16.08.2022 | Clinical Investigation

Agreement in the detection of chiasmal and postchiasmal visual field defects between imo binocular random single-eye test and Humphrey monocular test

verfasst von: Mari Sakamoto, Hiromasa Sawamura, Makoto Aihara, Toshiaki Goseki, Tetsuya Ikeda, Hitoshi Ishikawa, Makoto Nakamura

Erschienen in: Japanese Journal of Ophthalmology | Ausgabe 5/2022

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the ability of imo binocular random single-eye test (BRSET) to detect visual field (VF) defects due to chiasmal and postchiasmal lesions (C/PCLs) with a Humphrey Field Analyzer (HFA) monocular test.

Study design

Prospective multicenter study

Methods

This study enrolled 40 patients with C/PCLs and measured their VFs using both imo BRSET and HFA monocular test. The VFs were classified into three groups using the cluster criterion: 1) bitemporal group, 2) homonymous group, and 3) others. The agreement and correlation of VF results between imo and HFA were analyzed using the Bland–Altman plot and Spearman correlation coefficient.

Results

The VFs of 34 patients were analyzed and classified. There were 13 patients in the bitemporal, 6 in the homonymous, and 15 in the others group. BRSET showed a significantly shorter test duration than HFA. The imo systematically yielded a lower sensitivity than HFA. The average sensitivity at each test location correlated well between the perimeters in all groups, with the correlation coefficients ranging from 0.89 to 0.98. Bland–Altman plots showed wider limits of agreement in the affected quadrants compared to the unaffected quadrants in the bitemporal and homonymous groups. The fixation loss rate did not differ between the perimeters, but there were significant differences in the false positive and false negative rates between perimeters.

