Erschienen in: BMC Ophthalmology 1/2023

Open Access 01.12.2023 | Research

Clinical outcomes of implantation of posterior chamber phakic intraocular lens for pathologic and non-pathologic myopia

verfasst von: Lin Caixia, Bai Yawen, Fang Yuxin, Li Xiaoxia, Wang Yuhan, Yang Ke, Qiao Liya

Erschienen in: BMC Ophthalmology | Ausgabe 1/2023

Abstract

Background

To compare the clinical outcomes of posterior chamber phakic intraocular lens (pIOL) implantation for non-pathological myopia and pathological myopia.

Methods

This retrospective case series study which were conducted in Beijing Tongren Eye Center between July 2017 and Oct 2021 comprised 192 eyes of 100 consecutive patients undergoing pIOL implantation. Eyes were divided into two groups based on having pathological myopia or not. Predictability, efficacy, safety, and adverse events were compared at 6 months after pIOL implantation.

Results

Our study included 86 non-pathological myopes (171 eyes, group1) and 14 pathological myopes (21eyes, group2) to analysis. The average ages were 25.5 and 33.0, respectively, and the spherical equivalent (SE) were -9.31D and -17.50D pre-operation. Six months after pIOL implantation, the SE were 0.00 and -0.50, respectively, and the refraction changes were statistically significant (P ≤ 0.05). Six months after surgery, 76.92% and 80.41% were within ± 0.50 D of the target and 92.31% and 95.88% were within ± 1.00 D. All eyes had unchanged BCVA or gained 1 or more lines in both groups and mean BCVA both improved a line 6m after operation. The efficacy index in the two groups were 0.95 and 0.88 and the safety index were 1.20, 1.33, respectively which was significantly different (P ≤ 0.05). Over the 6-month follow-up, no cataract, pigment dispersion glaucoma, pupillary block, or other vision-threatening complications happened, either.

Conclusions

The pIOL performed well for the correction of both non-pathological and pathological myopia throughout the 6-month observation period. The clinical outcomes of pIOL implantation for non-pathological myopia are essentially equivalent to those for pathological myopia.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
pIOL
Posterior chamber phakic intraocular lens
CLE
Clear lens extraction
LASIK
Laser-assisted in-situ keratomileusis
IOP
Intraocular pressure
ACD
Anterior chamber depth
ECD
Endothelial cell density
UDVA
Uncorrected distance visual acuity
BCVA
Best corrected visual acuity
CDVA
Corrected distance visual acuity
CNV
Choroidal neovascularization
SE
Spherical equivalent

Background

Except for spectacles and contact lenses, there are many surgical options to correct myopia [1, 2]. Clear lens extraction (CLE), one procedure, has been used to treat high myopia for a long time [3, 4], however, after extracting the clear lens, people would lose natural accommodation. Refractive surgery, including laser-assisted in-situ keratomileusis (LASIK), small incision lenticule extraction and so on [57] has been widely accepted as a safe and effective surgical method for myopic correction, however, patients with high myopia or thin corneas have the risk of developing halos, glare and keratoectasia after undergoing refractive surgery. Besides, a large amount of laser ablation may lead to the decrease of superior intrinsic corneal optical performance.
In recent two decades, phakic intraocular lens (pIOL) implantation has gained widespread popularity as an effective refractive option for surgical correction of moderate to high ametropia [810]. Especially, the model V4c Visian Implantable Collamer Lens which was designed with a central hole of 0.36 mm could correct moderate or high myopia with little complications such as anterior subcapsular cataract, increased intraocular pressure (IOP), endothelial cell loss, pigment dispersion, pupillary block, and glaucoma. Besides, pIOL is a reversible procedure that could improve visual acuity and have excellent refractive stability while preserving accommodation.
Up to now, many clinical researches [1117] have proved pIOL is a safe, effective and predictable procedure to correct moderate or high myopia. Nevertheless, there are few studies reporting clinical outcome of pIOL for pathological myopia. The aim of our research was to demonstrate the predictability, efficacy and safety of the model V4c pIOL to correct non-pathological and pathological myopia and to analyze the possible occurrence of adverse events at 6m after operation.

Patients and methods

A retrospective chart review of a consecutive clinical case series study performed at Eye Center, Beijing Tongren Hospital, China between July 2017 and Oct 2021, consisting of 86 non-pathological myopia patients (171 eyes) and 14 pathological myopia (21 eyes) with Visian Implantable Collamer Lens (model V4c pIOL). This study followed the tenets of the Declaration of Helsinki and was approved by the institutional review board of Beijing Tongren Hospital. Written informed consent was obtained from all patients after they received a full explanation of the nature and possible consequences of the study.
The inclusion criteria were 1) age older than 18 years old, 2) a clear central cornea, 3) myopia less than -0.5D, 4) stable refraction for at least two years. The exclusion criteria included 1) anterior chamber depth (ACD) from the corneal endothelium of less than 2.8 mm, 2) endothelial cell density (ECD) less than 2000 cell/mm2, 3) mesopic pupil larger than 7.0 mm, 4) cataract, macular degeneration or retinopathy, retinal detachment, glaucoma, neuro-ophthalmic disease, amblyopia, and a history of ocular inflammation.

Preoperative assessment

Before pIOL implantation, patients had a complete ophthalmologic examination, which included uncorrected distance visual acuity (UDVA), best corrected visual acuity (BCVA), manifest refractions, noncontact intraocular pressure (Canon), slit lamp examination, corneal topography (Oculus pentacam HR), pachymetry (Lenstar LS 900), ECD measurement (Specular microscope, SP.3000P), fundus photographs (Canon CR-2 AF, Japan), and binocular indirect ophthalmoscopy through dilated pupils.

