Skip to main content
Erschienen in: Clinical Orthopaedics and Related Research® 11/2016

Open Access 27.09.2016 | Symposium: Advances in PEEK Technology

Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study

verfasst von: Ryan Siskey, MS, Lauren Ciccarelli, MS, Melissa K. C. Lui, MS, Steven M. Kurtz, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 11/2016

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Most contemporary total disc replacements (TDRs) use conventional orthopaedic bearing couples such as ultrahigh-molecular-weight polyethylene (polyethylene) and cobalt-chromium (CoCr). Cervical total disc replacements incorporating polyetheretherketone (PEEK) bearings (specifically PEEK-on-PEEK bearings) have been previously investigated, but little is known about PEEK-on-ceramic bearings for TDR.

Questions/purposes

(1) What is the tribologic behavior of a PEEK-on-ceramic bearing for cervical TDR under idealized, clean wear test conditions? (2) How does the PEEK-on-ceramic design perform under impingement conditions? (3) How is the PEEK-on-ceramic bearing affected by abrasive wear? (4) Is the particle morphology from PEEK-on-ceramic bearings for TDRs affected by adverse wear scenarios?

Methods

PEEK-on-ceramic cervical TDR bearings were subjected to a 10 million cycle ideal wear test based on ASTM F2423 and ISO 181912-1 using a six-station spine wear simulator (MTS, Eden Prairie, MN, USA) with 5 g/L bovine serum concentration at 23° ± 2° C (ambient temperature). Validated 1 million cycle impingement and 5 million cycle abrasive tests were conducted on the PEEK-on-ceramic bearings based, in part, on retrieval analysis of a comparable bearing design as well as finite element analyses. The ceramic-on-PEEK couple was characterized for damage modes, mass and volume loss, and penetration and the lubricant was subjected to particle analysis. The resulting mass wear rate, volumetric wear rate, based on material density, and particle analysis were compared with clinically available cervical disc bearing couples.

Results

The three modes of wear (idealized, impingement, and abrasive) resulted in mean mass wear rates of 0.9 ± 0.2 mg/MC, 1.9 ± 0.5 mg/MC, and 2.8 ± 0.6 mg/MC, respectively. The mass wear rates were converted to volumetric wear rates using density and found to be 0.7 ± 0.1 mm3/MC, 1.5 ± 0.4 mm3/MC, and 2.1 ± 0.5 mm3/MC, respectively. During each test, the PEEK endplates were the primary sources of wear and demonstrated an abrasive wear mechanism. Under idealized and impingement conditions, the ceramic core also demonstrated slight polishing of the articulating surface but the change in mass was unmeasurable. During abrasive testing, the titanium transfer on the core was shown to polish over 5 MC of testing. In all cases and consistent with previous studies of other PEEK bearing couples, the particle size was primarily < 2 µm and morphology was smooth and spheroidal.

Conclusions

Overall, the idealized PEEK-on-ceramic wear rate (0.7 ± 0.1 mm3/MC) appears comparable to the published wear rates for other polymer-on-hard bearing couples (0.3–6.7 mm3/MC) and within the range of 0.2 to 1.9 mm3/MC reported for PEEK-on-PEEK cervical disc designs. The particles, based on size and morphology, also suggest the wear mechanism is comparable between the PEEK-on-ceramic couple and other polymer-on-ceramic orthopaedic couples.

Clinical Relevance

The PEEK-on-ceramic bearing considered in this study is a novel bearing couple for use in total disc arthroplasty devices and will require clinical evaluation to fully assess the bearing couple and total disc design. However, the wear rates under idealized and adverse conditions, and particle size and morphology, suggest that PEEK-on-ceramic bearings may be a reasonable alternative to polyethylene-on-CoCr and metal-on-metal bearings currently used in cervical TDRs.
Hinweise
The research presented in this article was funded by Simplify Medical (Sunnyvale, CA, USA; SMK, Principal Investigator). One of the authors (MKCL) is an employee of Simplify Medical. One of the authors (SMK) is an employee and shareholder of Exponent, Inc (Philadelphia, PA, USA), a scientific and engineering consulting firm. Exponent has been paid fees by companies and suppliers for this author’s consulting services on behalf of such companies and suppliers. One of the authors (SMK) has received institutional support from Smith & Nephew (London, UK); Stryker (Kalamazoo, MI, USA); Zimmer Biomet (Warsaw, IN, USA); DePuy Synthes (Warsaw, IN, USA); Medtronic (Dublin, Ireland); Invibio (Lancashire, UK); Stelkast (McMurray, PA, USA); Formae (Paoli, PA, USA); Kyocera Medical (Osaka, Japan); Wright Medical Technology (Memphis, TN, USA); Ceramtec (Laurens, SC, USA); DJO (Vista, CA, USA); Celanese (Irving, TX, USA); Aesculap (Center Valley, PA, USA); SpinalMotion (Mountain View, CA, USA); Active Implants (Memphis, TN, USA); and Ferring Pharmaceuticals (Parsippany, NJ, USA).
All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research ® editors and board members are on file with the publication and can be viewed on request.
Clinical Orthopaedics and Related Research ® neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA-approval status, of any drug or device prior to clinical use.
This work was performed at Exponent, Inc, Philadelphia, PA, USA.
A comment to this article is available at http://​dx.​doi.​org/​10.​1007/​s11999-016-5096-5.

Introduction

Fixed-bearing and mobile-bearing cervical total disc replacements (TDRs) that incorporate traditional orthopaedic bearing materials such as ultrahigh-molecular-weight polyethylene (hereafter, polyethylene) and cobalt-chromium (CoCr) alloys have been used for several decades to treat patient pain associated with degenerative disc disease [34]. Polyethylene, CoCr, and stainless steel bearings for cervical TDR were approved for clinical use based on idealized, clean wear rates obtained from standardized testing [2, 3, 5, 11, 2025, 27, 29, 3436, 43]. Current industry standards for evaluating wear of TDRs are limited to idealized wear conditions, encompassing the intended bearing surfaces under anatomically justified ROMs. However, little is known about the tribologic behavior of cervical TDRs under adverse wear conditions such as impingement or third-body wear [36].
Recently, polyetheretherketone (PEEK) has been considered as an alternative bearing for all-polymer cervical TDRs [7, 911, 27, 31, 32, 51]. PEEK bearings are attractive candidates for TDR because they avoid the use of metallic components and do not interfere with MRI. However, under elevated contact stresses in vitro, a delamination wear mechanism has been observed in PEEK-on-PEEK TDRs [27].
PEEK-on-ceramic bearings represent another alternative for cervical TDRs. Although ceramic biomaterials are well accepted in large joint replacements [13, 15, 19, 38, 40, 45, 46, 49, 50], fracture risk is a concern [1, 44]. It is also unclear if PEEK-on-ceramic bearings will demonstrate acceptably low wear under idealized or aggressive tribologic conditions. Orthopaedic bearings may be exposed to adverse wear conditions that differ from the idealized conditions currently prescribed by industry standards. Cervical TDR polymer-on-ceramic bearings are no exception. THAs are assessed preclinically under conditions meant to simulate impingement based on retrieval studies that have shown ceramic femoral heads with evidence of metal transfer consistent with subluxation and/or impingement [6, 14, 17, 18, 30, 39, 41, 42]. Similarly, for TDRs, many designs have been shown to demonstrate impingement [28, 35, 36]. Therefore, assessing the effect of both impingement and metal transfer on a candidate polymer-on-ceramic bearing couple is essential. Furthermore, the particles generated under these adverse conditions may be different than the idealized conditions and therefore warrant investigation.
Consequently, we addressed the following research questions in a series of experiments: (1) What is the tribologic behavior of a PEEK-on-ceramic bearing for cervical TDR under idealized, clean wear test conditions? (2) How does the PEEK-on-ceramic design perform under impingement conditions? (3) How is the PEEK-on-ceramic bearing affected by abrasive wear? (4) Is the particle morphology from PEEK-on-ceramic bearings for TDRs affected by adverse wear scenarios?

