Analysis of Screening Electrocardiogram for the Subcutaneous Defibrillator in Adults with Congenital Heart Disease

verfasst von: Vincent C. Thomas, Mark Peterson, Martin McDaniel, Humberto Restrepo, Abraham Rothman, Amit Jain

Erschienen in: Pediatric Cardiology | Ausgabe 6/2017

Einloggen, um Zugang zu erhalten

Abstract

Candidates for the subcutaneous implantable cardioverter-defibrillator (S-ICD) are screened using an electrocardiogram (S-ECG) tool to measure appropriate detection. We sought to define the S-ICD candidacy of congenital heart disease patients using the S-ECG tool. We also analyzed the reliability of the (S-ECG) tool between measurers in this population. Patients above the age of 12 and with a diagnosis associated with either a higher incidence of cardiac arrest or vascular access challenges were asked to undergo screening. S-ECGs were then analyzed by a pediatric electrophysiologist, an S-ICD device engineer, and an S-ICD clinical representative for candidacy. Results were compared for interobserver variability and S-ECGs were analyzed by t test to determine variables that differ among passing and failing leads. Thirty-one patients underwent screening. Two of the 31 (6.5%) patients failed S-ICD screening. Analysis of the screening leads demonstrated the highest passing rates using lead III at a 5 mm/mV amplitude setting with 71 and 62% pass rate in the supine and standing positions, respectively. Interobserver analysis correlated well among the three measurers. There was a higher amplitude difference between QRS and T waves among passing versus failing S-ECG. Congenital heart disease patients have acceptable passage rates utilizing the S-ECG algorithm. Interobserver measurements were well correlated and these data suggest that the proximal coil to device (lead III) vector would be best utilized in this patient population. A larger difference between QRS and T wave amplitudes was associated with a higher S-ECG passing rate.
Literatur
1.
Zurück zum Zitat Khairy P (2016) Ventricular arrhythmias and sudden cardiac death in adults with congenital heart disease. Heart 102:1703–1709CrossRefPubMed Khairy P (2016) Ventricular arrhythmias and sudden cardiac death in adults with congenital heart disease. Heart 102:1703–1709CrossRefPubMed
2.
Zurück zum Zitat Silka MJ, Hardy BG, Menashe VD, Morris CD (1998) A population-based prospective evaluation of risk of sudden cardiac death after operation for common congenital heart defects. J Am Coll Cardiol 32:245–251CrossRefPubMed Silka MJ, Hardy BG, Menashe VD, Morris CD (1998) A population-based prospective evaluation of risk of sudden cardiac death after operation for common congenital heart defects. J Am Coll Cardiol 32:245–251CrossRefPubMed
3.
Zurück zum Zitat Khairy P, Fernandes SM, Mayer JE Jr, Triedman JK, Walsh EP, Lock JE, Landzberg MJ (2008) Long-term survival, modes of death, and predictors of mortality in patients with Fontan surgery. Circulation 117:85–92CrossRefPubMed Khairy P, Fernandes SM, Mayer JE Jr, Triedman JK, Walsh EP, Lock JE, Landzberg MJ (2008) Long-term survival, modes of death, and predictors of mortality in patients with Fontan surgery. Circulation 117:85–92CrossRefPubMed
4.
Zurück zum Zitat Khairy P, Harris L, Landzberg MJ, Fernandes SM, Barlow A, Mercier LA, Viswanathan S, Chetaille P, Gordon E, Dore A, Cecchin F (2008) Sudden death and defibrillators in transposition of the great arteries with intra-atrial baffles: a multicenter study. Circ Arrhythm Electrophysiol 1:250–257CrossRefPubMed Khairy P, Harris L, Landzberg MJ, Fernandes SM, Barlow A, Mercier LA, Viswanathan S, Chetaille P, Gordon E, Dore A, Cecchin F (2008) Sudden death and defibrillators in transposition of the great arteries with intra-atrial baffles: a multicenter study. Circ Arrhythm Electrophysiol 1:250–257CrossRefPubMed
5.
Zurück zum Zitat Groh CA, Sharma S, Pelchovitz DJ, Bhave PD, Rhyner J, Verma N, Arora R, Chicos AB, Kim SS, Lin AC, Passman RS, Knight BP (2014) Use of an electrocardiographic screening tool to determine candidacy for a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 11:1361–1366CrossRefPubMed Groh CA, Sharma S, Pelchovitz DJ, Bhave PD, Rhyner J, Verma N, Arora R, Chicos AB, Kim SS, Lin AC, Passman RS, Knight BP (2014) Use of an electrocardiographic screening tool to determine candidacy for a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 11:1361–1366CrossRefPubMed
6.
