Erschienen in: Cardiovascular Ultrasound 1/2013

Open Access 01.12.2013 | Research

Visualization of multimodal polymer-shelled contrast agents using ultrasound contrast sequences: an experimental study in a tissue mimicking flow phantom

verfasst von: Malin Larsson, Matilda Larsson, Letizia Oddo, Silvia Margheritelli, Gaio Paradossi, Jacek Nowak, Lars-Åke Brodin, Kenneth Caidahl, Anna Bjällmark

Erschienen in: Cardiovascular Ultrasound | Ausgabe 1/2013

Abstract

Background

A multimodal polymer-shelled contrast agent (CA) with target specific potential was recently developed and tested for its acoustic properties in a single element transducer setup. Since the developed polymeric CA has different chemical composition than the commercially available CAs, there is an interest to study its acoustic response when using clinical ultrasound systems. The aim of this study was therefore to investigate the acoustic response by studying the visualization capability and shadowing effect of three polymer-shelled CAs when using optimized sequences for contrast imaging.

Methods

The acoustic response of three types of the multimodal CA was evaluated in a tissue mimicking flow phantom setup by measuring contrast to tissue ratio (CTR) and acoustic shadowing using five image sequences optimized for contrast imaging. The measurements were performed over a mechanical index (MI) range of 0.2-1.2 at three CA concentrations (106, 105, 104 microbubbles/ml).

Results

The CTR-values were found to vary with the applied contrast sequence, MI and CA. The highest CTR-values were obtained when a contrast sequence optimized for higher MI imaging was used. At a CA concentration of 106 microbubbles/ml, acoustic shadowing was observed for all contrast sequences and CAs.

Conclusions

The CAs showed the potential to enhance ultrasound images generated by available contrast sequences. A CA concentration of 106 MBs/ml implies a non-linear relation between MB concentration and image intensity.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1476-7120-11-33) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

ML, LÅB, AB participated in the initiation and design of the study. LO, SM and GP produced the polymer-shelled contrast agent used in the study. ML, ML, JN, KC and AB participated in data collection and analysis of the results. All authors read and approved the final manuscript.

Background

In 1968, intravenously injectable air microbubbles (MBs) were introduced as a contrast agent (CA) for use with echocardiography in order to enhance echo from relatively weak echogenic regions such as the vascular lumen [1]. However, the instability of these free gas MBs with a lifetime of only a few seconds limited their practical use and their relatively large size hindered their passage through the pulmonary capillary. In order to improve the MB stability, attempts have been made to encapsulate gas within a thin shell and in 1984, the first MBs with ability to pass through the pulmonary circulation were developed [2]. However, the thin shell of these MBs still allowed air to diffuse across the shell structure. In order to optimize the MB stability, a low soluble gas, instead of air, was encapsulated within the shell and the gas leakage was thus limited. At present, CAs consisting of MBs with the improved stability are widely used during standard ultrasound examinations, enabling e.g. tissue perfusion studies and endocardial border delineation [3, 4].
Although the commercially available CAs are relatively stable and well functioning in different clinical applications, there is still a need for improvement and extended applicability, such as targeted imaging and fusion imaging. In fact, polymer-shelled CAs with multimodality potential are currently under development [5, 6]. The multimodality approach has the potential to increase the clinical value of contrast imaging by providing the possibility of retrieving anatomical and functional information from two or more imaging systems simultaneously. In addition, the attachment of specific antibodies and ligands to the shell surface could enable specific attraction of the MBs to intended targets. The CAs might also be employed as a carrier of drugs that could be released locally at chosen target sites by disruption of the MBs with high-energy ultrasound. Other advantages of these CAs are a long shelf life as well as a narrow size distribution that increases the image sensitivity, which is of primary importance during targeting imaging [79].
Nevertheless, it is important to remember that polymer-shelled MBs demonstrate a different acoustical behavior than lipid-shelled MBs due to differences in compressibility and visco-elastic properties. The acoustic properties of the polymeric CAs has so far been studied in-vitro using single element transducers (Poehlmann et al.: On the interplay of structural, mechanical and acoustic behaviour of multifunctional magnetic microbubbles, submitted) [10, 11] and by using a commercially ultrasound scanner with in-house developed contrast sequences [12, 13]. However, no in-vitro studies using available clinical ultrasound systems with optimized sequences for contrast imaging has been performed hitherto. Keeping this in mind, the aim of this study was to determine the acoustic response by studying the visualization capability of three polymer-shelled CAs – one plain ultrasound and two surface-modified MBs for combined ultrasound and magnetic resonance imaging - in a tissue mimicking flow phantom using different ultrasound contrast sequences. In addition, the acoustic shadowing at different concentrations of the developed CAs was studied.

Material and methods

An experimental setup including a tissue mimicking flow phantom, a reservoir tank and a peristaltic pump was designed as shown in Figure 1. The contrast to tissue ratio (CTR) and the acoustic shadowing were studied for three polymer-shelled CAs, using different ultrasound systems with associated contrast sequences. The tests were performed at three CA concentrations. For the CTR-measurements, the commercially available CA SonoVue® was used as reference.

