Temperature rise during removal of fractured components out of the implant body: an in vitro study comparing two ultrasonic devices and five implant types

verfasst von: Eric W Meisberger, Sjoerd J G Bakker, Marco S Cune

Erschienen in: International Journal of Implant Dentistry | Ausgabe 1/2015

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Ultrasonic instrumentation under magnification may facilitate mobilization of screw remnants but may induce heat trauma to surrounding bone. An increase of 5°C is considered detrimental to osseointegration. The objective of this investigation was to examine the rise in temperature of the outer implant body after 30 s of ultrasonic instrumentation to the inner part, in relation to implant type, type of ultrasonic equipment, and the use of coolants in vitro.

Methods

Two ultrasonic devices (Satelec Suprasson T Max and Electro Medical Systems (EMS) miniMaster) were used on five different implant types that were provided with a thermo couple (Astra 3.5 mm, bone level Regular CrossFit (RC) 4.1 mm, bone level Narrow CrossFit (NC) 3.3 mm, Straumann tissue level regular body regular neck 3.3 mm, and Straumann tissue level wide body regular neck 4.8 mm), either with or without cooling during 30 s. Temperature rise at this point in time is the primary outcome measure. In addition, the mean maximum rise in temperature (all implants combined) was assessed and statistically compared among devices, implant systems, and cooling mode (independent t-tests, ANOVA, and post hoc analysis).

Results

The Satelec device without cooling induces the highest temperature change of up to 13°C, particularly in both bone level implants (p < 0.05) but appears safe for approximately 10 s of continuous instrumentation, after which a cooling down period is rational. Cooling is effective for both devices. However, when the Satelec device is used with coolant for a longer period of time, a rise in temperature must be anticipated after cessation of instrumentation, and post-operational cooling is advised.

Conclusions

The in vitro setup used in this experiment implies that care should be taken when translating the observations to clinical recommendations, but it is carefully suggested that the EMS device causes limited rise in temperature, even without coolant.
Hinweise

Competing interests

This research was made possible by the support of the authors’ institutions. Eric W Meisberger, Sjoerd JG Bakker and Marco S Cune declare that they have no competing interests.

Authors’ contributions

EM and SB carried out the experiment. MC participated in the design of the study and performed the statistical analysis and helped to draft the manuscript. All authors read and approved the final manuscript.

Background

Complications in implant dentistry are generally divided into biological and mechanical complications. Mechanical complications include fracture of the implant body or prosthetic components such as chipping of ceramic material as well as loosening or fracture of implant abutments or fixation screws. This has a documented prevalence of 6 to 13% and 0.4 to 2% respectively after 5 years [1-7].
With respect to the etiology of screw loosening, several causes are to be considered [8]. Screw-nut systems generally become unstable when the load that is applied to the system exceeds that of the preload of the screw that causes a clamping force preventing separation of the joint [9]. Preload is proportional to the tightening torque at placement. Hence, tightening torque on the one hand has a significant effect on screw loosening [10]. Embedment relaxation or settling can be overcome by retorquing abutment screws after a certain period of time, increasing joint stability [9,11]. On the other hand, the magnitude of forces applied to the system is of major influence. The transfer of high forces can be generated by bruxers, through non-occlusal loading or because of a non-passive fit of suprastructures [8,12-15]. Other factors that are of influence of a systems’ resistance to screw loosening include lubrication, screw design, screw material, and surface characteristics [16-20]. Especially implant designs with an external hex configuration are prone to abutment screw loosening [21].
Fracture of abutments or abutment screws can be contributed to acute trauma, chronic overload, production flaws, or errors in screw-nut design. If fractured screw components cannot be removed, it may render the implant unrestorable or forces the dentist to creative solutions, such as cementable components. Fortunately, fractured screw components will generally be loose because the preload has not been retained. In that case, they may be removed by manipulating them counter clockwise with a straight probing instrument. On occasion, screw remnants cannot be mobilized, and removal remains a clinical challenge. Careful instrumentation when attempting to remove them should prevent damage to the internal thread of the implant and its surrounding tissues. The use of a fine-tipped hand instrument, a round burr turning counter clockwise, or drilling a slot in the screw in order to get more grip can be attempted to loosen the fragment. Specific instrumentation to remove broken screws from implants is available from most implant suppliers.
The use of ultrasonic equipment under adequate magnification may facilitate removal. It generates heat. Instrumentation without a coolant likely increases the temperature of the implant body and could cause tissue damage, in particular be harmful to osseointegration. The use of a coolant could be effective, but compromises visibility considerably, hence increases the risk of damaging the internal configuration of the implant.
Results from an experimental heat conduction model investigating the ranges of temperature gradients occurring in implants demonstrate that a 60°C heat source causes a heat front exceeding 47°C and advances more than 3 mm down an implant within 1 s [22]. Temperatures over 47°C for more than a minute cause necrosis of cortical bone [23-25]. Some studies investigating the potential harmfulness of intraoral abutment preparation or plaster on implants to osseointegration mention this threshold as harmful [26-28]. It has been postulated that a rise in temperature to 42°C causes denaturation of osteoblasts and should be considered the temperature threshold of transient changes in bone [25]. This threshold was used by others when investigating the potentially damaging effect to the implant-bone interface as a result of drinking hot beverages [29-31]. Also from endodontic literature regarding the removal of metal endodontic posts, concerns have been raised based on observations from in vitro experiments with respect to potentially detrimental heat transformation through dentine while ultrasonically manipulating the post [32-37].
The objective of this investigation is to examine the rise in temperature of the outer surface of an implant body after 30 s of ultrasonic instrumentation of its inner part in relation to type of ultrasonic equipment, implant type, and the use of coolants in vitro.

