Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 11/2019

Open Access 19.06.2019 | Trauma Surgery

The PRECICE magnetic IM compression nail for long bone nonunions: a preliminary report

verfasst von: Austin T. Fragomen, David Wellman, S. Robert Rozbruch

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 11/2019

Abstract

Introduction

The magnetic intramedullary (IM) compression nail is capable of providing sustained compression for the treatment of nonunions of long bones. This ability was previously only possible with the use of external fixation. We asked the following questions: How effective is the IM compression nail at achieving union? How do we know when adequate compression has been attained? Which types of nonunions are good candidates for this treatment?

Materials and methods

Fourteen patients with nonunions of the tibia (5) or femur (9) were treated with the PRECICE IM compression nail. The average age was 49 years number of previous surgeries was 1.9, 7 were atrophic and 7 normotrophic, 3 were metaphyseal and 11 diaphyseal. All PRECICE IM nails were pre-distracted prior to implantation. Compression was applied post-operatively until the locking bolts were seen on X-ray to be bending or the nail was no longer shortening despite applying the external magnet.

Results

Union was achieved in 13/14 cases. The time to union was 24.5 weeks (range 11–60). The two proximal tibia metaphyseal nonunions, both deformed into varus (4°) and flexion (10°) after compression was applied with one failing to unite. The distal tibia metaphyseal and diaphyseal nonunions did not deform upon compression. Three patients had positive cultures and were treated with IV antibiotics for 6 weeks followed by 3 months of oral suppression without subsequent infection. No mechanical nail failures were seen.

Conclusions

The IM compression nail was successful at applying compression, preventing deformity, and obtaining union in all diaphyseal and in distal tibia metaphyseal nonunions. Signs of active compression are bending of the locking bolts and failure of the nail to shorten. Proximal tibia metaphyseal nonunion may not be suited for this treatment.

Introduction

Nonunions of the femur and tibia occur in 2.5–10% of fracture cases [1, 2] and can be difficult to manage. Nonunion treatment often requires several surgeries, creates lost work time, and results in both financial burden [3, 4] and a poor health-related quality of life for the patient [5, 6]. The etiology of fracture nonunion is vast and multifactorial [7]. Treatments have primarily been surgical. Open bone grafting with plating has yielded high rates of union [8, 9], but plating is an extensive surgical intervention. Exchange intramedullary (IM) nailing has been the mainstay of treatment for diaphyseal nonunion in the femur and tibia with success rates ranging from 53 to 100% [10, 11] and 72 to 92% [12], respectively. Part of the range in union outcomes may be due to poor bone contact at the nonunion docking site [12]. The value of compression in the biomechanics of osteogenesis has been long recognized [13, 14]. IM tibial and femoral nails allow for mild compression to be delivered at the time of surgery [15, 16] but provide no mechanism to sustain the compression over time. Circular external fixation provides an alternative method for nonunion management with the advantage of providing sustained compression during the recovery period but requires wearing the frame for a prolonged post-operative course [1719]. Surgeons have gone to great lengths to achieve compression, even integrating IM nails with circular fixators as in the compression over a nail technique [20]. The advent of the PRECICE (NuVasive Specialized Orthopedics, San Diego, CA, USA) magnetic IM compression and distraction nail has created a treatment option that combines the convenience of IM nailing with the sustained compression that is ideal for fracture healing. Watson et al. [21] published a series of at-risk humeral fractures treated with the magnetic intramedullary compression nail (MICN) that demonstrated 100% union. A case report of a patient who had failed circular fixation-assisted compression for a tibial nonunion demonstrated rapid healing when treated with the MICN [22]. The present study asked the following questions: (1) How effective is the MICN at achieving union? (2) How do we know when adequate compression has been attained? (3) Which types of nonunions are good candidates for this treatment?