Zurück zum Zitat Aziz S, Leon AR, El-Chami MF (2014) The subcutaneous defibrillator: a review of the literature. J Am Coll Cardiol 63:1473–1479CrossRefPubMed Aziz S, Leon AR, El-Chami MF (2014) The subcutaneous defibrillator: a review of the literature. J Am Coll Cardiol 63:1473–1479CrossRefPubMed
7.
Zurück zum Zitat Burke MC, Gold MR, Knight BP, Barr CS, Theuns DA, Boersma LV, Knops RE, Weiss R, Leon AR, Herre JM, Husby M, Stein KM, Lambiase PD (2015) Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol 65:1605–1615CrossRefPubMed Burke MC, Gold MR, Knight BP, Barr CS, Theuns DA, Boersma LV, Knops RE, Weiss R, Leon AR, Herre JM, Husby M, Stein KM, Lambiase PD (2015) Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol 65:1605–1615CrossRefPubMed
8.
Zurück zum Zitat De Maria E, Olaru A, Cappelli S (2015) The entirely subcutaneous defibrillator (S-ICD): state of the art and selection of the ideal candidate. Curr Cardiol Rev 11:180–186CrossRefPubMedPubMedCentral De Maria E, Olaru A, Cappelli S (2015) The entirely subcutaneous defibrillator (S-ICD): state of the art and selection of the ideal candidate. Curr Cardiol Rev 11:180–186CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Bordachar P, Marquie C, Pospiech T, Pasquie JL, Jalal Z, Haissaguerre M, Thambo JB (2016) Subcutaneous implantable cardioverter defibrillators in children, young adults and patients with congenital heart disease. Int J Cardiol 203:251–258CrossRefPubMed Bordachar P, Marquie C, Pospiech T, Pasquie JL, Jalal Z, Haissaguerre M, Thambo JB (2016) Subcutaneous implantable cardioverter defibrillators in children, young adults and patients with congenital heart disease. Int J Cardiol 203:251–258CrossRefPubMed
10.
Zurück zum Zitat Bongiorni MG, Viani S, Zucchelli G, Di Cori A, Segreti L, Paperini L, Levorato D, Boem A, Branchitta G, Andreini D, Soldati E (2015) Subcutaneous implantable cardiac defibrillators: indications and limitations. Curr Heart Fail Rep 12:79–86CrossRefPubMed Bongiorni MG, Viani S, Zucchelli G, Di Cori A, Segreti L, Paperini L, Levorato D, Boem A, Branchitta G, Andreini D, Soldati E (2015) Subcutaneous implantable cardiac defibrillators: indications and limitations. Curr Heart Fail Rep 12:79–86CrossRefPubMed
11.
Zurück zum Zitat Pedersen SS, Mastenbroek MH, Carter N, Barr C, Neuzil P, Scholten M, Lambiase PD, Boersma L, Johansen JB, Theuns DA (2016) A comparison of the quality of life of patients with an entirely subcutaneous implantable defibrillator system versus a transvenous system (from the EFFORTLESS S-ICD quality of life substudy). Am J Cardiol 118:520–526CrossRefPubMed Pedersen SS, Mastenbroek MH, Carter N, Barr C, Neuzil P, Scholten M, Lambiase PD, Boersma L, Johansen JB, Theuns DA (2016) A comparison of the quality of life of patients with an entirely subcutaneous implantable defibrillator system versus a transvenous system (from the EFFORTLESS S-ICD quality of life substudy). Am J Cardiol 118:520–526CrossRefPubMed
12.
Zurück zum Zitat Tjong FVY, Brouwer TF, Smeding L, Kooiman KM, de Groot JR, Ligon D, Sanghera R, Schalij MJ, Wilde AAM, Knops RE (2016) Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance. Europace 18:1740–1747CrossRefPubMed Tjong FVY, Brouwer TF, Smeding L, Kooiman KM, de Groot JR, Ligon D, Sanghera R, Schalij MJ, Wilde AAM, Knops RE (2016) Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance. Europace 18:1740–1747CrossRefPubMed
13.