Materials and Methods

Test coupons were created for this study to be representative of an investigational mobile-bearing design (Simplify Medical, Sunnyvale, CA, USA). The geometry of this design is based on the KineFlex®|C Cervical TDR (SpinalMotion, Mountain View, CA, USA). The test coupons in the present study were comprised of two PEEK endplates (PEEK® Optima Natural; Invibio, West Conshohocken, PA, USA) separated by a mobile zirconia-toughened alumina ceramic core (CeraSurf®-p; CoorsTek® Medical, Grand Junction, CO, USA) (Fig. 1). The titanium plasma spray coating and other surface features that are normally present on the bone-contacting surfaces of the endplates were omitted per ASTM F2423 for ease of fixturing and to preclude third-body wear from compromising the evaluation of the articulating surfaces. The endplate footprints were all 14 × 16 mm with the exception of the impingement wear samples. Two sizes, small (12 × 15 mm) and large (16 × 18 mm), were evaluated under impingement conditions. All samples were subjected to 25 kGy terminal sterilization before testing.
Wear testing was conducted using a six-station spine wear simulator (MTS, Eden Prairie, MN, USA). Wear testing parameters for the idealized, impingement, and abrasive wear testing were established using ASTM F2423 and ISO 18192-1 as guides (Table 1) [5, 29]. A preliminary study using idealized conditions was performed to assess the effect of bovine serum concentration and solution temperature. A 1.0 million cycle (MC) test was used to assess the wear rate of the PEEK-on-ceramic bearing using the standard 20 g/L bovine serum concentration at 37° ± 2° C and a reduced concentration of 5 g/L at 23° ± 2° C (ambient temperature). Three bearings were evaluated under each lubrication condition, stopping the testing at 0.25 MC intervals to clean and inspect the bearing surfaces for wear mechanisms and to determine the mass loss of each component. The volumetric loss was calculated by dividing the mass loss of the entire device by the density of PEEK (1.3 mg/mm3). This assumes that the majority of the mass loss is coming from the PEEK endplates and not the ceramic cores. No difference was found in wear rate between the two lubrication conditions: 2.2 ± 0.3 mg/MC (37° C, 20 g/L) versus 2.4 ± 0.4 mg/MC (23° C, 5 g/L; t-test p = 0.76). Under both lubrication conditions, the endplates demonstrated microabrasive polishing and the core demonstrated no measurable change in mass. Therefore, the idealized wear test, using the reduced concentration/ambient temperature condition, was carried out on six samples to 10 MC, stopping at 0.25 MC intervals up to 1.0 MC and 0.5 MC intervals thereafter.
Table 1
The table summarizes the wear testing parameters by wear test scenario
Motions and loading profiles
Wear test parameters
Idealized*
Impingement*
Abrasive*
Flexion/extension (degrees)
± 7.5
17–18
± 7.5
Lateral bending (degrees)
± 6
N/A
± 6
Axial rotation (degrees)
± 6§
± 6§
± 6§
Axial load (N)|
50–150
150
50–150
* For each test, six samples were evaluated; for impingement testing, the six samples consisted of three small and three large samples; this includes a 10° extension bias of the fixtures; lateral bending was shifted 90° relative to flexion/extension and shifted 180° relative to axial rotation per Figure 2 found in ISO 18192-1; flexion/extension and axial load were in phase; §the magnitude of axial rotation as specified in ASTM F2423 was used in this test because it was more severe than the ± 4° magnitude specified in ISO 18192-1; |the magnitude of axial load as specified in ISO 18192-1 was used in this test because the peak load is more severe than the 100-N magnitude specified in ASTM F2423.
Impingement wear testing was conducted on three samples of two different size devices (six total) representing the smallest and largest available footplate geometries. Using a validated cervical spine finite element model, radiographs from retrieval studies of the KineFlex®|C Disc [16], and manufacturer-provided models of the Simplify Disc, the device was evaluated to determine the point at which impingement occurred in extension. Because no published standard exists for impingement testing, a validation experiment, using sample KineFlex®|C Discs, was conducted using the specified impingement motions and loading that were used in the experimental wear testing (Table 1). The resulting wear patterns on the KineFlex®|C Disc were found to mimic the wear patterns from in vivo retrieved KineFlex®|C devices, supporting the impingement testing protocol employed in this study. The three-dimensional models of the Simplify Disc were then used to assess the different footprint geometries and to determine the approximate angles of impingement. The smallest and largest endplates were selected as the worst-case contact stress and contact area impingement scenarios, respectively. After the impingement conditions were determined, a 1.0 MC test using the reduced concentration/ambient temperature condition was conducted with interval analysis at 0.05, 0.1, 0.25, 0.5, 0.75, and 1.0 MC.
Abrasive wear testing was conducted on six samples using the same parameters and sample geometry as described for the idealized testing. The test was conducted to 5 MC, stopping at 0.25 MC intervals up to 1.0 MC and 0.5 MC intervals thereafter. A titanium transfer scar, rather than slurry with particulate or scratched counterface [8, 26, 49], was used to create the abrasive tribologic test conditions. The titanium transfer scar was deposited on six cores before the test by articulating a titanium Grade 4 rod against both articulating faces of the ceramic cores. The motion of the rod was controlled axially using a servohydraulic load frame (858 MiniBionix II; MTS, Eden Prairie, MN, USA), while the core was rotated, to create circular deposits around the axis of symmetry of cores. The 10-point height (Rz) of titanium transfer on cores was monitored at each interval analysis through white light interferometry to ensure titanium transfer was still present.
At each interval, of each test, the samples were inspected for wear and damage modes and measured for mass loss. The PEEK endplates were imaged using a μCT 80 (Scanco Medical AG, Brüttisellen, Switzerland) at a maximum voxel resolution of 20 μm before the start and at the end of each test. Custom Matlab code (MathWorks, Natick, MA, USA) was developed to calculate the dimensional changes of the PEEK bearing surface and to estimate penetration [47]. This analysis resulted in a penetration map, which was used to visualize the combination of wear and deformation across the entire surface of the endplates.
For each mode of testing (ideal, impingement, abrasive), the mass loss by cycle count was tested to ensure it was normally distributed. The corresponding volume loss was determined using the density of the PEEK material (1.3 mg/mm3). Once normality was confirmed, the mass loss was analyzed using a regression analysis to assess linearity. The mean and SD mass wear rate were calculated for each mode of testing and compared using a Student’s t-test. For the regression analysis and Student’s t-test, a p value ≤ 0.05 was used to assess statistical significance. JMP 12.0.1 was used to conduct all statistical analyses (SAS, Cary, NC, USA).
Wear particle analysis was conducted using three bovine serum samples obtained at the final interval of each test scenario. The bovine serum samples were selected based on the mass loss of the devices; the serum from the stations demonstrating the average wear rate and two highest wear rates were used. Particle size, shape, and morphology were determined according to ASTM F1877 [4]. Enzymatic digestions were performed. Each serum sample was shaken by hand for 60 seconds, after which 10 mL of serum was removed from each sample and added to 100 mL of a 1% solution of enzyme in deionized water. The samples were digested at 37° C for a minimum of 3 days with periods of intermittent ultrasonication. After digestion, a 40 mL aliquot of the digested fluid was removed; each aliquot was added to 50 mL deionized water. The samples were passed through a 1 µm polycarbonate membrane (Whatman, Kent, UK). The filtrate was reserved and subsequently passed through a 100 nm polycarbonate membrane (Whatman, Kent, UK). Filters were gold/palladium sputtercoated (30 mA for 30 seconds) using a sputtercoater (208HR; Cressington, Watford, UK). Scanning electron microscopy was performed using an analytical scanning electron microscope (SEM) (Quanta 600 ESEM; FEI, Hillsboro, OR, USA). Coated samples were inserted into the SEM for imaging under high-vacuum mode at 5 kV. All 0.1- and 1.0-µm filters were imaged at 10,000× and 1000× magnification, respectively. For each filter, five fields of view were captured for image analysis. Each SEM micrograph was thresholded and then processed using a custom script in NIH ImageJ (National Institutes of Health, Bethesda, MD, USA). The particles were counted and, based on the scale bar in each SEM micrograph, a particle area was determined. Energy-dispersive X-ray spectroscopy (EDS) was used to determine representative particle elemental composition. The resulting particle equivalent circular diameter was plotted as a frequency distribution for each wear mode analyzed.