Methods

Two different types of commercially available ultrasonic devices, set at their lowest intensity for endodontic purpose, were used to instrument the internal portion of five different implant types, either with or without cooling. Intermittent anti-clockwise strokes were made, assuring that the tip was constantly in contact with the inner implant wall, as much as possible mimicking the motion that would have been used in clinical practice.
The ultrasonic devices used were the Satelec Suprasson T Max (Acteon Group, Merignac, France) and the EMS miniMaster (EMS, Electro Medical Systems SA, Nyon, Switzerland) with non-diamant, non-cutting tips ET 20, Satelec, and Instrument A (EMS). The device allowed for internal cooling of the tip with the cooling liquid at 31°C during instrumentation.
The implants used were from different brands; all 8-mm long but with various diameters and designs: Astra 3.5 mm (Dentsply Implants, Mölndal, Sweden), bone level Regular CrossFit (RC) 4.1 mm, bone level Narrow CrossFit (NC) 3.3 mm, Straumann tissue level regular body regular neck 3.3 mm (Straumann, Basel, Switzerland), and Straumann tissue level wide body regular neck 4.8 mm (Straumann AG, Basel, Switzerland). A single implant per group was used. They were embedded in epoxy resin, with a thermocouple (TC-08, Pico Technology, St. Neots, Great Britain) glued to the outer implant surface, at a level corresponding with the anticipated marginal bone level in uncompromised conditions. The change in temperature was registered for 30 s, followed by a 30-s cooling down period, at 5-s intervals (Figure 1).

Statistical analysis

The primary outcome variable was defined as the difference in temperature between the start of instrumentation and after 30 s when comparing the different implants and the maximum rise in temperature (deltaTmax) where results were averaged per experimental condition (type of device, with or without coolant). All tests were performed three times, and the results were averaged per condition. Differences between several experimental conditions were analyzed by means of independent t-tests and univariate analysis of variance, after verification of normal distribution by human eyeballing and the Kolmogorov-Smirnov test. Where appropriate, post hoc analysis was performed using the Student-Newman-Keuls multiple comparison test. The value for α was set at 0.05 to distinguish statistical significancy.

