07.06.2017 | Original Paper

Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography

verfasst von: Valérie Malfroy Camine, Hannes A. Rüdiger, Dominique P. Pioletti, Alexandre Terrier

Erschienen in: International Orthopaedics | Ausgabe 1/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The aim of this study is to quantitatively compare the difference in primary stability between collarless and collared versions of the same femoral stem. Specifically, we tested differences in subsidence and micromotion.

Methods

Collarless and collared versions of the same cementless femoral stem were implanted in two groups of six fresh-frozen cadaveric femurs. Each implanted femur was then subsequently tested for axial compressive and torsional loadings. A micro-CT based technique was applied to quantify implant subsidence and compute the map of local micromotion around the femoral stems. Micromotion of collarless and collared stems was compared in each Gruen zone.

Results

Subsidence was higher but not significantly (p = 0.352) with collarless (41.0 ± 29.9 μm) than with collared stems (37.0 ± 44.6 μm). In compression, micromotion was lower (p = 0.257) with collarless (19.5 ± 5 μm) than with collared stems (43.3 ± 33.1 μm). In torsion, micromotion was also lower (p = 0.476) with collarless (96.9 ± 59.8 μm) than collared stems (118.7 ± 45.0 μm). Micromotion was only significantly lower (p = 0.001) in Gruen zone 1 and for compression with collarless (7.0 ± 0.6 μm) than with collared stems (22.6 ± 25.5 μm).

Conclusions

Primary stability was achieved for both stem designs, with a mean micromotion below the osseointegration threshold. Under loading conditions similar to those observed in normal daily activity and with good press-fit, the collar had no influence on subsidence or micromotion. Further studies are required to test the potential advantage of collar with higher loads, undersized stems, or osteoporotic femurs.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
2.
Zurück zum Zitat Soballe K, Hansen ES, B-R H et al (1992) Tissue ingrowth into titanium and hydroxyapatite-coated implants during stable and unstable mechanical conditions. J Orthop Res 10:285–299CrossRefPubMed Soballe K, Hansen ES, B-R H et al (1992) Tissue ingrowth into titanium and hydroxyapatite-coated implants during stable and unstable mechanical conditions. J Orthop Res 10:285–299CrossRefPubMed
4.
Zurück zum Zitat Demey G, Fary C, Lustig S et al (2011) Does a collar improve the immediate stability of Uncemented femoral hip stems in Total hip arthroplasty? A bilateral comparative cadaver study. J Arthroplast 26:1549–1555. doi:10.1016/j.arth.2011.03.030 CrossRef Demey G, Fary C, Lustig S et al (2011) Does a collar improve the immediate stability of Uncemented femoral hip stems in Total hip arthroplasty? A bilateral comparative cadaver study. J Arthroplast 26:1549–1555. doi:10.​1016/​j.​arth.​2011.​03.​030 CrossRef
5.
Zurück zum Zitat Ebramzadeh E, Sangiorgio SN, Longjohn DB et al (2004) Initial stability of cemented femoral stems as a function of surface finish, collar, and stem size. J Bone Jt Surg Am 86–A:106–115CrossRef Ebramzadeh E, Sangiorgio SN, Longjohn DB et al (2004) Initial stability of cemented femoral stems as a function of surface finish, collar, and stem size. J Bone Jt Surg Am 86–A:106–115CrossRef
8.
11.
Zurück zum Zitat Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop 17–27 Gruen TA, McNeice GM, Amstutz HC (1979) “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop 17–27
12.
Zurück zum Zitat Jameson SS, Baker PN, Mason J et al (2013) Independent predictors of failure up to 7.5 years after 35 386 single-brand cementless total hip replacements: a retrospective cohort study using National Joint Registry data. Bone Jt J 95 B:747–757. doi:10.1302/0301-620X.95B6.31378 CrossRef Jameson SS, Baker PN, Mason J et al (2013) Independent predictors of failure up to 7.5 years after 35 386 single-brand cementless total hip replacements: a retrospective cohort study using National Joint Registry data. Bone Jt J 95 B:747–757. doi:10.​1302/​0301-620X.​95B6.​31378 CrossRef
13.
17.
Zurück zum Zitat Engh CA, O’Connor D, Jasty M, et al (1992) Quantification of implant micromotion, strain shielding, and bone resorption with porous-coated anatomic medullary locking femoral prostheses. Clin Orthop Relat Res 285:13–29 Engh CA, O’Connor D, Jasty M, et al (1992) Quantification of implant micromotion, strain shielding, and bone resorption with porous-coated anatomic medullary locking femoral prostheses. Clin Orthop Relat Res 285:13–29
18.
Zurück zum Zitat Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113 Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113
22.
Zurück zum Zitat Kang Q, An YH, Friedman RJ (1997) Effects of multiple freezing-thawing cycles on ultimate indentation load and stiffness of bovine cancellous bone. Am J Vet Res 58:1171–1173PubMed Kang Q, An YH, Friedman RJ (1997) Effects of multiple freezing-thawing cycles on ultimate indentation load and stiffness of bovine cancellous bone. Am J Vet Res 58:1171–1173PubMed
Metadaten
Titel
Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography
verfasst von
Valérie Malfroy Camine
Hannes A. Rüdiger
Dominique P. Pioletti
Alexandre Terrier
Publikationsdatum
07.06.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 1/2018
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-017-3524-0

Weitere Artikel der Ausgabe 1/2018

International Orthopaedics 1/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Lever-Sign-Test hilft beim Verdacht auf Kreuzbandriss

15.05.2024 Vordere Kreuzbandruptur Nachrichten

Mit dem Hebelzeichen-Test lässt sich offenbar recht zuverlässig feststellen, ob ein vorderes Kreuzband gerissen ist. In einer Metaanalyse war die Vorhersagekraft vor allem bei positivem Testergebnis hoch.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.