Vertebroplasty with self-locking hexagonal metal implants shows comparable primary and secondary stiffness to PMMA cement augmentation techniques in a biomechanical vertebral compression fracture model

verfasst von: W. Schmoelz, A. C. Disch, J. F. Huber

Erschienen in: European Spine Journal | Ausgabe 6/2010

Einloggen, um Zugang zu erhalten

Abstract

With the growing incidence of vertebral compression fractures in elderly patients having a fair overall health condition, minimal-invasive treatment techniques are getting in focus of surgical therapy. Cement augmentation is widely performed and its complications and mechanical limitations are well described. Implants avoiding the side effects of cement augmentation while reaching the same level of stability would be desirable. The primary and secondary stability of a new augmentation method with self-locking hexagonal metal implants were investigated and compared with the performance of established augmentation options. 18 fresh-frozen human spinal specimens (Th12–L2/L3–L5) were tested with pure moments of 7.5 Nm in a six-degree-of-freedom spine simulator to investigate primary and secondary stability of three augmentation techniques: (1) vertebroplasty, (2) PMMA filled cavity and (3) hexagonal metal implants. An increasing three-step cyclic loading model was included. Elastic displacement and height loss under loading did not show significant differences between the three test groups. Investigation of primary and secondary stability evenly demonstrated comparable results for all techniques indicating an insufficiency to stabilise the fracture with higher load cycles. The newly introduced method for augmentation with the metal implant Spine Pearls achieved comparable results to bone cement based techniques in a biomechanical in vitro study. Midterm and longterm reduction preservation and ingrowth of the implants have to be proven in further studies.
Literatur
1.
Zurück zum Zitat Belkoff SM, Mathis JM, Jasper LE, Deramond H (2001) The biomechanics of vertebroplasty. The effect of cement volume on mechanical behavior. Spine 26:1537–1541CrossRefPubMed Belkoff SM, Mathis JM, Jasper LE, Deramond H (2001) The biomechanics of vertebroplasty. The effect of cement volume on mechanical behavior. Spine 26:1537–1541CrossRefPubMed
2.
Zurück zum Zitat Bergot C, Laval-Jeantet AM, Hutchinson K, Dautraix I, Caulin F, Genant HK (2001) A comparison of spinal quantitative computed tomography with dual energy X-ray absorptiometry in European women with vertebral and nonvertebral fractures. Calcif Tissue Int 68:74–82CrossRefPubMed Bergot C, Laval-Jeantet AM, Hutchinson K, Dautraix I, Caulin F, Genant HK (2001) A comparison of spinal quantitative computed tomography with dual energy X-ray absorptiometry in European women with vertebral and nonvertebral fractures. Calcif Tissue Int 68:74–82CrossRefPubMed
3.
Zurück zum Zitat Boger A, Heini P, Windolf M, Schneider E (2007) Adjacent vertebral failure after vertebroplasty: a biomechanical study of low-modulus PMMA cement. Eur Spine J 16:2118–2125CrossRefPubMed Boger A, Heini P, Windolf M, Schneider E (2007) Adjacent vertebral failure after vertebroplasty: a biomechanical study of low-modulus PMMA cement. Eur Spine J 16:2118–2125CrossRefPubMed
4.
Zurück zum Zitat Cheung KM, Lu WW, Luk KD, Wong CT, Chan D, Shen JX, Qiu GX, Zheng ZM, Li CH, Liu SL, Chan WK, Leong JC (2005) Vertebroplasty by use of a strontium-containing bioactive bone cement. Spine 30:S84–S91CrossRefPubMed Cheung KM, Lu WW, Luk KD, Wong CT, Chan D, Shen JX, Qiu GX, Zheng ZM, Li CH, Liu SL, Chan WK, Leong JC (2005) Vertebroplasty by use of a strontium-containing bioactive bone cement. Spine 30:S84–S91CrossRefPubMed
5.
