12.06.2021 | Hip Arthroplasty

Cementless femoral stems with lower canal fill ratio have similar mid-term to long-term outcomes to those with adequate fill ratio in Dorr type C femurs

verfasst von: Mingyang Li, Yi Zeng, Yuangang Wu, Yuan Liu, Limin Wu, Bin Shen

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 6/2022

Einloggen, um Zugang zu erhalten

Abstract

Background

Lower canal fill ratio was reported to correlate with aseptic loosening in many studies. However, the most widely used standard of fill ratio seemed inapplicable to Dorr type C femurs. We aimed to adapt the method of measuring the fill ratio in Dorr type C femurs and compare the outcomes among patients with different fill ratios.

Methods

Twenty patients with Corail stems implanted in their Dorr type C femurs received spectrum CT to evaluate the whole-stem's fill ratio. Pearson Correlation Coefficient was calculated to assess the correlation between the fill ratio in X-ray film and spectrum CT. Then 87 THAs were involved in this study, divided into the fill ratio ≤ 80% group and the fill ratio > 80% group. Clinical and radiological outcomes were evaluated with a mean follow-up of 8.2 years.

Results

Fill ratio at 2 cm below the lesser trochanter in anterior–posterior X-ray film correlated with the whole-stem's fill ratio (r = 0.50, P = 0.02). Survival rate of stem, function scores, and radiological outcomes between the two groups showed no significant difference. In the fill ratio > 80% group, intraoperative fracture was significantly higher (19% VS 5%, P < 0.05).