Phakic intraocular lens

The V4c Visian Implantable Collamer Lens model used in this study is a new model that has an artificial hole of 0.36 mm diameter in the optic center which is designed to improve the circulation of aqueous humor and eliminate the need for preoperative laser iridotomy or intra-operative peripheral iridectomy. Details about the model V4c pIOL have been described previously [11]. In all eyes, emmetropia or the minimum myopia dioptor was selected as the postoperative target refraction. The pIOL power calculation was performed using the modified vertex formula of the pIOL power table software which was provided by the manufacturer. The pIOL size was individually chosen according to the horizontal white-to-white distance and ACD and also following the pIOL manufacturer’s recommendations. The pIOL was implanted by the same experienced doctor (Q.L.Y) uneventfully according to the pIOL implantation technique [12].

Fundus photograph classification

According to fundus photographs we put the patients into non-pathological myopia (group 1) and pathological myopia (group 2), those with normal fundus or tessellation (Fig. 1A and B) were classified into group 1, those with diffuse chorioretinal atrophy (Fig. 1C), patchy chorioretinal atrophy (Fig. 1D), macular atrophy, lacquer cracks or myopic choroidal neovascularization (CNV) at the posterior pole were classified into group 2. Because Fuchs spot is a pigmented spot representing the scarring phase of myopic CNV, thus we put Fuchs spot into the myopic CNV category [18]. Fundus photographs were graded by two graders (L.C.X., B.Y.W) who had been trained for pathological myopia grading carefully. And the agreement in grading pathological myopia between them was excellent (weighted kappa values for all myopic maculopathy features were ≥ 0.8). When grading fundus photographs, if there was disagreement between the two graders, a third grader (F.Y.X) who had a lot of experience in grading pathological myopia also graded the photograph. If there was still no agreement after discussion, a retinal specialist (Q.L.Y) reassessed the relevant photographs and made a final diagnosis.

Statistics

Statistical analyses were performed using SAS9.3. Patients with at least one eye satisfying the inclusion and exclusion criteria were enrolled in the study. All the statistical analyses were performed based on eyes. For the continuous outcomes, median and quartile ranges were presented due to non-normal distributions. For the categorical outcomes, the frequency and percentage were reported. To adjust the relationship between eyes, generalized estimated equation was used for all statistical comparisons. Patients were considered as the cluster, which had two levels (eyes). The correlation between measurements for the same subject was setting as exchangeable. Considering confounding effects, all the comparisons were further adjusted age and gender. Visual acuity data were converted to logMar values. Differences were considered statistically significant when the P value was less than 0.05.

Results

Our study included 86 non-pathological myopes (171 eyes, group1) and 14 pathological myopes (21eyes, group2) to analysis. All the pathological myopes had diffuse and/or patchy chorioretinal atrophy, and didn’t have macular atrophy or other pathological changes. Table 1 showed demographics and pIOL characteristics (pre-operation) of non-pathological and pathological myopia patients. The average ages of these two groups were 25.5 (22, 30) and 33 (29, 43), respectively, and patients in group 1 were much younger than those in group 2 and the difference was significant (P ≤ 0.05). There were less males in group 1 than in group 2 (18.5% (10) vs 54.6% (6)). UCVA in group 1(3.22 (2.59, 3.91)) was poorer than that in group 2 (2.30 (2.30, 3.51)), however the difference was not significant. While BCVA in group 1(0.00(-0.18, 0.00)) was better than that in group 2(0.51 (0.11, 0.69)), and the difference was significant. The spheric, cylinder and SE in group 1 were much higher than those in group 2(-8.63D, -1.25D, -9.31D vs -17.5D, -1.5D, -17.5D), and there was a significant difference. The SE pIOL residual in group 1 was lower than that in group 2(-0.04D vs -0.15D, P = 0.0491), however the difference was not significant after adjusted age and gender (P = 0.1642). While there was no significant difference in VA, IOP, RF, RS, RM, white to white and ECD between the two groups.
Table 1
Demographics and ICL characteristics (pre-operation) of patients in group1 and 2
Characteristic
Group1 (N = 171 eyes)
Group2 (N = 21 eyes)
P value
P *
Age
25.5 (22, 30)
33 (29, 43)
0.0182
 