Materials and methods

Study group

Fourteen patients with nonunions of the tibia (5) or femur (9) were treated with the PRECICE MICN. Patient demographics are listed (Table 1). Eleven of the 14 patients had post-traumatic fracture nonunion of either the femur or tibia, 1 patient had nonunion of a femoral allograft used for reconstruction after tumor resection, and 2 had femur osteotomy nonunions. Nonunion was confirmed on CT scan when questionable. Ten of 14 patients had failed prior attempts at nonunion repair (average 2.3 previous surgeries, range 2–5). Indications for MICN implant used in this practice included cases of nonunion in long bones where treatment with standard internal fixation or circular external fixation was already attempted or such treatment could be expected to fail. These selective indications are responsible for the limited size of this cohort. All MICNs were pre-distracted an average of 13.5 mm (range 10–18) prior to implantation. Five of the patients had been treated for fracture-related infection and were thought to be infection free at the time of MICN insertion. All patients received a standard pre-operative workup including serum ESR, CRP, and WBC counts. Active infection was a relative contraindication for this surgery. Need for hardware removal, deformity correction, and for later limb lengthening due to limb length inequality were assessed. Appropriateness of treatment was considered. Access to and experience with state-of-the-art hexapod external fixation and internal fixation from multiple vendors were unlimited, and surgeons chose to use the MICN strictly based on clinical experience.
Table 1
Patient demographics
Case #
Age
Sex
Seg
Etiology
Comorbidity
Fx type (GA)
Rx (#; type)
Time: initial Rx failure to ICN (months)
NU type
AIM
1
49
M
TB
Trauma
Smoking, opioid, treated infection
O, 3A
5; IMN, CP, BT
72
N
11
2
27
M
TB
Trauma
Highly contaminated Fx
O, 3A
2; CP, CEF
14
A
11
3
54
M
FR
Tumor resection
treated infection
C, Fx Allogrft
2; IMN and CP
84
N
5
4
27
M
FR
Trauma
Obesity, treated infection, bone defect
C
2; IMN
8
A
8
5
35
M
FR
Trauma
Smoking
O, 3A
1; IMN
23
N
5
6
71
M
TB
Trauma
DM2, ESRD, CHF, treated infection, bone defect
O, 3B
2; IMN, CEF
15
A
19
7
47
F
FR
Trauma
DM1, opioid, blindness, osteoporosis
C
1; IMN
15
N
7
8
40
F
FR
Osteotomy
Rickets
C
1; IMN
10
A
6
9
83
F
FR
Trauma
Treated infection, DM
C
2; IMN and MEF
60
A
8
10
24
M
FR
Trauma
none
C
2; IMN
13
A
5
11
56
F
FR
Osteotomy
none
C
2; IMN
13
N
5
12
75
F
FR
Trauma
Smoking, opioid
C
2; CP and IMN
16
A
6
13
41
M
TB
Trauma
none
0, 3A
1; IMN
10
N
4
14
62
M
TB
Trauma
RA, enbrel, opioid
C
2; CP, BMAC
13
N
8
Mean
49.3
M- 9/14
FR-9
TB-5
-
-
C-9
O-5
1.9
26.1
A-7/14
7.7
Case # patient case #, M male, F female, Seg segment, T tibia, F femur, Opioid opioid addiction/chronic pain, DM1 diabetes mellitus type 1, DM2 diabetes mellitus type 2, ESRD end stage renal disease, CHF congestive heart failure, Fx fracture, GA Gustilo–Anderson [23], O open fracture, C closed, Rx treatment, IMN intramedullary nail, CP compression plating, BT bone transport, CEF circular external fixation, ICN internal compression nailing, NU nonunion, N normotrophic, A atrophic, AIM AIM score [24] with 7.7 representing moderate complexity

Surgical technique

Existing hardware was removed either completely or at minimum from the path of the planned IM nail. In most cases, the nonunion site was approached with an open technique, debrided to bleeding bone, and irrigated. Cultures and pathology specimens were obtained. A two-pin external fixator was applied posterior to the path of the nail to control rotation and the reduction. Blocking screws were inserted to prevent deformity in all metaphyseal nonunions and in some diaphyseal cases. These screws were inserted prior to reaming to help direct the reamer. The nail entry point was localized with a Steinman pin, and a 12-mm cannulated acorn reamer was used to start the path of the nail. A rigid hand reamer was needed in many cases to recreate the IM canal across the sclerotic nonunion site prior to ball-tipped guide wire insertion. Once the ball-tipped wire was in position, reaming was performed using flexible reamers. Overreaming by 2 mm was done to ensure the bone’s ability to slide over the nail and compress across the nonunion site. The guide wire was removed, and the non-cannulated nail was inserted gently. Rough impaction or excessive bending of the nail can damage the internal gears and render the nail ineffective. If the bone was too tight to insert the nail, then further reaming was performed. Locking of the nail was performed in a standard fashion with a combination of a jig for the screws near the entry and free-hand technique for the screws at the far end.
Using the C-arm fluoroscopy, the internal magnet was localized and marked on the skin. A loose stitch was placed through the skin at this point as a more permanent indicator of where to place the external remote control (ERC) magnet post-operatively. The ERC placed on the thigh or leg spins the magnet in the nail leading to gradual compression. Compression was applied in the operating room in two cases and post-operatively in all cases until the locking bolts were seen to be bending under the force of the nail or the nail was no longer shortening despite applying the ERC (Table 2). The telltale sign that compression was present was bolt deflection (bending) as seen on radiographs. If bolt bending was not seen on a post-operative radiograph, then more compression was applied. (Fig. 1a–c) Since this system was originally designed for bone lengthening, to compress the nail the magnet was programed to lengthen in the opposite direction. For example, for an antegrade femur nail to compress, the ERC was programed for a retrograde lengthening. Compression intervals were set for 1 mm per session.
Table 2
Compression schedules used
Case #
Intra-op compression
POD 2
POD 3
POD 4
Home
Office
Total compression
1
0
2
2
1
0
1 PRN
8
2
2
2
2
0
0
1 PRN
8
3
0
0
0
0
2/day
 
