A fatal combination of situs inversus, pregnancy and cardiac arrest treated with an automated external defibrillator

verfasst von: S. Calle, M. De Leeuw, N. Mpotos, P. Calle, B. De Turck

Erschienen in: Netherlands Heart Journal | Ausgabe 11/2016

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Answer

The AED delivered wrongful shocks, probably because the tachyarrhythmia with QRS complexes of low amplitude was considered to be ventricular fibrillation. Evidence for this comes from the ECG strips from the first AED rhythm analysis, clearly showing QRS complexes of a similar shape and amplitude, but at a lower rate. Furthermore, these small QRS complexes were continuously found from the start of the resuscitation attempt until the fourth shock (data not shown).
This case illustrates that the specificity of shock/no-shock decisions by the AEDs is not 100 % [1, 2]. Cardiologists should also be aware that AED sensitivity figures are only within a 90–95 % range, and that external artefacts during the analysis process (such as chest compressions) decrease the accuracy significantly. This implies that shock/no-shock decisions should be scrutinised before being used for diagnostic/therapeutic decision making and prognostication.
Remarkably, no cause of death could be detected on autopsy. Therefore, further research on the cardiac arrest risk stratification in dextrocardia in general, and during pregnancy in particular, is needed.
A third lesson concerns the low amplitude of QRS complexes detected by the AED electrodes placed in the conventional sternal-apical position [3]. Therefore, it seems reasonable in known dextrocardia cases to change to bi-axillary electrodes placement or a mirror-like approach (i. e. placement to the left of the sternum and in the right mid-axillary line). The reduced QRS amplitude in this dextrocardia case also argues against the use of the above-mentioned mirror-like approach in a standard cardiac arrest patient with a medical device implanted below the right clavicle.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
download
DOWNLOAD
print
DRUCKEN
Literatur
1.
Zurück zum Zitat Calle PA, Mpotos N, Calle SP, Monsieurs KG. Inaccurate treatment decisions of automated external defibrillators used by emergency medical services personnel: incidence, cause and impact on outcome. Resuscitation. 2015;88:68–74.CrossRefPubMed Calle PA, Mpotos N, Calle SP, Monsieurs KG. Inaccurate treatment decisions of automated external defibrillators used by emergency medical services personnel: incidence, cause and impact on outcome. Resuscitation. 2015;88:68–74.CrossRefPubMed
2.
Zurück zum Zitat Calle PA, Paepe P De, Van Sassenbroeck D, Monsieurs KG. External artifacts by advanced life support providers misleading automated external defibrillators. Resuscitation. 2008;79:482–89.CrossRefPubMed Calle PA, Paepe P De, Van Sassenbroeck D, Monsieurs KG. External artifacts by advanced life support providers misleading automated external defibrillators. Resuscitation. 2008;79:482–89.CrossRefPubMed
Metadaten
Titel
A fatal combination of situs inversus, pregnancy and cardiac arrest treated with an automated external defibrillator
verfasst von
S. Calle
M. De Leeuw
N. Mpotos
P. Calle
B. De Turck
Publikationsdatum
23.05.2016
Verlag
Bohn Stafleu van Loghum
Erschienen in
Netherlands Heart Journal / Ausgabe 11/2016
Print ISSN: 1568-5888
Elektronische ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-016-0852-4

Weitere Artikel der Ausgabe 11/2016

Netherlands Heart Journal 11/2016 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.