Erschienen in: Clinical Orthopaedics and Related Research® 5/2017

22.06.2016 | Symposium: Advances in Polyethylene Biomaterials

A Novel Technique for Assessing Antioxidant Concentration in Retrieved UHMWPE

verfasst von: Barbara H. Currier, MChE, Douglas W. Van Citters, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 5/2017

Einloggen, um Zugang zu erhalten

Abstract

Background

Antioxidants added to UHMWPE to prevent in vivo oxidation are important to the long-term performance of hip and knee arthroplasty. Diffused vitamin E antioxidant polyethylene raised questions about potential in vivo elution that could cause inflammatory reactions in periprosthetic tissues and also potentially leave the implant once again prone to oxidation. Currently, there is no information on the elution, if any, of antioxidants from implant polyethylene materials in vivo.

Questions/purposes

(1) Do antioxidants, especially diffused vitamin E, elute from antioxidant polyethylene in vivo? (2) Can extraction of the retrieved antioxidant polyethylene (to remove absorbed species from the in vivo environment near the articular and nonarticular surfaces) improve the identification of antioxidant content? (3) Can actual antioxidant content be estimated from calculated antioxidant indices by accounting for ester content (from absorbed species) near the articular and nonarticular surfaces?

Methods

An institutional review board-approved retrieval laboratory received 39 antioxidant polyethylene hip and knee retrievals at revision from 25 surgeons with in vivo time of 0.02 to 3.6 years (median, 1.3 years). These consecutive antioxidant polyethylene retrievals, received between May 2010 and February 2016, were made from three different antioxidant highly crosslinked polyethylene materials: diffused vitamin E, blended vitamin E, and hindered phenol antioxidant pentaerythritol tetrakis[3-(3,5-di-tert-butyl-4-hydroxyphenyl)] propionate (here and after referred to as PBHP). Retrievals were analyzed using Fourier transform infrared (FTIR) spectroscopy. Absorbed ester index (1725–1740 cm−1 normalized to 1365–1371 cm−1), and vitamin E index (1245–1275 cm−1) and PBHP index (1125–1150 cm−1), normalized to 1850–1985 cm−1, were defined. Microtomed thin sections of PBHP and vitamin E retrievals were hexane-extracted to remove absorbed species from the in vivo environment in an effort to improve identification of antioxidant content. Paired Student’s t-tests were used to compare as-retrieved articular antioxidant index with expected antioxidant index (the bulk value for blended antioxidants where constant antioxidant content is expected throughout and the extrapolated original vitamin E concentration at the articular surface based on the as-manufactured vitamin E concentration gradient). Linear regression was used for each of the retrievals to evaluate the correlation of antioxidant index to ester content with the goal of extrapolation to the antioxidant index at zero ester content.

Results

On average, vitamin E index at the articular surface (0.04 ± 0.03) was reduced compared with expected vitamin E index (0.09 ± 0.04; 95% confidence interval [CI] of the difference, 0.04-0.07; p < 0.001), and PBHP index at the articular surface (0.06 ± 0.02) was elevated compared with the average PBHP index from the bulk (0.03 ± 0.00; 95% CI of the difference, 0.03-0.05; p < 0.001). Extraction returned the PBHP index at the articular surface (0.03 ± 0.00) to bulk values (95% CI of the difference, -0.001 to 0.004; p = 0.326); diffused vitamin E was removed by extraction. Crossplots of vitamin E index and PBHP index with ester index showed significant (p < 0.001 for 32 of the 35 retrievals with sufficient data) linear trends (r ≥ 0.89) that allowed extrapolation of the articular surface antioxidant indices at zero absorbed ester index.

Conclusions

Absorbed esters from time in vivo caused erroneous values of antioxidant index to be calculated. However, hexane extraction to remove absorbed species also removed diffused vitamin E. Correlating antioxidant indices with ester content, measured by FTIR in unextracted antioxidant retrievals, provides a nonaltered method for estimating actual articular surface vitamin E index and demonstrates that there was no measurable elution in these short-term retrievals.