10
4
0
1
1
1
0
1 PRN
10
5
0
0
0
0
3
 
3
6
0
2
2
2
2/day
 
10
7
0
2
2
2
1
 
7
8
0
2
2
1
2/day
 
11
9
0
1
1
1
0
1 PRN
9
10
0
1
1
1
0
1 PRN
6
11
1
1
1
1
1/day
 
13
12
0
2
2
2
0
2 PRN
10
13
0
2
2
2
0
2 PRN
10
14
0
2
2
2
3
1 PRN
18
Mean
0.21
1.4
1.4
1.1
  
9.5
The applied compression in this table refers to the quantity and frequency of ERC applications. It is a measure of how much compression the patient attempted to deliver to the nail and bone
Patients were placed on venous thromboembolic (VTE) prophylaxis, typically rivaroxaban 10 mg, starting post-operative day 2 and lasting for 2 weeks. Weight bearing was allowed but varied by patient bone quality, surgeon experience, and surgeon preference (Table 3). Patients were followed monthly until consolidation. An ERC was available in the outpatient clinic, and additional compression was applied as indicated at these follow-up visits.
Table 3
Peri-operative details results
Case #
Nail diamtr (mm)
Nail Pre-distraction (mm)
WB (% BW)
Cmprs applied (mm)
Distancenail shortened (mm)
Distance bone shortened (mm)
Bolt bending (deg)
FRI + 
Time to union (wks)
Final LLD (mm)
1
12.5
10
50
8
8
5
2
 
20
25
2
10.7
10
50
8
2
0
1
 
12
19
3
12.5
15
50
10
10
7
6
 
45
112
4
12.5
15
50
10
10
3
4
Y
32
100
5
12.5
15
AT
3
3
2
0
 
19
0
6
12.5
15
AT
10
9
5
5
 
15
60
7
12.5
13
50
7
4
2
3
 
11
25
8
10.7
13
AT
11
6
4
3
 
34
51
9
12.5
13
50
9
8
4
0
 
60
50
10
12.5
13
70lbs
6
6
3
6
 
13
10
11
12.5
13
70lbs
13
9
3
5
 
13
20
12
12.5
13
AT
10
6
0
0
 
28
40
13
10.7
13
AT
10
4
0
0
Y
16
0
14
12.5
18
70lbs
18
10
5
0
 
19
Mean
13.5
9.5
6.7
3.1
2.5
24.5
37.9
Diamtr diameter, Cmprs compression, WB weight bearing, %BW % of body weight allowed, AT as tolerated, wks weeks, XR X-ray, FRI fracture-related infection, LLD limb length discrepancy

Study design

A retrospective analysis was performed on a consecutive series of 14 patients treated by 3 surgeons at 1 academic orthopedic center in an urban setting. Inclusion criteria for this study included treatment of an established nonunion of a long bone with an MICN. Patients that had the MICN used for purposes other than nonunion repair were excluded. The surgical details were recorded (Table 3). Outcome measures included successful bony union with the implant and time to union. Union was defined as bridging callus on three of four cortices on radiographs or on CT scan. The amounts of compression applied, distance the nail shortened, distance the bone shortened, and angle of locking bolt bending were measured. (Fig. 2a, b) All complications were recorded. AIM scores were collected to gauge the complexity of the impending reconstruction [24].

Statistics

Statistical analysis was not employed for this limited study. Any conclusions for significance would be unreliable in this cohort.

Results

Union, as defined by bridging callus on 3 of 4 cortices on either radiographs or CT scan, was achieved in 13/14 cases. The time to union was 24.5 weeks (range 11–60). Follow-up was carried to final union or revision surgery in all cases. The average follow-up was 19 months (range 6–33). Two patients had less than 12-month follow-up. One united after 4 months. The other was censured at 8 months due to persistent nonunion, increasing progressive deformity, and loose hardware that required revision to external fixation.
On average, the ERC was used to compress the nail 9.5 mm (range), the nail actually shortened 6.7 mm (range), the bone shortened 3.1 mm (range), and the locking bolts deflected 2.5° (Table 3, Fig. 3a, b). Limb length discrepancy (LLD) after nonunion treatment averaged 3.7 cm (range 0–11.2 cm).
The two proximal tibial metaphyseal nonunions both deformed into varus and flexion after compression was applied with one failing to unite (Table 4). This deformity occurred despite the appropriate use of blocking screws [25]. The patient who failed to unite also had backing out of the locking screws and was revised with a hexapod frame. The distal tibia metaphyseal and the diaphyseal nonunions did not deform upon compression. Two patients (Cases #4 and 13) had positive intra-operative cultures and were treated with IV antibiotics for 6 weeks followed by 3 months of oral suppression without subsequent infection. One patient (Case #2) had an early post-operative infection at the nonunion incision site. He was treated with irrigation and debridement, IM nail retention, 6 weeks IV antibiotics, and oral suppression until union after which point all antibiotics were terminated. No mechanical nail failures were seen through a range of weight bearing loads with 11/14 patients bearing 50% body weight or greater immediately after surgery (Table 5).
Table 4
Diaphyseal vs. metaphyseal NU
Patient
Location
Distance from closest joint (mm)
Deformity after compression
Union
1
M
94 Knee
MPTA decreased 4°, PPTA decreased 3°
Y
2
D
270
None
Y
3
D
184
None
Y
4
D
172
None
Y
5
D
215
None
Y
6
M
74 Ankle
None
Y
7
D
153
None
Y
8
D
122
None
Y
9
D
209
None
Y
10
D
182
None
Y
11
D
290
None
Y
12
D
154
None
Y
13
D
233
None
Y
14
M
91 Knee
MPTA decreased 4°, PPTA decreased 18°
N
Mean
 