Results

The wear performance of a PEEK-on-ceramic TDR bearing under idealized, standard cervical disc conditions resulted in a microabrasive wear mode of the PEEK endplates and subtle abrasive polishing of the ceramic core (Fig. 2). The endplates demonstrated no evidence of fracture or risk of endplate penetration resulting from wear under these conditions. The ideal mass loss across the six samples was confirmed to be normally distributed. The mean wear rate for 10 MC of testing was found to be 0.9 ± 0.2 mg/MC (mean volumetric wear rate of 0.7 ± 0.1 mm3/MC for 10 MC of testing), was linear (R2 = 0.86, p < 0.0001), and no samples demonstrated runaway wear (Fig. 3). The PEEK endplates demonstrated measurable mass loss, whereas the mass of the ceramic cores remained constant through 10 MC. The penetration maps of the superior and inferior endplates indicated that the maximum penetration occurred at the articulating surface of the inferior endplate. The average penetration for this region was 0.08 ± 0.03 mm after 10 MC (Fig. 4).
Under impingement wear conditions, both the inferior and superior PEEK endplates, of both size devices, demonstrated impingement wear patches in regions consistent with the contact patches predicted by finite element analysis and solid modeling. These regions, located on the posterior rims of the endplates, were burnished and consistent with a microabrasive mechanism between the PEEK surfaces. Penetration maps confirmed the penetration was generally most apparent in the smaller size and less in the larger size configuration. The maximum penetration at the area of impingement for 1.0 MC ranged from 0.16 mm to 0.33 mm across both sizes of endplates and occurred at the rim away from the intended articulating surface (Figs. 2, 4). Despite this location, the endplates demonstrated no evidence of fracture. The mass loss from the six samples was confirmed to be normally distributed. The mean mass wear rate for 1 MC of impingement testing was linear (Size 1: R2 = 0.72, p < 0.0001; Size 3: R2 = 0.76, p < 0.0001) and found to be 1.0 ± 0.2 mg/MC and 1.9 ± 0.5 mg/MC for the small and large designs, respectively (Fig. 3). The wear rate for the larger size under impingement conditions was higher than the idealized conditions (0.98; confidence interval [CI], 0.23–1.73; p = 0.014) and comparable for the smaller size (0.13; CI, −0.62 to 0.87; p = 0.724). The volumetric wear rate for the small and large designs was 0.8 ± 0.2 mm3/MC and 1.5 ± 0.4 mm3/MC, respectively.
Under abrasive conditions, the PEEK endplates demonstrated polishing of the superior articulating surface (Fig. 2). The titanium transfer was confirmed to be present throughout the duration of the 5 MC tests but did demonstrate a reduction in height (Fig. 2). The penetration maps of the superior and inferior endplates indicated that the maximum penetration occurred at the articulating surface of the superior endplate. The mean penetration for this region was 0.13 ± 0.06 mm for 5 MC (Fig. 4). Given the initial thickness of the device is approximately 1 mm, it did not appear the PEEK endplates were readily susceptible to complete wear through under abrasive conditions. The mass loss for the six samples was normally distributed. The mean mass wear rate for 5 MC of abrasive testing was found to be 2.8 ± 0.6 mg/MC, was linear (R2 = 0.86, p < 0.0001), and, like the idealized test, no samples demonstrated runaway wear (Fig. 3). The volumetric wear rate for 5 MC of abrasive testing was 2.1 ± 0.5 mm3/MC. The abrasive conditions therefore produced the highest wear rate for the PEEK-on-ceramic bearing that was higher than both the idealized and impingement wear rates (abrasive versus ideal: 1.82, CI, 1.21–2.42, p < 0.001; abrasive versus impingement, smaller size: 1.69, CI, 0.94-2.44, p = 0.0003; abrasive versus impingement, larger size: 0.83, CI, 0.087–1.58, p = 0.0313).
Particles generated during idealized conditions were similar in size, shape, and morphology to those generated during impingement and abrasive conditions. The size distributions for all three tests were compared and found to have mean equivalent circular diameter values ranging from 0.10 to 16.69 µm (Fig. 5). During the last interval of testing for all modes, the size distributions were bimodal with the two main subgroups of particles found in the 0.1 to 0.2 µm and 1.0 to 2.0 µm size ranges (Fig. 5). Overall, 97% of the particles were less than 5.0 µm in diameter. For the idealized wear test, 65% of the particles produced were less than 1.0 µm and were 0.22 ± 0.15 µm in size. The EDS spectra for the particles for all tests indicated the particles were primarily polymeric with a small number of ceramic particles also present. Using the nomenclature for describing particle morphology according to ASTM F1877, the larger particulate tended to be smooth flakes and globular particles along with a few fibrils. The submicron particulate was primarily smooth with spheroidal granules.

