Open Access 01.12.2017 | Research article

An objective assessment of the variability in number of drops per bottle of glaucoma medication

verfasst von: Daniel B. Moore, Judy Beck, Richard J Kryscio

Erschienen in: BMC Ophthalmology | Ausgabe 1/2017

Abstract

Background

The purpose of this study is to evaluate the number of eyedrops available per bottle of a variety of commonly prescribed glaucoma medications.

Methods

Six bottles of each glaucoma medication were tested: three each in the vertical and horizontal orientations. Bottles were housed in a customized force gauge apparatus designed to mimic ballpoint fingertip contact with a bottle. At a standard rate, all drops were expressed from each bottle and counted with an automated drop counter. Simultaneously, bottle volume was measured and drop size and number were also estimated. The main outcome measures were: total number of drops, volume per bottle and drops per milliliter (mL) of glaucoma medication.

Results

A total of 192 bottles from 32 bottle designs and manufacturers were tested. Twenty-two of the 32 bottle designs had a significantly different mean number of drops in the vertical and horizontal positions, with 10 designs have more drops dispensed in the horizontal orientation and 12 in the vertical orientation. Six of the 32 bottle designs had a significantly different mean total bottle volume in the vertical and horizontal positions, with all designs having greater volume in the vertical position. An adjusted ratio of mean number of drops/mean bottle volume demonstrated a range from 20.9 drops/mL to 40.8 drops/mL.

Conclusions

There is significant variability in drops and volume available per bottle of glaucoma medication depending on both the bottle position and manufacturer. These data point to the need for circumspection in prescribing glaucoma medications and caution in evaluating therapeutic outcomes.
Abkürzungen
mL
milliliter
mm
millimeter
μl
microliter

Background

Although data demonstrate improved outcomes with lowered intraocular pressure from appropriate pharmacotherapy [1, 2], many patients with glaucoma struggle to adhere to their prescribed regimens. Several studies suggest patients comply with 70% or fewer of their glaucoma medications [35]. Many factors have been implicated [6], but difficulty instilling drops has garnered recent attention [7]. Proper eyedrop administration requires eye-hand coordination and dexterity, linking visual acuity with a steady hand and accurate proprioception [8]. Not surprisingly, videotape evidence demonstrated that glaucoma patient used an average of 1.4–1.8 drops when trying to instill a single eyedrop [9, 10]. A recent cross-sectional patient survey revealed that 25% of patients reported problems with early eyedrop bottle exhaustion and associated compliance with therapy. One-third of patients reported the reason for early bottle exhaustion was due to bottle related problems such as “more than one drop comes out” or “size of drops is too large”. The latter complaint is supported by literature that demonstrates the volume of an eyedrop in an ophthalmic solution may vary from 25 to 70 μl [11]. Given the normal tear film volume is 7 μl and only capable of containing 30 μl without overflow, a significant portion of an eyedrop is wasted [12]. These data suggest that a significant number of glaucoma patients run out of eyedrops prior to a scheduled prescription refill and that bottle related mechanics play a role [13].
Despite the importance of bottle design in proper use of topical therapeutics, there exists no standardization of manufacture in regards to drop instillation dynamics [14], and the only dosing requirement is to accurately label and package medication volume [15]. Being manufactured and sold on the basis of volume, eyedrops are distinct from most other forms of pharmacotherapy, which are prescribed with a discrete number of doses to guide dispensing and refill rates. While the minimum volume of medication consumers should anticipate per container of medication is available, this does not necessarily translate to number of applications, and without regulation, leads to the possibility of inconsistency in the number of drops of medicine available per bottle. Several previous studies evaluating a smaller number of medications in a non-standardized fashion have found significant variability in the number of drops between both brands of medication and position of the bottle [1618]. As such, the current study was designed to further evaluate this possibility by objectively and systematically measuring the number of eyedrops in each bottle of many common glaucoma medications.

