Acute and long-term results with the 3rd generation visually guided laser balloon ablation system for pv isolation

verfasst von: Moritoshi Funasako, Jan Petrů, Pavel Hála, Marek Janotka, Jan Škoda, Milan Chovanec, Lucie Šedivá, Vivek Y. Reddy, Petr Neužil

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 1/2024

Abstract

Background

Visually guided laser balloon ablation is known as an effective pulmonary vein (PV) isolation device. The third-generation laser balloon ablation system (X3) equipped with compliant balloon and an automated motor-driven laser output mechanism, namely RAPID mode, has been clinically proven for PV isolation.

Methods

PV isolation with X3 was performed in all the patients with paroxysmal and early-stage persistent atrial fibrillation (AF). Acute data for PV isolation and clinical outcomes including supraventricular tachyarrhythmia (SVT: AF, atrial flutter, or atrial tachycardia)-free survival rate beyond 1 year were analyzed.

Results

A total of 110 patients (62 ± 13 years old, 80% of paroxysmal AF) were treated with X3. RAPID mode with was utilized to achieve PV isolation in all cases. In combination with RAPID mode and spot mode laser ablation, 91.1% (380/417) of veins were isolated on the first circumferential lesion set and did not require touch-up ablation and during the index procedure 100% of attempted veins were isolated. The mean procedure time was 77.0 ± 22.7 min and LA dwell time was 61.9 ± 22.0 min. Total duration of laser application was 5.1 ± 2.3 min per vein. At 1 year, SVT-free survival rate was 93.7% in paroxysmal AF patients, and 81.1% in persistent AF patients.

Conclusions

A novel continuous automatic laser balloon ablation system was proved to be safe and effective for both paroxysmal and persistent AF patients. The clinical result demonstrated that PV isolation with X3 could achieve a high SVT-free survival rate.
Hinweise

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AF
Atrial fibrillation
AAD
Anti-arrhythmic drag
AMD
Amiodarone
BB
Beta blocker
LCPV
Left common pulmonary vein
PVI
Pulmonary vein isolation
RF
Radiofrequency
RSPV
Right superior pulmonary vein
SVT
Supraventricular arrhythmia
TIA
Transient ischaemic attack
X3
The third-generation laser balloon ablation system

1 Introduction

Pulmonary vein (PV) isolation is widely recognized as the gold standard therapy for paroxysmal and persistent atrial fibrillation (AF). Radiofrequency (RF) ablation was first established and played a key role in AF ablation, especially RF PV isolation which has been recognized as an optimal therapeutic option [13]. The conventional RF procedure requires experience in catheter handling and output power control, but PV reconnections have been recently reported to occur between 30 and 50% among recurrent AF patients even with highly integrated lesion quality markers/index RF ablation [46]. Moreover, overheating of RF applications may create thrombus or steam pops leading to major complications such as ischemic stroke or cardiac tamponade [7]. More recently, balloon ablation technologies have also been developed to simplify the PV isolation procedure and improve clinical outcomes for AF patients [811].
Endoscopic visually guided laser ablation for PV isolation (Heartlight®, CardioFocus, Inc., Marlborough, MA, USA) first emerged in 2004 and proved to be a preferable tool to create a transmural lesion in the left atrium with effective and properly titrated laser energy delivery. Laser balloon PV isolation has been known for its safety and durability [1215]. Combined with a second generation, more compliant balloon compared to the first generation, the latest third-generation laser balloon ablation system (X3) offers visualized tissue contact via an endoscope positioned in the catheter shaft, has proven essential for creating irreversible or durable ablation lesions [16]. Another major feature of the X3 technology is the ability to create continuous lesions with an automated motor-driven, self-rotation laser delivery system, which can create a gapless “single-sweep” lesion around the circumference of the PV in as few as three minutes [17] (Fig. 1).
Our study aims to evaluate the safety and efficacy of PV isolation procedures and clinical results with a novel automatic motor-driven laser balloon X3 ablation system featuring RAPID mode in routine clinical practice.

2 Methods

2.1 Patient selection

One hundred and ten (110) consecutive paroxysmal and persistent AF patients were included in the study. The inclusion criteria were defined the same as the standard RF ablation procedure. Exclusion criteria were left atrial (LA) thrombus, left atrium diameter > 60 mm, left ventricular ejection fraction (LVEF) < 35%, previous ablation for AF including PV isolation, NYHA class IV symptoms, recent myocardial infarction within 60 days, unstable angina, uncontrolled bleeding, contraindication of anticoagulation therapy, and active infectious disease. Anatomical characteristics of the LA or targeted pulmonary veins (i.e., left common vein or separated right middle vein/branches) were not considered as an exclusion criterion.

2.2 Ablation procedure

The laser balloon catheter was introduced through the right femoral vein utilizing a dedicated 14 Fr. single deflectable sheath via atrial transseptal puncture. ACT was maintained over 300 s by heparin injection throughout the procedure. The catheter tip was positioned into each vein and confirmed by fluoroscopy and intracardiac echography, then the balloon was inflated by operators with the controller which is located next to the catheter handle. The tissue contact of the balloon was only observed through the endoscopic view. The standard laser output power was set at 13W with RAPID mode where stable contact was obtained without any blood flow on the surface of the balloon, while set at 8.5W, 20 s or 5.5W, 30 s with spot mode ablation was applied where the contact was poor, unstable, or blood flow or blood pool was close to the targeted tissue. Operators were not supposed to deliver laser energy directly on blood. The esophageal temperature was monitored in all the cases and laser energy delivery was terminated when the temperature hit over 41.0°. Phrenic nerve pacing was performed during right PVs isolation to detect phrenic nerve injury as quickly as possible. After all the PV procedures were completed, PV isolation was confirmed by a 20-electrode circular mapping catheter. If PV potentials were still detected after the first attempted encirclement of the PV, the same laser balloon ablation system was used to reisolate the uncompleted vein. All the timeline was recorded from the first groin puncture to the last sheath extraction from the LA.

2.3 Patients’ follow-ups

Oral anticoagulation therapy was continued at least 2 months after the procedure even if the patient was completely free from AF recurrence. The first 3 months after the index procedure was considered as a blanking period when the patient with early recurrence of supraventricular tachyarrhythmia (SVT: i.e., atrial fibrillation, atrial flutter, or atrial tachycardia) was not counted as having a clinically significant recurrence. All the patients with refracting AF who were not able to be treated with an anti-arrhythmic drug or electrical cardioversion were scheduled for a second procedure with a 3D mapping system and standard RF catheter ablation. In the redo procedure, all the PVs were re-evaluated with a 20-electrode circular mapping catheter after the transseptal puncture and insertion of a long sheath. In case of any PV reconnection, RF touch-up ablation of the residual gaps was required to archive PV isolation. Beyond PV isolation strategies (i.e., additional linear ablation and substrate ablation) were considered depending on the operators’ decision.

2.4 Statistical analysis

All the data are expressed as mean ± SD in every table. Differences in frequencies were analyzed by the Chi-square test and a two-sided P < 0.05 was considered to be statistically significant (JMP ver. 12.0, SAS Institute Inc.).

2.5 Ethical clearance and data transparency

This study follows the principles of the Declaration of Helsinki. All the patients were informed of and agreed with the purpose of the study and the privacy policy in written form. Ethical approval was acquired by the local institutional ethical committee. The original and additional study data will be made available upon reasonable request.

3 Results

3.1 Patient characteristics and PV anatomical features

A total of 110 patients underwent PV isolation with X3. Four-fifths of the patients were paroxysmal (n = 88, 80%), the average age was 62 ± 13 years old (paroxysmal: 61 ± 14 years, persistent: 65 ± 11 years), and 37 patients were female (33.6%). More than half of the patients suffered from AF over 1 year (paroxysmal: 64.5%, persistent: 77.8%) and the mean LA size was 43 ± 4.3 mm (42 ± 3.9 mm vs. 45 ± 4.5 mm, paroxysmal vs. persistent AF patients, respectively; P = 0.002.). All the patients underwent PV isolation for the first time and 13 patients (12.5%) underwent CT isthmus RF ablation due to prior history of common flutter. The patient characteristics are summarized in Table 1. According to a CT scan or intracardiac echography, the left common vein was confirmed in 20 patients.
Table 1
Patient characteristics
Type of AF
Paroxysmal
Persistent
Total
 
(n = 88)
(n = 22)
(n = 110)
Age (y.o.)
61 ± 14
65 ± 11
62 ± 13
Male
64.8% (57/88)
72.7% (16/22)
66.4% (n = 73)
AF < 1 year
35.5% (27/76)
22.2% (4/18)
33.0% (31/94)
AF > 1 year
64.5% (49/76)
77.8% (14/18)
67.0% (63/94)
Previous AF ablation
0 (0%)
0 (0%)
0 (0%)
Past electrical cardioversion
20.5% (18/88)
40.9% (9/22)
24.5% (27/110)
Hypertension
54.5% (48/88)
40.9% (9/22)
51.8% (57/110)
Coronary artery diseases
5.7% (5/88)
0% (0/22)
4.5% (5/110)
Diabetes
13.6% (12/88)
0% (0/22)
10.9% (12/110)
Past heart failure
2.3% (2/88)
0% (0/22)
1.8% (2/110)
Past cardiac surgery
0% (0/88)
4.5% (1/22)
0.9% (1/110)
Stroke/TIA
3.4% (3/88)
4.5% (1/22)
3.6% (4/110)
Sleep apnea
0% (0/88)
4.5% (1/22)
0.9% (1/110)
AAD before X3 PVI
62.5% (55/88)
94.5% (21/22)
69.1% (76/110)
AAD at the last follow-up
23.9% (21/88)
27.3% (6/22)
24.5% (27/110)

