Clinical outcomes after mix-and-match implantation of diffractive multifocal intraocular lenses with + 2.75 and + 4.00 diopter add powers

verfasst von: Jae Hyuck Lee, Hun Lee, Jin Ah. Lee, Aeri Yoo, Jae Yong Kim, Hungwon Tchah

Erschienen in: BMC Ophthalmology | Ausgabe 1/2020

Abstract

Background

To evaluate the clinical outcomes of bilateral mix-and-match implantation of diffractive multifocal intraocular lenses (IOLs) with different add powers.

Methods

We retrospectively reviewed the medical records of 18 patients who underwent bilateral mix-and-match implantation of diffractive multifocal IOLs with different add powers. Multifocal IOLs with add powers of + 2.75 diopters (D) and + 4.00 D were implanted into the patients’ dominant and nondominant eyes, respectively. At 1 and 3-month postoperatively, monocular and binocular visual acuity was measured using logMAR charts and manifest refraction was performed. Specifically, logMAR charts were used to measure uncorrected distance visual acuity (UDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA), and corrected distance visual acuity (CDVA). Defocus curves, contrast sensitivity, and patient satisfaction were assessed at 3-month postoperatively.

Results

Binocular logMAR measurements (mean ± standard deviation) at 3-month postoperatively were 0.01 ± 0.04 (UDVA), 0.16 ± 0.05 (UIVA), and 0.11 ± 0.07 (UNVA). Postoperative spherical equivalent was − 0.43 ± 0.35 D and − 0.39 ± 0.21 D in the dominant and nondominant eyes, respectively. Defocus curves showed significant differences between − 1.50 and − 4.00 D among binocular, dominant, and nondominant eye measurements, except between − 2.50 and − 3.00 D. Eyes implanted with + 2.75 and + 4.00 D IOLs showed good contrast sensitivity under photopic and mesopic conditions. Over 80% of patients reported high satisfaction with their near vision.

Conclusions

Bilateral mix-and-match implantation of diffractive multifocal IOLs with add powers of + 2.75 D and + 4.00 D showed good near, intermediate, and far vision.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
CDVA
Corrected distance visual acuity
CPD
Cycles per degree
D
Diopters
EDOF
Extended depth of focus
IOL
Intraocular lenses
logMAR
Logarithm of the minimal angle of resolution
MR
Manifest refraction
SE
Spherical equivalent
UDVA
Uncorrected distance visual acuity
UIVA
Uncorrected intermediate visual acuity
UNVA
Uncorrected near visual acuity

Background

Technological developments, including the development of multifocal intraocular lenses (IOLs), have resulted in maximization of vision quality via cataract surgery. Bifocal IOLs show improvements in near and far distance visual acuity and emphasize near visual acuity [1]. In general, bifocal IOLs of the same type and with the same add power for each eye are bilaterally implanted [2].
However, several studies reported insufficient intermediate visual acuity with bilateral bifocal IOLs [3, 4]. The need for better intermediate distance visual acuity has led to efforts to improve the uncorrected intermediate visual acuity (UIVA) of these IOLs. For example, multifocal IOLs with extended depth of focus (EDOF) and trifocal IOLs have been developed and widely commercialized [5, 6]. However, trifocal IOLs showed more prominent background shadows than did bifocal IOLs, and EDOF IOLs have are limited in their ability to improve uncorrected near visual acuity (UNVA) [7, 8]. Thus, there are still no IOLs available that improve vision across all ranges.
Since the CONCERTO study used monovision approach with multifocal IOL [9], to overcome the limitations of available IOLs, efforts have been made to expand the range of vision by implanting different types of multifocal IOLs in each eye of a subject [10]. Several studies have reported a good range of vision after bilateral mix-and-match implantation of multifocal IOLs [2, 11, 12]. However, studies that evaluate clinical outcomes after implantation of diffractive multifocal IOLs with different add powers are scarce. Therefore, we aimed to evaluate clinical outcomes after implantation of TECNIS® IOLs with near addition powers of + 2.75 diopters (D) in the dominant eye and + 4.00 D in the nondominant eye.

Methods

We conducted this retrospective observational case series with the approval of the Institutional Review Board of the Asan Medical Center and the University of Ulsan College of Medicine (Seoul, South Korea). The study adhered to the tenets of the Declaration of Helsinki and followed good clinical practice guidelines. All patients provided written informed consent to allow their medical information to be included for analysis and publication.
This retrospective study included all patients who underwent cataract surgery with bilateral mix-and-match implantation of a TECNIS® + 2.75 D multifocal IOL (ZKB00, Johnson & Johnson Vision, Santa Ana, CA, USA) into the dominant eye and a TECNIS® + 4.00 D multifocal IOL (ZMB00, Johnson & Johnson Vision) into the nondominant eye by one surgeon (HT) at the Cataract and Refractive Surgery Clinic of Asan Medical Center from March 2015 to February 2016. Tecnis ZKB00 is a diffractive bifocal IOL which has an anterior aspheric and diffractive profile posterior surface [13]. It features 50/50 light distribution between the distance and near foci independent of the pupillary size [13]. Tecnis ZMB00 shares same characteristics with ZKB00, which has + 4 D near addition (approximately + 3.0 D at the spectacle plane) [14]. Patients who met the following inclusion criteria were included: (1) older than 18 years, (2) preexisting corneal astigmatism less than + 1.00 D, and (3) visual acuity greater than 0.1 logMAR as measured with a potential acuity meter. Patients were excluded if they had (1) optical opacities or pathology on slit-lamp examination, (2) previous corneal surgeries, (3) ocular trauma, (4) intraocular surgery, (5) severe dry eyes, (6) corneal disease, (7) ocular infection, or (8) collagen vascular disease or other autoimmune diseases. Two subjects were excluded from the review because they had refractive surgery before. And for more precise verification of the surgery outcomes, within the limited data available, we additionally compared preoperative characteristics and postoperative visual acuities with bilateral TECNIS® 1-piece monofocal IOL (ZCB00, Johnson & Johnson Vision, Santa Ana, CA, USA) implantation group.

