpretation of iron deposition, especially at low iron concentrations, as reported in model and patient data [15, 21, 22].
Another way to handle the signal plateau is by truncating higher TE data when the signal intensity reaches the noise level [15, 18], known as signal to noise ratio (SNR) truncation. Evaluating the goodness of the fit (R2), known as the R2 truncation method, can also be a criterion to exclude data points at longer TEs [10, 12]. Comparisons between either truncation method with an offset method were conducted previously [15, 22, 23]. Nevertheless, no comparison was conducted between truncation methods raising the question of which method best replaces the classic monoexponential method in quantifying myocardial T2* from clinical MGE data.
It is known that different coronary arteries supply different circumferential regions of the left ventricle (LV) [24], as shown in the American Heart Association (AHA) 16-segment model, reproduced in Fig. 2a [25]. Detailed analysis of the artery supply, reproduced in Fig. 2b, shows that most myocardial segments are served by more than one coronary artery [26]. Therefore, this study aimed to compare the SNR and R2 truncation methods in evaluating the pixel-wise T2* value for LV myocardial iron detection and analyze and compare the T2* values of the different coronary artery territories in the low iron deposition category, especially in the absence of major iron deposition before its progression.

Materials and methods

Patients

Short-axis images of 40 patients were acquired using a clinical routine MGE cardiac MRI protocol between February 2009 and May 2016. The data were analyzed retrospectively, and the hospital review board waived the informed consent requirement. Motion correction was not performed, but images from ten patients were excluded due to respiratory motion blurring. The remainder of 30 included patients consisted of six patients with thalassemia (age average of 15 years with four male), six with hemochromatosis (age average of 55 years with two male), thirteen with suspected cardiomyopathy (age average of 46 years with eight male), and five patients with sickle cell disease (16 year, male), congenital dyserythropoietic anemia (25 year, male), chemotherapy (14 year), blood transfusion (19 year, male), or congenital heart disease (38 year, male).

Cardiac iron assessment by magnetic resonance imaging

A bright blood MGE sequence with a single breath-hold was used at 8 TEs with a repetition time of 200 ms and a flip angle of 20° on either one of two 1.5 T MRI systems (Siemens Medical Solutions, Erlangen, Germany) without enabling parallel imaging. From 2009 till 2011, the Avanto scanner (TEs of 2.59–18.20 at 2.23 ms increments, pixel bandwidth set at 814 Hz) was used on 18 patients. From 2012 till 2016, the Aera scanner (TEs of 2.69–18.86 at 2.31 ms increments, pixel bandwidth set at 815 Hz) was used on the subsequent 12 included patients. Depending on the field of view of 275–362 × 400 mm2 that reflects the patient size, a body matrix coil of 6–9 elements and a spine matrix coil of 12–24 elements were used. The MGE sequence used a number of excitation (NEX) of 1 and a reconstructed voxel size of 1.56 × 1.56 × 10 mm3. Phase resolution sampling of 50% was applied using 18–24 cardiac cycles per breath-hold with five segments in each heartbeat. Eighteen patients were scanned at the apical, mid-ventricular, and basal short-axis slices, while the rest were only scanned at the mid-ventricular slice. Four patients undergoing multiple follow-ups were acquired four complete short-axis slices at the three standard locations and three at mid-ventricular. Therefore, a total of 22 apical, 37 mid-ventricular, and 22 basal short-axis slices were acquired (Fig. 3, top).
Custom-written software (developed in MATLAB version 7.14, The MathWorks, Natick, MA, USA) was used to apply pixel-wise myocardial T2* quantification on all segments by a classic monoexponential fitting [27] as follows:
$$y = Ke^{{ - {\text{TE/T}}2^{*} }} ,$$
(1)
where y, K, TE, and T2* represent the signal intensity, a fitting constant, echo time of the MGE image series, and myocardium transverse relaxation time. A default non-linear least square curve fitting (MATLAB lsqcurvefit function) with trust-region-reflective was used to perform the fitting with a minimum K, and T2* constraint values were kept to be 0.
Two monoexponential truncation methods were adopted in this study. The SNR truncation method used pixel's SNR as a longer TE data point exclusion criterion with the SNR defined by the following equation [28]:
$$\mathrm{SNR}=\mathrm{NF}\frac{SI}{{\sigma }_{b}},$$
(2)
where NF, SI, σb represent the noise factor, pixel-wise signal intensity, and the standard deviation of a region of interest of air background, respectively. The noise factor accounts for underestimating noise derived from complex-magnitude data, assessed using the root of the sum of square (SoS) reconstruction and equals 0.71 for up to 32 coil elements according to Table 2 in the previous study [29].
Using the custom-written software, the SNR truncation method automatically excluded any pixel-wise TE data point with SNR below 2.5 [18] starts from longer TE data points of monoexponential analysis with maintaining TE below 10 ms or, in this study, four early TE points. Until TE 10 ms, the noise's influence is minimal and gives similar signal intensity for different NEX up to 32 (21) while maintaining enough information of the tissue's fast relaxation component [30, 31]. The R2 truncation used pixel's fitting R2 as the longer TE data point exclusion criterion. The exclusion will automatically exclude the point until the monoexponential fitting R2 exceeds a threshold value of 0.995 [10, 12] or, when not achieved, until four early TE points.

Analysis and statistical methods

A cardiac radiologist with more than 10 years of experience performing semi-automatic LV epicardial and endocardial contours segmentation. The segmentation is conducted using custom-written software developed in MATLAB version 7.14; The MathWorks, Natick, MA, USA adapted from another study [32]. The AHA 16-segment model was used to segment the short-axis slices [27]. This model was used to analyze iron deposition in the myocardium as a whole (16 segments), at the suggested septum (i.e., apical septal, mid-ventricular anteroseptal, mid-ventricular inferoseptal, basal anteroseptal, and basal inferoseptal), and at inferolateral, that is prone to susceptibility artifact [13] (i.e., mid-ventricular inferolateral, and basal inferolateral).
Invasive iron deposition measurement was not conducted in this study. Therefore, iron deposition pathology was indicated by normal, moderate iron deposition, or severe iron deposition based on medians of pixel-wise T2* in the AHA segments. T2* classification for normal pathology (i.e., without iron deposition) was T2* > 20 ms, while for moderate iron deposition was 10 ms ≤ T2* ≤ 20 ms and for severe iron deposition was T2* < 10 ms [27, 33].
A flow chart of data assessment is shown in Fig. 3. The classic monoexponential fitting method (Eq. 1) was used to calculate the median pixel-wise T2* and its median absolute deviation (MAD) [27, 34] on each AHA segment (Fig. 3, step 2). The same T2* calculation was also conducted by the two alternative T2* monoexponential truncation methods. Then the segments were grouped according to the iron deposition classification (Fig. 3, step 3) at three different locations, i.e., global myocardium, septum, and inferolateral. Multiple median segmental pixel-wise T2* comparisons were made between the three methods for different iron deposition classification and different segment locations (Fig. 3, step 4). The pixel-wise T2* from the two truncation evaluation was grouped further to understand our observations on each iron deposition classification. Each group consists of pixel-wise T2* evaluated using different TE points from a minimum of 4 TE until 8 TE. (Fig. 3, step 5). Also, pixel-wise T2* analysis by the two truncation methods was conducted at different coronary artery territories, i.e., left anterior descending (LAD), right coronary artery (RCA), and left circumflex artery (LCx) territories [26] for patients without global iron deposition.
IBM SPSS Statistics software version 23 (IBM Corporation, Somer, NY, USA) was used for statistical testing. Multiple comparisons by one-way analysis of variance with Bonferroni post hoc test was used for normally distributed data. Kruskal–Wallis test analysis with Dun-Bonferroni post hoc test was used for non-normally distributed data. Shapiro–Wilk test was performed to test the normality of data distribution. P < 0.05 was considered statistically significant.