Male
10 (18.5%)
6 (54.6%)
0.0200
 
UCVA (logMAR)
3.22 (2.59, 3.91)
2.30 (2.30, 3.51)
0.5092
0.2109
BCVA (logMAR)
0.00 (-0.18, 0.00)
0.51 (0.11, 0.69)
 < 0.001
 < 0.001
IOP
15.50 (13.75, 17.30)
18.00 (14.50, 19.00)
0.2272
0.3936
Spheric
-8.63 (-10.75, -6.75)
-17.50 (-18.50, -13.50)
 < 0.001
 < 0.001
Cylinder
-1.25 (-1.75, -0.75)
-1.50 (-3.00, -1.50)
0.0343
0.0318
SE
-9.31 (-11.75, -7.25)
-17.50 (-18.75, -14.75)
 < 0.001
 < 0.001
SE pIOL residual
-0.04 (-0.14, 0.04)
-0.15 (-0.52, -0.08)
0.0491
0.1642
RF
7.77 (7.63, 8.00)
7.84 (7.74, 7.97)
0.5071
0.4815
RS
7.52 (7.38, 7.70)
7.57 (7.47, 7.84)
0.9634
0.8423
RM
7.63 (7.51, 7.83)
7.66 (7.63, 8.11)
0.2924
0.2897
White to White
11.86 (11.60, 12.09)
11.85 (11.66, 11.93)
0.5373
0.7090
ECD
2871.5 (2731.0, 3060.0)
3050.5 (2687.5, 3148.5)
0.6436
0.3900
The parameters are presented as mean (lower quartile, upper quartile) and number (percent), UCVA stands for uncorrected visual acuity, BCVA stands for best corrected visual acuity, IOP stands for intraocular pressure, SE stands for spherical equivalent, RF stands for corneal flat axil radius of curvature, RS stands for corneal steep axil radius of curvature, RM stands for mean corneal radius of curvature, ECD stands for endothelial cell density. P * stands for P value after adjusting for age and gender
Ocular parameters in the two groups 6 months (6m) after pIOL implantation were shown in Table 2, and from the table we can see that 6m after operation, patients with non-pathological myopia had better UCVA and BCVA (0.00, -0.18) than those with pathological myopia (0.51, 0.00), and the differences were significant. The spheric lens between the two groups (-0.13D vs -0.50D) was significantly different, however, after adjusted for age and gender, the difference was not meaningful further. The SE was also significantly different between the two groups (0.00D vs -0.50D). And there was no significant difference in IOP, cylinder, RF, RS, RM, white to white and ECD between the two groups.
Table 2
pIOL characteristics (post operation) of pathological and non-pathological myopia patients
Characteristic
Group1 (N = 171 eyes)
Group2 (N = 21 eyes)
P value
P *
UCVA (logMAR)
0.00 (-0.18, 0.11)
0.51 (0.00, 0.92)
< 0.001
0.0014
BCVA (logMAR)
-0.18 (-0.18, 0.00)
0.00 (0.00, 0.22)
0.0052
0.0114
IOP
15.45 (14.00, 17.40)
15.80 (14.30, 18.20)
0.2122
0.1244
Cylinder
-0.50 (-0.50, 0.50)
-1.00 (-1.00, -0.50)
0.1576
0.1576
Spheric
-0.13 (-0.25, 0.00)
-0.50 (-1.63, -0.13)
0.0451
0.1200
SE
0.00 (-0.38, 0.00)
-0.50 (-1.00, -0.50)
0.0133
0.0298
RF
7.77 (7.66, 8.14)
7.79 (7.68, 8.24)
0.4550
0.3214
RS
7.55 (7.45, 7.92)
7.64 (7.39, 7.94)
0.7242
0.5763
RM
7.70 (7.56, 8.02)
7.71 (7.51, 8.09)
0.5662
0.4215
White to White
11.86 (11.74, 12.20)
11.85 (11.68, 11.96)
0.4237
0.4236
ECD
2903.0 (2612.0, 2996.0)
2860.5 (2615.0, 3075.0)
0.5988
0.9598
The parameters are presented as mean (lower quartile, upper quartile) and number (percent), UCVA stands for uncorrected visual acuity, BCVA stands for best corrected visual acuity, IOP stands for intraocular pressure, SE stands for spherical equivalent, RF stands for corneal flat axil radius of curvature, RS stands for corneal steep axil radius of curvature, RM stands for mean corneal radius of curvature, ECD stands for endothelial cell density. P * stands for P value after adjusting for age and gender
Table 3 demonstrated ocular parameters pre-operation and 6m post operation in group 1. From the table, we can see that compared to pre-operation, patients had better UCVA and BCVA (3.22 and 0.00 vs 0.00 and -0.18) 6m after pIOL implantation, and the improvement was statistically significant. After pIOL plantation, spheric, cylinder and SE all obviously increased (from -8.63, -1.25D, -9.31D to -0.13D, -0.50D, 0.0D, respectively). In addition, RS became lager after operation compared to pre-operation (7.52 vs 7.55) and the difference was significant. While in IOP, RF, RM, white to white and ECD, there were no differences between pre-operation and post operation. From the table we also could see that ECD become lager post operation than pre-operation (2871.5vs 2903.0).
Table 3
Comparisons of ocular characteristics between pre-operation and 6m Post operation in group1 (N = 171 eyes)
Characteristics
pre-operation
Post operation
P value
P*
UCVA (logMAR)
3.22 (2.59, 3.91)
0.00 (-0.18, 0.11)
 < 0.001
 < 0.001
BCVA (logMAR)
0.00 (-0.18, 0.00)
-0.18 (-0.18, 0.00)
 < 0.001
 < 0.001
IOP
15.50 (13.75, 17.30)
15.45 (14.00, 17.40)
0.9335
0.9586
Spheric
-8.63 (-10.75, -6.75)
-0.13 (-0.25, 0.00)
 < 0.001
 < 0.001
Cylinder
-1.25 (-1.75, -0.75)
-0.50 (-0.50, 0.50)
 < 0.001
 < 0.001
SE
-9.31 (-11.75, -7.25)
0.00 (-0.38, 0.00)
 < 0.001
 < 0.001
RF
7.77 (7.63, 8.00)
7.77 (7.66, 8.14)
0.3798
0.3144
RS
7.52 (7.38, 7.70)
7.55 (7.45, 7.92)
0.0069
0.0061
RM
7.63 (7.51, 7.83)
7.70 (7.56, 8.02)
0.4062
0.4167
White to White
11.86 (11.60, 12.09)
11.86 (11.74, 12.20)
0.2425
0.2425
ECD
2871.5 (2731.0, 3060.0)
2903.0 (2612.0, 2996.0)
0.2520
0.2598