Conclusion

Patients with lower fill ratios at 2 cm below the lesser trochanter did not have poorer functional scores or more subsidence, but had a lower intraoperative fracture rate. The revision rates of the two groups presented no significant difference, but this result need to be confirmed in larger cohort in the future. In Dorr type C femurs, risk of fracture and the special morphology of the femur should be noted, and high fill ratio is not the most decisive factor for stem size selecting.
Literatur
1.
Zurück zum Zitat Achten J, Parsons NR, Edlin RP, Griffin DR, Costa ML (2010) A randomised controlled trial of total hip arthroplasty versus resurfacing arthroplasty in the treatment of young patients with arthritis of the hip joint. BMC Musculoskelet Disord 11:8CrossRef Achten J, Parsons NR, Edlin RP, Griffin DR, Costa ML (2010) A randomised controlled trial of total hip arthroplasty versus resurfacing arthroplasty in the treatment of young patients with arthritis of the hip joint. BMC Musculoskelet Disord 11:8CrossRef
2.
Zurück zum Zitat Al-Najjim M, Khattak U, Sim J, Chambers I (2016) Differences in subsidence rate between alternative designs of a commonly used uncemented femoral stem. J Orthop 13(4):322–326CrossRef Al-Najjim M, Khattak U, Sim J, Chambers I (2016) Differences in subsidence rate between alternative designs of a commonly used uncemented femoral stem. J Orthop 13(4):322–326CrossRef
3.
Zurück zum Zitat Al-Dirini RM, Huff D, Zhang J, Besier T, Clement JG, Taylor M (2018) Influence of collars on the primary stability of cementless femoral stems: a finite element study using a diverse patient cohort. J Orthop Res 36(4):1185–1195CrossRef Al-Dirini RM, Huff D, Zhang J, Besier T, Clement JG, Taylor M (2018) Influence of collars on the primary stability of cementless femoral stems: a finite element study using a diverse patient cohort. J Orthop Res 36(4):1185–1195CrossRef
4.
Zurück zum Zitat Aldinger PR, Jung AW, Breusch SJ, Ewerbeck V, Parsch D (2009) Survival of the cementless Spotorno® stem in the second decade. Clin Orthop Relat Res 467(9):2297–2304CrossRef Aldinger PR, Jung AW, Breusch SJ, Ewerbeck V, Parsch D (2009) Survival of the cementless Spotorno® stem in the second decade. Clin Orthop Relat Res 467(9):2297–2304CrossRef
5.
Zurück zum Zitat Biedermann R, Krismer M, Stöckl B, Mayrhofer P, Ornstein E, Franzén H (1999) Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement Einzel-Bild-Röntgen-Analyse-femoral component analysis. J Bone Jt Surg Br Vol 81(2):266–272CrossRef Biedermann R, Krismer M, Stöckl B, Mayrhofer P, Ornstein E, Franzén H (1999) Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement Einzel-Bild-Röntgen-Analyse-femoral component analysis. J Bone Jt Surg Br Vol 81(2):266–272CrossRef
6.
Zurück zum Zitat Camine VM, Rüdiger HA, Pioletti DP, Terrier A (2018) Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography. Int Orthop 42(1):49–57CrossRef Camine VM, Rüdiger HA, Pioletti DP, Terrier A (2018) Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography. Int Orthop 42(1):49–57CrossRef
7.
Zurück zum Zitat Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH (1993) Structural and cellular assessment of bone quality of proximal femur. Bone 14(3):231–242CrossRef Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH (1993) Structural and cellular assessment of bone quality of proximal femur. Bone 14(3):231–242CrossRef
8.
Zurück zum Zitat Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Jt Surg Br Vol 69(1):45–55CrossRef Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Jt Surg Br Vol 69(1):45–55CrossRef
9.
Zurück zum Zitat Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 257:107–128CrossRef Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 257:107–128CrossRef
10.
Zurück zum Zitat Evola F, Evola G, Graceffa A et al (2014) Performance of the CLS Spotorno uncemented stem in the third decade after implantation. Bone Jt J 96(4):455–461CrossRef Evola F, Evola G, Graceffa A et al (2014) Performance of the CLS Spotorno uncemented stem in the third decade after implantation. Bone Jt J 96(4):455–461CrossRef
11.
Zurück zum Zitat Fottner A, Woiczinski M, Kistler M et al (2017) Influence of undersized cementless hip stems on primary stability and strain distribution. Arch Orthop Trauma Surg 137(10):1435–1441CrossRef Fottner A, Woiczinski M, Kistler M et al (2017) Influence of undersized cementless hip stems on primary stability and strain distribution. Arch Orthop Trauma Surg 137(10):1435–1441CrossRef
12.
Zurück zum Zitat Gromov K, Bersang A, Nielsen CS, Kallemose T, Husted H, Troelsen A (2017) Risk factors for post-operative periprosthetic fractures following primary total hip arthroplasty with a proximally coated double-tapered cementless femoral component. Bone Jt J 99-b(4):451–457CrossRef Gromov K, Bersang A, Nielsen CS, Kallemose T, Husted H, Troelsen A (2017) Risk factors for post-operative periprosthetic fractures following primary total hip arthroplasty with a proximally coated double-tapered cementless femoral component. Bone Jt J 99-b(4):451–457CrossRef
13.
Zurück zum Zitat Hulley SBCS, Browner WS, Grady D, Newman TB (2013) Designing clinical research: an epidemiologic approach, 4th edn. Lippincott Williams & Wilkins, Philadelphia, p 75 (Appendix 6B) Hulley SBCS, Browner WS, Grady D, Newman TB (2013) Designing clinical research: an epidemiologic approach, 4th edn. Lippincott Williams & Wilkins, Philadelphia, p 75 (Appendix 6B)
14.
Zurück zum Zitat Hwang K-T, Kim Y-H, Kim Y-S, Choi I-Y (2012) Total hip arthroplasty using cementless grit-blasted femoral component: a minimum 10-year follow-up study. J Arthroplasty 27(8):1554–1561CrossRef Hwang K-T, Kim Y-H, Kim Y-S, Choi I-Y (2012) Total hip arthroplasty using cementless grit-blasted femoral component: a minimum 10-year follow-up study. J Arthroplasty 27(8):1554–1561CrossRef
15.
