Comparison of force loss during sliding of low friction and conventional TMA orthodontic archwires

An in vitro study

verfasst von: Nouf Alsabti, DScD candidate, Professor Christoph Bourauel, Professor Nabeel Talic

Erschienen in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Ausgabe 4/2021

Abstract

Objective

The goal was to measure and compare the amount of force loss during tooth movement guided by archwires, including a newly introduced low-friction titanium molybdenum alloy (TMA), conventional TMA, and stainless steel archwires.

Methods

The force loss was measured using a specialized biomechanical set-up, the orthodontic measurement and simulation system (OMSS). A total of 30 specimen were used (10 low-friction TMA (TMA-Low), 10 conventional TMA (TMA-C), and 10 stainless steel (SS) archwires, each having a dimension of 0.016 × 0.022 inches). The conventional and low friction TMA archwires served as test groups, while the SS archwires served as the control group.

Results

The mean values of force loss between the three types of wires (TMA‑C, TMA-Low, and SS) were significantly different (p < 0.0001). The highest mean force loss during sliding movement was found in the conventional TMA group (72.1%), followed by low friction TMA (48.8%) and stainless steel wires (33.7%) in a descending order.

Conclusion

The friction property of the low friction TMA archwire was superior to the conventional TMA archwire but was still inferior to the stainless steel archwire.

Introduction

Sliding mechanics for orthodontic space closure is a convenient technique that is widely accepted by many orthodontic clinicians. Thus, knowing the role of friction during treatment is a subject of concern for many orthodontists, as the friction generated between the bracket–archwire interface is integral to the time required for space closure. In orthodontics, friction is considered a clinical challenge especially if sliding mechanics are utilized and it must be controlled efficiently to ensure optimum orthodontic outcomes [1, 11, 24]. Friction is defined as the force that resists movement of one surface against another when two surfaces come into contact [8, 13]. When two surfaces slide, one over the other, an interacting force with two components results: a force perpendicular to the contacting surface, which is the actual force applied (N), and a force tangential to the contacting surface, which is the frictional force (F). The ratio of the tangential force to the applied force is called the coefficient of friction (μ), and it varies according to the nature of the two surfaces and their superficial characteristics. The magnitude of the frictional force is directly proportional to the applied force and dependent on the coefficient of friction such that F = μN [8, 13]. When friction-based mechanics (sliding mechanics) are utilized, a frictional force is generated between the bracket, archwire, and ligature, impeding tooth movement during the retraction phase and transmitting forces to the posterior teeth, thus, negatively affecting the anchorage requirement and resulting in potential loss of anchorage [13]. A percentage of the orthodontic force applied to teeth is lost due to friction when the orthodontic wire slides through the bracket slot and tubes, especially during the space closure stage [21]. It has been estimated that the percentage of the applied force that is lost due to friction is approximately 50%, which means that the total force applied by the clinician has to be twice as large in order to deliver an effective force in the absence of friction [23]. However, the application of heavy forces is nonproductive due to the outcomes of more friction and the risk of anchorage loss [15, 20].
TMA archwires generally have higher frictional properties when compared to other selected alloys which could be attributed to the high reactivity of the wire surface and the production of adhesive and abrasive wear [16]. Stainless steel (SS) archwires are widely used and known for their low friction properties compared to other orthodontic archwire materials [28]. Conventionally, the use of a combination of SS archwires and brackets has been the gold standard for the orthodontists to utilize during sliding mechanics. SS archwires are also considered to be the reference material for researchers assessing the mechanical properties of newly introduced archwires [26]. Given their strength, rigidity, and low-friction smooth-surface characteristics, SS archwires remain the mainstay of orthodontic mechanotherapy [14].
Over the past 100 years, developments in both mechanotherapy and treatment philosophy have resulted in huge improvements in all aspects of patient care. New advances have led to the development and introduction of new orthodontic materials with different properties, adding versatility to orthodontic treatment. Archwire materials comprise a significant part of this change and knowing which wire is appropriate for treatment requires a comprehensive understanding of the archwire biomechanical properties for certain clinical applications. Low-friction TMA wires were introduced in 2014 by the company Ormco to improve the properties of the conventional TMA wires. The company claimed that they used technical innovations to develop new low friction materials by altering the surface treatment and refining the TMA manufacturing process. By changing the processing at the vendor the surface finish was optimized and the frictional properties were improved. These low friction archwires seem to be beneficial to minimize the amount of friction in specific clinical applications. However, the cost for these materials is still considerably higher than for the traditionally used materials and their real cost–benefit ratio remains scientifically questionable.
The aim of this study was to measure and compare the amount of force loss in tooth movement guided by archwires, including a newly introduced low-friction titanium molybdenum alloy (TMA), conventional TMA, and stainless steel archwires.

