Skip to main content
Erschienen in: BMC Ophthalmology 1/2016

Open Access 01.12.2016 | Research article

Changes in intraocular pressure values measured with noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology tonometer (CST) in the early phase after small incision lenticule extraction (SMILE)

verfasst von: Yang Shen, Xiangjian Su, Xiu Liu, Huamao Miao, Xuejun Fang, Xingtao Zhou

Erschienen in: BMC Ophthalmology | Ausgabe 1/2016

Abstract

Background

Corneal biomechanical properties are always compromised after corneal refractive surgeries thus leading to underestimated intraocular pressure (IOP) that complicates the management of IOP. We investigated the changes in postoperative baseline of IOP values measured with noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology (CST) in the early phase after small incision lenticule extraction (SMILE).

Methods

Twenty-two eyes (−6.76 ± 1.39D) of 22 moderate and high myopes, (28.36 ± 7.14 years, 12 male and 10 female) were involved in this prospective study. IOP values were measured using a non-contact tomometer (NCT-IOP), an ocular response analyzer (corneal-compensated IOP, IOPcc and Goldmann-correlated IOP, IOPg) and a Corvis scheimpflug technology tonometer (CST-IOP) preoperatively, at 20 min and 24 h, postoperatively. Repeated measures analysis of variance (RM-ANOVA), Pearson’s correlation analysis and multiple linear regression models (stepwise) were performed. Cut-off P values were 0.05.

Results

Except for IOPcc, NCT-IOP, IOPg, and CST-IOP values significantly decreased after SMILE procedure (All P values <0.05). ΔCCT, as well as ΔMRSE and ΔKm, did not significantly correlated with ΔNCT-IOP, ΔIOPcc, ΔIOPg or ΔCST-IOP, (all P values >0.05). Multiple linear regression models (stepwise) showed that the practical post-operative IOP value was the main predictor of the theoretical post-operative NCT-IOP, IOPcc and IOPg values (all P values <0.001). The postoperative applanation time 1 (AT1) value (B = 8.079, t = 4.866, P < 0.001), preoperative central corneal thickness (CCT) value (B = 0.035, t = 2.732, P = 0.014) and postoperative peak distance (PD) value (B = 0.515, t = 2.176, P = 0.043) were the main predictors of the theoretical post-operative CST-IOP value.

Conclusions

IOP values are underestimated when assessed after SMILE by using NCT-IOP, IOPg and CST-IOP. The practical postoperative IOPcc value and theoretical post-operative CST-IOP value may be more preferable for IOP assessment in the early phase after SMILE.

Trial registration

Current Controlled Trials ChiCTRONRC130031​14. Retrospectively registered 17 March 2013
Abkürzungen
AL1
Applination length (time 1) is the cord length recorded at the moment that a cornea just deformed for the first time
AL2
Applination length (time 2) is the cord length recorded at the moment that a corneal deformation just completely resumed
AT1
Applination Time (time 1) is the duration from the initiation to the moment that a cornea just deformed for the first time
AT2
Applination Time (time 2) is the duration from the initiation to the moment that a corneal deformation just completely resumed
ATh
Applination Time (hightest concavity) is the duration from the initiation to the moment that a cornea is depressed to the highest concavity
AV1
Applination velocity (time 1) is the instantaneous velocity recorded when a cornea just deformed
AV2
Applination velocity (time 2): the instantaneous velocity recorded when a cornea just completely resumed
DA
Deformation Amplitude is the maximum amplitude of corneal deformation recorded when a cornea is depressed to the highest concavity
PD
Peak distance is the distance between the two corneal peaks recorded when a cornea is depressed to the highest concavity
Radius
Is the corneal radius obtained when a cornea is depressed to the highest concavity

Background

With the booming popularity of refractive surgeries, nowadays, intraocular pressure (IOP) management is not only essential for patients with glaucoma but also for those myopes who underwent refractive surgeries as long-term use of topical steroid may cause steroid-induced ocular hypertension, and primary open-angle glaucoma is a common complication accompanying myopia [1, 2]. However, corneal refractive surgeries remove corneal tissue, modify corneal shape and compromise corneal biomechanical properties thus leading to underestimated IOP values [3] and obscure the diagnosis of ocular hypertension. Noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology tonometer (CST) are three most commonly employed instruments for clinical IOP assessment. Goldmann applanation tonometer is the gold standard method for IOP assessment; nonetheless, its accuracy still depends on central corneal thickness (CCT), anterior corneal curvature and other potential factors that may affect corneal biomechanical properties [46]. The permanent corneal flap and the incomplete Bowman’s layer, caused by laser-assisted in situ keratomileusis (LASIK) or surface ablation techniques, are another two major factors that weaken corneal stiffness and affect postoperative IOP assessment [7, 8].
Previously, large quantities of formulas have been raised to correct IOP values following corneal refractive surgeries [911]. However, various confounding factors including surgical designs (i.e., flap thickness, residual stromal bed thickness, optic zone diameter and ablation depth), individual differences (i.e., age, gender, race, refractions, corneal curvature corneal hydration and post-operative wound healing response) [12] and long-term postoperative topical steroid usage make these formulas widely divergent.
Femtosecond laser small incision lenticule extraction (SMILE) is a flapless and minimally invasive corneal refractive surgery [3, 13]. With a refractive stromal lenticule extracted from a 2 mm-long side-cut, the integrity of corneal structure (including the Bowman’s layer) and corneal biomechanical properties are maximally maintained [14]. Technically, SMILE procedure only modifies corneal curvature and corneal thickness.
To minimize the interference of those confounding factors, in the present study, IOP values were obtained preoperatively, at 20 min postoperatively and at 24 h post-operatively by using the three frequently employed noncontact tonometers. We hypothesis the theoretical post-operative IOP values should be similar with the preoperative values. The gap between preoperative and postoperative IOP values should be dominantly caused by the surgery itself. As IOP values are always underestimated following corneal refractive surgeries [15], to investigate and establish a statistical model for compensating the gap between the pre-operative and the postoperative IOP values by involving corneal biomechanical parameters should be meaningful for clinical IOP management.