Methods

The number of eyedrops dispensed from various common glaucoma medications was measured. All medications were purchased at cost from the University of Kentucky Research Pharmacy and represented available regional brand and generic medications. A force gauge apparatus consisting of a Mecmesin M500E Motorised Tension and Compression Test Stand, Mecmesin 100 N Advanced Force Gauge (Mecmesin Corporation, Sterling, VA, USA) and custom grips and compressors were designed and calibrated by JA King & Company (Whitsett, NC, USA) (Fig. 1). The compressors were designed to mimic ballpoint fingertip contact with a bottle. For each medication, the bottle was housed in the apparatus and clamps were adjusted until the ballpoint compressors were located at mid bottle length. For bottles with a rectangular instead of round shape, the thinner dimensions were chosen for compression, as this represents the method most likely to be utilized by patients when instilling drops. Starting at 0 kg-force (kgf) and 0 mm (mm) displacement, the gauge was advanced in 0.1 mm increments until a drop of liquid fell from the bottle, as observed subjectively and confirmed with an automated VCD-BTD drop counter (Vernier Software and Technology, Beaverton, OR, USA) and LabQuest 2 display (Vernier Software and Technology, Beaverton, OR, USA). At a rate of approximately one drop/s, 10 drops were expressed, then the apparatus was retracted to 0 kgf. This was repeated until all drops were exhausted from the bottle.
Simultaneously, drop size and number was also estimated using the densitometric method for volume determination [19]. In twenty drop increments, the total volume of solution expressed was measured with a 0.0001 g analytical balance (Ohaus Corporation, Parsippany, NJ, USA). This was repeated until all drops were exhausted from the bottle. A 200 uL pipette (Zhejiang Huawei Scientific Instrument Co, LTD, Zhe Jiang, China) was used to remove four 100-uL aliquots of each bottle. The mean of the samples was divided by 0.1 mL to estimate the volume of each drop and each bottle by dividing the mass of each by the calculated density. For any bottles with residual liquid in the container lid, this was removed with the pipette and volume was measured separately.
Six bottles of each medication were tested. Three bottles were tested in the vertical orientation with the bottle tip at 180 degrees and three bottles were tested in the near horizontal orientation with the bottle tip at 30 degrees. The vertical and horizontal orientations were the starting position for the bottle tip during each measurement, as compression of the bottle variably and slightly changed the tip position.

Statistical analysis

Mean response was compared by constructing an analysis of variance for a two way layout with factors: position (horizontal versus vertical) and bottle (all combinations of medication name and formulation). A highly significant interaction between position and bottle was obtained (p < 0.0001). Post hoc comparison of means was done by comparing means between positions for each bottle by using two sample t-tests. To compare different bottle designs in the same orientation, Fisher’s least significant differences allowance was computed. Statistical significance was determined at the 0.01 level to minimize the Type I error rate.

