Skip to main content
Erschienen in: Osteoporosis International 4/2004

01.04.2004 | Original Article

Variations in diagnostic performances of dual-energy X-ray absorptiometry in the northwest of The Netherlands

verfasst von: Klaas P. Staal, Jan C. Roos, Radu A. Manoliu, Piet J. Kostense, Paul Lips, and the Densitometry Study Group

Erschienen in: Osteoporosis International | Ausgabe 4/2004

Einloggen, um Zugang zu erhalten

Abstract

Between-center variation in bone densitometry may influence the frequency of the diagnosis of osteoporosis. To evaluate this problem, dual-energy X-ray absorptiometry (DXA) machines of the medical centers in the northwest of The Netherlands were evaluated. Four phantoms were used to test the 17 DXA machines of 16 participating centers. Each phantom was measured 10 times and the data were analyzed on the corresponding DXA machine using the software delivered by the manufacturer. The analyses were done with the reference population as used in daily practice. There were DXA devices of seven different brands and types, using four different reference populations for the lumbar spine and seven for the hip. The observed differences in bone mineral densities (BMD) were up to 0.20 g/cm2 for the lumbar spine, 0.15 g/cm2 for the femoral neck, and 0.12 g/cm2 for the total hip. The coefficients of variation (CV) of the repeated phantom measurements ranged between 0.3% and 1.3% for the lumbar spine, 1.6% and 4.6% for the femoral neck, and 0.3% and 0.9% for the total hip. The mean female T-scores of 10 phantom measurements differed up to 0.6 SD between the DXA machines for the lumbar spine and up to 0.8 SD for the total hip. Mathematically, replacing a Hologic 2000 DXA machine with a newer type of the same brand (a Hologic 4500) caused a shift in diagnosis from osteoporosis to osteopenia of +1.1% for the lumbar spine and −4.5% for the total hip. When the Hologic 2000 was replaced by a Hologic 4500 with NHANES reference values, the shift from osteoporosis to osteopenia was also +1.1% for the lumbar spine, and −13.4% for the total hip. The clinical impact of the observed differences is difficult to estimate. One may conclude that the differences of the tested DXA devices are partly based on differences in DXA machines, but for the most part on the use of different reference populations. It is recommended to standardize the reference population, although the consequent shift in diagnosis will be confusing for physicians and patients, and adaptation of the reference values on the DXA devices of different brands with different technical qualities and measurement specifications will be difficult.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Consensus Development Conference (1991) Diagnosis, prophylaxis and treatment of osteoporosis. Am J Med 90:170–210 Consensus Development Conference (1991) Diagnosis, prophylaxis and treatment of osteoporosis. Am J Med 90:170–210
2.
Zurück zum Zitat Blake GM, Fogelman I (2001) Bone densitometry and the diagnosis of osteoporosis. Semin Nucl Med 31:69–81PubMed Blake GM, Fogelman I (2001) Bone densitometry and the diagnosis of osteoporosis. Semin Nucl Med 31:69–81PubMed
3.
Zurück zum Zitat Cummings SR, Black DM, Nevitt MC et al (1993) Bone density at various sites for prediction of hip fractures. Lancet 341:72–75PubMed Cummings SR, Black DM, Nevitt MC et al (1993) Bone density at various sites for prediction of hip fractures. Lancet 341:72–75PubMed
4.
Zurück zum Zitat WHO (1994) Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. Technical Report Series, No. 843. World Health Organization, Geneva WHO (1994) Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. Technical Report Series, No. 843. World Health Organization, Geneva
5.
Zurück zum Zitat Marshall D, Johnell O, Wedel H (1996) Meta-analysis of how well measures of bone mineral of bone mineral density predict occurrence of osteoporotic fractures. Br Med J 312:1254–1259 Marshall D, Johnell O, Wedel H (1996) Meta-analysis of how well measures of bone mineral of bone mineral density predict occurrence of osteoporotic fractures. Br Med J 312:1254–1259
6.
Zurück zum Zitat Lees B, Garland SW, Walton C, Stevenson JC (1997) Evaluation of the European spine phantom in a multi center trial. Osteoporos Int 7:570–574PubMed Lees B, Garland SW, Walton C, Stevenson JC (1997) Evaluation of the European spine phantom in a multi center trial. Osteoporos Int 7:570–574PubMed
7.
Zurück zum Zitat Cawte SA, Pearson D, Green DJ, Maslanka WB, Miller CG, Rogers AT (1999) Cross-calibration, precision and patient dose measurements in preparation for clinical trials using dual X-ray absorptiometry of the lumbar spine. Br J Radiol 72:354–362PubMed Cawte SA, Pearson D, Green DJ, Maslanka WB, Miller CG, Rogers AT (1999) Cross-calibration, precision and patient dose measurements in preparation for clinical trials using dual X-ray absorptiometry of the lumbar spine. Br J Radiol 72:354–362PubMed