Phantom design

A tissue mimicking flow phantom with a vessel size of the human carotid artery was designed, see Figure 2. The vessel phantom was constructed by heating a mixture of 15% (w/w) polyvinyl alcohol (PVA) (Sigma-Aldrich, St. Louise, MO, US), 3% (w/w) graphite powder (particle size < 50 μm, Merck KGaA, Darmstadt, Germany) and de-ionized water to 90°C. Graphite powder was added as acoustic scatters [14]. The mixture was then poured into a cylindrical vessel mould of acrylic plastic (diameter 12 mm, length 100 mm) and the vessel lumen was created by inserting a metal rod (diameter 6 mm) into the vessel mould. Thereafter, it was stored in a freezer at −20°C for 12 h. During the subsequent 12 h, the vessel mould was stored in room temperature. In order to obtain acoustical properties similar to tissue, the vessel phantom underwent three freeze-thaw cycles [15].
After the three freeze-thaw cycles, the vessel phantom was attached to plastic connectors on each side of a polyvinyl chloride (PVC) box (100 mm × 60 mm), see Figure 1. The inner part of the PVC-box was covered with a 4 mm thick rubber layer that acted as an ultrasound absorbing material. A surrounding tissue mimicking mixture of 3% (w/w) agar (Merck KGaA, Darmstadt, Germany), 4% (w/w) graphite powder (particle size < 50 μm, Merck KGaA, Darmstadt, Germany) and de-ionized water was heated to 85°C. To prevent the vessel phantom from dissolving, the tissue mimicking mixture was cooled in room temperature to approximately 50°C before pouring it into the PVC-box. In addition, the vessel phantom lumen was filled with de-ionized water to prevent it from collapsing when pouring the tissue mimicking mixture into the PVC-box.

Contrast agents

Three in-house made polymer-shelled CAs were employed in the study (Table 1). One type was a plain polymer-shelled CA (Plain PVA) useful solely with ultrasound imaging [6]. The other two types had a modified shell or shell surface, both containing superparamagnetic iron oxide nanoparticles (SPIONs) for combined ultrasound and magnetic resonance imaging. However, the binding technique of the SPIONs differed. In the modified CA Type A (PVA Type A), the SPIONs were covalently linked to the shell surface through chitosan molecules, whereas in the modified CA Type B (PVA Type B), the SPIONs were embedded in the shell [5, 16]. In addition, the commercially available lipid-shelled CA SonoVue® was used as a reference [17].
Table 1
Properties of the contrast agents
Type of CA
Shell material
Gas composition
Plain PVA
Polymer
Air
PVA Type A
Polymer + SPION
Air
PVA Type B
Polymer + SPION
Air
SonoVue®
Phospholipid
Sulfur hexafluoride
PVA = polyvinyl alcohol, CA = contrast agent, SPION = superparamagnetic iron oxide nanoparticles.
The stock solution concentrations of each CA were determined using a light microscope (Olympus, Hamburg, Germany) and a counting chamber (Marienfeld Superior, Lauda-Königshofen, Germany). The mean value of the number of MBs within 16 squares of the chamber, each with a volume of 6.25 × 10-3 mm3, was calculated and displayed as the stock solution concentration. Subsequently, each stock solution was diluted with de-ionized water to desired concentration. In total, three concentrations of each type of CA were prepared, i. e. 104, 105 and 106 MBs/ml.

Experimental procedure

The tissue mimicking flow phantom was connected to a peristaltic pump (Watson Marlow, Falmouth, United Kingdom) generating a pulsatile flow with a mean velocity of 8.8 cm/s. The 500 ml solution circulating in the closed system consisted of de-ionized water and CA, which underwent constant stirring in the reservoir tank to obtain an even distribution of CA in the solution. A cleaning procedure was performed after each experiment with a specific CA at a given concentration in order to remove all MBs in the closed system. During the cleaning procedure, the vessel phantom and the plastic tubes were washed with de-ionized water. In total, three cleaning procedures were performed before a new batch of CA was added to the system.
The employed ultrasound transducers were fixed by a tripod holder at two different distances from the vessel phantom lumen depending on the applied frequency. The low frequency transducers (M3S (GE Healthcare, Wisconsin, US) and S5-1 (Philips Healthcare, Amsterdam, The Netherlands)) were fixed at a distance of 6 cm from the vessel lumen while the high frequency transducers (M12L (GE Healthcare, Wisconsin, US) and 9 L4 (Siemens Healthcare, Erlangen, Germany)) were fixed 3 cm from the vessel lumen.

Contrast sequences

Four commonly used contrast sequences were employed in the study; pulse inversion (PI), power modulation (PM), contrast pulse sequence (CPS) and power pulse inversion (PPI) (Table 2). The common feature for all contrast sequences used, was the detection of the nonlinear response from the studied CA upon exposure to acoustic pressure. A brief description of each contrast sequences follows. In PI every second transmitted pulse was inverted compared to the previous, while in PM two pulses were transmitted in phase but with amplitude modifications [18]. For CPS a combination of both phase and amplitude modification was applied [19]. When applying PPI, a sequence of three pulses was transmitted for each echo line. The first and the third transmitted pulses are identical while the second transmitted pulse is an inverted replica of them [20].
Table 2
Characteristics and settings of the contrast sequences
Ultrasound system
Contrast sequence
Transducer
Frequency (MHz)
Frame rate (Hz)
MI-range
Vivid7, GE
PI
M12L
5/10
32.7
0.2-1.2
Vivid7, GE
PI
M3S
1.5/3.1
25.7
0.2-0.8
iE33, Philips
PM
S5-1
1.5/3.2
39
0.2-1.2
iE33, Philips
PPI
S5-1
1.3/2.6
39
0.2-1.0
Acuson sequoia 512, Siemens
CPS
9 L4
4/8
17
0.2-1.0
PI = pulse inversion, PM = power modulation, PPI = power pulse inversion, CPS = contrast pulse subtraction, MI = mechanical index.