Clinical Relevance

Assessing antioxidant content in retrieved polyethylene inserts is important to determine how much of the antioxidant remains in place to prevent oxidation of the polyethylene over time in vivo. Retrieval analyses reporting antioxidant content must account for absorbed species to be valid. Because standard hexane extraction removes both absorbed species and vitamin E from diffused vitamin E retrievals, the correlation method presented in this study is the recommended analysis alternative.
Literatur
1.
Zurück zum Zitat ASTM Standard F2102-13, 2013. Standard Guide for Evaluating the Extent of Oxidation in Ultra-high-molecular-weight Polyethylene Fabricated Forms Intended for Surgical Implants. West Conshohocken, PA, USA: ASTM International; 2013. ASTM Standard F2102-13, 2013. Standard Guide for Evaluating the Extent of Oxidation in Ultra-high-molecular-weight Polyethylene Fabricated Forms Intended for Surgical Implants. West Conshohocken, PA, USA: ASTM International; 2013.
2.
Zurück zum Zitat Berry D, Currier B, Mayor M, Collier J. Gamma-irradiation sterilization in an inert environment: a partial solution. Clin Orthop Relat Res. 2011;470:1805–1813.CrossRefPubMedCentral Berry D, Currier B, Mayor M, Collier J. Gamma-irradiation sterilization in an inert environment: a partial solution. Clin Orthop Relat Res. 2011;470:1805–1813.CrossRefPubMedCentral
3.
Zurück zum Zitat Bracco P, Oral E. Vitamin E-stabilized UHMWPE for total joint implants: a review. Clin Orthop Relat Res. 2011;469:2286–2293.CrossRefPubMed Bracco P, Oral E. Vitamin E-stabilized UHMWPE for total joint implants: a review. Clin Orthop Relat Res. 2011;469:2286–2293.CrossRefPubMed
4.
Zurück zum Zitat Currier BH, Van Citters DW, Currier JH, Carlson EM, Tibbo ME, Collier JP. In vivo oxidation in retrieved highly crosslinked tibial inserts. J Biomed Mater Res Part B. 2013;101:441–448. Currier BH, Van Citters DW, Currier JH, Carlson EM, Tibbo ME, Collier JP. In vivo oxidation in retrieved highly crosslinked tibial inserts. J Biomed Mater Res Part B. 2013;101:441–448.
5.
Zurück zum Zitat Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.CrossRefPubMed Dumbleton JH, D’Antonio JA, Manley MT, Capello WN, Wang A. The basis for a second-generation highly cross-linked UHMWPE. Clin Orthop Relat Res. 2006;453:265–271.CrossRefPubMed
6.
Zurück zum Zitat Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked Marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21(Suppl 2):17.CrossRefPubMed Engh CA Jr, Stepniewski AS, Ginn SD, Beykirch SE, Sychterz-Terefenko CJ, Hopper RH Jr, Engh CA. A randomized prospective evaluation of outcomes after total hip arthroplasty using cross-linked Marathon and non-cross-linked Enduron polyethylene liners. J Arthroplasty. 2006;21(Suppl 2):17.CrossRefPubMed
7.
Zurück zum Zitat Haider H, Weisenburger JN, Kurtz SM, Rimnac CM, Freedman J, Schroeder DW, Garvin KL. Does vitamin E-stabilized ultrahigh-molecular-weight polyethylene address concerns of cross-linked polyethylene in total knee arthroplasty? J Arthroplasty. 2012;27:461–469.CrossRefPubMed Haider H, Weisenburger JN, Kurtz SM, Rimnac CM, Freedman J, Schroeder DW, Garvin KL. Does vitamin E-stabilized ultrahigh-molecular-weight polyethylene address concerns of cross-linked polyethylene in total knee arthroplasty? J Arthroplasty. 2012;27:461–469.CrossRefPubMed
8.
Zurück zum Zitat Jacobs CA, Christensen CP, Greenwald AS, McKellop H. Clinical performance of highly crosslinked polyethylenes in total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2779–2786.CrossRefPubMed Jacobs CA, Christensen CP, Greenwald AS, McKellop H. Clinical performance of highly crosslinked polyethylenes in total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2779–2786.CrossRefPubMed
9.
Zurück zum Zitat James SP, Blazka S, Merrill EW, Jasty M, Lee KR, Bragdon CR, Harris WH. Challenge to the concept that UHMWPE acetabular components oxidize in vivo. Biomaterials. 1993;14:643–647.CrossRefPubMed James SP, Blazka S, Merrill EW, Jasty M, Lee KR, Bragdon CR, Harris WH. Challenge to the concept that UHMWPE acetabular components oxidize in vivo. Biomaterials. 1993;14:643–647.CrossRefPubMed
10.
Zurück zum Zitat Jarrett BT, Cofske J, Rosenberg AE, Oral E, Muratoglu OK, Malchau H. In vivo biological response to vitamin E and vitamin-E-doped polyethylene. J Bone Joint Surg Am. 2010;92:2672–2681.CrossRefPubMed Jarrett BT, Cofske J, Rosenberg AE, Oral E, Muratoglu OK, Malchau H. In vivo biological response to vitamin E and vitamin-E-doped polyethylene. J Bone Joint Surg Am. 2010;92:2672–2681.CrossRefPubMed