Methods

This prospective study was registered in Chinese Clinical Trial Registry (Trial registration: Current Controlled Trials ChiCTRONRC13003114. Retrospectively registered 17 March 2013), approved by ethics committee of Eye and ENT Hospital, Fudan University and was conducted with due regard to the tenets of the Declaration of Helsinki. Written informed consent was obtained from the participants after explanation of procedure was given.

Participants

Twenty-two moderate and high myopes (28.4 ± 7.1 years, 12 male and 10 female) were recruited in this prospective study at the Department of Ophthalmology, Eye and ENT Hospital, Fudan University. All the right eyes (−6.76 ± 1.39D) were analyzed.

Pre-operative examinations

Each participant underwent routine preoperative ophthalmologic examinations, including uncorrected distance visual acuity (UDVA), manifest refraction, best-corrected distance visual acuity (BDVA), slit lamp examination and fundus examination. Corneal topography was measured with a three-dimensional anterior segment analyzer (Pentacam HR, Typ70900, Oculus Optikgeräte GmbH, Wetzlar, Germany). The mean anterior corneal curvature (Km) and CCT were recorded. Preoperative IOP was measured with a noncontact tonometer (NCT, TX-20 Full Auto Tonometers, Canon, Japan), an ocular response analyzer (ORA, Reichert Inc, Depew, New York, USA) and a corvis scheimpflug technology tonometer (CST, Oculus Optikgeräte GmbH, Wetzlar, Germany). The sequence of measurement using these three instruments was arranged randomly. All the IOP measurements were obtained in a sitting position. NCT continuously obtained three valid readings and the mean value was calculated automatically. ORA provided values of corneal-compensated IOP (IOPcc), Goldmann-correlated IOP (IOPg), corneal hysteresis (CH) and corneal resistance factor (CRF). Four measurements were obtained in each eye and the one with the highest waveform score (WS) was recorded. CST-IOP and corneal deformation parameters were measured using CST. The measurement with an “OK” in Quality Specification (QS) was recorded [13].

Surgical technique

SMILE procedures were performed under topical anaesthesia using three drops of 0.4% Oxybuprocaine Hydrochloride (Santen Pharmaceutical Co., Ltd., Japan). One surgeon (ZXT) performed all the procedures with the VisuMax femtosecond laser system (Carl Zeiss Meditec AG, Germany). The intended thickness of the upper tissue arcade was set 120 μm, and its diameter was 7.5 mm. The diameter of the lenticule was 6.7 mm. The side cuts were set 90° apart at a width of 2 mm. The refractive lenticule of the intra-stromal corneal tissue was extracted through the side-cut opening using a modified serrated McPherson forceps (Geuder, GmbH, Heidelberg, Germany) [3].

Post-operative examinations and topical eye drops usage

UDVA, manifest refraction, BDVA, slit lamp examination, NCT-IOP, IOPcc, IOPg, CST-IOP were measured again at postoperative 20 min and postoperative 24 h. Corneal topography was measured again at 24 h postoperatively. Topical steroid (fluorometholone 0.1%; Santen Pharmaceutical Co., Ltd.), topical antibiotics (ofloxacin ophthalmic solutions 0.5%; Santen Pharmaceutical Co., Ltd.) and artificial tear (hypromellose 2910, dextran 70, glycerol eye drops; Alcon Laboratories, Inc., Fort Worth, TX) were used at 3-h intervals after SMILE procedure. At the first day post-operatively, topical steroid, topical antibiotics and artificial tear were employed for 4 times per day [13].

Data analysis and statistical evaluation

Statistical analysis was performed using SPSS 19 (SPSS Inc., IBM, USA). All the data were tested for normality using the Kolmogorov-Smirnov test. Repeated measures analysis of variance (RM-ANOVA) with LSD post hoc comparisons was performed to evaluate the changes in NCT-IOP, IOPcc, IOPg and CST-IOP over time. Pearson’s correlation analysis was applied to detect the potential correlations between these variables. Stepwise multiple linear regression model analysis was performed to predict theoretical post-operative IOP values. To minimize the effect of corneal epithelium edema, the IOP value obtained at 20 min post-operatively was excluded from the analysis. Cut-off P values were 0.05.

Results

All surgical procedures were successful and uneventful. The stromal layers of all the corneas were clear. Although mild epithelial edema could be observed in some cases by using a silt-lamp bio-microscopy at 20 min postoperatively but this sign disappeared at 24 h postoperatively. The main demographic and topographic data were shown in the Table 1.
Table 1
The main demographic and topographic data (n = 22)
Variables
Pre-operative
Postoperative 24 h
F valuea
P value
Mean ± SD
Range
Mean ± SD
Range
Age (year)
28.4 ± 7.1
18 to 42
28.4 ± 7.1
18 to 42
MRSE (D)
−6.76 ± 1.39
−10.00 to −4.13
−0.22 ± 0.48
−1.38 to 0.75
508.580
<0.001b
K1 (D)
43.23 ± 1.40
40.1 to 45.2
38.35 ± 2.12
32.3 to 40.9
510.718
<0.001b
K2 (D)
44.56 ± 1.84
41.0 to 47.6
39.03 ± 2.23
32.7 to 42.3
557.309
<0.001b
Km (D)
43.88 ± 1.54
40.6 to 46.4
38.69 ± 2.15
32.5 to 41.6
622.613
<0.001b
Pentacam-CCT (μm)
546.6 ± 23.4
515 to 592
431.9 ± 28.0
399 to 501
521.421
<0.001b
MRSE manifest refraction spherical equivalent, D diopter, K1 flat curvature power, K2 flat curvature power, Km mean curvature power, CCTcentral corneal thickness
aRepeated measures analysis of variance (RM-ANOVA)
bSignificant difference was detected