Data acquisition

Ultrasound short-axis images (n = 38) were acquired for four different contrast sequences (PI, PM, PPI and CPS) and three concentrations for each CA at a mechanical index (MI) ranging from 0.2 to maximal MI-value (0.8-1.2) of each ultrasound system tested (Table 2). A wide MI-range was used since the nonlinear responses from the CAs vary with applied acoustic pressure and the visco-elastic properties of the CA [21]. Before image acquisition, the focus point was placed in the middle of the vessel phantom lumen. The order of the image acquisition with the different contrast sequences was determined by a random number generator in Matlab® (MathWorks, Natick, US) to minimize the influence of possible concentration decrease of the CAs due to ultrasound exposure.

Data analysis

All the image sequences were analyzed offline using a specific work station for each ultrasound system; EchoPAC (GE Healthcare, Wisconsin, US), QLAB (Philips Healthcare Amsterdam, The Netherlands), Syngo (Siemens Healthcare, Erlangen, Germany).

Contrast to tissue ratio

Two regions of interest (ROIs) with an area of approximately 12.5 mm2 were manually placed in the middle of the vessel phantom lumen (ROIVessel) and in the surrounding tissue surrogate (ROITissue) at the same depth for all stored imaging sequences (Figure 3A). The ratio between the mean intensity (I), measured in decibel (dB), within the ROIVessel and the ROITissue was calculated and displayed as CTR (Equation 1).
CTR = I RO I Vessel I RO I Tissue
(1)

Acoustic shadowing

In order to assess the acoustic shadowing, two identical ROIs, with an area of approximately 3 mm2, were manually placed in the upper and the lower part of the vessel phantom lumen (Figure 3B) [22]. A shadowing effect was assumed if a significant higher intensity was observed for the upper ROI compared with the lower ROI. This was tested for each contrast sequence and CA concentration at the MI at which the maximal CTR-value was obtained using one-sided paired t-test (confidence interval of 95%).

Results

Contrast to tissue ratio

The distribution of the CTR-values expressed as a function of MI for different CAs at a concentration of 105 MBs/ml obtained with different contrast sequences are presented in Figure 4. As can be seen, using the PI and PM sequences, the highest CTR-values were obtained at a lower MI (0.4), whereas with PPI sequence, the highest CTR-values appeared at higher MI (1.2). On the other hand, CPS sequence tended to produce the highest CTR-values at intermediate to higher MI (0.6 – 1.0). A similar distribution pattern of the CTR-values was found for the two other CA concentrations (104 and 106 MBs/ml) that were also tested (data not shown). This is further elucidated in Figure 5 where it shows the maximum CTR-values for each studied CA at a concentration of 105 MBs/ml and the corresponding MI obtained with each contrast sequence employed. As can be seen from the figure, the PPI sequence provides the overall best performance, with the highest CTR-values for all studied CAs. A high dependency of the highest CTR values produced with different contrast sequences on the applied MI was also apparent in this study, a higher MI being optimal for use with PPI and CPS and a lower MI optimal for use with PI and PM sequences. Furthermore, PI and PM provided higher CTR-values for SonoVue® compared with the polymer-shelled CAs.
Figure 6 shows examples of ultrasound images of the tissue mimicking flow phantom containing Plain PVA at a concentration of 105 MBs/ml obtained with different contrast sequences at maximum CTR-values. The figure illustrates in further detail the high CTR produced by PPI sequence (Figure 6C) as well as a clear frequency-dependent difference in the images generated by PI with a high frequency transducer (f = 5/10 MHz) and low frequency transducer (f = 1.5/3.1 MHz), the higher CTR-values being obtained with the low frequency transducer (compare Figure 6A and 6B).

Acoustic shadowing

At a concentration of 106 MBs/ml, significant acoustic shadowing was observed for all three polymer-shelled CAs and contrast sequences (see Table 3). For concentrations lower than 105 MBs/ml, significant acoustic shadowing was only observed when using PPI.
Table 3
Acoustic shadowing
CA
PI
PI
PM
PPI
CPS
(f = 5/10 MHZ)
(f = 1.5/3.1 MHz)
(f = 1.5/3.2 MHz)
(f = 1.3/2.6 MHz)
(f = 4 MHz)
 
106
105
104
106
105
104
106
105
104
106
105
104
106
105
104
Plain PVA
p < 0.001
NS
NS
p < 0.001
NS
NS
p < 0.001
NS
NS
p < 0.001
p < 0.01
NS
p < 0.001
NS
NS
 
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(1.2)
(1.2)
(1.2)
(1.0)
(0.8)
(0.8)
PVA Type A
p < 0.001
NS
NS
p < 0.001
p < 0.001
NS
p < 0.001
p < 0.001
NS
p < 0.001
p < 0.001
p < 0.001
p < 0.001
NS
NS
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.4)
(0.6)
(1.2)
(1.2)
(1.2)
(0.8)
(0.6)
(0.6)
PVA Type B
p < 0.001
NS
NS
p < 0.001
NS
NS
P = 0.01
NS
NS
p < 0.001
p < 0.001
p < 0.05
p < 0.001
NS
NS
(0.4)
(0.4)
(0.4)
(0.4)
(0.6)
(0.4)
(0.6)
(0.4)
(0.6)
(1.2)
(1.2)
(1.2)
(1.0)
(1.0)
(1.0)
Observations of acoustic shadowing, at the maximal contrast to tissue ratio (dB) value and its associated mechanical index (brackets) for the different types of the polymer-shelled shelled CA at three concentrations; 106 MBs/ml, 105 MBs/ml, 104 MBs/ml. A significant shadowing effect is marked with its p-value and a non significant shadowing effect with NS (not significant).
PI = Pulse Inversion, PM = Power modulation, PPI = power pulse inversion, CPS = Contrast pulse sequence, PVA = Polyvinyl alcohol, NS = not significant.