Results

A total of 192 bottles from 32 bottle designs and manufacturers of medication were tested (Table 1). Further reference to medications will include brand name, manufacturer if more than one generic and bottle volume for identification. Fifteen formulations came from the same lot, sixteen came from a combination of two different lots and one came from a combination of three different lots. When comparing the mean value of variance between formulations from the same or different lots, there was no statistical difference (p = 0.62). Comparing mean number of drops per bottle using observed measurements versus the densitometric method to calculate number of drops in the bottle, there was no difference in the vertical (p = 0.35, paired t-test), but a significant difference between measurements in the horizontal (p = 0.02, paired t-test) position. Observed measurements via the automated drop counter were used for subsequent analysis.
Table 1
Description of Glaucoma Medications Tested
Medication Name
Formulation
Manufacturer
2.5 ml
 Travatan
travaprost 0.004%
Alcon Laboratories, Inc. Fort Worth, TX 76134
 Travaprost
travaprost 0.004%
Par Pharmaceutical Cos. Inc. Spring Valley, NY 10977
 Xalatan
latanoprost 0.005%
Pharmacia&Upjohn Co, Division of Pfizer, Inc. NY, NY 10017
 Latanoprost
latanoprost 0.005%
Akorn, Inc. Lake Forest, IL 60045
 Lumigan
bimatoprost 0.001%
Allergan, Inc. Irvine, CA 92612
5 ml
 Lumigan
bimatoprost 0.001%
Allergan, Inc. Irvine, CA 92612
 Travaprost
travaprost 0.004%
Par Pharmaceutical Cos. Inc. Spring Valley, NY 10977
 Travatan
travaprost 0.004%
Alcon Laboratories, Inc. Fort Worth, TX 76134
 Alphagan P 0.1%
brimonidine tartrate 0.1%
Allergan, Inc. Irvine, CA 92612
 Combigan
brimonidine tartrate 0.2%, timolol maleate 0.5%
Allergan, Inc. Irvine, CA 92612
 Timolol  Pacific
timolol maleate 0.5%
Pacific Pharma, Irvine CA, 92,612
 Timolol Hi-Tech
timolol maleate 0.5%
Hi-Tech Pharmacal Co, Inc. Amityville, NY 11701
 Timolol Sandoz
timolol maleate 0.5%
Alcon Laboratories, Inc. Fort Worth, TX 76134 for Sandoz Inc. Pricenton, NJ 08540
 Betimol
timolol maleate 0.5%
Akorn, Inc. Lake Forest, IL 60045 for Oak Pharmaceuticals, Inc.
 Timoptic XE
timolol maleate 0.5%
Merck Sharp & Dohme-Chibret 63,963 Clermont-Ferrand Cedex 9, France; Distributed by Valeant Ophthalmics, a division of Valeant Pharmaceuticals North America LLC, Bridgewater, NJ 08807
 Timoptic
timolol maleate 0.5%
Merck Sharp & Dohme-Chibret 63,963 Clermont-Ferrand Cedex 9, France; Distributed by Valeant Ophthalmics, a division of Valeant Pharmaceuticals North America LLC, Bridgewater, NJ 08807
 Istalol
timolol maleate 0.5%
Bausch&Lomb Incorporated Tampa, FL 33637 under License from Senju Pharmaceutical Co, Ltd. Osaka, Japan 541–0046
7.5 ml
 Lumigan
bimatoprost 0.001%
Allergan, Inc. Irvine, CA 92612
8 ml
 Simbrinza
brinzolamide 1%, brimonidine tartrate 0.2%
Alcon Laboratories, INC 6201 South Freeway, Fort Worth, TX 76134
10 ml
 Cosopt
dorzolamide HCL 22.3 mg/ml, timolol maleate 6.8 mg/ml
Akorn, Inc. Lake Forest, IL 60045
 Dorz/Tim Hi-Tech
dorzolamide HCL 22.3 mg/ml, timolol maleate 6.8 mg/ml
Hi-Tech Pharmacal Co, Inc. Amityville, NY 11701
 Dorz/Tim Bausch
dorzolamide HCL 22.3 mg/ml, timolol maleate 6.8 mg/ml
Bausch & Lomb Incorporated Tampl, FL 33637
 Dorz/Tim Sandoz
dorzolamide HCL 22.3 mg/ml, timolol maleate 6.8 mg/ml
Alcon Laboratories Inc. Fort Worth Tx 76,134 for Sandoz Inc. Princeton, NJ 08540
 Trusopt
dorzolamide HCL 2%
Merck Sharp & Dohme Corp.,a subsidiary of Merck & Co, Inc. Whitehouse Station, NJ 08889
 Dorz Hi Tech
dorzolamide HCL 2%
Hi-Tech Pharmacal Co, Inc. Amityville, NY 11701
 Dorz Bausch
dorzolamide HCL 2%
Bausch & Lomb Incorporated Tampa, FL 33637
 Dorz Teva
dorzolamide HCL 2%
Teva Pharmaceutical Ind, Ltd. Jerusalem, 91,010, Israel for Teva Pharmaceuticals USA Sellersville, PA 18960
 Dorz Sandoz
dorzolamide HCL 2%
Alcon Laboratires, Inc. Fort Worth, TX 76134 for Sandoz Inc. Princeton, NJ 08540
 Azopt
brinzolamide 1%
Alcon Laboratories, Inc. Fort Worth, TX 76134
 Combigan
brimonidine tartrate 0.2%, timolol maleate 0.5%
Allergan, Inc. Irvine, CA 92612
 Alphagan P 0.15%
brimonidine tartrate 0.15%
Allergan, Inc. Irvine, CA 92612
 Alphagan P 0.1%
brimonidine tartrate 0.1%
Allergan, Inc. Irvine, CA 92612
The medication name and volume used in the study, the formulation and manufacturer are listed
The mean number and standard deviation of drops per bottle in the horizontal, vertical and summative positions of all formulations are provided in Table 2. For 2.5 mL bottles, the mean number of drops ranged from 75.3–101.7 and 72–102.3 in the vertical and horizontal orientations, respectively. For 5 mL bottles, the range was 111–209.3 and 115–189 drops in the vertical and horizontal orientations, respectively. For 10 mL bottles, the range was 193.7–313.3 and 234–323.7 drops in the vertical and horizontal orientations, respectively. Twenty-two of the 32 bottle designs had a significantly different mean number of drops in the vertical and horizontal positions, with 10 designs have more drops dispensed in the horizontal orientation and 12 in the vertical.
Table 2
Mean Number of Drops, Volume and Drop/Volume Ratio per Glaucoma Medication Tested
Drop
Mean # Drops
 
Mean Bottle Volume
Mean Drop Volume
Mean Drops/mL
vert
horz
p value
vert
horz
p value
vert
horz
vert
horz
p value
2.5 ml
 Travatan
76.3 ± 4.2
102.3 ± 1.5
<0.01
2.57 ± 0.07
2.53 ± 0.04
 
0.032
0.025
29.66 ± 1.35
40.4 ± 1.07
0.0004
 Travaprost
75.3 ± 5.1
72 ± 1
 
2.46 ± 0.04
2.36 ± 0.07
 
0.032
0.032
30.6 ± 2.51
30.47 ± 1.14
0.9445
 Xalatan
101.3 ± 0.6
83.3 ± 3.1
 
2.69 ± 0.05
2.70 ± 0.05
 
0.027
0.032
37.68 ± 0.73
30.90 ± 0.60
0.0002
 Latanoprost
86.7 ± 9.3
76.7 ± 5.0
 
2.37 ± 0.26
2.42 ± 0.13
 
0.027
0.032
36.59 ± 0.98
31.63 ± 0.76
0.0023
 Lumigan
101.7 ± 2.3
81.7 ± 4.0
<0.01
2.50 ± 0.05
2.38 ± 0.08
 