Discussion

Replacing the diseased cervical disc with an arthroplasty device provides surgeons with an alternative to fusion for some patients. Although many arthroplasty devices are comprised of CoCr or titanium articulating against ultrahigh-molecular-weight polyethylene, other less traditional biomaterials for cervical TDRs may improve the clinical performance of these devices. A PEEK-on-ceramic bearing is an example of a possible bearing combination, but little is known about the wear performance under cervical disc conditions. The results of this study suggest that the PEEK-on-ceramic bearing combination is sufficiently robust and predictable for cervical TDR applications to warrant further investigation in a clinical study. The PEEK-on-ceramic bearing couple demonstrated no evidence of runaway wear, endplate perforation, or component fracture in any of the idealized or aggressive wear simulations.
This study has several limitations. The displacements and loads that were used for idealized and adverse testing were based on testing standards, and therefore the results of wear testing may not be indicative of the full range of tribologic conditions that may be encountered in vivo. Although the wear rate for each set of conditions appears repeatable within each test, the overall sample sizes and limitation of six wear stations make adding statistical power to these studies extremely resource-intensive and impractical. The test coupons used for testing were representative of the final bearing geometry intended for cervical disc replacement; however, the coupons did not have the final backside geometry or titanium plasma spray so the performance of that interface could not be assessed. The impingement and abrasive testing conditions are adverse scenarios that were based on reported methods and observations in the literature [28, 36, 37] and were derived from experience with retrievals and wear testing from other devices. This study did not explore other potential adverse scenarios that could potentially lead to different wear modes or bearing couple performance.
When compared with other cervical disc bearing couples, the idealized PEEK-on-ceramic wear rate (0.7 ± 0.1 mm3/MC) appears comparable to the published wear rates for other polymer-on-hard bearing couples (0.3–6.7 mm3/MC) and within the range of 0.2 to 1.9 mm3/MC reported for PEEK-on-PEEK cervical disc designs (Table 2). The bearing couple demonstrated no evidence of delamination wear as has been shown for some cervical disc PEEK-on-PEEK bearing couples [27].
Table 2
Summary of available wear testing data for cervical disc replacements under idealized (Mode I) conditions
Reference
Design
Bearing couple
Standard referenced
Relevant test inputs
Wear rate
Particle size
SSED for PMA Number: P060023 (Bryan® Cervical Disc) [21]
Bryan® (Medtronic Sofamor Danek, Memphis, TN, USA)
Polycarbonate urethane on titanium
N/A
FE: ± 4.9°
LB: N/A
AR: ± 3.8°
AL: 130 N and 300 N
F: 2, 4, and 6 Hz
As published:
0.4–1.4 mg/ MC
Converted (density = 1.2 mg/mm3): 0.3–1.2 mm3/MC
Greater than 90% of particles were less than 1 µm
SSED for PMA Number: P070001 (ProDisc™-C Total Disc Replacement) [20]
ProDisc™-C (Synthes Spine, West Chester, PA, USA)
UHMWPE on CoCr
N/A
FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 150 N
F: 1.0 Hz
As published:
2.59 ± 0.36 mg/MC
Converted (density = 0.94 mg/mm3): 2.8 ± 0.36 mm3/MC
Average size 0.17–0.35 µm
Nectow et al., 2008 [43]
ProDisc™-C (Synthes Spine)
UHMWPE on CoCr
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
1.99 ± 0.15 mg/MC
Converted (density = 0.94 mg/mm3): 2.12 ± 0.15 mm3/MC
Greater than 90% of particles were less than 1 µm
Bushelow et al., 2008 [12]
ProDisc™-C (Synthes Spine)
UHMWPE on CoCr
N/A
FE: ± 5°
LB: ± 5.0°
AR: ± 3.5°
AL: 15-150 N
F: 1.0 Hz
As published:
0.88 ± 0.07 mg/MC
Converted (density = 0.94 mg/mm3): 0.94 ± 0.07 mm3/MC
Greater than 90% of particles were less than 1 µm; average diameter between 0.22 and 0.37 µm
N/A
FE: ± 10°
LB: ± 5.0°
AR: ± 4.0°
AL:150 N
F: 1.0 Hz
As published:
2.59 ± 0.36 mg/MC
Converted (density = 0.94 mg/mm3): 2.75 ± 0.38 mm3/MC
ISO 18192-1
FE: ± 7.5°
LB: ± 4.0°
AR: ± 6.0°
AL: 15-150 N
F: 1.0 Hz
As published:
1.82 ± 0.11 mg/MC
Converted (density = 0.94 mg/ mm3): 1.94 ± 0.12 mm3/MC
SSED for PMA Number: P110002 (Mobi-C® Cervical Disc Prosthesis) [24]
Mobi-C® (LDR Spine, Austin, TX, USA)
UHMWPE on CoCr
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
1.55 ± 0.08 mg/MC
Converted (density = 0.94 mg/mm3): 1.64 ± 0.09 mm3/MC
Average size 0.77 µm
SSED for PMA Number: P100003 (SECURE®-C Cervical Artificial Disc) [23]
SECURE®-C (Globus Medical, Audubon, PA, USA)
UHMWPE on CoCr
N/A
FE: ± 7.0°
LB: ± 7.0°
AR: ± 1.5° (Test I) and ± 6.0° (Test II)
AL: 150 N
F: 2.0 Hz
As published:
Test I: 2.57 ± 1.21 mg/MC
Test II: 0.89 ± 0.3 mg/MC
Converted (density = 0.94 mg/mm3):
Test I: 2.73 ± 1.29 mm3/MC
Test I: 0.95 ± 0.32 mm3/MC
N/A
SSED for PMA Number: P100012 (PCM® Cervical Disc) [22]
PCM® (NuVasive, San Diego, CA, USA)
UHMWPE on CoCr
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
6.33 mg/MC
Converted (density = 0.94 mg/mm3): 6.73 mm3/MC
Average size: 0.43 ± 0.04 µm and 0.53 ± 0.15 µm for two samples, respectively
Grupp et al., 2015 [28]
activ® C (Aesculap AG, Tuttlingen, Germany)
UHMWPE on CoCr
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 1.0 Hz
As published:
1.0 ± 0.1 mg/MC
Converted (1.29 mg/mm3):
0.78 ± 0.08 mm3/MC
99% of particles were less than 1 µm
PEEK on PEEK
As published:
1.4 ± 0.4 mg/MC
Converted (1.29 mg/mm3):
1.1 ± 0.3 mm3/MC
99% of particles were less than 1 µm
CFR-PEEK on CFR-PEEK
As published:
0.02 ± 0.02 mg/MC
Converted (1.29 mg/mm3):
0.02 ± 0.02 mm3/MC
99% of particles were less than 1 µm
PEK on PEK
As published:
0.8 ± 0.1 mg/MC
Converted (1.29 mg/mm3):
0.6 ± 0.1 mm3/MC
90% of particles were less than 1 µm
Brown et al., 2010 [11]
NuNec (Pioneer Surgical, Marquette, MI, USA)
PEEK on PEEK
ASTM F2423
FE: ± 7.5°
LB: ± 6.0°
AR: ± 4.0°
AL: 50-150 N
F: 2.0 Hz
As published:
0.26 ± 0.01 mg/MC
Converted (1.29 mg/mm3):