Discussion

The CTR-values from the different CAs was strongly dependent on the contrast sequence and MI applied. As in previous studies [5, 10, 13] (Poehlmann et al.: On the interplay of structural, mechanical and acoustic behaviour of multifunctional magnetic microbubbles, submitted), variations in acoustic response for the polymer-shelled CAs were observed, thus suggesting that surface modification has an impact on the acoustic properties.
Overall, the PPI generated the highest CTR-values for all CAs included in the study. PPI is a decorrelation detection contrast sequence with the ability to effectively cancel the fundamental signal and enhance the harmonic signals due to the transmission and detection of three pulses for each echo line [20]. Both the oscillation and the destruction of CAs between the pulses, results in a high degree dissimilarities of the harmonic components, especially at high pressure, i.e. at higher MI. However, not only CA, but tissue also generates a nonlinear response when exposed to higher MI. Therefore, in order to obtain an image with minimal tissue influence, the received signal is filtered between the first and the second harmonic peak for the PPI contrast sequence. The main reason for this is that the harmonic signals from the CA have a wider bandwidth than those from the surrounding tissues [18]. The use of flexible and easily destroyed CA would result in a high decorrelation of the returning ultrasound pulses, which appears to be optimal for PPI sequence. This fact is confirmed by the high CTR-values for the thin-shelled SonoVue® using PPI. However, the thick polymer-shelled CAs used in this study produced CTR-values of the same magnitude as SonoVue®, even though the thick-shelled CAs fracture at higher MIs and this process is different compared to thin-shelled MBs [11]. The strong acoustic signal from the polymer-shelled CAs using PPI may be a result of shell defect caused by the ultrasound exposure resulting in release of free gas [18].
The ability to generate harmonics is highly dependent on the visco-elastic properties of the shell, which are determined by the shell thickness and its composition [21]. The polymer-shelled CAs have a stiffer and thicker shell than SonoVue®, implying that a higher MI is needed to obtain a sufficient nonlinear response from these CAs [23, 24]. But, in the case for PM and PI, a higher MI generates too much non-linear tissue signal to obtain high CTR-values. The difference in shell properties appears therefore to be the main reason for the fact that PI and PM perform better with SonoVue® than with the polymeric shelled CAs.
The two tested contrast sequences based on PI (low frequency f = 1.5/3.1 MHz and high frequency f = 5/10 MHz) generated different acoustic response, the lower frequency being superior for all CAs tested. The flexible shell of SonoVue® allows for large radius excursions giving rise to a well-defined resonance frequency [25]. This resonance frequency corresponds to the frequency range of the low frequency transducer, which in fact explains the higher CTR-values in this case. The polymer-shelled CAs are stiffer, which results in a considerable damping and a higher and less well-defined resonance frequency. Nevertheless, the low frequency transducer generated the highest CTR-values also with these CAs. One probable explanation to the observed phenomena, is that the high frequency signal results in higher pressure amplitudes at a certain MI, which cause a higher nonlinear response from tissue and hence, a decreased CTR-value. The frequency-dependant progressive distortion of the wave as it travels distally is an additional explanation. This influence was, however, reduced in the present experimental setup by the different imaging depths for the low and high frequency transducers.
The penetration of the ultrasound beam at a high concentration of MBs is reduced due to scattering and absorption [26]. It has been shown that acoustic shadowing is less pronounced at concentrations where there is a linear relationship between the concentration of CA and backscattered intensity [27]. In the present study, acoustic shadowing was observed for concentrations of 106 MBs/ml, irrespective of the CA and contrast sequence used, indicating a non-linear relation between CA concentration and image intensity at this concentration. Acoustic shadowing was most pronounced for PPI, where significant shadowing was observed even at low CA concentrations. This can, at least partly, be explained by the high pressure amplitudes for this contrast sequence, which in other studies have shown to highly increase the attenuation in a contrast media suspension [2830]. These findings are contradictory to the observation for CPS, where acoustic shadowing was not observed for the lower CA concentrations of the polymeric CA even though the pressure amplitude and the frequency was higher than for PPI. However, it should be mentioned in this context that the attenuation also has shown to be dependent on the transducer design and incident frequency spectra [28]. These results were found in a single element transducer setting but related findings was also obtained in a study utilizing clinical ultrasound systems [31]. In that study, machine-specific pre-and post processing algorithms were pointed out as an explanation for variation in the relationship between CA concentration and video intensity between ultrasound systems. In the current performed study, broadband transducers were used and the impact of the transducer design in combination with other machine-specific factors remains to be defined.
The in-vitro flow phantom setup offers standardized conditions for image acquisition. However, some limitations must be taken into account. It would have been optimal to use one single ultrasound system equipped with different contrast sequences over a wide frequency range. This was not possible since the clinical ultrasound systems available on the market today are limited to a few specific frequencies and contrast sequences. Still, this study provides an indication of which settings to use in order to visualize the polymeric CAs. A closed system was employed in the present study, implying that concentration loss due to MB destruction might have influenced the obtained CTR-values. However, the short ultrasound exposure time in combination with a relatively large volume of the CA solution, limited the effect of any MB destruction. Furthermore, the image acquisition was also performed in a randomized order in order to minimize possible influence of MB destruction. Different frame rates were used in the study. As a consequence, the acquisition time differed between the ultrasound systems tested. However, variations within the stored image sequences were limited as a homogenous CA mixture was used. In addition, a relatively high flow was applied which minimize the effect of eventual MB destruction in the field of view. Furthermore, if air bubbles were observed, a new image sequence was obtained. MBs motion is known to influence multi-pulse techniques using phase modulations. Although it has been demonstrated that the influence of axial motion of the MBs is much higher than that of lateral motion [32], the lateral motion of the MBs in our experimental setting might have influenced the tested contrast sequences. The present experimental setup mimicked the in-vivo imaging situation in the macrocirculation, in which contrast enhanced imaging can be used for detection of various clinical scenarios, e.g. left ventricle opacification, endocardial border delineation and detection of atherosclerotic plaques. Nevertheless, in order to fully evaluate the potential of the polymer-shelled CAs, future studies need to focus on the ability to visualize the microcirculation and to quantify tissue perfusion.
The polymer-shelled CAs has high mechanical and chemical stability [8], facilitating a possible use of the MBs as a carrier for different substances that can be incorporated into the shell or attached to the shell surface. It has thereby the potential to be of value in new clinical applications as, for example, multimodality imaging, targeted imaging and drug delivery. In particular, multimodal imaging where ultrasound and MRI are combined can add incremental value since these imaging methods are complementary in the diagnostic practice. The results of this study provide a clear indication of the optimal settings and contrast sequences for the polymer-shelled CAs, however, in-vivo imaging will certainly be more challenging and factors such as injection speed and concentration focus position and gain levels remain to be optimized. There is therefore a need for new in-vitro and in-vivo studies in order to further evaluate the applicability of the developed CAs, especially in the evaluation of microcirculation and tissue perfusion.