Results

The results for all implants instrumented with the two tested ultrasonic devices, either with or without cooling, are presented in Figure 2a,b,c,d.
For the Satelec device, applied without coolant, only the temperature of the Straumann wide body regular neck implant never exceeded the 50 threshold. The data proved normally distributed (Kolmogorov-Smirnov test, p > 0.05). Analysis of variance indicated statistically significant differences among the implant types (F = 33.3, df 4, p < 0.001). The highest mean temperature increase at 30 s was seen around the 3.3-mm bone level implant (13.0°C), followed by the 4.1-mm bone level implant (9.5°C), and subsequently by the other three implant types (Student-Newman-Keuls (SNK) test, Figure 2a). When coolant was used during instrumentation, an increase of temperature exceeding 5°C was not seen. There were some differences between the implant types (ANOVA, F = 6.4, df 4, p < 0.01), primarily between the bone level 3.3 mm and Astra 3.5-mm implant on the one hand and both Straumann tissue level implants on the other hand (SNK test). Interestingly, when instrumentation and cooling had stopped, the outer temperature of all implants raised markedly above the 5°C critical threshold during the following 10 s, for both bone level implant types up to approximately 10°C (Figure 2b).
For the EMS device when applied without cooling, an increase at 30 s above the threshold was only seen for the 3.3- and 4.1-mm bone level implant types (approximately 6°C), which was statistically significantly higher than the rise in temperature seen in the other three implant types (ANOVA, F = 3.4, df = 4, p = 0.04, and SNK test, Figure 2c). Cooling the EMS device proved pretty efficient at 30 s, without differences between the groups. Only a mild increase of the outer temperature was observed to a maximum of 2°C. It takes some time for cooling to take effect. When instrumentation and cooling is stopped, no increase of outer implant temperature occurs, which contrasts the findings with the Satelec device (Figure 2d).
The mean maximum rise in temperature (deltaTmax, all implants averaged per experimental condition) was reached at 30 to 40 s for all conditions. There was a significant difference for the Satelec device without cooling (mean 9.6, SD 1.6°C) and the EMS device without cooling (mean 4.3, SD 2.0°C) (t-test, t(28) = 4.7, p < 0.001). When using coolant, the deltaTmax for the Satelec device (mean 8.3, SD 1.3°C) was significantly higher than for the EMS device (mean 1.6, SD 0.6°C) (t-test, t(28) = 17.9, p < 0.001).