0.20 ± 0.01 mm3/MC
N/A
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 6.0°
AL: 100 N
F: 2.0 Hz
As published:
0.32 ± 0.02 mg/MC
Converted (1.29 mg/mm3):
0.25 ± 0.02 mm3/MC
N/A
Xin et al., 2013 [51]
NuNec (Pioneer Surgical)
PEEK on PEEK
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 6.0°
AL: 50-150 N
F: 1.0 Hz
As published:
2.5 ± 0.1 mg/MC
As published:
1.9 ± 0.1 mm3/MC
N/A
Design History File Report for the Kineflex|C
Kineflex|C (SpinalMotion, Mountain View, CA)
CoCr on CoCr
ISO 18192-1
FE: ± 7.5°
LB: ± 6.0°
AR: ± 6.0°
AL: 100 N
F: 5.0 Hz
As published:
3.2 mg/ MC
Converted (Density = 8.5 mg/mm3): 0.4 mm3/MC
Average size: 0.4–0.55 µm
Kurtz et al., 2012 [35]
Prestige ST (Medtronic Sofamor Danek)
Stainless steel on stainless steel
ASTM F2423
FE: ± 9.7° (Phase II)
LB: ± 4.7° (Phase I)
AR: ± 3.8° (Phase I)
AL: 49 N (Phase I)/149 N (Phase II)
F: 1.0 Hz
As published:
Phase I: 0.72-0.76 mm3/MC
Phase II: 0.004-0.059 mm3/MC
Average size: 0.3–0.6 µm
SSED for PMA Number: P090029 (Prestige® LP Cervical Disc) [25]
Prestige® LP (Medtronic Sofamor Danek)
Titanium carbide on Titanium carbide
ASTM F2423*
FE: ± 7.5° (Phase II)
LB: ± 6.0° (Phase I)
AR: ± 6.0° (Phase I)
AL: 100 N (Phase I/II)
F: 1.0 Hz
As published:
Phase I then Phase II: 0.35 ± 0.3 mm3/MC*
Average size less than 0.2 µm
* Four wear tests were described in the SSED for the Prestige® LP Cervical Disc; the maximum standard wear rate has been reported; ASTM = American Society for Testing and Materials; ISO = International Standards Organization; FE = flexion/extension; LB = lateral bending; AR = axial rotation; AL = axial load; F = frequency; SSED = Summary of Safety and Effectiveness Data; N/A = not available.
Few studies have reported impingement wear rates for cervical disc replacements and the data that do exist are primarily for polyethylene-on-cobalt-chrome designs. The summary of safety and effectiveness for LDR’s Mobi-C device [24] indicates the impingement wear rate for the cobalt chrome-on-cobalt chrome articulation was 0.23 ± 0.20 mg/MC and 0.87 ± 1.1 mg/MC for the superior and inferior endplates, respectively. Grupp et al. [28] reported an impingement wear rate of the active-L lumbar TDR and found the impingement wear rate to range from 0.06 ± 0.17 mm3/MC to 1.44 ± 0.54 mm3/MC, depending on the impingement conditions analyzed. For these designs, the idealized wear conditions lead to polyethylene wear rather than CoCr wear; therefore, comparing the wear rates between idealized and impingement for these designs is not possible.
Although third-body damage on retrieved TDRs has been reported [37], to our knowledge, there are no other reports of cervical disc bearings being tested under abrasive conditions. Third-body damage can be assessed using a variety of methods that have been suggested in the orthopaedic literature [18, 26, 33]. For this design, the combination of titanium plasma-sprayed endplates and the ceramic core made using titanium transfer relevant. However, for other total disc designs, this method of simulating third-body wear may not be appropriate.
For the idealized wear test, 65% of the particles produced were less than 1.0 µm. In comparison to published data for polymer-on-hard couples, four studies indicated that greater than 90% of the particles were less than 1.0 µm and reported average particle sizes from other studies ranged from 0.17 to 0.77 µm (Table 2). Only one publication provided a distribution for a PEEK-on-PEEK articulation, and it stated that 99% of the particles produced were less than 1.0 µm (Table 2). Overall, it appears the PEEK material, when articulated against ceramic or PEEK under cervical disc loading conditions, produces particles that are consistently bimodally distributed with the majority of particles occurring in the 1 to 2 µm and < 0.5 µm size ranges.
Overall, the wear performance of the PEEK-on-ceramic bearing under idealized and adverse in vitro wear testing conditions suggests the bearing couple may be robust for cervical total disc applications and promising for further clinical evaluation. The polymer-on-hard bearing performance appears microabrasive in nature and not substantially different from other traditional polymer-on-hard bearing couples. However, no in vitro assessment of wear can replicate all of the clinical conditions the end device will experience. Ultimately, the in vivo response to wear debris from the couple will need to be monitored to understand the clinical impact of the debris generated.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Allain J, Le Mouel S, Goutallier D, Voisin M. Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br. 1999;81:835–842.CrossRefPubMed Allain J, Le Mouel S, Goutallier D, Voisin M. Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br. 1999;81:835–842.CrossRefPubMed
2.
Zurück zum Zitat Anderson PA, Rouleau JP, Bryan VE, Carlson CS. Wear analysis of the Bryan Cervical Disc prosthesis. Spine. 2003;28:S186–194.CrossRefPubMed Anderson PA, Rouleau JP, Bryan VE, Carlson CS. Wear analysis of the Bryan Cervical Disc prosthesis. Spine. 2003;28:S186–194.CrossRefPubMed
3.
Zurück zum Zitat Anderson PA, Rouleau JP, Toth JM, Riew KD. A comparison of simulator-tested and -retrieved cervical disc prostheses. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004. J Neurosurg Spine. 2004;1:202–210. Anderson PA, Rouleau JP, Toth JM, Riew KD. A comparison of simulator-tested and -retrieved cervical disc prostheses. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004. J Neurosurg Spine. 2004;1:202–210.
4.
Zurück zum Zitat ASTM. ASTM F1877-05(2010) Standard Practice for Characterization of Particles. West Conshohocken, PA, USA: ASTM International; 2010. ASTM. ASTM F1877-05(2010) Standard Practice for Characterization of Particles. West Conshohocken, PA, USA: ASTM International; 2010.
5.
Zurück zum Zitat ASTM. ASTM F2423-11 Standard Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses. West Conshohocken, PA, USA: ASTM International; 2011. ASTM. ASTM F2423-11 Standard Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses. West Conshohocken, PA, USA: ASTM International; 2011.