Results

According to the classic monoexponential fitting method, from a total of 442 segments, 360 segments of 29 patients were classified as without iron deposition, 64 segments of 21 patients with moderate iron, and 18 segments of three patients with severe iron deposition, as shown in Fig. 3 and grouped in Table 1. From these segments, in the septal region, 124 segments of 29 patients were classified as without iron deposition, 12 segments of seven patients with moderate iron, and four segments of one patient with severe iron deposition. Meanwhile, in the inferolateral region, 43 segments of 25 patients were classified as without iron deposition, 13 segments of 10 patients with moderate iron, and three segments of two patients with severe iron deposition.
Table 1
Comparison of segmental pixel-wise T2* for three pixel-wise monoexponential fitting methods differentiated into categories of iron deposition
T2* quantification
ns
Monoexponential
R2 truncation
SNR truncation
T2* (ms)
T2* (ms)
T2* (ms)
AHA 16 segments having
 T2* > 20 ms
360
31.96 ± 6.23
36.57 ± 6.95*
31.96 ± 6.22
 10 ms ≤ T2* ≤ 20 ms
64
17.63 ± 2.15
22.47 ± 4.34*
17.74 ± 2.13
 T2* < 10 ms
18
8.53 ± 1.46
8.22 ± 2.10
8.08 ± 2.02
Septal segments having
 T2* > 20 ms
124
36.52 ± 5.93
40.07 ± 5.89*
36.52 ± 5.93
 10 ms ≤ T2* ≤ 20 ms
12
16.99 ± 2.65
22.34 ± 4.68
16.99 ± 2.65
 T2* < 10 ms
4
6.97 ± 0.85
6.16 ± 0.26
6.21 ± 0.24
Inferolateral segments having
 T2* > 20 ms
43
25.73 ± 4.42
32.14 ± 5.78*
25.74 ± 4.41
 10 ms ≤ T2* ≤ 20 ms
13
19.14 ± 1.21
26.14 ± 3.59*
19.52 ± 0.56
 T2* < 10 ms
3
8.78 ± 0.38
9.54 ± 0.94
10.73 ± 0.52
Data are presented as median ± median absolute deviation. ns number of segments according to T2* classification by monoexponential fitting method, SNR signal-to-noise ratio, AHA American Heart Association. Both truncation methods were performed with a minimum of four echo times
*Dun-Bonferroni post hoc test P < 0.05 compared to the T2* classic monoexponential method
Dun-Bonferroni post hoc test P < 0.05 compared to the T2* monoexponential with R2 truncation method
Of 22 patients with data allowing for a complete AHA 16-segment model analysis, 20 had a global monoexponential T2* without iron deposition. One patient with congenital dyserythropoietic anemia had severe iron deposition in 14 segments, with only two segments classified as moderate. Another patient with thalassemia had a majority of moderate iron deposition segments with three severe iron deposition segments and only one segment at the borderline without iron classification. After a one-year follow-up, the patient had five segments classified as moderate and eleven segments without iron classification. Visual inspection showed a minor region of susceptibility artifact at the inferolateral region on 53 out of 59 collected segments.
Evaluated by the three monoexponential fitting methods on 20 patients without global iron deposition, the RCA territory had a higher T2* value than the LAD and the LCx territories (P < 0.01), as shown in Fig. 4. However, when using R2 truncation, no significant difference was found between the RCA and the LAD (P > 0.05). No significant difference was found between the LCx and the LAD by the three fittings (P > 0.05).
Histogram analysis on each territory in Fig. 4 showed multiple histogram peaks indicating multiple T2* components. Therefore, further analysis in Fig. 5a, c revealed T2* heterogeneity, using the two truncation methods, on AHA segments in each coronary artery territory. In Fig. 5b, d, similar T2* heterogeneity between the three coronary arteries was shown when comparing segments supported by a dominant coronary artery. A significant difference of T2* value was found on each territory between dominant coronary artery segments compared to segments supported by more arteries (P < 0.05).
Compared to the classic monoexponential and the SNR truncation method, the R2 truncation method consistently produced higher segmental T2* values in the absence of iron deposition at all investigated regions (P < 0.01), as shown in Table 1 and Fig. 6. Tables 2 and 3 express this significant difference between the two truncation methods on all AHA 16 segments and at recommended septal and at prone to artifact, inferolateral locations.The pixel-wise R2 goodness of fitting maps, presented in Fig. 7, show the two truncation methods' relatively high fitting performance on different myocardial regions. However, at the susceptibility artifact region, the R2 fitting of the two methods tends to reduce. On its majority pixels, the R2 truncation method used four early TE points to quantify T2*, i.e., 87.83%, 83.97%, and 93.99% of total pixels at the global myocardium, septal, and inferolateral regions, respectively. Meanwhile, the SNR truncation method used all TE points for the T2* quantification at the exact locations, i.e., 98.16%, 99.81%, and 92.81% of total pixels.
Table 2
Pixel-wise T2* quantification by the two monoexponential truncation methods evaluated by using different TE on all AHA 16 segments
 
T2* (ms)
Using 8 TE
Using 7 TE
Using 6 TE
Using 5 TE
Using 4 TE
T2* > 20 ms¥
R2 truncation
32.99 ± 6.72 (5.42%)
31.96 ± 6.84 (1.84%)
32.89 ± 7.13 (2.13%)
33.30 ± 7.21 (2.78%)
37.71 ± 10.88 (87.83%)
(Total 100% pixels: 36,596)
SNR truncation
32.45 ± 8.37 (98.16%)
14.82 ± 2.82 (0.87%)
13.09 ± 2.81 (0.50%)
10.07 ± 2.35 (0.21%)
11.80 ± 4.95 (0.26%)
10 ms < T2* ≤ 20 ms¥
R2 truncation
14.60 ± 3.09 (6.03%)
16.72 ± 5.50 (1.80%)
20.14 ± 7.09 (2.42%)
21.80 ± 6.78 (3.20%)
23.24 ± 6.24 (86.56%)
(Total 100% pixels: 5787)
SNR truncation
18.28 ± 3.80 (88.92%)
12.55 ± 1.64 (4.56%)
12.92 ± 1.80 (2.78%)
10.84 ± 2.44 (2.07%)
7.82 ± 2.38 (1.66%)
T2* ≤ 10 ms¥
R2 truncation
7.67 ± 2.00 (9.25%)
6.47 ± 1.44 (2.89%)
7.03 ± 1.87 (3.78%)
6.13 ± 1.80 (5.04%)
8.40 ± 2.97 (79.03%)
(Total 100% pixels: 1903)
SNR truncation
14.45 ± 4.00 (19.92%)
11.39 ± 2.02 (7.41%)
11.67 ± 1.91 (7.83%)
8.89 ± 1.44 (11.25%)
5.97 ± 1.40 (53.60%)
Data are presented as median ± median absolute deviation. Data in parentheses are the percentage of quantified pixels from its total in each iron deposition classification
TE echo time, SNR signal to noise ratio
¥Pixels selection was based on T2* classification by the classic monoexponential fitting method on Table 1
Table 3
Pixel-wise T2* quantification by the two monoexponential truncation methods evaluated by using different TE at recommended septal and at prone to artifact, inferolateral locations
 