0.024
0.029
40.61 ± 0.18
34.25 ± 0.63
<0.0001
 overall mean
88.3
83.2
 
2.52
2.48
 
0.029
0.03
   
 overall median
91
81
 
2.53
2.46
 
0.027
0.031
  
 overall ST DEV
12.7
11.08
 
0.15
0.15
 
0.004
0.003
  
5 ml
 Lumigan
195.6 ± 4.2
155 ± 1
<0.01
4.97 ± 0.06
4.73 ± 0.06
 
0.025
0.03
39.4 ± 0.69
32.75 ± 0.22
<0.001
 Travaprost
150.3 ± 6.5
139.7 ± 13.7
 
4.91 ± 0.02
4.78 ± 0.24
 
0.032
0.035
30.62 ± 1.42
29.15 ± 1.50
0.2859
 Travatan
146.3 ± 0.6
189 ± 1
<0.01
4.8 ± 0.04
4.78 ± 0.07
 
0.033
0.026
30.49 ± 0.82
29.66 ± 1.35
0.0004
 Alphagan P
111 ± 1
117 ± 2.6
 
4.89 ± 0.05
4.72 ± 0.18
 
0.044
0.04
22.7 ± 0.16
24.82 ± 0.78
0.0102
 Combigan
148 ± 2
161.3 ± 3.8
 
5.02 ± 0.16
4.67 ± 0.07
0.01
0.033
0.029
25.26 ± 1.53
34.57 ± 0.50
0.0003
 Timolol Pacific
123 ± 1
147.7 ± 9.6
<0.01
4.88 ± 0.04
4.69 ± 0.10
 
0.04
0.031
25.21 ± 0.09
25.21 ± 0.09
0.0206
 Timolol Hi-Tech
126.7 ± 6.7
149.3 ± 4.0
<0.01
5.09 ± 0.15
5.04 ± 0.13
 
0.041
0.034
24.89 ± 0.56
29.61 ± 0.633
0.0006
 Timolol Sandoz
161.3 ± 3.5
156.3 ± 2.5
 
4.89 ± 0.04
4.84 ± 0.08
 
0.03
0.031
33.02 ± 0.92
32.28 ± 0.21
0.2459
 Betimol
165.3 ± 0.6
160.7 ± 1.2
 
5.17 ± 0.035
5.08 ± 0.06
 
0.031
0.031
32.00 ± 0.14
31.63 ± 0.56
0.3282
 Timoptic XE
130.7 ± 7.5
115 ± 9.8
0.004
5.53 ± 0.24
4.9 ± 0.24
 
0.042
0.043
23.71 ± 2.41
23.45 ± 1.16
0.8734
 Timoptic
209.3 ± 5.7
155.3 ± 7.1
<0.01
5.93 ± 0.03
5.66 ± 0.03
0.0003
0.089
0.036
35.28 ± 0.82
27.6 ± 1.21
0.0008
 Istalol
154.6 ± 2.1
165.6 ± 2.5
 
4.72 ± 0.03
4.74 ± 0.09
 
0.03
0.029
32.77 ± 0.22
34.95 ± 0.32
0.0006
 overall mean
151.9
151
 
5.06
4.89
 
0.039
0.033
   
 overall median
149
155.2
 
4.93
4.78
 
0.033
0.031
  
 overall ST DEV
28.2
20.26
 
0.34
0.28
 
0.03
0.005
  
7.5 ml
 Lumigan
294 ± 2.6
236 ± 2
<0.01
7.21 ± 0.06
7.18 ± 0.03
 
0.024
0.03
40.76 ± 0.56
32.85 ± 0.18
<0.001
8 ml
 Simbrinza
180.7 ± 3.1
219.7 ± 6.7
<0.01
7.51 ± 0.03
6.74 ± 0.12
0.0004
0.167
0.031
24.07 ± 0.47
32.58 ± 1.22
0.0003
10 ml
 Cosopt
301.7 ± 39.6
242 ± 12.5
<0.01
10.2 ± 0.17
9.58 ± 0.18
0.01
0.033
0.039
29.56 ± 3.66
25.26 ± 1.53
0.1341
 Dorz/Tim Hi-Tech
303 ± 1
251.7 ± 10.0
<0.01
9.23 ± 0.34
9.19 ± 0.26
 
0.031
0.038
32.85 ± 1.22
27.39 ± 0.39
0.0018
 Dorz/Tim Bausch
313.3 ± 1.5
293.7 ± 1.5
<0.01
9.22 ± 0.23
8.91 ± 0.21
 