The parameters are presented as mean (lower quartile, upper quartile) and number (percent), UCVA stands for uncorrected visual acuity, BCVA stands for best corrected visual acuity, IOP stands for intraocular pressure, SE stands for spherical equivalent, RF stands for corneal flat axil radius of curvature, RS stands for corneal steep axil radius of curvature, RM stands for mean corneal radius of curvature, ECD stands for endothelial cell density. P * stands for P value after adjusting for age and gender
Table 4 displayed that compared to ocular parameters pre-operation, patients in group 2 had better UCVA and BCVA (2.30, 0.51 vs 0.51, 0.00), lower spheric and SE (-17.5D, -17.5D vs -0.5D, -0.5D) 6m after operation. Besides, after adjusted for age and gender, cylinder difference became significant between pre-operation and post operation (-1.5D, -1.0D). And there were no differences between pre-operation and post operation in IOP, RF, RS, RM, white to white and ECD.
Table 4
Comparisons of characteristics between pre-operation and post operation in group2 (N = 21 eyes)
Outcome
Pre-operation
Post operation
P value
P*
UCVA (logMAR)
2.30 (2.30, 3.51)
0.51 (0.00, 0.92)
 < 0.001
 < 0.001
BCVA (logMAR)
0.51 (0.11, 0.69)
0.00 (0.00, 0.22)
 < 0.001
 < 0.001
IOP
18.00 (14.50, 19.00)
15.80 (14.30, 18.20)
0.4960
0.6716
Spheric
-17.50 (-18.50, -13.50)
-0.50 (-1.63, -0.13)
 < 0.001
 < 0.001
Cylinder
-1.50 (-3.00, -1.50)
-1.00 (-1.00, -0.50)
0.2989
0.0244
SE
-17.50 (-18.75, -14.75)
-0.50 (-1.00, -0.50)
 < 0.001
 < 0.001
RF
7.84 (7.74, 7.97)
7.79 (7.68, 8.24)
0.7183
0.7436
RS
7.57 (7.47, 7.84)
7.64 (7.39, 7.94)
0.7989
0.7995
RM
7.66 (7.63, 8.11)
7.71 (7.51, 8.09)
0.3184
0.3213
White to White
11.85 (11.66, 11.93)
11.85 (11.68, 11.96)
0.8072
0.6368
ECD
3050.5 (2687.5, 3148.5)
2860.5 (2615.0, 3075.0)
0.2989
0.2988
The parameters are presented as mean (lower quartile, upper quartile) and number (percent), UCVA stands for uncorrected visual acuity, BCVA stands for best corrected visual acuity, IOP stands for intraocular pressure, SE stands for spherical equivalent, RF stands for corneal flat axil radius of curvature, RS stands for corneal steep axil radius of curvature, RM stands for mean corneal radius of curvature, ECD stands for endothelial cell density. P *stands for P value adjusted for age and gender
Figure 2 showed a scatterplot of the attempted versus the achieved SE correction. Six months after surgery, 76.92% (10/13eyes) in group 1 and 80.41% (78 /97eyes) in group 2 were within ± 0.50 D of the target and 92.31% (12/13 eyes) and 95.88% (93/97 eyes) were within ± 1.00 D, respectively.
Clinical outcomes 6m after pIOL implantation in group 1 and group 2 were shown in Table 5. The efficacy index in the two groups were 0.95(0.83, 1.00) and 0.88(0.67, 1.00) and the safety index were 1.20(1.00, 1.20), 1.33(1.11, 1.43), respectively and the safety index between the two groups were significantly different. BCVA in the two groups both improved one line and the exact increase numbers were -0.18(-0.18, 0.00) and -0.29(-0.36, -0.11), separately. ECD increased in group 1 (0.01(-0.04, 0.04) while ECD loss happened in group 2 (-0.01(-0.02,0.07), however, the difference was not significant. The Arch in the two groups were 500.50 (360.50, 750.50) and 573.00 (333.00, 623.00), separately, and the difference was large, however, there was no clinical significance. There was more percentile (72.5% (71) vs 30.8% (4)) to have good VA (≥ 1.0) in group 1 than in group 2 and the difference were significant.
Table 5
Comparison of outcomes 6m after pIOL implantation in group1 and group2
Characteristic
Group1(N = 171 eyes)
Group2 (N = 21 eyes)
P value
P*
Efficacy
0.95 (0.83, 1.00)
0.88 (0.67, 1.00)
0.1275
0.3972
Safety
1.20 (1.00, 1.20)
1.33 (1.11, 1.43)
0.0466
0.0413
BCVA Increase (logMAR)
-0.18 (-0.18, 0.00)
-0.29 (-0.36, -0.11)
0.0592
0.0680
BCVA Increase (Line)
1.00 (0.00, 1.00)
1.00 (0.00, 1.00)
0.1654
0.1916
ECD Loss
0.01 (-0.04, 0.04)
-0.01 (-0.02, 0.07)
0.8832
0.0728
Arch
500.50 (360.50, 750.50)
573.00 (333.00, 623.00)
0.3894
0.1690
VA (> = 1.0)
71 (72.5%)
4 (30.8%)
0.0059
0.0292
The parameters are presented as mean (lower quartile, upper quartile) and number (percent), BCVA stands for best corrected visual acuity, ECD stands for endothelial cell density, VA stands for visual acuity. P *stands for P value adjusted for age and gender
Figure 3 showed the changes in BCVA in group 1 and group 2 six months after pIOL implantation. From the figure we could see that no eye lost 1 or more lines, 30.8% and 48% did not change from preoperatively, 46.2% and 41% gained 1 line, 23.1% and 11% gained 2 lines or more, respectively, in the two groups. There was more percentile in group 1 to gain 1 line or more compared to that in group 2, yet the difference was not significant.
There were no intraoperative complications, and no eye required pIOL explantation or repositioning in the two groups. Over the 6-month follow-up, no cataract, pigment dispersion glaucoma, pupillary block, or other vision-threatening complications happened, either.