T2* (ms)
Using 8 TE
Using 7 TE
Using 6 TE
Using 5 TE
Using 4 TE
Pixel-wise T2* at septal segments ¥
 T2* > 20 ms
R2 truncation
33.92 ± 6.67 (7.71%)
33.37 ± 7.49 (2.25%)
34.12 ± 6.95 (2.64%)
34.76 ± 7.45 (3.43%)
41.02 ± 11.15 (83.97%)
 (Total 100% pixels: 14,027)
SNR truncation
35.73 ± 9.05 (99.81%)
14.13 ± 1.28 (0.15%)
11.53 ± 1.23 (0.02%)
13.06 ± 0.94 (0.01%)
47.91 (1 pixel)
 10 ms < T2* ≤ 20 ms
R2 truncation
13.38 ± 2.81 (13.01%)
15.92 ± 4.23 (3.41%)
18.79 ± 7.04 (4.80%)
21.54 ± 5.92 (3.58%)
24.71 ± 5.35 (75.20%)
 (Total 100% pixels: 1145)
SNR truncation
17.79 ± 3.74 (96.94%)
10.80 ± 1.65 (1.83%)
10.16 ± 0.41 (0.52%)
6.82 ± 0.30 (0.52%)
5.75 ± 0.14 (0.17%)
 T2* ≤ 10 ms
R2 truncation
6.62 ± 1.20 (21.92%)
6.19 ± 1.66 (6.58%)
5.82 ± 1.54 (7.95%)
5.75 ± 0.95 (9.32%)
5.80 ± 1.18 (54.25%)
 (Total 100% pixels: 365)
SNR truncation
14.34 ± 2.09 (16.16%)
11.09 ± 0.91 (4.66%)
8.12 ± 0.77 (3.56%)
7.23 ± 0.93 (11.23%)
5.29 ± 0.71 (64.38%)
Pixel-wise T2* at inferolateral segments ¥
 T2* > 20 ms
R2 truncation
28.57 ± 4.30 (1.74%)
28.52 ± 6.51 (1.10%)
23.51 ± 4.21 (1.49%)
30.19 ± 8.70 (1.69%)
32.80 ± 11.64 (93.99%)
 (Total 100% pixels: 4090)
SNR truncation
27.83 ± 7.84 (92.81%)
14.22 ± 2.40 (2.96%)
13.21 ± 2.83 (2.54%)
9.02 ± 1.89 (0.93%)
10.42 ± 2.88 (0.76%)
 10 ms < T2* ≤ 20 ms
R2 truncation
12.83 ± 4.36 (2.10%)
16.84 ± 5.75 (1.05%)
20.73 ± 7.11 (0.79%)
18.68 ± 6.24 (1.84%)
23.09 ± 7.92 (94.22%)
 (Total 100% pixels: 1142)
SNR truncation
20.05 ± 4.93 (81.79%)
13.24 ± 2.02 (7.27%)
13.80 ± 2.10 (4.20%)
13.03 ± 2.67 (2.80%)
7.57 ± 2.42 (3.94%)
 T2* ≤ 10 ms
R2 truncation
7.00 ± 0.44 (0.94%)
9.24 ± 1.33 (0.94%)
7.49 ± 0.01 (0.47%)
8.84 ± 1.36 (1.88%)
9.80 ± 3.01 (95.76%)
 Total 100% pixels: 425)
SNR truncation
15.72 ± 5.21 (7.53%)
15.58 ± 3.49 (8.24%)
13.63 ± 2.41 (14.82%)
10.34 ± 2.03 (17.65%)
7.95 ± 1.97 (51.76%)
Data are presented as median ± median absolute deviation. Data in parentheses are the percentage of quantified pixels from its total in each iron deposition classification
TE echo time, SNR signal to noise ratio
¥Pixels selection was based on T2* classification by the classic monoexponential fitting method on Table 1
In moderate iron deposition classification at the global myocardium region, the T2* values from the R2 truncation method were significantly higher than the T2* resulted from the other two fitting methods (P < 0.001). Table 2 shows that the R2 truncation method used four TE points in the majority of 86.56% pixels to quantify T2* while the SNR truncation method used all TE points in 88.92% pixels. Meanwhile, at septal and inferolateral regions, even though a different use of TE points in the majority of pixels remains (Table 3), the T2* values produced by the two truncation methods are not significantly different (P > 0.05). Segmental T2* quantification by the three monoexponential fitting methods was not significantly different in severe iron deposition segments at any investigated region (P > 0.05). Table 2 shows that both truncation methods used four early TE values in a majority of pixels for quantifying T2*.
In all iron deposition classification and at different analyzed locations, the elimination of later TE points increased the T2* values when quantified by the R2 truncation, as shown in Fig. 8, Tables 2, and 3. However, when the SNR truncation is used, the elimination of later TE points will decrease the quantified T2*.

Discussion

This study demonstrated T2* heterogeneity in different circumferential regions according to coronary artery territories. T2* differences were also apparent when the analysis focused on the coronary arteries' contribution in supplying blood on segments. A consistent T2* heterogeneity by the two truncation methods was also found in the territories analysis based on its contribution on segments.
T2* variation occurred from different pixel-wise monoexponential curve fitting truncation methods. These apparent discrepancies apply either to the myocardium (AHA 16-segment model) or at the septal and inferolateral regions.. Therefore, the difference can influence clinical decision-making, especially in the myocardial region indicated with moderate iron deposition and without iron deposition, for the early treatment of patients with thalassemia, hemochromatosis, cardiomyopathy, and sickle cell disease.
It is known that some center prefers mid-ventricular septal to detect iron loading [2, 10, 14], while others choose to detect its progression at the global myocardium [1, 4, 8] with performing correction maps [35, 36]. Early detection of T2* is beneficial to understand its relation with iron distribution mechanisms on different diseases [3742]. Meanwhile, both truncation methods produce similar results in identifying a severe iron loading region. Therefore, it is essential to understand the causes of the measurement discrepancies as a base for selecting the most appropriate analysis method to identify iron loading in a wide range of conditions.
The nature of R2 truncation likelihood in eliminating later TE points shows that the method is sensitive to pixel-wise signal intensity fluctuation to produce optimum R2 fitting. It is known that the source of the signal's fluctuation can be from partial volume voxel, cardiac motion, flow artifact, or fat and water phase oscillation [20]. Therefore, implementing the fitting's R2 until its maximum threshold, in this study 0.995, will push the TE point's elimination to meet the required R2 setting and leads to bias T2* measurement for iron deposition determination. An example is a possible missing chance of early iron deposition identification, as shown in the comparison between Fig. 6c, d at the anterior, anteroseptal, and anterolateral regions. A comparison between Fig. 8c, d, at the anterior region, show that the signal drops at early TE tend to be shallow at a higher T2* range [9, 33]. At the myocardial areas with higher T2*, this condition might be influenced by the in-phase and opposed phase fat signal that makes the TE signal fluctuate [43]. Therefore, excluding later TE reduces the constant K value of Eq. 1 and produces a higher T2* value. Some study has suggested avoiding TE points exclusion when classic monoexponential evaluation shows normal T2* range when performing R2 truncation [44].
TE point elimination's main idea was to eliminate the influence of signal plateau at longer TE due to noise and artifact [15, 22]. Therefore, the truncation method's implementation should reflect this idea by performing elimination only when the source of the signal plateau becomes apparent and not due to signal fluctuation. Moreover, eliminating TE points until their minimum early TEs does not always produce better T2* R2 fitting, as shown in the comparison between Fig. 8g, h. Analysis in Fig. 7g, h shows the reduction of the pixel-wise R2 fitting of the R2 truncation method in the presence of susceptibility artifact. TE points elimination can lead to a high T2* value, as shown in Fig. 6g, h at the inferolateral qwregion.
The SNR truncation eliminates later TE points when its signal's SNR is below a certain threshold. This procedure will dominantly eliminate TE points in the presence of signal plateau at longer TE. This truncation means that the method will selectively eliminate TE points when the signal's SNR is below the susceptibility artifact regions' threshold. Several studies have reported that TE truncation, at 1.5 T was mainly apparent after 10 ms [9, 10, 22, 31, 33, 45, 46]. Besides, aliasing effects due to rapid signal decay and limited TE data points also mainly occurred at TE after 10 ms [20]. Table 3 shows that in its application, in the susceptibility artifact region, i.e., inferolateral, the number of evaluated pixel-wise T2* by using fewer TE points is higher than the recommended artifact-free region, i.e., septal. This TE points’ selection showed the SNR truncation method's ability to detect the presence of possible artifacts.
At severe iron classification, even though truncated until 4 points, the similarity of T2* between the two truncated methods with the classic monoexponential is acquired due to the capture of fast relaxation component at early TE as shown in Fig. 8i–l. The bright-blood SNR truncation method has been reported to yield T2* values closest to those obtained by black-blood mode at severe iron identification [22]. It is known that the black-blood mode has the advantage of reducing the influence of artifacts and partial volume effect by suppressing the blood signal to produce clear endocardial and epicardial borders.
It is important to realize that T2* heterogeneity on circumferential regions can be explained due to the different contributions of coronary artery territories on each segment [26]. The previous study also mentioned that segments 1, 7, 13 were exclusively supported by LAD and are shown by the lower T2* threshold in Fig. 5 [26]. Further investigation in Fig. 5 showed that higher T2* at segments 2, 8, and 14 in the LAD territory is relevant to the RCA's higher contribution on the territory. At segments 2 and 8, higher T2* values might be influenced by segments 3 and 9, partly supported by the LAD and RCA, as shown in Fig. 2b. The T2* heterogeneity can be explained due to coronary collateral circulation, which provides alternative blood supply pathways in segment's supported by multiple coronary artery territories [4749].
In the RCA territory, it seems that other territories also support segments 4 and 10. Therefore its T2* was lower compared to segments 3 and 9. Broader T2* range at segment 15 was explained due to the support of all territories at this segment. Meanwhile, in the LCx territory, segment 5 has lower T2* variability due to its exclusive support only by the LCx, as shown in Fig. 2b [26]. The rest LCx segments were supported by LAD and are explained by no significant difference between LAD and LCx territories in this study. It is known that the RCA has a longer artery length than the LAD and LCx, with a higher pressure gradient received from other arteries. Therefore, it receives higher collateral supplies than LAD and LCx, as shown with higher T2* in Figs. 4 and 5 [47].
Similar T2* heterogeneity by the two truncation methods on coronary artery territories and circumferential regions reflects a different iron deposition progression on different locations. This clear finding is observed when the analysis used the dominant coronary artery segments as suggested in Fig. 2b [26] rather than the traditional territories in Fig. 2a. The analysis of the coronary artery contribution on the circumferential regions might also be used to explain the heterogeneity of T1 value across circumferential regions observed by several studies [34, 50, 51]. Several post mortem and prospective studies [1, 11, 17] had reported a lower T2* value at the anterior, followed by inferior and lateral. Therefore, early detection of iron deposition is advised to start from these locations. The suggestion is also supported by the coronary artery analysis showing that LCx and LAD arteries had a lower T2* than RCA. It is known that LCx and LAD arteries were responsible for supplying blood to the most anterior and lateral segments while RCA supplies blood to inferior and septal segments [26]. Higher T2* value on RCA compared to other arteries indicates a lower progression of early iron deposition at the septal region, as detected previously [1, 4, 52].
When later TE points elimination is conducted to avoid any influence of signal plateau due to artifacts, the SNR truncation method is preferable to achieve the goal, as demonstrated in this study. However, it is essential to consider that the accuracy of the T2* values obtained with different fitting methods does not directly reflect tissue iron status [22]. A difference in T2* value, as a parameter to early identify iron deposition, by the two truncation methods is evident in this study. For that, comparing the two truncation methods with a different iron detection method [11, 17] in ex-vivo subjects would be advised at the higher end range of T2* values. Moreover, even though benefits from is lower variability on intraobserver and interobserver agreement, the validation of the two methods in higher magnetic field strengths is also needed due to its significantly higher susceptibility artifact at different locations [13, 5355].