Conclusions

The present results demonstrate that the three polymer-shelled CAs have the potential to enhance ultrasound images when using commercially available contrast sequences, the highest CTR-values being obtained with PPI at higher MI. Surface modification seems to have an impact on the acoustic properties, as illustrated by the differences in CTR-values and acoustic shadowing between the three polymer-shelled CAs currently tested. A significant acoustical shadowing observed with all contrast sequences at a CA concentration of 106 MBs/ml implies that this CA concentration is too high for a linear relation between MB concentration and image intensity.

Acknowledgements

This study was supported by the European commission project 3MiCRON (245572), within the frame of the seventh framework program.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

ML, LÅB, AB participated in the initiation and design of the study. LO, SM and GP produced the polymer-shelled contrast agent used in the study. ML, ML, JN, KC and AB participated in data collection and analysis of the results. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Gramiak R, Shah PM: Echocardiography of the aortic root. Invest Radiol. 1968, 3: 356-366. 10.1097/00004424-196809000-00011.CrossRefPubMed Gramiak R, Shah PM: Echocardiography of the aortic root. Invest Radiol. 1968, 3: 356-366. 10.1097/00004424-196809000-00011.CrossRefPubMed
2.
Zurück zum Zitat Feinstein SB, Shah PM, Bing RJ, Meerbaum S, Corday E, Chang BL, Santillan G, Fujibayashi Y: Microbubble dynamics visualized in the intact capillary circulation. J Am Coll Cardiol. 1984, 4: 595-600. 10.1016/S0735-1097(84)80107-2.CrossRefPubMed Feinstein SB, Shah PM, Bing RJ, Meerbaum S, Corday E, Chang BL, Santillan G, Fujibayashi Y: Microbubble dynamics visualized in the intact capillary circulation. J Am Coll Cardiol. 1984, 4: 595-600. 10.1016/S0735-1097(84)80107-2.CrossRefPubMed
3.
Zurück zum Zitat Sboros V, Tang MX: The assessment of microvascular flow and tissue perfusion using ultrasound imaging. Proc Inst Mech Eng H. 2010, 224: 273-290. 10.1243/09544119JEIM621.CrossRefPubMed Sboros V, Tang MX: The assessment of microvascular flow and tissue perfusion using ultrasound imaging. Proc Inst Mech Eng H. 2010, 224: 273-290. 10.1243/09544119JEIM621.CrossRefPubMed
4.
Zurück zum Zitat Senior R, Andersson O, Caidahl K, Carlens P, Herregods MC, Jenni R, Kenny A, Melcher A, Svedenhag J, Vanoverschelde JL, et al.: Enhanced left ventricular endocardial border delineation with an intravenous injection of SonoVue, a new echocardiographic contrast agent: a European multicenter study. Echocardiogr. 2000, 17: 705-711. 10.1111/j.1540-8175.2000.tb01223.x.CrossRef Senior R, Andersson O, Caidahl K, Carlens P, Herregods MC, Jenni R, Kenny A, Melcher A, Svedenhag J, Vanoverschelde JL, et al.: Enhanced left ventricular endocardial border delineation with an intravenous injection of SonoVue, a new echocardiographic contrast agent: a European multicenter study. Echocardiogr. 2000, 17: 705-711. 10.1111/j.1540-8175.2000.tb01223.x.CrossRef
5.
Zurück zum Zitat Brismar TB, Grishenkov D, Gustafsson B, Harmark J, Barrefelt A, Kothapalli SV, Margheritelli S, Oddo L, Caidahl K, Hebert H, Paradossi G: Magnetite nanoparticles can be coupled with microbubbles to support multimodal imaging. Biomacromolecules. 2012, 13: 1390-1399. 10.1021/bm300099f.CrossRefPubMed Brismar TB, Grishenkov D, Gustafsson B, Harmark J, Barrefelt A, Kothapalli SV, Margheritelli S, Oddo L, Caidahl K, Hebert H, Paradossi G: Magnetite nanoparticles can be coupled with microbubbles to support multimodal imaging. Biomacromolecules. 2012, 13: 1390-1399. 10.1021/bm300099f.CrossRefPubMed
6.
Zurück zum Zitat Paradossi G, Cavalieri F, Chiessi E, Ponassi V, Martorana V: Tailoring of physical and chemical properties of macro- and microhydrogels based on telechelic PVA. Biomacromolecules. 2002, 3: 1255-1262. 10.1021/bm0256247.CrossRefPubMed Paradossi G, Cavalieri F, Chiessi E, Ponassi V, Martorana V: Tailoring of physical and chemical properties of macro- and microhydrogels based on telechelic PVA. Biomacromolecules. 2002, 3: 1255-1262. 10.1021/bm0256247.CrossRefPubMed