Discussion

Several techniques have been described to deal with biomechanical complications. Acquiring adequate visibility and access is essential to success which will require the use of a dental microscope.
Several mechanical approaches to remove screw remnants can be employed. Generally, after identifying the position and condition of the screw remnant, it can be carefully removed using manual instrumentation. A fine-tipped hand instrument may be wedged between screw and implant. If this fails to do the trick, the use of a round burr applying occlusal pressure and turning counter clockwise may loosen the fragment. When a flared burr is wedged between implant and screw remnant, the drill should make a clockwise action, to direct the screw in the counterclockwise direction. Cutting a deep slot in the screw and subsequently removing it with a bladed screw driver has also been advised. The use of rotary instrumentation may cause damage to the cranial inner portion of the implant body. If that occurs, even when the fragment (finally) loosens, it will not screw out, and this complicates matters considerably.
Most implant suppliers, and also some third parties, provide specific instrumentation to remove broken screws from implants (i.e., Certain screw removal kit, Biomet 3I, Palm Beach Gardens, United States; screw removal kit NobelReplace, Nobel Biocare, Göteborg, Sweden; Neo screw remover kit, Neobiotech Co, South Korea). In general, they screw into the center of the screw remnant, which facilitates another burr to grip it and remove it counter clockwise or fragmentate the screw remnant.
The use of ultrasonic equipment has also proven to be effective in dislodging fixed screw remnants at the risk of damaging the inner portion or overheating the implant. Hence, the use of magnification is a must and cooling may be advisable, but compromises the vision of the operator. The present study evaluated heat accumulation in vitro and the efficiency of cooling when using two types of ultrasonic equipment to the inner portion of several implants.
The degree to which a material is able to transfer heat is called thermal conductivity. It can be defined as the time rate of transfer by conduction, through unit thickness, across unit area for unit temperature gradient. Differences in design and wall thickness of the implants used in the present study account for the variation in outcome (i.e., Straumann tissue level 4.8 implant less effected), considering the fact that all implant types used are made from the same material: grade 4 commercially pure (CP) titanium. The thermal conductivity of CP titanium is relatively low compared to, for instance, that of steel (1/4) and aluminum (1/13). On the other hand, it is approximately 60% higher than that of grade 5 titanium alloy (16.3 W · m−1 · K−1 for grade 4 CP titanium compared to 7.2 W · m−1 · K−1 for grade 5 Ti-6Al-4 V). The former material is used in several other implant brands than the ones used here [38], and as a consequence, the results cannot be extrapolated to those systems. The data from this in vitro experiment can only be generalized to the clinical situation bearing in mind some inherent limitations and assumptions. The epoxy resin used does not resemble alveolar bone, its structure, water content, and potential to cope with thermal challenges. It is unlikely that wall thickness, design, and material between different implants from the same implant type will vary because of the high degree of current precision and standardization achieved during the fabrication of implants. Consequently, only one specimen per brand was used. To correct for variation during the instrumentation of the ultrasonic device, the experiment was performed three times.
The results show that both 1. the type of implant and 2. the type of ultrasonic device (and in especially the use of coolant) affect the amount of temperature rise to the outer implant surface. Both bone level implants in particular appear to heat up the most. Without the use of coolant, the heat accumulation was much higher with the Satelec compared to the EMS device and exceeded the theoretical threshold for permanent biological damage after 10 to 15 s of continuous instrumentation (Figure 2a,c) and differed statistically significant at deltaTmax. There may be several explanations for this, but most likely, the produced energy at the point of the tip for the Satelec device was higher. Both devices were set at their lowest possible level. The present authors have not been able to verify what energy levels are actually produced. The former is presumed linear to the frequency and the deflection/amplitude of the tip. The frequency is mentioned in the product documentation (Satelec: 27 to 33 kHz and the EMS: 24 to 32 kHz) but the amplitudes differ per tip and are not disclosed. The effect of the difference in coolant spray between the two instruments is also a factor that needs considering when interpreting the data.
Cooling proves effective for both systems, increasing the outer implant temperature by an acceptable 1°C to 3°C during continued instrumentation; however, as already stated, the spray will blur the vision of the operator.
One would expect the temperature to drop immediately after cessation of instrumentation and cooling, but the peak temperature is reached some seconds later for all experimental conditions, so regardless of device, use of coolant, or implant type (Figure 2a,b,c,d). Others, instrumenting endodontic posts in natural teeth [32,33], also saw a ‘lag’ period with a rise in temperature after cessation that lasted up to 9 s post-instrumentation. They left it unexplained. We offer two possible explanations. Firstly, ultrasonic instrumentation causes the implant wall and its surrounding tissues (in this case, resin and thermocouple, but in vivo, this would be surrounding alveolar bone) to vibrate and generate heat. Vibration continues some time after cessation of instrumentation, explaining the continued rise in temperature. Secondly, after cooling has stopped, generated and stored heat in the surrounding ‘tissues’ may flow back to the outer wall of the implant that is no longer cooled from the inside and raises the temperature. The potentially damaging rise in temperature (‘explosion’) seen for the Satelec device, but not for the EMS device when instrumentation and cooling was stopped, could be explained by these phenomena on the one hand and by the fact that energy produced by the Satelec device was presumably higher on the other hand (Figure 2b).

Conclusions

It is concluded from this in vitro study that heat accumulation and transfer is dependent on the type of ultrasonic device, the use of coolant, and the implant type. The highest rise in temperature is seen when using the Satelec device without coolant on the smaller diameter implants. The EMS device causes limited rise in temperature when used without coolant for less than 10 s, but presumably delivers less energy to the tip, and consequently may be not as effective. A cooling down period is sensible. When used with coolant for a longer period of time, the clinician should anticipate a considerable rise in temperature after cessation of instrumentation, and post-operational cooling is advised.

Acknowledgements

The input of Prashant K. Sharma PhD, material scientist of the Department of Biomedical Engineering (Kolff Institute) of the UMC Groningen is greatly appreciated for his help interpreting the data. Michael Kars, DDS assisted during the collection of the data.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://​creativecommons.​org/​licenses/​by/​4.​0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Competing interests

This research was made possible by the support of the authors’ institutions. Eric W Meisberger, Sjoerd JG Bakker and Marco S Cune declare that they have no competing interests.

