Skip to main content
Erschienen in: BMC Ophthalmology 1/2023

Open Access 01.12.2023 | Research

Comparison between the CASIA SS-1000 and Pentacam in measuring corneal curvatures and corneal thickness maps

verfasst von: Robert M. Feldman, Gene Kim, Alice Z. Chuang, Atsushi Shiraishi, Keiichiro Okamoto, Makoto Tsukamoto

Erschienen in: BMC Ophthalmology | Ausgabe 1/2023

Abstract

Purpose

To compare the intra-device repeatability and inter-device reproducibility between two anterior segment imaging instruments, the CASIA SS-1000 (Tomey Corp., Nagoya, Japan) and Pentacam (OCULUS, Arlington, WA) in measuring anterior segment parameters.

Methods

Single-center, prospective clinical trial. Participants ≥20 years of age were included. One eye was randomly selected, each imaged by three CASIA SS-1000 devices and three Pentacam devices by three different examiners. Each photographer operated a pair of devices, one CASIA SS-1000 and one Pentacam. The image order for each participant was determined by a random permutation table. Three images were taken from each device. A total of 18 images were taken for each eye. Ratios of the standard deviations, referenced as (CASIA/Pentacam), were calculated to compare the repeatability and reproducibility of the two imaging instruments.

Results

In all, 66 participants with a mean age of 46.4 years (±21.7) were enrolled in the study. All repeatability ratios and intra-device variability were less than 1 (anterior corneal curvature: flat = 0.86, steep = 0.85; posterior corneal curvature: flat = 0.43, steep = 0.61; and map: thinnest = 0.22; central = 0.24, 2 mm = 0.26, 4 mm = 0.27, and 6 mm = 0.30). All reproducibility ratios, which measure the inter-device variability, were less than 1 (anterior corneal curvature: flat = 0.58, steep = 0.73; posterior corneal curvature: flat = 0.25, steep = 0.31; and pachymetry map: thinnest = 0.20; central = 0.20; 2 mm = 0.20; 4 mm = 0.19; and 6 mm = 0.22). A ratio of less than 1 indicates that the CASIA SS-1000 has more consistent measurements.

Conclusions

The CASIA SS-1000 was found to have better repeatability and reproducibility compared to the Pentacam for both corneal curvature and pachymetry maps. This greater consistency may require further study to determine whether the decreased variability can be translated into improved clinical results.
Hinweise
Meeting Presentation: Previously presented (in part) at the 2016 American Society of Cataract and Refractive Surgery in New Orleans, Louisiana, United States.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AKf
Anterior curvature flat axis
AKs
Anterior curvature steep axis
CCT
Central corneal thickness
CV
Coefficient of variation
IOL
Intraocular lens
LOA
Limits of agreement
OCT
Optical coherence tomography
PCT
Peripheral corneal thickness
PKf
Posterior curvature flat axis
PKs
Posterior curvature steep axis
SD
Standard deviation
TCP(X)
Horizontal distance from the apex to the thinnest corneal point
TCP(Y)
Vertical distance from the apex to the thinnest corneal point
TCT
Thinnest corneal thickness

Introduction

Accurate measurements of the cornea are becoming increasingly important. Clinicians are using precise quantitative measurements of epithelium, stroma and endothelium to evaluate patients and improve surgical outcomes. Surgical planning before femtosecond laser-assisted treatments now requires precise stromal thickness to avoid sacrificing the stromal bed (< 250 μm). Pachymetry maps are relied on for any pre- and post-laser surgery evaluation, especially in subclinical and early diagnosis of corneal ectasia or form fruste keratoconus [1, 2]. Furthermore, morphological measurements, such as AC depth and angle, can prove vital in placing phakic intraocular lens (IOL) and recognizing complications of refractive surgeries [3].
Due to operator-dependent accuracy of contact tomography with ultrasound, accurate noncontact topography and tomography have evolved quickly. Scheimpflug systems, such as the Pentacam, allow for analysis of distorted and aberrated corneas, with true elevation measurements of both the anterior and posterior surfaces of the cornea. However, this technique is limited by cornea clarity, pupil diameter, and iris reflections [4]. Within the past decade, optical coherence tomography (OCT) imaging has increasingly been used to image the anterior chamber. Compared with Scheimpflug imaging, it boasts higher resolution and faster scans. With a newer swept-source OCT, such as the CASIA SS-1000, the increased frequency wavelength source decreases eye motion artifacts and scans deep corneal structures, due to little signal decay. The high-quality, cross-sectional images produce topographical maps of both the anterior and posterior corneal surfaces [5].
Since very small differences in corneal measurements alter clinical decisions, reproducibility and repeatability are extremely important in obtaining measurements from noncontact tomography. Studies have examined the inter-operator reproducibility and repeatability of a single Scheimpflug machine with a single OCT machine in corneal thickness and anterior chamber measurements in both healthy and postoperative eyes [611]. However, there has been no study looking at the inter-device reproducibility using multiple machines of the same type with the same operators. The purpose of this study is to compare both the intra-device repeatability and inter-device reproducibility between two anterior segment imaging instruments, the CASIA SS-1000 (OCT) and Pentacam (Scheimpflug imaging), in measuring anterior segment parameters.

