01.12.2010 | Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

The Use of a Cemented Dual Mobility Socket to Treat Recurrent Dislocation

verfasst von: Moussa Hamadouche, MD, PhD, David J. Biau, MD, Denis Huten, MD, Thierry Musset, MD, François Gaucher, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 12/2010

Einloggen, um Zugang zu erhalten

Abstract

Background

The treatment of recurrent dislocation after total hip arthroplasty remains challenging. Dual mobility sockets have been associated with a low rate of dislocation but it is not known whether they are useful for treating recurrent dislocation.

Questions/purposes

We therefore asked whether a cemented dual mobility socket would (1) restore hip stability following recurrent dislocation; (2) provide a pain-free and mobile hip; and (3) show durable radiographic fixation.

Methods

We retrospectively reviewed 51 patients treated with a cemented dual mobility socket for recurrent dislocation after total hip arthroplasty between August 2002 and June 2005. The mean age at the time of the index procedure of was 71.3 years. Of the 51 patients, 47 have had complete clinical and radiographic evaluation data at a mean followup of 51.4 months (range, 25–76.3 months).

Results

The cemented dual mobility socket restored complete stability of the hip in 45 of the 47 patients (96%). The mean Merle d’Aubigné hip score was 16 ± 2 at the latest followup. Radiographic analysis revealed no or radiolucent lines less than 1 mm thick located in a single acetabular zone in 43 of 47 hips (91.5%). The cumulative survival rate of the acetabular component at 72 months using revision for dislocation and/or mechanical failure as the end point was 96% ± 4% (95% confidence interval, 90%–100%).

Conclusions

A cemented dual mobility socket was able to restore hip stability in 96% of recurrent dislocating hips. However, longer-term followup is needed to ensure that dislocation and loosening rates will not increase.

Level of Evidence

Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants [in French]. Rev Chir Orthop. 2005;91:627–636.PubMed Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants [in French]. Rev Chir Orthop. 2005;91:627–636.PubMed
2.
Zurück zum Zitat Anderson MJ, Murray WR, Skinner HB. Constrained acetabular components. J Arthroplasty. 1994;9:17–23.CrossRefPubMed Anderson MJ, Murray WR, Skinner HB. Constrained acetabular components. J Arthroplasty. 1994;9:17–23.CrossRefPubMed
3.
Zurück zum Zitat Aubriot J, Lesimple P, Leclercq S. Study of Bousquet’s non-cemented acetabular implant in 100 hybrid total hip prostheses (Charnley type cemented femoral component). Average 5-year follow-up [in French]. Acta Orthop Belg. 1993;59(Suppl 1):267–271.PubMed Aubriot J, Lesimple P, Leclercq S. Study of Bousquet’s non-cemented acetabular implant in 100 hybrid total hip prostheses (Charnley type cemented femoral component). Average 5-year follow-up [in French]. Acta Orthop Belg. 1993;59(Suppl 1):267–271.PubMed
4.
Zurück zum Zitat Barrack RL, Mulroy RD, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br. 1992;74:385–389.PubMed Barrack RL, Mulroy RD, Harris WH. Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br. 1992;74:385–389.PubMed
5.
Zurück zum Zitat Berend KR, Lombardi AV, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty. 2005;20(Suppl 3):93–102.CrossRefPubMed Berend KR, Lombardi AV, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty examining the successes and failures. J Arthroplasty. 2005;20(Suppl 3):93–102.CrossRefPubMed
6.
Zurück zum Zitat Berry DJ. Unstable total hip arthroplasty: detailed overview. Instr Course Lect. 2001;50:265–274.PubMed Berry DJ. Unstable total hip arthroplasty: detailed overview. Instr Course Lect. 2001;50:265–274.PubMed
7.
Zurück zum Zitat Bousquet G, Gazielly DF, Giradin P, Debiesse JL, Relave M, Israeli A. The ceramic coated cementless total hip arthroplasty. Basic concepts and surgical technique. J Orthop Surg Tech. 1985;1:15–28. Bousquet G, Gazielly DF, Giradin P, Debiesse JL, Relave M, Israeli A. The ceramic coated cementless total hip arthroplasty. Basic concepts and surgical technique. J Orthop Surg Tech. 1985;1:15–28.
8.
Zurück zum Zitat Cates HE, Faris PM, Keating EM, Ritter MA. Polyethylene wear in cemented metal-backed acetabular cups. J Bone Joint Surg Br. 1993;75:249–253.PubMed Cates HE, Faris PM, Keating EM, Ritter MA. Polyethylene wear in cemented metal-backed acetabular cups. J Bone Joint Surg Br. 1993;75:249–253.PubMed
9.
Zurück zum Zitat Chandler RW, Dorr LD, Perry J. The functional cost of dislocation following total hip arthroplasty. Clin Orthop Relat Res. 1982;168:168–172.PubMed Chandler RW, Dorr LD, Perry J. The functional cost of dislocation following total hip arthroplasty. Clin Orthop Relat Res. 1982;168:168–172.PubMed
10.
Zurück zum Zitat Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH. Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty. 2003;18:827–833.CrossRefPubMed Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH. Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty. 2003;18:827–833.CrossRefPubMed
11.
Zurück zum Zitat Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am. 1992;74:1334–1343.PubMed Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am. 1992;74:1334–1343.PubMed
12.
Zurück zum Zitat De Lee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32. De Lee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.
13.
Zurück zum Zitat Della Valle CJ, Chang D, Sporer S, Berger RA, Rosenberg AG, Paprosky WG. High failure rate of a constrained acetabular liner in revision total hip arthroplasty. J Arthroplasty. 2005;20(Suppl 3):103–107.CrossRefPubMed Della Valle CJ, Chang D, Sporer S, Berger RA, Rosenberg AG, Paprosky WG. High failure rate of a constrained acetabular liner in revision total hip arthroplasty. J Arthroplasty. 2005;20(Suppl 3):103–107.CrossRefPubMed
14.
Zurück zum Zitat Farizon F, de LR, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup with dual mobility. A twelve-year follow-up study. Int Orthop. 1998;22:219–224.CrossRefPubMed Farizon F, de LR, Azoulai JJ, Bousquet G. Results with a cementless alumina-coated cup with dual mobility. A twelve-year follow-up study. Int Orthop. 1998;22:219–224.CrossRefPubMed
15.
Zurück zum Zitat Fisher DA, Kiley K. Constrained acetabular cup disassembly. J Arthroplasty. 1994;9:325–329.CrossRefPubMed Fisher DA, Kiley K. Constrained acetabular cup disassembly. J Arthroplasty. 1994;9:325–329.CrossRefPubMed
16.
Zurück zum Zitat Goetz DD, Bremner BR, Callaghan JJ, Capello WN, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A concise follow-up of a previous report. J Bone Joint Surg Am. 2004;86:2419–2423.PubMed Goetz DD, Bremner BR, Callaghan JJ, Capello WN, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A concise follow-up of a previous report. J Bone Joint Surg Am. 2004;86:2419–2423.PubMed
17.
Zurück zum Zitat Goetz DD, Capello WN, Callaghan JJ, Brown TD, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases. J Bone Joint Surg Am. 1998;80:502–509.PubMed Goetz DD, Capello WN, Callaghan JJ, Brown TD, Johnston RC. Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases. J Bone Joint Surg Am. 1998;80:502–509.PubMed
18.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
19.
