Insulin injection is the main treatment in patients with type 1 diabetes mellitus (T1DM). Even though continuous glucose monitoring has significantly improved the conditions of these patients, limitations still exist. To further enhance glucose control in patients with T1DM, an artificial pancreas has been developed. We aimed to systematically compare artificial pancreas with its control group during a 24-h basis in patients with T1DM.
Methods
Electronic databases were carefully searched for English publications comparing artificial pancreas with its control group. Overall daytime and nighttime glucose parameters were considered as the endpoints. Data were evaluated by means of weighted mean differences (WMDs) and 95% confidence intervals (CIs) generated by RevMan 5.3 software.
Results
A total number of 354 patients were included. Artificial pancreas significantly maintained a better mean concentration of glucose (WMD − 1.03, 95% CI − 1.32 to − 0.75; P = 0.00001). Time spent in the hypoglycemic phase was also significantly lower (WMD − 1.23, 95% CI − 1.56 to − 0.91; P = 0.00001). Daily insulin requirement also significantly favored artificial pancreas (WMD − 3.43, 95% CI − 4.27 to − 2.59; P = 0.00001). Time spent outside the euglycemic phase and hyperglycemia phase (glucose > 10.0 mmol/L) also significantly favored artificial pancreas. Also, the numbers of hypoglycemic events were not significantly different.
Conclusion
Artificial pancreas might be considered an effective and safe alternative to be used during a 24-h basis in patients with T1DM.
Type 1 diabetes mellitus (T1DM) is still a major concern in this new era. This chronic disorder occurs when beta cells of the pancreas are destroyed by autoimmune antibodies at a younger age (childhood) or it can even develop in adults, during their late 30s or 40s [1]. Insulin injection has been the main treatment in these patients with T1DM [2]. However, despite concerted efforts of patients and physicians/caregivers, control of blood glucose has often been difficult to achieve [3].
With the development of new techniques and devices in clinical medicine, continuous glucose monitoring has significantly improved conditions of patients with T1DM [4]. To further enhance glucose control in these patients, an artificial pancreas has recently been developed [5].
Anzeige
Even though this artificial pancreas has already been approved for use by the US Food and Drug Administration (FDA) [5], the existence and benefits of this “expected to be” effective device are not well known to the general population.
In this analysis, we aimed to systematically compare artificial pancreas with its control group in terms of effectiveness and safety, during a 24-h treatment of patients with T1DM.
Methods
Searched Databases
The Medline database of medical publications, the Cochrane library of randomized controlled trials, and EMBASE database were carefully searched by the five authors for English publications comparing artificial pancreas with its control group (any control group). Reference lists of selective publications (most relevant ones) were also carefully reviewed for appropriate articles. In addition, official websites of specific journals such as Lancet Diabetes and Endocrinology, Diabetes Care, and Cardiovascular Diabetology were also searched for any relevant publication.
Search Strategies
During this process, we searched for the terms “artificial pancreas”, “bionic pancreas”, “closed loop glucose control”, “type 1 diabetes mellitus”, “diabetes mellitus”, “glucose control”, and “glucose monitoring”, one at a time and in combination. Publications comparing artificial pancreas with its control group were considered relevant to this analysis.
Anzeige
Inclusion Criteria
Studies were included if they compared artificial pancreas with its control group (any relevant control group); they reported glucose control parameters or other outcomes related to glucose monitoring (during daytime and overnight/24-h basis); they involved patients with T1DM.
Exclusion Criteria
Studies were excluded if they did not compare artificial pancreas with its control group; they did not report glucose monitoring parameters as their endpoints or they reported only daytime or overnight measurement, but not both in combination; they involved patients with type 2 diabetes mellitus instead of patients with T1DM; they were duplicates.
Type of Participants, Endpoints, and Follow-Up Periods
This analysis included patients with T1DM.
The endpoints (during daytime and nighttime) which were analyzed included:
(a)
Mean and median glucose concentration.
(b)
Time spent outside the euglycemic phase (outside the glucose range 3.90 to 8.0 mmol/L).
(c)
Time spent in the hypoglycemia phase (blood glucose < 3.9 mmol/L).
(d)
Insulin required/delivered per day.
