Introduction
Corneal refractive surgery has become an extremely popular procedure to correct ametropia, such as myopia and hyperopia [
1]. Corneal refractive surgery is mainly divided into surface refractive surgery and lamellar refractive surgery. The former mainly includes Photorefractive Keratectomy (PRK), Laser-assisted Subepithelial Keratomileusis (LASEK) and Epipolis Laser in Situ Keratomileusis (EPI-LASIK), while Laser-assisted in Situ Keratomileusis (LASIK) and Femtosecond Laser-assisted LASIK (FS-LASIK) belong to the latter [
2].
No matter what kind of refractive surgery the patient had undergone, their central corneal thicknesses (CCT) decreased, and corneal thickness affected the measured intraocular pressure (IOP) [
3]. Therefore, the accurate measurement of IOP after refractive surgery is one of the most challenging problems. In addition, there are risks of steroid-induced glaucoma and secondary keratoconus after corneal refractive surgery, so it is of great clinical significance to accurately measure postoperative IOP for the diagnosis and treatment of ophthalmology [
4].
At present, there are several devices for measuring IOP, such as Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), iCare rebound tonometer (iCare RBT) and Ocular Response Analyzer (ORA) etc. [
5] Among them, the GAT (Haag Streit, Könitz, Switzerland) is regarded as the gold standard for IOP measurement [
6]. It follows the Imbert-Fick principle which is based on the relationship between IOP, the outlet force, and the applanation area to measure IOP in contact. But when Goldmann [
7] introduced this tonometer to measure IOP, he clearly pointed out the defects of his equipment, which measurement value (IOP
GAT) was inevitably affected by CCT. And then some researchers also proposed that GAT could provide a correct value for the IOP when the corneal thickness was about 520 μm [
8], its accuracy would gradually deteriorate when the corneal thickness deviated from this size. Therefore, the IOP
GAT is known to be significantly affected by CCT [
9]. ORA (Reichert, Depew, NY, USA) is a kind of noncontact tonometer. During the measurement process, the amplitude of the air pulse pressure at the corneal apex change over time, and the corneal movement is in response to increased and decreased pressure amplitude. Two air pulse pressure values (P1 and P2) are recorded at the inward and outward applanation events. Due to the indentation of the cornea by an air-puff causing a dynamic time-dependent response, ORA can provide extra information about IOP and corneal biomechanics. In the output parameters of ORA, Goldmann-correlated IOP (IOPg) is the mean of these applanation pressures (IOPg = (P1 + P2)/2). Corneal hysteresis (CH) is the difference in applanation pressures (CH = C × (P1 − P2)) and is an indication of viscous damping in the cornea. The corneal resistance factor (CRF) captures the overall viscoelastic behavior of the cornea [
10]. The correction of IOP according to CH could reduce the measurement of IOP by corneal factors, that is, Corneal-Compensated Intraocular Pressure (IOPcc) [
11]. IOPcc is claimed to measure IOP independent of CCT and takes certain biomechanical properties of cornea into account.
Many studies have been done on comparing the IOP measured by the GAT and ORA after the corneal refractive surgery. But each study is just for one or two types of refractive surgery, so the conclusions of each study are lack of integrity. Therefore, in this work, we gave a systematic review and meta-analysis on the three types of IOP (IOPcc, IOPg, IOPGAT) measured by ORA and GAT after corneal refractive surgery, and hoped to draw a more comprehensive conclusion on IOP of post operation.
Methods
Search strategy
We searched foreign language electronic databases of MEDLINE, EMBASE, Web of science, Cochrane library and Chinese electronic databases of CNKI and Wanfang. The search terms used were “ocular response analyzer” or “ORA”, “Goldmann applanation tonometer” or “GAT”, “intraocular pressure” or “IOP”. The publication period was from Jan. 2005 to Jan. 2019, and references to all of the retrieved literature were supplemented.
Two investigators (HZ and ZS) independently searched the studies, screened identified abstracts and articles in duplicate, extracted the available data from eligible studies, and assessed the study quality.
Inclusion and exclusion criteria
Studies describing IOP comparisons measured by GAT and ORA in their title or abstract were retrieved for full text review. Inclusions for analysis were restricted to: 1) study participants underwent corneal refractive surgery and IOP was measured with ORA and GAT after surgery; 2) mean and standard deviation of three IOP measurements (IOPcc, IOPg and IOPGAT) could be extracted from studies. Exclusion criteria applied were as follows: 1) studies done before 2005; 2) reviews or animal studies; 3) studies with no definite follow-up time; 4) studies comparing IOP with other conditions such as glaucoma, keratoconus, diabetes; 5) studies reported by other language (non-Chinese, non-English).
Data extraction and quality assessment
The following available data were extracted from eligible studies: the name of first author and the year of publication (name/y), country in which the study was carried out (country), study design (retrospective or prospective), the number of eyes included in the study (sample size), mean and standard deviation of age (mean age ± SD), surgical method, surgical procedure (lamellar corneal refraction surgery or surface corneal refractive surgery), postoperative follow-up time (post-op follow-up), and mean and standard deviation of IOP measurements (IOPcc, IOPg and IOPGAT) after corneal refractive surgery. Any differences in data abstraction were resolved by consensus and discussion with the other authors.
The study quality was assessed by using the Quality Assessment for Diagnostic Accuracy Studies 2 (QUADAS2) checklist [
12]. The patient selection risk of bias question 2 (“Was a case-control design avoided?”), and index test risk of bias question 2 (“If a threshold was used, was it prespecified?”) were excluded from the checklist because they did not apply to the current review [
13].
The effect was expressed by weighted mean difference (WMD) and 95% confidence interval (CI). The heterogeneity test was performed by chi-square test, and heterogeneity index (
I2) was used to assess heterogeneity quantitatively. If
p ≥ 0.05 and
I2 < 50%, multiple sets of data were considered to be homogeneous, and fixed effect model was selected for calculation and combined effect quantity. On the contrary, it was considered that there was heterogeneity, and random effect model was selected for correction [
14]. Publication bias was assessed via Egger precision-weighted linear regression. We also performed subgroup analysis and univariate meta-regression to explain possible sources of statistical heterogeneity when there were differences. The prespecified subgroups of interest were study design (retrospective compared with prospective), surgical procedure (lamellar corneal refractive surgery compared with surface corneal refractive surgery), and post-op follow-up (time ≤ 1 month compared with time > 1 month, time ≤ 3 months compared with time > 3 months). And meta-regression was also performed for these three study characteristics respectively. The quality assessment process was completed by Review Manager (version 5.3), and the rest of the analysis was performed by R programming language software (version 3.5.2).
Discussion
With the rapid development of corneal refractive surgery technology and the improvement of social living standards, more and more myopic patients choose to undergo the refractive surgery to improve their vision. Meanwhile, the importance of accurately measuring the IOP after corneal refractive surgery for guiding clinical medication and timely discovering secondary diseases is gradually recognized by more and more ophthalmologists [
1,
3]. Therefore we conducted a systematic review and meta-analysis on comparison of IOPcc, IOPg and IOP
GAT after corneal refractive surgery to gain a more comprehensive conclusion on postoperative IOP.
In the matter of comparison between IOPg and IOP
GAT after corneal refractive surgery, our result showed that there was no significant difference between IOPg and IOP
GAT. That is consistent with most previous studies [
3,
17‐
20,
22]. In addition, almost all studies have suggested that IOP
GAT is highly dependent on corneal thickness [
17,
21‐
23], so it is reasonable to believe that IOPg is also associated with corneal thickness. In terms of the comparison between IOPcc and IOP
GAT after corneal refractive surgery, our study showed that IOPcc was 2.56 mmHg higher than IOP
GAT in general, and the difference between IOPcc and IOP
GAT was statistically significant. The relationship between IOPcc and IOP
GAT after surgery is also consistent with the existing studies [
17‐
20,
24].