Zurück zum Zitat Guyen O, Chen QS, Bejui-Hugues J, Berry DJ, An KN. Unconstrained tripolar hip implants: effect on hip stability. Clin Orthop Relat Res. 2007;455:202–208.CrossRefPubMed Guyen O, Chen QS, Bejui-Hugues J, Berry DJ, An KN. Unconstrained tripolar hip implants: effect on hip stability. Clin Orthop Relat Res. 2007;455:202–208.CrossRefPubMed
20.
Zurück zum Zitat Guyen O, Lewallen DG, Cabanela ME. Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am. 2008;90:1553–1560.CrossRefPubMed Guyen O, Lewallen DG, Cabanela ME. Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants. J Bone Joint Surg Am. 2008;90:1553–1560.CrossRefPubMed
21.
Zurück zum Zitat Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res. 2009;467:465–472.CrossRefPubMed Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res. 2009;467:465–472.CrossRefPubMed
22.
Zurück zum Zitat Guyen O, Pibarot V, Vaz G, Chevillotte C, Carret JP, Bejui-Hugues J. Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation. J Arthroplast. 2007;22:849–58. Guyen O, Pibarot V, Vaz G, Chevillotte C, Carret JP, Bejui-Hugues J. Unconstrained tripolar implants for primary total hip arthroplasty in patients at risk for dislocation. J Arthroplast. 2007;22:849–58.
23.
Zurück zum Zitat Harris WH, Penenberg BL. Further follow-up on socket fixation using a metal-backed acetabular component for total hip replacement. A minimum ten-year follow-up study. J Bone Joint Surg Am. 1987;69:1140–1143.PubMed Harris WH, Penenberg BL. Further follow-up on socket fixation using a metal-backed acetabular component for total hip replacement. A minimum ten-year follow-up study. J Bone Joint Surg Am. 1987;69:1140–1143.PubMed
24.
Zurück zum Zitat Kaper BP, Bernini PM. Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases. J Bone Joint Surg Am. 1998;80:561–565.PubMed Kaper BP, Bernini PM. Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases. J Bone Joint Surg Am. 1998;80:561–565.PubMed
25.
Zurück zum Zitat Kerboull M, Hamadouche M, Kerboull L. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res. 2000;378:155–168.CrossRefPubMed Kerboull M, Hamadouche M, Kerboull L. The Kerboull acetabular reinforcement device in major acetabular reconstructions. Clin Orthop Relat Res. 2000;378:155–168.CrossRefPubMed
26.
Zurück zum Zitat Khan RJ, Fick D, Alakeson R, Haebich S, de Cruz M, Nivbrant B, Wood D. A constrained acetabular component for recurrent dislocation. J Bone Joint Surg Br. 2006;88:870–876.CrossRefPubMed Khan RJ, Fick D, Alakeson R, Haebich S, de Cruz M, Nivbrant B, Wood D. A constrained acetabular component for recurrent dislocation. J Bone Joint Surg Br. 2006;88:870–876.CrossRefPubMed
27.
Zurück zum Zitat Krein SW, Chao EY. Biomechanics of bipolar hip endoprostheses. J Orthop Res. 1984;4:356–368.CrossRef Krein SW, Chao EY. Biomechanics of bipolar hip endoprostheses. J Orthop Res. 1984;4:356–368.CrossRef
28.
Zurück zum Zitat Langlais F, Ropars M, Gaucher F, Musset T, Chaix O. Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res. 2008;466:389–395.CrossRefPubMed Langlais F, Ropars M, Gaucher F, Musset T, Chaix O. Dual mobility cemented cups have low dislocation rates in THA revisions. Clin Orthop Relat Res. 2008;466:389–395.CrossRefPubMed
29.
Zurück zum Zitat Lautridou C, Lebel B, Burdin G, Vielpeau C. Survival of the cementless Bousquet dual mobility cup: Minimum 15-year follow-up of 437 total hip arthroplasties [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:731–739.PubMed Lautridou C, Lebel B, Burdin G, Vielpeau C. Survival of the cementless Bousquet dual mobility cup: Minimum 15-year follow-up of 437 total hip arthroplasties [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:731–739.PubMed
30.
Zurück zum Zitat Leclercq S, El Bidi S, Aubriot JH. Bousquet’s device in the treatment of recurrent dislocation of a total hip prosthesis. Apropos of 13 cases [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1995;81:389–394.PubMed Leclercq S, El Bidi S, Aubriot JH. Bousquet’s device in the treatment of recurrent dislocation of a total hip prosthesis. Apropos of 13 cases [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1995;81:389–394.PubMed
31.
Zurück zum Zitat Lecuire F, Benareau I, Rubini J, Basso M. [Intra-prosthetic dislocation of the Bousquet dual mobility socket] [in French].Rev Chir Orthop Reparatrice Appar Mot. 2004;90:249–255. Lecuire F, Benareau I, Rubini J, Basso M. [Intra-prosthetic dislocation of the Bousquet dual mobility socket] [in French].Rev Chir Orthop Reparatrice Appar Mot. 2004;90:249–255.
32.
Zurück zum Zitat Lombardi AV, Mallory TH, Kraus TJ, Vaughn BK. Preliminary report on the S-ROM constraining acetabular insert: a retrospective clinical experience. Orthopedics. 1991;14:297–303.PubMed Lombardi AV, Mallory TH, Kraus TJ, Vaughn BK. Preliminary report on the S-ROM constraining acetabular insert: a retrospective clinical experience. Orthopedics. 1991;14:297–303.PubMed
33.
Zurück zum Zitat Mantel N, Haenzel W. Statistical aspects of the analysis of data from retrospective studies of disease. J Natl Cancer Inst. 1959;22:719–748.PubMed Mantel N, Haenzel W. Statistical aspects of the analysis of data from retrospective studies of disease. J Natl Cancer Inst. 1959;22:719–748.PubMed
34.
Zurück zum Zitat McCarthy JC, Lee JA. Constrained acetabular components in complex revision total hip arthroplasty. Clin Orthop Relat Res. 2005;441:210–215.CrossRefPubMed McCarthy JC, Lee JA. Constrained acetabular components in complex revision total hip arthroplasty. Clin Orthop Relat Res. 2005;441:210–215.CrossRefPubMed