Zurück zum Zitat Kuschyk J, Stach K, Tulumen E, Rudic B, Liebe V, Schimpf R, Borggrefe M, Roger S (2015) Subcutaneous implantable cardioverter-defibrillator: first single-center experience with other cardiac implantable electronic devices. Heart Rhythm 12:2230–2238CrossRefPubMed Kuschyk J, Stach K, Tulumen E, Rudic B, Liebe V, Schimpf R, Borggrefe M, Roger S (2015) Subcutaneous implantable cardioverter-defibrillator: first single-center experience with other cardiac implantable electronic devices. Heart Rhythm 12:2230–2238CrossRefPubMed
14.
Zurück zum Zitat Pettit SJ, McLean A, Colquhoun I, Connelly D, McLeod K (2013) Clinical experience of subcutaneous and transvenous implantable cardioverter defibrillators in children and teenagers. Pacing Clin Electrophysiol 36:1532–1538CrossRefPubMed Pettit SJ, McLean A, Colquhoun I, Connelly D, McLeod K (2013) Clinical experience of subcutaneous and transvenous implantable cardioverter defibrillators in children and teenagers. Pacing Clin Electrophysiol 36:1532–1538CrossRefPubMed
15.
Zurück zum Zitat Jarman JW, Lascelles K, Wong T, Markides V, Clague JR, Till J (2012) Clinical experience of entirely subcutaneous implantable cardioverter-defibrillators in children and adults: cause for caution. Eur Heart J 33:1351–1359CrossRefPubMed Jarman JW, Lascelles K, Wong T, Markides V, Clague JR, Till J (2012) Clinical experience of entirely subcutaneous implantable cardioverter-defibrillators in children and adults: cause for caution. Eur Heart J 33:1351–1359CrossRefPubMed
16.
Zurück zum Zitat Bettin M, Lascelles F, Rath B, Kobe J, Eckardt L (2015) Recent advances in the entirely subcutaneous ICD system. F1000Prime Rep 7:46 eCollection 2015 CrossRefPubMedPubMedCentral Bettin M, Lascelles F, Rath B, Kobe J, Eckardt L (2015) Recent advances in the entirely subcutaneous ICD system. F1000Prime Rep 7:46 eCollection 2015 CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Zeb M, Curzen N, Veldtman G, Yue A, Roberts P, Wilson D, Morgan J (2015) Potential eligibility of congenital heart disease patients for subcutaneous implantable cardioverter-defibrillator based on surface electrocardiogram mapping. Europace 17:1059–1067PubMed Zeb M, Curzen N, Veldtman G, Yue A, Roberts P, Wilson D, Morgan J (2015) Potential eligibility of congenital heart disease patients for subcutaneous implantable cardioverter-defibrillator based on surface electrocardiogram mapping. Europace 17:1059–1067PubMed
18.
Zurück zum Zitat Zeb M, Curzen N, Allavatam V, Wilson D, Yue A, Roberts P, Morgan J (2015) Sensitivity and specificity of the subcutaneous implantable cardioverter defibrillator pre-implant screening tool. Int J Cardiol 195:205–209CrossRefPubMed Zeb M, Curzen N, Allavatam V, Wilson D, Yue A, Roberts P, Morgan J (2015) Sensitivity and specificity of the subcutaneous implantable cardioverter defibrillator pre-implant screening tool. Int J Cardiol 195:205–209CrossRefPubMed
19.
Zurück zum Zitat Francia P, Adduci C, Palano F, Semprini L, Serdoz A, Montesanti D, Santini D, Musumeci B, Salvati A, Volpe M, Autore C (2015) Eligibility for the subcutaneous implantable cardioverter-defibrillator in patients with hypertrophic cardiomyopathy. J Cardiovasc Electrophysiol 26:893–899CrossRefPubMed Francia P, Adduci C, Palano F, Semprini L, Serdoz A, Montesanti D, Santini D, Musumeci B, Salvati A, Volpe M, Autore C (2015) Eligibility for the subcutaneous implantable cardioverter-defibrillator in patients with hypertrophic cardiomyopathy. J Cardiovasc Electrophysiol 26:893–899CrossRefPubMed
20.