0.03
0.031
34.01 ± 0.92
32.96 ± 0.65
0.1804
 Dorz/Tim Sandoz
315 ± 4.6
271.3 ± 5.8
<0.01
8.93 ± 0.15
8.39 ± 0.29
 
0.115
0.031
35.27 ± 1.09
32.37 ± 1.03
0.0288
 Trusopt
313.3 ± 20.0
236 ± 5.6
<0.01
10.18 ± 0.20
9.60 ± 0.33
 
0.221
0.051
30.77 ± 1.86
24.59 ± 0.94
0.0068
 Dorz Hi Tech
292.7 ± 10.3
261.7 ± 12.9
<0.01
9.54 ± 0.19
9.13 ± 0.29
 
0.033
0.035
20.92 ± 17.91
28.70 ± 2.21
0.5313
 Dorz Bausch
289.7 ± 4.2
289 ± 4.6
 
9.18 ± 0.26
8.78 ± 0.15
 
0.032
0.03
31.55 ± 0.61
32.91 ± 0.06
0.0594
 Dorz Teva
282 ± 11.5
260 ± 1
<0.01
9.31 ± 0.58
8.69 ± 0.12
 
0.033
0.034
30.33 ± 0.79
29.91 ± 0.31
0.4423
 Dorz Sandoz
193.7 ± 3.79
253.7 ± 15.3
<0.01
9.19 ± 0.12
8.38 ± 0.08
0.0006
0.332
0.034
21.06 ± 0.16
30.26 ± 1.94
0.014
 Azopt
254.3 ± 2.1
322 ± 8.9
<0.01
9.42 ± 0.08
8.59 ± 0.04
<0.0001
0.037
0.027
27.01 ± 0.38
37.5 ± 1.08
<0.001
 Combigan
270.3 ± 28.4
323.7 ± 17.8
<0.01
9.86 ± 0.21
9.79 ± 0.07
 
0.034
0.03
27.4 ± 2.50
33.07 ± 1.64
0.0303
 Alphagan 0.15%
212.7 ± 3.1
234 ± 3.6
<0.01
9.78 ± 0
9.53 ± 0.09
 
0.046
0.041
21.75 ± 0.31
24.56 ± 0.16
0.0002
 Alphagan 0.1%
216 ± 4
283.3 ± 5.5
<0.01
9.93 ± 0.19
9.61 ± 0.07
 
0.046
0.037
21.75 ± 0.11
24.80 ± 0.41
0.0002
 overall mean
273.7
267.5
 
9.54
9.09
 
0.079
0.034
   
 overall median
290
261
 
9.51
9.02
 
0.034
0.033
  
 overall ST DEV
42.69
31.07
 
0.45
0.51
 
0.119
0.005
  
The mean number of drops, bottle volume, drop volume and ratio of mean number of drops per bottle volume of three bottles tested for each medication in the vertical and horizontal positions. Medications are grouped by volume and an overall mean, median and standard deviation for all medications of similar volumes are provided. A p-value for medications with a significant difference in mean drop number between the horizontal and vertical position is listed, with significance determined at p = 0.01. A ratio of mean number of drops/mean bottle volume is listed, and any two means that differ by a value of 5.27 or greater are significantly different (p < 0.01). Volume measurements are in milliliters (mL). A bold value indicates a p value of statistical significance
The mean bottle and drop volume are also listed in Table 2. Bottle volumes were measured with and without accounting for residual volume in the cap of the container, with less than a 2.2% difference of the overall mean in 90% of the bottles. Presented bottle volume measurements include the cap volume. For 2.5 mL bottles, the range was 2.37–2.69 mL and 2.38–2.70 mL in the vertical and horizontal orientations, respectively. For 5 mL bottles, the range was 4.72–5.93 mL and 4.67–5.66 mL in the vertical and horizontal orientations, respectively. For 10 mL bottles, the range was 8.93–10.2 mL and 8.38–9.79 mL in the vertical and horizontal orientations, respectively. Six of the 32 bottle designs had a significantly different mean total bottle volume in the vertical and horizontal positions, with all designs having greater volume in the vertical position. There were no significant differences between the horizontal and vertical measured drop volume for any designs.
Given the multiple different comparisons possible between bottle type, size and volume, an adjusted ratio of mean number of drops/mean bottle volume was created (Table 2), with a range from 20.9 drops/mL to 40.8 drops/mL among the bottle designs and positions. Nineteen formulations had a significantly different ratio between the horizontal and vertical positions (p < 0.01). For comparing means between two bottle designs in the same orientation, any two means that differ by a value of 5.27 or greater are significantly different (p < 0.01).