Measurements

All subjects underwent comprehensive ophthalmological examinations preoperatively, including logMAR visual acuity measurements of monocular and binocular uncorrected distance visual acuity (UDVA), UIVA, UNVA, corrected distance visual acuity (CDVA), corrected intermediate visual acuity (CIVA), and corrected near visual acuity (CNVA). Preoperative assessments also included autorefraction and keratometry (Canon R-50, Canon USA Inc., Huntington, NY, USA), slit-lamp examinations (Haag-Streit, Gartenstadtstrasse, Köniz, Switzerland), biometry (IOL Master 500, Carl Zeiss Meditec, Jena, Germany), and corneal topography (Orbscan, Bausch & Lomb, Rochester, NY, USA). Each patient’s dominant eye was determined prior to surgery using the hole-in-the-card test wherein the patient looks at a target through a 1 in. hole in the center of a card held at one arm’s length, with only one eye open at a time, to determine which eye saw the target.
The ophthalmic examinations conducted at 1 and 3-month after surgery included logMAR measurements of monocular and binocular UDVA, UIVA, UNVA, and CDVA. Autorefraction and keratometry were also performed. Intermediate visual acuity was measured at 60 cm. Near visual acuity was measured at 33, 40, and 50 cm, with near visual acuity expressed as the average of visual acuity at these distances. In many previous studies, near VA was measured only at 40 cm [15, 16]. But we focused on that there were delicately different needs for near target distance in various situations such as reading books or watching mobile phones, therefore, we defined near VA more broadly as the average VA at 33, 40, and 50 cm in this study. In addition, monocular and binocular defocus curves were obtained at 3-month postoperatively by measuring monocular or binocular visual acuity at 4 m starting from distance correction and then defocusing with added lenses in half-diopter steps from − 4.50 D to 0.00 D. In monofocal IOL group, defocus curves were obtatined only binocularly according to our clinic’s own protocol .
Contrast sensitivity was measured monocularly under uncorrected condition at 3-month postoperatively, using the Functional Acuity Contrast Test of the Ophtec 6500 view-in test system (Stereo Optical Co, Inc., Chicago, IL, USA) with stimulus spatial frequencies between 1.5 and 18 cycles per degree under photopic (target luminance = 85 cd per square meter [cd/m2]) and mesopic (target luminance = 3 cd/m2) conditions.
Finally, patients were asked to complete a questionnaire regarding their overall satisfaction, the occurrence of visual symptoms, and their dependence on spectacles for near and far vision. Overall satisfaction was assessed using a 5-point Likert scale: 1 = very dissatisfied, 2 = dissatisfied, 3 = neither satisfied nor dissatisfied, 4 = satisfied, and 5 = very satisfied. Visual symptoms (glare, halo, and visual disturbances at night or in the dark) were scored on a 5-point scale from 1 (absent symptoms) to 5 (severe symptoms). Patients were also asked if they would recommend bilateral mix-and-match implantation of multifocal IOLs to their friends or relatives, with allowed responses being yes or no.

Surgical technique

After instillation of topical anesthesia (0.5% proparacaine hydrochloride), the phacoemulsification surgery was performed. A continuous curvilinear capsulorrhexis marker with a 6.0-mm diameter was used to reference the corneal plane. The main clear corneal incision was made using a 2.2-mm keratome, followed by capsulorrhexis using a capsulorrhexis needle. Phacoemulsification was performed using either the Infiniti® or Centurion® phacoemulsifier (Alcon Laboratories, Inc., Fort Worth, TX, USA). Using an injector, a + 2.75 D multifocal IOL was implanted into the capsular bag of the dominant eye, and a + 4.00 D multifocal IOL was implanted into the capsular bag of the nondominant eye. The dominant eye was first implanted with Tecnis + 2.75 D multifocal IOL (ZKB00), and 1 week after surgery in the dominant eye, Tecnis + 4.00 D multifocal IOL (ZMB00) was implanted in the non-dominant eye. The target postoperative refraction was emmetropia in both eyes. All patients were administered 0.5% gatifloxacin ophthalmic solution (Gatiflo®, HANDOK, Seoul, South Korea) and prednisolone eye drops (Pred-Forte®, Allergan, Dublin, Ireland) for 1-month postoperatively. All of the above surgical protocols were equally applied to the bilaterally monofocal IOLs implanted group.

Statistical analysis

Results are expressed as the mean ± standard deviation with range. Differences between preoperative and postoperative data were assessed using the Wilcoxon signed-rank test. Values for the defocus curves for both eyes, the dominant eye, and the nondominant eye were analyzed by the Kruskal-Wallis test with the Bonferroni correction. All statistical analyses were performed using SPSS® version 21 software (IBM, SPSS Inc., Chicago, IL, USA). Differences were considered statistically significant for P values of less than 0.05.

Results

The study included 18 patients (10 female and 8 male), of mean age 65.8 ± 5.7 years (range, 55–76 years). Preoperative subject and ocular characteristics are summarized in Table 1. Statistically significant difference was not found in all parameters between bilateral multifocal IOL group and bilateral monofocal IOL group Table 2 shows preoperative and postoperative spherical equivalent (SE) and monocular visual acuity. At 3-month postoperatively, monocular logMAR UDVA, logMAR UIVA and logMAR UNVA of the dominant eye were 0.04 ± 0.05 (range: 0–0.20), 0.16 ± 0.05 (range: 0–0.50) and 0.17 ± 0.10 (range: 0–0.40), respectively. And monocular logMAR UDVA, logMAR UIVA and logMAR UNVA of the nondominant eye were 0.04 ± 0.05 (range: 0–0.30), 0.30 ± 0.12 (range: 0–0.50) and 0.11 ± 0.07 (range: 0–0.20), respectively. Postoperative monocular SE, logMAR UDVA, logMAR UIVA, logMAR UNVA and logMAR CDVA were all significantly better than preoperative values. Also significant differences were found between preoperative and postoperative binocular visual acuity for logMAR UDVA, logMAR UIVA, and logMAR UNVA in bilateral multifocal IOL group and only for logMAR UDVA in bilateral monofocal IOL group. (Tables 3 and 4). At 3-month postoperatively, binocular logMAR UDVA, logMAR UIVA and logMAR UNVA of bilateral multifocal IOL group were 0.01 ± 0.04 (range: 0–0.10), 0.16 ± 0.05 (range: 0–0.20) and 0.11 ± 0.07 (range: 0–0.20), respectively.
Table 1
Demographic and clinical characteristics of multifocal IOL group and monofocal IOL group
Parameter
Multifocal IOL group
Monofocal IOL group
P value*
Number of eyes / patients
36 / 18
34 / 17
 
Sex (male / female)
8 / 10
8 / 9
 
Age (years)
65.8 ± 5.7 (range: 55–76)
67.0 ± 5.9 (range: 56–78)
0.797
Mean corneal astigmatism (D)
0.60 ± 0.25 (range: 0–1.00) (dominant eye), 0.60 ± 0.30 (range: 0–1.00) (non-dominant eye)
0.58 ± 0.23 (range: 0–1.00) (dominant eye), 0.61 ± 0.27 (range: 0–1.00) (non-dominant eye)
0.688
0.724
Mean spherical equivalent (D)
0.55 ± 1.99 (range: − 5.75–2.25) (dominant eye), 0.21 ± 2.05 (range: − 5.50–3.50) (non-dominant eye)
0.61 ± 1.78 (range: − 4.75–2.00) (dominant eye), 0.25 ± 1.85 (range: − 5.00–3.25) (non-dominant eye)
0.556
0.473
Mean axial length (mm)
23.46 ± 1.08 (range: 20.84–26.00) (dominant eye), 23.52 ± 1.09 (range: 20.91–26.13) (non-dominant eye)
23.38 ± 0.99 (range: 21.01–25.87) (dominant eye), 23.50 ± 1.03 (range: 20.99–26.02) (non-dominant eye)
0.522
0.610
Results reported as means ± standard deviations
D Diopters, IOL Intraocular lenses
*Statistically significant at P < 0.05
Table 2
Refractive results and monocular visual acuity in patients with mix-and-match implantation of diffractive multifocal intraocular lenses with different add power
  
MRSE
LogMAR UDVA
LogMAR UIVA
LogMAR UNVA
LogMAR CDVA
Preoperative
Dominant eye
0.55 ± 1.99
(range: − 5.75–2.25)
0.36 ± 0.31
(range: 0–1.30)
0.64 ± 0.25
(range: 0.30–1.20)
0.58 ± 0.30
(range: 0–1.00)
0.10 ± 0.10
(range: 0–0.30)
Non-dominant eye
0.21 ± 2.05
(range: − 5.50–3.50)
0.38 ± 0.35
(range: 0–1.30)
0.62 ± 0.27
(range: 0.30–1.30)
0.58 ± 0.30
(range: 0–0.82)
0.13 ± 0.16
(range: − 0.10–0.52)
1-month postoperative
Dominant eye
−0.43 ± 0.44
(range: − 1.00–0)
0.06 ± 0.07
(range: 0–0.20)
0.16 ± 0.10
(range:0–0.50)
0.19 ± 0.09
(range: 0–0.50)
0.03 ± 0.06
(range: 0–0.20)
P value*
< 0.001
0.002
< 0.001
0.003
0.021
Non-dominant eye
−0.35 ± 0.37 (range: − 0.75–0.5)
0.07 ± 0.12 (range: 0–0.30)
0.29 ± 0.08 (range: 0.10–0.60)
0.13 ± 0.09 (range: 0–0.50)
0.02 ± 0.04 (range: 0–0.10)
P value*
< 0.001
0.001
0.001
0.001
0.015
3-month postoperative
Dominant eye
−0.43 ± 0.35 (range: − 0.88–0)
0.04 ± 0.05 (range: 0–0.20)
0.16 ± 0.05 (range: 0–0.50)
0.17 ± 0.10 (range: 0–0.40)
0.01 ± 0.03 (range: 0–0.10)
P value*
< 0.001
0.002
0.001
0.005
0.005
Non-dominant eye
−0.39 ± 0.21
(range: − 0.75–0.25)
0.04 ± 0.05
(range: 0–0.30)
0.30 ± 0.12
(range: 0–0.50)
0.11 ± 0.07
(range: 0–0.20)
0.01 ± 0.03
(range: 0–0.10)
P value*
< 0.001
0.003
0.002
0.002
0.019
Results reported as means ± standard deviations
MRSE Manifest refraction spherical equivalent, UDVA Uncorrected distance visual acuity; UIVA Uncorrected intermediate visual acuity, UNVA Uncorrected near visual acuity, CDVA Corrected distance visual acuity
*Compared with preoperative values
Table 3
Binocular visual acuity in multifocal IOL group and monofocal IOL group at 1 month postoperatively
 