Zurück zum Zitat Chan NY, Yuen HC, Mok NS (2015) Right parasternal electrode configuration converts a failed electrocardiographic screening to a pass for subcutaneous implantable cardioverter-defibrillator implantation. Heart Lung Circ 24:e203–e205CrossRefPubMed Chan NY, Yuen HC, Mok NS (2015) Right parasternal electrode configuration converts a failed electrocardiographic screening to a pass for subcutaneous implantable cardioverter-defibrillator implantation. Heart Lung Circ 24:e203–e205CrossRefPubMed
21.
Zurück zum Zitat Wilson DG, Zeb M, Veldtman G, Dimitrov BD, Morgan JM (2016) Left and right parasternal sensing for the S-ICD in adult congenital heart disease patients and normal controls. Pacing Clin Electrophysiol 39:282–290CrossRefPubMed Wilson DG, Zeb M, Veldtman G, Dimitrov BD, Morgan JM (2016) Left and right parasternal sensing for the S-ICD in adult congenital heart disease patients and normal controls. Pacing Clin Electrophysiol 39:282–290CrossRefPubMed
22.
Zurück zum Zitat Randles DA, Hawkins NM, Shaw M, Patwala AY, Pettit SJ, Wright DJ (2014) How many patients fulfil the surface electrocardiogram criteria for subcutaneous implantable cardioverter-defibrillator implantation? Europace 16:1015–1021CrossRefPubMed Randles DA, Hawkins NM, Shaw M, Patwala AY, Pettit SJ, Wright DJ (2014) How many patients fulfil the surface electrocardiogram criteria for subcutaneous implantable cardioverter-defibrillator implantation? Europace 16:1015–1021CrossRefPubMed
23.
Zurück zum Zitat Frommeyer G, Dechering DG, Zumhagen S, Loher A, Kobe J, Eckardt L, Reinke F (2016) Long-term follow-up of subcutaneous ICD systems in patients with hypertrophic cardiomyopathy: a single-center experience. Clin Res Cardiol 105:89–93CrossRefPubMed Frommeyer G, Dechering DG, Zumhagen S, Loher A, Kobe J, Eckardt L, Reinke F (2016) Long-term follow-up of subcutaneous ICD systems in patients with hypertrophic cardiomyopathy: a single-center experience. Clin Res Cardiol 105:89–93CrossRefPubMed
24.
Zurück zum Zitat Maurizi N, Olivotto I, Olde Nordkamp LR, Baldini K, Fumagalli C, Brouwer TF, Knops RE, Cecchi F (2015) Prevalence of subcutaneous implantable cardioverter-defibrillator candidacy based on template ECG screening in patients with hypertrophic cardiomyopathy. Heart Rhythm 13:457–463CrossRefPubMed Maurizi N, Olivotto I, Olde Nordkamp LR, Baldini K, Fumagalli C, Brouwer TF, Knops RE, Cecchi F (2015) Prevalence of subcutaneous implantable cardioverter-defibrillator candidacy based on template ECG screening in patients with hypertrophic cardiomyopathy. Heart Rhythm 13:457–463CrossRefPubMed
25.
Zurück zum Zitat Lambiase PD, Gold MR, Hood M, Boersma L, Theuns DA, Burke MC, Weiss R, Russo AM, Kaab S, Knight BP (2016) Evaluation of subcutaneous ICD early performance in hypertrophic cardiomyopathy from the pooled EFFORTLESS and IDE cohorts. Heart Rhythm 13:1066–1074CrossRefPubMed Lambiase PD, Gold MR, Hood M, Boersma L, Theuns DA, Burke MC, Weiss R, Russo AM, Kaab S, Knight BP (2016) Evaluation of subcutaneous ICD early performance in hypertrophic cardiomyopathy from the pooled EFFORTLESS and IDE cohorts. Heart Rhythm 13:1066–1074CrossRefPubMed
26.
Zurück zum Zitat Olde Nordkamp LR, Brouwer TF, Barr C, Theuns DA, Boersma LV, Johansen JB, Neuzil P, Wilde AA, Carter N, Husby M, Lambiase PD, Knops RE (2015) Inappropriate shocks in the subcutaneous ICD: incidence, predictors and management. Int J Cardiol 195:126–133CrossRefPubMed Olde Nordkamp LR, Brouwer TF, Barr C, Theuns DA, Boersma LV, Johansen JB, Neuzil P, Wilde AA, Carter N, Husby M, Lambiase PD, Knops RE (2015) Inappropriate shocks in the subcutaneous ICD: incidence, predictors and management. Int J Cardiol 195:126–133CrossRefPubMed
27.