IOP measurements

As demonstrated in the Table 2, the measurements of NCT-IOP, IOPg, and CST-IOP significantly decreased after SMILE procedure (All P values <0.05). Fisher’s least significant difference (LSD) post hoc comparisons (Fig. 1) revealed that at 20 min postoperatively, the mean values of NCT-IOP (post hoc P < 0.001), IOPg (post hoc P < 0.001) and CST-IOP (post hoc P < 0.001) were all decreased dramatically when compared with the pre-operative values, but all of the values kept stable in the next 24 h (post hoc P = 0.365, post hoc P = 0.050 and post hoc P = 0.585, respectively). The mean values of IOPcc slightly increased at the 20-min mark (post hoc P = 0.056), but then deceased significantly at the 24-h mark (P = 0.028) however, no significant difference was detected between the IOPcc values obtained before and 24 h after SMILE (post hoc P = 0.051).
Table 2
Changes in IOP measurements (n = 22)
Variables
Pre-operative
Postoperative 20 min
Postoperative 24 h
F valuea
P value
Mean ± SD
Range
Mean ± SD
Range
Mean ± SD
Range
NCT-IOP (mmHg)
15.04 ± 2.91
11.1 to 20.9
9.19 ± 2.52
4.6 to 15.2
9.70 ± 2.95
5.8 to 17.2
81.173
<0.001b
IOPcc (mmHg)
14.99 ± 2.46
10.2 to 19.0
16.65 ± 4.29
12.1 to 29.4
14.01 ± 3.10
6.8 to 20.3
4.975
0.030b
IOPg (mmHg)
15.10 ± 3.21
9.3 to 21.9
11.20 ± 3.38
4.7 to 18.2
9.52 ± 2.85
4.5 to 15.9
40.128
<0.001b
CST-IOP (mmHg)
17.74 ± 2.16
13.0 to 23.0
14.29 ± 2.09
10.0 to 17.5
14.57 ± 1.58
10.5 to 18.0
31.886
<0.001b
IOPcc corneal-compensated IOP, IOPg Goldmann-correlated IOP
aRepeated measures analysis of variance (RM-ANOVA)
bSignificant difference was detected over time

Changes in corneal biomechanical and deformation parameters

The definition of each deformation parameter was listed in the Table 3 [3, 13, 16]. As shown in the Table 4, the mean values of corneal hysteresis (CH), corneal resistance factor (CRF), AT1, AV1, AV2, Radius, DA all changed significantly after SMILE procedure (All P values <0.001). CH (post hoc P < 0.001), CRF (post hoc P < 0.001), AT1 (post hoc P < 0.001), AV1 (post hoc P = 0.007), AV2 (post hoc P = 0.016) and Radius (post hoc P = 0.001) decreased but DA increased remarkably (post hoc P < 0.001) at 20 min postoperatively. While the remaining parameters in the Table 4 all kept unchanged (All P values >0.05). Expect for the mean value of CH, which kept decreasing at 24 h postoperatively (post hoc P = 0.025), CRF, AT1, AV1, AV2, Radius and DA all remained stable at 24-h mark (All post hoc P values >0.05).
Table 3
Abbreviations for corneal deformation parameters
Abbreviation
Definition
AT1
Applination Time (time 1): the duration from the initiation to the moment that a cornea just deformed for the first time
ATh
Applination Time (hightest concavity): the duration from the initiation to the moment that a cornea is depressed to the highest concavity
AT2
Applination Time (time 2): the duration from the initiation to the moment that a corneal deformation just completely resumed
AL1
Applination length (time 1): the cord length recorded at the moment that a cornea just deformed for the first time
AL2
Applination length (time 2): the cord length recorded at the moment that a corneal deformation just completely resumed
AV1
Applination velocity (time 1): the instantaneous velocity recorded when a cornea just deformed
AV2
Applination velocity (time 2): the instantaneous velocity recorded when a cornea just completely resumed
Radius
Radius: the corneal radius obtained when a cornea is depressed to the highest concavity
PD
Peak Distance: the distance between the two corneal peaks recorded when a cornea is depressed to the highest concavity
DA
Deformation Amplitude: the maximum amplitude of corneal deformation recorded when a cornea is depressed to the highest concavity
Table 4
Changes in corneal biomechanical and deformation parameters (n = 22)
Variables
Pre-operative
Postoperative 20 min
Postoperative 24 h
F valuea
P value
Mean ± SD
Range
Mean ± SD
Range
Mean ± SD
Range
  
