Open Access 03.03.2018 | Original Research

User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors

verfasst von: Guido Freckmann, Nina Jendrike, Annette Baumstark, Stefan Pleus, Christina Liebing, Cornelia Haug

Erschienen in: Diabetes Therapy | Ausgabe 2/2018

Abstract

Introduction

The international standard ISO 15197:2013 requires a user performance evaluation to assess if intended users are able to obtain accurate blood glucose measurement results with a self-monitoring of blood glucose (SMBG) system. In this study, user performance was evaluated for four SMBG systems on the basis of ISO 15197:2013, and possibly related insulin dosing errors were calculated. Additionally, accuracy was assessed in the hands of study personnel.

Methods

Accu-Chek® Performa Connect (A), Contour® plus ONE (B), FreeStyle Optium Neo (C), and OneTouch Select® Plus (D) were evaluated with one test strip lot. After familiarization with the systems, subjects collected a capillary blood sample and performed an SMBG measurement. Study personnel observed the subjects’ measurement technique. Then, study personnel performed SMBG measurements and comparison measurements. Number and percentage of SMBG measurements within ± 15 mg/dl and ± 15% of the comparison measurements at glucose concentrations < 100 and ≥ 100 mg/dl, respectively, were calculated. In addition, insulin dosing errors were modelled.

Results

In the hands of lay-users three systems fulfilled ISO 15197:2013 accuracy criteria with the investigated test strip lot showing 96% (A), 100% (B), and 98% (C) of results within the defined limits. All systems fulfilled minimum accuracy criteria in the hands of study personnel [99% (A), 100% (B), 99.5% (C), 96% (D)]. Measurements with all four systems were within zones of the consensus error grid and surveillance error grid associated with no or minimal risk. Regarding calculated insulin dosing errors, all 99% ranges were between dosing errors of − 2.7 and + 1.4 units for measurements in the hands of lay-users and between − 2.5 and + 1.4 units for study personnel. Frequent lay-user errors were not checking the test strips’ expiry date and applying blood incorrectly.

Conclusions

Data obtained in this study show that not all available SMBG systems complied with ISO 15197:2013 accuracy criteria when measurements were performed by lay-users.

Trial Registration

The study was registered at ClinicalTrials.gov (NCT02916576).

Funding

Ascensia Diabetes Care Deutschland GmbH.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s13300-018-0392-6) contains supplementary material, which is available to authorized users.

Enhanced content

To view enhanced content for this article go to https://​doi.​org/​10.​6084/​m9.​figshare.​5909179.

Introduction

Accurate measurement results obtained with blood glucose (BG) monitoring systems intended for self-testing by diabetes patients are important for correct insulin dosing calculations and adequate therapeutic decisions [14]. Several self-monitoring of blood glucose (SMBG) measurements per day are recommended especially for patients on intensive insulin regimens [5]. Regular SMBG is also recommended for patients with less intensive insulin therapy [6].
Previous accuracy evaluations often analyzed accuracy of SMBG systems in the hands of professionals who are trained in the use of SMBG systems [710]. However, accuracy achieved by professionals does not necessarily reflect accuracy achieved by lay-users, i.e., people with diabetes [1114]. The International Organization for Standardization’s (ISO) standard ISO 15197:2013, clause 8 [15], requires a user performance evaluation to assess if intended users are able to obtain accurate BG measurement results with an SMBG system. ISO 15197:2013 was harmonized with the regulations of the European Union as EN ISO 15197:2015 [16]. This harmonization had no impact on the requirements and procedures in ISO 15197:2013; changes were made to the foreword and an informative annex. According to ISO 15197 clause 8, a system’s user performance shall be evaluated with one test strip lot applying the following acceptance criteria: at least 95% of SMBG measurement results shall fall within ± 15 mg/dl (0.83 mmol/l) of comparison measurement results at glucose concentrations < 100 mg/dl (5.55 mmol/l) and within ± 15% at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
In this study, accuracy of four SMBG systems in the hands of intended users was evaluated on the basis of testing procedures and acceptance criteria of ISO 15197:2013, clause 8. Accuracy was also evaluated in the hands of trained personnel. Potential insulin dosing errors were calculated by using a model described in detail by Baumstark et al. [17] in order to estimate possible influences of SMBG accuracy on glycemic control.

Methods

The study was performed between September and October 2016 at the Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm (IDT), Germany. The study was conducted in compliance with the German Medical Devices Act, approved by the responsible Ethics Committee in Stuttgart, Germany and exempted from approval by the German Federal Institute for Drugs and Medical Devices. All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the 1964 Declaration of Helsinki, as revised in 2013. Informed consent was obtained from all patients for being included in the study.

Subjects

To obtain at least 100 evaluable data sets, in total 158 subjects were enrolled. Table 1 shows demographic data for subjects included in data analysis. Included subjects did not use the investigated SMBG systems for the last 3 years according to their own statement.
Table 1
Subjects demographics
Included subjects
System A
System B
System C
System D
n = 100
n = 100
n = 100
n = 100
Male, female (n)
54, 46
60, 40
59, 41
58, 42
Mean age (min–max) (years)
60.8 (29–80)
60.3 (27–79)
60.7 (25–80)
61.7 (25–80)
Diabetes type 1, 2 (n)
36, 64
35, 65
38, 62
36, 64
Education level (n):
 Secondary education grade
70
66
68
67
 University entrance qualification
11
17
17
17
 University degree
19
17
15
16
Time since diabetes diagnosis, mean (min–max) (years)
17.8 (1–66)
18 (1–55)
19.3 (1–66)
19.3 (1–66)
Subjects performing at least 1 SMBG measurement per week (n)
95
94
95
94
For each system, 100 different subjects were included in data analysis
Before study participation, a study physician reviewed the subjects’ self-reported medical history and medication, the inclusion and exclusion criteria for study participation (e.g., pregnancy or lactation period, severe acute disease, and/or chronic disease), and checked for interfering substances given in the manufacturers’ labelling.