174.5
 
13/14
NU nonunion, M metaphyseal, D diaphyseal, MPTA medial proximal tibia angle, PPTA posterior proximal tibial angle
Table 5
Complications
Complication
N (Case #)
Management and outcome
Positive intra-op Cx
2 (Case #4 and 13)
Nail retained, 6 weeks IV ABX, oral suppression until union, no recurrence after union
Post-op infection
1 (Case #2)
Occurred 2 weeks post-op: nail retained, I&D performed, 6 weeks IV ABX, oral suppression until union, no recurrence after union
DVT
0
 
Implant failure
0
 
Nonunion
1 (Case #14)
Revised with hexapod circular external fixation
Compression-induced deformity
2 (Cases #1 and 14)
Case #1 had no intervention and resulted in malunion. Case #14 was revised and realigned
ABX antibiotics, I&D irrigation and debridement

Discussion

The MICN offers a novel, all-internal method to induce healing of recalcitrant long bone nonunions through sustained compression. The optimal indications for this technology will be culled over time for maximum impact and cost optimization. This study does not prove that compression nailing is superior to simple exchange nailing or that it prevents further surgery. If found to be more effective than simple exchange nailing with a classic static nail for complex nonunions in a future controlled trial, the MICN implant’s expense may be offset by the savings of avoiding additional surgery. A recent investigation showed that the same implant used for lengthening introduced no additional cost in the medical system when compared with lengthening over nail which required an additional ambulatory surgery [26]. When compared to a prolonged course wearing a hexapod circular fixator which averages 216 days [18], the MICN with WBAT ambulation may result in earlier return to work, and the cost of the two options may be equivalent; however, there is no data to date to confirm this. The risk of refracture may be less with a IM nail than after immediate removal of an external fixator.
The primary outcome measure was the MICN’s ability to unite nonunions of the femur and tibia. Thirteen of 14 patients united with this treatment over an average of 24.5 weeks. This cohort included many recalcitrant nonunions for which rapid consolidation was not expected. Patients were allowed to weight bear on the implant, and many returned to work by 6 weeks after surgery. Unlike circular fixation reconstructions, the time spent waiting for union was not as cumbersome for the patient. We expected to find a correlation between degrees of screw bending and time to union, but this study lacked the power for this analysis.
The mechanics of applying compression was studied in this group of patients. The goal was simply to compress the nonunion site until bending of the locking screws was seen on radiographs. While the average amount of compression applied to the nail with the ERC was 9.5 mm, the distance that the nail actually shortened was only 6.7 mm. This difference can be explained by the increasing resistance to shortening at the bone docking site resulting in a counter force that the gears in the nail were unable to overcome. Once the bone was well compressed, the external magnet could no longer activate the nail. There was no evidence that any additional attempts to compress the nail beyond this point damaged the implant in any way. In fact, in one patient (Case #4) the same nail was used for a subsequent lengthening without mechanical problems by creating an osteotomy around the nail after the nonunion was healed. The average distance the bone shortened at the nonunion site was 3.1 mm. With the nail shortening 6.7 mm and the bone moving only 3.1 mm, the difference represents the amount of screw bending which allowed the nail to continue shortening when the bone ends were already opposed. The average angle of screw bending was 2.5°. The bending of the screws is the only objective indication of the presence of compression at the bony interface. The degree to which a screw will bend is highly variable and is a function of not only the force applied to it but also: the length of the screw, whether the nail rests in the center of the screw or at one end of the screw, the diameter of the screw (5.0 mm vs. 4.0 mm), etc. A subjective sign of bone compression is pain. The patient will typically experience pain at the nonunion site during and after a compression adjustment with the external magnet. This can last from minutes to days. In one patient, the screws were not seen to bend (Case #13), yet additional ERC compression was painful, and the nail failed to shorten. The failure of the nail to shorten after applying ERC compression is due to strong resistance at the nonunion site, and is therefore another metric to confirm the presence of compression.
Based on this preliminary study, the types of nonunions that may be best suited for the MICN technique are femoral and tibia diaphyseal and distal tibial metaphyseal. Proximal and distal femoral metaphyseal nonunions were not evaluated, and we cannot comment on the use of this technique for these locations. Proximal tibia metaphyseal fractures are difficult to treat with IM nailing, and it follows that nonunions in this location are equally challenging. The compression nail is unique in that it exacerbates the tendency for the proximal fragment to bend medially leading to varus and flexion as the nail pulls the fragments together. The use of blocking screws posterior to the nail and medial to the nail near the nonunion site did not prevent this. Radiographic comparison taken pre- and post-compression showed migration of the nail at the entry point in the proximal tibia suggesting a possible explanation. Osteopenia would likely intensify this nail drift phenomena. The types of nonunions that will benefit from this method need further study.
