Skip to main content
Erschienen in: Clinical Orthopaedics and Related Research® 4/2014

01.04.2014 | Basic Research

Angle Stable Nails Provide Improved Healing for a Complex Fracture Model in the Femur

verfasst von: Meghan R. Kubacki, MS, Christopher A. Verioti, DO, Savan D. Patel, MD, Adam N. Garlock, MS, David Fernandez, MD, Patrick J. Atkinson, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 4/2014

Einloggen, um Zugang zu erhalten

Abstract

Background

Conventional nails are being used for an expanding range of fractures from simple to more complex. Angle stable designs are a relatively new innovation; however, it is unknown if they will improve healing for complex fractures.

Questions/purposes

When comparing traditional and angle stable nails to treat complex open canine femur fractures, the current study addressed the following questions: do the two constructs differ in (1) radiographic evidence of bone union across the cortices; (2) stability as determined by toggle (torsional motion with little accompanying torque) and angular deformation; (3) biomechanical properties, including stiffness in bending, axial compression, and torsional loading, and construct failure properties in torsion; and (4) degree of bone tissue mineralization?

Methods

Ten hounds with a 1-cm femoral defect and periosteal stripping were treated with a reamed titanium angle stable or nonangle stable nail after the creation of a long soft tissue wound. Before the study, the animals were randomly assigned to receive one of the nails and to be evaluated with biomechanical testing or histology. After euthanasia at 16 weeks, all operative femora were assessed radiographically. Histological or biomechanical evaluation was conducted of the operative bones with nails left in situ compared with the nonoperative contralateral femora.

Results

Radiographic and gross inspection demonstrated hypertrophic nonunion in all 10 animals treated with the nonangle stable nail, whereas six of 10 animals treated with the angle stable nail bridged at least one cortex (p = 0.023). The angle stable nail construct demonstrated no toggle in nine of 10 animals, whereas all control femora exhibited toggle. The angle stable nail demonstrated less angular deformation and toggle (p ≤ 0.005) and increased compressive stiffness (p = 0.001) compared with the conventional nonangle stable nail. Histology demonstrated more nonmineralized tissue in the limbs treated with the conventional nail (p = 0.005).

Conclusions

Angle stable nails that eliminate toggle lead to enhanced yet incomplete fracture healing in a complex canine fracture model.

Clinical Relevance

Care should be taken in tailoring the nail design features to the characteristics of the fracture and the patient.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Bottlang M, Lesser M, Koerber J, Doornink J, von Rechenberg B, Augat P, Fitzpatrick DC, Madey SM, Marsh JL. Far cortical locking can improve healing of fractures stabilized with locking plates. J Bone Joint Surg Am. 2010;92:1652–1660.PubMedCentralPubMedCrossRef Bottlang M, Lesser M, Koerber J, Doornink J, von Rechenberg B, Augat P, Fitzpatrick DC, Madey SM, Marsh JL. Far cortical locking can improve healing of fractures stabilized with locking plates. J Bone Joint Surg Am. 2010;92:1652–1660.PubMedCentralPubMedCrossRef
2.
Zurück zum Zitat Bullens PH, Hannink G, Verdonschot N, Buma P. No effect of dynamic loading on bone graft healing in femoral segmental defect reconstructions in the goat. Injury. 2010;41:1284–1291.PubMedCrossRef Bullens PH, Hannink G, Verdonschot N, Buma P. No effect of dynamic loading on bone graft healing in femoral segmental defect reconstructions in the goat. Injury. 2010;41:1284–1291.PubMedCrossRef
3.
Zurück zum Zitat Bullens PH, Schreuder HW, de Waal Malefijt MC, Verdonschot N, Buma P. The presence of periosteum is essential for the healing of large diaphyseal segmental bone defects reconstructed with trabecular metal: a study in the femur of goats. J Biomed Mater Res B Appl Biomater. 2010;92:24–31.PubMedCrossRef Bullens PH, Schreuder HW, de Waal Malefijt MC, Verdonschot N, Buma P. The presence of periosteum is essential for the healing of large diaphyseal segmental bone defects reconstructed with trabecular metal: a study in the femur of goats. J Biomed Mater Res B Appl Biomater. 2010;92:24–31.PubMedCrossRef
4.
Zurück zum Zitat Carter DR, Vasu R, Harris WH. The plated femur-relationships between the changes in bone stresses and bone loss. Acta Orthop Scand. 1981;52:241–248.PubMedCrossRef Carter DR, Vasu R, Harris WH. The plated femur-relationships between the changes in bone stresses and bone loss. Acta Orthop Scand. 1981;52:241–248.PubMedCrossRef
5.
Zurück zum Zitat Carter DR, Vasu R, Spengler DM, Dueland RT. Stress fields in the unplated and plated canine femur calculated from in vivo strain measurements. J Biomech. 1981;14:63–70.PubMedCrossRef Carter DR, Vasu R, Spengler DM, Dueland RT. Stress fields in the unplated and plated canine femur calculated from in vivo strain measurements. J Biomech. 1981;14:63–70.PubMedCrossRef
6.
Zurück zum Zitat Cheung G, Zalzal P, Bhandari M, Spelt JK, Papini M. Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading. Med Eng Phys. 2004;26:93–108.PubMedCrossRef Cheung G, Zalzal P, Bhandari M, Spelt JK, Papini M. Finite element analysis of a femoral retrograde intramedullary nail subject to gait loading. Med Eng Phys. 2004;26:93–108.PubMedCrossRef
7.
Zurück zum Zitat Claes L, Augat P, Suger G, Wilke HJ. Influence of size and stability of the osteotomy gap on the success of fracture healing. J Orthop Res. 1997;15:577–584.PubMedCrossRef Claes L, Augat P, Suger G, Wilke HJ. Influence of size and stability of the osteotomy gap on the success of fracture healing. J Orthop Res. 1997;15:577–584.PubMedCrossRef
8.
Zurück zum Zitat Déjardin LM, Guillou RP, Ting D, Sinnott MT, Meyer E, Haut RC. Effect of bending direction on the mechanical behavior of interlocking nail systems. Vet Comp Orthop Traumatol. 2009;22:264–269.PubMed Déjardin LM, Guillou RP, Ting D, Sinnott MT, Meyer E, Haut RC. Effect of bending direction on the mechanical behavior of interlocking nail systems. Vet Comp Orthop Traumatol. 2009;22:264–269.PubMed
9.
Zurück zum Zitat Déjardin LM, Guiot LP, von Pfeil DJ. Interlocking nails and minimally invasive osteosynthesis. Vet Clin North Am Small Anim Pract. 2012;42:935–936.PubMedCrossRef Déjardin LM, Guiot LP, von Pfeil DJ. Interlocking nails and minimally invasive osteosynthesis. Vet Clin North Am Small Anim Pract. 2012;42:935–936.PubMedCrossRef
10.
Zurück zum Zitat Déjardin LM, Lansdowne JL, Sinnott MT, Sidebotham CG, Haut RC. In vitro mechanical evaluation of torsional loading in simulated canine tibiae for a novel hourglass-shaped interlocking nail with a self-tapping tapered locking design. Am J Vet Res. 2006;67:678–685.PubMedCrossRef Déjardin LM, Lansdowne JL, Sinnott MT, Sidebotham CG, Haut RC. In vitro mechanical evaluation of torsional loading in simulated canine tibiae for a novel hourglass-shaped interlocking nail with a self-tapping tapered locking design. Am J Vet Res. 2006;67:678–685.PubMedCrossRef
11.
Zurück zum Zitat den Boer FC, Patka P, Bakker FC, Wippermann BW, van Lingen A, Vink GQ, Boshuizen K, Haarman HJ. New segmental long bone defect model in sheep: quantitative analysis of healing with dual energy x-ray absorptiometry. J Orthop Res. 1999;17:654–660.CrossRef den Boer FC, Patka P, Bakker FC, Wippermann BW, van Lingen A, Vink GQ, Boshuizen K, Haarman HJ. New segmental long bone defect model in sheep: quantitative analysis of healing with dual energy x-ray absorptiometry. J Orthop Res. 1999;17:654–660.CrossRef
12.
Zurück zum Zitat Donovan JP, Garlock A, Lecronier D, Houghtaling J, Burton S, Atkinson PJ. Assessment of the stability of an alternative interlocking nail for segmental lone bone defects. Proc Ann Mich Orthop Soc. 2012;226:469–476. Donovan JP, Garlock A, Lecronier D, Houghtaling J, Burton S, Atkinson PJ. Assessment of the stability of an alternative interlocking nail for segmental lone bone defects. Proc Ann Mich Orthop Soc. 2012;226:469–476.
13.
Zurück zum Zitat Epari DR, Kassi JP, Schell H, Duda GN. Timely fracture-healing requires optimization of axial fixation stability. J Bone Joint Surg Am. 2007;89:1575–1585.PubMedCrossRef Epari DR, Kassi JP, Schell H, Duda GN. Timely fracture-healing requires optimization of axial fixation stability. J Bone Joint Surg Am. 2007;89:1575–1585.PubMedCrossRef
14.
Zurück zum Zitat Garlock AN, Donovan J, LeCronier DJ, Houghtaling J, Burton S, Atkinson PJ. A modified intramedullary nail interlocking design yields improved stability for fatigue cycling in a canine femur fracture model. Proc Inst Mech Eng H. 2012;226:469–476.PubMedCrossRef Garlock AN, Donovan J, LeCronier DJ, Houghtaling J, Burton S, Atkinson PJ. A modified intramedullary nail interlocking design yields improved stability for fatigue cycling in a canine femur fracture model. Proc Inst Mech Eng H. 2012;226:469–476.PubMedCrossRef
15.
Zurück zum Zitat Geiger F, Bertram H, Berger I, Lorenz H, Wall O, Eckhardt C, Simank HG, Richter W. Vascular endothelial growth factor gene-activated matrix (VEGF165-GAM) enhances osteogenesis and angiogenesis in large segmental bone defects. J Bone Miner Res. 2005;20:2028–2035.PubMedCrossRef Geiger F, Bertram H, Berger I, Lorenz H, Wall O, Eckhardt C, Simank HG, Richter W. Vascular endothelial growth factor gene-activated matrix (VEGF165-GAM) enhances osteogenesis and angiogenesis in large segmental bone defects. J Bone Miner Res. 2005;20:2028–2035.PubMedCrossRef
16.
Zurück zum Zitat Goel VK, Drinker H, Panjabi MM, Strongwater A. Selection of an animal model for implant fixation studies: anatomical aspects. Yale J Biol Med. 1982;55:113–122.PubMedCentralPubMed Goel VK, Drinker H, Panjabi MM, Strongwater A. Selection of an animal model for implant fixation studies: anatomical aspects. Yale J Biol Med. 1982;55:113–122.PubMedCentralPubMed
17.
Zurück zum Zitat Gueorguiev B, Wähnert D, Albrecht D, Ockert B, Windolf M, Schwieger K. Effect on dynamic mechanical stability and interfragmentary movement of angle-stable locking of intramedullary nails in unstable distal tibia fractures: a biomechanical study. J Trauma. 2011;70:358–365.PubMedCrossRef Gueorguiev B, Wähnert D, Albrecht D, Ockert B, Windolf M, Schwieger K. Effect on dynamic mechanical stability and interfragmentary movement of angle-stable locking of intramedullary nails in unstable distal tibia fractures: a biomechanical study. J Trauma. 2011;70:358–365.PubMedCrossRef
18.
Zurück zum Zitat Gustilo RB, Anderson JT. Prevention of infection in the treatment of one thousand and twenty-five open fractures of long bones. Retrospective and prospective analyses. J Bone Joint Surg Am. 2002;84:682.PubMed Gustilo RB, Anderson JT. Prevention of infection in the treatment of one thousand and twenty-five open fractures of long bones. Retrospective and prospective analyses. J Bone Joint Surg Am. 2002;84:682.PubMed