Test Systems

In this study, four SMBG systems were evaluated (Table 2): Accu-Chek® Performa Connect (system A), Contour® plus ONE (system B), FreeStyle Optium Neo (system C), and OneTouch Select® Plus (system D). All systems were CE-marked. System B is a new system; systems A, C and D are available in Eastern Europe and Asia. All systems displayed plasma-equivalent glucose concentrations in milligrams per deciliter (mg/dl).
Table 2
System characteristics
System
Manufacturer’s comparison method
Enzyme test strips
Measurement conditions
Manufacturer
#
Name
Test strip lot (expiry date)
Temperature (°C)
Humidity (%)
Hematocrit (%)
A
Accu-Chek Performa Connect
475098 (2017-09)
Hexokinase
Glucose dehydrogenase
8–44
10–90
10–65
Roche Diabetes Care GmbH, Mannheim, Germany
B
Contour plus ONE
DP6ELHD01C (2018-05)
Glucose oxidase
Glucose dehydrogenase
5–45
10–93
0–70
Bayer Consumer Care AG, Basel, Switzerland
C
FreeStyle Optium Neo
45001 67624 (2017-12)
Glucose oxidase
Glucose dehydrogenase
15–40
10–90
15–65
Abbott Diabetes Care Ltd., Witney, Oxon, UK
D
OneTouch Select Plus
3997403 (2017-08)
Glucose oxidase
Glucose oxidase
10–44
10–90
30–55
LifeScan Europe, Division of Cilag GmbH International Zug, Switzerland
Meters and test strips of system A were purchased from a pharmacy in Austria. For system B, five meters and test strips were purchased from a pharmacy in Poland; 115 meters were provided by the manufacturer. Meters of system C were bought online from amazon.co.uk, and test strips were purchased from a pharmacy in the UK. Meters and test strips of system D were purchased from a pharmacy in Germany.
Meters and test strips were stored, maintained, and used in compliance with the manufacturer’s labelling. Control measurements according to the manufacturer’s labelling were performed daily prior to the test procedure and for each test strip vial to ensure the proper function of each system.
Since, for systems B and D, the same meter was used for the familiarization and the measurement procedure (as described in Sect. “Test Procedure”), control measurements were also performed retrospectively after the measurement procedure to confirm the proper function of the system.

Instructions for Use

According to ISO 15197:2013, clause 8, subjects shall be given the instructions for use and any training material that are routinely provided with the SMBG system.
For system A, C, and D, instructions for use routinely provided with the SMBG system were available. For system B, instructions for use were provided by the manufacturer. Since not all instructions for use were available in German, relevant information, i.e., preparation and performance of BG measurements, possible error sources, and error messages, was summarized in German in a comparable layout for each system.

Comparison Measurement Methods

Comparison measurements were performed with a glucose oxidase (GOD) method (YSI 2300 STAT Plus glucose analyzer, YSI Incorporated, Yellow Springs, OH, USA) and a hexokinase (HK) method (Cobas Integra® 400 plus, Roche Instrument Center, Rotkreuz, Switzerland). Comparison measurements were performed in duplicate with both comparison methods. Both methods provided BG values in milligrams per deciliter (mg/dl). For the YSI 2300 STAT Plus glucose analyzer and the Cobas Integra 400 plus instruments, conformity to traceability requirements of ISO 17511 [18] was confirmed by the manufacturers. Trueness and precision of the two analyzers were verified by regular internal and external quality control measures as required by the German national guideline (Rili-BÄK) [19]. In addition, daily quality control measurements were performed applying IDT internal standard operating procedures by using National Institute of Standards and Technology (NIST) standard reference material (SRM) 965b. The manufacturer’s comparison measurement method and/or the method used by the manufacturer for SMBG accuracy evaluation as indicated in the manufacturer’s labelling is shown in Table 2.

Test Procedure

The test procedure was performed on the basis of ISO 15197:2013, clause 8, with one test strip lot for each system. For each subject, testing procedures were performed with three different SMBG systems; the evaluation of one system was completed before the evaluation of another system began. The number of SMBG systems used by the same subject was limited to three because the procedures were very time-consuming and attention-demanding for individual subjects. To minimize possible order effects, the systems were rotated according to a predefined pattern. The familiarization procedure (as described below) was performed with a designated training meter and the measurement procedure (as described below) with a designated test meter. Test meters were cleaned and disinfected after each subject. According to the systems’ labelling, system B and D might be infectious after use even if they were disinfected. For these systems, every subject got their own meter for the familiarization and the measurement procedure.
For each subject, the capillary blood’s hematocrit value was checked to be within the range indicated in the SMBG system’s labelling. For this purpose, samples were collected in heparinized capillaries and the capillaries were centrifuged. Hematocrit values were determined by using an alignment chart.