Preoperative
P value+
1-month postoperative
P value*
P value+
LogMAR UDVA
(multifocal IOL group)
0.31 ± 0.31
(range: 0.10–0.60)
0.824
0.02 ± 0.04
(range: 0–0.10)
0.007
0.779
LogMAR UDVA
(monofocal IOL group)
0.34 ± 0.28
(range: 0.10–0.60)
0.03 ± 0.05
(range: 0–0.20)
0.010
LogMAR UIVA
(multifocal IOL group)
0.56 ± 0.23
(range: 0.30–0.80)
0.659
0.14 ± 0.07
(range: 0–0.20)
0.007
0.004
LogMAR UIVA
(monofocal IOL group)
0.49 ± 0.28
(range: 0.20–0.80)
0.37 ± 0.21
(range: 0.20–0.70)
0.102
LogMAR UNVA
(multifocal IOL group)
0.48 ± 0.25
(range: 0.20–0.80)
0.701
0.10 ± 0.10
(range: 0–0.30)
0.017
0.001
LogMAR UNVA
(monofocal IOL group)
0.43 ± 0.26
(range: 0.20–0.80)
0.36 ± 0.22
(range: 0.20–0.70)
0.338
LogMAR CDVA
(multifocal IOL group)
0.09 ± 0.13
(range: 0–0.30)
0.871
0.01 ± 0.03
(range: 0–0.10)
0.084
0.921
LogMAR CDVA
(monofocal IOL group)
0.10 ± 0.10
(range: 0–0.20)
0.01 ± 0.02
(range: 0–0.10)
0.077
Results reported as means ± standard deviations
UDVA Uncorrected distance visual acuity, UIVA Uncorrected intermediate visual acuity, UNVA Uncorrected near visual acuity, CDVA Corrected distance visual acuity, IOL Intraocular lenses
*Compared with preoperative values
+Compared between multifocal IOL group and monofocal IOL group
Table 4
Binocular visual acuity in multifocal IOL group and monofocal IOL group at 3 month postoperatively
 
Preoperative
P value+
3-month postoperative
P value*
P value+
LogMAR UDVA
(multifocal IOL group)
0.31 ± 0.31
(range: 0.10–0.60)
0.824
0.01 ± 0.04
(range: 0–0.10)
0.011
0.702
LogMAR UDVA
(monofocal IOL group)
0.34 ± 0.28
(range: 0.10–0.60)
0.03 ± 0.04
(range: 0–0.20)
0.009
LogMAR UIVA
(multifocal IOL group)
0.56 ± 0.23
(range: 0.30–0.80)
0.659
0.16 ± 0.05
(range: 0–0.20)
0.011
0.006
LogMAR UIVA
(monofocal IOL group)
0.49 ± 0.28
(range: 0.20–0.80)
0.35 ± 0.22
(range: 0.20–0.70)
0.120
LogMAR UNVA
(multifocal IOL group)
0.48 ± 0.25
(range: 0.20–0.80)
0.701
0.11 ± 0.07
(range: 0–0.30)
0.017
0.002
LogMAR UNVA
(monofocal IOL group)
0.43 ± 0.26
(range: 0.20–0.80)
0.37 ± 0.21
(range: 0.20–0.70)
0.305
LogMAR CDVA
(multifocal IOL group)
0.09 ± 0.13
(range: 0–0.30)
0.871
0.01 ± 0.03
(range: 0–0.10)
0.084
0.888
LogMAR CDVA
(monofocal IOL group)
0.10 ± 0.10
(range: 0–0.20)
0.01 ± 0.03
(range: 0–0.10)
0.086
Results reported as means ± standard deviations
UDVA Uncorrected distance visual acuity, UIVA Uncorrected intermediate visual acuity, UNVA Uncorrected near visual acuity, CDVA Corrected distance visual acuity, IOL Intraocular lenses
*Compared with preoperative values
+Compared between multifocal IOL group and monofocal IOL group
The binocular and monocular defocus curves are shown in Fig. 1. Binocular defocus curves of bilateral multifocal IOL group showed a better range of postoperative vision when compared with each monocular defocus curve across all distances. Visual acuity between − 1.50 D and − 4.00 D differed significantly among the 3 defocus curves (+ 2.75 D multifocal IOL, + 4.00 D multifocal IOL, and binocular multifocal IOL), with the exception of between − 2.50 D and − 3.00 D (P = 0.001 for − 1.50 D, 0.003 for − 2.00 D, 0.003 for − 3.50 D, and 0.002 for − 4.00 D). Defocus curves between − 3.50 D and − 4.00 D differed significantly for comparisons of binocular multifocal IOL and + 2.75 D multifocal IOL monocular vision (P = 0.003 for − 3.50 D and 0.001 for − 4.00 D) and between − 1.50 D and − 2.00 D for comparisons of binocular multifocal IOL and + 4.00 D multifocal IOL monocular vision (P = 0.001 for − 1.50 D and 0.002 for − 2.00 D). Defocus curves between − 1.50 D and − 4.00 D also differed significantly for comparisons of + 2.75 D multifocal IOL and + 4.00 D multifocal IOL monocular vision, with the exception of between − 2.50 D and − 3.00 D (P = 0.001 for − 1.50 D, 0.007 for − 2.00 D, 0.006 for − 3.50 D, and 0.007 for − 4.00 D). Binocular defocus curves of bilateral monofocal IOL groups exhibited a visual acuity of 0.10 logMAR or better between 0 and − 0.5 D, then there was a sharp decrease in visual acuity over − 1.0 D.
As shown in Fig. 2, contrast sensitivity was demonstrated under both photopic and mesopic conditions in eyes implanted with + 2.75 D and + 4.00 D multifocal IOLs, with no significant differences at any spatial frequency.
Sixteen subjects completed satisfaction questionnaires at 3-month postoperatively. Thirteen subjects (81.3%) reported that they were satisfied or very satisfied with their near vision, with an average satisfaction score of 4.4 ± 0.9 (range: 2–5). Only one subject (6.3%) reported occasionally needing glasses for near vision after surgery. Regarding the rate of visual symptoms, 4 (25.0%) subjects reported glare and halo symptom scores > 3 (average score: 2.4 ± 0.9) (range: 1–5) and 3 (18.8%) patients scored symptoms of visual disturbances at night or in the dark at ≥3 (average score: 2.3 ± 0.9) (range: 1–5) (Table 5).
Table 5
Results for questionnaire about overall satisfaction, visual symptoms and dependence on spectacles
Questionnaire
Response (average score/rate)
Overall satisfaction
4.4 ± 0.9 (range: 2–5), very satisfied or satisfied: 81.3%
Needing for near glasses after surgery
4.8 ± 0.6 (range: 3–5), occasionally need near glasses: 6.3%
Glare and halo symptoms
2.4 ± 0.9 (range: 1–5), over score 3: 25.0%
Visual disturbance at night or dark place
2.3 ± 0.9 (range: 1–5), over score 3: 18.8%
Recommendation for mix-and-match implantation
Yes: 93.8%
Results reported as means ± standard deviations
IOL Intraocular lenses
Satisfaction scale; 5 = very satisfied; 1 = very dissatisfied; Need for near glasses; 5 = not at all; 1 = always needed. Scale of discomfort due to visual symptom; 5 = severe symptoms; 1 = absent symptoms