6.
Zurück zum Zitat ASTM. ASTM F2582-08 Standard Test Method for Impingement of Acetabular Prostheses. West Conshohocken, PA, USA: ASTM International; 2012. ASTM. ASTM F2582-08 Standard Test Method for Impingement of Acetabular Prostheses. West Conshohocken, PA, USA: ASTM International; 2012.
7.
Zurück zum Zitat Bao Q-B, Songer M, Pimenta L, Werner D, Reyes-Sanchez A, Balsano M, Agrillo U, Coric D, Davenport K, Yuan H. Nubac disc arthroplasty: preclinical studies and preliminary safety and efficacy evaluations. SAS J. 2007;1:36–45.CrossRefPubMedPubMedCentral Bao Q-B, Songer M, Pimenta L, Werner D, Reyes-Sanchez A, Balsano M, Agrillo U, Coric D, Davenport K, Yuan H. Nubac disc arthroplasty: preclinical studies and preliminary safety and efficacy evaluations. SAS J. 2007;1:36–45.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Bragdon CR, Jasty M, Muratoglu OK, O’Connor DO, Harris WH. Third-body wear of highly cross-linked polyethylene in a hip simulator. J Arthroplasty. 2003;18:553–61.CrossRefPubMed Bragdon CR, Jasty M, Muratoglu OK, O’Connor DO, Harris WH. Third-body wear of highly cross-linked polyethylene in a hip simulator. J Arthroplasty. 2003;18:553–61.CrossRefPubMed
9.
Zurück zum Zitat Brockett C, Carbone S, Fisher J, Jennings L. PEEK and CFR-PEEK as alternative bearing materials to polyethylene. Bone Joint J. 2016;98(Suppl 2):78. Brockett C, Carbone S, Fisher J, Jennings L. PEEK and CFR-PEEK as alternative bearing materials to polyethylene. Bone Joint J. 2016;98(Suppl 2):78.
10.
Zurück zum Zitat Brown T, Bao Q-B. The use of self-mating PEEK as an alternative bearing material for cervical disc arthroplasty: a comparison of different simulator inputs and tribological environments. Eur Spine J. 2012;21:717–26.CrossRefPubMedCentral Brown T, Bao Q-B. The use of self-mating PEEK as an alternative bearing material for cervical disc arthroplasty: a comparison of different simulator inputs and tribological environments. Eur Spine J. 2012;21:717–26.CrossRefPubMedCentral
11.
Zurück zum Zitat Brown T, Bao Q-B, Kilpela T, Songer M. An in vitro biotribological assessment of NUBAC, a polyetheretherketone-on-polyetheretherketone articulating nucleus replacement device: methodology and results from a series of wear tests using different motion profiles, test frequencies, and environmental conditions. Spine. 2010;35:E774–E781.CrossRefPubMed Brown T, Bao Q-B, Kilpela T, Songer M. An in vitro biotribological assessment of NUBAC, a polyetheretherketone-on-polyetheretherketone articulating nucleus replacement device: methodology and results from a series of wear tests using different motion profiles, test frequencies, and environmental conditions. Spine. 2010;35:E774–E781.CrossRefPubMed
12.
Zurück zum Zitat Bushelow M, Nechtow W, Hinter M, Dressel H, Kaddick C. Wear testing of a cervical total disc replacement: effect of motion and load parameters on wear rate and particle morphology. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008. Bushelow M, Nechtow W, Hinter M, Dressel H, Kaddick C. Wear testing of a cervical total disc replacement: effect of motion and load parameters on wear rate and particle morphology. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008.
13.
Zurück zum Zitat Cash DJ, Khanduja V. The case for ceramic-on-polyethylene as the preferred bearing for a young adult hip replacement. Hip Int. 2014;24:421–427.CrossRefPubMed Cash DJ, Khanduja V. The case for ceramic-on-polyethylene as the preferred bearing for a young adult hip replacement. Hip Int. 2014;24:421–427.CrossRefPubMed
14.
Zurück zum Zitat Chang CB, Yoo JJ, Song WS, Kim DJ, Koo KH, Kim HJ. Transfer of metallic debris from the metal surface of an acetabular cup to artificial femoral heads by scraping: Comparison between alumina and cobalt–chrome heads. J Biomed Mater Res B Appl Biomater. 2008;85:204–209.CrossRefPubMed Chang CB, Yoo JJ, Song WS, Kim DJ, Koo KH, Kim HJ. Transfer of metallic debris from the metal surface of an acetabular cup to artificial femoral heads by scraping: Comparison between alumina and cobalt–chrome heads. J Biomed Mater Res B Appl Biomater. 2008;85:204–209.CrossRefPubMed
15.
Zurück zum Zitat Choudhury D, Roy T, Krupka I, Hartl M, Mootanah R. Tribological investigation of ultra-high molecular weight polyethylene against advanced ceramic surfaces in total hip joint replacement. Proc Inst Mech Eng. 2015;229:410–419.CrossRef Choudhury D, Roy T, Krupka I, Hartl M, Mootanah R. Tribological investigation of ultra-high molecular weight polyethylene against advanced ceramic surfaces in total hip joint replacement. Proc Inst Mech Eng. 2015;229:410–419.CrossRef
16.
Zurück zum Zitat Coric D, Nunley PD, Guyer RD, Musante D, Carmody CN, Gordon CR, Lauryssen C, Ohnmeiss DD, Boltes MO. Prospective, randomized, multicenter study of cervical arthroplasty: 269 patients from the Kineflex|C artificial disc investigational device exemption study with a minimum 2-year follow-up: clinical article. J Neurosurg Spine. 2011;15:348–358.CrossRefPubMed Coric D, Nunley PD, Guyer RD, Musante D, Carmody CN, Gordon CR, Lauryssen C, Ohnmeiss DD, Boltes MO. Prospective, randomized, multicenter study of cervical arthroplasty: 269 patients from the Kineflex|C artificial disc investigational device exemption study with a minimum 2-year follow-up: clinical article. J Neurosurg Spine. 2011;15:348–358.CrossRefPubMed
17.
Zurück zum Zitat Della Valle AG, Doty S, Gradl G, Labissiere A, Nestor BJ. Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head. J Arthroplasty. 2004;19:532–536.CrossRefPubMed Della Valle AG, Doty S, Gradl G, Labissiere A, Nestor BJ. Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head. J Arthroplasty. 2004;19:532–536.CrossRefPubMed
18.