Familiarization with SMBG Systems

As requested by ISO 15197:2013, each subject had the opportunity to review the instructions for use and to practice testing with a given system in a manner that represents how lay-users learn to use a new SMBG system. For this purpose, subjects were allowed to review the instructions for use and to perform up to three training measurements with control solution.
No additional instructions, training, and assistance were provided to the subjects. In addition, the familiarization procedure was supervised by study personnel to prevent any influence on a subject, e.g., by other subjects.

Measurements

Measurements were performed in a controlled laboratory setting (temperature, 19.8–24.3 °C; humidity, 28.5–56.4%). Each subject was allowed to use the SMBG system’s instructions for use. Measurements were supervised by trained study personnel.
Each subject washed and dried their hands before the measurement procedure. Then, the subject collected his/her own capillary blood sample from a fingertip by skin puncture and performed a measurement using the test meter. For hygiene reasons, the skin puncture was performed with a single-use lancing device. For systems A, B, and D, study personnel removed each test strip from the vial and put it on the closed vial. For system C each test strip was packaged separately and was removed from the package by the subject. Test strip vials or packages were changed after approximately 10 subjects.
After the measurement with a given test system, subjects were asked if they thought that they performed the measurement correctly. Subjects repeated the measurement up to three times if they reported making a mistake or if no valid result was obtained. The first quantitative measurement result for which a subject did not report making a mistake in the measurement procedure was included in the accuracy analysis.
After the SMBG measurement by the subject, study personnel collected a capillary blood sample from the subject’s fingertip and performed measurements in duplicate with the same SMBG system, consecutively using the subject’s test meter and an additional meter. Before and after these duplicate SMBG measurements, samples for comparison measurements (first and second comparison measurement) were collected in lithium heparin tubes, aliquots were centrifuged within 7 min, and measurements were performed on separated plasma.

Human Factors

The subjects’ measurement technique was observed and documented by study personnel with regard to the manufacturer’s labelling, i.e., preparation of the measurement, insertion of test strips, blood sampling, application of blood, and mistakes in device handling. Errors documented only by the study personnel but not by the subjects did not lead to exclusion from accuracy analysis.

Data Analysis

Data Exclusions

Data were excluded from analysis for the following reasons: the subject reported making a mistake in his/her measurement procedure and the measurement was repeated; the number of 100 valid measurement results was already reached; the subject obtained no valid result and the measurement was repeated; hemolysis in plasma samples for comparison measurements, errors regarding comparison measurements; quality control measurement results obtained with the comparison method were outside predefined limits or no valid measurement result was obtained; the difference between the first and second comparison measurements exceeded the acceptance criteria for sample stability (≤ 4 mg/dl at glucose concentrations < 100 mg/dl or ≤ 4% at glucose concentrations ≥ 100 mg/dl).

User Performance Evaluation

For each system, data from 100 subjects (n = 100) were included in the evaluation. Accuracy was evaluated by comparing the SMBG system’s measurement result with the respective comparison measurement result (first comparison measurement) according to ISO 15197:2013, clause 8, by calculating the number and percentage of results within ± 15 mg/dl of the mean comparison measurement result at BG concentrations < 100 mg/dl and within ± 15% at BG concentrations ≥ 100 mg/dl. In addition, the percentage of results within the clinically acceptable zones A and B of the consensus error grid (CEG) and the bias (systematic measurement difference) according to Bland and Altman [20] were calculated.
Data are also presented in radar plots [21, 22], an alternative approach to visualize the measurement accuracy of an SMBG system, and in surveillance error grids (SEG) [23], which is an alternative approach to assess the clinical risk associated with SMBG systems’ accuracy.

Evaluation of Human Factors

For each system, the occurrence of user errors in performing the measurement procedure was analyzed.

Evaluation of Accuracy in the Hands of Study Personnel

For each system, data from 100 subjects (n = 200, duplicate measurements) were included in the evaluation. Accuracy was evaluated with one test strip lot for each system by comparing the study personnel’s SMBG measurement result with the respective mean comparison measurement result (first and second comparison measurement). The same parameters as for lay-users were calculated and are presented as described above.

Calculation of Possibly Related Insulin Dosing Errors

All data included in the accuracy evaluation (lay-users n = 100, study personnel n = 200) were assessed regarding possibly related insulin doses based on the model described in detail by Baumstark et al. [17]. In this model, only short-term insulin doses (i.e., meal-related doses and doses for BG correction), but not long-term (basal) insulin doses were covered.
The model utilizes the mealtime insulin dosing formula used by the Diabetes Teaching Center at the University of California San Francisco (UCSF), San Francisco, CA [24]. Insulin doses were calculated for comparison method results and SMBG system results and the following fixed parameters: 60 g carbohydrate intake, 1 unit per 15 g insulin-to-carb ratio, a target BG value of 100 mg/dL, and an insulin sensitivity of 25 mg/dL per unit.