Limitations

Here, the comparison of pixel-wise fitting methods and the influence of noise and artifact in T2* quantification of the myocardium was limited to the bright blood mode. Even though a low NEX was performed in this study, an attempt to increase NEX might reduce noise but at the expense of increasing scanning time [15, 44]. This study was conducted without enabling parallel imaging. Therefore the image SNR measurement can be straightforward on the MGE images. When parallel imaging is applied, a different approach in measuring the image's SNR is needed [18] and is beyond this study's scope.
Furthermore, rather than correcting for artifact contributions in the MGE data analysis, one might prefer to optimize the MGE imaging method to reduce and eliminate artifacts before the curve fitting (beyond the scope of this study). Another limitation is that this study used different scanners and slightly different TEs for acquiring the MGE series, relying on previous studies confirming the MRI sequence's reproducibility [2, 10]. The use of minimum TE close to 1 ms will increase the ability to capture the signal intensity fast relaxation component before the susceptibility artifact at mostly 10 ms [20, 46].
A limited number of patients each year who follow the MGE MRI clinical routine in our center also holds the strength of the analysis. The validation of the T2* measurement with the patient’s total body iron or other cardiac MR evaluations [5658] is recommended to determine iron status and understand the progression of the disease. However, it is beyond the scope of this study.

Conclusion

In conclusion, this study shows that focusing the coronary artery territories analysis based on the artery contribution on the segments produces a consistent T2* heterogeneity pattern by the two truncation methods. Moreover, the indication of severe and moderate iron deposition at the recommended septal region can be classified interchangeably by the two T2* truncation methods. Nevertheless, T2* differences remain in the region with no iron indication suggesting further investigation to avoid misleading early iron progression investigated by the two methods.

Acknowledgements

Pandji Triadyaksa receives scholarship from the Directorate General of Higher Education of Republic Indonesia (617/E4.4/K/2011) to conduct this study

Declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and later amendments or comparable ethical standards.
For this type of study, formal consent is not required.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Meloni A, Positano V, Ruffo GB, Spasiano A, D’Ascola DG, Peluso A, Keilberg P, Restaino G, Valeri G, Renne S, Midiri M, Pepe A (2015) Improvement of heart iron with preserved patterns of iron store by CMR-guided chelation therapy. Eur Heart J Cardiovasc Imaging 16:325–334PubMed Meloni A, Positano V, Ruffo GB, Spasiano A, D’Ascola DG, Peluso A, Keilberg P, Restaino G, Valeri G, Renne S, Midiri M, Pepe A (2015) Improvement of heart iron with preserved patterns of iron store by CMR-guided chelation therapy. Eur Heart J Cardiovasc Imaging 16:325–334PubMed
2.
Zurück zum Zitat Kirk P, He T, Anderson LJ, Roughton M, Tanner MA, Lam WWM, Au WY, Chu WCW, Chan G, Galanello R, Matta G, Fogel M, Cohen AR, Tan RS, Chen K, Ng I, Lai A, Fucharoen S, Laothamata J, Chuncharunee S, Jongjirasiri S, Firmin DN, Smith GC, Pennell DJ (2010) International reproducibility of single breathhold T2* MR for cardiac and liver iron assessment among five thalassemia centers. J Magn Reson Imaging 32:315–319PubMedPubMedCentral Kirk P, He T, Anderson LJ, Roughton M, Tanner MA, Lam WWM, Au WY, Chu WCW, Chan G, Galanello R, Matta G, Fogel M, Cohen AR, Tan RS, Chen K, Ng I, Lai A, Fucharoen S, Laothamata J, Chuncharunee S, Jongjirasiri S, Firmin DN, Smith GC, Pennell DJ (2010) International reproducibility of single breathhold T2* MR for cardiac and liver iron assessment among five thalassemia centers. J Magn Reson Imaging 32:315–319PubMedPubMedCentral
3.
Zurück zum Zitat Assis RA, Kay FU, Conti FM, Campregher PV, Szarf G, Diniz MS, Rodrigues M, Helman R, Funari MBG, Wood J, Hamerschlak N (2015) The role of magnetic resonance imaging-T2* in the evaluation of iron overload early in hereditary hemochromatosis. A cross-sectional study with 159 patients. Am J Hematol 90:E220–E221PubMed Assis RA, Kay FU, Conti FM, Campregher PV, Szarf G, Diniz MS, Rodrigues M, Helman R, Funari MBG, Wood J, Hamerschlak N (2015) The role of magnetic resonance imaging-T2* in the evaluation of iron overload early in hereditary hemochromatosis. A cross-sectional study with 159 patients. Am J Hematol 90:E220–E221PubMed
4.
Zurück zum Zitat Meloni A, Restaino G, Borsellino Z, Caruso V, Spasiano A, Zuccarelli A, Valeri G, Toia P, Salvatori C, Positano V, Midiri M, Pepe A (2014) Different patterns of myocardial iron distribution by whole-heart T2* magnetic resonance as risk markers for heart complications in thalassemia major. Int J Cardiol 177:1012–1019PubMed Meloni A, Restaino G, Borsellino Z, Caruso V, Spasiano A, Zuccarelli A, Valeri G, Toia P, Salvatori C, Positano V, Midiri M, Pepe A (2014) Different patterns of myocardial iron distribution by whole-heart T2* magnetic resonance as risk markers for heart complications in thalassemia major. Int J Cardiol 177:1012–1019PubMed
5.
Zurück zum Zitat Meloni A, Puliyel M, Pepe A, Berdoukas V, Coates TD, Wood JC (2014) Cardiac iron overload in sickle-cell disease. Am J Hematol 89:678–683PubMed Meloni A, Puliyel M, Pepe A, Berdoukas V, Coates TD, Wood JC (2014) Cardiac iron overload in sickle-cell disease. Am J Hematol 89:678–683PubMed
6.
Zurück zum Zitat Kremastinos DT, Farmakis D (2011) Iron overload cardiomyopathy in clinical practice. Circulation 124:2253–2263PubMed Kremastinos DT, Farmakis D (2011) Iron overload cardiomyopathy in clinical practice. Circulation 124:2253–2263PubMed
7.
Zurück zum Zitat Ghugre NR, Enriquez CM, Gonzalez I, Nelson MD, Coates TD, Wood JC (2006) MRI detects myocardial iron in the human heart. Magn Reson Med 56:681–686PubMedPubMedCentral Ghugre NR, Enriquez CM, Gonzalez I, Nelson MD, Coates TD, Wood JC (2006) MRI detects myocardial iron in the human heart. Magn Reson Med 56:681–686PubMedPubMedCentral
8.
Zurück zum Zitat Meloni A, Positano V, Pepe A, Rossi G, Dell’Amico M, Salvatori C, Keilberg P, Filosa A, Sallustio G, Midiri M, D’Ascola D, Santarelli MF, Lombardi M (2010) Preferential patterns of myocardial iron overload by multislice multiecho T2* CMR in thalassemia major patients. Magn Reson Med 64:211–219PubMed Meloni A, Positano V, Pepe A, Rossi G, Dell’Amico M, Salvatori C, Keilberg P, Filosa A, Sallustio G, Midiri M, D’Ascola D, Santarelli MF, Lombardi M (2010) Preferential patterns of myocardial iron overload by multislice multiecho T2* CMR in thalassemia major patients. Magn Reson Med 64:211–219PubMed
9.
Zurück zum Zitat Carpenter JP, He T, Kirk P, Roughton M, Anderson LJ, De Noronha SV, Sheppard MN, Porter JB, Walker JM, Wood JC, Galanello R, Forni G, Catani G, Matta G, Fucharoen S, Fleming A, House MJ, Black G, Firmin DN, Pierre St TG, Pennell DJ (2011) On T2* magnetic resonance and cardiac iron. Circulation 123:1519–1528PubMedPubMedCentral Carpenter JP, He T, Kirk P, Roughton M, Anderson LJ, De Noronha SV, Sheppard MN, Porter JB, Walker JM, Wood JC, Galanello R, Forni G, Catani G, Matta G, Fucharoen S, Fleming A, House MJ, Black G, Firmin DN, Pierre St TG, Pennell DJ (2011) On T2* magnetic resonance and cardiac iron. Circulation 123:1519–1528PubMedPubMedCentral
10.
Zurück zum Zitat He T, Zhang J, Carpenter J-P, Feng Y, Smith GC, Pennell DJ, Firmin DN (2013) Automated truncation method for myocardial T2* measurement in thalassemia. J Magn Reson imaging 37:479–483PubMed He T, Zhang J, Carpenter J-P, Feng Y, Smith GC, Pennell DJ, Firmin DN (2013) Automated truncation method for myocardial T2* measurement in thalassemia. J Magn Reson imaging 37:479–483PubMed
11.
Zurück zum Zitat House MJ, Fleming AJ, de Jonge MD, Paterson D, Howard DL, Carpenter J-P, Pennell DJ, St Pierre TG (2014) Mapping iron in human heart tissue with synchrotron x-ray fluorescence microscopy and cardiovascular magnetic resonance. J Cardiovasc Magn Reson 16:80PubMedPubMedCentral House MJ, Fleming AJ, de Jonge MD, Paterson D, Howard DL, Carpenter J-P, Pennell DJ, St Pierre TG (2014) Mapping iron in human heart tissue with synchrotron x-ray fluorescence microscopy and cardiovascular magnetic resonance. J Cardiovasc Magn Reson 16:80PubMedPubMedCentral
12.
Zurück zum Zitat Positano V, Meloni A, Santarelli MF, Gerardi C, Bitti PP, Cirotto C, De Marchi D, Salvatori C, Landini L, Pepe A (2015) Fast generation of T2* maps in the entire range of clinical interest: application to thalassemia major patients. Comput Biol Med 56:200–210PubMed Positano V, Meloni A, Santarelli MF, Gerardi C, Bitti PP, Cirotto C, De Marchi D, Salvatori C, Landini L, Pepe A (2015) Fast generation of T2* maps in the entire range of clinical interest: application to thalassemia major patients. Comput Biol Med 56:200–210PubMed
13.
Zurück zum Zitat Meloni A, Hezel F, Positano V, Keilberg P, Pepe A, Lombardi M, Niendorf T (2014) Detailing magnetic field strength dependence and segmental artifact distribution of myocardial effective transverse relaxation rate at 1.5, 3.0, and 7.0 T. Magn Reson Med 71:2224–2230PubMed Meloni A, Hezel F, Positano V, Keilberg P, Pepe A, Lombardi M, Niendorf T (2014) Detailing magnetic field strength dependence and segmental artifact distribution of myocardial effective transverse relaxation rate at 1.5, 3.0, and 7.0 T. Magn Reson Med 71:2224–2230PubMed
14.
Zurück zum Zitat Feng Y, He T, Carpenter J-P, Jabbour A, Alam MH, Gatehouse PD, Greiser A, Messroghli D, Firmin DN, Pennell DJ (2013) In vivo comparison of myocardial T1 with T2 and T2* in thalassaemia major. J Magn Reson Imaging 38:588–593PubMed Feng Y, He T, Carpenter J-P, Jabbour A, Alam MH, Gatehouse PD, Greiser A, Messroghli D, Firmin DN, Pennell DJ (2013) In vivo comparison of myocardial T1 with T2 and T2* in thalassaemia major. J Magn Reson Imaging 38:588–593PubMed
15.
Zurück zum Zitat He T, Gatehouse PD, Kirk P, Mohiaddin RH, Pennell DJ, Firmin DN (2008) Myocardial T2* measurement in iron-overloaded thalassemia: an ex vivo study to investigate optimal methods of quantification. Magn Reson Med 60:350–356PubMed He T, Gatehouse PD, Kirk P, Mohiaddin RH, Pennell DJ, Firmin DN (2008) Myocardial T2* measurement in iron-overloaded thalassemia: an ex vivo study to investigate optimal methods of quantification. Magn Reson Med 60:350–356PubMed
16.
Zurück zum Zitat Kirk P, Sheppard M, Carpenter J-P, Anderson L, He T, St Pierre T, Galanello R, Catani G, Wood J, Fucharoen S, Porter JB, Walker JM, Forni GL, Pennell DJ (2017) Post-mortem study of the association between cardiac iron and fibrosis in transfusion dependent anaemia. J Cardiovasc Magn Reson 19:36PubMedPubMedCentral Kirk P, Sheppard M, Carpenter J-P, Anderson L, He T, St Pierre T, Galanello R, Catani G, Wood J, Fucharoen S, Porter JB, Walker JM, Forni GL, Pennell DJ (2017) Post-mortem study of the association between cardiac iron and fibrosis in transfusion dependent anaemia. J Cardiovasc Magn Reson 19:36PubMedPubMedCentral
17.
Zurück zum Zitat Meloni A, Maggio A, Positano V, Leto F, Angelini A, Putti MC, Maresi E, Pucci A, Basso C, Marra MP, Pistoia L, Marchi De D, Pepe A (2020) CMR for myocardial iron overload quantification: calibration curve from the MIOT Network. Eur Radiol 30:3217–3225PubMed Meloni A, Maggio A, Positano V, Leto F, Angelini A, Putti MC, Maresi E, Pucci A, Basso C, Marra MP, Pistoia L, Marchi De D, Pepe A (2020) CMR for myocardial iron overload quantification: calibration curve from the MIOT Network. Eur Radiol 30:3217–3225PubMed
18.
Zurück zum Zitat Sandino CM, Kellman P, Arai AE, Hansen MS, Xue H (2015) Myocardial T2* mapping: influence of noise on accuracy and precision. J Cardiovasc Magn Reson 17:7PubMedPubMedCentral Sandino CM, Kellman P, Arai AE, Hansen MS, Xue H (2015) Myocardial T2* mapping: influence of noise on accuracy and precision. J Cardiovasc Magn Reson 17:7PubMedPubMedCentral
19.
Zurück zum Zitat Taylor BA, Loeffler RB, Song R, McCarville ME, Hankins JS, Hillenbrand CM (2013) Automated T2 * measurements using supplementary field mapping to assess cardiac iron content. J Magn Reson Imaging 38:441–447PubMedPubMedCentral Taylor BA, Loeffler RB, Song R, McCarville ME, Hankins JS, Hillenbrand CM (2013) Automated T2 * measurements using supplementary field mapping to assess cardiac iron content. J Magn Reson Imaging 38:441–447PubMedPubMedCentral
20.
Zurück zum Zitat Ghugre NR, Enriquez CM, Coates TD, Nelson MD, Wood JC (2006) Improved R2* measurements in myocardial iron overload. J Magn Reson Imaging 23:9–16PubMedPubMedCentral Ghugre NR, Enriquez CM, Coates TD, Nelson MD, Wood JC (2006) Improved R2* measurements in myocardial iron overload. J Magn Reson Imaging 23:9–16PubMedPubMedCentral
21.
Zurück zum Zitat Feng Y, He T, Gatehouse PD, Li X, Harith Alam M, Pennell DJ, Chen W, Firmin DN (2013) Improved MRI R2* relaxometry of iron-loaded liver with noise correction. Magn Reson Med 70:1765–1774PubMed Feng Y, He T, Gatehouse PD, Li X, Harith Alam M, Pennell DJ, Chen W, Firmin DN (2013) Improved MRI R2* relaxometry of iron-loaded liver with noise correction. Magn Reson Med 70:1765–1774PubMed
22.
Zurück zum Zitat He T, Gatehouse PD, Smith GC, Mohiaddin RH, Pennell DJ, Firmin DN (2008) Myocardial T2* measurements in iron-overloaded thalassemia: an in vivo study to investigate optimal methods of quantification. Magn Reson Med 60:1082–1089PubMedPubMedCentral He T, Gatehouse PD, Smith GC, Mohiaddin RH, Pennell DJ, Firmin DN (2008) Myocardial T2* measurements in iron-overloaded thalassemia: an in vivo study to investigate optimal methods of quantification. Magn Reson Med 60:1082–1089PubMedPubMedCentral
23.
Zurück zum Zitat Tanner MA, He T, Westwood MA, Firmin DN, Pennell DJ (2006) Multi-center validation of the transferability of the magnetic resonance T2* technique for the quantification of tissue iron. Haematologica 91:1388–1391PubMed Tanner MA, He T, Westwood MA, Firmin DN, Pennell DJ (2006) Multi-center validation of the transferability of the magnetic resonance T2* technique for the quantification of tissue iron. Haematologica 91:1388–1391PubMed
24.
Zurück zum Zitat Partridge JB, Anderson RH (2009) Left ventricular anatomy: its nomenclature, segmentation, and planes of imaging. Clin Anat 22:77–84PubMed Partridge JB, Anderson RH (2009) Left ventricular anatomy: its nomenclature, segmentation, and planes of imaging. Clin Anat 22:77–84PubMed
25.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS (2002) Standardized myocardial sementation and nomenclature for tomographic imaging of the heart: a Statement for Healthcare Professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 105:539–542PubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, Pennell DJ, Rumberger JA, Ryan T, Verani MS (2002) Standardized myocardial sementation and nomenclature for tomographic imaging of the heart: a Statement for Healthcare Professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Circulation 105:539–542PubMed
26.
Zurück zum Zitat Donato P, Coelho P, Santos C, Bernardes A, Caseiro-Alves F (2012) Correspondence between left ventricular 17 myocardial segments and coronary anatomy obtained by multi-detector computed tomography: an ex vivo contribution. Surg Radiol Anat 34:805–810PubMed Donato P, Coelho P, Santos C, Bernardes A, Caseiro-Alves F (2012) Correspondence between left ventricular 17 myocardial segments and coronary anatomy obtained by multi-detector computed tomography: an ex vivo contribution. Surg Radiol Anat 34:805–810PubMed
27.
Zurück zum Zitat Triadyaksa P, Oudkerk M, Sijens PE (2020) Cardiac T 2 * mapping: techniques and clinical applications. J Magn Reson Imaging 52:1340–1351PubMed Triadyaksa P, Oudkerk M, Sijens PE (2020) Cardiac T 2 * mapping: techniques and clinical applications. J Magn Reson Imaging 52:1340–1351PubMed
28.