11.
Zurück zum Zitat Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.CrossRefPubMed Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.CrossRefPubMed
12.
Zurück zum Zitat McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.CrossRefPubMed McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2009;91:773–782.CrossRefPubMed
13.
Zurück zum Zitat McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.CrossRefPubMed McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–167.CrossRefPubMed
14.
Zurück zum Zitat Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. HAP Paul Award: A novel method of crosslinking UHMWPE to improve wear, reduce oxidation, and retain mechanical properties. J Arthroplasty. 2001;16:149–160.CrossRefPubMed Muratoglu OK, Bragdon CR, O’Connor DO, Jasty M, Harris WH. HAP Paul Award: A novel method of crosslinking UHMWPE to improve wear, reduce oxidation, and retain mechanical properties. J Arthroplasty. 2001;16:149–160.CrossRefPubMed
15.
Zurück zum Zitat Narayan VS. Spectroscopic and chromatographic quantification of an antioxidant-stabilized ultrahigh-molecular-weight polyethylene. Clin Orthop Relat Res. 2015;473:952–959.CrossRefPubMed Narayan VS. Spectroscopic and chromatographic quantification of an antioxidant-stabilized ultrahigh-molecular-weight polyethylene. Clin Orthop Relat Res. 2015;473:952–959.CrossRefPubMed
16.
Zurück zum Zitat Oral E, Beckos CG, Malhi AS, Muratoglu OK. The effects of high dose irradiation on the cross-linking of vitamin E-blended UHMWPE. Biomaterials. 2008;29:3557–3560.CrossRefPubMedPubMedCentral Oral E, Beckos CG, Malhi AS, Muratoglu OK. The effects of high dose irradiation on the cross-linking of vitamin E-blended UHMWPE. Biomaterials. 2008;29:3557–3560.CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Oral E, Neils AL, Rowell SL, Lozynsky AJ, Muratoglu OK. Increasing irradiation temperature maximizes vitamin E grafting and wear resistance of ultrahigh molecular weight polyethylene. J Biomed Mater Res Part B. 2013;101:436–440. Oral E, Neils AL, Rowell SL, Lozynsky AJ, Muratoglu OK. Increasing irradiation temperature maximizes vitamin E grafting and wear resistance of ultrahigh molecular weight polyethylene. J Biomed Mater Res Part B. 2013;101:436–440.
18.
Zurück zum Zitat Oral E, Wannomae KK, Rowell SL, Muratoglu OK. Migration stability of α-tocopherol in irradiated UHMWPE. Biomaterials. 2006;27:2434–2439.CrossRefPubMed Oral E, Wannomae KK, Rowell SL, Muratoglu OK. Migration stability of α-tocopherol in irradiated UHMWPE. Biomaterials. 2006;27:2434–2439.CrossRefPubMed
19.
Zurück zum Zitat Oral E, Wannomae KK, Rowell SL, Muratoglu OK. Diffusion of vitamin E in ultra-high molecular weight polyethylene. Biomaterials. 2007;28:5225–5237.CrossRefPubMedPubMedCentral Oral E, Wannomae KK, Rowell SL, Muratoglu OK. Diffusion of vitamin E in ultra-high molecular weight polyethylene. Biomaterials. 2007;28:5225–5237.CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Reinitz SD, Currier BH, Van Citters DW, Levine RA, Collier JP. Oxidation and other property changes of retrieved sequentially annealed UHMWPE acetabular and tibial bearings. J Biomed Mater Res Part B. 2015;103:578–586.CrossRef Reinitz SD, Currier BH, Van Citters DW, Levine RA, Collier JP. Oxidation and other property changes of retrieved sequentially annealed UHMWPE acetabular and tibial bearings. J Biomed Mater Res Part B. 2015;103:578–586.CrossRef
21.
Zurück zum Zitat Rowell SL, Muratoglu OK. Investigation of surgically retrieved, vitamin E-stabilized, crosslinked UHMWPE implants after short-term in vivo service. J Biomed Mater Res Part B. 3 June 2015. [Epub ahead of print] Rowell SL, Muratoglu OK. Investigation of surgically retrieved, vitamin E-stabilized, crosslinked UHMWPE implants after short-term in vivo service. J Biomed Mater Res Part B. 3 June 2015. [Epub ahead of print]
22.
Zurück zum Zitat Thomas GER, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.CrossRefPubMed Thomas GER, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.CrossRefPubMed
Metadaten
Titel
A Novel Technique for Assessing Antioxidant Concentration in Retrieved UHMWPE
verfasst von
Barbara H. Currier, MChE
Douglas W. Van Citters, PhD
Publikationsdatum
22.06.2016
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 5/2017
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-016-4939-4

Weitere Artikel der Ausgabe 5/2017

Clinical Orthopaedics and Related Research® 5/2017 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.