Zurück zum Zitat Colon-Emeric CS, Saag KG (2006) Osteoporotic fractures in older adults. Best Pract Res Clin Rheumatol 20:695–706CrossRefPubMed Colon-Emeric CS, Saag KG (2006) Osteoporotic fractures in older adults. Best Pract Res Clin Rheumatol 20:695–706CrossRefPubMed
6.
Zurück zum Zitat Ferguson SJ, Winkler F, Nolte LP (2002) Anterior fixation in the osteoporotic spine: cut-out and pullout characteristics of implants. Eur Spine J 11:527–534CrossRefPubMed Ferguson SJ, Winkler F, Nolte LP (2002) Anterior fixation in the osteoporotic spine: cut-out and pullout characteristics of implants. Eur Spine J 11:527–534CrossRefPubMed
7.
Zurück zum Zitat Fourney DR, Schomer DF, Nader R, Chlan-Fourney J, Suki D, Ahrar K, Rhines LD, Gokaslan ZL (2003) Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients. J Neurosurg 98:21–30CrossRefPubMed Fourney DR, Schomer DF, Nader R, Chlan-Fourney J, Suki D, Ahrar K, Rhines LD, Gokaslan ZL (2003) Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients. J Neurosurg 98:21–30CrossRefPubMed
8.
Zurück zum Zitat Fribourg D, Tang C, Sra P, Delamarter R, Bae H (2004) Incidence of subsequent vertebral fracture after kyphoplasty. Spine 29:2270–2276 (discussion 2277)CrossRefPubMed Fribourg D, Tang C, Sra P, Delamarter R, Bae H (2004) Incidence of subsequent vertebral fracture after kyphoplasty. Spine 29:2270–2276 (discussion 2277)CrossRefPubMed
9.
Zurück zum Zitat Gaitanis IN, Hadjipavlou AG, Katonis PG, Tzermiadianos MN, Pasku DS, Patwardhan AG (2005) Balloon kyphoplasty for the treatment of pathological vertebral compressive fractures. Eur Spine J 14:250–260CrossRefPubMed Gaitanis IN, Hadjipavlou AG, Katonis PG, Tzermiadianos MN, Pasku DS, Patwardhan AG (2005) Balloon kyphoplasty for the treatment of pathological vertebral compressive fractures. Eur Spine J 14:250–260CrossRefPubMed
10.
Zurück zum Zitat Goldhahn J, Reinhold M, Stauber M, Knop C, Frei R, Schneider E, Linke B (2006) Improved anchorage in osteoporotic vertebrae with new implant designs. J Orthop Res 24:917–925CrossRefPubMed Goldhahn J, Reinhold M, Stauber M, Knop C, Frei R, Schneider E, Linke B (2006) Improved anchorage in osteoporotic vertebrae with new implant designs. J Orthop Res 24:917–925CrossRefPubMed
11.
Zurück zum Zitat Heini PF (2005) The current treatment—a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon’s perspective. Osteoporos Int 16(Suppl 2):S85–S92CrossRefPubMed Heini PF (2005) The current treatment—a survey of osteoporotic fracture treatment. Osteoporotic spine fractures: the spine surgeon’s perspective. Osteoporos Int 16(Suppl 2):S85–S92CrossRefPubMed
12.
Zurück zum Zitat Heini PF, Berlemann U (2001) Bone substitutes in vertebroplasty. Eur Spine J 10(Suppl 2):S205–S213PubMed Heini PF, Berlemann U (2001) Bone substitutes in vertebroplasty. Eur Spine J 10(Suppl 2):S205–S213PubMed
13.
Zurück zum Zitat Heini PF, Orler R (2004) Kyphoplasty for treatment of osteoporotic vertebral fractures. Eur Spine J 13:184–192CrossRefPubMed Heini PF, Orler R (2004) Kyphoplasty for treatment of osteoporotic vertebral fractures. Eur Spine J 13:184–192CrossRefPubMed
14.
Zurück zum Zitat Heo DH, Kuh SU (2007) Progressive, repeated lumbar compression fracture at the same level after vertebral kyphoplasty with calcium phosphate cement. Case report. J Neurosurg Spine 6:559–562CrossRefPubMed Heo DH, Kuh SU (2007) Progressive, repeated lumbar compression fracture at the same level after vertebral kyphoplasty with calcium phosphate cement. Case report. J Neurosurg Spine 6:559–562CrossRefPubMed
15.