Discussion

The V4c Visian Implantable Collamer Lens (pIOL) has been improved to have good predictability, stability, efficacy, and safety for the correction of moderate to high myopia in previous studies [12, 15, 19]. However, there is few researches previously to study how the pIOL implantation to correct pathological myopia. The present study showed that the pIOL implantation is an effective treatment for both non-pathological and pathological myopia, with excellent safety and predictability throughout a 6-month observation period.
Our study demonstrated that spheric, cylinder and SE all increased significantly in both groups 6m after pIOL implantation and patients all get good UCVA (from 3.22 to 0.00 and from 2.30 to 0.51) and BCVA (from 0.00 to -0.18 and from 0.51 to 0.00). The UCVA and BCVA improvement may be contributed to that compared to frame glasses, pIOL has less spherical aberration, astigmatism and other off-axis aberrations, less prism effect, wider field of vision and smaller reduction of retinal imaging, thus after pIOL implantation patients can gain better VA no matter UCVA or BCVA. Compared with patients in group 1, those in group 2 had poorer UCVA and BCVA after pIOL implantation, the reason may be that patients with pathological myopia had pathological changes such as diffuse chorioretinal dystrophy which could decrease the resolution of retina, yet the difference is not significant.
In group 1, we found that RS became lager 6m after operation, this may be caused by during pIOL implantation, we need to make a transparent limbal incision which flattens the steep axis of the cornea and increases the radius of corneal curvature. Thus, we can design the surgical incision location combined with the corneal astigmatism axis to reduce corneal astigmatism during the operation.
Both the postoperative UDVA and BCVA in group 1 were significantly better than in group 2, possibly because patients in group 1 had lower rates of myopia than those in group 2. And there was more percentile who gained 1.0 or better UCVA in group 1 than in group 2, this may be due to patients with pathological myopia have some pathological change which affect the VA improvement after pIOL implantation.
We obtained stable (0.00 (-0.38, 0.00) and -0.50 (-1.00, -0.50)) and predictable refractive outcomes in both groups 6m after surgery, with 76.92% (10/13eyes) in group 1 and 80.41% (78 /97eyes) in group 2 within ± 0.50 D of the target and 92.31% (12/13 eyes) and 95.88% (93/97 eyes) within ± 1.00 D. The percentiles were similar to those of Kazutaka Kamiya’s research [20] and lower than those of other previous studies [1113], the reason may be that we included some patients whose myopia were less than -18.0D, the lowest diopter of the pIOL.
The efficacy and safety of non-pathologic myopia and pathologic myopia were 0.95 and 0.88, 1.20 and 1.33, respectively, the efficacy in the two groups was comparable and the safety in group 2 was higher than that in group 1, which means patients with pathological myopia could get a lager BCVA improvement after pIOL implantation, the reason may be that patients with pathological myopia usually have higher myopia, therefore the aberration and reduction of retinal imaging were more obvious caused by frame glasses, after pIOL implantation, the visual improvement was more significant than those with non-pathological myopia. And the efficacy and safety of both non-pathologic myopia and pathologic myopia groups were similar to previous studies [11, 13, 15] which means that pIOL is an effective and safe operation to correct non-pathological myopia as well as pathological myopia. Previous studies [2123] indicated that pIOL implantation provided long-term stability and good refractive outcomes for high myopia.
About ECD change, one common complication after pIOL implantation, we found that ECD decreased in group 2 while it increased in group 1 six months after pIOL implantation, however, the difference between pre-operation and post operation was little (1%) and not significant. the ECD increase may be caused by that the ECD in different corneal endothelial regions is different, and the measurement areas chosen before and after operation were different which led to the ECD increase after operation. In addition, the measurement error also played a certain role. And the specific reasons need to be further explored. About ECD reduction, Yan Ju et al. [24] found that the mean 3-month postoperative ECD decreased but had no statistically difference compared with the preoperative ECD after pIOL implantation. Whose result was similar to ours. Regarding the long-term endothelial cell loss, Jime´nez-Alfaro et al. [25] showed that the percentage of endothelial cells lost was 6.57% 2 years after surgery [26]. Pineda-Fernandez et al. [26] found 6.09% of endothelial cell loss 3 years after surgery. Kazutaka Kamiya et al. [17] reported that the percentage of endothelial cell loss was 3.7% 4 years postoperatively, and Alfonso et al. [21] stated that the total endothelial cell loss was 7.7% 5 years postoperatively. Akihito Igarashi et al. [22] found that the mean percentage of endothelial cell loss was 6.2% 8 years after surgery. Jae Hwan Choi et al. [16] found that the rate of ECD decrease was 4.8% at 10 years after surgery. However, no eyes reduced to less than 2000 cell/mm2 or had a significant loss over 30%. A decrease in the mean ECD in acute stage could be aggravated by endothelial damage during pIOL implantation, inflammation after implantation, or physical contact between the pIOL and corneal endothelium. And the main cause of endothelial cell loss over the long term is the aging process [16]. How the number of endothelial cells change over long time about our patients need to be observe.
In our research, we didn’t find any case with other postoperative complications including IOP rise (covering pupillary block) and cataract formation in both non-pathological and pathological myopia groups 6m after pIOL implantation. These findings were consistent with those of previous studies on pIOL implantation, even without laser or surgical peripheral iridotomy or iridectomy performed [15, 27, 28]. Kawamorita et al. [29] used computational fluid dynamics to demonstrate that hole pIOLs may improve the circulation of aqueous humor to the anterior surface of the crystalline lens which may prevent postoperative pupillary block and cataract formation. Previously Kamiya et al. reported that the incidence of cataract formation with V4 pIOL was 10.7% four years after pIOL implantation if traumatic cataract formation was excluded [17]. While, Packer et al. [30] in a Meta-analysis and review on Hole pIOL implantation including data on 1291 eyes followed for up to 5 years reported no incidence of asymptomatic anterior subcapsular cataract formation. And they also reported that the age at surgery and degree of myopia are risk factors for cataract formation after pIOL implantation. Jae Hwan Choi et al. [16] found that there was a higher cumulative incidence of lens opacity in patients with higher preoperative age or more severe preoperative myopia. And in the future more prolonged and careful follow-up researches are still required to determine the exact rate of cataract formation in Hole pIOL implantation eyes.
This study has several limitations. First, the study was performed in a retrospective way, a prospective randomised study would be ideal for confirming our results. Second, there were a relatively low patient population, especially in the pathological group, thus our data may have a possible selection bias. Third, a 6m of follow-up might not be sufficient to evaluate the long-term complications of a pIOL. The last but not the least, the number of patients in the pathological group was small and the pathological myopia mainly contained diffuse and patchy chorioretinal atrophy, how does other pathological myopia such as macular atrophy, lack cracks and so on with pIOL implantation, we don’t know. Nevertheless, to our knowledge, the current study was the first to evaluate refractive and visual outcomes of V4c pIOL for correction of non-pathological and pathological myopia.
In conclusion, our study supports the view that the pIOL performed equally well in correction of pathological myopia as it did in non-pathological myopia during the 6-month observation period. pIOL implantation for pathological myopia is clinically, if not statistically, equivalent to the results in non-pathological myopia in terms of the predictability, efficacy and safety. pIOL implantation is a viable surgical option for the treatment of pathological myopia.