3.2 Procedure details of PV isolation

Of the 420 PVs where treatment was attempted/treated, 417 veins were assessed post-ablation. One patient had four veins, including one untreated vein, and the other three veins were treated but not checked after PV isolation due to interruption of the procedure. The RAPID mode was applied to 418/420 (99.5%) PVs and 138/420 (32.8%) PVs were treated only with the RAPID mode. In combination with RAPID mode and spot mode laser ablation, 91.1% (380/417) of veins were isolated on the first circumferential lesion set. All the veins were isolated only with X3 and no vein required an RF touch-up.
The mean procedure time from the first groin puncture to the sheath extraction from the LA was 77.0 ± 22.7 min. The mean left atrial treatment time (LA dwell time) was 61.9 ± 22.0 min and the mean treatment procedure time was 54.0 ± 20.2 min. In 88.3% of the PVs, RAPID mode was used in most (> 50%) or all of the ablation lesion sets. The mean fluoroscopy time was 3.7 ± 3.2 min. There was no significant difference in procedure time, left atrial treatment time, and fluoroscopy time between paroxysmal AF and persistent AF patients. Total duration of laser application was 5.1 ± 2.3 min (paroxysmal AF: 5.0 ± 2.0 min, vs. persistent AF: 5.8 ± 3.3 min, p = 0.03) per vein. In 8 paroxysmal AF patients (7.3%), the catheter was exchanged due to a pinhole during laser application (Table 2).
Table 2
Acute results of PV isolation procedure with X3
 
Paroxysmal
(n = 88)
Persistent
(n = 22)
Total X3 population (n = 110)
Procedure time (min) [1]
76.2 ± 20.7 (88)
80.2 ± 29.5 (22)
77.0 ± 22.7 (110)
Treatment left atrial time (min) [2]
61.8 ± 20.5 (88)
62.1 ± 28.0 (22)
61.9 ± 22.0 (110)
Treatment time (min) [3]
54.4 ± 19.8 (88)
52.7 ± 22.5 (22)
54.0 ± 20.2 (110)
Fluoroscopy time (min)
3.7 ± 3.1 (88)
3.6 ± 3.4 (22)
3.7 ± 3.2 (110)
No. PVs attempted per patient
3.8 ± 0.5 (88)
3.7 ± 0.6 (22)
3.8 ± 0.5 (110)
No. PVs isolated/attempted
100.0% (335/335)
100.0% (82/82)
100.0% (417/417)
No. mapping attempts per PV to achieve isolation
  1
91.6% (307/335)
89.0% (73/82)
91.1% (380/417)
  2
7.2% (24/335)
11.0% (9/82)
7.9% (33/417)
   ≥ 3
1.2% (4/335)
0.0% (0/82)
1.0% (4/417)
No. ablation catheters used
  1
90.9% (80/88)
100.0% (22/22)
92.7% (102/110)
  2
9.1% (8/88)
0.0% (0/22)
7.3% (8/110)
  Acute isolation rate [4]
100.0% (87/87)
100.0% (22/22)
100.0% (109/109)
  First attempt at all veins isolated (FAAVI) [5]
75.9% (66/87)
68.2% (15/22)
74.3% (81/109)
Percent of the vein circumference treated in RAPID
  0%
0.2% (1/419)
0.0% (0/337)
1.2% (1/82)
  1–25%
1.4% (6/419)
1.8% (6/337)
0.0% (0/82)
  26–50%
9.8% (41/419)
9.2% (31/337)
12.2% (10/82)
  51–75%
21.0% (88/419)
20.5% (69/337)
23.2% (19/82)
  76–99%
34.6% (145/419)
34.4% (116/337)
35.4% (29/82)
  100%
32.9% (138/419)
34.1% (115/337)
28.0% (23/82)
  Total therapy time (min) [6]
5.1 ± 2.3 (413)
5.0 ± 2.0 (331)
5.8 ± 3.3 (82)
[1] Defined as the time from venous access to the time of withdrawal of the HeartLight Catheter
[2] Defined as the time from transseptal puncture to withdrawal of the HeartLight Catheter
[3] Defined as the time from the insertion of the HeartLight Catheter to the time of withdrawal of the HeartLight Catheter
[4] Defined as the % of patients where all veins attempted were isolated with the X3 catheter
[5] Defined as the % of patients where all veins attempted were isolated with the X3 catheter and only 1 mapping/vein
[6] Defined as the total number of minutes the laser is active in the vein
The only complications were transient phrenic nerve palsy (PNP) observed in three patients. One of them was the patient without the completed PVI mentioned above, but all PNP recovered at the 6-month visit after the procedures. No other major complications occurred during and after the procedure through 1 year of follow-up.