7.
Zurück zum Zitat Cavalieri F, El Hamassi A, Chiessi E, Paradossi G: Stable polymeric microballoons as multifunctional device for biomedical uses: synthesis and characterization. Langmuir. 2005, 21: 8758-8764. 10.1021/la050287j.CrossRefPubMed Cavalieri F, El Hamassi A, Chiessi E, Paradossi G: Stable polymeric microballoons as multifunctional device for biomedical uses: synthesis and characterization. Langmuir. 2005, 21: 8758-8764. 10.1021/la050287j.CrossRefPubMed
8.
Zurück zum Zitat Cavalieri F, El Hamassi A, Chiessi E, Paradossi G, Villa R, Zaffaroni N: Tethering functional ligands onto shell of ultrasound active polymeric microbubbles. Biomacromolecules. 2006, 7: 604-611. 10.1021/bm050723g.CrossRefPubMed Cavalieri F, El Hamassi A, Chiessi E, Paradossi G, Villa R, Zaffaroni N: Tethering functional ligands onto shell of ultrasound active polymeric microbubbles. Biomacromolecules. 2006, 7: 604-611. 10.1021/bm050723g.CrossRefPubMed
9.
Zurück zum Zitat Talu E, Hettiarachchi K, Zhao S, Powell RL, Lee AP, Longo ML, Dayton PA: Tailoring the size distribution of ultrasound contrast agents: possible method for improving sensitivity in molecular imaging. Mol Imaging. 2007, 6: 384-392.PubMedPubMedCentral Talu E, Hettiarachchi K, Zhao S, Powell RL, Lee AP, Longo ML, Dayton PA: Tailoring the size distribution of ultrasound contrast agents: possible method for improving sensitivity in molecular imaging. Mol Imaging. 2007, 6: 384-392.PubMedPubMedCentral
10.
Zurück zum Zitat Grishenkov D, Pecorari C, Brismar TB, Paradossi G: Characterization of acoustic properties of PVA-shelled ultrasound contrast agents: linear properties (part I). Ultrasound Med Biol. 2009, 35: 1127-1138. 10.1016/j.ultrasmedbio.2009.02.002.CrossRefPubMed Grishenkov D, Pecorari C, Brismar TB, Paradossi G: Characterization of acoustic properties of PVA-shelled ultrasound contrast agents: linear properties (part I). Ultrasound Med Biol. 2009, 35: 1127-1138. 10.1016/j.ultrasmedbio.2009.02.002.CrossRefPubMed
11.
Zurück zum Zitat Grishenkov D, Pecorari C, Brismar TB, Paradossi G: Characterization of acoustic properties of PVA-shelled ultrasound contrast agents: ultrasound-induced fracture (part II). Ultrasound Med Biol. 2009, 35: 1139-1147. 10.1016/j.ultrasmedbio.2009.03.006.CrossRefPubMed Grishenkov D, Pecorari C, Brismar TB, Paradossi G: Characterization of acoustic properties of PVA-shelled ultrasound contrast agents: ultrasound-induced fracture (part II). Ultrasound Med Biol. 2009, 35: 1139-1147. 10.1016/j.ultrasmedbio.2009.03.006.CrossRefPubMed
12.
Zurück zum Zitat Sciallero C, Paradossi G, Trucco A: A preliminary in vitro assessment of polymer-shelled microbubbles in contrast-enhanced ultrasound imaging. Ultrasonics. 2012, 52: 456-464. 10.1016/j.ultras.2011.10.008.CrossRefPubMed Sciallero C, Paradossi G, Trucco A: A preliminary in vitro assessment of polymer-shelled microbubbles in contrast-enhanced ultrasound imaging. Ultrasonics. 2012, 52: 456-464. 10.1016/j.ultras.2011.10.008.CrossRefPubMed
13.
Zurück zum Zitat Sciallero C, Trucco A: Ultrasound assessment of polymer-shelled magnetic microbubbles used as dual contrast agents. J Acoust Soc Am. 2013, 133: EL478-484. 10.1121/1.4804942.CrossRefPubMed Sciallero C, Trucco A: Ultrasound assessment of polymer-shelled magnetic microbubbles used as dual contrast agents. J Acoust Soc Am. 2013, 133: EL478-484. 10.1121/1.4804942.CrossRefPubMed
14.
Zurück zum Zitat Madsen EL, Zagzebski JA, Banjavie RA, Jutila RE: Tissue mimicking materials for ultrasound phantoms. Med Phys. 1978, 5: 391-394. 10.1118/1.594483.CrossRefPubMed Madsen EL, Zagzebski JA, Banjavie RA, Jutila RE: Tissue mimicking materials for ultrasound phantoms. Med Phys. 1978, 5: 391-394. 10.1118/1.594483.CrossRefPubMed