(e)
Time spent in the hyperglycemia phase (blood glucose > 10.0 mmol/L).
(f)
Number of hypoglycemic events.
The follow-up period ranged from less than 1 week to 2 months.
Data Extraction, Review, and Statistical Analysis
After the search process, which was conducted in accordance with the PRISMA guideline [6], the same reviewers assessed the titles and abstracts and independently selected the most suitable articles which satisfied the inclusion and exclusion criteria of this analysis and then data were extracted. The studies which were included in this analysis were judged as having low to moderate risk of bias [7].
This is a meta-analysis and therefore inconsistency across the studies was evident [8]. Hence, heterogeneity was assessed by two statistical methods:
(a)
The Q statistic test, whereby a P value less or equal to 0.05 was considered statistically significant.
(b)
The I2 statistic test; a high percentage value indicated high heterogeneity (whereby a random effects model was used) and low percentage value denoted low heterogeneity (whereby a fixed effects model was used).
Since continuous data was used in this analysis, i.e., mean and standard deviation (SD), data were evaluated by means of weighted mean differences (WMDs) and 95% confidence intervals (CIs). In case the SD value was not provided, but a p value was given, SD was calculated using the formula SD = √n × p × (1 – p). The analysis was carried out by RevMan 5.3 software.
Anzeige
Publication bias was assessed by visually observing funnel plots.
Compliance with Ethics Guidelines
This meta-analysis is based on previously conducted studies and does not contain any studies with human participants or animals performed by any of the authors.
Results
Search Outcomes, Main and Baseline Features of the Studies
Eight studies [9‐16] were selected to be used in this analysis as shown in Fig. 1.
×
Table 1 summarizes the main features of the studies included in this analysis. As per the criteria of this analysis which required an experimental and a control group, a total of 354 patients were included (177 patients in each group) as shown in Table 1.
AP artificial pancreas, C control group, HBA1c glycosylated hemoglobin, BMI body mass index, DM diabetes mellitus
Analysis Results
Results of this current analysis showed artificial pancreas to significantly maintain a better mean concentration of glucose with WMD − 1.03, 95% CI − 1.32 to − 0.75; P = 0.00001, I2 = 46% compared to the control group during a 24-h basis. The median glucose concentration was similar in both groups with WMD − 0.30, 95% CI − 1.03 to 0.44; P = 0.43, I2 = 0% as shown in Fig. 2.
×
In addition, the time spent in the hypoglycemic phase (glucose < 3.9 mmol/L) also significantly favored artificial pancreas with WMD − 1.23, 95% CI − 1.56 to − 0.91; P = 0.00001, I2 = 19% meaning the patients experienced less time in the hypoglycemia phase with this artificial pancreas as compared to the control group (Fig. 2).
Daily insulin required (24-h basis) also significantly favored artificial pancreas with WMD − 3.43, 95% CI − 4.27 to − 2.59; P = 0.00001, I2 = 0% compared to the control group, indicating that good glucose control was continuously maintained without requiring an excess of insulin (Fig. 2).
Anzeige
Also, the numbers of hypoglycemic events were not significantly different with WMD − 0.83, 95% CI − 1.76 to 0.10; P = 0.08, I2 = 0% (Fig. 2).
Time spent outside the euglycemic phase (24 h-basis) also significantly favored artificial pancreas with WMD − 6.28, 95% CI − 10.67 to − 1.88; P = 0.005. This meant that most of the time, patients using artificial pancreas were in the euglycemic phase (neither experiencing hypoglycemia nor hyperglycemia) as shown in Fig. 3. However, the results were moderately heterogeneous.
×
Moreover, the time spent in the hyperglycemia phase (glucose > 10.0 mmol/L) also significantly favored artificial pancreas with WMD − 13.20, 95% CI − 16.47 to − 9.94; P = 0.00001 meaning that patients using artificial pancreas hardly suffered any hyperglycemic stage as shown in Fig. 3. However, this result was also moderately heterogeneous.
Discussion
The use of a fully integrated artificial pancreas in patients with T1DM was previously demonstrated [17].