Materials and methods

Ethical approval was obtained from King Saud University, College of Dentistry Research Center (PR 0073).

Sample description

A total of 30 archwires (10 low-friction TMA (TMA-Low), 10 conventional TMA (TMA-C), and 10 stainless steel (SS), from Ormco, Orange, CA, USA; 0.016 × 0.022″ (inch) each) were used. The conventional and low friction TMA archwires served as the test groups, while the stainless steel archwire served as the control group. Stainless steel brackets (0.018″, Roth prescription) were used. The brackets were conventionally ligated to the different archwires using stainless steel ligatures (Table 1).
Table 1
Sample description
Tab. 1
Beschreibung der Proben
Appliance
Product
Dimension
(inches)
Manufacturer/Location
Archwire
Stainless steel archwire
Package of Kleen Pack™ system
0.016 × 0.022″
(Preformed)
(Ormco, Glendora, CA, USA)
Conventional titanium molybdenum archwire
TMA®
Package of Kleen Pack™ system
0.016 × 0.022″
(Preformed)
(Ormco, Glendora, CA, USA)
Low friction titanium molybdenum archwire
Low Friction TMA®
Package of Kleen Pack™ system
0.016 × 0.022″
(Preformed)
(Ormco, Glendora, CA, USA)
Bracket
Stainless steel bracket, (Roth prescription)
Victory series
0.018″
(3M Unitek, Monrovia, CA, USA)
Ligature
Stainless steel ligature
0.010″
(Ormco, Glendora, CA, USA)
TMA titanium molybdenum alloy

Experimental setup

A resin replica model of a normally aligned upper arch (Palavit G 4004, Heraeus Kulzer, Hanau, Germany) constructed from a Frasaco model (Franz Sachs, Tettnang, Germany) was used. A normally aligned upper arch model was selected because the space closure and sliding stage typically starts when complete leveling and alignment of teeth is achieved. Stainless steel brackets (0.018″, Roth prescription) were used and bonded to the resin model from the second premolar on one side to the second premolar on the opposite side using cyanoacrylate adhesive. All teeth were bonded except for the extracted canine and first premolar teeth on the left side. For standardization purpose all the bonding procedures were carried out by one investigator. Following the complete bonding procedure, the model was mounted on the Orthodontic Measurement and Simulation System (OMSS; Fig. 1). The OMSS comprises two force–moment sensors mounted on motor-driven positioning tables with three-dimensional mobility, thus, enabling the system to simultaneously measure forces and moments in all three spatial planes acting on the bracket. All the mechanical components of the system are built in a temperature-controlled chamber which is especially important when testing temperature-dependent wires. The components are connected to a personal computer, through which orthodontic tooth movement simulation directions are delivered [9]. The OMSS utilizes a personal computer connected to two microcomputer-based sensors to deliver the commands needed to initiate the experiment and the digital output of the force–moment vectors to the control software.
A bracket holder was fixed to the first sensor where the upper left canine bracket was bonded. The sensor was then adjusted in the position of the upper left canine in the resin model. After that, the wire was ligated to the brackets. All the wires were ligated by the same investigator. Stainless steel ligatures (0.010″) were used for ligation using a needle holder. During ligation of the archwire to the canine bracket, the ligature was initially ligated tightly around the canine bracket wings then loosened by 90° to reduce the pressure generated by the ligature on the wire and permit free movement of the bracket along the archwire. After that, the initial forces and moments generated by the guiding arch on the canine bracket were recorded by the sensor and manually adjusted to be as close to zero as possible.
Over the second sensor, a closed Sentalloy nickel-titanium (NiTi) coil spring (GAC, Central Islip, NY, USA) was attached and hooked over the upper left canine bracket to apply about 1 N of retraction force (approximately 101 g, Fig. 1). The center of resistance of the retracted canine was set by the system software at 8.5 mm apical to the bracket and 4.5 mm lingual to it [18, 27]. The location of the center of resistance in relation to the bracket critically affects the movement of the tooth by producing moments of forces. The software utilized by the OMSS automatically calculates these moments from the measured force system at the bracket and transforms this into the simulated tooth movement that is subsequently executed by the positioning tables of the OMSS.
The OMSS then performed the simulation of the canine sliding movement by dividing it into a maximum of 200 movement increments along an approximately 4 mm retraction path. Each increment consisted of a cycle that starts with simultaneous measurement of the forces and moments delivered by the NiTi coil spring and those acting on the canine bracket. The resultant tooth movement from these forces was then calculated by the system and the canine bracket was retracted along the premolar extraction space. The amount of force loss during each movement increment was calculated as the difference between the force applied by the coil spring and the actual force reaching the canine bracket and used to commence the movement as detailed elsewhere [9]. Wire change and readjustment of the system was performed after each successful simulation run, until all 10 specimens per group were measured.

Statistical analysis

The statistical analysis was undertaken using SPSS software (IBM SPSS Inc., version 20, Chicago, IL, USA) and the level of significance was set at P < 0.05. Assuming an effect size of f = 0.6 [7] with α = 0.05 and β = 0.20 (power 80%), the required sample size was estimated to be 10 specimens in each of the three groups. Descriptive data including the means, standard deviations, minimum and maximum readings were calculated for all groups’ comparison. The differences between the weighted means of force loss percentage were analyzed with one-way analysis of variance (ANOVA) and group differences were further analyzed with Tukey’s post hoc comparisons test.

Results

Descriptive statistics showed that the highest mean of force loss of 72.1% during sliding movement was found in the TMA‑C group followed by TMA-Low with 48.8% and SS archwires with 33.7% in descending order (Fig. 2, Table 2). The mean values of force loss across the three type of wires (TMA‑C, TMA-Low, and SS) were significantly different (p < 0.0001; Table 3). The pairwise comparisons among the three mean values indicated that the mean force loss value of TMA‑C was significantly higher than the mean values of the SS and TMA-Low wires (Table 4).
Table 2
Descriptive statistics of force loss with regard to wire type
Tab. 2
Beschreibende Statistik des Kraftverlustes in Abhängigkeit vom Drahttyp
Wire type
N
Mean
Std. deviation
Minimum
Maximum
SS
10
33.7
3.9
27.9
38.5
TMA‑C
10
72.1
12.2
53.5
86.3
TMA-Low
10
48.8
7.4
43.1
69.0
Total
30
51.5
18.0
27.9
86.3
SS stainless steel archwire, TMA‑C conventional titanium molybdenum alloy (TMA) archwire, TMA-Low low-friction TMA archwire, Std. deviation standard deviation
Table 3
One-way analysis of variance of force loss with regard to wire type
Tab. 3
Einfaktorielle Varianzanalyse des Kraftverlustes in Abhängigkeit vom Drahttyp
 
Sum of squares
Dff
Mean square
F‑value
p-value
Between groups
7464.568
2
3732.284
51.273
<0.0001
Within groups
1965.404
27
72.793
Total
9429.972
29
Table 4
Multiple comparisons of mean values of force loss with regard to wire type
Tab. 4
Mehrfachvergleiche der Mittelwerte der Kraftverluste in Abhängigkeit vom Drahttyp
(I) Type of wire
(J) Type of wire
Mean Difference (I−J) in %
Std. error
p-value
95% confidence interval
Lower bound
Upper bound
SS
TMA‑C
−38.3*
3.8
0.0001
−47.8
−28.9
TMA-Low
−15.1*
3.8
0.001
−24.5
−5.6
TMA‑C
SS
38.3*
3.8
0.0001
28. 9
47.8
TMA-Low
23.3*
3.8
0.0001
13.8
32.7
TMA-Low
SS
15.1*
3.8
0.001
5.6
24.5
TMA‑C
−23.3*
3.8
0.0001
−32.7
−13.8
*The mean difference is significant at the 0.05 level
SS stainless steel archwire, TMA‑C conventional titanium molybdenum alloy (TMA) archwire, TMA-Low low-friction TMA archwire, Std. error standard error