Authors’ contributions

EM and SB carried out the experiment. MC participated in the design of the study and performed the statistical analysis and helped to draft the manuscript. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Sailer I, Philipp A, Zembic A, Pjetursson BE, Hammerle CH, Zwahlen M. A systematic review of the performance of ceramic and metal implant abutments supporting fixed implant reconstructions. Clin Oral Implants Res. 2009;20 Suppl 4:4–31.CrossRefPubMed Sailer I, Philipp A, Zembic A, Pjetursson BE, Hammerle CH, Zwahlen M. A systematic review of the performance of ceramic and metal implant abutments supporting fixed implant reconstructions. Clin Oral Implants Res. 2009;20 Suppl 4:4–31.CrossRefPubMed
2.
Zurück zum Zitat Romeo E, Storelli S. Systematic review of the survival rate and the biological, technical, and aesthetic complications of fixed dental prostheses with cantilevers on implants reported in longitudinal studies with a mean of 5 years follow-up. Clin Oral Implants Res. 2012;23 Suppl 6:39–49.CrossRefPubMed Romeo E, Storelli S. Systematic review of the survival rate and the biological, technical, and aesthetic complications of fixed dental prostheses with cantilevers on implants reported in longitudinal studies with a mean of 5 years follow-up. Clin Oral Implants Res. 2012;23 Suppl 6:39–49.CrossRefPubMed
3.
Zurück zum Zitat Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res. 2008;19:119–30.CrossRefPubMed Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res. 2008;19:119–30.CrossRefPubMed
4.
Zurück zum Zitat Hsu YT, Fu JH, Al-Hezaimi K, Wang HL. Biomechanical implant treatment complications: a systematic review of clinical studies of implants with at least 1 year of functional loading. Int J Oral Maxillofac Implants. 2012;27:894–904.PubMed Hsu YT, Fu JH, Al-Hezaimi K, Wang HL. Biomechanical implant treatment complications: a systematic review of clinical studies of implants with at least 1 year of functional loading. Int J Oral Maxillofac Implants. 2012;27:894–904.PubMed
5.
Zurück zum Zitat Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003;90:121–32.CrossRefPubMed Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent. 2003;90:121–32.CrossRefPubMed
6.
Zurück zum Zitat Pjetursson BE, Bragger U, Lang NP, Zwahlen M. Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs). Clin Oral Implants Res. 2007;18 Suppl 3:97–113.CrossRefPubMed Pjetursson BE, Bragger U, Lang NP, Zwahlen M. Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs). Clin Oral Implants Res. 2007;18 Suppl 3:97–113.CrossRefPubMed
7.
Zurück zum Zitat Kreissl ME, Gerds T, Muche R, Heydecke G, Strub JR. Technical complications of implant-supported fixed partial dentures in partially edentulous cases after an average observation period of 5 years. Clin Oral Implants Res. 2007;18:720–6.CrossRefPubMed Kreissl ME, Gerds T, Muche R, Heydecke G, Strub JR. Technical complications of implant-supported fixed partial dentures in partially edentulous cases after an average observation period of 5 years. Clin Oral Implants Res. 2007;18:720–6.CrossRefPubMed
8.
Zurück zum Zitat Cranin AN, Dibling JB, Simons A, Klein M, Sirakian A. Report of the incidence of implant insert fracture and repair of Core-Vent dental implants. J Oral Implantol. 1990;16:184–8.PubMed Cranin AN, Dibling JB, Simons A, Klein M, Sirakian A. Report of the incidence of implant insert fracture and repair of Core-Vent dental implants. J Oral Implantol. 1990;16:184–8.PubMed
9.
Zurück zum Zitat Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol. 2002;28:67–73.CrossRefPubMed Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol. 2002;28:67–73.CrossRefPubMed
10.
