Adjustable loop ACL suspension devices demonstrate less reliability in terms of reproducibility and irreversible displacement

Zeitschrift:
Knee Surgery, Sports Traumatology, Arthroscopy > Ausgabe 5/2018
Autoren:
Sufian S. Ahmad, Michael T. Hirschmann, Benjamin Voumard, Sandro Kohl, Philippe Zysset, Takura Mukabeta, Dimitrios S. Evangelopoulos, Atesch Ateschrang
Wichtige Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00167-017-4769-2) contains supplementary material, which is available to authorized users.

Abstract

Purpose

The aim of this study was to perform a comprehensive biomechanical examination of frequently applied femoral cortical suspension devices, comparing the properties of both fixed and adjustable fixation mechanisms. It was hypothesized that adjustable loop devices demonstrate less consistent fixation properties with increased variability compared to fixed loop devices.

Methods

Nine frequently applied fixation button types were tested, six adjustable and three rigid loop devices. Six samples of each device type were purchased. Each device was installed in a servo-hydraulic mechanical testing machine, running a 2000 cycle loading protocol at force increments between 50 and 500 N. Irreversible displacement in mm was measured for all of the tested samples of each implant. Ultimately, maximum load to failure was applied and measured in Nm. An irreversible displacement of 3 mm was considered failure of the implant.

Results

Three of the six adjustable devices (GraftMax™, TightRope® ToggleLoc™) demonstrated a median displacement above the threshold of clinical failure before completion of the cycles. All adjustable loop devices showed a wide intragroup variation in terms of irreversible displacement, compared to fixed-loop devices. Fixed-loop devices provided consistent reproducible results with narrow ranges and significantly lower irreversible displacement (p < 0.05), the maximum being 1.4 mm. All devices withstood an ultimate force of more than 500 N.

Conclusion

Adjustable loop devices still show biomechanical inferiority and demonstrate heterogeneity of fixation properties with wide- and less-reproducible displacement ranges resultant to the mechanism of adjustment, denoting less reliability. However, three adjustable devices (RIGIDLOOP™ Adjustable, Ultrabutton , ProCinch™) demonstrate fixation capacities within the margins of clinical acceptance. RIGIDLOOP™ Adjustable provides the most comparable fixation properties to fixed loop devices.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der Fachzeitschriften, inklusive eines Print-Abos.

Weitere Produktempfehlungen anzeigen
Zusatzmaterial
Supplementary material 1 (PDF 35 KB)
167_2017_4769_MOESM1_ESM.pdf
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 5/2018

Knee Surgery, Sports Traumatology, Arthroscopy 5/2018Zur Ausgabe
  1. Das kostenlose Testabonnement läuft nach 14 Tagen automatisch und formlos aus. Dieses Abonnement kann nur einmal getestet werden.

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Orthopädie und Unfallchirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise