Clinical impact of different cut-off values in high-resolution manometry systems on diagnosing esophageal motility disorders

verfasst von: Shiko Kuribayashi, Katsuhiko Iwakiri, Tomohiro Shinozaki, Hiroko Hosaka, Akiyo Kawada, Noriyuki Kawami, Shintaro Hoshino, Nana Takenouchi, Yasuyuki Shimoyama, Osamu Kawamura, Motoyasu Kusano, Toshio Uraoka

Erschienen in: Journal of Gastroenterology | Ausgabe 12/2019

Einloggen, um Zugang zu erhalten

Abstract

Background

The values of the parameters in the Chicago classification measured by a high-resolution manometry (HRM) system with the Unisensor catheter (Starlet) are significantly different from those measured by the ManoScan. The contraction vigor is categorized by values of the distal contractile integral (DCI) in the Chicago classification v3.0; however, reference values of the DCI in the Starlet and the clinical impact of the different reference values in the Starlet and ManoScan on diagnosing esophageal motility disorders are not known.

Methods

We evaluated data from a previous report in which ManoScan and Starlet were compared in the same subjects. The DCI values in each system were compared and reference DCI values were calculated. Moreover, diagnoses assessed by Starlet using reference values in ManoScan were compared with those using calculated reference values and those assessed by ManoScan.

Results

There was a significant positive correlation between the DCI values measured by ManoScan and those measured by Starlet (r = 0.80, p < 0.01). Based on a linear functional relationship considering measurement errors, the reference DCI values for diagnosing failed, weak and hypercontractile contraction vigor were calculated as 590.6, 1011.3 and 10,085.8 mmHg-s-cm, respectively, in the Starlet. Therefore, the proposed reference values in the Starlet were 500, 1000 and 10,000 mmHg-s-cm, respectively. When the reference values in the ManoScan were used in the Starlet data, approximately 30% of subjects were diagnosed inappropriately. This issue was resolved using the proposed reference values in the Starlet.

Conclusion

Recognizing systemic differences in HRM systems is important.
Literatur
1.
Zurück zum Zitat Pandolfino JE, Ghosh SK, Rice J, et al. Classifying esophageal motility by pressure topography characteristics: a study of 400 patients and 75 controls. Am J Gastroenterol. 2008;103:27–37.CrossRef Pandolfino JE, Ghosh SK, Rice J, et al. Classifying esophageal motility by pressure topography characteristics: a study of 400 patients and 75 controls. Am J Gastroenterol. 2008;103:27–37.CrossRef
2.
Zurück zum Zitat Bredenoord AJ, Fox M, Kahrilas PJ, et al. Chicago classification criteria of esophageal motility disorders defined in high resolution esophageal pressure topography. Neurogastroenterol Motil. 2012;24(Suppl 1):57–655.CrossRef Bredenoord AJ, Fox M, Kahrilas PJ, et al. Chicago classification criteria of esophageal motility disorders defined in high resolution esophageal pressure topography. Neurogastroenterol Motil. 2012;24(Suppl 1):57–655.CrossRef
3.
Zurück zum Zitat Kahrilas PJ, Bredenoord AJ, Fox M, et al. The Chicago classification of esophageal motility disorders, v.30. Neurogastroenterol Motil. 2015;27:160–74.CrossRef Kahrilas PJ, Bredenoord AJ, Fox M, et al. The Chicago classification of esophageal motility disorders, v.30. Neurogastroenterol Motil. 2015;27:160–74.CrossRef
4.
Zurück zum Zitat Bogte A, Bredenoord AJ, Oors J, et al. Normal values for esophageal high-resolution manometry. Neurogastroenterol Motil. 2013;25:762-e579.CrossRef Bogte A, Bredenoord AJ, Oors J, et al. Normal values for esophageal high-resolution manometry. Neurogastroenterol Motil. 2013;25:762-e579.CrossRef
5.
Zurück zum Zitat do Carmo GC, Jafari J, Sifrim D, et al. Normal esophageal pressure topography metrics for data derived from the Sandhill-Unisensor high-resolution manometry assembly in supine and sitting positions. Neurogastroenterol Motil. 2015;27:285–92.CrossRef do Carmo GC, Jafari J, Sifrim D, et al. Normal esophageal pressure topography metrics for data derived from the Sandhill-Unisensor high-resolution manometry assembly in supine and sitting positions. Neurogastroenterol Motil. 2015;27:285–92.CrossRef
6.
Zurück zum Zitat Shi Y, Xiao Y, Peng S, et al. Normative data of high-resolution impedance manometry in the Chinese population. J Gastroenterol Hepatol. 2013;28:1611–5.CrossRef Shi Y, Xiao Y, Peng S, et al. Normative data of high-resolution impedance manometry in the Chinese population. J Gastroenterol Hepatol. 2013;28:1611–5.CrossRef
7.
Zurück zum Zitat Kuribayashi S, Iwakiri K, Kawada A, et al. Variant parameter values-as defined by the Chicago Criteria-produced by ManoScan and a new system with Unisensor catheter. Neurogastroenterol Motil. 2015;27:188–94.CrossRef Kuribayashi S, Iwakiri K, Kawada A, et al. Variant parameter values-as defined by the Chicago Criteria-produced by ManoScan and a new system with Unisensor catheter. Neurogastroenterol Motil. 2015;27:188–94.CrossRef
8.
Zurück zum Zitat Kusano M, Shimoyama Y, Sugimoto S, et al. Development and evaluation of FSSG: frequency scale for the symptoms of GERD. J Gastroenterol. 2004;39:888–91.CrossRef Kusano M, Shimoyama Y, Sugimoto S, et al. Development and evaluation of FSSG: frequency scale for the symptoms of GERD. J Gastroenterol. 2004;39:888–91.CrossRef
Metadaten
Titel
Clinical impact of different cut-off values in high-resolution manometry systems on diagnosing esophageal motility disorders
verfasst von
Shiko Kuribayashi
Katsuhiko Iwakiri
Tomohiro Shinozaki
Hiroko Hosaka
Akiyo Kawada
Noriyuki Kawami
Shintaro Hoshino
Nana Takenouchi
Yasuyuki Shimoyama
Osamu Kawamura
Motoyasu Kusano
Toshio Uraoka
Publikationsdatum
06.08.2019
Verlag
Springer Japan
Erschienen in
Journal of Gastroenterology / Ausgabe 12/2019
Print ISSN: 0944-1174
Elektronische ISSN: 1435-5922
DOI
https://doi.org/10.1007/s00535-019-01608-3

Weitere Artikel der Ausgabe 12/2019

Journal of Gastroenterology 12/2019 Zur Ausgabe

Original Article—Liver, Pancreas, and Biliary Tract

Familial pancreatic cancer risk: a population-based study in Utah

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.