The complications included one deep infection (Case #2) 2 weeks after conversion from a hexapod frame to the ICN despite a 1-month “frame holiday”. This was treated with open I&D, culture-specific antibiotics (6 weeks IV and 6 months PO), and nail retention. Two other patients were found to have positive intra-operative cultures at the time of nonunion bone debridement and nail insertion. These fracture-related infections [27] were treated with 6 weeks of culture-specific antibiotics, and then oral suppression until bony union. Amorosa et al. found that benign-appearing nonunions had a 28% chance of yielding positive cultures; fortunately, few of those required implant removal [28]. None of our patients required implant removal to control infection.
Limb length discrepancy (LLD) is an expected outcome of any compression technique. (Table 3) Three patients in this series have had subsequent limb lengthening surgery. In one case (Case #4), a corticotomy was made around the nail and the same implant was used. The regenerate was notably quite slow to form (BHI 2.1mo/cm). The patient who had the post-operative infection (Case #2) underwent osteoplasty lengthening 1 year later with a new IM lengthening nail. Although the implanted compression nail could have been used, it was removed due to concerns about chronic contamination from the previous deep infection. A third patient (Case #3) underwent lengthening osteoplasty in the distal femoral metaphysis which required removal of the antegrade compression nail and insertion of a retrograde lengthening nail. Most of the patients with LLD were too metabolically challenged, and therefore poor candidates, for lengthening surgery (Cases #1, 4, 6, 7, 8, 9, 12, and 14). They were treated with shoe lifts. Other healthy patients were simply elected to use a shoe lift (Cases #10 and 11). An alternative approach to nonunion repair and LLD that entails creation of an osteotomy adjacent to the nonunion site to redistribute stress and heal the nonunion was attempted in 1 patient. (Case #5) He underwent osteotomy adjacent to a stiff nonunion in the femur to induce healing at the nonunion site as well as to lengthen the femur through the new osteotomy site. The nail lengthening mode was used to reestablish limb length through the osteotomy. Although the adjacent nonunion was stiff enough that it did not distract during lengthening, unfortunately, it failed to unite. Once the lengthening regenerate bone was mature at the lengthening site, the nail was compressed to facilitate healing at the nonunion site resulting in a strong union.
Bone defects are a common sequela of nonunion surgery. If these defects can be acutely shortened (Cases #4, #9), then the MICN can be used to compress the repair site. The amount of shortening that can be done acutely depends on the soft tissue compliance and neurovascular function. At the time of this study, pre-distraction of the nail was limited to 15 mm so it could not gradually shorten a defect greater than a few millimeters. Recently, a “rapid distraction” tool has become available allowing for 8 cm of pre-distraction which can handle a gradual shortening of several centimeters followed by compression upon docking. How much shortening a patient can tolerate is patient specific. Young and active adult patients will use a shoe lift for LLD of less than 1 cm and will often want lengthening surgery for greater limb length inequality. This technique can be used in either scenario since a staged lengthening surgery can be performed in the same bone. Patients who are poor lengthening candidates can be treated with a shoe lift for any LLD.
Large bone defects may require bone transport rather than shortening which can now be accomplished with an all-internal approach using the novel bone transport nail (PRECICE, NuVasive Specialized Orthopedics, San Diego, CA, USA) or the plate-assisted bone transport technique [29]. Both of these transport methods rely on compression at the docking site, and the preliminary findings of the present study will improve the understanding of the mechanics of MICN compression universally.
The retrospective nature of this single cohort study is a limitation. We were unable to rigorously study the clinical judgement indices used to decide which patients are treated with this MICN technique and which were managed with a different approach. A randomized controlled trial is needed to prove any of the observations noted in this study. The trail would need to compare MICN to simple exchange nailing to show the actual value of sustained compression. Anecdotally, some of the patients studied (Cases #3, #4) lost compression early in treatment and had no progression of union over 3 months. Compression was added and the bones united quickly thereafter. The same trial would be needed to look at the cost of treatment between the two methods. The MICN would need to show a significantly lower time to union and incidence of nonunion to justify the expense of the implant, a result which may vary between the tibia and femur. Such a study is the only way to truly know which patients require the MICN technique.

Conclusion

The magnetic intramedullary compression nail has resulted in the union of several difficult to treat fracture nonunion cases. Confirming that enough compression has been applied can be inferred from the presence of locking bolt bending on radiographs. The ideal angle of screw bending needs to be further delineated and may be correlated to rate of consolidation, a finding that could not be proven from this small series. Proximal tibia metaphyseal nonunions may be ill-suited for this treatment due to inability to control deformity. Weight bearing as tolerated ambulation did not negatively impact the results and is our standard rehabilitation protocol. Compression can be administered in many patterns. We recommend establishing early compression and adding more as needed during follow-up. Staged lengthening with the same internal lengthening nail is a future option but may require slow distraction. The lessons learned from this experience will help us understand docking site compression when using a magnetic IM bone transport nail in the near future.