Zurück zum Zitat Lee J, Kim YS, Kim CW, Han JS. Wave analysis of implant screw loosening using an air cylindrical cyclic loading device. J Prosthet Dent. 2002;88:402–8.CrossRefPubMed Lee J, Kim YS, Kim CW, Han JS. Wave analysis of implant screw loosening using an air cylindrical cyclic loading device. J Prosthet Dent. 2002;88:402–8.CrossRefPubMed
11.
Zurück zum Zitat Bakaeen LG, Winkler S, Neff PA. The effect of implant diameter, restoration design, and occlusal table variations on screw loosening of posterior single-tooth implant restorations. J Oral Implantol. 2001;27:63–72.CrossRefPubMed Bakaeen LG, Winkler S, Neff PA. The effect of implant diameter, restoration design, and occlusal table variations on screw loosening of posterior single-tooth implant restorations. J Oral Implantol. 2001;27:63–72.CrossRefPubMed
12.
Zurück zum Zitat Assenza B, Tripodi D, Scarano A, Perrotti V, Piattelli A, Iezzi G, et al. Bacterial leakage in implants with different implant-abutment connections: an in vitro study. J Periodontol. 2011;83:491–7.CrossRefPubMed Assenza B, Tripodi D, Scarano A, Perrotti V, Piattelli A, Iezzi G, et al. Bacterial leakage in implants with different implant-abutment connections: an in vitro study. J Periodontol. 2011;83:491–7.CrossRefPubMed
13.
Zurück zum Zitat Ha CY, Lim YJ, Kim MJ, Choi JH. The influence of abutment angulation on screw loosening of implants in the anterior maxilla. Int J Oral Maxillofac Implants. 2011;26:45–55.PubMed Ha CY, Lim YJ, Kim MJ, Choi JH. The influence of abutment angulation on screw loosening of implants in the anterior maxilla. Int J Oral Maxillofac Implants. 2011;26:45–55.PubMed
14.
Zurück zum Zitat Spazzin AO, Henrique GE, Nobilo MA, Consani RL, Correr-Sobrinho L, Mesquita MF. Effect of retorque on loosening torque of prosthetic screws under two levels of fit of implant-supported dentures. Braz Dent J. 2010;21:12–7.CrossRefPubMed Spazzin AO, Henrique GE, Nobilo MA, Consani RL, Correr-Sobrinho L, Mesquita MF. Effect of retorque on loosening torque of prosthetic screws under two levels of fit of implant-supported dentures. Braz Dent J. 2010;21:12–7.CrossRefPubMed
15.
Zurück zum Zitat Khraisat A. bu-Hammad O, Dar-Odeh N, Al-Kayed AM. Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading. Clin Implant Dent Relat Res. 2004;6:157–64.CrossRefPubMed Khraisat A. bu-Hammad O, Dar-Odeh N, Al-Kayed AM. Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading. Clin Implant Dent Relat Res. 2004;6:157–64.CrossRefPubMed
16.
Zurück zum Zitat Farina AP, Spazzin AO, Pantoja JM, Consani RL, Mesquita MF. An in vitro comparison of joint stability of implant-supported fixed prosthetic suprastructures retained with different prosthetic screws and levels of fit under masticatory simulation conditions. Int J Oral Maxillofac Implants. 2012;27:833–8.PubMed Farina AP, Spazzin AO, Pantoja JM, Consani RL, Mesquita MF. An in vitro comparison of joint stability of implant-supported fixed prosthetic suprastructures retained with different prosthetic screws and levels of fit under masticatory simulation conditions. Int J Oral Maxillofac Implants. 2012;27:833–8.PubMed
17.
Zurück zum Zitat Diez JS, Brigagao VC, Cunha LG, Neves AC, da Silva-Concilio LR. Influence of diamondlike carbon-coated screws on the implant-abutment interface. Int J Oral Maxillofac Implants. 2012;27:1055–60.PubMed Diez JS, Brigagao VC, Cunha LG, Neves AC, da Silva-Concilio LR. Influence of diamondlike carbon-coated screws on the implant-abutment interface. Int J Oral Maxillofac Implants. 2012;27:1055–60.PubMed
18.
Zurück zum Zitat Kim SK, Koak JY, Heo SJ, Taylor TD, Ryoo S, Lee SY. Screw loosening with interchangeable abutments in internally connected implants after cyclic loading. Int J Oral Maxillofac Implants. 2012;27:42–7.PubMed Kim SK, Koak JY, Heo SJ, Taylor TD, Ryoo S, Lee SY. Screw loosening with interchangeable abutments in internally connected implants after cyclic loading. Int J Oral Maxillofac Implants. 2012;27:42–7.PubMed
19.