15.
Zurück zum Zitat Jea F: Estimation of polyvinyl alcohol cryogel mechanical properties with four ultrasound elastography methods and comparison with gold standard testings. IEEE Trans Ultrason Ferroelectr Freq Control. 2007, 54: 498-509.CrossRef Jea F: Estimation of polyvinyl alcohol cryogel mechanical properties with four ultrasound elastography methods and comparison with gold standard testings. IEEE Trans Ultrason Ferroelectr Freq Control. 2007, 54: 498-509.CrossRef
16.
Zurück zum Zitat Dähne L, Egri G: Preparation of magnetic microbubbles with PVA wall and use of them. German patent, DE. 10 2011 000 264.2-43. Dähne L, Egri G: Preparation of magnetic microbubbles with PVA wall and use of them. German patent, DE. 10 2011 000 264.2-43.
17.
Zurück zum Zitat Schneider M: Characteristics of SonoVuetrade mark. Echocardiogr. 1999, 16: 743-746. 10.1111/j.1540-8175.1999.tb00144.x.CrossRef Schneider M: Characteristics of SonoVuetrade mark. Echocardiogr. 1999, 16: 743-746. 10.1111/j.1540-8175.1999.tb00144.x.CrossRef
18.
Zurück zum Zitat Kaufmann BA, Wei K, Lindner JR: Contrast Echocardiography. Curr Probl Cardiol. 2007, 32: 51-96. 10.1016/j.cpcardiol.2006.10.004.CrossRefPubMed Kaufmann BA, Wei K, Lindner JR: Contrast Echocardiography. Curr Probl Cardiol. 2007, 32: 51-96. 10.1016/j.cpcardiol.2006.10.004.CrossRefPubMed
19.
Zurück zum Zitat Phillips P, Gardner E: Contrast-agent detection and quantification. Eur Radiol. 2004, 14 (Suppl 8): P4-10.PubMed Phillips P, Gardner E: Contrast-agent detection and quantification. Eur Radiol. 2004, 14 (Suppl 8): P4-10.PubMed
20.
Zurück zum Zitat Averkiou M, Powers J, Skyba D, Bruce M, Jensen S: Ultrasound contrast imaging research. Ultrasound Q. 2003, 19: 27-37. 10.1097/00013644-200303000-00004.CrossRefPubMed Averkiou M, Powers J, Skyba D, Bruce M, Jensen S: Ultrasound contrast imaging research. Ultrasound Q. 2003, 19: 27-37. 10.1097/00013644-200303000-00004.CrossRefPubMed
21.
Zurück zum Zitat Leong-Poi H, Song J, Rim SJ, Christiansen J, Kaul S, Lindner JR: Influence of microbubble shell properties on ultrasound signal: implications for low-power perfusion imaging. J Am Soc Echocardiogr. 2002, 15: 1269-1276. 10.1067/mje.2002.124516.CrossRefPubMed Leong-Poi H, Song J, Rim SJ, Christiansen J, Kaul S, Lindner JR: Influence of microbubble shell properties on ultrasound signal: implications for low-power perfusion imaging. J Am Soc Echocardiogr. 2002, 15: 1269-1276. 10.1067/mje.2002.124516.CrossRefPubMed
22.
Zurück zum Zitat Tiemann K, Schlosser T, Pohl C, Bimmel D, Wietasch G, Hoeft A, Likungu J, Vahlhaus C, Kuntz S, Nanda NC, et al.: Are microbubbles free flowing tracers through the myocardium? Comparison of indicator-dilution curves obtained from dye dilution and echo contrast using harmonic power Doppler imaging. Echocardiogr. 2000, 17: 17-27. 10.1111/j.1540-8175.2000.tb00989.x.CrossRef Tiemann K, Schlosser T, Pohl C, Bimmel D, Wietasch G, Hoeft A, Likungu J, Vahlhaus C, Kuntz S, Nanda NC, et al.: Are microbubbles free flowing tracers through the myocardium? Comparison of indicator-dilution curves obtained from dye dilution and echo contrast using harmonic power Doppler imaging. Echocardiogr. 2000, 17: 17-27. 10.1111/j.1540-8175.2000.tb00989.x.CrossRef
23.
Zurück zum Zitat Cavalieri F, FInelli I, Tortora M, Mozetic P, Chiessi E, Polizio F, Brismar TB, Paradossi G: Polymer micobubbles as diagnostic and therapeutic Gas delivery device. Chem Mater. 2008, 20: 3254-3258. 10.1021/cm703702d.CrossRef Cavalieri F, FInelli I, Tortora M, Mozetic P, Chiessi E, Polizio F, Brismar TB, Paradossi G: Polymer micobubbles as diagnostic and therapeutic Gas delivery device. Chem Mater. 2008, 20: 3254-3258. 10.1021/cm703702d.CrossRef
24.
Zurück zum Zitat Hoff L: Acoustic Characterization of Contrast Agents for Medical Ultrasound Imaging. 2001, AA Dordrecht: Kluwer Academic PublisherCrossRef Hoff L: Acoustic Characterization of Contrast Agents for Medical Ultrasound Imaging. 2001, AA Dordrecht: Kluwer Academic PublisherCrossRef