CH (mmHg)
11.05 ± 1.81
8.7 to 16.7
6.82 ± 0.98
5.3 to 8.4
7.49 ± 1.32
5.6 to 11.0
90.283
<0.001b
CRF (mmHg)
10.84 ± 2.20
7.1 to 17.3
6.07 ± 1.26
2.9 to 9.3
6.09 ± 1.27
4.1 to 9.1
131.239
<0.001b
CST-CCT (μm)
545.8 ± 24.8
515.0 to 618.0
469.7 ± 46.4
389 to 580
440.9 ± 29.5
407 to 522
136.093
<0.001b
AT1 (ms)
7.28 ± 0.26
6.71 to 7.97
6.91 ± 0.23
6.52 to 7.32
6.90 ± 0.18
6.54 to 7.37
36.484
<0.001b
ATh (ms)
16.94 ± 0.53
16.17 to 17.79
16.77 ± 0.86
14.32 to 18.02
16.94 ± 0.63
15.71 to 18.02
0.465
0.631
AT2 (ms)
21.64 ± 1.29
16.89 to 22.82
22.13 ± 0.41
21.36 to 22.86
22.17 ± 0.29
21.61 to 22.87
3.042
0.089
AL1 (mm)
1.78 ± 0.04
1.71 to 1.87
1.72 ± 0.17
1.26 to 1.92
1.74 ± 0.20
1.19 to 1.94
0.798
0.429
AL2 (mm)
1.49 ± 0.39
0.44 to 2.00
1.35 ± 0.44
0.73 to 2.00
1.22 ± 0.41
0.73 to 1.89
2.281
0.131
AV1 (m/s)
0.16 ± 0.02
0.09 to 0.22
0.14 ± 0.02
0.08 to 0.19
0.15 ± 0.02
0.11 to 0.20
5.528
0.007b
AV2 (m/s)
−0.44 ± 0.14
−0.92 to −0.20
−0.56 ± 0.13
−0.75 to −0.33
−0.61 ± 0.16
−1.11 to −0.33
8.910
0.001b
PD (mm)
4.27 ± 1.19
2.36 to 5.43
4.51 ± 1.25
2.62 to 6.19
4.53 ± 1.27
2.57 to 6.23
0.355
0.704
Radius (mm)
6.97 ± 1.46
1.66 to 9.45
5.51 ± 1.00
1.78 to 6.88
5.69 ± 0.54
4.89 to 7.05
13.208
<0.001b
DA (mm)
1.09 ± 0.09
0.89 to 1.37
1.17 ± 0.10
1.03 to 1.43
1.14 ± 0.07
0.99 to 1.34
10.100
<0.001b
MRSE manifest refraction spherical equivalent, D diopter, CH corneal hysteresis, CRF corneal resistance factor, CCT central corneal thickness, AT1 applanation time 1, ATh applanation time at the highest concavity; AT2 applanation time 2, AL1 applanation length 1; AL2 applanation length 2, AV1 applanation velocity 1, AV2 applanation velocity 2, PD peak distance, DA deformation amplitude
aRepeated measures analysis of variance (RM-ANOVA)
bSignificant difference was detected over time

Correlations

Pearson’s correlation analysis showed that neither ΔNCT-IOP, ΔIOPcc, ΔIOPg nor ΔCST-IOP (“Δ” refers to the difference between the value obtained preoperatively and at 24 h postoperatively) significantly correlated with ΔCCT, ΔMRSE or ΔKm (all P values >0.05) But the post-operative CCT value measured with CST (CST-CCT) significantly correlated with CH (R = 0.511, P = 0.015) and CRF (R = 0.674, P = 0.001).

Stepwise multiple linear regression models

The Table 5 demonstrated the statistically significant stepwise linear multiple regression models for predicting the theoretical postoperative values of NCT-IOP (Adjusted R2 = 0.400, F = 15.000, P = 0.001), IOPcc (Adjusted R2 = 0.472, F = 19.755, P < 0.001), IOPg (Adjusted R2 = 0.542, F = 25.850, P < 0.001) and CST-IOP (Adjusted R2 = 0.596, F = 11.336, P < 0.001).
Table 5
The stepwise multiple linear regression models for predicting theoretical post-op IOP values (n = 22)
Dependent Variables
Main Predictors
Ba
SEb
tc
Sig.
βd
Regression Equation
Adjusted R2
Fe
Sig.
Pre-op NCT-IOP
Practical postop
NCT-IOP
0.646
0.167
3.873
<0.001
0.655
Theoretical post-op NCT-IOP (mmHg) = 0.646 × Practical post-op NCT-IOP(mmHg) + 8.774(mmHg)
0.400
15.000
<0.001
 
Constant
8.774
1.688
5.198
0.001
-
Pre-op ORAIOPcc
Practical postop
IOPcc
0.558
0.126
4.445
<0.001
0.705
Theoretical post-op IOPcc (mmHg) = 0.558 × practical postoperative IOPcc (mmHg) + 7.166(mmHg)
0.472
19.755
<0.001
 
Constant
7.166
1.800
3.981
0.001
-
Pre-op ORAIOPg
Practical postopIOPg
0.848
0.167
5.084
<0.001
0.751
Theoretical post-op IOPg (mmHg) = 0.848 × practical postoperative IOPg (mmHg) + 7.024 (mmHg)
0.542
25.850
<0.001
 
Constant
7.024
1.654
4.247
<0.001
-
Pre-op CST-IOP
Post-op AT1
8.079
1.660
4.866
<0.001
0.680
Theoretical post-op CST-IOP
(mmHg) = 8.079 × post-op AT1 (ms) + 0.035 × pre-op
CCT(μm) + 0.515 × post-op
PD (mm) - 59.47 mmHg
0.596
11.336
<0.001
 