Participants and Methods

This prospective, single-center study was approved by the Institutional Review Board/Ethics at Ehime University Hospital, Shitsukawa, Toon, Ehime, Japan, and was registered on clinicaltrials.​gov (NCT02077790) on 04/03/2014. All research adhered to the tenets of the Declaration of Helsinki and was HIPAA compliant.

Participants

In all, 66 consecutive adult subjects (at least 20 years of age) were recruited. After obtaining informed consent, demographics, medical/ocular history, and present ocular medications were recorded. Eyes were excluded if they had any previous intraocular or corneal surgery (except laser refractive surgery), used a rigid contact lens within 1 week or soft contact lens within 24 hours, were currently on mydriatic and/or miotic medications, or were incapable of fixating at the internal fixation light or opening the eye sufficient to image a full image area. When both eyes of the participant were eligible, one eye was randomly selected by coin flip.
Study eyes underwent a series of imaging by three CASIA SS-1000 devices and three Pentacam devices. These devices were placed in three separate rooms. In each room, one CASIA SS-1000 and one Pentacam were installed and one examiner was assigned to operate the devices. Three images were taken by each of six devices. A computer-generated, randomization schedule was used to determine the order of the rooms/examiners and the order of the devices (CASIA SS-1000 vs. Pentacam) in a room. The exam procedure took approximately 1 h for imaging a total of 18 images in each eye. With this design, the device effect and examiner effect were confounded, and is called the [device+examiner] effect throughout the rest of this paper.

Study instruments

CASIA SS-1000

The CASIA SS-1000 is a three-dimensional, noncontact, noninvasive, imaging system based on the principle of “swept-source” OCT. This system achieves high-resolution imaging of 10 μm (axial) and 30 μm (transverse) and high-speed scanning of 30,000 A-scans per second. All images were taken in 3D mode using the “corneal map” scan.

Pentacam

The Pentacam is a rotating Scheimpflug camera. It is a noncontact, noninvasive system that uses a monochromatic slit-light source (blue LED at 475 nm) for measuring anterior segment tomography. Twenty-five images with 500 measurement points on the front and the back of the corneal surface are acquired over a 180-degree rotation in 2 s. All images were taken by 3D scanning of 25 Picture/1 Sec. In addition, to reduce examiner-dependent variability, the CASIA SS-1000 auto-alignment/auto shot and Pentacam’s automatic release mode were used. This way each instrument automatically determines when correct focus and alignment is achieved and then performs a scan.

Measurements

All measurements are performed in constant, dim-lighting conditions. Several corneal curvature and corneal thickness measurements were obtained from each image by each instrument:
1.
Anterior axial power of the flattest meridian (AKf, D)
 
2.
Anterior axial power of the steepest meridian (AKs, D)
 
3.
Posterior axial power of the flattest meridian (PKf, D)
 
4.
Posterior axial power of the steepest meridian (PKs, D)
 
5.
Central corneal thickness (CCT, μm)
 
6.
Thinnest corneal thickness (TCT, μm)
 
7.
Horizontal location of thinnest corneal point (TCP(X, mm), in which 0 indicates the apex, positive value for the temporal direction, and negative value for the nasal direction (TCP(X), mm)
 
8.
Vertical location of thinnest corneal point from apex (TCP(Y), mm), in which 0 indicates the apex, positive value for the superior direction, and negative value for the inferior direction
 
9.
Peripheral corneal thickness at intersection between the 2-mm diameter circle and the nasal horizontal meridian (PCT2, μm)
 
10.
Peripheral corneal thickness at intersection between the 4-mm diameter circle and the nasal horizontal meridian (PCT4, μm)
 