3.3 Clinical course after PV isolation

All the patients after the first laser balloon PV isolation were investigated employing 12 leads ECG (3 M, 6 M, 12 M), 24 h Holter ECG (6 M, 12 M), and an interview survey on AF symptoms. Recurrent AF is defined as documented sustained AF of more than 30 s by 12 leads ECG or 24 h Holter ECG. Out of 109 patients after PVI, SVT recurrence was observed in 20 patients (14 paroxysmal AF and 6 persistent AF patients) 1 year after the initial procedure and 5 patients underwent redo AF procedures. Of all the 20 recurrent patients, one paroxysmal AF patient underwent electrical cardioversion due to common flutter after the blanking period (109 days after the index procedure) and kept sinus rhythm up to the end of the follow-up period at 15 months. All the other 19 patients developed recurrent AF during follow-up period. At 1 year, SVT-free survival rate was 93.7% in paroxysmal AF patients and 81.1% in persistent AF patients (Fig. 2).

3.4 The findings from the second procedures

The recurrent AF patients were mainly treated by anti-arrhythmic medications or electrical cardioversion. A total of five patients (4 paroxysmal and 1 persistent AF patients) underwent the second ablation procedure due to the recurrence of AF. Among four paroxysmal patients, two patients had one reconnected PV each (LCPV and RSPV, respectively). One persistent AF patient confirmed reconnected RSPV. Sixteen out of 19 PVs (84.2%) in the recurrent AF patients confirmed durable PV isolation as described in Fig. 3. In the redo cases, additional posterior box isolation, linear ablation, and substrate ablation were performed after PV re-isolation depending on the findings during procedures.

4 Discussion

4.1 Rational for durable PV isolation

PV isolation has been recognized as the fundamental procedure for AF treatment. Although “beyond PV” strategies such as linear ablation or substrate-based ablations in addition to PV isolation have been tried, the main factor for clinical AF-free survival results is durable and antral PV isolation [18, 19]. The past studies repeatedly referred to the importance of durable PV isolation as a strong factor for the high AF-free survival rate in any type of AF [20]. Balloon ablation devices have been developed to make PV isolation procedures simple and short; however, durable PV isolation is still challenging as both conventional RF ablation and balloon ablation require experience and technique to control the catheter to get stability and safety profile simultaneously, or to occlude PV with the device in any anatomical conditions, respectively. Different energy sources need specific approaches for PV isolation and laser balloon ablation is known for its safety and durability of PV isolation with a short learning curve [2, 16].
Cryoballoon ablation requires basically total PV occlusion to maximize cryo-thermal effect or a relatively large area with optimal tissue contact as the cryo ablation creates lesions mainly with conductive thermal energy from the balloon surface to the targeted tissue. Laser balloon ablation requires a specifically targeted tissue view via an endoscope as the laser energy through the balloon runs straight directly onto the visualized targeted tissue and generates the heating energy slightly beneath the surface of the tissue with minimal energy loss as long as there is no blood flow or blood pool that may cause intermediate energy attenuation. Laser energy seems to be more concentrated and effective for stable localized transmural ablation lesions wherever the targeted tissue is clearly visible and away from the blood. As an ablation energy source, laser energy has such a unique character that ablation lesions would be minimal but sufficient for durable PV isolation once delivered to the tissue. As seen in this study, this energy-specific lesion formation may have contributed to a high SVT-free survival rate with durable PV isolation and to no left atrial AT recurrences with less reversible ablation lesions that often cause abnormal automaticity, triggered activity, or slow conduction formation leading to clinical ATs.
The latest laser balloon ablation can also reduce the risk of phrenic nerve injury as well when the ablation line is selectively created out of the PV ostium while cryoballoon ablation creates wide lesions all around PV ostium as the northern hemisphere of the balloon generates freezing thermal energy [21]. In this study, three different operators experienced one transient phrenic nerve palsy each during the right superior PV isolation despite phrenic nerve pacing, all three patients recovered to normal function. This implies that the laser energy reached the phrenic nerve to affect the function, but the heating seemed to have been conducive away from the treated tissue where the laser energy was applied. Laser balloon ablation is the only balloon ablation method that enables the design of the isolation line on the actual tissue. Highly selectively designed antral PV isolation lines to avoid the specific location can reduce phrenic nerve injury and esophageal complication in contrast to other single-shot balloon ablation modalities, and at the same time, more antral PV isolation could be archived like RF ablation if necessary.