The current results showed artificial pancreas to be significantly more effective compared to its control group in terms of glucose concentration, time spent in the hypoglycemic phase, and insulin delivery during a 24-h period. Artificial pancreas was also safer to use owing to its association with a significantly lower time period in the hyperglycemia phase, its significant maintenance of a longer euglycemic period, and its lack of association with any significantly higher episode of hypoglycemic event compared to its control.
Similarly, Hovorka et al. showed that artificial pancreas improved overnight control of glucose level and decreased the rate of nocturnal hypoglycemia in patients with T1DM within a study time period of 3 months [18]. Another multicenter study showed this artificial pancreas to be very effective and safe to use in patients with T1DM [19]. Similarly, through a multicenter 6-month trial of 24/7 automated insulin delivery in 2014, Kovatchev et al. recently showed closed-loop control technology to have matured and to appear safe for long-term use in patients with T1DM [20].
This new device was even considered effective in pediatric participants. Weinzimer et al. recently demonstrated fully automated closed-loop insulin delivery versus semi-automated hybrid control in pediatric candidates with T1DM [21]. Insulin delivery using artificial pancreas was further illustrated in the Virginia experience, wherein the participants were studied twice, once using their personal open-loop technique, and then a second time using the closed-loop (artificial) system [22]. Other research further complemented the closed-loop insulin delivery technique [23, 24].
Further improvement is being considered in relation to this artificial pancreas [25]. Also, useful tools have already been devised to improve the assessment of glycemic variability in patients with artificial pancreas [26].
Limitations
This analysis also has limitations: (a) The number of participants was extremely limited; however, when compared to other previously published studies, this analysis included a large number of patients. (b) The different follow-up time periods could have had an impact on the results obtained. (c) The range of the euglycemic phase was supposed to be a glucose level ranging between 3.9 and 8.0 mmol/L; however, a few studies recorded a glucose level varying between 4.4 and 8.0 mmol/L or 3.9 to 10.0 mmol/L which might have contributed to the moderate level of heterogeneity in this particular subgroup. (d) The inclusion of one non-randomized study might have introduced bias, contributing to the limitations in this analysis. (e) The control groups were not similar in all the studies, which might be another limitation of this analysis. (f) Utilizing sensor augmented pump as the control group is the current clinical golden standard which artificial pancreas needs to be able to outperform if clinically relevant. So, another limitation of this study might be the lack of an analysis strictly dealing with studies comparing artificial pancreas and sensor-augmented pump. However, the number of studies reporting this control was too small.
Conclusion
According to the results of this analysis, artificial pancreas might be considered an effective and safe alternative to be used during a 24-h basis in patients with T1DM. Several benefits of the artificial pancreas in maintaining and improving glucose levels were observed in comparison to its control. Nevertheless, a major shortcoming of this analysis is the extremely limited number of patients analyzed.
Acknowledgements
Funding
This research was supported by National Natural Science Foundation of China (No. 81560046), Guangxi Natural Science Foundation (No. 2016GXNSFAA380002), Scientific Project of Guangxi Higher Education (No. KY2015ZD028), Science Research and Technology Development Project of Qingxiu District of Nanning (No. 2016058), and Lisheng Health Foundation pilotage fund of Peking (No. LHJJ20158126). No funding or sponsorship was received for the publication of this article. The article processing charges were funded by the authors.
Authorship
All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this article, take responsibility for the integrity of the work as a whole, and have given their approval for this version to be published.
Authorship Contributions
Xia Dai, Zu-chun Luo, Lu Zhai, Wen-piao Zhao, and Feng Huang were responsible for the conception and design, acquisition of data, analysis and interpretation of data, drafting the initial manuscript and revising it critically for important intellectual content. Xia Dai wrote the final manuscript. All the authors approved the manuscript as it is.
Disclosures
Xia Dai, Zu-chun Luo, Lu Zhai, Wen-piao Zhao, and Feng Huang declare that they have nothing to disclose. They do not have any personal, financial, commercial, or academic conflicts of interest.
Compliance with Ethics Guidelines
This meta-analysis is based on previously conducted studies and does not contain any studies with human participants or animals performed by any of the authors.
Data Availability
All data generated or analyzed during this study are included in this published article.
Open Access
This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.
Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.
Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.
Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.
Update Innere Medizin
Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.