Zurück zum Zitat Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J. 2009;28:373–81.CrossRefPubMed Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J. 2009;28:373–81.CrossRefPubMed
20.
Zurück zum Zitat Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J Prosthet Dent. 2002;88:183–91.CrossRefPubMed Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J Prosthet Dent. 2002;88:183–91.CrossRefPubMed
21.
Zurück zum Zitat Bidra AS, Rungruanganunt P. Clinical outcomes of implant abutments in the anterior region: a systematic review. J Esthet Restor Dent. 2013;25:159–76.CrossRefPubMed Bidra AS, Rungruanganunt P. Clinical outcomes of implant abutments in the anterior region: a systematic review. J Esthet Restor Dent. 2013;25:159–76.CrossRefPubMed
22.
Zurück zum Zitat Wong K, Boyde A, Howell PG. A model of temperature transients in dental implants. Biomaterials. 2001;22:2795–7.CrossRefPubMed Wong K, Boyde A, Howell PG. A model of temperature transients in dental implants. Biomaterials. 2001;22:2795–7.CrossRefPubMed
23.
Zurück zum Zitat Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent. 1983;50:101–7.CrossRefPubMed Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent. 1983;50:101–7.CrossRefPubMed
24.
Zurück zum Zitat Benington IC, Biagioni PA, Briggs J, Sheridan S, Lamey PJ. Thermal changes observed at implant sites during internal and external irrigation. Clin Oral Implants Res. 2002;13:293–7.CrossRefPubMed Benington IC, Biagioni PA, Briggs J, Sheridan S, Lamey PJ. Thermal changes observed at implant sites during internal and external irrigation. Clin Oral Implants Res. 2002;13:293–7.CrossRefPubMed
25.
Zurück zum Zitat Li S, Chien S, Branemark PI. Heat shock-induced necrosis and apoptosis in osteoblasts. J Orthop Res. 1999;17:891–9.CrossRefPubMed Li S, Chien S, Branemark PI. Heat shock-induced necrosis and apoptosis in osteoblasts. J Orthop Res. 1999;17:891–9.CrossRefPubMed
26.
Zurück zum Zitat Huh JB, Eckert SE, Ko SM, Choi YG. Heat transfer to the implant-bone interface during preparation of a zirconia/alumina abutment. Int J Oral Maxillofac Implants. 2009;24:679–83.PubMed Huh JB, Eckert SE, Ko SM, Choi YG. Heat transfer to the implant-bone interface during preparation of a zirconia/alumina abutment. Int J Oral Maxillofac Implants. 2009;24:679–83.PubMed
27.
Zurück zum Zitat Nissan J, Gross M, Ormianer Z, Barnea E, Assif D. Heat transfer of impression plasters to an implant-bone interface. Implant Dent. 2006;15:83–8.CrossRefPubMed Nissan J, Gross M, Ormianer Z, Barnea E, Assif D. Heat transfer of impression plasters to an implant-bone interface. Implant Dent. 2006;15:83–8.CrossRefPubMed
28.
Zurück zum Zitat Gross M, Laufer BZ, Ormianer Z. An investigation on heat transfer to the implant-bone interface due to abutment preparation with high-speed cutting instruments. Int J Oral Maxillofac Implants. 1995;10:207–12.PubMed Gross M, Laufer BZ, Ormianer Z. An investigation on heat transfer to the implant-bone interface due to abutment preparation with high-speed cutting instruments. Int J Oral Maxillofac Implants. 1995;10:207–12.PubMed
29.
Zurück zum Zitat Feuerstein O, Zeichner K, Imbari C, Ormianer Z, Samet N, Weiss EI. Temperature changes in dental implants following exposure to hot substances in an ex vivo model. Clin Oral Implants Res. 2008;19:629–33.CrossRefPubMed Feuerstein O, Zeichner K, Imbari C, Ormianer Z, Samet N, Weiss EI. Temperature changes in dental implants following exposure to hot substances in an ex vivo model. Clin Oral Implants Res. 2008;19:629–33.CrossRefPubMed
30.
Zurück zum Zitat Ormianer Z, Laufer BZ, Nissan J, Gross M. An investigation of heat transfer to the implant-bone interface related to exothermic heat generation during setting of autopolymerizing acrylic resins applied directly to an implant abutment. Int J Oral Maxillofac Implants. 2000;15:837–42.PubMed Ormianer Z, Laufer BZ, Nissan J, Gross M. An investigation of heat transfer to the implant-bone interface related to exothermic heat generation during setting of autopolymerizing acrylic resins applied directly to an implant abutment. Int J Oral Maxillofac Implants. 2000;15:837–42.PubMed