4.2 Utility of compliant balloon for various anatomical conditions

For balloon ablation modalities, variability in PV anatomy can be a procedural limitation. These balloon techniques require occlusion of the PVs when the energy is delivered to the tissue. Poor tissue contact may lead to unsuccessful PV isolation or reversible lesion formation, and thereby poor clinical results. While cryoballoon ablation is designed to occlude the PV and make wide and circumferential lesions in a single-shot manner with fixed balloon size, laser balloon ablation enables operators to design the antral PV isolation lesion lines via endoscopic view with an expandable compliant balloon adjustable to various PV size from 8 to 41 mm. Specifically, one of the most challenging anatomical profiles is the left common vein [22]. In this study, 20 common veins were all treated only with X3, which is compatible with various PV sizes and shapes. X3 is the only single-sweep device for left common veins compared to other balloon techniques that require multiple segment applications per vein. Right inferior PV is also sometimes challenging for balloon ablations due to poor contact on the bottom aspect. One of the advantages of X3 is the compliant balloon that sits well around the PV ostium and helps create circumferential continuous ablation using RAPID mode even for small veins, common PV trunks, and veins with small or early branches. The laser ablation provides sufficient lesion depth without any pushing maneuver as long as the targeted tissue is visible. Our study showed high flexibility of the X3 system against anatomical anomalies without any RF touch-up ablation.

4.3 Advantage of seamless continuous ablation compared to point-by-point spot mode ablation

The first- and second-generation laser balloon systems did not include the automated rotational RAPID mode ablation feature. Although those prior generations recorded excellent clinical outcomes, each 20- to 30-s single application needs 20 to 50% lesion overlap [13, 14]. The necessity of the repeated overlaps with spot mode ablations between the lesions may make the procedure time long and a higher possibility of a gap or reversible lesion formation would be expected. The newly established RAPID mode provides (1) fast PV isolation as few as three minutes per vein, (2) continuous lesion formation with less possibility to have a gap between two different ablation lesions, and (3) qualitatively stable and durable lesions. Indeed, not only the total procedure time and ablation time but also the first-pass isolation rate was improved compared to the first- or second-generation laser balloon ablation system. As seen in this study, catheter exchanges were required due to a pinhole in 8 cases (7.3%). Since the X3 procedures have been reported as a pivotal study, a relatively high pinhole rate (8–13%) was reported [17, 23]. The RAPID mode may damage the balloon material when applied directly to the blood. Operators should be cautious to choose spot mode ablation if the targeted lesion is next to the blood pool or intermittently covered by blood flow. Despite this limitation, the past study demonstrated that the RAPID mode improved acute PV isolation results with a highly successful PV isolation rate and shorter procedure time compared to the former generation [23]. More recently, the balloon material has been modified to be more robust for RAPID mode usage, which was not clinically available in this study. The new balloon material could improve the percentage of RAPID mode usage, shorten the procedure time, and reduce catheter exchanges, leading to better acute procedural results and long-term patient clinical results.

5 Limitations

This study includes some limitations. First, the protocol is a non-randomized, single-arm, and single-center study. Second, the results were not compared with other specific procedures such as RF PVI or cryoballoon PVI. Third, the procedure was limited to PV isolation even for persistent AF patients. Although the number of patients was relatively small, few recurrent arrhythmias were observed beyond 1-year observation. According to the redo cases, in most of the cases, PVs were isolated and non-PV triggers or substrates were considered as the source of AF. Based on this fact, the high durability of the PV isolation with X3 seemed to have contributed to the superb SVT-free clinical result. At the same time, even for persistent AF patients, durable PV isolation should be the essential strategy for better clinical outcomes [24].

6 Conclusion

A novel continuous automatic laser balloon ablation system for PV isolation was found to be safe and effective for both paroxysmal and persistent AF patients. The combination of effective laser energy and seamless lesion archives quick and durable PV isolation without serious complication, which was proved in the clinical outcome.

Acknowledgements

We would like to thank Mr. Jiří Baroch and our technicians for their assistance with collecting all the acute procedural data and follow-up data. We also appreciate the demonstrative pictures provided by CardioFocus, Inc.