Compliance with ethical standards

Conflict of interest

No author has any conflict related to this study or the preparation of this manuscript. Dr. Austin Fragomen, MD is a consultant for Smith & Nephew, Synthes, and NuVasive and serves on the Board of the Limb Lengthening & Reconstruction Society. Dr. S Robert Rozbruch, MD is a consultant for Smith & Nephew, Stryker, and NuVasive and serves on the Board of the Limb Lengthening & Reconstruction Society. He receives royalty payments from Stryker. Dr David Wellman, MD has no conflict.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Zura R, Xiong Z, Einhorn T, Watson JT, Ostrum RF, Prayson MJ, Della Rocca GJ, Mehta S, McKinley T, Wang Z, Steen RG (2016) Epidemiology of fracture nonunion in 18 human bones. JAMA Surg 151(11):e162775CrossRef Zura R, Xiong Z, Einhorn T, Watson JT, Ostrum RF, Prayson MJ, Della Rocca GJ, Mehta S, McKinley T, Wang Z, Steen RG (2016) Epidemiology of fracture nonunion in 18 human bones. JAMA Surg 151(11):e162775CrossRef
2.
Zurück zum Zitat Zura R, Braid-Forbes MJ, Jeray K, Mehta S, Einhorn TA, Watson JT, Della Rocca GJ, Forbes K, Steen RG (2017) Bone fracture nonunion rate decreases with increasing age: a prospective inception cohort study. Bone 95:26–32CrossRef Zura R, Braid-Forbes MJ, Jeray K, Mehta S, Einhorn TA, Watson JT, Della Rocca GJ, Forbes K, Steen RG (2017) Bone fracture nonunion rate decreases with increasing age: a prospective inception cohort study. Bone 95:26–32CrossRef
3.
Zurück zum Zitat Hak DJ, Fitzpatrick D, Bishop JA, Marsh JL, Tilp S, Schnettler R, Simpson H, Alt V (2014) Delayed union and nonunions: epidemiology, clinical issues, and financial aspects. Injury 45(Suppl 2):S3–7CrossRef Hak DJ, Fitzpatrick D, Bishop JA, Marsh JL, Tilp S, Schnettler R, Simpson H, Alt V (2014) Delayed union and nonunions: epidemiology, clinical issues, and financial aspects. Injury 45(Suppl 2):S3–7CrossRef
4.
Zurück zum Zitat Patil S, Montgomery R (2006) Management of complex tibial and femoral nonunion using the Ilizarov technique, and its cost implications. J Bone Jt Surg Br. 88(7):928–932CrossRef Patil S, Montgomery R (2006) Management of complex tibial and femoral nonunion using the Ilizarov technique, and its cost implications. J Bone Jt Surg Br. 88(7):928–932CrossRef
5.
Zurück zum Zitat Brinker MR, Trivedi A, O'Connor DP (2017) Debilitating effects of femoral nonunion on health-related quality of life. J Orthop Trauma 31(2):e37–e42CrossRef Brinker MR, Trivedi A, O'Connor DP (2017) Debilitating effects of femoral nonunion on health-related quality of life. J Orthop Trauma 31(2):e37–e42CrossRef
6.
Zurück zum Zitat Schottel PC, O'Connor DP, Brinker MR (2015) Time trade-off as a measure of health-related quality of life: long bone nonunions have a devastating impact. J Bone Jt Surg Am 97(17):1406–1410CrossRef Schottel PC, O'Connor DP, Brinker MR (2015) Time trade-off as a measure of health-related quality of life: long bone nonunions have a devastating impact. J Bone Jt Surg Am 97(17):1406–1410CrossRef
7.
Zurück zum Zitat Brinker MR, O'Connor DP, Monla YT, Earthman TP (2007) Metabolic and endocrine abnormalities in patients with nonunions. J Orthop Trauma 21(8):557–570CrossRef Brinker MR, O'Connor DP, Monla YT, Earthman TP (2007) Metabolic and endocrine abnormalities in patients with nonunions. J Orthop Trauma 21(8):557–570CrossRef
8.
Zurück zum Zitat Konda SR, Christiano A, Fisher N, Leucht P, Egol KA (2017) Femoral nonunion with iliac crest bone graft. J Orthop Trauma 31(Suppl 3):S19–S20CrossRef Konda SR, Christiano A, Fisher N, Leucht P, Egol KA (2017) Femoral nonunion with iliac crest bone graft. J Orthop Trauma 31(Suppl 3):S19–S20CrossRef
9.
Zurück zum Zitat Ru J, Xu H, Kang W, Chang H, Niu Y, Zhao J (2016) Augmentative compression plating versus exchanging reamed nailing for nonunion of femoral shaft fracture after intramedullary nailing: a retrospective cohort study. Acta Orthop Belg 82(2):249–257PubMed Ru J, Xu H, Kang W, Chang H, Niu Y, Zhao J (2016) Augmentative compression plating versus exchanging reamed nailing for nonunion of femoral shaft fracture after intramedullary nailing: a retrospective cohort study. Acta Orthop Belg 82(2):249–257PubMed