Pre-op CCT
0.035
0.013
2.732
0.014
0.379
 
Post-op PD
0.515
0.237
2.176
0.043
0.304
 
Constant
−59.47
13.524
−4.397
<0.001
-
Pre-op pre-operative, Post-op post-operative, AT1 applanation time 1, CCT central corneal thickness, PDpeak distance, Sig. Significance
aUnstandardized Coefficients
bStandard Error of Unstandardized Coefficients
cUnstandardized Coefficients/Standard Error
dStandardized Coefficients (Beta)
eMultiple Linear Regression Model (Stepwise)
The practical post-operative NCT-IOP value was the main predictor of the theoretical post-operative NCT-IOP value (B = 0.646,t = 3.873,P = 0.001), but the theoretical post-operative NCT-IOP value did not associated with age, preoperative CCT, postoperative CCT, corneal curvature or ΔMRSE (All P values > 0.05).
For IOPcc and IOPg, the practical post-operative IOPcc and IOPg values were the main predictors (B = 0.558, t = 4.445, P < 0.001; B = 0.848, t = 5.084, P < 0.001, respectively) for predicting theoretical post-operative IOPcc and IOPg value. But both the theoretical post-operative IOPcc and IOPg values did not associated with age, preoperative CCT, postoperative CCT, postoperative CH, postoperative CRF, corneal curvature or ΔMRSE (All P values > 0.05).
For CST-IOP, the postoperative AT1 value (B = 8.079, t = 4.866, P < 0.001), preoperative CCT value (B = 0.035, t = 2.732, P = 0.014) and postoperative PD value (B = 0.515, t = 2.176, P = 0.043) were the main predictors of the theoretical post-operative CST-IOP value. But the theoretical post-operative CST-IOP value did not associated with age, postoperative CCT, corneal curvature, ΔMRSE or other postoperative CST parameters (All P values >0.05).

Discussion

Corneal refractive surgeries compromise corneal mechanical strength, thus leading to underestimated IOP, which may obscure the diagnosis of steroid-induced glaucoma or open-angle glaucoma during postoperative follow-ups [17, 18].
In the present study, the mean values of NCT-IOP, IOPcc, IOPg and CST-IOP significantly deceased at 24 h postoperatively. Post-operative CCT significantly correlated with CH and CRF, but ΔCCT, as well as ΔMRSE and ΔKm, did not correlated with ΔNCT-IOP, ΔIOPcc, ΔIOPg or ΔCST-IOP, indicating that ΔCCT, ΔMRSE and corneal curvature may not be the main factors affecting IOP measurements. CH and CRF reflect the characters of force absorption and the resistance of the cornea, respectively. They are both dependent to corneal thickness [19]. Vestergaard AH, et al. [20] reported CH and CRF still significantly correlated with CCT after SMILE procedure, while Shah S [15] reported that ΔCH and ΔCRF did not correlated with ΔCCT after corneal refractive surgeries. A possible explanation is that comparing with the removed corneal tissue, the residual cornea contributed much more effect in maintaining the entire corneal biomechanical strength.
The stepwise multiple linear regression model analysis revealed that the practical post-operative IOP values obtained using NCT and ORA were the main predictors of the theoretical post-operative IOP values (Pre-operative IOP values). The models explained 40.0, 47.2 and 54.2% of the variance assessed using NCT-IOP, IOPcc and IOPg respectively. For CST, the postoperative AT1 value, preoperative CCT value and postoperative PD value were the main predictors of the theoretical post-operative CST-IOP value. CST-IOP model explained 59.6% of the variance.
We found it is interesting that the change in CCT seems to be independent of the change in IOP value after SMILE procedure as neither ΔCCT value was correlated with ΔNCT-IOP, ΔIOPcc, ΔIOPg or ΔCST-IOP nor CCT was a predictor of theoretical post-operative IOP value. So far, various instruments including NCT [21], ORA [22], CST, Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) [23] have been employed to investigate the changes in IOP values after LASIK. Cheng AC, et al. [21] reported the preoperative NCT-IOP significantly correlated with the postoperative one after LASIK. Moreover, postoperative CCT and ablation depth were included into their model, which was different from ours. Chen S and his colleagues [22] studied the changes in ORA parameters after myopic LASIK. They found that ΔIOPg was positively correlated with ΔCCT, but ΔIOPcc was not. In addition, they found ablation depth was correlated with ΔCRF and ΔCH. It is also reported that CST and GAT would underestimate IOP following LASIK procedure [24]. The underlying reasons for these discrepancies may be the difference in surgical techniques and their efforts on corneal biomechanical structure. During LASIK procedure, stromal flap creation cuts almost all the collagen fibers in the anterior stroma of a cornea and compromised the integrality of the Bowman’s layer, moreover, excimer laser ablated the stromal tissue, which contributes most of the biomechanics of the entire cornea. So the residual stromal bed undertakes the main role to maintain the corneal shape and biomechanical stability, in addition, the more corneal tissue removed, the lower the postoperative IOP value will be. However, SMILE procedure neither creates a stromal flap nor ablates Bowman’s layer and the anterior stroma. As the corneal collagen fibers in the anterior stromal layer is much compact than that in the middle or posterior layer [25], most of the corneal biomechanical properties are remained. Reinstein DZ, et al. [14] developed a mathematical model to estimate the postoperative stromal tensile strength following SMILE, PRK and LASIK. They found that SMILE lenticule thickness could be approximately 100 μm greater than the LASIK ablation depth, which is equivalent to approximately 7.75 diopters, and still have equivalent corneal strength. Wang D, et al. [26] found an interesting phenomenon that in myopia of −6.00D or less, the ΔCH and the ΔCRF value between SMILE and LASIK were not significant. But in myopia greater than −6.00D, CH and CRF decreased more in LASIK than in SMILE. Another reason is that the NCT, ORA and CST may not be sensitive enough to detect the linear correlations between ΔIOP and ΔCCT as the residual cornea following SMILE contributes much more biochemical stability than the removed leuticule. We noticed that although the IOPcc values obtained at 20-min mark increased slightly but returned to pre-operative level at 24 h postoperatively, indicating corneal suction and surgical operation might also affect the postoperative IOP assessment [27]. As IOPcc is adjusted to account for CH, which is significantly correlated with CCT value, the IOP gap can be compensated [28, 29]. Osman IM [16] reported similar findings, indicating that practical post-operative IOPcc value may be used to assess IOP after SMILE procedure.
The limitations of the present study are as follows. As the Goldmann applanation tonometer (GAT) IOP measurement has potential risk of infection, additionally, fluorescein sodium eye drops may infiltrate into the interlayer when GAT is employed for IOP assessment during the early post-operative phase, we were not able to obtain the GAT-IOP values. But the purpose of the study is to investigate the changes in IOP before and after SMILE with the same tonometer rather than to investigate the interrelations among the IOP values obtain with different instruments, GAT is not indispensable in the present study. As the biomechanical properties of the corneas underwent SMILE procedure may fluctuate during long-term follow-up, further studies are required to validate if the models were still effective in long-term outcomes of SMILE.

Conclusions

IOP values were underestimated when assessed by using NCT-IOP, IOPg and CST-IOPg after SMILE procedure. Both practical postoperative IOPcc values and theoretical post-operative CST-IOP values may be more preferable for IOP assessment or management following SMILE procedure.

Acknowledgments

Not applicable.

Funding

(1) National Natural Science Foundation of China (Grant No.81570879)
(2) Outstanding Academic Leaders Program of Shanghai Health System (Grant No. XBR2013098).
(3) Liaoning Province Science and Technology Project (Grant No.2013225303)
The funders had no role in study design, data collection and analysis, decision to publish or preparation of the manuscript.

Availability of data and materials

Available upon request from the co-first authors Yang Shen and Xiangjian Su.

Authors’ contributions

YS and XJS were the major contributors in writing the manuscript; YS, XJS and XL collected the data. HMM analyzed and interpreted the data; XTZ and XJF designed the study; XTZ operated the surgeries. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.
Not applicable.
This prospective study was registered in Chinese Clinical Trial Registry (Trial registration: Current Controlled Trials ChiCTRONRC13003114. Retrospectively registered 17 March 2013) website: http://​www.​chictr.​org.​cn/​showproj.​aspx?​proj=​6445), approved by ethics committee of Eye and ENT Hospital, Fudan University and was conducted with due regard to the tenets of the Declaration of Helsinki. Written informed consent was obtained from the participants after explanation of procedure was given.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Levinger E, Slomovic A, Bahar I, Slomovic AR. Diagnosis of steroid-induced elevated intraocular pressure and associated lamellar keratitis after laser in situ keratomileusis using optical coherence tomography. J Cataract Refract Surg. 2009;35:386–8.CrossRefPubMed Levinger E, Slomovic A, Bahar I, Slomovic AR. Diagnosis of steroid-induced elevated intraocular pressure and associated lamellar keratitis after laser in situ keratomileusis using optical coherence tomography. J Cataract Refract Surg. 2009;35:386–8.CrossRefPubMed
2.
Zurück zum Zitat Worley A, Grimmer-Somers K. Risk factors for glaucoma: what do they really mean? Aust J Prim Health. 2011;17:233–9.CrossRefPubMed Worley A, Grimmer-Somers K. Risk factors for glaucoma: what do they really mean? Aust J Prim Health. 2011;17:233–9.CrossRefPubMed
3.
Zurück zum Zitat Shen Y, Zhao J, Yao P, Miao H, Niu L, Wang X, Zhou X. Changes in corneal deformation parameters after lenticule creation and extraction during small incision lenticule extraction (SMILE) procedure. PLoS One. 2014;9:e103893.CrossRefPubMedPubMedCentral Shen Y, Zhao J, Yao P, Miao H, Niu L, Wang X, Zhou X. Changes in corneal deformation parameters after lenticule creation and extraction during small incision lenticule extraction (SMILE) procedure. PLoS One. 2014;9:e103893.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Kim NR, Kim CY, Kim H, Seong GJ, Lee ES. Comparison of goldmann applanation tonometer, noncontact tonometer, and TonoPen XL for intraocular pressure measurement in different types of glaucomatous, ocular hypertensive, and normal eyes. Curr Eye Res. 2011;36:295–300.CrossRefPubMed Kim NR, Kim CY, Kim H, Seong GJ, Lee ES. Comparison of goldmann applanation tonometer, noncontact tonometer, and TonoPen XL for intraocular pressure measurement in different types of glaucomatous, ocular hypertensive, and normal eyes. Curr Eye Res. 2011;36:295–300.CrossRefPubMed
5.
Zurück zum Zitat Lee M, Ahn J. Effects of central corneal stromal thickness and epithelial thickness on intraocular pressure using goldmann applanation and non-contact tonometers. PLoS One. 2016;11:e0151868.CrossRefPubMedPubMedCentral Lee M, Ahn J. Effects of central corneal stromal thickness and epithelial thickness on intraocular pressure using goldmann applanation and non-contact tonometers. PLoS One. 2016;11:e0151868.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Kohlhaas M, Boehm AG, Spoerl E, Pürsten A, Grein HJ, Pillunat LE. Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol. 2006;124:471–6.CrossRefPubMed Kohlhaas M, Boehm AG, Spoerl E, Pürsten A, Grein HJ, Pillunat LE. Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol. 2006;124:471–6.CrossRefPubMed
7.
Zurück zum Zitat Schallhorn JM, Schallhorn SC, Ou Y. Factors that influence intraocular pressure changes after myopic and hyperopic LASIK and photorefractive keratectomy: a large population study. Ophthalmology. 2015;122:471–9.CrossRefPubMed Schallhorn JM, Schallhorn SC, Ou Y. Factors that influence intraocular pressure changes after myopic and hyperopic LASIK and photorefractive keratectomy: a large population study. Ophthalmology. 2015;122:471–9.CrossRefPubMed
8.
Zurück zum Zitat Uzbek AK, Kamburoglu G, Mahmoud AM, Roberts CJ. Change in biomechanical parameters after flap creation using the Intralase femtosecond laser andsubsequent excimer laser ablation. Curr Eye Res. 2011;36:614–9.CrossRefPubMed Uzbek AK, Kamburoglu G, Mahmoud AM, Roberts CJ. Change in biomechanical parameters after flap creation using the Intralase femtosecond laser andsubsequent excimer laser ablation. Curr Eye Res. 2011;36:614–9.CrossRefPubMed
9.
Zurück zum Zitat Silva TG, Polido JG, Pinheiro MV, Silva AL, Goldbach L, Mascaro VL, Serracarbassa PD, Araújo ME. Application of corrective formula for intraocular pressure changes in patients that underwent LASIK. Arq Bras Oftalmol. 2011;74:102–5.CrossRefPubMed Silva TG, Polido JG, Pinheiro MV, Silva AL, Goldbach L, Mascaro VL, Serracarbassa PD, Araújo ME. Application of corrective formula for intraocular pressure changes in patients that underwent LASIK. Arq Bras Oftalmol. 2011;74:102–5.CrossRefPubMed
10.
11.
Zurück zum Zitat Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS. Effect of corneal thickness on intraocular pressuren measurements with the pneumotonometer, Goldmann applanation tonometer and Tono-pen. Invest Ophthalmol Vis Sci. 2002;43:1389–92.PubMed Bhan A, Browning AC, Shah S, Hamilton R, Dave D, Dua HS. Effect of corneal thickness on intraocular pressuren measurements with the pneumotonometer, Goldmann applanation tonometer and Tono-pen. Invest Ophthalmol Vis Sci. 2002;43:1389–92.PubMed
12.
Zurück zum Zitat Hatami-Marbini H, Rahimi A. Interrelation of hydration, collagen cross-linking treatment, and biomechanical properties of the cornea. Curr Eye Res. 2015;30:1–7.CrossRef Hatami-Marbini H, Rahimi A. Interrelation of hydration, collagen cross-linking treatment, and biomechanical properties of the cornea. Curr Eye Res. 2015;30:1–7.CrossRef
13.
Zurück zum Zitat Shen Y, Chen Z, Knorz MC, Li M, Zhao J, Zhou X. Comparison of corneal deformation parameters after SMILE, LASEK, and femtosecond laser-assisted LASIK. J Refract Surg. 2014;30:310–8.CrossRefPubMed Shen Y, Chen Z, Knorz MC, Li M, Zhao J, Zhou X. Comparison of corneal deformation parameters after SMILE, LASEK, and femtosecond laser-assisted LASIK. J Refract Surg. 2014;30:310–8.CrossRefPubMed
14.
Zurück zum Zitat Reinstein DZ, Archer TJ, Randleman JB. Mathematical model to compare the relative tensile strength of the cornea after PRK, LASIK, and small incision lenticule extraction. J Refract Surg. 2013;29:454–60.CrossRefPubMed Reinstein DZ, Archer TJ, Randleman JB. Mathematical model to compare the relative tensile strength of the cornea after PRK, LASIK, and small incision lenticule extraction. J Refract Surg. 2013;29:454–60.CrossRefPubMed
15.
Zurück zum Zitat Shah S, Laiquzzaman M, Yeung I, Pan X, Roberts C. The use of the ocular response analyser to determine corneal hysteresis in eyes before and after excimer laser refractive surgery. Cont Lens Anterior Eye. 2009;32:123–8.CrossRefPubMed Shah S, Laiquzzaman M, Yeung I, Pan X, Roberts C. The use of the ocular response analyser to determine corneal hysteresis in eyes before and after excimer laser refractive surgery. Cont Lens Anterior Eye. 2009;32:123–8.CrossRefPubMed
16.
Zurück zum Zitat Osman IM, Helaly HA, Abdalla M, Shousha MA. Corneal biomechanical changes in eyes with small incision lenticule extraction and laser assisted in situ keratomileusis. BMC Ophthalmol. 2016;16:123.CrossRefPubMedPubMedCentral Osman IM, Helaly HA, Abdalla M, Shousha MA. Corneal biomechanical changes in eyes with small incision lenticule extraction and laser assisted in situ keratomileusis. BMC Ophthalmol. 2016;16:123.CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Osman E. Laser refractive surgery in glaucoma patients. Saudi J Ophthalmol. 2011;25:169–73.CrossRefPubMed Osman E. Laser refractive surgery in glaucoma patients. Saudi J Ophthalmol. 2011;25:169–73.CrossRefPubMed
18.
Zurück zum Zitat Shrivastava A, Madu A, Schultz J. Refractive surgery and the glaucoma patient. Curr Opin Ophthalmol. 2011;22:215–21.CrossRefPubMed Shrivastava A, Madu A, Schultz J. Refractive surgery and the glaucoma patient. Curr Opin Ophthalmol. 2011;22:215–21.CrossRefPubMed
19.
Zurück zum Zitat Detry-Morel M, Jamart J, Pourjavan S. Evaluation of corneal biomechanical properties with the Reichert ocular response analyzer. Eur J Ophthalmol. 2011;21:138–48.CrossRefPubMed Detry-Morel M, Jamart J, Pourjavan S. Evaluation of corneal biomechanical properties with the Reichert ocular response analyzer. Eur J Ophthalmol. 2011;21:138–48.CrossRefPubMed