Results

Accuracy in the Hands of Lay-users

Accuracy results for each system when evaluated against the manufacturer’s comparison method are presented in Table 3, Fig. 1 (difference plot and radar plot), and Fig. 2 (SEG). Results when evaluated against the alternative comparison method are provided in Supplemental Table 1 and Supplemental Fig. 1).
Table 3
Accuracy results based on ISO 15197:2013 criteria, results within consensus error grid (CEG) zones A and B, results within surveillance error grid zones with “no risk” or “slight lower risk”, relative bias according to Bland and Altman, and calculated insulin dosing error
System
User
Results within
Relative bias
Calculated insulin dosing error
±15 mg/dl/± 15%
CEG zones A + B
SEG zones
No risk
Slight, lower risk
Median (50th percentile)
99% ranges between the 0.5th and 99.5th percentile
Risk score
0–0.5
> 0.5–1.0
%
%
n
%
Units
A
Lay-users
96
100
100
0
− 7.6
− 0.9
− 2.7 to − 0.1
Study personnel
99
100
197
3
− 6.8
− 0.8
− 2.4 to 0.0
B
Lay-users
100
100
100
0
− 2.2
− 0.3
− 1.0 to + 0.3
Study personnel
100
100
200
0
− 2.2
− 0.2
− 1.0 to + 0.2
C
Lay-users
98
100
97
3
+ 2.8
+ 0.3
− 0.6 to + 1.4
Study personnel
99.5
100
198
2
+ 1.6
+ 0.2
− 0.7 to +1.4
D
Lay-users
92
100
97
3
− 5.6
− 0.6
− 2.4 to +0.6
Study personnel
96
100
193
7
− 5.3
− 0.7
− 2.5 to + 0.3
Results are shown for measurements performed by lay-users (n = 100) and by study personnel (N = 200) when evaluated by using the manufacturer’s comparison method (hexokinase for system A, glucose oxidase for systems B, C, and D)
Three of the four systems fulfilled ISO 15197:2013 accuracy criteria in the hands of lay-users with the investigated test strip lot showing 96% (A), 100% (B), and 98% (C) of results within ± 15 mg/dl and ± 15% of the comparison method results at BG concentrations < 100 mg/dl and ≥ 100 mg/dl, respectively. One system (D) did not fulfill ISO 15197:2013 accuracy criteria with the investigated test strip lot showing 92% of results within the limits (± 15 mg/dl and ± 15%). All four systems showed 100% of results within CEG zones A and B. Regarding the SEG, all measurements with systems A and B were within the “no risk” zone (risk score 0–0.5). The relative bias ranged from − 7.6% (system A) to + 2.8% (system C) with a smallest bias of − 2.2% (system B). Median modelled insulin dosing errors ranged from − 0.9 units (system A) to + 0.3 units (system C) with all 99% ranges between dosing errors of − 2.7 and + 1.4 units (Table 3, Fig. 3).

Human Factors

Frequently observed lay-user errors were not checking the test strip’s expiry date and incorrect blood application, e.g., application of an insufficient blood amount, pressing the fingertip onto the test strip, or applying the blood incorrectly onto the test field (Table 4). Because all test strips were used in this study before the expiry date, this error had no impact on measurement results.
Table 4
Human factors: number of errors in the measurement technique of 100 lay-users observed by study personnel
 
Number of lay-user errors
System A
System B
System C
System D
Preparation of the measurement—testa strip’s expiry date was not checked
90
97
87
101
Insertion of test strips—e.g., not inserted correctly, kinked, bent
7
4
5
4
Blood sampling—e.g., blood drop too small, too old, smeared
3
3
11
8
Blood application—blood was applied incorrectly, e.g., insufficient amount, fingertip was pressed onto the test strip; blood was not applied onto the test field correctly
17
20
38
35
Device handling—e.g., blood application before meter was ready for use, test strip was removed before completion of the measurement, fingertip was not removed from the test trip after the signal tone
8
6
12
7
Number of incorrectly performed measurements as reported by subjects
4
2
17
7
aOnly non-expired test strips were used in this study; therefore, not checking the test strip’s expiry date had no impact on measurement results

Accuracy in the Hands of Study Personnel

Accuracy results for one lot of each system when evaluated against the manufacturer’s comparison method are presented in Table 3 and Fig. 1 (difference plot and radar plot). Results when evaluated against the alternative comparison method are provided in Supplemental Table 1 and Supplemental Fig. 1.
All four systems fulfilled ISO 15197:2013 accuracy criteria in the hands of study personnel with the investigated test strip lot showing 99% (A), 100% (B), 99.5% (C), and 96% (D) of results within ± 15 mg/dl and ± 15% of the comparison method results at BG concentrations < 100 mg/dl and ≥ 100 mg/dl, respectively. All four systems showed 100% of results within CEG zones A and B. Regarding the SEG, all measurements with system B were within the “no risk” zone (risk score 0–0.5). The relative bias ranged from − 6.8% (system A) to + 1.6% (system C) with a smallest bias of + 1.6% (system C). Median modelled insulin dosing errors ranged from − 0.8 units (system A) to + 0.2 units (system C) all with 99% ranges between dosing errors of − 2.5 and + 1.4 units.