19.
Zurück zum Zitat Kaspar K, Schell H, Seebeck P, Thompson MS, Schütz M, Haas NP, Duda GN. Angle stable locking reduces interfragmentary movements and promotes healing after unreamed nailing. Study of displaced osteotomy model in sheep tibia. J Bone Joint Surg Am. 2005;87:2028–2037.PubMedCrossRef Kaspar K, Schell H, Seebeck P, Thompson MS, Schütz M, Haas NP, Duda GN. Angle stable locking reduces interfragmentary movements and promotes healing after unreamed nailing. Study of displaced osteotomy model in sheep tibia. J Bone Joint Surg Am. 2005;87:2028–2037.PubMedCrossRef
20.
Zurück zum Zitat Kasten P, Vogel J, Geiger F, Niemeyer P, Luginbühl R, Szalay K. The effect of platelet-rich plasma on healing in critical-size long-bone defect. Biomaterials. 2008;29:3983–3992.PubMedCrossRef Kasten P, Vogel J, Geiger F, Niemeyer P, Luginbühl R, Szalay K. The effect of platelet-rich plasma on healing in critical-size long-bone defect. Biomaterials. 2008;29:3983–3992.PubMedCrossRef
21.
Zurück zum Zitat Keating JF, Orfaly R, O’Brien PJ. Knee pain after tibial nailing. J Orthop Trauma. 1997;11:10–13.PubMedCrossRef Keating JF, Orfaly R, O’Brien PJ. Knee pain after tibial nailing. J Orthop Trauma. 1997;11:10–13.PubMedCrossRef
22.
Zurück zum Zitat Klein P, Opitz M, Schell H, Taylor WR, Heller MO, Kassi JP, Kandziora F, Duda GN. Comparison of unreamed nailing and external fixation of tibial diastases—mechanical conditions during healing and biological outcome. J Orthop Res. 2004;22:1072–1078.PubMedCrossRef Klein P, Opitz M, Schell H, Taylor WR, Heller MO, Kassi JP, Kandziora F, Duda GN. Comparison of unreamed nailing and external fixation of tibial diastases—mechanical conditions during healing and biological outcome. J Orthop Res. 2004;22:1072–1078.PubMedCrossRef
23.
Zurück zum Zitat Kraus KH, Kadiyala S, Wotton H, Kurth A, Shea M, Hannan M, Hayes WC, Kirker-Head CA, Bruder S. Critically sized osteo-periosteal femoral defects: a dog model. J Invest Surg. 1999;12:115–124.PubMedCrossRef Kraus KH, Kadiyala S, Wotton H, Kurth A, Shea M, Hannan M, Hayes WC, Kirker-Head CA, Bruder S. Critically sized osteo-periosteal femoral defects: a dog model. J Invest Surg. 1999;12:115–124.PubMedCrossRef
24.
Zurück zum Zitat Lansdowne JL, Sinnott MT, Déjardin LM, Ting D, Haut RC. In vitro mechanical comparison of screwed, bolted, and novel interlocking nail systems to buttress plate fixation in torsion and mediolateral bending. Vet Surg. 2007;36:368–377.PubMedCrossRef Lansdowne JL, Sinnott MT, Déjardin LM, Ting D, Haut RC. In vitro mechanical comparison of screwed, bolted, and novel interlocking nail systems to buttress plate fixation in torsion and mediolateral bending. Vet Surg. 2007;36:368–377.PubMedCrossRef
25.
Zurück zum Zitat Lindsey RW, Gugala Z, Milne E, Sun M, Gannon FH, Latta LL. The efficacy of cylindrical titanium mesh gage for the reconstruction of a critical-size canine segmental femoral diaphyseal defects. J Orthop Res. 2006;24:1438–1453.PubMedCrossRef Lindsey RW, Gugala Z, Milne E, Sun M, Gannon FH, Latta LL. The efficacy of cylindrical titanium mesh gage for the reconstruction of a critical-size canine segmental femoral diaphyseal defects. J Orthop Res. 2006;24:1438–1453.PubMedCrossRef
26.
Zurück zum Zitat McDuffee LA, Stover SM, Taylor KT. In vitro cyclic biomechanical properties of an interlocking equine tibial nail. Vet Surg. 2000;29:163–172.PubMedCrossRef McDuffee LA, Stover SM, Taylor KT. In vitro cyclic biomechanical properties of an interlocking equine tibial nail. Vet Surg. 2000;29:163–172.PubMedCrossRef
27.
Zurück zum Zitat Mückley T, Hoffmeier K, Klos K, Petrovitch A, von Oldenburg G, Hofmann GO. Angle-stable and compressed angle-stable locking for tibiotalocalcaneal arthrodesis with retrograde intramedullary nails. J Bone Joint Surg Am. 2008;90:620–627.PubMedCrossRef Mückley T, Hoffmeier K, Klos K, Petrovitch A, von Oldenburg G, Hofmann GO. Angle-stable and compressed angle-stable locking for tibiotalocalcaneal arthrodesis with retrograde intramedullary nails. J Bone Joint Surg Am. 2008;90:620–627.PubMedCrossRef
28.
29.