Declarations

Conflict of interest

This project was supported by scientific grant of Ministry of Health, Czech Republic - MHCZ-DRO (NNH, 00023884) IG220501. PN and VR have received speaking/honoraria from, and are consultants to CardioFocus. VR has disclosures with various cardiovascular companies unrelated to the current study. Consultant: Abbott, AtiAN, Biosense-Webster, BioTel Heart, Biotronik, Cardionomic, CoreMap, EBR, Fire1, Gore & Associates, Impulse Dynamics, Medtronic, Philips, and Pulse Biosciences; Consultant, Equity: Ablacon, Acutus Medical, Affera, Apama Medical-Boston Scientific, APN Health, Aquaheart, Atacor, Autonomix, Axon Therapies, Backbeat, BioSig, CardiaCare, CardioNXT/AFTx, Circa Scientific, CoRISMA, Corvia Medical, Dinova-Hangzhou DiNovA EP Technology, East End Medical, EPD-Philips, EP Frontiers, Epix Therapeutics, EpiEP, Eximo, Farapulse-Boston Scientific, HRT, Intershunt, Javelin, Kardium, Keystone Heart, LuxMed, Medlumics, Middlepeak, Neutrace, Nuvera-Biosense Webster, Oracle Health, Restore Medical, Sirona Medical, and Valcare; Equity: Manual Surgical Sciences, Newpace, Surecor, and Vizaramed. All other authors have no disclosures to report.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat January CT, Wann LS, Alpert JS, et al. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the Heart Rhythm Society. Circulation. 2014;130(23):e199-267.PubMedPubMedCentral January CT, Wann LS, Alpert JS, et al. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the Heart Rhythm Society. Circulation. 2014;130(23):e199-267.PubMedPubMedCentral
2.
Zurück zum Zitat January CT, Wann LS, Calkins H, et al. 2019 AHA/ACC/HRS focused update of the 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society in Collaboration With the Society of Thoracic Surgeons. Circulation. 2019;140(2):e125–51.CrossRefPubMed January CT, Wann LS, Calkins H, et al. 2019 AHA/ACC/HRS focused update of the 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society in Collaboration With the Society of Thoracic Surgeons. Circulation. 2019;140(2):e125–51.CrossRefPubMed
3.
Zurück zum Zitat Morillo CA, Verma A, Connolly SJ, et al. Radiofrequency ablation vs antiarrhythmic drugs as first-line treatment of paroxysmal atrial fibrillation (RAAFT-2): a randomized trial. JAMA. 2014;311(7):692–9.CrossRefPubMed Morillo CA, Verma A, Connolly SJ, et al. Radiofrequency ablation vs antiarrhythmic drugs as first-line treatment of paroxysmal atrial fibrillation (RAAFT-2): a randomized trial. JAMA. 2014;311(7):692–9.CrossRefPubMed
4.
Zurück zum Zitat Das M, Loveday JJ, Wynn GJ, et al. Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values. Europace. 2017;19(5):775–83.PubMed Das M, Loveday JJ, Wynn GJ, et al. Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values. Europace. 2017;19(5):775–83.PubMed
5.
Zurück zum Zitat Molina-Ramos AI, Ruiz-Salas A, Medina-Palomo C, et al. Reablation in atrial fibrillation recurrence and pulmonary vein reconnection: cryoballoon versus radiofrequency as index ablation procedures. J Clin Med. 2022;11(19):5862.CrossRefPubMedPubMedCentral Molina-Ramos AI, Ruiz-Salas A, Medina-Palomo C, et al. Reablation in atrial fibrillation recurrence and pulmonary vein reconnection: cryoballoon versus radiofrequency as index ablation procedures. J Clin Med. 2022;11(19):5862.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Lennon SJ, Mannion J, Keelan E, et al. Ablation index outcome in redo persistent atrial fibrillation ablation: results of a short-term study. Cardiol Res. 2022;13(2):97–103.CrossRefPubMedPubMedCentral Lennon SJ, Mannion J, Keelan E, et al. Ablation index outcome in redo persistent atrial fibrillation ablation: results of a short-term study. Cardiol Res. 2022;13(2):97–103.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Chun KRJ, Perrotta L, Bordignon S, et al. Complications in catheter ablation of atrial fibrillation in 3,000 consecutive procedures: balloon versus radiofrequency current ablation. JACC Clin Electrophysiol. 2017;3(2):154–61.CrossRefPubMed Chun KRJ, Perrotta L, Bordignon S, et al. Complications in catheter ablation of atrial fibrillation in 3,000 consecutive procedures: balloon versus radiofrequency current ablation. JACC Clin Electrophysiol. 2017;3(2):154–61.CrossRefPubMed
8.