Discussion

In this study, system accuracy of four SMBG systems in the hands of lay-users, i.e., diabetes patients, was evaluated with one test strip lot on the basis of testing procedures and accuracy criteria of ISO 15197:2013, clause 8. Additionally, accuracy in the hands of study personnel trained in operating the SMBG systems was evaluated with one test strip lot for each of the four systems. According to ISO 15197, user performance evaluation shall demonstrate that intended users are able to obtain accurate measurement results when operating the SMBG system, given only the instructions for use and training materials routinely provided with the systems.
Three out of four systems fulfilled accuracy criteria in the hands of lay-users with the evaluated test strip lot showing 96% (system A), 100% (system B), and 98% (system C) of results within ± 15 mg/dl for BG concentrations < 100 mg/dl and ± 15% for BG concentrations ≥ 100 mg/dl. All systems fulfilled ISO 15197:2013 accuracy criteria in the hands of trained study personnel. One system (B) showed 100% of results in the hand of lay-users and study personnel, and similar accuracy results for this system were obtained in a previous user performance evaluation [25].
The other three systems showed slightly better accuracy results for study personnel measurements compared to lay-user measurements. Different studies showed that accuracy achieved by trained study personnel is often better than accuracy achieved by lay-persons [11, 12, 26, 27]; however, only few user performance evaluations were performed on the basis of procedures described in ISO 15197:2013 [25, 26, 2830].
For all four systems, all measurements were within zones of the CEG and SEG that are associated with no or minimal risk, irrespective of the user group. ISO 15197:2013, clause 6.3, requires a CEG analysis for system accuracy evaluation. The CEG is developed for type 1 diabetes patients and is based on data collected in 1994 [31]. The CEG is divided into eight zones indicating the estimated risk to the patient. The SEG also assesses the clinical risk but is based on much more recent survey data [23]. The SEG is designed as an assessment tool of SMBG systems’ postmarket risk. SEG risk zones are color-coded and data points are assigned to continuous individual risk scores.
For both comparison measurement methods used in our study, compliance with traceability requirements was confirmed by the manufacturers. In this study, the comparison measurement method/system applied had no impact as to whether the data obtained comply with ISO 15197:2013 accuracy criteria or not. However, differences in accuracy data depending on the applied comparison method/system were found which is consistent with results of two previous ISO 15197-based studies performed at our institute [9, 32]. This influence might be reduced if manufacturers would adjust calibration of their reference methods [33].
This study design has several limitations. According to ISO 15197:2013, the user performance evaluation shall also include the evaluation of the instructions for use and the messages displayed on the meter by a questionnaire in order to assess whether the instructions and messages are adequate and easy to understand. Since not all instructions for use were available in German, subjects received a summarized instruction for use (in German) for each system with relevant information for the preparation and performance of BG measurements. Therefore, instructions for use and the messages displayed on the respective meters were not evaluated. In this study, only one lot was evaluated on the basis of ISO 15197:2013, clause 8. According to ISO 15197:2013, clause 6.3, system accuracy evaluation in the hands of study personnel shall be performed with three test strip lots, since lot-to-lot variations between multiple test strip lots used for the same system can markedly affect a system’s accuracy results [11, 34, 35].
Regarding the subjects’ measurement technique, not checking the expiry date of test strips before the measurement was the most frequent error. It might be that subjects assumed that they were only provided with non-expired test strips. Because test strips were used in this study before their expiry date, this error had no impact on measurement results. However, different studies showed that the use of deteriorated test strips can affect SMBG measurement results [36, 37].
For all four systems, an incorrect blood application, e.g., applying an insufficient amount of blood, pressing the fingertip onto the test strip, or applying the blood not onto the test field, was the most frequent error during the measurement procedure. Since incorrect blood application can potentially affect SMBG measurement results, adequate SMBG training and education should be an integral part of diabetes therapy [36, 38].
Accurate SMBG measurement results are crucial for adequate insulin dosing decisions by patients on insulin therapy. Modelling analyses showed that inaccurate BG measurements can lead to insulin dosing errors [1, 3, 4, 39, 40] which can adversely affect glycemic control and increase the risk of long-term complications in patients on intensive insulin therapy. In this study, negative median insulin dosing errors (too small insulin doses) were calculated for three systems (A, B, D) and positive median insulin dosing errors (too high insulin doses) were calculated for one system (C). Except for one system (D), median calculated insulin dosing errors were larger when SMBG measurements were performed by lay-persons. In addition, the four systems showed differences in the width of the error range (99% ranges between the 0.5th percentile and the 99.5th percentile).

Conclusions

In this user performance evaluation, three out of four systems fulfilled ISO 15197:2013 accuracy criteria with the investigated test strip lot. However, the systems showed slight differences in the number of results within ISO 15197:2013 accuracy limits. All four systems fulfilled ISO 15197:2013 accuracy criteria with the investigated test strip lot in the hands of study personnel, with three systems showing better accuracy results when measurements were performed by trained study personnel. One system (B) showed 100% of results within the ISO 15197:2013 accuracy limits in the hands of lay-users and in the hands of study personnel. In both user groups, calculated median insulin dosing errors were within ± 0.9 units and tended to be larger when measurements were performed by lay-users. Inaccurate SMBG measurements can result in insulin dosing errors and adversely affect glycemic control. Therefore, manufacturers should provide high quality SMBG systems that are easy to use and resistant to user errors.

Acknowledgments

Funding

This investigator-initiated trial and the medical writing were supported by Ascensia Diabetes Care Deutschland GmbH, Germany. Article processing charges were funded by the authors’ institute Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm (IDT), Ulm.

Medical Writing, Editorial, and Other Assistance

We would like to thank the study personnel and other employees of the IDT who conducted the study or helped in preparing the manuscript. We would like to thank Scott Pardo, PhD, PStat®, for his assistance in modeling insulin dosing errors.

Authorship

All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this manuscript, take responsibility for the integrity of the work as a whole, and have given final approval for the version to be published. All authors had full access to all of the data in this study and take complete responsibility for the integrity of the data and accuracy of the data analysis.

Previous Presentations

Parts of this study’s results were presented at the 77th Scientific Sessions of the American Diabetes Association in San Diego, CA, June 9–13, 2017 and at the International Diabetes Federation’s congress in Abu Dhabi, UAE, December 4–8, 2017.

Thanking Patients

We thank the participants of the study.