We all know that normal and stable IOP depends on the dynamic balance of volume of ocular contents, rate of aqueous humor production and rate of aqueous humor discharge [
25]. After refractive surgery, in spite of the cornea becomes thinner, there is no significant change in the generation and flow of aqueous humor, that is, there have little effect on the aqueous humor circulation, so, the actual IOP will not have a great change theoretically [
26]. Based on the assumption that IOP remained basically unchanged before and after refractive surgery, this study also referred to the method of previous researches [
3,
27], that was, using the difference of pre- and post-operative IOP (∆IOP) to evaluate which IOP measurements is closer to the real IOP after the surgery. Our study showed that mean-∆IOPg > mean-∆IOP
GAT > mean-∆IOPcc, and mean-∆IOP
GAT was 1.853 times that of mean-∆IOPcc. So considering the high dependence between IOP
GAT and CCT, it is reasonable to speculate that IOPcc may indeed be less dependent on corneal thickness. Previous studies [
4,
18,
19] have also mentioned lower percent change in IOPcc measured before and after surgery, and it suggests that IOPcc could partially compensate for the corneal properties of the cornea. Therefore, we inferred that IOPcc may be closer to real IOP after corneal refractive surgery than others.
Although this study was focused on the influence of corneal thickness on IOP after corneal refractive surgery, the measurement value of IOP was actually also affected by mechanical properties [
28,
29]. Corneal Visualisation Scheimpflug Technology (Corvis ST; Oculus Optikgerte GmbH, Wetzlar, Germany) is a new noncontact tonometer characterized with high-speed Scheimpflug technology, which facilitated the measurement of IOP. And the biomechanical corrected IOP (bIOP) [
30] is purported to be less dependent on biomechanical properties. At present, there are some researches to further explore and verify the IOP measurement of bIOP in healthy people [
31], glaucoma [
32] and even keratoconus patients [
31]. In addition, it has been proposed [
33] that bIOP is able to reduce the known correlation between CCT and IOP readings before and after refractive surgery, such as LASIK and small-incision lenticule extraction (SMILE) [
33]. However, as Corvis is a newly developed device in recent years, there are few studies on the measurement and comparison of IOP after refractive surgery, so it has not been included in the meta-analysis as ORA and GAT for overall comparison. However, it is believed that with the increase in the number of related literatures, it is also of great significance to conduct meta-analysis of Corvis and other tonometers after refractive surgery.
The choice of the type of corneal refractive surgery is limited by many factors such as the diopter of patient and corneal thickness [
34]. Generally, the corrected diopter of corneal lamellar refractive surgery is higher than that of surface refractive surgery. Moreover, the higher the corrected diopter is, the larger the amount of corneal cutting need to take, and the thinner the residual bed thickness of cornea are left [
35]. And relatively, the greater the amount of corneal cutting is, the more significant the changes of corneal biomechanical properties will be [
29]. IOP measurements affected by CCT and corneal mechanical properties. The stability of biomechanical properties of cornea after the refractive surgery is another possible error source in IOP measurement. In the subgroup analysis of postoperative IOPcc and IOP
GAT on surgical procedures, the heterogeneity among the data after corneal surface refractive surgery was zero, while the heterogeneity after corneal lamellar refractive surgery was 87%. Compared with the IOPcc after the corneal lamellar refractive surgery, IOPcc after corneal surface refractive surgery was relatively stable. This result reminds us that the stability of IOPcc after different refractive surgery may be related to the stability of corneal biomechanics. Therefore, it is assumed that the biomechanical properties of cornea after surface refractive surgery may be more stable than those after lamellar surgery. Of course, more clinical data are needed to further substantiate the conclusion.
The use of hormones after corneal refractive surgery is the main influencing factor of steroid-type glaucoma, so monitoring of IOP is particularly important during the postoperative use of hormones [
36]. Although the types of hormones used and the time of administration for different surgical methods are not the same, our literature-based research showed that there was no significant correlation between the follow-up time and the size of heterogeneity, regardless of the follow-up time of dividing line for 1 month or 3 months. Considering that there are many factors influencing the occurrence of glaucoma after corneal refractive surgery, the conclusion of this paper needs further verification of clinical cases.
Our reporting also had certain limitations. For instance, we did not consider the effect of sample size on the combined effect size, which might lead to the relatively dominant effect of large sample size data on the results. We also did not conduct meta-regression or subgroup analysis on the age of patients, the main reason was the corneal refractive surgery for patients with a certain age requirements. In general, 18 to 45 was the optimal age for the procedure, so the age of the patients in this study was too concentrated to be grouped again.