Discussion

Friction in orthodontic treatment is considered a clinical challenge, especially with sliding mechanics and it must be well understood and controlled since this component cannot be eliminated from the materials. Because previous studies showed that TMA archwires exhibited significantly higher frictional values during sliding mechanics when compared to stainless steel wires [2, 8, 13, 14, 25, 26], the manufacturers introduced TMA archwires with different surface treatments attempting to reduce their frictional characteristics [6].
The results of the present study showed that mean values of force loss of three archwires were significantly different from each other. The mean force loss value for the TMA‑C group was found to be significantly higher than the mean values of the other two wires (SS and TMA-Low). This is in agreement with previous studies that showed high frictional values associated with TMA wires during sliding mechanics [2, 8, 13, 14, 25, 26]. The standard deviation in the obtained TMA‑C data is considered high which is due to the extremely high numbers in the data set. This could be due to a high variability of the archwires surface finish as TMA wires tend to show different patterns of surface irregularities due to the complex manufacturing process [3, 19, 22].
The TMA-Low archwire had an intermediate value of force loss between the other two groups (TMA‑C and SS archwires). The results indicate that the amount of force loss in the TMA-Low wire was significantly lower and higher than in the TMA‑C and SS groups, respectively. Although our findings indicate that the TMA-Low archwire exhibits less force loss during sliding when compared to the TMA‑C archwire, it is still considered inferior to the stainless steel archwire. This is in accordance with a reported study that evaluated the properties of the newly introduced TiMolium alloy which is a modified TMA alloy containing both aluminum and vanadium as stabilizing agents to improve the mechanical properties of the conventional TMA alloy. It was found that this alloy provided better strength, less friction, and smoother surface features than conventional TMA alloys. However, it was still inferior to SS archwires [14]. In addition, colored TMA archwires were introduced by passing a direct electrical current through the titanium alloy in order to have an improved surface finish and to reduce the frictional characteristics of the wire [6]. Furthermore, ion implanted TMA is another modified TMA archwire that was treated by accelerating the vapor flux of ions against the archwire via an electron beam evaporator to improve the surface finish as the ions penetrate the surface [6]. Evaluating the frictional characteristics of these archwires has shown that the ion implanted TMA wires showed minimum frictional forces and most of the colored TMA archwires demonstrated frictional characteristics similar to the conventional TMA archwire with both groups having no advantage over the stainless steel archwire group [6].
Among the three mean values, the SS wire had a lower mean value of force loss when compared with the other two TMA wires (TMA‑C and TMA-Low). In another previous study, it was found that SS wires generated the least amount of friction, followed by cobalt–chromium (CoCr), nickel–titanium (NiTi), and beta-titanium (β-Ti) wires [25]. These findings are consistent with those of multiple studies that have found that SS wires produce the least amount of friction compared with other alloys [13, 17].
In our study, a stainless steel ligature was used to ligate the upper left canine bracket for two reasons. First, many authors agreed that using loosely tied stainless steel ligatures generate less friction when compared to the standard elastomeric ligatures [10, 12]. Second and more importantly, using a stainless steel ligature allowed to control the force of ligation and minimize the frictional forces generated by the ligature on the wire. By loosening the ligature by 90°, it was ensured that the force applied on the canine bracket over the first sensor equals the force generated by the attached NiTi coil over the second sensor. Equal forces on the two sensors ensured that no additional frictional forces from the steel ligature were placed over the first sensor. During upper left canine bracket ligation and NiTi coil attachment, the amount of forces in the Z direction in both sensors representing the generated and applied forces were monitored via the connected personal computer. The simulation of tooth movement was not initiated until the forces appeared equal by adjusting the sensors to eliminate any additional forces resulting from the ligation procedure.
The experimental set-up of any study has a critical effect on the resulting outcomes. Multiple previous studies evaluating frictional values have utilized a testing model designed without representing the complete dental arch [4, 5, 13]. In our study, the OMSS device was used which utilizes a testing model representing the complete dental arch. The set-up included a resin replica where archwire-guided retraction of the canine bracket was performed simulating the clinical tooth retraction during sliding mechanics in which the position of the center of resistance (CR) of the retracted bracket is located away from the bracket’s plane. The OMSS enables any anticipated static and dynamic orthodontic movement to be analyzed at the level of a two-tooth model [9]. As the measurements are dynamic, the simulated tooth resembles the clinical reaction during sliding with tipping and rotation until the wire gets deformed and produces counteracting moments. The experimental set-up used in this study allowed for easy observation of the amount of force loss among the three alloys simulating the actual clinical situation during canine retraction. As a result, the TMA‑C arch showed the highest value of force loss followed by the TMA-Low and stainless steel wires.

Conclusions

The differences of force loss during sliding movement between the three types of wire were significant, i.e., the resistance of sliding of the newly introduced TMA-Low archwire was superior to the TMA-C archwire but inferior to the SS archwire.

Acknowledgements

We thank King Abdulaziz City for Science and Technology for providing funds for this work (1-18-03-001-0041). We would also like to thank Anna Weber for providing help throughout the research project.