Zurück zum Zitat Andrade JG, Khairy P, Macle L, et al. Incidence and significance of early recurrences of atrial fibrillation after cryoballoon ablation: insights from the multicenter Sustained Treatment of Paroxysmal Atrial Fibrillation (STOP AF) Trial. Circ Arrhythm Electrophysiol. 2014;7(1):69–75.CrossRefPubMed Andrade JG, Khairy P, Macle L, et al. Incidence and significance of early recurrences of atrial fibrillation after cryoballoon ablation: insights from the multicenter Sustained Treatment of Paroxysmal Atrial Fibrillation (STOP AF) Trial. Circ Arrhythm Electrophysiol. 2014;7(1):69–75.CrossRefPubMed
9.
Zurück zum Zitat Chun JKR, Bordignon S, Chen S, et al. Current status of atrial fibrillation ablation with balloon strategy. Korean Circ J. 2019;49(11):991–1001.CrossRefPubMedPubMedCentral Chun JKR, Bordignon S, Chen S, et al. Current status of atrial fibrillation ablation with balloon strategy. Korean Circ J. 2019;49(11):991–1001.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Wazni OM, Dandamudi G, Sood N, et al. Cryoballoon ablation as initial therapy for atrial fibrillation. N Engl J Med. 2020;384(4):316–24.CrossRefPubMed Wazni OM, Dandamudi G, Sood N, et al. Cryoballoon ablation as initial therapy for atrial fibrillation. N Engl J Med. 2020;384(4):316–24.CrossRefPubMed
11.
Zurück zum Zitat Heeger CH, Pott A, Sohns C, et al. Novel cryoballoon ablation system for pulmonary vein isolation: multicenter assessment of efficacy and safety-ANTARCTICA study. Europace. 2022;24(12):1917–25.CrossRefPubMedPubMedCentral Heeger CH, Pott A, Sohns C, et al. Novel cryoballoon ablation system for pulmonary vein isolation: multicenter assessment of efficacy and safety-ANTARCTICA study. Europace. 2022;24(12):1917–25.CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Reddy VY, Neuzil P, Themistoclakis S, et al. Visually-guided balloon catheter ablation of atrial fibrillation: experimental feasibility and first-in-human multicenter clinical outcome. Circulation. 2009;120(1):12–20.CrossRefPubMed Reddy VY, Neuzil P, Themistoclakis S, et al. Visually-guided balloon catheter ablation of atrial fibrillation: experimental feasibility and first-in-human multicenter clinical outcome. Circulation. 2009;120(1):12–20.CrossRefPubMed
13.
Zurück zum Zitat Dukkipati SR, Neuzil P, Skoda J, et al. Visual balloon-guided point-by-point ablation: reliable, reproducible, and persistent pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2010;3(3):266–73.CrossRefPubMed Dukkipati SR, Neuzil P, Skoda J, et al. Visual balloon-guided point-by-point ablation: reliable, reproducible, and persistent pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2010;3(3):266–73.CrossRefPubMed
14.
Zurück zum Zitat Dukkipati SR, Kuck KH, Neuzil P, et al. Pulmonary vein isolation using a visually guided laser balloon catheter: the first 200-patient multicenter clinical experience. Circ Arrhythm Electrophysiol. 2013;6(3):467–72.CrossRefPubMed Dukkipati SR, Kuck KH, Neuzil P, et al. Pulmonary vein isolation using a visually guided laser balloon catheter: the first 200-patient multicenter clinical experience. Circ Arrhythm Electrophysiol. 2013;6(3):467–72.CrossRefPubMed
15.
Zurück zum Zitat Sediva L, Petru J, Skoda J, et al. Visually guided laser ablation: a single-centre long-term experience. Europace. 2014;16(12):1746–51.CrossRefPubMed Sediva L, Petru J, Skoda J, et al. Visually guided laser ablation: a single-centre long-term experience. Europace. 2014;16(12):1746–51.CrossRefPubMed
16.
Zurück zum Zitat Heeger CH, Phan HL, Meyer-Saraei R, et al. Second-generation visually guided laser balloon ablation system for pulmonary vein isolation: learning curve, safety and efficacy- the MERLIN registry. Circ J. 2019;83(12):2443–51.CrossRefPubMed Heeger CH, Phan HL, Meyer-Saraei R, et al. Second-generation visually guided laser balloon ablation system for pulmonary vein isolation: learning curve, safety and efficacy- the MERLIN registry. Circ J. 2019;83(12):2443–51.CrossRefPubMed
17.
Zurück zum Zitat Schmidt B, Petru J, Chun KRJ, et al. Pivotal study of a novel motor-driven endoscopic ablation system. Circ Arrhythm Electrophysiol. 2021;14(3): e009544.CrossRefPubMed Schmidt B, Petru J, Chun KRJ, et al. Pivotal study of a novel motor-driven endoscopic ablation system. Circ Arrhythm Electrophysiol. 2021;14(3): e009544.CrossRefPubMed
18.
