Skip to main content
Erschienen in: Journal of Interventional Cardiac Electrophysiology 3/2020

06.03.2020 | Cardiac Resynchronization Therapy

Extending multipoint pacing CRT battery longevity by swapping left ventricular pulse configurations

verfasst von: Juan José García Guerrero, Joaquín Fernández de la Concha Castañeda, Antonio Chacón Piñero, Carlos Hidalgo Martínez, Nima Badie, Luke McSpadden, Kyungmoo Ryu

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 3/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Cardiac resynchronization therapy (CRT) with left ventricular (LV) MultiPoint™ pacing (MPP) has been shown to improve CRT response by pacing two LV sites (LV1, LV2). While an additional LV pacing site reduces battery longevity, this cost can be minimized by leveraging an existing device-based capture management algorithm (LVCap™ Confirm). The purpose of this study was to evaluate the MPP battery longevity improvement achieved by configuring LV pacing sites to properly leverage LVCap Confirm.

Methods

Patients previously enrolled in the MORE-CRT MPP trial with existing MPP-enabled CRT-D devices (Abbott Quadra Assura MP™ CD3371-40QC, Quartet™ LV lead) underwent device interrogation. Device electrical characteristics and estimated battery longevities were compared for various MPP settings.

Results

At 2.1 ± 1.1 years post-implant, the estimated remaining battery longevity in 65 patients was 70 ± 14 months with MPP Off (LV pacing from minimum capture threshold). Enabling MPP with maximal anatomical separation between LV1 and LV2 cathodes reduced longevity by 15 ± 14%. However, swapping the LV1 and LV2 cathodes, such that the LV1 threshold was the higher of the two, allowed the device to take full advantage of the LVCap™ Confirm capture management algorithm, resulting in significantly lower longevity reduction of 9 ± 11% (p < 0.001). Ultimately, a mean MPP battery longevity improvement of 7.7 ± 10.3% (p < 0.001) was achieved by simply swapping LV1/LV2 configurations.

Conclusions

By properly leveraging device-based capture management features, the impact of MPP on battery longevity can be significantly reduced.
Literatur
4.
Zurück zum Zitat Pappone C, Calovic Z, Vicedomini G, Cuko A, McSpadden LC, Ryu K, … Santinelli V. Multipoint left ventricular pacing improves acute hemodynamic response assessed with pressure-volume loops in cardiac resynchronization therapy patients. Heart Rhythm. 2014;11(3):394–401 https://doi.org/10.1016/j.hrthm.2013.11.023. Pappone C, Calovic Z, Vicedomini G, Cuko A, McSpadden LC, Ryu K, … Santinelli V. Multipoint left ventricular pacing improves acute hemodynamic response assessed with pressure-volume loops in cardiac resynchronization therapy patients. Heart Rhythm. 2014;11(3):394–401 https://​doi.​org/​10.​1016/​j.​hrthm.​2013.​11.​023.
6.
Zurück zum Zitat Osca J, Alonso P, Cano O, Andres A, Miro V, Tello MJ, … Salvador A. The use of multisite left ventricular pacing via quadripolar lead improves acute haemodynamics and mechanical dyssynchrony assessed by radial strain speckle tracking: initial results. Europace. 2016;18:560–567 https://doi.org/10.1093/europace/euv211. Osca J, Alonso P, Cano O, Andres A, Miro V, Tello MJ, … Salvador A. The use of multisite left ventricular pacing via quadripolar lead improves acute haemodynamics and mechanical dyssynchrony assessed by radial strain speckle tracking: initial results. Europace. 2016;18:560–567 https://​doi.​org/​10.​1093/​europace/​euv211.
9.
Zurück zum Zitat Forleo GB, Santini L, Giammaria M, Potenza D, Curnis A, Calabrese V, et al. Multipoint pacing via a quadripolar left-ventricular lead: preliminary results from the Italian registry on multipoint left-ventricular pacing in cardiac resynchronization therapy (IRON-MPP). Europace. 2017;19(7):1170–7. https://doi.org/10.1093/europace/euw094.CrossRefPubMed Forleo GB, Santini L, Giammaria M, Potenza D, Curnis A, Calabrese V, et al. Multipoint pacing via a quadripolar left-ventricular lead: preliminary results from the Italian registry on multipoint left-ventricular pacing in cardiac resynchronization therapy (IRON-MPP). Europace. 2017;19(7):1170–7. https://​doi.​org/​10.​1093/​europace/​euw094.CrossRefPubMed
10.
Zurück zum Zitat Zanon F, Marcantoni L, Baracca E, Pastore G, Lanza D, Aggio S, … Prinzen FW. Optimization of left ventricular pacing site plus multipoint pacing improves remodeling and clinical response to cardiac resynchronization therapy at 1 year. Heart Rhythm. 2016;13(8):1644–1651 https://doi.org/10.1016/S0735-1097(16)31281-5. Zanon F, Marcantoni L, Baracca E, Pastore G, Lanza D, Aggio S, … Prinzen FW. Optimization of left ventricular pacing site plus multipoint pacing improves remodeling and clinical response to cardiac resynchronization therapy at 1 year. Heart Rhythm. 2016;13(8):1644–1651 https://​doi.​org/​10.​1016/​S0735-1097(16)31281-5.
11.
Metadaten
Titel
Extending multipoint pacing CRT battery longevity by swapping left ventricular pulse configurations
verfasst von
Juan José García Guerrero
Joaquín Fernández de la Concha Castañeda
Antonio Chacón Piñero
Carlos Hidalgo Martínez
Nima Badie
Luke McSpadden
Kyungmoo Ryu
Publikationsdatum
06.03.2020
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 3/2020
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-020-00720-2

Weitere Artikel der Ausgabe 3/2020

Journal of Interventional Cardiac Electrophysiology 3/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.