Zurück zum Zitat Eberhardt AW, McKee RT, Cuckler JM, Peterson DW, Beck PR, Lemons JE. Surface roughness of CoCr and ZrO2 femoral heads with metal transfer: a retrieval and wear simulator study. Int J Biomater. 2009;2009:185456.CrossRefPubMedPubMedCentral Eberhardt AW, McKee RT, Cuckler JM, Peterson DW, Beck PR, Lemons JE. Surface roughness of CoCr and ZrO2 femoral heads with metal transfer: a retrieval and wear simulator study. Int J Biomater. 2009;2009:185456.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Epinette JA, Manley MT. No differences found in bearing related hip survivorship at 10–12 years follow-up between patients with ceramic on highly cross-linked polyethylene bearings compared to patients with ceramic on ceramic bearings. J Arthroplasty. 2014;29:1369–1372.CrossRefPubMed Epinette JA, Manley MT. No differences found in bearing related hip survivorship at 10–12 years follow-up between patients with ceramic on highly cross-linked polyethylene bearings compared to patients with ceramic on ceramic bearings. J Arthroplasty. 2014;29:1369–1372.CrossRefPubMed
26.
Zurück zum Zitat Fisher J, Firkins P, Reeves E, Hailey J, Isaac G. The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene. Proc Inst Mech Eng H. 1995;209:263–264.CrossRefPubMed Fisher J, Firkins P, Reeves E, Hailey J, Isaac G. The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene. Proc Inst Mech Eng H. 1995;209:263–264.CrossRefPubMed
27.
Zurück zum Zitat Grupp TM, Meisel H-J, Cotton JA, Schwiesau J, Fritz B, Blömer W, Jansson V. Alternative bearing materials for intervertebral disc arthroplasty. Biomaterials. 2010;31:523–531.CrossRefPubMed Grupp TM, Meisel H-J, Cotton JA, Schwiesau J, Fritz B, Blömer W, Jansson V. Alternative bearing materials for intervertebral disc arthroplasty. Biomaterials. 2010;31:523–531.CrossRefPubMed
28.
Zurück zum Zitat Grupp TM, Yue JJ, Garcia R Jr, Kaddick C, Fritz B, Schilling C, Schwiesau J, Blomer W. Evaluation of impingement behaviour in lumbar spinal disc arthroplasty. Eur Spine J. 2015;24:2033–2046.CrossRefPubMed Grupp TM, Yue JJ, Garcia R Jr, Kaddick C, Fritz B, Schilling C, Schwiesau J, Blomer W. Evaluation of impingement behaviour in lumbar spinal disc arthroplasty. Eur Spine J. 2015;24:2033–2046.CrossRefPubMed
29.
Zurück zum Zitat ISO. ISO 18192-1:2011–Implants for Surgery–Wear of Total Intervertebral Spinal Disc Prostheses–Part 1: Loading and Displacement Parameters for Wear Testing and Corresponding Environmental Conditions for Test. Geneva, Switzerland: International Organization for Standardization; 2011. ISO. ISO 18192-1:2011–Implants for Surgery–Wear of Total Intervertebral Spinal Disc Prostheses–Part 1: Loading and Displacement Parameters for Wear Testing and Corresponding Environmental Conditions for Test. Geneva, Switzerland: International Organization for Standardization; 2011.
30.
Zurück zum Zitat Kim Y-H, Ritchie A, Hardaker C. Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear. J Bone Joint Surg Am. 2005;87:577–582.PubMed Kim Y-H, Ritchie A, Hardaker C. Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear. J Bone Joint Surg Am. 2005;87:577–582.PubMed
31.
Zurück zum Zitat Kraft M. Wear performance of a PEEK-on-PEEK bearing under cervical loading conditions and at different test frequencies. Global Spine J. 2012;2:S120–S121.CrossRef Kraft M. Wear performance of a PEEK-on-PEEK bearing under cervical loading conditions and at different test frequencies. Global Spine J. 2012;2:S120–S121.CrossRef
32.
Zurück zum Zitat Kraft M. Impact of the processing history on the wear performance of a PEEK-on-PEEK bearing for cervical total disc replacement. International Journal of Mechanics and Applications. 2013;3:1–7. Kraft M. Impact of the processing history on the wear performance of a PEEK-on-PEEK bearing for cervical total disc replacement. International Journal of Mechanics and Applications. 2013;3:1–7.
33.
Zurück zum Zitat Kubo K, Clarke IC, Sorimachi T, Williams PA, Donaldson TK, Yamamoto K. Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: a hip simulator model. Wear. 2009;267:734–742.CrossRef Kubo K, Clarke IC, Sorimachi T, Williams PA, Donaldson TK, Yamamoto K. Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: a hip simulator model. Wear. 2009;267:734–742.CrossRef
34.
Zurück zum Zitat Kurtz SM. Total disc arthroplasty. In: Kurtz SM, Edidin AA, eds. Spine Technology Handbook. New York, NY, USA: Academic Press; 2006:1–10.CrossRef Kurtz SM. Total disc arthroplasty. In: Kurtz SM, Edidin AA, eds. Spine Technology Handbook. New York, NY, USA: Academic Press; 2006:1–10.CrossRef
35.
Zurück zum Zitat Kurtz SM, Ciccarelli L, Harper ML, Siskey R, Shorez J, Chan FW. Comparison of in vivo and simulator-retrieved metal-on-metal cervical disc replacements. Int J Spine Surg. 2012;6:145–156.CrossRefPubMedPubMedCentral Kurtz SM, Ciccarelli L, Harper ML, Siskey R, Shorez J, Chan FW. Comparison of in vivo and simulator-retrieved metal-on-metal cervical disc replacements. Int J Spine Surg. 2012;6:145–156.CrossRefPubMedPubMedCentral
36.
Zurück zum Zitat Kurtz SM, Toth JM, Siskey R, Ciccarelli L, MacDonald D, Isaza J, Lanman T, Punt I, Steinbeck M, Goffin J, van Ooij A. The latest lessons learned from retrieval analyses of ultra-high molecular weight polyethylene, metal-on-metal, and alternative bearing total disc replacements. Semin Spine Surg. 2012;24:57–70.CrossRefPubMedPubMedCentral Kurtz SM, Toth JM, Siskey R, Ciccarelli L, MacDonald D, Isaza J, Lanman T, Punt I, Steinbeck M, Goffin J, van Ooij A. The latest lessons learned from retrieval analyses of ultra-high molecular weight polyethylene, metal-on-metal, and alternative bearing total disc replacements. Semin Spine Surg. 2012;24:57–70.CrossRefPubMedPubMedCentral
37.
Zurück zum Zitat Lebl DR, Cammisa FP, Girardi FP, Wright T, Abjornson C. In vivo functional performance of failed Prodisc-L devices: retrieval analysis of lumbar total disc replacements. Spine (Phila Pa 1976). 2012;37:E1209–1217. Lebl DR, Cammisa FP, Girardi FP, Wright T, Abjornson C. In vivo functional performance of failed Prodisc-L devices: retrieval analysis of lumbar total disc replacements. Spine (Phila Pa 1976). 2012;37:E1209–1217.
38.