Acknowledgements

Not applicable.

Declarations

This study followed the tenets of the Declaration of Helsinki and was approved by the institutional review board of Beijing Tongren Hospital. Written informed consent was obtained from all patients after they received a full explanation of the nature and possible consequences of the study.
Not applicable.

Competing interests

The authors declare no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Emarah AM, El-Helw MA, Yassin HM. Comparison of clear lens extraction and collamer lens implantation in high myopia. Clin Ophthalmol. 2010;4:447–54.PubMedPubMedCentral Emarah AM, El-Helw MA, Yassin HM. Comparison of clear lens extraction and collamer lens implantation in high myopia. Clin Ophthalmol. 2010;4:447–54.PubMedPubMedCentral
2.
Zurück zum Zitat Choi JH, Lim DH, Nam SW, Yang CM, Chung ES, Chung TY. Ten-year clinical outcomes after implantation of a posterior chamber phakic intraocular lens for myopia. J Cataract Refr Surg. 2019;45(11):1555–61.CrossRef Choi JH, Lim DH, Nam SW, Yang CM, Chung ES, Chung TY. Ten-year clinical outcomes after implantation of a posterior chamber phakic intraocular lens for myopia. J Cataract Refr Surg. 2019;45(11):1555–61.CrossRef
3.
Zurück zum Zitat Colin J, Robinet A. Clear lensectomy and implantation of low-power posterior chamber intraocular lens for the correction of high myopia. Ophthalmology. 1994;101(1):107–12.CrossRefPubMed Colin J, Robinet A. Clear lensectomy and implantation of low-power posterior chamber intraocular lens for the correction of high myopia. Ophthalmology. 1994;101(1):107–12.CrossRefPubMed
4.
Zurück zum Zitat Goldberg MF. Clear lens extraction for axial myopia. An appraisal. Ophthalmology. 1987;94(5):571–82.CrossRefPubMed Goldberg MF. Clear lens extraction for axial myopia. An appraisal. Ophthalmology. 1987;94(5):571–82.CrossRefPubMed
5.
Zurück zum Zitat Ang M, Farook M, Htoon HM, Mehta JS. Randomized clinical trial comparing femtosecond LASIK and small-incision lenticule extraction. Ophthalmology. 2020;127(6):724–30.CrossRefPubMed Ang M, Farook M, Htoon HM, Mehta JS. Randomized clinical trial comparing femtosecond LASIK and small-incision lenticule extraction. Ophthalmology. 2020;127(6):724–30.CrossRefPubMed
6.
Zurück zum Zitat Blum M, Lauer AS, Kunert KS, Sekundo W. 10-year results of small incision lenticule extraction. J Refract Surg. 2019;35(10):618–23.CrossRefPubMed Blum M, Lauer AS, Kunert KS, Sekundo W. 10-year results of small incision lenticule extraction. J Refract Surg. 2019;35(10):618–23.CrossRefPubMed
7.
Zurück zum Zitat Sutton G, Lawless M, Hodge C. Laser in situ keratomileusis in 2012: a review. Clin Exp Optom. 2014;97(1):18–29.CrossRefPubMed Sutton G, Lawless M, Hodge C. Laser in situ keratomileusis in 2012: a review. Clin Exp Optom. 2014;97(1):18–29.CrossRefPubMed
8.
Zurück zum Zitat Chen H, Liu Y, Niu G, Ma J. Excimer laser versus phakic intraocular lenses for myopia and astigmatism: a meta-analysis of randomized controlled trials. Eye Contact Lens. 2018;44(3):137–43.CrossRefPubMed Chen H, Liu Y, Niu G, Ma J. Excimer laser versus phakic intraocular lenses for myopia and astigmatism: a meta-analysis of randomized controlled trials. Eye Contact Lens. 2018;44(3):137–43.CrossRefPubMed
9.
Zurück zum Zitat Kohnen T. Phakic intraocular lenses: where are we now? J Cataract Refract Surg. 2018;44(2):121–3.CrossRefPubMed Kohnen T. Phakic intraocular lenses: where are we now? J Cataract Refract Surg. 2018;44(2):121–3.CrossRefPubMed
10.
11.
Zurück zum Zitat Lisa C, Naveiras M, Alfonso-Bartolozzi B, Belda-Salmeron L, Montes-Mico R, Alfonso JF. Posterior chamber collagen copolymer phakic intraocular lens with a central hole to correct myopia: one-year follow-up. J Cataract Refract Surg. 2015;41(6):1153–9.CrossRefPubMed Lisa C, Naveiras M, Alfonso-Bartolozzi B, Belda-Salmeron L, Montes-Mico R, Alfonso JF. Posterior chamber collagen copolymer phakic intraocular lens with a central hole to correct myopia: one-year follow-up. J Cataract Refract Surg. 2015;41(6):1153–9.CrossRefPubMed
12.
Zurück zum Zitat Alfonso JF, Lisa C, Fernandez-Vega Cueto L, Belda-Salmeron L, Madrid-Costa D, Montes-Mico R. Clinical outcomes after implantation of a posterior chamber collagen copolymer phakic intraocular lens with a central hole for myopic correction. J Cataract Refract Surg. 2013;39(6):915–21.CrossRefPubMed Alfonso JF, Lisa C, Fernandez-Vega Cueto L, Belda-Salmeron L, Madrid-Costa D, Montes-Mico R. Clinical outcomes after implantation of a posterior chamber collagen copolymer phakic intraocular lens with a central hole for myopic correction. J Cataract Refract Surg. 2013;39(6):915–21.CrossRefPubMed