Discussion

In the current study, we demonstrated that bilateral mix-and-match implantation of multifocal IOLs with add powers of + 2.75 D and + 4.00 D showed good near, intermediate, and far vision. We implanted multifocal IOL with add power of + 4.00 D into the nondominant eye for near visual acuity, which was relatively high add diopter compared to previous studies, based on the fact that Asians have lower amplitudes of accommodation and thus need higher add powers compared to Caucasians [17]. Previously, patients who underwent implantation of bilateral + 4.00 D multifocal IOLs have reported problems with intermediate vision, although their near visual acuity was good [16]. Bilateral implantation of + 2.50 D and + 3.00 D multifocal IOLs resulted in good near vision and noninferior intermediate and distance vision compared with bilateral implantation of + 2.50 D multifocal IOLs [2]. Unilateral implantation of TECNIS® + 2.75 D, + 3.25 D and + 4.00 D multifocal IOLs resulted in similar monocular UDVA, and UNVA was best in patients who underwent diffractive multifocal IOL implantation with add power of + 2.75 D at 50 cm [18]. Bilateral implantation of the TECNIS® + 2.75 D and + 3.25 D multifocal IOLs resulted in good binocular UIVA (0.07 ± 0.11 logMAR) but relatively inferior binocular UNVA (0.25 ± 0.11 logMAR) [11]. On the basis that mix-and-match implantation of multifocal IOLs with different add powers may be more beneficial than bilateral implantation of multifocal IOLs with the same add power for subjects who desire spectacle independence, we evaluated clinical outcomes after bilateral mix-and-match implantation of diffractive multifocal IOLs with different add powers. A + 2.75 D multifocal IOL was implanted into the dominant eye, and a + 4.00 D multifocal IOL was implanted into the nondominant eye.
The spectacle plane add power of multifocal IOLs differs from the IOL plane add power. For example, the spectacle plane add powers of the TECNIS® + 2.75 D, + 3.25 D, and + 4.00 D multifocal IOLs are + 2.01 D, + 2.37 D, and + 3.00 D, respectively. In the present study, depending on the spectacle plane add powers of the IOLs, visual acuity in the 0.00 D to − 3.00 D range of binocular defocus curves was 0.1 logMAR or better. Binocular visual acuity at − 3.50 D was better than 0.2 logMAR. On the other hand, a previous study found that the second peak of the binocular defocus curve was at − 2.00 D, and the visual acuity at − 2.50 D was about 0.0 logMAR [11]. Then, visual acuity dropped sharply at values below − 2.50 D, being 0.2 logMAR and 0.3 logMAR at − 3.00 D and − 3.50 D [11].
Recently introduced TECNIS® EDOF IOLs provide an elongated focal area but not multiple foci. Therefore, these IOLs could provide better intermediate vision than other currently available multifocal IOLs. However, UIVA from our study was better than that of subjects who underwent bilateral implantation of TECNIS® EDOF and TECNIS® + 4.00 D multifocal IOLs [19]. Although binocular defocus curves showed that the visual acuity in the 0.00 D to − 1.50 D range in a previous study were similar to those in the present study, the earlier study found that visual acuity decreased at values below − 2.00 D [20]. Several studies have evaluated methods to overcome the inferior UNVA following implantation of EDOF IOLs [9, 20, 21]. For example, according to the CONCERTO prospective case series study in which EDOF IOLs were implanted into both eyes of 411 subjects, 299 had emmetropia target (non-monovision) and 112 had micro-monovision. The mean UNVA of the micro-monovision and non-monovision groups were 0.17 logMAR and 0.21 logMAR, respectively, with the former being inferior to UNVA in our study [9]. To sum up, bilateral mix-and-match implantation of diffractive multifocal IOLs with add powers of + 2.75 D and + 4.00 D can be the good alternatives of implantion of EDOF IOL for intermediate visual acuity, with better near visual acuity, too.
According to the results for contrast sensitivity test of the current study, eyes implanted with diffractive multifocal IOLs with + 4.00 D add powers showed better contrast sensitivity than eyes implanted with diffractive multifocal IOLs with + 2.75 D add powers, albeit the differences were not significant. Contrast sensitivity under photopic and mesopic conditions without glare was similar to the results of previous studies of bilateral implantation of the TECNIS® bifocal IOLs and of bilateral implantation of an EDOF IOL and a TECNIS® + 4.00 D multifocal IOL [2, 18]. Furthermore, similar results at all spatial frequencies under photopic conditions were reported between the mix-and-match group and the bilaterally + 2.50 D implanted group [10].
In terms of subject satisfaction, we found that 25.0% had moderate glare and halo symptoms, while 18.8% reported night vision problems. These findings were inferior to those of the CONCERTO study [8]. On the other hand, these results are superior to those of previous studies with implantation of bilateral multifocal IOLs [22, 23]. In addition, the rate of spectacle independence was higher in our study than in the CONCERTO study. Our findings that 93.8% of subjects reported that they would recommend mix-and-match implantation of diffractive multifocal IOLs with + 2.75 D and + 4.00 D add powers to their friends and relatives indicated that despite some visual problems, these symptoms did not have a significant impact on overall satisfaction.
Our study had several limitations, including the small number of eyes and the lack of a control group. Nonetheless, we demonstrated that UIVA and UNVA following mix-and-match implantation of diffractive multifocal IOLs with + 2.75 D and + 4.00 D add powers was not inferior to those following bilateral implantation of EDOF IOLs or bilateral implantation of an EDOF IOL and a diffractive trifocal IOL [24, 25]. Further research is needed on the bilateral implantation of other types of multifocal IOLs, especially trifocal and EDOF IOLs, to determine which combination of IOLs could provide superior UIVA, extended visual acuity range on defocus curves, and high spectacle independence at all distances.