20.
Zurück zum Zitat Vestergaard AH, Grauslund J, Ivarsen AR, Hjortdal JØ. Central corneal sublayer pachymetry and biomechanical properties after refractive femtosecond lenticule extraction. J Refract Surg. 2014;30:102–8.CrossRefPubMed Vestergaard AH, Grauslund J, Ivarsen AR, Hjortdal JØ. Central corneal sublayer pachymetry and biomechanical properties after refractive femtosecond lenticule extraction. J Refract Surg. 2014;30:102–8.CrossRefPubMed
21.
Zurück zum Zitat Cheng AC, Fan D, Tang E, Lam DS. Effect of corneal curvature and corneal thickness on the assessment of intraocular pressure using noncontact tonometry in patients after myopic LASIK surgery. Cornea. 2006;25:26–8.CrossRefPubMed Cheng AC, Fan D, Tang E, Lam DS. Effect of corneal curvature and corneal thickness on the assessment of intraocular pressure using noncontact tonometry in patients after myopic LASIK surgery. Cornea. 2006;25:26–8.CrossRefPubMed
22.
Zurück zum Zitat Chen S, Chen D, Wang J, Lu F, Wang Q, Qu J. Changes in ocular response analyzer parameters after LASIK. J Refract Surg. 2010;26:279–88.CrossRefPubMed Chen S, Chen D, Wang J, Lu F, Wang Q, Qu J. Changes in ocular response analyzer parameters after LASIK. J Refract Surg. 2010;26:279–88.CrossRefPubMed
23.
Zurück zum Zitat Aristeidou AP, Labiris G, Katsanos A, Fanariotis M, Foudoulakis NC, Kozobolis VP. Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery. Graefes Arch Clin Exp Ophthalmol. 2011;249:767–73.CrossRefPubMed Aristeidou AP, Labiris G, Katsanos A, Fanariotis M, Foudoulakis NC, Kozobolis VP. Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery. Graefes Arch Clin Exp Ophthalmol. 2011;249:767–73.CrossRefPubMed
24.
Zurück zum Zitat Hong J, Yu Z, Jiang C, Zhou X, Liu Z, Sun X, Xu J. Corvis ST Tonometer for Measuring Postoperative IOP in LASIK Patients. Optom Vis Sci. 2015;92:589–95.CrossRefPubMed Hong J, Yu Z, Jiang C, Zhou X, Liu Z, Sun X, Xu J. Corvis ST Tonometer for Measuring Postoperative IOP in LASIK Patients. Optom Vis Sci. 2015;92:589–95.CrossRefPubMed
25.
Zurück zum Zitat Kohlhaas M, Spoerl E, Schilde T, Unger G, Wittig C, Pillunat LE. Biomechanical evidence of the distribution of cross-links in corneas treated with riboflavin and ultraviolet A light. J Cataract Refract Surg. 2006;32:279–83.CrossRefPubMed Kohlhaas M, Spoerl E, Schilde T, Unger G, Wittig C, Pillunat LE. Biomechanical evidence of the distribution of cross-links in corneas treated with riboflavin and ultraviolet A light. J Cataract Refract Surg. 2006;32:279–83.CrossRefPubMed
26.
Zurück zum Zitat Wang D, Liu M, Chen Y, Zhang X, Xu Y, Wang J, To CH, Liu Q. Differences in the corneal biomechanical changes after SMILE and LASIK. J Refract Surg. 2014;30:702–7.CrossRefPubMed Wang D, Liu M, Chen Y, Zhang X, Xu Y, Wang J, To CH, Liu Q. Differences in the corneal biomechanical changes after SMILE and LASIK. J Refract Surg. 2014;30:702–7.CrossRefPubMed
27.
Zurück zum Zitat Reinstein DZ, Archer TJ, Gobbe M. Small incision lenticule extraction (SMILE) history, fundamentals of a new refractive surgery technique and clinical outcomes. Eye Vis (Lond). 2014;16:1:3. Reinstein DZ, Archer TJ, Gobbe M. Small incision lenticule extraction (SMILE) history, fundamentals of a new refractive surgery technique and clinical outcomes. Eye Vis (Lond). 2014;16:1:3.
28.
Zurück zum Zitat Chou CY, Jordan CA, McGhee CN, Patel DV. Comparison of intraocular pressure measurement using 4 different instruments following penetrating keratoplasty. Am J Ophthalmol. 2012;153:412–8.CrossRefPubMed Chou CY, Jordan CA, McGhee CN, Patel DV. Comparison of intraocular pressure measurement using 4 different instruments following penetrating keratoplasty. Am J Ophthalmol. 2012;153:412–8.CrossRefPubMed
29.
Zurück zum Zitat Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31:156–62.CrossRefPubMed Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31:156–62.CrossRefPubMed
Metadaten
Titel
Changes in intraocular pressure values measured with noncontact tonometer (NCT), ocular response analyzer (ORA) and corvis scheimpflug technology tonometer (CST) in the early phase after small incision lenticule extraction (SMILE)
verfasst von
Yang Shen
Xiangjian Su
Xiu Liu
Huamao Miao
Xuejun Fang
Xingtao Zhou
Publikationsdatum
01.12.2016
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2016
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-016-0381-3

Weitere Artikel der Ausgabe 1/2016

BMC Ophthalmology 1/2016 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.