8.
Zurück zum Zitat Chen Z, Maricic M, Lund P, Tesser J, Gluck O (1998) How the new Hologic reference values affect the densitometric diagnosis of osteoporosis. Osteoporos Int 8:423–427PubMed Chen Z, Maricic M, Lund P, Tesser J, Gluck O (1998) How the new Hologic reference values affect the densitometric diagnosis of osteoporosis. Osteoporos Int 8:423–427PubMed
9.
Zurück zum Zitat Petley GW, Cotton AM, Murrils AJ, Taylor PA, Cooper C, Cawley MID, Wilkin TJ (1996) Reference ranges of bone mineral density for women in southern England: the impact of local data on the diagnosis of osteoporosis. Br J Radiol 69:655–660PubMed Petley GW, Cotton AM, Murrils AJ, Taylor PA, Cooper C, Cawley MID, Wilkin TJ (1996) Reference ranges of bone mineral density for women in southern England: the impact of local data on the diagnosis of osteoporosis. Br J Radiol 69:655–660PubMed
10.
Zurück zum Zitat Kalender WA, Felsenberg D, Genant HK, Fischer M, Dequeker J, Reeve J (1995) The European Spine Phantom: a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT. Eur J Radiol 20:83–92CrossRefPubMed Kalender WA, Felsenberg D, Genant HK, Fischer M, Dequeker J, Reeve J (1995) The European Spine Phantom: a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT. Eur J Radiol 20:83–92CrossRefPubMed
11.
Zurück zum Zitat Looker AC, Wahner HW, Dunn WL, Calvo MS, Harris TB, Heyse SP, Johnston CC Jr, Lindsay RL (1995) Proximal femur bone mineral levels of US adults. Osteoporos Int 5:389–409PubMed Looker AC, Wahner HW, Dunn WL, Calvo MS, Harris TB, Heyse SP, Johnston CC Jr, Lindsay RL (1995) Proximal femur bone mineral levels of US adults. Osteoporos Int 5:389–409PubMed
12.
Zurück zum Zitat Genant H, Grammp S, Gluer C et al (1994) Universal standardization for dual X-ray absorptiometry: patient and phantom cross-calibration results. J Bone Miner Res 9:1503–1514PubMed Genant H, Grammp S, Gluer C et al (1994) Universal standardization for dual X-ray absorptiometry: patient and phantom cross-calibration results. J Bone Miner Res 9:1503–1514PubMed
13.
Zurück zum Zitat Pocock NA, Sambrook PN, Nguyen T, Kelly P, Freund J, Eisman JA (1992) Assessment of spinal and femoral bone density by dual X-ray absorptiometry: comparison of Lunar and Hologic instruments. J Bone Miner Res 7:1081–1084PubMed Pocock NA, Sambrook PN, Nguyen T, Kelly P, Freund J, Eisman JA (1992) Assessment of spinal and femoral bone density by dual X-ray absorptiometry: comparison of Lunar and Hologic instruments. J Bone Miner Res 7:1081–1084PubMed
14.
Zurück zum Zitat Barthe N, Braillon P, Ducassou, Basse-Cathilinat (1997) Comparison of two DXA systems (QDR 1000 and QDR 4500/A). Br J Radiol 70:728–739 Barthe N, Braillon P, Ducassou, Basse-Cathilinat (1997) Comparison of two DXA systems (QDR 1000 and QDR 4500/A). Br J Radiol 70:728–739
15.
Zurück zum Zitat Lu Y, Fuerst T, Hui S, Genant HK (2001) Standardization of bone mineral density at femoral neck, trochanter and Ward’s triangle. Osteoporos Int 12:438–444PubMed Lu Y, Fuerst T, Hui S, Genant HK (2001) Standardization of bone mineral density at femoral neck, trochanter and Ward’s triangle. Osteoporos Int 12:438–444PubMed
16.
Zurück zum Zitat Simmons A. Barrington SF, O’Doherty MJ, Coakley AJ (1995) DEXA normal reference ranges on the assesment of bone density. Br J Radiol 68:809 Simmons A. Barrington SF, O’Doherty MJ, Coakley AJ (1995) DEXA normal reference ranges on the assesment of bone density. Br J Radiol 68:809
17.
Zurück zum Zitat Ahmed AIH, Blake GM, Rymer JM, Fogelman I (1997) Screening for osteopenia and osteoporosis: do the accepted normal ranges lead to overdiagnosis? Osteoporos Int 7:432–438CrossRefPubMed Ahmed AIH, Blake GM, Rymer JM, Fogelman I (1997) Screening for osteopenia and osteoporosis: do the accepted normal ranges lead to overdiagnosis? Osteoporos Int 7:432–438CrossRefPubMed
18.
Zurück zum Zitat Laskey MA, Crisp AJ, Cole TJ, Compston JE (1992) Comparison of the effect of different reference data in Lunar DPX and Hologic QDR 1000 dual X-ray absorptiometers. Br J Radiol 65:1124–1129PubMed Laskey MA, Crisp AJ, Cole TJ, Compston JE (1992) Comparison of the effect of different reference data in Lunar DPX and Hologic QDR 1000 dual X-ray absorptiometers. Br J Radiol 65:1124–1129PubMed
Metadaten
Titel
Variations in diagnostic performances of dual-energy X-ray absorptiometry in the northwest of The Netherlands
verfasst von
Klaas P. Staal
Jan C. Roos
Radu A. Manoliu
Piet J. Kostense
Paul Lips
and the Densitometry Study Group
Publikationsdatum
01.04.2004
Verlag
Springer-Verlag
Erschienen in
Osteoporosis International / Ausgabe 4/2004
Print ISSN: 0937-941X
Elektronische ISSN: 1433-2965
DOI
https://doi.org/10.1007/s00198-003-1541-x

Weitere Artikel der Ausgabe 4/2004

Osteoporosis International 4/2004 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.