Zurück zum Zitat Herren T, Remagen W, Schenk R (1987) Histology of the implant-bone interface in cemented and uncemented endoprostheses. Orthopade 16:239–251PubMed Herren T, Remagen W, Schenk R (1987) Histology of the implant-bone interface in cemented and uncemented endoprostheses. Orthopade 16:239–251PubMed
16.
Zurück zum Zitat Higgins KB, Harten RD, Langrana NA, Reiter MF (2003) Biomechanical effects of unipedicular vertebroplasty on intact vertebrae. Spine 28:1540–1547 (discussion 1548)CrossRefPubMed Higgins KB, Harten RD, Langrana NA, Reiter MF (2003) Biomechanical effects of unipedicular vertebroplasty on intact vertebrae. Spine 28:1540–1547 (discussion 1548)CrossRefPubMed
17.
Zurück zum Zitat Kettler A, Schmoelz W, Shezifi Y, Ohana N, Ben-Arye A, Claes L, Wilke HJ (2006) Biomechanical performance of the new BeadEx implant in the treatment of osteoporotic vertebral body compression fractures: restoration and maintenance of height and stability. Clin Biomech (Bristol, Avon) 21:676–682CrossRef Kettler A, Schmoelz W, Shezifi Y, Ohana N, Ben-Arye A, Claes L, Wilke HJ (2006) Biomechanical performance of the new BeadEx implant in the treatment of osteoporotic vertebral body compression fractures: restoration and maintenance of height and stability. Clin Biomech (Bristol, Avon) 21:676–682CrossRef
18.
Zurück zum Zitat Kim MJ, Lindsey DP, Hannibal M, Alamin TF (2006) Vertebroplasty versus kyphoplasty: biomechanical behavior under repetitive loading conditions. Spine 31:2079–2084CrossRefPubMed Kim MJ, Lindsey DP, Hannibal M, Alamin TF (2006) Vertebroplasty versus kyphoplasty: biomechanical behavior under repetitive loading conditions. Spine 31:2079–2084CrossRefPubMed
19.
Zurück zum Zitat King AB, Saag KG, Burge RT, Pisu M, Goel N (2005) Fracture reduction affects medicare economics (FRAME): impact of increased osteoporosis diagnosis and treatment. Osteoporos Int 16:1545–1557CrossRefPubMed King AB, Saag KG, Burge RT, Pisu M, Goel N (2005) Fracture reduction affects medicare economics (FRAME): impact of increased osteoporosis diagnosis and treatment. Osteoporos Int 16:1545–1557CrossRefPubMed
20.
Zurück zum Zitat Knop C, Lange U, Bastian L, Blauth M (2000) Three-dimensional motion analysis with Synex. Comparative biomechanical test series with a new vertebral body replacement for the thoracolumbar spine. Eur Spine J 9:472–485CrossRefPubMed Knop C, Lange U, Bastian L, Blauth M (2000) Three-dimensional motion analysis with Synex. Comparative biomechanical test series with a new vertebral body replacement for the thoracolumbar spine. Eur Spine J 9:472–485CrossRefPubMed
21.
Zurück zum Zitat Lavelle WF, Cheney R (2006) Recurrent fracture after vertebral kyphoplasty. Spine J 6:488–493CrossRefPubMed Lavelle WF, Cheney R (2006) Recurrent fracture after vertebral kyphoplasty. Spine J 6:488–493CrossRefPubMed
22.
Zurück zum Zitat Leslie-Mazwi T, Deen HG (2006) Repeated fracture of a vertebral body after treatment with balloon kyphoplasty: case illustration. J Neurosurg Spine 4:270CrossRefPubMed Leslie-Mazwi T, Deen HG (2006) Repeated fracture of a vertebral body after treatment with balloon kyphoplasty: case illustration. J Neurosurg Spine 4:270CrossRefPubMed
23.