31.
Zurück zum Zitat Ormianer Z, Feuerstein O, Assad R, Samet N, Weiss EI. In vivo changes in dental implant temperatures during hot beverage intake: a pilot study. Implant Dent. 2009;18:38–45.CrossRefPubMed Ormianer Z, Feuerstein O, Assad R, Samet N, Weiss EI. In vivo changes in dental implant temperatures during hot beverage intake: a pilot study. Implant Dent. 2009;18:38–45.CrossRefPubMed
32.
Zurück zum Zitat Huttula AS, Tordik PA, Imamura G, Eichmiller FC, McClanahan SB. The effect of ultrasonic post instrumentation on root surface temperature. J Endod. 2006;32:1085–7.CrossRefPubMed Huttula AS, Tordik PA, Imamura G, Eichmiller FC, McClanahan SB. The effect of ultrasonic post instrumentation on root surface temperature. J Endod. 2006;32:1085–7.CrossRefPubMed
33.
Zurück zum Zitat Davis S, Gluskin AH, Livingood PM, Chambers DW. Analysis of temperature rise and the use of coolants in the dissipation of ultrasonic heat buildup during post removal. J Endod. 2010;36:1892–6.CrossRefPubMed Davis S, Gluskin AH, Livingood PM, Chambers DW. Analysis of temperature rise and the use of coolants in the dissipation of ultrasonic heat buildup during post removal. J Endod. 2010;36:1892–6.CrossRefPubMed
34.
Zurück zum Zitat Ettrich CA, Labossiere PE, Pitts DL, Johnson JD. An investigation of the heat induced during ultrasonic post removal. J Endod. 2007;33:1222–6.CrossRefPubMed Ettrich CA, Labossiere PE, Pitts DL, Johnson JD. An investigation of the heat induced during ultrasonic post removal. J Endod. 2007;33:1222–6.CrossRefPubMed
35.
Zurück zum Zitat Satterthwaite JD, Stokes AN, Frankel NT. Potential for temperature change during application of ultrasonic vibration to intra-radicular posts. Eur J Prosthodont Restor Dent. 2003;11:51–6.PubMed Satterthwaite JD, Stokes AN, Frankel NT. Potential for temperature change during application of ultrasonic vibration to intra-radicular posts. Eur J Prosthodont Restor Dent. 2003;11:51–6.PubMed
36.
Zurück zum Zitat Horan BB, Tordik PA, Imamura G, Goodell GG. Effect of dentin thickness on root surface temperature of teeth undergoing ultrasonic removal of posts. J Endod. 2008;34:453–5.CrossRefPubMed Horan BB, Tordik PA, Imamura G, Goodell GG. Effect of dentin thickness on root surface temperature of teeth undergoing ultrasonic removal of posts. J Endod. 2008;34:453–5.CrossRefPubMed
37.
Zurück zum Zitat Dominici JT, Clark S, Scheetz J, Eleazer PD. Analysis of heat generation using ultrasonic vibration for post removal. J Endod. 2005;31:301–3.CrossRefPubMed Dominici JT, Clark S, Scheetz J, Eleazer PD. Analysis of heat generation using ultrasonic vibration for post removal. J Endod. 2005;31:301–3.CrossRefPubMed
38.
Zurück zum Zitat Jokstad A, Braegger U, Brunski JB, Carr AB, Naert I, Wennerberg A. Quality of dental implants. Int Dent J. 2003;53:409–43.CrossRefPubMed Jokstad A, Braegger U, Brunski JB, Carr AB, Naert I, Wennerberg A. Quality of dental implants. Int Dent J. 2003;53:409–43.CrossRefPubMed
Metadaten
Titel
Temperature rise during removal of fractured components out of the implant body: an in vitro study comparing two ultrasonic devices and five implant types
verfasst von
Eric W Meisberger
Sjoerd J G Bakker
Marco S Cune
Publikationsdatum
01.12.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Implant Dentistry / Ausgabe 1/2015
Elektronische ISSN: 2198-4034
DOI
https://doi.org/10.1186/s40729-015-0008-0

Weitere Artikel der Ausgabe 1/2015

International Journal of Implant Dentistry 1/2015 Zur Ausgabe

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.