01.12.2014 | Research article | Ausgabe 1/2014 Open Access

BMC Family Practice 1/2014

The diagnostic accuracy of the MyDiagnostick to detect atrial fibrillation in primary care

Zeitschrift:
BMC Family Practice > Ausgabe 1/2014
Autoren:
Bert Vaes, Silke Stalpaert, Karen Tavernier, Britt Thaels, Daphne Lapeire, Wilfried Mullens, Jan Degryse
Wichtige Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2296-15-113) contains supplementary material, which is available to authorized users.

Competing interests

The author declares that they have no competing interests.

Authors’ contributions

BV, SS, KT, BT, DL drafted the manuscript. JD is responsible for the design, conduct and analysis of the study. All authors participated in the critical revision of the manuscript. All authors read and approved the final manuscript.

Abstract

Background

Atrial fibrillation is very common in people aged 65 or older. This condition increases the risk of death, congestive heart failure and thromboembolic conditions. Many patients with atrial fibrillation are asymptomatic and a cerebrovascular accident (CVA) is often the first clinical presentation. Guidelines concerning the prevention of CVA recommend monitoring the heart rate in patients aged 65 or older. Recently, the MyDiagnostick (Applied Biomedical Systems BV, Maastricht, The Netherlands) was introduced as a new screening tool which might serve as an alternative for the less accurate pulse palpation. This study was designed to explore the diagnostic accuracy of the MyDiagnostick for the detection of atrial fibrillation.

Methods

A phase II diagnostic accuracy study in a convenience sample of 191 subjects recruited in primary care. The majority of participants were patients with a known history of atrial fibrillation (n = 161). Readings of the MyDiagnostick were compared with electrocardiographic recordings. Sensitivity and specificity and their 95% confidence interval were calculated using 2x2 tables.

Results

A prevalence of 54% for an atrial fibrillation rhythm was found in the study population at the moment of the study. A combination of three measurements with the MyDiagnostick for each patient showed a sensitivity of 94% (95% CI 87 – 98) and a specificity of 93% (95% CI 85 – 97).

Conclusion

The MyDiagnostick is an easy-to-use device that showed a good diagnostic accuracy with a high sensitivity and specificity for atrial fibrillation in a convenience sample in primary care. Future research is needed to determine the place of the MyDiagnostick in possible screening or case-finding strategies for atrial fibrillation.