Conclusion

BRSET detected VF defects due to C/PCLs as accurately as the HFA monocular test with a shorter test duration.
Literatur
1.
Zurück zum Zitat Matsumoto C, Yamao S, Nomoto H, Takada S, Okuyama S, Kimura S, et al. Visual field testing with head-mounted perimeter “imo.” PLoS ONE. 2016;11: e0161974.CrossRef Matsumoto C, Yamao S, Nomoto H, Takada S, Okuyama S, Kimura S, et al. Visual field testing with head-mounted perimeter “imo.” PLoS ONE. 2016;11: e0161974.CrossRef
2.
Zurück zum Zitat Kitagawa A, Shimizu M, Yamanaka M. Experience in using “imo” 24plus (1) and comparison with HFA. Atarashii Ganka. 2018;35:1117–21 (in Japanese). Kitagawa A, Shimizu M, Yamanaka M. Experience in using “imo” 24plus (1) and comparison with HFA. Atarashii Ganka. 2018;35:1117–21 (in Japanese).
3.
Zurück zum Zitat Hayashi Y, Sakamoto M, Murai Y, Nishisho R, Hayashida M, Mori S, et al. Utility of the imoⓇ “binocular random single-eye test” in glaucoma practice. Nippon Ganka Gakkai Zasshi. 2021;125:530–8 (in Japanese). Hayashi Y, Sakamoto M, Murai Y, Nishisho R, Hayashida M, Mori S, et al. Utility of the imo “binocular random single-eye test” in glaucoma practice. Nippon Ganka Gakkai Zasshi. 2021;125:530–8 (in Japanese).
4.
Zurück zum Zitat Kimura T, Matsumoto C, Nomoto H. Comparison of head-mounted perimeter (imo((R))) and Humphrey Field Analyzer. Clin Ophthalmol. 2019;13:501–13.CrossRef Kimura T, Matsumoto C, Nomoto H. Comparison of head-mounted perimeter (imo((R))) and Humphrey Field Analyzer. Clin Ophthalmol. 2019;13:501–13.CrossRef
5.
Zurück zum Zitat Goseki T, Ishikawa H, Shoji N. Bilateral Concurrent eye examination with a head-mounted perimeter for diagnosing functional visual loss. Neuroophthalmology. 2016;40:281–5.CrossRef Goseki T, Ishikawa H, Shoji N. Bilateral Concurrent eye examination with a head-mounted perimeter for diagnosing functional visual loss. Neuroophthalmology. 2016;40:281–5.CrossRef
6.
Zurück zum Zitat Wall M, Punke SG, Stickney TL, Brito CF, Withrow KR, Kardon RH. SITA standard in optic neuropathies and hemianopias: a comparison with full threshold testing. Invest Ophthalmol Vis Sci. 2001;42:528–37.PubMed Wall M, Punke SG, Stickney TL, Brito CF, Withrow KR, Kardon RH. SITA standard in optic neuropathies and hemianopias: a comparison with full threshold testing. Invest Ophthalmol Vis Sci. 2001;42:528–37.PubMed
7.
Zurück zum Zitat Fujimoto N, Saeki N, Miyauchi O, Adachi-Usami E. Criteria for early detection of temporal hemianopia in asymptomatic pituitary tumor. Eye (Lond). 2002;16:731–8.CrossRef Fujimoto N, Saeki N, Miyauchi O, Adachi-Usami E. Criteria for early detection of temporal hemianopia in asymptomatic pituitary tumor. Eye (Lond). 2002;16:731–8.CrossRef
8.
Zurück zum Zitat Szatmary G, Biousse V, Newman NJ. Can Swedish interactive thresholding algorithm fast perimetry be used as an alternative to goldmann perimetry in neuro-ophthalmic practice? Arch Ophthalmol. 2002;120:1162–73.CrossRef Szatmary G, Biousse V, Newman NJ. Can Swedish interactive thresholding algorithm fast perimetry be used as an alternative to goldmann perimetry in neuro-ophthalmic practice? Arch Ophthalmol. 2002;120:1162–73.CrossRef
9.
Zurück zum Zitat Danesh-Meyer HV, Carroll SC, Gaskin BJ, Gao A, Gamble GD. Correlation of the multifocal visual evoked potential and standard automated perimetry in compressive optic neuropathies. Invest Ophthalmol Vis Sci. 2006;47:1458–63.CrossRef Danesh-Meyer HV, Carroll SC, Gaskin BJ, Gao A, Gamble GD. Correlation of the multifocal visual evoked potential and standard automated perimetry in compressive optic neuropathies. Invest Ophthalmol Vis Sci. 2006;47:1458–63.CrossRef
10.
Zurück zum Zitat Feinstein AR, Cicchetti DV. High agreement but low kappa: I. The problems of two paradoxes. J Clin Epidemiol. 1990;43:543–9.CrossRef Feinstein AR, Cicchetti DV. High agreement but low kappa: I. The problems of two paradoxes. J Clin Epidemiol. 1990;43:543–9.CrossRef
11.