10.
Zurück zum Zitat Somford MP, van den Bekerom MP, Kloen P (2013) Operative treatment for femoral shaft nonunions, a systematic review of the literature. Strategies Trauma Limb Reconstr. 8(2):77–88CrossRef Somford MP, van den Bekerom MP, Kloen P (2013) Operative treatment for femoral shaft nonunions, a systematic review of the literature. Strategies Trauma Limb Reconstr. 8(2):77–88CrossRef
11.
Zurück zum Zitat Swanson EA, Garrard EC, Bernstein DT, O'Connor DP, Brinker MR, (2015) Results of a systematic approach to exchange nailing for the treatment of aseptic femoral nonunions. J Orthop Trauma 29(1):21–27CrossRef Swanson EA, Garrard EC, Bernstein DT, O'Connor DP, Brinker MR, (2015) Results of a systematic approach to exchange nailing for the treatment of aseptic femoral nonunions. J Orthop Trauma 29(1):21–27CrossRef
12.
Zurück zum Zitat Litrenta J, Tornetta P 3rd, Vallier H, Firoozabadi R, Leighton R, Egol K, Kruppa C, Jones CB, Collinge C, Bhandari M, Schemitsch E, Sanders D, Mullis B (2015) Dynamizations and exchanges: success rates and indications. J Orthop Trauma 29(12):569–573CrossRef Litrenta J, Tornetta P 3rd, Vallier H, Firoozabadi R, Leighton R, Egol K, Kruppa C, Jones CB, Collinge C, Bhandari M, Schemitsch E, Sanders D, Mullis B (2015) Dynamizations and exchanges: success rates and indications. J Orthop Trauma 29(12):569–573CrossRef
13.
Zurück zum Zitat Buckwalter JA, Grodzinsky AJ (1999) Loading of healing bone, fibrous tissue, and muscle: implications for orthopaedic practice. J Am Acad Orthop Surg 7:291–299CrossRef Buckwalter JA, Grodzinsky AJ (1999) Loading of healing bone, fibrous tissue, and muscle: implications for orthopaedic practice. J Am Acad Orthop Surg 7:291–299CrossRef
14.
Zurück zum Zitat Gubin AV, Borzunov DY, Marchenkova LO, Malkova TA, Smirnova IL (2016) Contribution of GA Ilizarov to bone reconstruction: historical achievements and state of the art. Strateg Trauma Limb Reconstr 11(3):145–152CrossRef Gubin AV, Borzunov DY, Marchenkova LO, Malkova TA, Smirnova IL (2016) Contribution of GA Ilizarov to bone reconstruction: historical achievements and state of the art. Strateg Trauma Limb Reconstr 11(3):145–152CrossRef
15.
Zurück zum Zitat Fenton P, Qureshi F, Bejjanki N, Potter D (2011) Management of non-union of humeral fractures with the Stryker T2 compression nail. Arch Orthop Trauma Surg 131(1):79–84CrossRef Fenton P, Qureshi F, Bejjanki N, Potter D (2011) Management of non-union of humeral fractures with the Stryker T2 compression nail. Arch Orthop Trauma Surg 131(1):79–84CrossRef
16.
Zurück zum Zitat Huckstep RL (1986) The Huckstep intramedullary compression nail. Indications, technique, and results. Clin Orthop Relat Res 212:48–61 Huckstep RL (1986) The Huckstep intramedullary compression nail. Indications, technique, and results. Clin Orthop Relat Res 212:48–61
17.
Zurück zum Zitat Ferreira N, Marais LC, Aldous C (2015) Mechanobiology in the management of mobile atrophic and oligotrophic tibial nonunions. J Orthop 12(Suppl 2):S182–S187CrossRef Ferreira N, Marais LC, Aldous C (2015) Mechanobiology in the management of mobile atrophic and oligotrophic tibial nonunions. J Orthop 12(Suppl 2):S182–S187CrossRef
18.
Zurück zum Zitat Rozbruch SR, Pugsley J, Fragomen A, Ilizarov S (2008) Repair of tibial nonunions and bone defects with the Taylor spatial frame. J Orthop Trauma 22(2):88–95CrossRef Rozbruch SR, Pugsley J, Fragomen A, Ilizarov S (2008) Repair of tibial nonunions and bone defects with the Taylor spatial frame. J Orthop Trauma 22(2):88–95CrossRef
19.
Zurück zum Zitat Yin P, Ji Q, Li T, Li J, Li Z, Liu J, Wang G, Wang S, Zhang L, Mao Z, Tang P (2015) A systematic review and meta-analysis of ilizarov methods in the treatment of infected nonunion of tibia and femur. PLoS ONE 10(11):e0141973CrossRef Yin P, Ji Q, Li T, Li J, Li Z, Liu J, Wang G, Wang S, Zhang L, Mao Z, Tang P (2015) A systematic review and meta-analysis of ilizarov methods in the treatment of infected nonunion of tibia and femur. PLoS ONE 10(11):e0141973CrossRef
20.
Zurück zum Zitat Brinker MR, O'Connor DP (2003) Ilizarov compression over a nail for aseptic femoral nonunions that have failed exchange nailing: a report of five cases. J Orthop Trauma 17(10):668–676CrossRef Brinker MR, O'Connor DP (2003) Ilizarov compression over a nail for aseptic femoral nonunions that have failed exchange nailing: a report of five cases. J Orthop Trauma 17(10):668–676CrossRef
21.
Zurück zum Zitat Watson JT, Sanders RW (2017) Controlled compression nailing for at risk humeral shaft fractures. J Orthop Trauma 31(Suppl 6):S25–S28CrossRef Watson JT, Sanders RW (2017) Controlled compression nailing for at risk humeral shaft fractures. J Orthop Trauma 31(Suppl 6):S25–S28CrossRef
22.
Zurück zum Zitat Fragomen AT (2017) Transitioning to an intramedullary lengthening and compression nail. J Orthop Trauma 31(Suppl 2):S7–S13CrossRef Fragomen AT (2017) Transitioning to an intramedullary lengthening and compression nail. J Orthop Trauma 31(Suppl 2):S7–S13CrossRef
23.
Zurück zum Zitat Gustilo RB, Anderson JT (1976) Prevention of infection in the treatment of one thousand and twenty-five open fractures of long bones: retrospective and prospective analyses. J Bone Jt Surg Am 58:453–458CrossRef Gustilo RB, Anderson JT (1976) Prevention of infection in the treatment of one thousand and twenty-five open fractures of long bones: retrospective and prospective analyses. J Bone Jt Surg Am 58:453–458CrossRef
24.
Zurück zum Zitat McCarthy J, Iobst CA, Rozbruch SR, Sabharwal S, Eismann EA (2012) limb lengthening and reconstruction society AIM index reliably assesses lower limb deformity. Clin Orthop Relat Res 471(2):621–627CrossRef McCarthy J, Iobst CA, Rozbruch SR, Sabharwal S, Eismann EA (2012) limb lengthening and reconstruction society AIM index reliably assesses lower limb deformity. Clin Orthop Relat Res 471(2):621–627CrossRef
25.
Zurück zum Zitat Muthusamy S, Rozbruch SR, Fragomen AT (2016) The use of blocking screws with internal lengthening nail and reverse rule of thumbs for blocking screws in limb lengthening and deformity correction surgery. Strateg Trauma Limb Reconstruct 11(3):199–205CrossRef Muthusamy S, Rozbruch SR, Fragomen AT (2016) The use of blocking screws with internal lengthening nail and reverse rule of thumbs for blocking screws in limb lengthening and deformity correction surgery. Strateg Trauma Limb Reconstruct 11(3):199–205CrossRef
27.
Zurück zum Zitat Metsemakers WJ, Moregenstern M, McNally MA et al (2017) Fracture related infection: a consensus on definition from an international expert group. Injury 49(3):505–510 [Epub ahead of print]CrossRef Metsemakers WJ, Moregenstern M, McNally MA et al (2017) Fracture related infection: a consensus on definition from an international expert group. Injury 49(3):505–510 [Epub ahead of print]CrossRef
28.
Zurück zum Zitat Amorosa LF, Buirs LD, Bexkens R, Wellman DS, Kloen P, Lorich DG, Helfet DL (2013) A single-stage treatment protocol for presumptive aseptic diaphyseal nonunions: a review of outcomes. J Orthop Trauma 27(10):582–586CrossRef Amorosa LF, Buirs LD, Bexkens R, Wellman DS, Kloen P, Lorich DG, Helfet DL (2013) A single-stage treatment protocol for presumptive aseptic diaphyseal nonunions: a review of outcomes. J Orthop Trauma 27(10):582–586CrossRef
29.
Zurück zum Zitat Barinaga G, Beason AM, Gardner MP (2018) Novel surgical approach to segmental bone transport using a magnetic intramedullary limb lengthening system. J Am Acad Orthop Surg 26(22):e477–e482CrossRef Barinaga G, Beason AM, Gardner MP (2018) Novel surgical approach to segmental bone transport using a magnetic intramedullary limb lengthening system. J Am Acad Orthop Surg 26(22):e477–e482CrossRef
Metadaten
Titel
The PRECICE magnetic IM compression nail for long bone nonunions: a preliminary report
verfasst von
Austin T. Fragomen
David Wellman
S. Robert Rozbruch
Publikationsdatum
19.06.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 11/2019
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-019-03225-4

Weitere Artikel der Ausgabe 11/2019

Archives of Orthopaedic and Trauma Surgery 11/2019 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.