25.
Zurück zum Zitat Greis C: Technology overview: SonoVue (Bracco, Milan). Eur Radiol. 2004, 14 (Suppl 8): P11-15.PubMed Greis C: Technology overview: SonoVue (Bracco, Milan). Eur Radiol. 2004, 14 (Suppl 8): P11-15.PubMed
26.
Zurück zum Zitat de Jong N, Hoff L, Skotland T, Bom N: Absorption and scatter of encapsulated gas filled microspheres: theoretical considerations and some measurements. Ultrasonics. 1992, 30: 95-103. 10.1016/0041-624X(92)90041-J.CrossRefPubMed de Jong N, Hoff L, Skotland T, Bom N: Absorption and scatter of encapsulated gas filled microspheres: theoretical considerations and some measurements. Ultrasonics. 1992, 30: 95-103. 10.1016/0041-624X(92)90041-J.CrossRefPubMed
27.
Zurück zum Zitat Lampaskis M, Averkiou M: Investigation of the relationship of nonlinear backscattered ultrasound intensity with microbubble concentration at low MI. Ultrasound Med Biol. 2010, 36: 306-312. 10.1016/j.ultrasmedbio.2009.09.011.CrossRefPubMed Lampaskis M, Averkiou M: Investigation of the relationship of nonlinear backscattered ultrasound intensity with microbubble concentration at low MI. Ultrasound Med Biol. 2010, 36: 306-312. 10.1016/j.ultrasmedbio.2009.09.011.CrossRefPubMed
28.
Zurück zum Zitat Chen Q, Zagzebski J, Wilson T, Stiles T: Pressure-dependent attenuation in ultrasound contrast agents. Ultrasound Med Biol. 2002, 28: 1041-1051. 10.1016/S0301-5629(02)00546-X.CrossRefPubMed Chen Q, Zagzebski J, Wilson T, Stiles T: Pressure-dependent attenuation in ultrasound contrast agents. Ultrasound Med Biol. 2002, 28: 1041-1051. 10.1016/S0301-5629(02)00546-X.CrossRefPubMed
29.
Zurück zum Zitat Stride E, Saffari N: Investigating the significance of multiple scattering in ultrasound contrast agent particle populations. IEEE Trans Ultrason Ferroelectr Freq Control. 2005, 52: 2332-2345.CrossRefPubMed Stride E, Saffari N: Investigating the significance of multiple scattering in ultrasound contrast agent particle populations. IEEE Trans Ultrason Ferroelectr Freq Control. 2005, 52: 2332-2345.CrossRefPubMed
30.
Zurück zum Zitat Tang MX, Eckersley RJ: Frequency and pressure dependent attenuation and scattering by microbubbles. Ultrasound Med Biol. 2007, 33: 164-168. 10.1016/j.ultrasmedbio.2006.07.031.CrossRefPubMed Tang MX, Eckersley RJ: Frequency and pressure dependent attenuation and scattering by microbubbles. Ultrasound Med Biol. 2007, 33: 164-168. 10.1016/j.ultrasmedbio.2006.07.031.CrossRefPubMed
31.
Zurück zum Zitat Jayaweera AR, Skyba DM, Kaul S: Technical factors that influence the determination of microbubble transit rate during contrast echocardiography. J Am Soc Echocardiogr. 1995, 8: 198-206. 10.1016/S0894-7317(05)80408-0.CrossRefPubMed Jayaweera AR, Skyba DM, Kaul S: Technical factors that influence the determination of microbubble transit rate during contrast echocardiography. J Am Soc Echocardiogr. 1995, 8: 198-206. 10.1016/S0894-7317(05)80408-0.CrossRefPubMed
32.
Zurück zum Zitat Fanglue Lin CC, Riccardo M, Jacopo V, François V, Francesco G, Olivier B: Influences of bubble motion to second-harmonic inversion imaging. Proceedings of the. 2012, Dresden, 675-678. IEEE Ultrasonics symposium: 7-10 October 2012. Fanglue Lin CC, Riccardo M, Jacopo V, François V, Francesco G, Olivier B: Influences of bubble motion to second-harmonic inversion imaging. Proceedings of the. 2012, Dresden, 675-678. IEEE Ultrasonics symposium: 7-10 October 2012.
Metadaten
Titel
Visualization of multimodal polymer-shelled contrast agents using ultrasound contrast sequences: an experimental study in a tissue mimicking flow phantom
verfasst von
Malin Larsson
Matilda Larsson
Letizia Oddo
Silvia Margheritelli
Gaio Paradossi
Jacek Nowak
Lars-Åke Brodin
Kenneth Caidahl
Anna Bjällmark
Publikationsdatum
01.12.2013
Verlag
BioMed Central
Erschienen in
Cardiovascular Ultrasound / Ausgabe 1/2013
Elektronische ISSN: 1476-7120
DOI
https://doi.org/10.1186/1476-7120-11-33