Disclosures

Guido Freckmann is general manager of the IDT, which carries out clinical studies on the evaluation of BG meters and medical devices for diabetes therapy on its own initiative and on behalf of various companies. Guido Freckmann/IDT have received speakers’ honoraria or consulting fees from Abbott, Ascensia, Bayer, LifeScan, Menarini Diagnostics, Novo Nordisk, Roche, Sanofi, Sensile, and Ypsomed. Nina Jendrike is an employee of the IDT. Annette Baumstark is an employee of the IDT. Stefan Pleus is an employee of the IDT. Christina Liebing is an employee of the IDT. Cornelia Haug is an employee of the IDT.

Compliance with Ethics Guidelines

All procedures followed were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the 1964 Declaration of Helsinki, as revised in 2013. Informed consent was obtained from all patients for being included in the study.

Data Availability

The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.

Open Access

This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://​creativecommons.​org/​licenses/​by/​4.​0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Boyd JC, Bruns DE. Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose. Clin Chem. 2001;47:209–14.PubMed Boyd JC, Bruns DE. Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose. Clin Chem. 2001;47:209–14.PubMed
2.
Zurück zum Zitat Raine CH III, Schrock LE, Edelman SV, et al. Significant insulin dose errors may occur if blood glucose results are obtained from miscoded meters. J Diabetes Sci Technol. 2007;1:205–10.CrossRefPubMedPubMedCentral Raine CH III, Schrock LE, Edelman SV, et al. Significant insulin dose errors may occur if blood glucose results are obtained from miscoded meters. J Diabetes Sci Technol. 2007;1:205–10.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Breton MD, Kovatchev BP. Impact of blood glucose self-monitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: an in silico study. J Diabetes Sci Technol. 2010;4:562–70.CrossRefPubMedPubMedCentral Breton MD, Kovatchev BP. Impact of blood glucose self-monitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: an in silico study. J Diabetes Sci Technol. 2010;4:562–70.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat ADA. Standards of medical care in diabetes—2017. Diabetes Care. 2017;40:1–135.CrossRef ADA. Standards of medical care in diabetes—2017. Diabetes Care. 2017;40:1–135.CrossRef
6.
Zurück zum Zitat Elgart JF, Gonzalez L, Prestes M, Rucci E, Gagliardino JJ. Frequency of self-monitoring blood glucose and attainment of HbA1c target values. Acta Diabetol. 2016;53:57–62.CrossRefPubMed Elgart JF, Gonzalez L, Prestes M, Rucci E, Gagliardino JJ. Frequency of self-monitoring blood glucose and attainment of HbA1c target values. Acta Diabetol. 2016;53:57–62.CrossRefPubMed
7.
Zurück zum Zitat Klaff LJ, Brazg R, Hughes K, et al. Accuracy evaluation of contour next compared with five blood glucose monitoring systems across a wide range of blood glucose concentrations occurring in a clinical research setting. Diabetes Technol Ther. 2015;17:8–15.CrossRefPubMed Klaff LJ, Brazg R, Hughes K, et al. Accuracy evaluation of contour next compared with five blood glucose monitoring systems across a wide range of blood glucose concentrations occurring in a clinical research setting. Diabetes Technol Ther. 2015;17:8–15.CrossRefPubMed
8.
Zurück zum Zitat Brazg RL, Klaff LJ, Sussman AM. New generation blood glucose monitoring system exceeds international accuracy standards. J Diabetes Sci Technol. 2016;10:1414–5.CrossRefPubMedPubMedCentral Brazg RL, Klaff LJ, Sussman AM. New generation blood glucose monitoring system exceeds international accuracy standards. J Diabetes Sci Technol. 2016;10:1414–5.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Freckmann G, Link M, Schmid C, et al. System accuracy evaluation of different blood glucose monitoring systems following ISO 15197:2013 by using two different comparison methods. Diabetes Technol Ther. 2015;17:635–48.CrossRefPubMed Freckmann G, Link M, Schmid C, et al. System accuracy evaluation of different blood glucose monitoring systems following ISO 15197:2013 by using two different comparison methods. Diabetes Technol Ther. 2015;17:635–48.CrossRefPubMed
10.
Zurück zum Zitat Hasslacher C, Kulozik F, Platten I. Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting. J Diabetes Sci Technol. 2014;8:466–72.CrossRefPubMedPubMedCentral Hasslacher C, Kulozik F, Platten I. Analytical performance of glucose monitoring systems at different blood glucose ranges and analysis of outliers in a clinical setting. J Diabetes Sci Technol. 2014;8:466–72.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Kristensen GB, Monsen G, Skeie S, Sandberg S. Standardized evaluation of nine instruments for self-monitoring of blood glucose. Diabetes Technol Ther. 2008;10:467–77.CrossRefPubMed Kristensen GB, Monsen G, Skeie S, Sandberg S. Standardized evaluation of nine instruments for self-monitoring of blood glucose. Diabetes Technol Ther. 2008;10:467–77.CrossRefPubMed
12.
Zurück zum Zitat Littmann K, Petersen ER, Pussinen C, et al. Evaluation of OneTouch Verio(®), a new blood glucose self-monitoring system for patients with diabetes. Scand J Clin Lab Invest. 2013;73:286–92.CrossRefPubMed Littmann K, Petersen ER, Pussinen C, et al. Evaluation of OneTouch Verio(®), a new blood glucose self-monitoring system for patients with diabetes. Scand J Clin Lab Invest. 2013;73:286–92.CrossRefPubMed