13.
Zurück zum Zitat Kamiya K, Shimizu K, Igarashi A, Kitazawa Y, Kojima T, Nakamura T, et al. Posterior chamber phakic intraocular lens implantation: comparative, multicentre study in 351 eyes with low-to-moderate or high myopia. Br J Ophthalmol. 2018;102(2):177–81.CrossRefPubMed Kamiya K, Shimizu K, Igarashi A, Kitazawa Y, Kojima T, Nakamura T, et al. Posterior chamber phakic intraocular lens implantation: comparative, multicentre study in 351 eyes with low-to-moderate or high myopia. Br J Ophthalmol. 2018;102(2):177–81.CrossRefPubMed
14.
Zurück zum Zitat Gomez-Bastar A, Jaimes M, Graue-Hernandez EO, Ramirez-Luquin T, Ramirez-Miranda A, Navas A. Long-term refractive outcomes of posterior chamber phakic (spheric and toric implantable collamer lens) intraocular lens implantation. Int Ophthalmol. 2014;34(3):583–90.CrossRefPubMed Gomez-Bastar A, Jaimes M, Graue-Hernandez EO, Ramirez-Luquin T, Ramirez-Miranda A, Navas A. Long-term refractive outcomes of posterior chamber phakic (spheric and toric implantable collamer lens) intraocular lens implantation. Int Ophthalmol. 2014;34(3):583–90.CrossRefPubMed
15.
Zurück zum Zitat Shimizu K, Kamiya K, Igarashi A, Shiratani T. Early clinical outcomes of implantation of posterior chamber phakic intraocular lens with a central hole (Hole ICL) for moderate to high myopia. Br J Ophthalmol. 2012;96(3):409–12.CrossRefPubMed Shimizu K, Kamiya K, Igarashi A, Shiratani T. Early clinical outcomes of implantation of posterior chamber phakic intraocular lens with a central hole (Hole ICL) for moderate to high myopia. Br J Ophthalmol. 2012;96(3):409–12.CrossRefPubMed
16.
Zurück zum Zitat Choi JH, Lim DH, Nam SW, Yang CM, Chung ES, Chung TY. Ten-year clinical outcomes after implantation of a posterior chamber phakic intraocular lens for myopia. J Cataract Refract Surg. 2019;45(11):1555–61.CrossRefPubMed Choi JH, Lim DH, Nam SW, Yang CM, Chung ES, Chung TY. Ten-year clinical outcomes after implantation of a posterior chamber phakic intraocular lens for myopia. J Cataract Refract Surg. 2019;45(11):1555–61.CrossRefPubMed
17.
Zurück zum Zitat Kamiya K, Shimizu K, Igarashi A, Hikita F, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol (Chicago, Ill : 1960). 2009;127(7):845–50.CrossRef Kamiya K, Shimizu K, Igarashi A, Hikita F, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol (Chicago, Ill : 1960). 2009;127(7):845–50.CrossRef
18.
Zurück zum Zitat Ohno-Matsui K, Kawasaki R, Jonas JB, Cheung CMG, Saw SM, Verhoeven VJM, et al. International photographic classification and grading system for myopic maculopathy. Am J Ophthalmol. 2015;159(5):877–83.CrossRefPubMed Ohno-Matsui K, Kawasaki R, Jonas JB, Cheung CMG, Saw SM, Verhoeven VJM, et al. International photographic classification and grading system for myopic maculopathy. Am J Ophthalmol. 2015;159(5):877–83.CrossRefPubMed
19.
Zurück zum Zitat Shimizu K, Kamiya K Fau - Igarashi A, Igarashi A Fau - Shiratani T, Shiratani T. Intraindividual comparison of visual performance after posterior chamber phakic intraocular lens with and without a central hole implantation for moderate to high myopia. Am J Ophthalmol. 2012;154(3):486–494.e1. Shimizu K, Kamiya K Fau - Igarashi A, Igarashi A Fau - Shiratani T, Shiratani T. Intraindividual comparison of visual performance after posterior chamber phakic intraocular lens with and without a central hole implantation for moderate to high myopia. Am J Ophthalmol. 2012;154(3):486–494.e1.
20.
Zurück zum Zitat Kamiya K, Shimizu K Fau - Igarashi A, Igarashi A Fau - Hikita F, Hikita F Fau - Komatsu M, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol. 2009;127(7):845–50. Kamiya K, Shimizu K Fau - Igarashi A, Igarashi A Fau - Hikita F, Hikita F Fau - Komatsu M, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol. 2009;127(7):845–50.
21.
Zurück zum Zitat Alfonso JF, Baamonde B, Fernández-Vega L, Fernandes P, González-Méijome JM, Montés-Micó R. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J Cataract Refract Surg. 2011;37(5):873–80.CrossRefPubMed Alfonso JF, Baamonde B, Fernández-Vega L, Fernandes P, González-Méijome JM, Montés-Micó R. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J Cataract Refract Surg. 2011;37(5):873–80.CrossRefPubMed
22.
Zurück zum Zitat Igarashi A, Shimizu K, Kamiya K. Eight-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Am J Ophthalmol. 2014;157(3):532-9 e1.CrossRefPubMed Igarashi A, Shimizu K, Kamiya K. Eight-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Am J Ophthalmol. 2014;157(3):532-9 e1.CrossRefPubMed
23.