Conclusions

In summary, we evaluated clinical outcomes after mix-and-match implantation of diffractive multifocal IOLs with + 2.75 D and + 4.00 D add powers and demonstrated good near, intermediate, and far vision with a high degree of patient satisfaction. Therefore, mix-and-match implantation of + 2.75 D and + 4.00 D multifocal IOLs can be a good option for subjects who do not want to depend on glasses after cataract surgery.

Acknowledgements

None.
This study was performed in accordance with the tenets of the Declaration of Helsinki. Approval to conduct this study was obtained from the Institutional Review Board of the Asan Medical Center (Seoul, Republic of Korea). Informed written consent was obtained from all participants.
Not applicable.

Competing interests

No conflicts of interest and have no proprietary interest in any of the materials mentioned in this article.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Monaco G, Gari M, Di Censo F, Poscia A, Ruggi G, Scialdone A. Visual performance after bilateral implantation of 2 new presbyopia-correcting intraocular lenses: trifocal versus extended range of vision. J Cataract Refract Surg. 2017;43:737–47.CrossRef Monaco G, Gari M, Di Censo F, Poscia A, Ruggi G, Scialdone A. Visual performance after bilateral implantation of 2 new presbyopia-correcting intraocular lenses: trifocal versus extended range of vision. J Cataract Refract Surg. 2017;43:737–47.CrossRef
2.
Zurück zum Zitat Nuijts RM, Jonker SM, Kaufer RA, Lapid-Gortzak R, Mendicute J, Martinez CP, Schmickler S, Kohnen T. Bilateral implantation of +2.5 D multifocal intraocular lens and contralateral implantation of +2.5 D and +3.0 D multifocal intraocular lenses: clinical outcomes. J Cataract Refract Surg. 2016;42:194–202.CrossRef Nuijts RM, Jonker SM, Kaufer RA, Lapid-Gortzak R, Mendicute J, Martinez CP, Schmickler S, Kohnen T. Bilateral implantation of +2.5 D multifocal intraocular lens and contralateral implantation of +2.5 D and +3.0 D multifocal intraocular lenses: clinical outcomes. J Cataract Refract Surg. 2016;42:194–202.CrossRef
3.
Zurück zum Zitat Cochener B, Lafuma A, Khoshnood B, Courouve L, Berdeaux G. Comparison of outcomes with multifocal intraocular lenses: a meta-analysis. Clin Ophthalmol. 2011;5:45–56.PubMedPubMedCentral Cochener B, Lafuma A, Khoshnood B, Courouve L, Berdeaux G. Comparison of outcomes with multifocal intraocular lenses: a meta-analysis. Clin Ophthalmol. 2011;5:45–56.PubMedPubMedCentral
4.
Zurück zum Zitat Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37:720–6.CrossRef Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37:720–6.CrossRef
5.
Zurück zum Zitat Attia MSA, Auffarth GU, Kretz FTA, Tandogan T, Rabsilber TM, Holzer MP, Khoramnia R. Clinical evaluation of an extended depth of focus intraocular Lens with the Salzburg Reading desk. J Refract Surg. 2017;33:664–9.CrossRef Attia MSA, Auffarth GU, Kretz FTA, Tandogan T, Rabsilber TM, Holzer MP, Khoramnia R. Clinical evaluation of an extended depth of focus intraocular Lens with the Salzburg Reading desk. J Refract Surg. 2017;33:664–9.CrossRef
6.
Zurück zum Zitat Carballo-Alvarez J, Vazquez-Molini JM, Sanz-Fernandez JC, Garcia-Bella J, Polo V, Garcia-Feijoo J, Martinez-de-la-Casa JM. Visual outcomes after bilateral trifocal diffractive intraocular lens implantation. BMC Ophthalmol. 2015;15:26.CrossRef Carballo-Alvarez J, Vazquez-Molini JM, Sanz-Fernandez JC, Garcia-Bella J, Polo V, Garcia-Feijoo J, Martinez-de-la-Casa JM. Visual outcomes after bilateral trifocal diffractive intraocular lens implantation. BMC Ophthalmol. 2015;15:26.CrossRef
7.
Zurück zum Zitat Ruiz-Mesa R, Abengozar-Vela A, Aramburu A, Ruiz-Santos M. Comparison of visual outcomes after bilateral implantation of extended range of vision and trifocal intraocular lenses. Eur J Ophthalmol. 2017;27:460–5.CrossRef Ruiz-Mesa R, Abengozar-Vela A, Aramburu A, Ruiz-Santos M. Comparison of visual outcomes after bilateral implantation of extended range of vision and trifocal intraocular lenses. Eur J Ophthalmol. 2017;27:460–5.CrossRef
8.
Zurück zum Zitat Carson D, Hill WE, Hong X, Karakelle M. Optical bench performance of AcrySof((R)) IQ ReSTOR((R)), AT LISA((R)) tri, and FineVision((R)) intraocular lenses. Clin Ophthalmol. 2014;8:2105–13.PubMedPubMedCentral Carson D, Hill WE, Hong X, Karakelle M. Optical bench performance of AcrySof((R)) IQ ReSTOR((R)), AT LISA((R)) tri, and FineVision((R)) intraocular lenses. Clin Ophthalmol. 2014;8:2105–13.PubMedPubMedCentral