13.
Zurück zum Zitat Bailey TS, Wallace JF, Parkes JL, et al. Performance of a new blood glucose monitoring system in the hands of intended users. Diabetes Technol Ther. 2012;14:783–9.CrossRefPubMed Bailey TS, Wallace JF, Parkes JL, et al. Performance of a new blood glucose monitoring system in the hands of intended users. Diabetes Technol Ther. 2012;14:783–9.CrossRefPubMed
14.
Zurück zum Zitat Solvik UO, Risa M, Jacobsen CE, Monsen G, Sandberg S. Performance of 10 systems for self-monitoring of blood glucose by trained healthcare professionals and in the hands of the users. Clin Chem. 2015;61:772–4.CrossRefPubMed Solvik UO, Risa M, Jacobsen CE, Monsen G, Sandberg S. Performance of 10 systems for self-monitoring of blood glucose by trained healthcare professionals and in the hands of the users. Clin Chem. 2015;61:772–4.CrossRefPubMed
15.
Zurück zum Zitat International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013.
16.
Zurück zum Zitat International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus (ISO 15197:2013). EN ISO 15197:2015. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus (ISO 15197:2013). EN ISO 15197:2015.
17.
Zurück zum Zitat Baumstark A, Jendrike N, Pleus S, Haug C, Freckmann G. Evaluation of accuracy of six blood glucose monitoring systems and modeling of possibly related insulin dosing errors. Diabetes Technol Ther. 2017;19:580–8.CrossRefPubMed Baumstark A, Jendrike N, Pleus S, Haug C, Freckmann G. Evaluation of accuracy of six blood glucose monitoring systems and modeling of possibly related insulin dosing errors. Diabetes Technol Ther. 2017;19:580–8.CrossRefPubMed
18.
Zurück zum Zitat International Organization for Standardization. In vitro diagnostic medical devices—measurement of quantities in biological samples—metrological traceability of values assigned to calibrators and control materials. EN ISO 17511:2003. International Organization for Standardization. In vitro diagnostic medical devices—measurement of quantities in biological samples—metrological traceability of values assigned to calibrators and control materials. EN ISO 17511:2003.
19.
Zurück zum Zitat Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Deutsches Ärzteblatt. 2014;111:1583-618. Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Deutsches Ärzteblatt. 2014;111:1583-618.
20.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
22.
Zurück zum Zitat Pardo S, Dunne N, Simmons DA. Using radar plots to demonstrate the accuracy and precision of 6 blood glucose monitoring systems. J Diabetes Sci Technol. 2017;11:966–9.CrossRefPubMed Pardo S, Dunne N, Simmons DA. Using radar plots to demonstrate the accuracy and precision of 6 blood glucose monitoring systems. J Diabetes Sci Technol. 2017;11:966–9.CrossRefPubMed
25.
Zurück zum Zitat Bailey TS, Wallace JF, Pardo S, et al. Accuracy and user performance evaluation of a new, wireless-enabled blood glucose monitoring system that links to a smart mobile device. J Diabetes Sci Technol. 2017;11:736–43.CrossRefPubMedPubMedCentral Bailey TS, Wallace JF, Pardo S, et al. Accuracy and user performance evaluation of a new, wireless-enabled blood glucose monitoring system that links to a smart mobile device. J Diabetes Sci Technol. 2017;11:736–43.CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Freckmann G, Baumstark A, Jendrike N, et al. Accuracy evaluation of four blood glucose monitoring systems in the hands of intended users and trained personnel based on ISO 15197 requirements. Diabetes Technol Ther. 2017;19:246–54.CrossRefPubMed Freckmann G, Baumstark A, Jendrike N, et al. Accuracy evaluation of four blood glucose monitoring systems in the hands of intended users and trained personnel based on ISO 15197 requirements. Diabetes Technol Ther. 2017;19:246–54.CrossRefPubMed
27.
Zurück zum Zitat Skeie S, Thue G, Nerhus K, Sandberg S. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician. Clin Chem. 2002;48:994–1003.PubMed Skeie S, Thue G, Nerhus K, Sandberg S. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician. Clin Chem. 2002;48:994–1003.PubMed
28.
Zurück zum Zitat Christiansen M, Greene C, Pardo S, et al. A new, wireless-enabled blood glucose monitoring system that links to a smart mobile device: accuracy and user performance evaluation. J Diabetes Sci Technol. 2017;11:567–73.CrossRefPubMedPubMedCentral Christiansen M, Greene C, Pardo S, et al. A new, wireless-enabled blood glucose monitoring system that links to a smart mobile device: accuracy and user performance evaluation. J Diabetes Sci Technol. 2017;11:567–73.CrossRefPubMedPubMedCentral
29.
Zurück zum Zitat Bailey T, Wallace JF, Greene C, et al. Accuracy and user performance evaluation of the Contour(®) Next Link 2.4 blood glucose monitoring system. Clin Chim Acta. 2015;448:139–45.CrossRefPubMed Bailey T, Wallace JF, Greene C, et al. Accuracy and user performance evaluation of the Contour(®) Next Link 2.4 blood glucose monitoring system. Clin Chim Acta. 2015;448:139–45.CrossRefPubMed
30.