Zurück zum Zitat Fernandez-Vega-Cueto L, Alfonso-Bartolozzi B, Lisa C, Madrid-Costa D, Alfonso JF. Seven-year follow-up of posterior chamber phakic intraocular lens with central port design. Eye Vis (Lond). 2021;8(1):23.CrossRefPubMed Fernandez-Vega-Cueto L, Alfonso-Bartolozzi B, Lisa C, Madrid-Costa D, Alfonso JF. Seven-year follow-up of posterior chamber phakic intraocular lens with central port design. Eye Vis (Lond). 2021;8(1):23.CrossRefPubMed
24.
Zurück zum Zitat Ju Y, Gao XW, Ren B. Posterior chamber phakic intraocular lens implantation for high myopia. Int J Ophthalmol. 2013;6(6):831–5.PubMedPubMedCentral Ju Y, Gao XW, Ren B. Posterior chamber phakic intraocular lens implantation for high myopia. Int J Ophthalmol. 2013;6(6):831–5.PubMedPubMedCentral
25.
Zurück zum Zitat Jimenez-Alfaro I, Benitez del Castillo JM, Garcia-Feijoo J, Gil de Bernabe JG, Serrano de La Iglesia JM. Safety of posterior chamber phakic intraocular lenses for the correction of high myopia: anterior segment changes after posterior chamber phakic intraocular lens implantation. Ophthalmology. 2001;108(1):90–9.CrossRefPubMed Jimenez-Alfaro I, Benitez del Castillo JM, Garcia-Feijoo J, Gil de Bernabe JG, Serrano de La Iglesia JM. Safety of posterior chamber phakic intraocular lenses for the correction of high myopia: anterior segment changes after posterior chamber phakic intraocular lens implantation. Ophthalmology. 2001;108(1):90–9.CrossRefPubMed
26.
Zurück zum Zitat Pineda-Fernandez A, Jaramillo J, Vargas J, Jaramillo M, Jaramillo J, Galindez A. Phakic posterior chamber intraocular lens for high myopia. J Cataract Refract Surg. 2004;30(11):2277–83.CrossRefPubMed Pineda-Fernandez A, Jaramillo J, Vargas J, Jaramillo M, Jaramillo J, Galindez A. Phakic posterior chamber intraocular lens for high myopia. J Cataract Refract Surg. 2004;30(11):2277–83.CrossRefPubMed
27.
Zurück zum Zitat Kamiya K, Shimizu K Fau - Saito A, Saito A Fau - Igarashi A, Igarashi A Fau - Kobashi H, Kobashi H. Comparison of optical quality and intraocular scattering after posterior chamber phakic intraocular lens with and without a central hole (Hole ICL and Conventional ICL) implantation using the double-pass instrument. PLoS One. 2013;8(6):e66846. Kamiya K, Shimizu K Fau - Saito A, Saito A Fau - Igarashi A, Igarashi A Fau - Kobashi H, Kobashi H. Comparison of optical quality and intraocular scattering after posterior chamber phakic intraocular lens with and without a central hole (Hole ICL and Conventional ICL) implantation using the double-pass instrument. PLoS One. 2013;8(6):e66846.
28.
Zurück zum Zitat Shimizu K, Kamiya K, Igarashi A, Kobashi H. Long-Term Comparison of Posterior Chamber Phakic Intraocular Lens With and Without a Central Hole (Hole ICL and Conventional ICL) Implantation for Moderate to High Myopia and Myopic Astigmatism Consort-Compliant Article. Medicine. 2016;95(14):e3270.CrossRefPubMedPubMedCentral Shimizu K, Kamiya K, Igarashi A, Kobashi H. Long-Term Comparison of Posterior Chamber Phakic Intraocular Lens With and Without a Central Hole (Hole ICL and Conventional ICL) Implantation for Moderate to High Myopia and Myopic Astigmatism Consort-Compliant Article. Medicine. 2016;95(14):e3270.CrossRefPubMedPubMedCentral
29.
Zurück zum Zitat Kawamorita T, Uozato H, Shimizu K. Fluid dynamics simulation of aqueous humour in a posterior-chamber phakic intraocular lens with a central perforation. Graef Arch Clin Exp. 2012;250(6):935–9.CrossRef Kawamorita T, Uozato H, Shimizu K. Fluid dynamics simulation of aqueous humour in a posterior-chamber phakic intraocular lens with a central perforation. Graef Arch Clin Exp. 2012;250(6):935–9.CrossRef
30.
Zurück zum Zitat Packer M. Meta-analysis and review: effectiveness, safety, and central port design of the intraocular collamer lens. Clin Ophthalmol. 2016;10:1059–77.CrossRefPubMedPubMedCentral Packer M. Meta-analysis and review: effectiveness, safety, and central port design of the intraocular collamer lens. Clin Ophthalmol. 2016;10:1059–77.CrossRefPubMedPubMedCentral
Metadaten
Titel
Clinical outcomes of implantation of posterior chamber phakic intraocular lens for pathologic and non-pathologic myopia
verfasst von
Lin Caixia
Bai Yawen
Fang Yuxin
Li Xiaoxia
Wang Yuhan
Yang Ke
Qiao Liya
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2023
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-023-02890-9

Weitere Artikel der Ausgabe 1/2023

BMC Ophthalmology 1/2023 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Die überschießende Wundheilung in der filtrierenden Glaukomchirurgie ist ein zentraler Faktor für ein operatives Versagen. Nach der Einführung der Trabekulektomie in den 1960er-Jahren wurden viele Faktoren erkannt, die mit einer vermehrten …

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.