Zurück zum Zitat Cicchetti DV, Feinstein AR. High agreement but low kappa: II. Resolving the paradoxes. J Clin Epidemiol. 1990;43:551–8.CrossRef Cicchetti DV, Feinstein AR. High agreement but low kappa: II. Resolving the paradoxes. J Clin Epidemiol. 1990;43:551–8.CrossRef
12.
Zurück zum Zitat Zec S, Soriani N, Comoretto R, Baldi I. High Agreement and High Prevalence: The paradox of Cohen’s Kappa. Open Nurs J. 2017;11:211–8.CrossRef Zec S, Soriani N, Comoretto R, Baldi I. High Agreement and High Prevalence: The paradox of Cohen’s Kappa. Open Nurs J. 2017;11:211–8.CrossRef
13.
Zurück zum Zitat Gwet KL. Computing inter-rater reliability and its variance in the presence of high agreement. Br J Math Stat Psychol. 2008;61:29–48.CrossRef Gwet KL. Computing inter-rater reliability and its variance in the presence of high agreement. Br J Math Stat Psychol. 2008;61:29–48.CrossRef
14.
Zurück zum Zitat Goukon H, Hirasawa K, Kasahara M, Matsumura K, Shoji N. Comparison of Humphrey Field Analyzer and imo visual field test results in patients with glaucoma and pseudo-fixation loss. PLoS ONE. 2019;14: e0224711.CrossRef Goukon H, Hirasawa K, Kasahara M, Matsumura K, Shoji N. Comparison of Humphrey Field Analyzer and imo visual field test results in patients with glaucoma and pseudo-fixation loss. PLoS ONE. 2019;14: e0224711.CrossRef
15.
Zurück zum Zitat Bengtsson B, Olsson J, Heijl A, Rootzen H. A new generation of algorithms for computerized threshold perimetry, SITA. Acta Ophthalmol Scand. 1997;75:368–75.CrossRef Bengtsson B, Olsson J, Heijl A, Rootzen H. A new generation of algorithms for computerized threshold perimetry, SITA. Acta Ophthalmol Scand. 1997;75:368–75.CrossRef
16.
Zurück zum Zitat Heijl A, Patella VM, Flanagan JG, Iwase A, Leung CK, Tuulonen A, et al. False positive responses in standard automated perimetry. Am J Ophthalmol. 2022;233:180–8.CrossRef Heijl A, Patella VM, Flanagan JG, Iwase A, Leung CK, Tuulonen A, et al. False positive responses in standard automated perimetry. Am J Ophthalmol. 2022;233:180–8.CrossRef
17.
Zurück zum Zitat Bengtsson B, Heijl A. False-negative responses in glaucoma perimetry: indicators of patient performance or test reliability? Invest Ophthalmol Vis Sci. 2000;41:2201–4.PubMed Bengtsson B, Heijl A. False-negative responses in glaucoma perimetry: indicators of patient performance or test reliability? Invest Ophthalmol Vis Sci. 2000;41:2201–4.PubMed
18.
Zurück zum Zitat Bengtsson B. Reliability of computerized perimetric threshold tests as assessed by reliability indices and threshold reproducibility in patients with suspect and manifest glaucoma. Acta Ophthalmol Scand. 2000;78:519–22.CrossRef Bengtsson B. Reliability of computerized perimetric threshold tests as assessed by reliability indices and threshold reproducibility in patients with suspect and manifest glaucoma. Acta Ophthalmol Scand. 2000;78:519–22.CrossRef
19.
Zurück zum Zitat Yohannan J, Wang J, Brown J, Chauhan BC, Boland MV, Friedman DS, et al. Evidence-based criteria for assessment of visual field reliability. Ophthalmology. 2017;124:1612–20.CrossRef Yohannan J, Wang J, Brown J, Chauhan BC, Boland MV, Friedman DS, et al. Evidence-based criteria for assessment of visual field reliability. Ophthalmology. 2017;124:1612–20.CrossRef
Metadaten
Titel
Agreement in the detection of chiasmal and postchiasmal visual field defects between imo binocular random single-eye test and Humphrey monocular test
verfasst von
Mari Sakamoto
Hiromasa Sawamura
Makoto Aihara
Toshiaki Goseki
Tetsuya Ikeda
Hitoshi Ishikawa
Makoto Nakamura
Publikationsdatum
16.08.2022
Verlag
Springer Japan
Erschienen in
Japanese Journal of Ophthalmology / Ausgabe 5/2022
Print ISSN: 0021-5155
Elektronische ISSN: 1613-2246
DOI
https://doi.org/10.1007/s10384-022-00935-y

Weitere Artikel der Ausgabe 5/2022

Japanese Journal of Ophthalmology 5/2022 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Die überschießende Wundheilung in der filtrierenden Glaukomchirurgie ist ein zentraler Faktor für ein operatives Versagen. Nach der Einführung der Trabekulektomie in den 1960er-Jahren wurden viele Faktoren erkannt, die mit einer vermehrten …

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.