9.
Zurück zum Zitat Cochener B, Concerto SG. Clinical outcomes of a new extended range of vision intraocular lens: international multicenter Concerto study. J Cataract Refract Surg. 2016;42:1268–75.CrossRef Cochener B, Concerto SG. Clinical outcomes of a new extended range of vision intraocular lens: international multicenter Concerto study. J Cataract Refract Surg. 2016;42:1268–75.CrossRef
10.
Zurück zum Zitat Goes FJ. Visual results following implantation of a refractive multifocal IOL in one eye and a diffractive multifocal IOL in the contralateral eye. J Refract Surg. 2008;24:300–5.CrossRef Goes FJ. Visual results following implantation of a refractive multifocal IOL in one eye and a diffractive multifocal IOL in the contralateral eye. J Refract Surg. 2008;24:300–5.CrossRef
11.
Zurück zum Zitat Yang CM, Lim DH, Hwang S, Hyun J, Chung TY. Prospective study of bilateral mix-and-match implantation of diffractive multifocal intraocular lenses in Koreans. BMC Ophthalmol. 2018;18:73.CrossRef Yang CM, Lim DH, Hwang S, Hyun J, Chung TY. Prospective study of bilateral mix-and-match implantation of diffractive multifocal intraocular lenses in Koreans. BMC Ophthalmol. 2018;18:73.CrossRef
12.
Zurück zum Zitat Yoo A, Kwag JY, Song IS, Kim MJ, Jeong H, Kim JY, Tchah H. Comparison of visual function after implantation of inferior sector-shaped intraocular lenses: low-add +1.5 D vs +3.0 D. Eur J Ophthalmol. 2016;26:607–11.CrossRef Yoo A, Kwag JY, Song IS, Kim MJ, Jeong H, Kim JY, Tchah H. Comparison of visual function after implantation of inferior sector-shaped intraocular lenses: low-add +1.5 D vs +3.0 D. Eur J Ophthalmol. 2016;26:607–11.CrossRef
13.
Zurück zum Zitat Vega F, Alba-Bueno F, Millán MS, Varón C, Gil MA, Buil JA. Halo and through-focus performance of four diffractive multifocal intraocular lenses. Invest Ophthalmol Vis Sci. 2015;56:3967–75.CrossRef Vega F, Alba-Bueno F, Millán MS, Varón C, Gil MA, Buil JA. Halo and through-focus performance of four diffractive multifocal intraocular lenses. Invest Ophthalmol Vis Sci. 2015;56:3967–75.CrossRef
14.
Zurück zum Zitat Liu X, Song X, Wang W, Zhu Y, Lyu D, Shentu X, Chen P, Yu Y, Yao K. Comparison of the clinical outcomes between Echelette extended range of vision and diffractive bifocal intraocular lenses. J Ophthalmol. 2019;2019:5815040.PubMedPubMedCentral Liu X, Song X, Wang W, Zhu Y, Lyu D, Shentu X, Chen P, Yu Y, Yao K. Comparison of the clinical outcomes between Echelette extended range of vision and diffractive bifocal intraocular lenses. J Ophthalmol. 2019;2019:5815040.PubMedPubMedCentral
15.
Zurück zum Zitat Son HS, Kim SH, Auffarth GU, Choi CY. Prospective comparative study of tolerance to refractive errors after implantation of extended depth of focus and monofocal intraocular lenses with identical aspheric platform in Korean population. BMC Ophthalmol. 2019;19:187.CrossRef Son HS, Kim SH, Auffarth GU, Choi CY. Prospective comparative study of tolerance to refractive errors after implantation of extended depth of focus and monofocal intraocular lenses with identical aspheric platform in Korean population. BMC Ophthalmol. 2019;19:187.CrossRef
16.
Zurück zum Zitat Yoon SY, Song IS, Kim JY, Kim MJ, Tchah H. Bilateral mix-and-match versus unilateral multifocal intraocular lens implantation: long-term comparison. J Cataract Refract Surg. 2013;39:1682–90.CrossRef Yoon SY, Song IS, Kim JY, Kim MJ, Tchah H. Bilateral mix-and-match versus unilateral multifocal intraocular lens implantation: long-term comparison. J Cataract Refract Surg. 2013;39:1682–90.CrossRef
17.
Zurück zum Zitat Edwards MH, Law LF, Lee CM, Leung KM, Lui WO. Clinical norms for amplitude of accommodation in Chinese. Ophthalmic Physiol Opt. 1993;13:199–204.CrossRef Edwards MH, Law LF, Lee CM, Leung KM, Lui WO. Clinical norms for amplitude of accommodation in Chinese. Ophthalmic Physiol Opt. 1993;13:199–204.CrossRef
18.
Zurück zum Zitat Kim JS, Jung JW, Lee JM, Seo KY, Kim EK, Kim TI. Clinical Outcomes Following Implantation of Diffractive Multifocal Intraocular Lenses With Varying Add Powers. Am J Ophthalmol. 2015;160:702–9 e1.CrossRef Kim JS, Jung JW, Lee JM, Seo KY, Kim EK, Kim TI. Clinical Outcomes Following Implantation of Diffractive Multifocal Intraocular Lenses With Varying Add Powers. Am J Ophthalmol. 2015;160:702–9 e1.CrossRef