Funding

This study was funded by the King Abdulaziz City for Science and Technology (1-18-03-001-0041).

Compliance with ethical guidelines

Conflict of interest

N. Alsabti, C. Bourauel and N. Talic declare that they have no competing interests.

Ethical standards

Ethical approval was obtained from King Saud University, College of Dentistry Research Center, Saudi Arabia (PR 0073). The manuscript does not include studies on humans or animals. Neither written informed consent to participate nor written informed consent for publication was needed for the type of this study.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Dent – Das Online-Abo der Zahnmedizin

Online-Abonnement

Mit e.Dent erhalten Sie Zugang zu allen zahnmedizinischen Fortbildungen und unseren zahnmedizinischen und ausgesuchten medizinischen Zeitschriften.

Journal of Orofacial Orthopedics - Fortschritte der Kieferorthopädie

Print-Titel

 

• Offizielles Organ der Deutschen Gesellschaft für Kieferorthopädie (DGKFO)

• Gelistet in: Science Citation Index Expanded (SciSearch), Journal Citation Reports/Science Edition, PubMed/Medline, SCOPUS, EMBASE, Google Scholar, EBSCO, Academic OneFile, CSA Environmental Sciences, EMCare, Gale, OCLC, SCImago, Summon by ProQuest

 

Literatur
1.
Zurück zum Zitat Andreasen GF, Quevedo FR (1970) Evaluation of friction forces in the 0· 022× 0· 028 edgewise bracket in vitro. J Biomech 3:151–160CrossRef Andreasen GF, Quevedo FR (1970) Evaluation of friction forces in the 0· 022× 0· 028 edgewise bracket in vitro. J Biomech 3:151–160CrossRef
4.
Zurück zum Zitat Braun S, Bluestein M, Moore BK, Benson G (1999) Friction in perspective. Am J Orthod Dentofacial Orthop 115(6):619–627CrossRef Braun S, Bluestein M, Moore BK, Benson G (1999) Friction in perspective. Am J Orthod Dentofacial Orthop 115(6):619–627CrossRef
5.
Zurück zum Zitat Cacciafesta V, Sfondrini MF, Ricciardi A, Scribante A, Klersy C, Auricchio F (2003) Evaluation of friction of stainless steel and esthetic self-ligating brackets in various bracket-archwire combinations. Am J Orthod Dentofacial Orthop 124:395–402CrossRef Cacciafesta V, Sfondrini MF, Ricciardi A, Scribante A, Klersy C, Auricchio F (2003) Evaluation of friction of stainless steel and esthetic self-ligating brackets in various bracket-archwire combinations. Am J Orthod Dentofacial Orthop 124:395–402CrossRef
7.
Zurück zum Zitat Cohen J (1988) Statistical power analysis for the behavioural sciences, 2nd edn. Lawrence Erlbaum, Hillsdale Cohen J (1988) Statistical power analysis for the behavioural sciences, 2nd edn. Lawrence Erlbaum, Hillsdale
8.
Zurück zum Zitat Drescher D, Bourauel C, Schumacher H‑A (1989) Frictional forces between bracket and arch wire. Am J Orthod Dentofacial Orthop 96:397–404CrossRef Drescher D, Bourauel C, Schumacher H‑A (1989) Frictional forces between bracket and arch wire. Am J Orthod Dentofacial Orthop 96:397–404CrossRef
10.
Zurück zum Zitat Edwards GD, Davies EH, Jones SP (1995) The ex vivo effect of ligation technique on the static frictional resistance of stainless steel brackets and archwires. Br J Orthod 22:45–153CrossRef Edwards GD, Davies EH, Jones SP (1995) The ex vivo effect of ligation technique on the static frictional resistance of stainless steel brackets and archwires. Br J Orthod 22:45–153CrossRef
11.
Zurück zum Zitat Frank CA, Nikolai RJ (1980) A comparative study of frictional resistances between orthodontic bracket and arch wire. Am J Orthod Dentofacial Orthop 78:593–609CrossRef Frank CA, Nikolai RJ (1980) A comparative study of frictional resistances between orthodontic bracket and arch wire. Am J Orthod Dentofacial Orthop 78:593–609CrossRef
12.
Zurück zum Zitat Hain M, Dhopatkar A, Rock P (2003) The effect of ligation method on friction in sliding mechanics. Am J Orthod Dentofacial Orthop 123:416–422CrossRef Hain M, Dhopatkar A, Rock P (2003) The effect of ligation method on friction in sliding mechanics. Am J Orthod Dentofacial Orthop 123:416–422CrossRef