Zurück zum Zitat Triadyaksa P, Handayani A, Dijkstra H, Aryanto KYE, Pelgrim GJ, Xie X, Willems TP, Prakken NHJ, Oudkerk M, Sijens PE (2016) Contrast-optimized composite image derived from multigradient echo cardiac magnetic resonance imaging improves reproducibility of myocardial contours and T2* measurement. Magn Reson Mater Phy 29:17–27 Triadyaksa P, Handayani A, Dijkstra H, Aryanto KYE, Pelgrim GJ, Xie X, Willems TP, Prakken NHJ, Oudkerk M, Sijens PE (2016) Contrast-optimized composite image derived from multigradient echo cardiac magnetic resonance imaging improves reproducibility of myocardial contours and T2* measurement. Magn Reson Mater Phy 29:17–27
29.
Zurück zum Zitat Dietrich O, Raya JG, Reeder SB, Ingrisch M, Reiser MF, Schoenberg SO (2008) Influence of multichannel combination, parallel imaging and other reconstruction techniques on MRI noise characteristics. Magn Reson Imaging 26:754–762PubMed Dietrich O, Raya JG, Reeder SB, Ingrisch M, Reiser MF, Schoenberg SO (2008) Influence of multichannel combination, parallel imaging and other reconstruction techniques on MRI noise characteristics. Magn Reson Imaging 26:754–762PubMed
30.
Zurück zum Zitat Clark PR, Chua-anusorn W, St. Pierre TG (2003) Bi-exponential proton transverse relaxation rate (R2) image analysis using RF field intensity-weighted spin density projection: potential for R2 measurement of iron-loaded liver. Magn Reson Imaging 21:519–530PubMed Clark PR, Chua-anusorn W, St. Pierre TG (2003) Bi-exponential proton transverse relaxation rate (R2) image analysis using RF field intensity-weighted spin density projection: potential for R2 measurement of iron-loaded liver. Magn Reson Imaging 21:519–530PubMed
31.
Zurück zum Zitat Menacho K, Abdel-Gadir A, Moon JC, Fernandes JL (2019) T2* mapping techniques: iron overload assessment and other potential clinical applications. Magn Reson Imaging Clin N Am 27:439–451PubMed Menacho K, Abdel-Gadir A, Moon JC, Fernandes JL (2019) T2* mapping techniques: iron overload assessment and other potential clinical applications. Magn Reson Imaging Clin N Am 27:439–451PubMed
32.
Zurück zum Zitat Triadyaksa P, Prakken NHJ, Overbosch J, Peters RB, van Swieten JM, Oudkerk M, Sijens PE (2017) Semi-automated myocardial segmentation of bright blood multi-gradient echo images improves reproducibility of myocardial contours and T2* determination. Magn Reson Mater Phy 30:239–254 Triadyaksa P, Prakken NHJ, Overbosch J, Peters RB, van Swieten JM, Oudkerk M, Sijens PE (2017) Semi-automated myocardial segmentation of bright blood multi-gradient echo images improves reproducibility of myocardial contours and T2* determination. Magn Reson Mater Phy 30:239–254
33.
Zurück zum Zitat Fernandes JL (2018) MRI for iron overload in Thalassemia. Hematol Oncol Clin North Am 32:277–295PubMed Fernandes JL (2018) MRI for iron overload in Thalassemia. Hematol Oncol Clin North Am 32:277–295PubMed
34.
Zurück zum Zitat Triadyaksa P, Kuijpers D, Akinci D’Antonoli T, Overbosch J, Rook M, van Swieten JM, Oudkerk M, Sijens PE (2020) Early detection of heart function abnormality by native T1: a comparison of two T1 quantification methods. Eur Radiol 30:652–662PubMed Triadyaksa P, Kuijpers D, Akinci D’Antonoli T, Overbosch J, Rook M, van Swieten JM, Oudkerk M, Sijens PE (2020) Early detection of heart function abnormality by native T1: a comparison of two T1 quantification methods. Eur Radiol 30:652–662PubMed
35.
Zurück zum Zitat Positano V, Pepe A, Santarelli MF, Ramazzotti A, Meloni A, De Marchi D, Favilli B, Cracolici E, Midiri M, Spasiano A, Lombardi M, Landini L (2009) Multislice multiecho T2* cardiac magnetic resonance for the detection of heterogeneous myocardial iron distribution in thalassaemia patients. NMR Biomed 22:707–715PubMed Positano V, Pepe A, Santarelli MF, Ramazzotti A, Meloni A, De Marchi D, Favilli B, Cracolici E, Midiri M, Spasiano A, Lombardi M, Landini L (2009) Multislice multiecho T2* cardiac magnetic resonance for the detection of heterogeneous myocardial iron distribution in thalassaemia patients. NMR Biomed 22:707–715PubMed
36.
Zurück zum Zitat Positano V, Pepe A, Santarelli MF, Scattini B, De Marchi D, Ramazzotti A, Forni G, Borgna-Pignatti C, Lai ME, Midiri M, Maggio A, Lombardi M, Landini L (2007) Standardized T2* map of normal human heart in vivo to correct T2* segmental artefacts. NMR Biomed 20:578–590PubMed Positano V, Pepe A, Santarelli MF, Scattini B, De Marchi D, Ramazzotti A, Forni G, Borgna-Pignatti C, Lai ME, Midiri M, Maggio A, Lombardi M, Landini L (2007) Standardized T2* map of normal human heart in vivo to correct T2* segmental artefacts. NMR Biomed 20:578–590PubMed
38.
Zurück zum Zitat De Valk B, Marx JJM (1999) Iron, atherosclerosis, and ischemic heart disease. Arch Intern Med 159:1542–1548PubMed De Valk B, Marx JJM (1999) Iron, atherosclerosis, and ischemic heart disease. Arch Intern Med 159:1542–1548PubMed
39.
Zurück zum Zitat Cohen AR, Galanello R, Pennell DJ, Cunningham MJ, Vichinsky E (2004) Thalassemia. Hematology 2004:14–34 Cohen AR, Galanello R, Pennell DJ, Cunningham MJ, Vichinsky E (2004) Thalassemia. Hematology 2004:14–34
40.
Zurück zum Zitat Wood JC, Enriquez C, Ghugre N, Otto-Duessel M, Aguilar M, Nelson MD, Moats R, Coates TD (2005) Physiology and pathophysiology of iron cardiomyopathy in Thalassemia. Ann N Y Acad Sci 1054:386–395PubMedPubMedCentral Wood JC, Enriquez C, Ghugre N, Otto-Duessel M, Aguilar M, Nelson MD, Moats R, Coates TD (2005) Physiology and pathophysiology of iron cardiomyopathy in Thalassemia. Ann N Y Acad Sci 1054:386–395PubMedPubMedCentral
41.
Zurück zum Zitat Kohgo Y, Ikuta K, Ohtake T, Torimoto Y, Kato J (2008) Body iron metabolism and pathophysiology of iron overload. Int J Hematol 88:7–15PubMedPubMedCentral Kohgo Y, Ikuta K, Ohtake T, Torimoto Y, Kato J (2008) Body iron metabolism and pathophysiology of iron overload. Int J Hematol 88:7–15PubMedPubMedCentral
42.
Zurück zum Zitat Murphy CJ, Oudit GY (2010) Iron-overload cardiomyopathy: pathophysiology, diagnosis, and treatment. J Cardiac Fail 16:888–900 Murphy CJ, Oudit GY (2010) Iron-overload cardiomyopathy: pathophysiology, diagnosis, and treatment. J Cardiac Fail 16:888–900
43.
Zurück zum Zitat Reeder SB, Sirlin CB (2010) Quantification of liver fat with magnetic resonance imaging. Magn Reson Imaging Clin N Am 18:337–357PubMedPubMedCentral Reeder SB, Sirlin CB (2010) Quantification of liver fat with magnetic resonance imaging. Magn Reson Imaging Clin N Am 18:337–357PubMedPubMedCentral
44.
Zurück zum Zitat Hedström E, Voigt T, Greil G, Schaeffter T, Nagel E (2017) Automatic T2* determination for quantification of iron load in heart and liver: a comparison between automatic inline Maximum Likelihood Estimate and the truncation and offset methods. Clin Physiol Funct Imaging 37:299–304PubMed Hedström E, Voigt T, Greil G, Schaeffter T, Nagel E (2017) Automatic T2* determination for quantification of iron load in heart and liver: a comparison between automatic inline Maximum Likelihood Estimate and the truncation and offset methods. Clin Physiol Funct Imaging 37:299–304PubMed
45.
Zurück zum Zitat Reeder SB, Faranesh AZ, Boxerman JL, McVeigh ER (1998) In vivo measurement of T2* and field inhomogeneity maps in the human heart at 1.5 T. Magn Reson Med 39:988–998PubMedPubMedCentral Reeder SB, Faranesh AZ, Boxerman JL, McVeigh ER (1998) In vivo measurement of T2* and field inhomogeneity maps in the human heart at 1.5 T. Magn Reson Med 39:988–998PubMedPubMedCentral
46.
Zurück zum Zitat Beaumont M, Odame I, Babyn PS, Vidarsson L, Kirby-Allen M, Cheng HLM (2009) Accurate liver T*2 measurement of iron overload: a simulations investigation and in vivo study. J Magn Reson Imaging 30:313–320PubMed Beaumont M, Odame I, Babyn PS, Vidarsson L, Kirby-Allen M, Cheng HLM (2009) Accurate liver T*2 measurement of iron overload: a simulations investigation and in vivo study. J Magn Reson Imaging 30:313–320PubMed
48.
Zurück zum Zitat Koerselman J, Van der Graaf Y, De Jaegere PPT, Grobbee DE (2003) Coronary collaterals: an important and underexposed aspect of coronary artery disease. Circulation 107:2507–2511PubMed Koerselman J, Van der Graaf Y, De Jaegere PPT, Grobbee DE (2003) Coronary collaterals: an important and underexposed aspect of coronary artery disease. Circulation 107:2507–2511PubMed
49.
Zurück zum Zitat Faber JE, Chilian WM, Deindl E, Van Royen N, Simons M (2014) A brief etymology of the collateral circulation. Arterioscler Thromb Vasc Biol 34:1854–1859PubMedPubMedCentral Faber JE, Chilian WM, Deindl E, Van Royen N, Simons M (2014) A brief etymology of the collateral circulation. Arterioscler Thromb Vasc Biol 34:1854–1859PubMedPubMedCentral
50.
Zurück zum Zitat Piechnik SK, Ferreira VM, DallArmellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD (2010) Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 15 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson 12:69PubMedPubMedCentral Piechnik SK, Ferreira VM, DallArmellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD (2010) Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 15 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson 12:69PubMedPubMedCentral
51.
Zurück zum Zitat Reiter U, Reiter G, Dorr K, Greiser A, Maderthaner R, Fuchsjäger M (2014) Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology 271:365–372PubMed Reiter U, Reiter G, Dorr K, Greiser A, Maderthaner R, Fuchsjäger M (2014) Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology 271:365–372PubMed
52.
Zurück zum Zitat Borgna-Pignatti C, Meloni A, Guerrini G, Gulino L, Filosa A, Ruffo GB, Casini T, Chiodi E, Lombardi M, Pepe A (2014) Myocardial iron overload in thalassaemia major. How early to check? Br J Haematol 164:579–585PubMed Borgna-Pignatti C, Meloni A, Guerrini G, Gulino L, Filosa A, Ruffo GB, Casini T, Chiodi E, Lombardi M, Pepe A (2014) Myocardial iron overload in thalassaemia major. How early to check? Br J Haematol 164:579–585PubMed
53.
Zurück zum Zitat Meloni A, Positano V, Keilberg P, De Marchi D, Pepe P, Zuccarelli A, Campisi S, Romeo MA, Casini T, Bitti PP, Gerardi C, Lai ME, Piraino B, Giuffrida G, Secchi G, Midiri M, Lombardi M, Pepe A (2012) Feasibility, reproducibility, and reliability for the T*2 iron evaluation at 3 T in comparison with 1.5 T. Magn Reson Med 68:543–551PubMed Meloni A, Positano V, Keilberg P, De Marchi D, Pepe P, Zuccarelli A, Campisi S, Romeo MA, Casini T, Bitti PP, Gerardi C, Lai ME, Piraino B, Giuffrida G, Secchi G, Midiri M, Lombardi M, Pepe A (2012) Feasibility, reproducibility, and reliability for the T*2 iron evaluation at 3 T in comparison with 1.5 T. Magn Reson Med 68:543–551PubMed
54.
Zurück zum Zitat Moon BF, Iyer SK, Josselyn NJ, Hwuang E, Swago S, Keeney SJ, Castillero E, Ferrari G, Pilla JJ, Gorman JH, Gorman RC, Tschabrunn CM, Shou H, Matthai W, Wehrli FW, Ferrari VA, Han Y, Litt H, Witschey WR (2022) Magnetic susceptibility and R2* of myocardial reperfusion injury at 3T and 7T. Magn Reson Med 87:323–336PubMed Moon BF, Iyer SK, Josselyn NJ, Hwuang E, Swago S, Keeney SJ, Castillero E, Ferrari G, Pilla JJ, Gorman JH, Gorman RC, Tschabrunn CM, Shou H, Matthai W, Wehrli FW, Ferrari VA, Han Y, Litt H, Witschey WR (2022) Magnetic susceptibility and R2* of myocardial reperfusion injury at 3T and 7T. Magn Reson Med 87:323–336PubMed
55.
Zurück zum Zitat Storey P, Thompson AA, Carqueville CL, Wood JC, de Freitas RA, Rigsby CK (2007) R2* imaging of transfusional iron burden at 3T and comparison with 1.5T. J Magn Reson Imaging 25:540–547PubMedPubMedCentral Storey P, Thompson AA, Carqueville CL, Wood JC, de Freitas RA, Rigsby CK (2007) R2* imaging of transfusional iron burden at 3T and comparison with 1.5T. J Magn Reson Imaging 25:540–547PubMedPubMedCentral
56.
Zurück zum Zitat Torlasco C, Cassinerio E, Roghi A, Faini A, Capecchi M, Abdel-Gadir A, Giannattasio C, Parati G, Moon JC, Cappellini MD, Pedrotti P (2018) Role of T1 mapping as a complementary tool to T2* for non-invasive cardiac iron overload assessment. PLoS ONE 13:e0192890PubMedPubMedCentral Torlasco C, Cassinerio E, Roghi A, Faini A, Capecchi M, Abdel-Gadir A, Giannattasio C, Parati G, Moon JC, Cappellini MD, Pedrotti P (2018) Role of T1 mapping as a complementary tool to T2* for non-invasive cardiac iron overload assessment. PLoS ONE 13:e0192890PubMedPubMedCentral
57.
Zurück zum Zitat Sado DM, Maestrini V, Piechnik SK, Banypersad SM, White SK, Flett AS, Robson MD, Neubauer S, Ariti C, Arai A, Kellman P, Yamamura J, Schoennagel BP, Shah F, Davis B, Trompeter S, Walker M, Porter J, Moon JC (2015) Noncontrast myocardial T1 mapping using cardiovascular magnetic resonance for iron overload. J Magn Reson Imaging 41:1505–1511PubMed Sado DM, Maestrini V, Piechnik SK, Banypersad SM, White SK, Flett AS, Robson MD, Neubauer S, Ariti C, Arai A, Kellman P, Yamamura J, Schoennagel BP, Shah F, Davis B, Trompeter S, Walker M, Porter J, Moon JC (2015) Noncontrast myocardial T1 mapping using cardiovascular magnetic resonance for iron overload. J Magn Reson Imaging 41:1505–1511PubMed
58.
Zurück zum Zitat Krittayaphong R, Zhang S, Saiviroonporn P, Viprakasit V, Tanapibunpon P, Komoltri C, Wangworatrakul W (2017) Detection of cardiac iron overload with native magnetic resonance T1 and T2 mapping in patients with thalassemia. Int J Cardiol 248:421–426PubMed Krittayaphong R, Zhang S, Saiviroonporn P, Viprakasit V, Tanapibunpon P, Komoltri C, Wangworatrakul W (2017) Detection of cardiac iron overload with native magnetic resonance T1 and T2 mapping in patients with thalassemia. Int J Cardiol 248:421–426PubMed
Metadaten
Titel
T2* assessment of the three coronary artery territories of the left ventricular wall by different monoexponential truncation methods
verfasst von
Pandji Triadyaksa
Jelle Overbosch
Matthijs Oudkerk
Paul Eduard Sijens
Publikationsdatum
18.04.2022
Verlag
Springer International Publishing
Erschienen in
Magnetic Resonance Materials in Physics, Biology and Medicine / Ausgabe 5/2022
Print ISSN: 0968-5243
Elektronische ISSN: 1352-8661
DOI
https://doi.org/10.1007/s10334-022-01008-4

Weitere Artikel der Ausgabe 5/2022

Magnetic Resonance Materials in Physics, Biology and Medicine 5/2022 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.