11.
Peripheral corneal thickness at intersection between the 6-mm diameter circle and the nasal horizontal meridian (PCT6, μm)
 

Data analysis

Demographics were summarized by mean and standard deviation (SD) for continuous variables or by frequency (%) for discrete variables. The repeatability and reproducibility were evaluated by:
  • Repeatability (intra-device) SD—the variation of measurements taken by each device for each eye; and
  • Reproducibility (inter-device + intra-device) SD—the variation of measurements taken by all three devices, by each instrument, for each eye.
Repeatability SD and Reproducibility SD were calculated for each device using a random effect model with random effects, participants (66 participants) and devices + examiners (three devices + examiners). The ratio of CASIA SS-1000 SD to Pentacam SD for each measure was computed for repeatability, reproducibility, and agreement. A ratio of less than 1 indicates that CASIA SS-1000 is more consistent than Pentacam, and vice versa. The coefficients of variation (CVs) were also calculated by repeatability SD/mean and reproducibility SD/mean.
Agreement, mean difference (bias) and limits of agreement (LOA) between CASIA SS-1000 and Pentacam was calculated. In addition, Deming regression analysis, which allows measurement errors in both CASIA SS-1000 and Pentacam, was used to examine any systematic deviations and estimate the conversion equations. Bland-Altman agreement plots and Deming regression plots were graphed.
Statistical analysis was performed using PROC MIXED procedure in SAS for Window 9.4 (SAS Inc., Cary, NC). Bland-Altman agreement plot was generated using BA.plot() and Deming regression analysis was performed using Deming() in MethComp package using R × 64 Version 3.01.

Results

In all, 66 eyes of 66 participants were included in the study. The mean (±SD) age was 46 (±22) years (range 23–79 years). Of 66 study eyes, 32 (48%) were right eyes and 35 (53%) were from male subjects. Most eyes had no corneal pathology (54, 82%). Other ocular pathologies can be seen in Table 1.
Table 1
Demographics and Ocular Characteristics
Variable
ALL
(N = 66)
Age (year, mean ± SD [Range])
46.4 ± 21.7
[23 to 79]
Sex (male, %)
35 (53%)
Race (Asian, %)
66 (100%)
Study eye (Right, %)
32 (48%)
Corneal Pathology (%)
 None
54 (81.8%)
 Dry eye
4 (6.1%)
 Cataract
3 (4.6%)
 Fuchs dystrophy
2 (3.0%)
 Previous refractive surgery
1 (1.5%)
 Sjogren syndrome
1 (1.5%)
 Herpes keratitis
1 (1.5%)

Intra-device repeatability

Complete results are shown in Table 2. All repeatability SD ratios for corneal curvature and thickness measurements were less than 1, which indicates that the CASIA SS-1000 is more consistent within an eye measured by the same device and the same operator, especially the corneal thickness measures (0.22–0.30). The repeatability of anterior corneal curvature between the two instruments were similar (ratio = 0.86 and 0.85 for anterior flat and steep meridian curvature, respectively). All coefficients of variation (CV) (SD/mean) for corneal curvature and thickness measurements were less than 0.5% for CASIA SS-1000 and 1% for Pentacam. The repeatability for TCP(X) was similar between two instruments (SD ratio = 1.06 with CV = 48.8% for CASIA; 54.3% for Pentacam), and CASIA was more repeatable in TCP(Y) than Pentacam (SD ratio = 0.68 with CV = 35.6% for CASIA and 60.5% for Pentacam).
Table 2
Repeatability and Reproducibility
Parameter
Mean
Inter-device
Repeatability
(Intra-device)
Reproducibility
(Intra-device + Inter-device)
Repeatability
CV (%)
Reproducibility
CV (%)
CASIA
Pent
P
CASIA (SD)
Pent
(SD)
CASIA (SD)
Pent
(SD)
Ratio
CASIA (SD)
Pent
(SD)
Ratio
CASIA
Pent
CASIA
Pent
AKf (D)
43.29
43.29
0.71
0.06
0.21
0.14
0.17
0.86
0.16
0.27
0.58
0.3
0.4
0.4
0.6
AKs (D)
44.17
44.18
0.39
0.10
0.17
0.16
0.18
0.85
0.18
0.25
0.73
0.4
0.4
0.4
0.6
PKf (D)
51.01
52.14
< 0.001
0.06
0.57
0.16
0.38
0.43
0.17
0.69
0.25
0.3
0.7
0.3
1.3
PKs (D)
53.52
54.86
< 0.001
0.10
0.55
0.25
0.41
0.61
0.27
0.69
0.39
0.5
0.8
0.5
1.3
TCT (μm)
527.48
538.71
< 0.001
2.02
10.07
1.06
4.91
0.22
2.28
11.2
0.20
0.2
0.9
0.4
2.1
TCP(X) (mm)
0.41
0.35
0.001
0.06
0.13
0.20
0.19
1.06
0.21
0.23
0.92
48.8
54.3
51.2
65.7
TCP(Y) (mm)
−0.45
−0.38
0.004
0.07
0.24
0.16
0.23
0.68
0.17
0.33
0.52
35.6
60.5
37.8
86.8
CCT (μm)
533.68
544.13
< 0.001
1.83
9.70
1.03
4.26
0.24
2.10
10.59
0.20
0.2
0.8
0.4
1.9
PCT2 (μm)
543.69
554.96
< 0.001
1.93
10.26
1.23
4.69
0.26
2.29
11.28
0.20
0.2
0.8
0.4
2.0
PCT4 (μm)
565.68
582.89
< 0.001
2.00
11.60
1.44
5.32
0.27
2.46
12.76
0.19
0.3
0.9
0.4
2.2
PCT6 (μm)
602.03
632.55
< 0.001
2.36
12.51
1.96
6.62
0.30
3.07
14.16
0.22
0.3
1.0
0.5
2.2
SD Standard deviation, AKf Anterior curvature flat axis, AKs Anterior curvature steep axis, PKf Posterior curvature flat axis, PKs Posterior curvature steep axis, TCT Thinnest corneal thickness, TCP(X) Horizontal distance from the apex to the thinnest corneal point with positive value for the temporal direction and negative value for the nasal direction, TCP(Y) Vertical distance from the apex to the thinnest corneal point, CCT Central corneal thickness, PCT Peripheral corneal thickness, Pent Pentacam

Inter-device reproducibility

All reproducibility SD ratios were less than 1, indicating the CASIA SS-1000 is more consistent (Table 2). The ratios for corneal thickness ranged from 0.19–0.22, indicating the Pentacam’s inter-[device+examiner] variation was 5 times greater than the CASIA SS-1000. The ratios for posterior corneal curvature ranged from 0.25–0.39. All CVs for corneal thickness and curvature measurements were less than 0.5% for CASIA SS-1000 and 2.2% for Pentacam. The reproducibility of TCP(X) was similar between the two instruments (SD ratio = 0.92 and with CV = 51.2% for CASIA and 65.7% for Pentacam), while CASIA had a better reproducibility in TCP(Y) (SD ratio = 0.52 with CV = 37.8% for CASIA and 86.8% for Pentacam).

Agreement

Bland-Altman agreement plots and Deming regression plots are shown in Figs. 1 and 2, respectively. Anterior corneal curvature had excellent agreement between CASIA SS-1000 and Pentacam. The mean difference between the two instruments (see Table 2 and Fig. 1) was 0.01 D (P = 0.71) and 0.02 D (P = 0.39) for flat and steep meridian, respectively, and 95% of the differences for the flat meridian were within 0.70 D and 0.87 D for the steep meridian. The CASIA SS-1000 posterior corneal curvature was flatter by 1.13 D (P < 0.001) and 1.35 D (P < 0.001) in flat and steep meridian, respectively, compared to the Pentacam. The widths of LOAs for posterior corneal curvatures were about 2.5 times wider than the LOA widths for anterior curvatures.
Corneal thickness measured by CASIA SS-1000 was thinner compared with the values obtained with the Pentacam in various locations. The mean difference in TCT between CASIA SS-1000 and Pentacam was − 11.20 μm (P < 0.001) with LOA of [− 34.9 12.5]. The mean difference in CCT between CASIA SS-1000 and Pentacam was − 10.45 μm (P < 0.001) with LOA of [− 33.75, 12.85]. The differences and LOA increased when the location of measurements was away from the central cornea (Fig. 1). The mean difference in TCP(X) and TCP(Y) between CASIA SS-1000 and Pentacam was 0.06 mm (P = 0.001) and − 0.07 mm (P = 0.004), respectively, and 95% of the differences for TCP(X) were within 0.70 mm, which was smaller than 1.07 mm for TCP(Y) (Fig. 1). Further, Bland-Altman plots showed no special distribution patterns, such as significant trend in mean difference or varying the LOA width.
Deming regression plots (Fig. 2) showed that the all slopes of Deming regression lines (blue lines) were parallel to the red lines (CASIA SS-1000 = Pentacam or slope = 1), except flat and steep posterior corneal curvatures and TCP(X). The slope of flat and steep was 0.83 and 0.87, respectively, which indicates that CASIA SS-1000 measurements are smaller than Pentacam when the posterior curvature increases. The slope of TCP(X) was 1.42 and intercepted with the line with slope = 1 around the apex (TCP(X) = 0), which implies that CASIA SS-1000 measured farther away from the apex than Pentacam. However, the difference may not be clinically significant. A linear conversion between measurements, taken by CASIA SS-1000 and Pentacam, was derived from Deming regression analysis (Table 3).
Table 3
Conversion equation between CASIA SS-1000 and Pentacam from Deming Regression Analysis
Measurement
Conversion Equation by
Deming Regression Equation
AKf (D)
CASIA = 0.032 + 0.999 x Pentacam
AKs (D)
CASIA = 0.717 + 0.983 x Pentacam
PKf (D)
CASIA = 7.905 + 0.827 x Pentacam
PKs (D)
CASIA = 5.758 + 0.870 x Pentacam
TCT (μm)
CASIA = 6.88 + 0.97 x Pentacam
TCP(X) (mm)
CASIA = -0.092 + 1.417 x Pentacam
TCP(Y) (mm)
CASIA = -0.048 + 1.051 x Pentacam
CCT (μm)
CASIA = 10.87 + 0.96 x Pentacam
PCT2 (μm)
CASIA = −4.63 + 0.99 x Pentacam
PCT4 (μm)
CASIA = −25.58 + 1.01 x Pentacam
PCT6 (μm)
CASIA = −3.96 + 0.96 x Pentacam
AKf Anterior curvature flat axis, AKs Anterior curvature steep axis, PKf Posterior curvature flat axis, PKs Posterior curvature steep axis, TCT Thinnest corneal thickness, TCP(X) Horizontal distance from the apex to the thinnest corneal point with positive value for the temporal direction and negative value for the nasal direction, TCP(Y) Vertical distance from the apex to the thinnest corneal point, CCT Central corneal thickness, PCT Peripheral corneal thickness

Discussion

This study aims to evaluate the repeatability, reproducibility and agreement of the CASIA SS-1000 and Pentacam in measuring the anterior segment parameters in an eye, including corneal curvature and corneal thickness maps. Of the eight parameters studied, flat and steep meridians of the anterior corneal curvature (AKf and AKs) and CCT/TCT are the most important parameters in making clinical decisions.

Anterior corneal curvatures

Anterior curvatures, along with axial length, are the key factors in calculating IOL power. A difference of more than 1 D in anterior corneal curvature is considered as clinically significant. The repeatability and reproducibility were excellent (SDs < 0.3 D). However, the CASIA SS-1000 had a better reproducibility in measuring flat axis (ratio = 0.58). The average difference in AKf and AKs between the two devices was 0.01 D and − 0.02 D, respectively. LOA limits are within 1 D. Thus, the two instruments are substantially equivalent, however, the CASIA SS-1000 had more consistent results.

Thinnest and central corneal thickness

TCT and CCT are important parameters for preoperative refractive surgery screening. In refractive surgery, based on the Munnerlyn formula [12], every 12 μm of ablation depth will correct 1 D of the spherical equivalent for a single 6 mm ablation zone. Depending on the lasers and ablation zones, the average corneal ablation depth ranged from 12 to 17 μm per diopter [12]. In addition, sufficient thickness of both flap and residual posterior stroma is required to prevent irregular astigmatism and to avoid diseases of corneal integrity and/or postoperative corneal ectasia. In general, a minimum of 110 μm flap will be created, leaving a minimum of 250 μm of residual posterior stroma. Thus, the accuracy of TCT and CCT are extremely important for thin corneas and high-myopia corrections, which require greater ablation depth. An overestimate of TCT and CCT during refractive surgery screening may result in postoperative complications. Therefore, an overestimate of 12–17 μm is considered clinically significant.
CASIA SS-1000 is 4x (repeatability ratio = 0.22–0.24) more precise than Pentacam when repeating TCT and CCT measurements using the same device, and 5x (repeatability ratio = 0.20) more precise than Pentacam when repeating TCT and CCT measurements using different devices.
The mean difference in CCT between CASIA SS-1000 and Pentacam of − 10.45 μm (~ 2% of average CCT for all participants) with LOA ranging 45 μm (− 33.75, 12.85) could be considered clinically significant. This difference may lead ophthalmologists to make a safer (more conservative) decision for refractive surgery. However, it may disqualify some patients who would be qualified if measured by Pentacam. The two measurements can be converted by the following equation: CASIA = 10.87 + 0.97 x Pentacam. For example, an eye measured 500 μm CCT by Pentacam can be converted to CASIA measurement as 490.87 μm. Therefore, decreasing the clinical difference in the values can be obtained. Similarly, TCT measurements can be converted by CASIA = 6.88 + 0.97 Pentacam. However, the decreased variability in measurements obtained with the CASIA SS-1000 results in more consistent clinical decisions.
In addition, after reviewing 94 papers published between 2000 and 2014, Rozema et al. reported that of the 14 instruments in which Pentacam was compared, five significantly overestimated CCT in normal eyes [13]. These included TMS-5 (Tomey Corp, Nagoya, Japan, by 17 μm), Orbscan with AF (Bausch & Lomb, Claremont, CA, by 11 μm), Visante/Stratus (Carl Zeiss, Dublin, CA, by 14 μm), SL-OCT (Heidelberg Engineering, Heidelberg, Germany, by 16 μm), and specular microscopes (by 20 μm). Other studies showed similar results, with Pentacam overestimating by an average of 17 μm, compared to Visante OMNI (Carl Zeiss, Dublin, CA) (17.8 ± 11.3 μm) [14], and Lenstar (Haag-Streit AG, Koeniz, Switzerland) (17.1 ± 8.5 μm) [15].

Posterior corneal curvature

Recent studies have shown that posterior corneal curvature, especially maximal point of elevation over the best-fit reference image, was statistically significant for keratoconus [16]. The ability to detect subclinical keratoconus earlier has important implications for improving refractive surgical screening to ensure that patients do not develop subsequent ectasia [16]. In theory, flat and steep posterior corneal curvature (PKf and PKs) may provide useful information for more accurate toric IOL calculations. However, measured vs. assumed posterior corneal curvatures have shown no difference in IOL calculation accuracy [17]. Our study showed that the CASIA SS-1000 underestimated the posterior corneal curvature comparing to the Pentacam. It is difficult to know what implications that an underestimate will have on screening for early keratoconus. The topographic morphologic distribution of the posterior corneal surface, specifically looking at the superior and inferior asymmetry within one topographic image, is a better early indicator of subclinical keratoconus than the raw absolute value of aggregate posterior corneal curvature being higher. Further studies are needed to analyze the topographic map of the posterior surface—and not just a solitary absolute value of the mean corneal curvature.
The absolute value of the posterior surface being smaller will affect the IOL calculation if the posterior corneal curvature value is used to calculate the total corneal power. However, recent studies using the Barret toric calculator showed that an estimated posterior curvature is just as accurate as a measured corneal curvature when looking at post-cataract refractive outcomes. As the Barrett formula is the current standard for toric IOL calculations, the difference in absolute value measurement of the posterior corneal surface between the Pentacam and CASIA will have little clinical difference in accuracy when correcting astigmatism, as the assumption has shown equivalent results to measured values. The anterior corneal surface measurements between the Pentacam and CASIA, which are more important in corneal measurement when performing IOL calculations, were not statistically different.

Peripheral corneal thickness

The corneal thickness and location data has less variability for the CASIA, compared to the Pentacam, in both repeatability and reproducibility. The CASIA corneal thickness measurements are also consistently lower in value than the Pentacam corneal thickness measurement in every point in the periphery: PCT2, PCT4, and PCT6. The pattern/distribution of corneal thickness (pachymetry map) and the location of the thinnest point in the corneal is important in planning for refractive surgery and detecting corneal diseases, such as keratoconus [18]. Thus, less variability in both the corneal thickness measurements and location of thinnest corneal point are all very important clinical measurements for screening laser refractive surgery patients.

Conclusion

This paper is the first to compare the inter-user repeatability and inter-device reproducibility of the Pentacam and CASIA SS-1000. The instruments were found to be substantially equivalent for planning IOL implantation surgery. Measurements in corneal thickness, although not identical, can be adjusted by the formula reported herein. Additionally, the variability is considerably less with the CASIA SS-1000 and, thus, offers more consistent results, an important factor in laser refractive surgery.

Acknowledgements

The authors thank Troy Brown for editorial assistance.

Declarations

This prospective, single-center study was approved by the Institutional Review Board/Ethics at Ehime University Hospital, Shitsukawa, Toon, Ehime, Japan, and was registered on clinicaltrials.​gov (NCT02077790) on 04/03/2014. All research adhered to the tenets of the Declaration of Helsinki and was HIPAA compliant. Informed consent was obtained from each participant.
Not applicable.

Competing interests

KO and MT are employees of Tomey Corp. The other authors have no relevant conflicts of interest.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Ramos JL, Li Y, Huang D. Clinical and research applications of anterior segment optical coherence tomography-a review. Clin Exp Ophthalmol. 2009;37(1):81–9.CrossRef Ramos JL, Li Y, Huang D. Clinical and research applications of anterior segment optical coherence tomography-a review. Clin Exp Ophthalmol. 2009;37(1):81–9.CrossRef
2.
Zurück zum Zitat Pang CE, Vanathi M, Tan D, Mehta J. Evaluation of corneal epithelial healing under contact lens with spectral domain anterior segment optical coherence tomography. Open Ophthalmol J. 2011;5:51–4.CrossRef Pang CE, Vanathi M, Tan D, Mehta J. Evaluation of corneal epithelial healing under contact lens with spectral domain anterior segment optical coherence tomography. Open Ophthalmol J. 2011;5:51–4.CrossRef
3.
Zurück zum Zitat Hurmeric V, Yoo SH, Mutlu FM. Optical coherence tomography in cornea and refractive surgery. Expert Rev Ophthalmol. 2012;7(3):241–50.CrossRef Hurmeric V, Yoo SH, Mutlu FM. Optical coherence tomography in cornea and refractive surgery. Expert Rev Ophthalmol. 2012;7(3):241–50.CrossRef
4.
Zurück zum Zitat Wegener A, Laser-Junga H. Photography of the anterior eye segment according to Scheimpflug's principle: options and limitations – a review. Clin Exp Ophthalmol. 2009;37(1):144–54.CrossRef Wegener A, Laser-Junga H. Photography of the anterior eye segment according to Scheimpflug's principle: options and limitations – a review. Clin Exp Ophthalmol. 2009;37(1):144–54.CrossRef
5.
Zurück zum Zitat Nakagawa T, Maeda N, Higashiura R, et al. Corneal topographic analysis in patients with keratoconus using 3-dimensional anterior segment optical coherence tomography. J Cataract Refract Surg. 2011;37(10):1871–8.CrossRef Nakagawa T, Maeda N, Higashiura R, et al. Corneal topographic analysis in patients with keratoconus using 3-dimensional anterior segment optical coherence tomography. J Cataract Refract Surg. 2011;37(10):1871–8.CrossRef
6.
Zurück zum Zitat Amano S, Honda N, Amano Y, et al. Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology. 2006;113(6):937–41.CrossRef Amano S, Honda N, Amano Y, et al. Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology. 2006;113(6):937–41.CrossRef
7.
Zurück zum Zitat Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol. 2006;141(1):7–12.CrossRef Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol. 2006;141(1):7–12.CrossRef
8.
Zurück zum Zitat McAlinden C, Khadka J, Pesudovs K. A comprehensive evaluation of the precision (repeatability and reproducibility) of the Oculus Pentacam HR. Invest Ophthalmol Vis Sci. 2011;52(10):7731–7.CrossRef McAlinden C, Khadka J, Pesudovs K. A comprehensive evaluation of the precision (repeatability and reproducibility) of the Oculus Pentacam HR. Invest Ophthalmol Vis Sci. 2011;52(10):7731–7.CrossRef
9.
Zurück zum Zitat Chen S, Huang J, Wen D, et al. Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry. Acta Ophthalmol. 2012;90(5):449–55.CrossRef Chen S, Huang J, Wen D, et al. Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry. Acta Ophthalmol. 2012;90(5):449–55.CrossRef
10.
Zurück zum Zitat Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005;31(9):1729–35.CrossRef Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005;31(9):1729–35.CrossRef
11.
Zurück zum Zitat Chan TCY, Biswas S, Yu M, Jhanji V. Longitudinal evaluation of cornea with swept-source optical coherence tomography and Scheimpflug imaging before and after Lasik. Medicine. 2015;94(30):e1219.CrossRef Chan TCY, Biswas S, Yu M, Jhanji V. Longitudinal evaluation of cornea with swept-source optical coherence tomography and Scheimpflug imaging before and after Lasik. Medicine. 2015;94(30):e1219.CrossRef
12.
Zurück zum Zitat Munnerlyn CR, Koons SJ, Marshall J. Photorefractive keratectomy: a technique for laser refractive surgery. J Cataract Refract Surg. 1988;14(1):46–52.CrossRef Munnerlyn CR, Koons SJ, Marshall J. Photorefractive keratectomy: a technique for laser refractive surgery. J Cataract Refract Surg. 1988;14(1):46–52.CrossRef
13.
Zurück zum Zitat Rozema JJ, Wouters K, Mathysen DG, Tassignon MJ. Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices. Am J Ophthalmol. 2014;158(6):1111–1120.e1.CrossRef Rozema JJ, Wouters K, Mathysen DG, Tassignon MJ. Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices. Am J Ophthalmol. 2014;158(6):1111–1120.e1.CrossRef
14.
Zurück zum Zitat Viswanathan D, Kumar NL, Males JJ, Graham SL. Comparative analysis of corneal measurements obtained from a Scheimpflug camera and an integrated Placido-optical coherence tomography device in normal and keratoconic eyes. Acta Ophthalmol. 2015;93(6):e488–94.CrossRef Viswanathan D, Kumar NL, Males JJ, Graham SL. Comparative analysis of corneal measurements obtained from a Scheimpflug camera and an integrated Placido-optical coherence tomography device in normal and keratoconic eyes. Acta Ophthalmol. 2015;93(6):e488–94.CrossRef
15.
Zurück zum Zitat Huerva V, Ascaso FJ, Soldevila J, Lavilla L. Comparison of anterior segment measurements with optical low-coherence reflectometry and rotating dual Scheimpflug analysis. J Cataract Refract Surg. 2014;40(7):1170–6.CrossRef Huerva V, Ascaso FJ, Soldevila J, Lavilla L. Comparison of anterior segment measurements with optical low-coherence reflectometry and rotating dual Scheimpflug analysis. J Cataract Refract Surg. 2014;40(7):1170–6.CrossRef
16.
Zurück zum Zitat Golan O, Hwang ES, Lang P, et al. Differences in posterior corneal features between Normal corneas and subclinical Keratoconus. J Refract Surg. 2018;34(10):664–70.CrossRef Golan O, Hwang ES, Lang P, et al. Differences in posterior corneal features between Normal corneas and subclinical Keratoconus. J Refract Surg. 2018;34(10):664–70.CrossRef
17.
Zurück zum Zitat Skrzypecki J, Patel MS, Suh LH. Performance of the Barrett Toric calculator with and without measurements of posterior corneal curvature. Eye (Lond). 2019;33(E11):1762–7.CrossRef Skrzypecki J, Patel MS, Suh LH. Performance of the Barrett Toric calculator with and without measurements of posterior corneal curvature. Eye (Lond). 2019;33(E11):1762–7.CrossRef
18.
Zurück zum Zitat Ambrósio R Jr, Caiado AL, Guerra FP, et al. Novel pachymetric parameters based on corneal tomography for diagnosing keratoconus. J Refract Surg. 2011;27(10):753–8.CrossRef Ambrósio R Jr, Caiado AL, Guerra FP, et al. Novel pachymetric parameters based on corneal tomography for diagnosing keratoconus. J Refract Surg. 2011;27(10):753–8.CrossRef
Metadaten
Titel
Comparison between the CASIA SS-1000 and Pentacam in measuring corneal curvatures and corneal thickness maps
verfasst von
Robert M. Feldman
Gene Kim
Alice Z. Chuang
Atsushi Shiraishi
Keiichiro Okamoto
Makoto Tsukamoto
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2023
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-023-02768-w

Weitere Artikel der Ausgabe 1/2023

BMC Ophthalmology 1/2023 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Die überschießende Wundheilung in der filtrierenden Glaukomchirurgie ist ein zentraler Faktor für ein operatives Versagen. Nach der Einführung der Trabekulektomie in den 1960er-Jahren wurden viele Faktoren erkannt, die mit einer vermehrten …

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.