Zurück zum Zitat Lübbeke A, Gonzalez A, Garavaglia G, Roussos C, Bonvin A, Stern R, Peter R, Hoffmeyer P. A comparative assessment of small-head metal-on-metal and ceramic-on-polyethylene total hip replacements. Bone Joint J. 2014;96:868–875.CrossRefPubMed Lübbeke A, Gonzalez A, Garavaglia G, Roussos C, Bonvin A, Stern R, Peter R, Hoffmeyer P. A comparative assessment of small-head metal-on-metal and ceramic-on-polyethylene total hip replacements. Bone Joint J. 2014;96:868–875.CrossRefPubMed
39.
Zurück zum Zitat Luchetti WT, Copley LA, Vresilovic EJ, Black J, Steinberg ME. Drain entrapment and titanium to ceramic head deposition. J Arthroplasty. 1998;13:713–717.CrossRefPubMed Luchetti WT, Copley LA, Vresilovic EJ, Black J, Steinberg ME. Drain entrapment and titanium to ceramic head deposition. J Arthroplasty. 1998;13:713–717.CrossRefPubMed
40.
Zurück zum Zitat Meftah M, Klingenstein GG, Yun RJ, Ranawat AS, Ranawat CS. Long-term performance of ceramic and metal femoral heads on conventional polyethylene in young and active patients. J Bone Joint Surg Am. 2013;95:1193–1197.CrossRefPubMed Meftah M, Klingenstein GG, Yun RJ, Ranawat AS, Ranawat CS. Long-term performance of ceramic and metal femoral heads on conventional polyethylene in young and active patients. J Bone Joint Surg Am. 2013;95:1193–1197.CrossRefPubMed
41.
Zurück zum Zitat Moussa ME, Esposito CI, Elpers ME, Wright TM, Padgett DE. Hip dislocation increases roughness of oxidized zirconium femoral heads in total hip arthroplasty: an analysis of 59 retrievals. J Arthroplasty. 2015;30:713–717.CrossRefPubMed Moussa ME, Esposito CI, Elpers ME, Wright TM, Padgett DE. Hip dislocation increases roughness of oxidized zirconium femoral heads in total hip arthroplasty: an analysis of 59 retrievals. J Arthroplasty. 2015;30:713–717.CrossRefPubMed
42.
Zurück zum Zitat Müller FA, Hagymási M, Greil P, Zeiler G, Schuh A. Transfer of metallic debris after dislocation of ceramic femoral heads in hip prostheses. Arch Orthop Trauma Surg. 2006;126:174–180.CrossRefPubMed Müller FA, Hagymási M, Greil P, Zeiler G, Schuh A. Transfer of metallic debris after dislocation of ceramic femoral heads in hip prostheses. Arch Orthop Trauma Surg. 2006;126:174–180.CrossRefPubMed
43.
Zurück zum Zitat Nechtow W, Bushelow M, Hinter M, Ochs A, Kaddick C. Cervical disc prosthesis polyethylene wear following the ISO cervical test. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008. Nechtow W, Bushelow M, Hinter M, Ochs A, Kaddick C. Cervical disc prosthesis polyethylene wear following the ISO cervical test. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008.
44.
45.
Zurück zum Zitat Rajpura A, Kendoff D, Board T. The current state of bearing surfaces in total hip replacement. Bone Joint J. 2014;96:147–156.CrossRefPubMed Rajpura A, Kendoff D, Board T. The current state of bearing surfaces in total hip replacement. Bone Joint J. 2014;96:147–156.CrossRefPubMed
46.
Zurück zum Zitat Rajpura A, Wroblewski B, Siney P, Board T, Jones HW. 28-year clinical and radiological follow-up of cross-linked polyethylene on alumina ceramic: a prospective study of modified Charnley low frictional torque arthroplasty. Bone Joint J. 2015;97(Suppl 12):43. Rajpura A, Wroblewski B, Siney P, Board T, Jones HW. 28-year clinical and radiological follow-up of cross-linked polyethylene on alumina ceramic: a prospective study of modified Charnley low frictional torque arthroplasty. Bone Joint J. 2015;97(Suppl 12):43.
47.
Zurück zum Zitat Shkolnikov YP, Bowden A, MacDonald D, Kurtz SM. Wear pattern observations from TDR retrievals using autoregistration of voxel data. J Biomed Mater Res B Appl Biomater. 2010;94:312–317.PubMedPubMedCentral Shkolnikov YP, Bowden A, MacDonald D, Kurtz SM. Wear pattern observations from TDR retrievals using autoregistration of voxel data. J Biomed Mater Res B Appl Biomater. 2010;94:312–317.PubMedPubMedCentral
48.
Zurück zum Zitat Sorimachi T, Clarke I, Williams P, Gustafson A, Yamamoto K. Third-body abrasive wear challenge of 32 mm conventional and 44 mm highly crosslinked polyethylene liners in a hip simulator model. Proc Inst Mech Eng H. 2009;223:607–23.CrossRefPubMed Sorimachi T, Clarke I, Williams P, Gustafson A, Yamamoto K. Third-body abrasive wear challenge of 32 mm conventional and 44 mm highly crosslinked polyethylene liners in a hip simulator model. Proc Inst Mech Eng H. 2009;223:607–23.CrossRefPubMed
49.
Zurück zum Zitat Tipper J, Firkins P, Besong A, Barbour P, Nevelos J, Stone M, Ingham E, Fisher J. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear. 2001;250:120–128.CrossRef Tipper J, Firkins P, Besong A, Barbour P, Nevelos J, Stone M, Ingham E, Fisher J. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear. 2001;250:120–128.CrossRef
50.
Zurück zum Zitat Wang A, Dumbleton JH. Ultra-high-molecular-weight polyethylene (UHMWPE) as a bearing material in hip joint replacements. In: Wang QJ, Chung Y-W, eds. Encyclopedia of Tribology. Boston, MA, USA: Springer; 2013:3933–3939.CrossRef Wang A, Dumbleton JH. Ultra-high-molecular-weight polyethylene (UHMWPE) as a bearing material in hip joint replacements. In: Wang QJ, Chung Y-W, eds. Encyclopedia of Tribology. Boston, MA, USA: Springer; 2013:3933–3939.CrossRef
51.
Zurück zum Zitat Xin H, Shepherd D, Dearn K. A tribological assessment of a PEEK based self-mating total cervical disc replacement. Wear. 2013;303:473–479.CrossRef Xin H, Shepherd D, Dearn K. A tribological assessment of a PEEK based self-mating total cervical disc replacement. Wear. 2013;303:473–479.CrossRef
Metadaten
Titel
Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study
verfasst von
Ryan Siskey, MS
Lauren Ciccarelli, MS
Melissa K. C. Lui, MS
Steven M. Kurtz, PhD
Publikationsdatum
27.09.2016
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 11/2016
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-016-5041-7

Weitere Artikel der Ausgabe 11/2016

Clinical Orthopaedics and Related Research® 11/2016 Zur Ausgabe

Symposium: Advances in PEEK Technology

Editorial Comment: Advances in PEEK Technology

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.