Zurück zum Zitat Caswell M, Frank J, Viggiani MT, et al. Accuracy and user performance evaluation of a blood glucose monitoring system. Diabetes Technol Ther. 2015;17:152–8.CrossRefPubMed Caswell M, Frank J, Viggiani MT, et al. Accuracy and user performance evaluation of a blood glucose monitoring system. Diabetes Technol Ther. 2015;17:152–8.CrossRefPubMed
31.
Zurück zum Zitat Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 2000;23:1143–8.CrossRefPubMed Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 2000;23:1143–8.CrossRefPubMed
32.
Zurück zum Zitat Link M, Schmid C, Pleus S, et al. System accuracy evaluation of four systems for self-monitoring of blood glucose following ISO 15197 using a glucose oxidase and a hexokinase-based comparison method. J Diabetes Sci Technol. 2015;9:1041–50.CrossRefPubMedPubMedCentral Link M, Schmid C, Pleus S, et al. System accuracy evaluation of four systems for self-monitoring of blood glucose following ISO 15197 using a glucose oxidase and a hexokinase-based comparison method. J Diabetes Sci Technol. 2015;9:1041–50.CrossRefPubMedPubMedCentral
33.
Zurück zum Zitat Bailey TS, Klaff LJ, Wallace JF, et al. Fundamental importance of reference glucose analyzer accuracy for evaluating the performance of blood glucose monitoring systems (BGMSs). J Diabetes Sci Technol. 2016;10:872–5.CrossRefPubMedPubMedCentral Bailey TS, Klaff LJ, Wallace JF, et al. Fundamental importance of reference glucose analyzer accuracy for evaluating the performance of blood glucose monitoring systems (BGMSs). J Diabetes Sci Technol. 2016;10:872–5.CrossRefPubMedPubMedCentral
34.
Zurück zum Zitat Baumstark A, Pleus S, Schmid C, et al. Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197. J Diabetes Sci Technol. 2012;6:1076–86.CrossRefPubMedPubMedCentral Baumstark A, Pleus S, Schmid C, et al. Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197. J Diabetes Sci Technol. 2012;6:1076–86.CrossRefPubMedPubMedCentral
35.
Zurück zum Zitat Kristensen GB, Christensen NG, Thue G, Sandberg S. Between-lot variation in external quality assessment of glucose: clinical importance and effect on participant performance evaluation. Clin Chem. 2005;51:1632–6.CrossRefPubMed Kristensen GB, Christensen NG, Thue G, Sandberg S. Between-lot variation in external quality assessment of glucose: clinical importance and effect on participant performance evaluation. Clin Chem. 2005;51:1632–6.CrossRefPubMed
36.
Zurück zum Zitat Müller UA, Hämmerlein A, Casper A, Schultz M. Community pharmacy-based intervention to improve self-monitoring of blood glucose in type 2 diabetic patients. Pharmacy Pract (Granada). 2006;4:195–203. Müller UA, Hämmerlein A, Casper A, Schultz M. Community pharmacy-based intervention to improve self-monitoring of blood glucose in type 2 diabetic patients. Pharmacy Pract (Granada). 2006;4:195–203.
37.
Zurück zum Zitat Bamberg R, Schulman K, MacKenzie M, Moore J, Olchesky S. Effect of adverse storage conditions on performance of glucometer test strips. Clin Lab Sci. 2005;18:203–9.PubMed Bamberg R, Schulman K, MacKenzie M, Moore J, Olchesky S. Effect of adverse storage conditions on performance of glucometer test strips. Clin Lab Sci. 2005;18:203–9.PubMed
38.
Zurück zum Zitat Bergenstal R, Pearson J, Cembrowski GS, et al. Identifying variables associated with inaccurate self-monitoring of blood glucose: proposed guidelines to improve accuracy. Diabetes Educ. 2000;26:981–9.CrossRefPubMed Bergenstal R, Pearson J, Cembrowski GS, et al. Identifying variables associated with inaccurate self-monitoring of blood glucose: proposed guidelines to improve accuracy. Diabetes Educ. 2000;26:981–9.CrossRefPubMed
39.
Zurück zum Zitat Bedini JL, Wallace JF, Pardo S, Petruschke T. Performance evaluation of three blood glucose monitoring systems using ISO 15197: 2013 accuracy criteria, consensus and surveillance error grid analyses, and insulin dosing error modeling in a hospital setting. J Diabetes Sci Technol. 2015;10:85–92.CrossRefPubMedPubMedCentral Bedini JL, Wallace JF, Pardo S, Petruschke T. Performance evaluation of three blood glucose monitoring systems using ISO 15197: 2013 accuracy criteria, consensus and surveillance error grid analyses, and insulin dosing error modeling in a hospital setting. J Diabetes Sci Technol. 2015;10:85–92.CrossRefPubMedPubMedCentral
40.
Zurück zum Zitat Pardo S, Dunne N, Simmons DA. Bolus insulin dose error distributions based on results from two clinical trials comparing blood glucose monitoring systems. J Diabetes Sci Technol. 2017;11:970–4.CrossRefPubMed Pardo S, Dunne N, Simmons DA. Bolus insulin dose error distributions based on results from two clinical trials comparing blood glucose monitoring systems. J Diabetes Sci Technol. 2017;11:970–4.CrossRefPubMed
Metadaten
Titel
User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors
verfasst von
Guido Freckmann
Nina Jendrike
Annette Baumstark
Stefan Pleus
Christina Liebing
Cornelia Haug
Publikationsdatum
03.03.2018
Verlag
Springer Healthcare
Erschienen in
Diabetes Therapy / Ausgabe 2/2018
Print ISSN: 1869-6953
Elektronische ISSN: 1869-6961
DOI
https://doi.org/10.1007/s13300-018-0392-6

Weitere Artikel der Ausgabe 2/2018

Diabetes Therapy 2/2018 Zur Ausgabe

Practical Approach

Diabetes in the Elderly

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.