Zurück zum Zitat Fink T, Schluter M, Heeger CH, et al. Stand-alone pulmonary vein isolation versus pulmonary vein isolation with additional substrate modification as index ablation procedures in patients with persistent and long-standing persistent atrial fibrillation: the Randomized Alster-Lost-AF Trial (Ablation at St. Georg Hospital for Long-Standing Persistent Atrial Fibrillation). Circ Arrhythm Electrophysiol. 2017;10(7):e005114. Fink T, Schluter M, Heeger CH, et al. Stand-alone pulmonary vein isolation versus pulmonary vein isolation with additional substrate modification as index ablation procedures in patients with persistent and long-standing persistent atrial fibrillation: the Randomized Alster-Lost-AF Trial (Ablation at St. Georg Hospital for Long-Standing Persistent Atrial Fibrillation). Circ Arrhythm Electrophysiol. 2017;10(7):e005114.
19.
Zurück zum Zitat Miyazaki S, Horie T, Hachiya H, et al. The mechanisms of recurrent atrial arrhythmias after second-generation cryoballoon ablation. Am Heart J. 2020;221:29–38.CrossRefPubMed Miyazaki S, Horie T, Hachiya H, et al. The mechanisms of recurrent atrial arrhythmias after second-generation cryoballoon ablation. Am Heart J. 2020;221:29–38.CrossRefPubMed
20.
Zurück zum Zitat Verma A, Jiang C-y, Betts TR, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372(19):1812–22.CrossRefPubMed Verma A, Jiang C-y, Betts TR, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372(19):1812–22.CrossRefPubMed
21.
Zurück zum Zitat Chun JKR, Bordignon S, Last J, et al. Cryoballoon versus laserballoon: insights from the first prospective randomized balloon trial in catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2021;14(2): e009294.CrossRefPubMed Chun JKR, Bordignon S, Last J, et al. Cryoballoon versus laserballoon: insights from the first prospective randomized balloon trial in catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2021;14(2): e009294.CrossRefPubMed
22.
Zurück zum Zitat Yamaguchi M, Miyazaki S, Kajiyama T, et al. Pulmonary vein isolation in patients with a left common pulmonary vein: comparison between second-generation cryoballoon and radiofrequency ablation. J Cardiol. 2019;73(4):292–8.CrossRefPubMed Yamaguchi M, Miyazaki S, Kajiyama T, et al. Pulmonary vein isolation in patients with a left common pulmonary vein: comparison between second-generation cryoballoon and radiofrequency ablation. J Cardiol. 2019;73(4):292–8.CrossRefPubMed
23.
Zurück zum Zitat Heeger CH, Tiemeyer CM, Phan HL, et al. Rapid pulmonary vein isolation utilizing the third-generation laserballoon - The PhoeniX registry. Int J Cardiol Heart Vasc. 2020;29: 100576.PubMedPubMedCentral Heeger CH, Tiemeyer CM, Phan HL, et al. Rapid pulmonary vein isolation utilizing the third-generation laserballoon - The PhoeniX registry. Int J Cardiol Heart Vasc. 2020;29: 100576.PubMedPubMedCentral
24.
Zurück zum Zitat Bordignon S, Boehmer MC, Klostermann A, et al. Visually guided pulmonary vein isolation in patients with persistent atrial fibrillation. Europace. 2016;18(4):538–42.CrossRefPubMed Bordignon S, Boehmer MC, Klostermann A, et al. Visually guided pulmonary vein isolation in patients with persistent atrial fibrillation. Europace. 2016;18(4):538–42.CrossRefPubMed
Metadaten
Titel
Acute and long-term results with the 3rd generation visually guided laser balloon ablation system for pv isolation
verfasst von
Moritoshi Funasako
Jan Petrů
Pavel Hála
Marek Janotka
Jan Škoda
Milan Chovanec
Lucie Šedivá
Vivek Y. Reddy
Petr Neužil
Publikationsdatum
20.02.2023
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 1/2024
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-023-01499-8

Weitere Artikel der Ausgabe 1/2024

Journal of Interventional Cardiac Electrophysiology 1/2024 Zur Ausgabe

Triglyzeridsenker schützt nicht nur Hochrisikopatienten

10.05.2024 Hypercholesterinämie Nachrichten

Patienten mit Arteriosklerose-bedingten kardiovaskulären Erkrankungen, die trotz Statineinnahme zu hohe Triglyzeridspiegel haben, profitieren von einer Behandlung mit Icosapent-Ethyl, und zwar unabhängig vom individuellen Risikoprofil.

Gibt es eine Wende bei den bioresorbierbaren Gefäßstützen?

In den USA ist erstmals eine bioresorbierbare Gefäßstütze – auch Scaffold genannt – zur Rekanalisation infrapoplitealer Arterien bei schwerer PAVK zugelassen worden. Das markiert einen Wendepunkt in der Geschichte dieser speziellen Gefäßstützen.

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.