Discussion

The results of this study suggest there is significant variability in the number of drops and volume per bottle of glaucoma medications, both in terms of which bottle design and manufacturer is utilized and the position the bottle is held when squeezed. For example, if a representative patient from this region is prescribed a 5 mL bottle of timolol 0.5%, there are 5 different generic and brand bottle designs that may be dispensed. The patient could anticipate a range of 123–209 and 147–166 mean drops per bottle in the horizontal and vertical positions, respectively. If instilled twice daily, this would suggest a difference between 25.5 to 43 days of available medication depending on the manufacturer and bottle position.
This study is the first to our knowledge to present an objective, automated and reproducible method to measure the number of drops available per bottle of medication. Further, we measured countable drops instead of calculating the number of drops based on volume, which was shown to be inaccurate. Several previous studies have evaluated small samples of bottle formulations with similar, variable results. A 1994 study of patients blinded to either use of a 5 mL bottle of timolol maleate versus levobunolol found a 21% greater length of use of timolol (37 versus 29 days) [16]. Another study evaluated 45 versus 90 degree administration of artificial tear bottles and found a significantly greater number of drops per bottle using densitometric analysis and smaller drop volume for 4 out of 5 formulations at 45 compared to 90 degrees. The authors suggest administration at 45 degrees would result in up to $1.93 savings per bottle compared to 90 degrees [17]. A study evaluating 2.5 ml bottles of prostaglandin analogues held vertically, at 45 degrees and horizontally found that vertical instillation resulted in more drops per bottle for bimatoprost and latanoprost, while 45 degrees was most efficient for travaprost. Assuming 1 year of bilateral therapy at 2006 costs, the authors determined use of the most efficient instillation method would result in yearly savings of $109–192 [18]. Lastly, a recent study found significant variability in the number of eyedrops per bottle of four regionally available formulations of latanoprost when measured by manually counting the number of drops expressed by hand, ranging from 77.6 to 88.7 drops per bottle. The authors estimated a similarly significant difference in estimated annual cost, ranging from $184 to $1198 per formulation [20].
Currently, there exist no federal guidelines to regulate bottle design or amount of drops available per volume of medication [14, 15]. It has been suggested that pharmacists often use a rough guideline of 0.05 mL per eyedrop or 20 drops per mL (written personal communication, Division of Drug Information, FDA, March 27, 2015). With that in mind, there are significantly more eyedrops per bottle in this analysis than recommended, ranging from 10.4% to 45.8% more mean drops per bottle tested. This may indicate that manufacturers “overfill” the bottles to allow for a margin of error during dispensing. If true, this serves an important purpose, since it is known that many patients require more than one eyedrop per application [9, 10]. However, this data suggests another problem has been created by this practice: significant variability in the number of doses per bottle.
Aside from bottle volume, an additional factor influence the number of drops available per bottle is drop size. We found significant variability in the estimated drop size of studied formulations, ranging from 0.024 to 0.221 mL. The size of drop dispensed from a bottle depends on three basic elements: the design of the dropper bottle and tip, properties of the contained solution and the position of the bottle. The surface area around the bottle tip and surface tension of the solution are both manufacturer controlled factors that influence drop size. Patient manipulations such as the angle and rate drops are produced are less predictable [11]. Further, the results of the current study and others suggest the most economical bottle position varies from one design to the next [1719]. A final potential determinant of drop size is the force required to squeeze the bottle, which unfortunately is significantly variable in both the experimental and clinical setting [21, 22]. With all of these influences, it may be difficult to design an ideal bottle for instillation. One expert suggestion has been to utilize a dropper tip with a smaller outer orifice diameter that provides consistent surface area for a smaller-volume drop to fall [11].
This study has several limitations. The experimental design was novel, and although measurements were automated, they have not been independently verified. Because the study was conducted in an objective and reproducible manner, it may not accurately reflect many of the patient related factors in dosing; it is likely that our patients experience even greater variability in the number of drops available per bottle. Only two bottle positions were tested and most ophthalmic containers are not intended to be delivered in a strictly horizontal position. While 192 bottles from 32 bottle designs were tested, they still represent a small sampling of all available brand and generic ophthalmic medications.

Conclusion

In summary, this study demonstrates the significant variability in drops and volume available per bottle of glaucoma medication depending on both the bottle position and manufacturer. This unregulated practice leaves prescribing physicians and pharmacists unable to accurately predict the quantity of medication to dispense. This may lead to patients running out of medication early or being left with excess and associated costs. The variability from one refill to the next could be a contributor to limited compliance. The experimental design in this study indicates an objective, reproducible method to determine drop number uniformly across different bottles and designs. It should compel further evaluation and consideration of standardization in the industry.

Acknowledgments

The authors would like to acknowledge Dr. Jayakrishna Ambati (University of Kentucky, Department of Ophthalmology & Visual Sciences) for his insight and guidance with the manuscript preparation.

Funding

The project described was supported by the National Center for Research Resources and the National Center for Advancing Translational Sciences, National Institutes of Health, through Grant UL1TR000117. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH.