Zurück zum Zitat Lewis G (2007) Percutaneous vertebroplasty and kyphoplasty for the stand-alone augmentation of osteoporosis-induced vertebral compression fractures: present status and future directions. J Biomed Mater Res B Appl Biomater 81:371–386PubMed Lewis G (2007) Percutaneous vertebroplasty and kyphoplasty for the stand-alone augmentation of osteoporosis-induced vertebral compression fractures: present status and future directions. J Biomed Mater Res B Appl Biomater 81:371–386PubMed
24.
Zurück zum Zitat Lieberman IH, Togawa D, Kayanja MM (2005) Vertebroplasty and kyphoplasty: filler materials. Spine J 5:305S–316SCrossRefPubMed Lieberman IH, Togawa D, Kayanja MM (2005) Vertebroplasty and kyphoplasty: filler materials. Spine J 5:305S–316SCrossRefPubMed
25.
Zurück zum Zitat Liebschner MA, Rosenberg WS, Keaveny TM (2001) Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty. Spine 26:1547–1554CrossRefPubMed Liebschner MA, Rosenberg WS, Keaveny TM (2001) Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty. Spine 26:1547–1554CrossRefPubMed
26.
Zurück zum Zitat Lim TH, Brebach GT, Renner SM, Kim WJ, Kim JG, Lee RE, Andersson GB, An HS (2002) Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty. Spine 27:1297–1302CrossRefPubMed Lim TH, Brebach GT, Renner SM, Kim WJ, Kim JG, Lee RE, Andersson GB, An HS (2002) Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty. Spine 27:1297–1302CrossRefPubMed
27.
Zurück zum Zitat McCann H, LePine M, Glaser J (2006) Biomechanical comparison of augmentation techniques for insufficiency fractures. Spine 31:E499–E502CrossRefPubMed McCann H, LePine M, Glaser J (2006) Biomechanical comparison of augmentation techniques for insufficiency fractures. Spine 31:E499–E502CrossRefPubMed
28.
Zurück zum Zitat Mehbod A, Aunoble S, Le Huec JC (2003) Vertebroplasty for osteoporotic spine fracture: prevention and treatment. Eur Spine J 12(Suppl 2):S155–S162CrossRefPubMed Mehbod A, Aunoble S, Le Huec JC (2003) Vertebroplasty for osteoporotic spine fracture: prevention and treatment. Eur Spine J 12(Suppl 2):S155–S162CrossRefPubMed
29.
Zurück zum Zitat Panjabi MM (1988) Biomechanical evaluation of spinal fixation devices: I. A conceptual framework. Spine 13:1129–1134CrossRefPubMed Panjabi MM (1988) Biomechanical evaluation of spinal fixation devices: I. A conceptual framework. Spine 13:1129–1134CrossRefPubMed
30.
Zurück zum Zitat Panjabi MM, Krag M, Summers D, Videman T (1985) Biomechanical time-tolerance of fresh cadaveric human spine specimens. J Orthop Res 3:292–300CrossRefPubMed Panjabi MM, Krag M, Summers D, Videman T (1985) Biomechanical time-tolerance of fresh cadaveric human spine specimens. J Orthop Res 3:292–300CrossRefPubMed
31.
Zurück zum Zitat Panjabi MM, Wrathall JR (1988) Biomechanical analysis of experimental spinal cord injury and functional loss. Spine 13:1365–1370CrossRefPubMed Panjabi MM, Wrathall JR (1988) Biomechanical analysis of experimental spinal cord injury and functional loss. Spine 13:1365–1370CrossRefPubMed
32.
Zurück zum Zitat Perry A, Mahar A, Massie J, Arrieta N, Garfin S, Kim C (2005) Biomechanical evaluation of kyphoplasty with calcium sulfate cement in a cadaveric osteoporotic vertebral compression fracture model. Spine J 5:489–493CrossRefPubMed Perry A, Mahar A, Massie J, Arrieta N, Garfin S, Kim C (2005) Biomechanical evaluation of kyphoplasty with calcium sulfate cement in a cadaveric osteoporotic vertebral compression fracture model. Spine J 5:489–493CrossRefPubMed
33.