Zurück zum Zitat El-Chami MF, Harbieh B, Levy M, Leon AR, Merchant FM (2016) Clinical and electrocardiographic predictors of T wave oversensing in patients with subcutaneous ICD. J Arrhythm 32:181–185CrossRefPubMedPubMedCentral El-Chami MF, Harbieh B, Levy M, Leon AR, Merchant FM (2016) Clinical and electrocardiographic predictors of T wave oversensing in patients with subcutaneous ICD. J Arrhythm 32:181–185CrossRefPubMedPubMedCentral
28.
Zurück zum Zitat Olde Nordkamp LR, Warnaars JL, Kooiman KM, de Groot JR, Rosenmoller BR, Wilde AA, Knops RE (2014) Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening. J Cardiovasc Electrophysiol 25:494–499CrossRefPubMed Olde Nordkamp LR, Warnaars JL, Kooiman KM, de Groot JR, Rosenmoller BR, Wilde AA, Knops RE (2014) Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening. J Cardiovasc Electrophysiol 25:494–499CrossRefPubMed
29.
Zurück zum Zitat Brisben AJ, Burke MC, Knight BP, Hahn SJ, Herrmann KL, Allavatam V, Mahajan D, Sanghera R, Gold MR (2015) A new algorithm to reduce inappropriate therapy in the S-ICD system. J Cardiovasc Electrophysiol 26:417–423CrossRefPubMed Brisben AJ, Burke MC, Knight BP, Hahn SJ, Herrmann KL, Allavatam V, Mahajan D, Sanghera R, Gold MR (2015) A new algorithm to reduce inappropriate therapy in the S-ICD system. J Cardiovasc Electrophysiol 26:417–423CrossRefPubMed
30.
Zurück zum Zitat Ziacchi M, Corzani A, Diemberger I, Martignani C, Marziali A, Mazzotti A, Massaro G, Rapezzi C, Biffi M, Boriani G (2016) Electrocardiographic eligibility for subcutaneous implantable cardioverter defibrillator: evaluation during bicycle exercise. Heart Lung Circ 25:476–483CrossRefPubMed Ziacchi M, Corzani A, Diemberger I, Martignani C, Marziali A, Mazzotti A, Massaro G, Rapezzi C, Biffi M, Boriani G (2016) Electrocardiographic eligibility for subcutaneous implantable cardioverter defibrillator: evaluation during bicycle exercise. Heart Lung Circ 25:476–483CrossRefPubMed
31.
Zurück zum Zitat Kooiman KM, Knops RE, Olde Nordkamp L, Wilde AA, de Groot JR (2014) Inappropriate subcutaneous implantable cardioverter-defibrillator shocks due to T-wave oversensing can be prevented: implications for management. Heart Rhythm 11:426–434CrossRefPubMed Kooiman KM, Knops RE, Olde Nordkamp L, Wilde AA, de Groot JR (2014) Inappropriate subcutaneous implantable cardioverter-defibrillator shocks due to T-wave oversensing can be prevented: implications for management. Heart Rhythm 11:426–434CrossRefPubMed
32.
Zurück zum Zitat Gold MR, Weiss R, Theuns DA, Smith W, Leon A, Knight BP, Carter N, Husby M, Burke MC (2014) Use of a discrimination algorithm to reduce inappropriate shocks with a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 11:1352–1358CrossRefPubMed Gold MR, Weiss R, Theuns DA, Smith W, Leon A, Knight BP, Carter N, Husby M, Burke MC (2014) Use of a discrimination algorithm to reduce inappropriate shocks with a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 11:1352–1358CrossRefPubMed
Metadaten
Titel
Analysis of Screening Electrocardiogram for the Subcutaneous Defibrillator in Adults with Congenital Heart Disease
verfasst von
Vincent C. Thomas
Mark Peterson
Martin McDaniel
Humberto Restrepo
Abraham Rothman
Amit Jain
Publikationsdatum
22.05.2017
Verlag
Springer US
Erschienen in
Pediatric Cardiology / Ausgabe 6/2017
Print ISSN: 0172-0643
Elektronische ISSN: 1432-1971
DOI
https://doi.org/10.1007/s00246-017-1635-4

Weitere Artikel der Ausgabe 6/2017

Pediatric Cardiology 6/2017 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.