Availability of data and materials

The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors’ contributions

DBM participated in all aspects of the study. JB aided in study design, acquisition of resources and review of the manuscript. RJK aided in study design, provided statistical support and analysis and review of the manuscript. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.
Not applicable.
Not applicable.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Leske MC, Heijl A, Hussein M, et al. Early manifest glaucoma trial group. Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol. 2003;121(1):48–56.CrossRefPubMed Leske MC, Heijl A, Hussein M, et al. Early manifest glaucoma trial group. Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol. 2003;121(1):48–56.CrossRefPubMed
2.
Zurück zum Zitat The AGIS Investigators. The advanced glaucoma interventional study (AGIS): 7, the relationship between control of intraocular pressure and visual field deterioration. Am J Ophthalmol. 2000;130(4):429–40.CrossRef The AGIS Investigators. The advanced glaucoma interventional study (AGIS): 7, the relationship between control of intraocular pressure and visual field deterioration. Am J Ophthalmol. 2000;130(4):429–40.CrossRef
3.
Zurück zum Zitat Gordon ME, Kass MA. Validity of standard compliance measures in glaucoma compared with an electronic eye drop monitor. In: Cramer JA, Spilker B, editors. Patient compliance in medical practice and clinical trials. New York: Raven Press; 1991. p. 163–73. Gordon ME, Kass MA. Validity of standard compliance measures in glaucoma compared with an electronic eye drop monitor. In: Cramer JA, Spilker B, editors. Patient compliance in medical practice and clinical trials. New York: Raven Press; 1991. p. 163–73.
4.
Zurück zum Zitat Kass MA, Meltzer DW, Gordon M, et al. Compliance with topical pilocarpine treatment. Am J Ophthalmol. 1986;101(5):515–23.CrossRefPubMed Kass MA, Meltzer DW, Gordon M, et al. Compliance with topical pilocarpine treatment. Am J Ophthalmol. 1986;101(5):515–23.CrossRefPubMed
5.
Zurück zum Zitat Okeke CO, Quigley HA, Jampel HD, et al. Interventions improve poor adherence with once daily glaucoma medications in electronically monitored patients. Ophthalmology. 2009;116(12):2286–93.CrossRefPubMedPubMedCentral Okeke CO, Quigley HA, Jampel HD, et al. Interventions improve poor adherence with once daily glaucoma medications in electronically monitored patients. Ophthalmology. 2009;116(12):2286–93.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Tsai JC, McClure CA, Ramos SE, et al. Compliance barriers in glaucoma: a systematic classification. J Glaucoma. 2003;12(5):393–8.CrossRefPubMed Tsai JC, McClure CA, Ramos SE, et al. Compliance barriers in glaucoma: a systematic classification. J Glaucoma. 2003;12(5):393–8.CrossRefPubMed
7.
Zurück zum Zitat Newman-Casey PA, Robin AL, Blachley T, et al. The most common barriers to glaucoma medication adherence: a cross-sectional survey. Ophthalmology. 2015;122(7):1308–16.CrossRefPubMedPubMedCentral Newman-Casey PA, Robin AL, Blachley T, et al. The most common barriers to glaucoma medication adherence: a cross-sectional survey. Ophthalmology. 2015;122(7):1308–16.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Winfield AJ, Jessiman D, Williams A, et al. A study of the causes of non-compliance by patients prescribed eyedrops. Br J Ophthalmol. 1990;74(8):477–80.CrossRefPubMedPubMedCentral Winfield AJ, Jessiman D, Williams A, et al. A study of the causes of non-compliance by patients prescribed eyedrops. Br J Ophthalmol. 1990;74(8):477–80.CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Stone JL, Robin AL, Novack GD, Covert DW, Cagle GD. An objective evaluation of eyedrop instillation in patients with glaucoma. Arch Ophthalmol. 2009;127(6):732–6.CrossRefPubMed Stone JL, Robin AL, Novack GD, Covert DW, Cagle GD. An objective evaluation of eyedrop instillation in patients with glaucoma. Arch Ophthalmol. 2009;127(6):732–6.CrossRefPubMed
10.
Zurück zum Zitat Hennessy AL, Katz J, Covert D, Protzko C, Robin AL. Videotaped evaluation of eyedrop instillation in glaucoma patients with visual impairment or moderate to severe visual field loss. Ophthalmology. 2010;117(12):2345–52.CrossRefPubMed Hennessy AL, Katz J, Covert D, Protzko C, Robin AL. Videotaped evaluation of eyedrop instillation in glaucoma patients with visual impairment or moderate to severe visual field loss. Ophthalmology. 2010;117(12):2345–52.CrossRefPubMed