Zurück zum Zitat Ishii S, Homma Y, Baba T, Ozaki Y, Matsumoto M, Kaneko K (2016) Does the canal fill ratio and femoral morphology of Asian females influence early radiographic outcomes of total hip arthroplasty with an uncemented proximally coated, tapered-wedge stem? J Arthroplasty 31(7):1524–1528CrossRef Ishii S, Homma Y, Baba T, Ozaki Y, Matsumoto M, Kaneko K (2016) Does the canal fill ratio and femoral morphology of Asian females influence early radiographic outcomes of total hip arthroplasty with an uncemented proximally coated, tapered-wedge stem? J Arthroplasty 31(7):1524–1528CrossRef
16.
Zurück zum Zitat Khalily C, Lester DK (2002) Results of a tapered cementless femoral stem implanted in varus. J Arthroplasty 17(4):463–466CrossRef Khalily C, Lester DK (2002) Results of a tapered cementless femoral stem implanted in varus. J Arthroplasty 17(4):463–466CrossRef
17.
Zurück zum Zitat Krismer M, Biedermann R, Stöckl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis. J Bone Jt Surg Br Vol 81(2):273–280CrossRef Krismer M, Biedermann R, Stöckl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis. J Bone Jt Surg Br Vol 81(2):273–280CrossRef
18.
Zurück zum Zitat McLaughlin JR, Lee KR (2016) Long-term results of uncemented total hip arthroplasty with the Taperloc femoral component in patients with Dorr type C proximal femoral morphology. Bone Jt J 98-b(5):595–600CrossRef McLaughlin JR, Lee KR (2016) Long-term results of uncemented total hip arthroplasty with the Taperloc femoral component in patients with Dorr type C proximal femoral morphology. Bone Jt J 98-b(5):595–600CrossRef
19.
Zurück zum Zitat Nakaya R, Takao M, Hamada H, Sakai T, Sugano N (2019) Reproducibility of the Dorr classification and its quantitative indices on plain radiographs. Orthop Traumatol Surg Res OTSR 105(1):17–21CrossRef Nakaya R, Takao M, Hamada H, Sakai T, Sugano N (2019) Reproducibility of the Dorr classification and its quantitative indices on plain radiographs. Orthop Traumatol Surg Res OTSR 105(1):17–21CrossRef
20.
Zurück zum Zitat Nash W, Harris A (2014) The Dorr type and cortical thickness index of the proximal femur for predicting peri-operative complications during hemiarthroplasty. J Orthop Surg 22(1):92–95CrossRef Nash W, Harris A (2014) The Dorr type and cortical thickness index of the proximal femur for predicting peri-operative complications during hemiarthroplasty. J Orthop Surg 22(1):92–95CrossRef
21.
Zurück zum Zitat Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS (1988) The anatomic basis of femoral component design. Clin Orthop Relat Res 235:148–165CrossRef Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS (1988) The anatomic basis of femoral component design. Clin Orthop Relat Res 235:148–165CrossRef
22.
Zurück zum Zitat Panichkul P, Bavonratanavech S, Arirachakaran A, Kongtharvonskul J (2019) Comparative outcomes between collared versus collarless and short versus long stem of direct anterior approach total hip arthroplasty: a systematic review and indirect meta-analysis. Eur J Orthop Surg Traumatol 29(8):1693–1704CrossRef Panichkul P, Bavonratanavech S, Arirachakaran A, Kongtharvonskul J (2019) Comparative outcomes between collared versus collarless and short versus long stem of direct anterior approach total hip arthroplasty: a systematic review and indirect meta-analysis. Eur J Orthop Surg Traumatol 29(8):1693–1704CrossRef
23.
Zurück zum Zitat Rattanaprichavej P, Laoruengthana A, Chotanaphuti T, Khuangsirikul S, Phreethanutt C, Pongpirul K (2019) Subsidence of hydroxyapatite-coated femoral stem in Dorr type C proximal femoral morphology. J Arthroplasty 34(9):2011–2015CrossRef Rattanaprichavej P, Laoruengthana A, Chotanaphuti T, Khuangsirikul S, Phreethanutt C, Pongpirul K (2019) Subsidence of hydroxyapatite-coated femoral stem in Dorr type C proximal femoral morphology. J Arthroplasty 34(9):2011–2015CrossRef
24.
Zurück zum Zitat Ries C, Boese CK, Dietrich F, Miehlke W, Heisel C (2019) Femoral stem subsidence in cementless total hip arthroplasty: a retrospective single-centre study. Int Orthop 43(2):307–314CrossRef Ries C, Boese CK, Dietrich F, Miehlke W, Heisel C (2019) Femoral stem subsidence in cementless total hip arthroplasty: a retrospective single-centre study. Int Orthop 43(2):307–314CrossRef
25.
Zurück zum Zitat Streit MR, Innmann MM, Merle C, Bruckner T, Aldinger PR, Gotterbarm T (2013) Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship. Clin Orthop Relat Res 471(10):3262–3269CrossRef Streit MR, Innmann MM, Merle C, Bruckner T, Aldinger PR, Gotterbarm T (2013) Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship. Clin Orthop Relat Res 471(10):3262–3269CrossRef
26.
Zurück zum Zitat Syed F, Hussein A, Katam K, Saunders P, Young SK, Faisal M (2018) Risk of subsidence and peri-prosthetic fractures using collared hydroxyapatite-coated stem for hip arthroplasty in the elderly. HIP Int 28(6):663–667CrossRef Syed F, Hussein A, Katam K, Saunders P, Young SK, Faisal M (2018) Risk of subsidence and peri-prosthetic fractures using collared hydroxyapatite-coated stem for hip arthroplasty in the elderly. HIP Int 28(6):663–667CrossRef
27.
Zurück zum Zitat Vidalain J-P, Selmi TAS, Beverland D et al (2011) The CORAIL® Hip System: a practical approach based on 25 years of experience. Springer, pp 23–29 Vidalain J-P, Selmi TAS, Beverland D et al (2011) The CORAIL® Hip System: a practical approach based on 25 years of experience. Springer, pp 23–29
28.
Zurück zum Zitat Young S, Michel MC, Selmi TAS et al (2011) The Corail® hip stemthe Corail® hip system. Springer, Berlin, pp 53–88CrossRef Young S, Michel MC, Selmi TAS et al (2011) The Corail® hip stemthe Corail® hip system. Springer, Berlin, pp 53–88CrossRef
Metadaten
Titel
Cementless femoral stems with lower canal fill ratio have similar mid-term to long-term outcomes to those with adequate fill ratio in Dorr type C femurs
verfasst von
Mingyang Li
Yi Zeng
Yuangang Wu
Yuan Liu
Limin Wu
Bin Shen
Publikationsdatum
12.06.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 6/2022
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-021-03916-x

Weitere Artikel der Ausgabe 6/2022

Archives of Orthopaedic and Trauma Surgery 6/2022 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.