Zurück zum Zitat Reinhold M, Schwieger K, Goldhahn J, Linke B, Knop C, Blauth M (2006) Influence of screw positioning in a new anterior spine fixator on implant loosening in osteoporotic vertebrae. Spine 31:406–413CrossRefPubMed Reinhold M, Schwieger K, Goldhahn J, Linke B, Knop C, Blauth M (2006) Influence of screw positioning in a new anterior spine fixator on implant loosening in osteoporotic vertebrae. Spine 31:406–413CrossRefPubMed
34.
Zurück zum Zitat Rotter R, Pflugmacher R, Kandziora F, Ewert A, Duda G, Mittlmeier T (2007) Biomechanical in vitro testing of human osteoporotic lumbar vertebrae following prophylactic kyphoplasty with different candidate materials. Spine 32:1400–1405CrossRefPubMed Rotter R, Pflugmacher R, Kandziora F, Ewert A, Duda G, Mittlmeier T (2007) Biomechanical in vitro testing of human osteoporotic lumbar vertebrae following prophylactic kyphoplasty with different candidate materials. Spine 32:1400–1405CrossRefPubMed
35.
Zurück zum Zitat Song Y, Beaupre G, Goodman SB (1999) Osseointegration of total hip arthroplasties: studies in humans and animals. J Long Term Eff Med Implants 9:77–112PubMed Song Y, Beaupre G, Goodman SB (1999) Osseointegration of total hip arthroplasties: studies in humans and animals. J Long Term Eff Med Implants 9:77–112PubMed
36.
Zurück zum Zitat Steens J, Verdonschot N, Aalsma AM, Hosman AJ (2007) The influence of endplate-to-endplate cement augmentation on vertebral strength and stiffness in vertebroplasty. Spine 32:E419–E422CrossRefPubMed Steens J, Verdonschot N, Aalsma AM, Hosman AJ (2007) The influence of endplate-to-endplate cement augmentation on vertebral strength and stiffness in vertebroplasty. Spine 32:E419–E422CrossRefPubMed
37.
Zurück zum Zitat Sundaresan N, Boriani S, Rothman A, Holtzman R (2004) Tumors of the osseous spine. J Neurooncol 69:273–290CrossRefPubMed Sundaresan N, Boriani S, Rothman A, Holtzman R (2004) Tumors of the osseous spine. J Neurooncol 69:273–290CrossRefPubMed
38.
Zurück zum Zitat Wilke HJ, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7:148–154CrossRefPubMed Wilke HJ, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7:148–154CrossRefPubMed
39.
Zurück zum Zitat Willert HG, Buchhorn GH (1999) Osseointegration of cemented and noncemented implants in artificial hip replacement: long-term findings in man. J Long Term Eff Med Implants 9:113–130PubMed Willert HG, Buchhorn GH (1999) Osseointegration of cemented and noncemented implants in artificial hip replacement: long-term findings in man. J Long Term Eff Med Implants 9:113–130PubMed
40.
Zurück zum Zitat Yuan HA, Brown CW, Phillips FM (2004) Osteoporotic spinal deformity: a biomechanical rationale for the clinical consequences and treatment of vertebral body compression fractures. J Spinal Disord Tech 17:236–242CrossRefPubMed Yuan HA, Brown CW, Phillips FM (2004) Osteoporotic spinal deformity: a biomechanical rationale for the clinical consequences and treatment of vertebral body compression fractures. J Spinal Disord Tech 17:236–242CrossRefPubMed
Metadaten
Titel
Vertebroplasty with self-locking hexagonal metal implants shows comparable primary and secondary stiffness to PMMA cement augmentation techniques in a biomechanical vertebral compression fracture model
verfasst von
W. Schmoelz
A. C. Disch
J. F. Huber
Publikationsdatum
01.06.2010
Verlag
Springer-Verlag
Erschienen in
European Spine Journal / Ausgabe 6/2010
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-010-1342-9

Weitere Artikel der Ausgabe 6/2010

European Spine Journal 6/2010 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.