11.
Zurück zum Zitat Van Santvliet L, Ludwig A. Determinants of eye drop size. Surv Ophthalmol. 2004;49(2):197–213.CrossRefPubMed Van Santvliet L, Ludwig A. Determinants of eye drop size. Surv Ophthalmol. 2004;49(2):197–213.CrossRefPubMed
12.
Zurück zum Zitat Mishima S, Gasset A, Kylce S, Baum J. Determination of tear volume and tear flow. Inv Ophthalmol. 1966;5:264–76. Mishima S, Gasset A, Kylce S, Baum J. Determination of tear volume and tear flow. Inv Ophthalmol. 1966;5:264–76.
13.
Zurück zum Zitat Moore DB, Walton C, Moeller KL, et al. Prevalence of early glaucoma eye drop bottle exhaustion and associated risk factors: a patient survey. BMC Ophthalmol. 2014;14:79.CrossRefPubMedPubMedCentral Moore DB, Walton C, Moeller KL, et al. Prevalence of early glaucoma eye drop bottle exhaustion and associated risk factors: a patient survey. BMC Ophthalmol. 2014;14:79.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Guidance for industry: container closure systems for packaging human drugs and biologics. US Department of Health and Human Services, Food and Drug Administration May 1999. Guidance for industry: container closure systems for packaging human drugs and biologics. US Department of Health and Human Services, Food and Drug Administration May 1999.
15.
Zurück zum Zitat Ophthalmic preparations and dispensers, 21 C.F.R § 200.50 (2015). Ophthalmic preparations and dispensers, 21 C.F.R § 200.50 (2015).
16.
Zurück zum Zitat Meyer MA, Savitt ML. A comparison of timolol maleate and levobunolol. Length of use per 5-ml bottle. Ophthalmology. 1994;101(10):1658–61.CrossRefPubMed Meyer MA, Savitt ML. A comparison of timolol maleate and levobunolol. Length of use per 5-ml bottle. Ophthalmology. 1994;101(10):1658–61.CrossRefPubMed
17.
Zurück zum Zitat Gaynes BI, Singa RM, Schaab G, Sorokin Y. Impact of administration angle on the cost of artificial tear solutions: does bottle positioning minimize wastage? J Ocul Pharmacol Ther. 2007;23(2):196–201.CrossRefPubMed Gaynes BI, Singa RM, Schaab G, Sorokin Y. Impact of administration angle on the cost of artificial tear solutions: does bottle positioning minimize wastage? J Ocul Pharmacol Ther. 2007;23(2):196–201.CrossRefPubMed
18.
Zurück zum Zitat Fiscella R, Wilensky JT, Chiang TH, Walt JG. Efficiency of instillation methods for prostaglandin medications. J Ocul Pharmacol Ther. 2006;22(6):477–82.CrossRefPubMed Fiscella R, Wilensky JT, Chiang TH, Walt JG. Efficiency of instillation methods for prostaglandin medications. J Ocul Pharmacol Ther. 2006;22(6):477–82.CrossRefPubMed
19.
Zurück zum Zitat Gaynes BI, Singa RM, Cao Y. Dosage variability of topical ocular hypotensive products: a densitometric assessment. J Glaucoma. 2009;18(2):149–52.CrossRefPubMed Gaynes BI, Singa RM, Cao Y. Dosage variability of topical ocular hypotensive products: a densitometric assessment. J Glaucoma. 2009;18(2):149–52.CrossRefPubMed
20.
Zurück zum Zitat Queen JH, Feldman RM, Lee DA. Variation in number of doses, bottle volume, and calculated yearly cost of generic and branded latanoprost for glaucoma. Am J Ophthalmol. 2016;163:70–4.CrossRefPubMed Queen JH, Feldman RM, Lee DA. Variation in number of doses, bottle volume, and calculated yearly cost of generic and branded latanoprost for glaucoma. Am J Ophthalmol. 2016;163:70–4.CrossRefPubMed
21.
Zurück zum Zitat Drew T, Wolffsohn JS. Usability of prostaglandin monotherapy eye droppers. Br J Ophthalmol 2015;(0):1-4. Drew T, Wolffsohn JS. Usability of prostaglandin monotherapy eye droppers. Br J Ophthalmol 2015;(0):1-4.
22.
Zurück zum Zitat Moore DB, Hammer JD, Akhtari R, et al. Squeeze me if you can: variability in force requirements to extract a drop from common glaucoma bottles. J Glaucoma. 2016;25:780–4.CrossRefPubMed Moore DB, Hammer JD, Akhtari R, et al. Squeeze me if you can: variability in force requirements to extract a drop from common glaucoma bottles. J Glaucoma. 2016;25:780–4.CrossRefPubMed
Metadaten
Titel
An objective assessment of the variability in number of drops per bottle of glaucoma medication
verfasst von
Daniel B. Moore
Judy Beck
Richard J Kryscio
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2017
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0473-8

Weitere Artikel der Ausgabe 1/2017

BMC Ophthalmology 1/2017 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.