19.
Zurück zum Zitat de Medeiros AL, de Araujo Rolim AG, Motta AFP, Ventura BV, Vilar C, Chaves M, Carricondo PC, Hida WT. Comparison of visual outcomes after bilateral implantation of a diffractive trifocal intraocular lens and blended implantation of an extended depth of focus intraocular lens with a diffractive bifocal intraocular lens. Clin Ophthalmol. 2017;11:1911–6.CrossRef de Medeiros AL, de Araujo Rolim AG, Motta AFP, Ventura BV, Vilar C, Chaves M, Carricondo PC, Hida WT. Comparison of visual outcomes after bilateral implantation of a diffractive trifocal intraocular lens and blended implantation of an extended depth of focus intraocular lens with a diffractive bifocal intraocular lens. Clin Ophthalmol. 2017;11:1911–6.CrossRef
20.
Zurück zum Zitat Cochener B, Boutillier G, Lamard M, Auberger-Zagnoli C. A comparative evaluation of a new generation of diffractive trifocal and extended depth of focus intraocular lenses. J Refract Surg. 2018;34:507–14.CrossRef Cochener B, Boutillier G, Lamard M, Auberger-Zagnoli C. A comparative evaluation of a new generation of diffractive trifocal and extended depth of focus intraocular lenses. J Refract Surg. 2018;34:507–14.CrossRef
21.
Zurück zum Zitat Hogarty DT, Russell DJ, Ward BM, Dewhurst N, Burt P. Comparing visual acuity, range of vision and spectacle independence in the extended range of vision and monofocal intraocular lens. Clin Exp Ophthalmol. 2018;46:854–60.CrossRef Hogarty DT, Russell DJ, Ward BM, Dewhurst N, Burt P. Comparing visual acuity, range of vision and spectacle independence in the extended range of vision and monofocal intraocular lens. Clin Exp Ophthalmol. 2018;46:854–60.CrossRef
22.
Zurück zum Zitat Nuzzi R, Tridico F. Comparison of visual outcomes, spectacles dependence and patient satisfaction of multifocal and accommodative intraocular lenses: innovative perspectives for maximal refractive-oriented cataract surgery. BMC Ophthalmol. 2017;17:12.CrossRef Nuzzi R, Tridico F. Comparison of visual outcomes, spectacles dependence and patient satisfaction of multifocal and accommodative intraocular lenses: innovative perspectives for maximal refractive-oriented cataract surgery. BMC Ophthalmol. 2017;17:12.CrossRef
23.
Zurück zum Zitat Maxwell A, Holland E, Cibik L, Fakadej A, Foster G, Grosinger L, Moyes A, Nielsen S, Silverstein S, Toyos M, Weinstein A, Hartzell S. Clinical and patient-reported outcomes of bilateral implantation of a +2.5 diopter multifocal intraocular lens. J Cataract Refract Surg. 2017;43:29–41.CrossRef Maxwell A, Holland E, Cibik L, Fakadej A, Foster G, Grosinger L, Moyes A, Nielsen S, Silverstein S, Toyos M, Weinstein A, Hartzell S. Clinical and patient-reported outcomes of bilateral implantation of a +2.5 diopter multifocal intraocular lens. J Cataract Refract Surg. 2017;43:29–41.CrossRef
24.
Zurück zum Zitat Torun Acar B, Duman E, Simsek S. Clinical outcomes of a new diffractive trifocal intraocular lens with enhanced depth of focus (EDOF). BMC Ophthalmol. 2016;16:208.CrossRef Torun Acar B, Duman E, Simsek S. Clinical outcomes of a new diffractive trifocal intraocular lens with enhanced depth of focus (EDOF). BMC Ophthalmol. 2016;16:208.CrossRef
25.
Zurück zum Zitat Savini G, Schiano-Lomoriello D, Balducci N, Barboni P. Visual performance of a new extended depth-of-focus intraocular Lens compared to a distance-dominant diffractive multifocal intraocular Lens. J Refract Surg. 2018;34:228–35.CrossRef Savini G, Schiano-Lomoriello D, Balducci N, Barboni P. Visual performance of a new extended depth-of-focus intraocular Lens compared to a distance-dominant diffractive multifocal intraocular Lens. J Refract Surg. 2018;34:228–35.CrossRef
Metadaten
Titel
Clinical outcomes after mix-and-match implantation of diffractive multifocal intraocular lenses with + 2.75 and + 4.00 diopter add powers
verfasst von
Jae Hyuck Lee
Hun Lee
Jin Ah. Lee
Aeri Yoo
Jae Yong Kim
Hungwon Tchah
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2020
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-020-01460-7

Weitere Artikel der Ausgabe 1/2020

BMC Ophthalmology 1/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.