35.
Zurück zum Zitat Merle d’aubigné R. Numerical classification of the function of the hip [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1970;56:481–486. Merle d’aubigné R. Numerical classification of the function of the hip [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1970;56:481–486.
36.
Zurück zum Zitat Murray DW, Carr AJ, Bulstrode C. Survival analysis of joint replacements. J Bone Joint Surg Br. 1993;75:697–704.PubMed Murray DW, Carr AJ, Bulstrode C. Survival analysis of joint replacements. J Bone Joint Surg Br. 1993;75:697–704.PubMed
37.
Zurück zum Zitat Philippot R, Adam P, Farizon F, Fessy MH, Bousquet G. Survival of cementless dual mobility sockets: ten-year follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2006;92:326–331.PubMed Philippot R, Adam P, Farizon F, Fessy MH, Bousquet G. Survival of cementless dual mobility sockets: ten-year follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2006;92:326–331.PubMed
38.
Zurück zum Zitat Philippot R, Adam P, Reckhaus M, Delangle F, Verdot FX, Curvale G, Farizon F. Prevention of dislocation in total hip revision surgery using a dual mobility design [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2009;95:407–413. Philippot R, Adam P, Reckhaus M, Delangle F, Verdot FX, Curvale G, Farizon F. Prevention of dislocation in total hip revision surgery using a dual mobility design [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2009;95:407–413.
39.
Zurück zum Zitat Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F. The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop. 2009;33:927–932.CrossRefPubMed Philippot R, Camilleri JP, Boyer B, Adam P, Farizon F. The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years. Int Orthop. 2009;33:927–932.CrossRefPubMed
40.
Zurück zum Zitat Philippot R, Farizon F, Camilleri JP, Boyer B, Derhi G, Bonnan J, Fessy MH, Lecuire F. Survival of dual mobility socket with a mean 17 years follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:43–48.PubMed Philippot R, Farizon F, Camilleri JP, Boyer B, Derhi G, Bonnan J, Fessy MH, Lecuire F. Survival of dual mobility socket with a mean 17 years follow-up [in French]. Rev Chir Orthop Reparatrice Appar Mot. 2008;94:43–48.PubMed
41.
Zurück zum Zitat Ritter MA. The cemented acetabular component of a total hip replacement. All polyethylene versus metal backing. Clin Orthop Relat Res. 1995;311:69–75. Ritter MA. The cemented acetabular component of a total hip replacement. All polyethylene versus metal backing. Clin Orthop Relat Res. 1995;311:69–75.
42.
Zurück zum Zitat Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552, vii. Sanchez-Sotelo J, Berry DJ. Epidemiology of instability after total hip replacement. Orthop Clin North Am. 2001;32:543–552, vii.
43.
Zurück zum Zitat Sanchez-Sotelo J, Haidukewych GJ, Boberg C. Hospital cost of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2006;88:290–294.CrossRefPubMed Sanchez-Sotelo J, Haidukewych GJ, Boberg C. Hospital cost of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2006;88:290–294.CrossRefPubMed
44.
Zurück zum Zitat Shapiro G, Weiland D, Markel D, Padgett D, Sculco T, Pellicci P. The use of a constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003;18:250–258.CrossRefPubMed Shapiro G, Weiland D, Markel D, Padgett D, Sculco T, Pellicci P. The use of a constrained acetabular component for recurrent dislocation. J Arthroplasty. 2003;18:250–258.CrossRefPubMed
45.
Zurück zum Zitat Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2179–2183.PubMed Shrader MW, Parvizi J, Lewallen DG. The use of a constrained acetabular component to treat instability after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2179–2183.PubMed
46.
Zurück zum Zitat Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop Relat Res. 2004;420:122–129.CrossRefPubMed Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop Relat Res. 2004;420:122–129.CrossRefPubMed
47.
Zurück zum Zitat Tufescu TV, Dust W. Failure of a new constrained acetabular insert: a report of 2 cases. J Arthroplasty. 2004;19:238–239. CrossRefPubMed Tufescu TV, Dust W. Failure of a new constrained acetabular insert: a report of 2 cases. J Arthroplasty. 2004;19:238–239. CrossRefPubMed
48.
Zurück zum Zitat Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed
49.
Zurück zum Zitat Yun AG, Padgett D, Pellicci P, Dorr LD. Constrained acetabular liners: mechanisms of failure. J Arthroplasty. 2005;20:536–541.CrossRefPubMed Yun AG, Padgett D, Pellicci P, Dorr LD. Constrained acetabular liners: mechanisms of failure. J Arthroplasty. 2005;20:536–541.CrossRefPubMed
Metadaten
Titel
The Use of a Cemented Dual Mobility Socket to Treat Recurrent Dislocation
verfasst von
Moussa Hamadouche, MD, PhD
David J. Biau, MD
Denis Huten, MD
Thierry Musset, MD
François Gaucher, MD
Publikationsdatum
01.12.2010
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 12/2010
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1404-7

Weitere Artikel der Ausgabe 12/2010

Clinical Orthopaedics and Related Research® 12/2010 Zur Ausgabe

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Constrained Liner in Neurologic or Cognitively Impaired Patients Undergoing Primary THA

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.