Zurück zum Zitat Pazzaglia UE, Bernini F, Zatti G, Di Nucci A. Histology of the metal-bone interface: interpretation of plastic embedded slides. Biomaterials. 1994;15:273–277.PubMedCrossRef Pazzaglia UE, Bernini F, Zatti G, Di Nucci A. Histology of the metal-bone interface: interpretation of plastic embedded slides. Biomaterials. 1994;15:273–277.PubMedCrossRef
30.
Zurück zum Zitat Perren SM. Physical and biological aspects of fracture healing with special reference to internal fixation. Clin Orthop Relat Res. 1979;138:175–196.PubMed Perren SM. Physical and biological aspects of fracture healing with special reference to internal fixation. Clin Orthop Relat Res. 1979;138:175–196.PubMed
31.
Zurück zum Zitat Reems MR, Pluhar GE, Wheeler DL. Ex vivo comparison of one versus two distal screws in 8 mm model 11 interlocking nails used to stabilize canine distal femoral fractures. Vet Surg. 2006;35:161–167.PubMedCrossRef Reems MR, Pluhar GE, Wheeler DL. Ex vivo comparison of one versus two distal screws in 8 mm model 11 interlocking nails used to stabilize canine distal femoral fractures. Vet Surg. 2006;35:161–167.PubMedCrossRef
32.
Zurück zum Zitat Reichert JC, Saifzadeh S, Wullschleger ME, Epari DR, Schütz MA, Duda GN, Schell H, van Griensven M, Redl H, Hutmacher DW. The challenge of establishing preclinical models for segmental bone defect research. Biomaterials. 2009;30:2149–2163.PubMedCrossRef Reichert JC, Saifzadeh S, Wullschleger ME, Epari DR, Schütz MA, Duda GN, Schell H, van Griensven M, Redl H, Hutmacher DW. The challenge of establishing preclinical models for segmental bone defect research. Biomaterials. 2009;30:2149–2163.PubMedCrossRef
34.
Zurück zum Zitat Takushima A, Kitano Y, Harii K. Osteogenic potential of cultured periosteal cells in a distracted bone gap in rabbits. J Surg Res. 1998;78:68–77.PubMedCrossRef Takushima A, Kitano Y, Harii K. Osteogenic potential of cultured periosteal cells in a distracted bone gap in rabbits. J Surg Res. 1998;78:68–77.PubMedCrossRef
35.
Zurück zum Zitat von Pfeil DJ, Déjardin LM, DeCamp CE, Meyer EG, Lansdowne JL, Weerts RJ, Haut RC. In vitro biomechanical comparison of a plate-rod combination-construct and an interlocking nail-construct for experimentally induced gap fractures in canine tibiae. Am J Vet Res. 2005;66:1536–1543.CrossRef von Pfeil DJ, Déjardin LM, DeCamp CE, Meyer EG, Lansdowne JL, Weerts RJ, Haut RC. In vitro biomechanical comparison of a plate-rod combination-construct and an interlocking nail-construct for experimentally induced gap fractures in canine tibiae. Am J Vet Res. 2005;66:1536–1543.CrossRef
36.
Zurück zum Zitat Weber BG, Čech O. Pseudarthrosis: Pathophysiology, Biomechanics, Therapy, Results. New York, NY, USA: Grune and Stratton; 1976. Weber BG, Čech O. Pseudarthrosis: Pathophysiology, Biomechanics, Therapy, Results. New York, NY, USA: Grune and Stratton; 1976.
37.
Zurück zum Zitat Winquist RA, Hansen ST Jr, Clawson DK. Closed intramedullary nailing of femoral fractures. A report of five hundred and twenty cases. J Bone Joint Surg Am. 1984;66:529–539.PubMed Winquist RA, Hansen ST Jr, Clawson DK. Closed intramedullary nailing of femoral fractures. A report of five hundred and twenty cases. J Bone Joint Surg Am. 1984;66:529–539.PubMed
38.
Zurück zum Zitat Zhu L, Liu W, Cui L, Cao Y. Tissue-engineered bone repair of goat femur defects with osteogenically induced bone marrow stromal cells. Tissue Eng. 2006;12:423–433.PubMedCrossRef Zhu L, Liu W, Cui L, Cao Y. Tissue-engineered bone repair of goat femur defects with osteogenically induced bone marrow stromal cells. Tissue Eng. 2006;12:423–433.PubMedCrossRef
Metadaten
Titel
Angle Stable Nails Provide Improved Healing for a Complex Fracture Model in the Femur
verfasst von
Meghan R. Kubacki, MS
Christopher A. Verioti, DO
Savan D. Patel, MD
Adam N. Garlock, MS
David Fernandez, MD
Patrick J. Atkinson, PhD
Publikationsdatum
01.04.2014
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 4/2014
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-013-3288-9

Weitere Artikel der Ausgabe 4/2014

Clinical Orthopaedics and Related Research® 4/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.