13.
Zurück zum Zitat Kapila S, Angolkar PV, Duncanson MG, Nanda RS (1990) Evaluation of friction between edgewise stainless steel brackets and orthodontic wires of four alloys. Am J Orthod Dentofacial Orthop 98(2):117–126CrossRef Kapila S, Angolkar PV, Duncanson MG, Nanda RS (1990) Evaluation of friction between edgewise stainless steel brackets and orthodontic wires of four alloys. Am J Orthod Dentofacial Orthop 98(2):117–126CrossRef
14.
Zurück zum Zitat Krishnan V, Kumar KJ, Rcs MO, Rcps MDO, Rcs FDS (2004) Mechanical properties and surface characteristics of three archwire alloys. Angle Orthod 74(6):825–831PubMed Krishnan V, Kumar KJ, Rcs MO, Rcps MDO, Rcs FDS (2004) Mechanical properties and surface characteristics of three archwire alloys. Angle Orthod 74(6):825–831PubMed
15.
Zurück zum Zitat Kusy RP, Whitley JQ (1997) Friction between different wire-bracketconfigurations and materials. In: Seminars in Orthodontics, vol 3, No 3. Elsevier, Amsterdam, pp 166–177 Kusy RP, Whitley JQ (1997) Friction between different wire-bracketconfigurations and materials. In: Seminars in Orthodontics, vol 3, No 3. Elsevier, Amsterdam, pp 166–177
17.
Zurück zum Zitat Kusy RP, Whitley JQ, Mayhew MJ, Buckthal JE (1988) Surface roughness of orthodontic archwires via laser spectroscopy. Angle Orthod 58:33–45PubMed Kusy RP, Whitley JQ, Mayhew MJ, Buckthal JE (1988) Surface roughness of orthodontic archwires via laser spectroscopy. Angle Orthod 58:33–45PubMed
19.
Zurück zum Zitat Neumann P, Bourauel C, Ja A (2002) Corrosion and permanent fracture resistance of coated and conventional orthodontic wires. J Mater Sci Mater Med 3:141–147CrossRef Neumann P, Bourauel C, Ja A (2002) Corrosion and permanent fracture resistance of coated and conventional orthodontic wires. J Mater Sci Mater Med 3:141–147CrossRef
20.
Zurück zum Zitat Omana HM (1992) Frictional properties of metal and ceramic brackets. J Clin Orthod 26:425–432PubMed Omana HM (1992) Frictional properties of metal and ceramic brackets. J Clin Orthod 26:425–432PubMed
22.
Zurück zum Zitat Pernier C, Grosgogeat B, Ponsonnet L, Benay G, Lissac M (2005) Influence of autoclave sterilization on the surface parameters and mechanical properties of six orthodontic wires. Eur J Orthod 27:72–81CrossRef Pernier C, Grosgogeat B, Ponsonnet L, Benay G, Lissac M (2005) Influence of autoclave sterilization on the surface parameters and mechanical properties of six orthodontic wires. Eur J Orthod 27:72–81CrossRef
23.
Zurück zum Zitat Proffit WR, Fields HW, Larson B, Sarver DM (1993) Contemporary orthodontics-e-book. Elsevier, Amsterdam Proffit WR, Fields HW, Larson B, Sarver DM (1993) Contemporary orthodontics-e-book. Elsevier, Amsterdam
27.
Zurück zum Zitat Vollmer D, Bourauel C, Maier K, Jäger A (1999) Determination of the centre of resistance in an upper human canine and idealized tooth model. Eur J Orthod 21:633–648CrossRef Vollmer D, Bourauel C, Maier K, Jäger A (1999) Determination of the centre of resistance in an upper human canine and idealized tooth model. Eur J Orthod 21:633–648CrossRef
Metadaten
Titel
Comparison of force loss during sliding of low friction and conventional TMA orthodontic archwires
An in vitro study
verfasst von
Nouf Alsabti, DScD candidate
Professor Christoph Bourauel
Professor Nabeel Talic
Publikationsdatum
02.12.2020
Verlag
Springer Medizin
Erschienen in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Ausgabe 4/2021
Print ISSN: 1434-5293
Elektronische ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-020-00266-y

Weitere Artikel der Ausgabe 4/2021

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 4/2021 Zur Ausgabe

Mitteilungen der DGKFO

Mitteilungen der DGKFO

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.