Corneal biomechanical properties in myopic eyes evaluated via Scheimpflug imaging

verfasst von: A-Yong Yu, Hui Shao, Anpeng Pan, Qinmei Wang, Zixu Huang, Benhao Song, Colm McAlinden, Jinhai Huang, Sisi Chen

Erschienen in: BMC Ophthalmology | Ausgabe 1/2020

Abstract

Background

To investigate the biomechanical properties of the cornea in myopic eyes using corneal visualization Scheimpflug technology (Corvis ST). The relationships between the biomechanical properties of the cornea and the degree of myopia were also investigated.

Methods

265 eyes of 265 subjects were included. Based on spherical equivalent (SE) in diopters (D), participants were divided into four groups: low myopia/control (SE: − 0.50 to − 3.00D), moderate myopia (SE: − 3.00 to − 6.00D), high myopia (SE: − 6.00 to − 10.00D) and severe myopia (SE greater than − 10.00D). Axial length (AL), anterior segment parameters, and corneal biomechanical properties were obtained with the Lenstar LS900, Pentacam HR and Corvis ST, respectively.

Results

Mean (±SD) SE was − 7.29 ± 4.31D (range: − 0.63 to − 25.75D). Mean AL was 26.31 ± 1.82 mm (range: 21.87 to 31.94 mm). Significant differences were detected within the four groups in terms of six corneal biomechanical parameters: deformation amplitude (DA), time from start until second applanation (A2-time), length of flattened cornea at the second applanation (A2-length), corneal velocity during the first and second applanation (A2-velocity), time from start to highest concavity (HC-time), and central curvature at highest concavity (HC radius). AL was positively associated with DA whereas negatively associated with A1-velocity and A2-length. SE was positively associated with A2-time, HC-time and A2-velocity, whereas negatively associated with DA. IOP was positively associated with four corneal biomechanical parameters and negatively associated with three parameters.

Conclusions

Eyes with severe myopia showed greater DA, lesser A2 time, HC time, and faster A2-velocity compared to low to high myopia. This suggests the cornea becomes weaker and more deformable with elongation of axial length with corresponding increases in myopia. DA, A2-time and A2-velocity could be useful corneal biomechanical indicators in patients with myopia.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
Corvis ST
Corneal visualization Scheimpflug technology
SE
Spherical equivalent
AL
Axial length
DA
Deformation amplitude
A2-time
Time from start until second applanation
A2-length
Length of flattened cornea at the second applanation
A2-velocity
Corneal velocity during the first and second applanation
HC-time
Time from start to highest concavity
HC radius
Central curvature at highest concavit
Ks
Steepest keratometry
Kf
Flattest keratometry
CCT
Central corneal thickness
CV
Corneal volume
ACD
Anterior chamber depth
IOP
Intraocular pressure
PD
Distance between the two peaks of the cornea at highest concavity

Background

Myopia is the most common vision condition, affecting 70–80% of the population in many East Asian countries [1]. High myopia in Taiwan increased from 26% in 1988 to 40% in 2005 [2]. Brien et al. [3] estimated that myopia and high myopia will influence nearly 5 billion people and 1 billion people, respectively, by 2050. More importantly, a series of complications such as glaucoma, cataract, myopic retinal degeneration and retinal detachment are associated with high myopia, some of which may eventually lead to blindness [4, 5]. High myopia is characterized by scleral thinning and a sclera that is elastic and deformable [68]. Whether or not similar biomechanical property alterations occur in the cornea of myopic eyes is still under debate. Assessment of the corneal biomechanical properties may be a way to understand the possible etiology of high myopia development, which is of great importance in its prevention of development and progression [9].
The development of the Corneal Visualization Scheimpflug Technology (Corvis ST, Oculus, Wetzlar, Germany), a relatively new noncontact tonometer combined with an ultra-high-speed Scheimpflug camera, makes it possible for us to intuitively assess the corneal deformation response during an air impulse indentation in vivo. Hon [10] and Nemeth [11] have investigated the repeatability and reproducibility of parameters provided by the Corvis ST. There were several studies focusing on the potential differences in corneal biomechanical parameters between myopia and emmetropia. Lee et al. [12] indicated that there were significant differences in outward applanation velocity, peak distance and deformation amplitude in high myopia, low myopia and emmetropia. However, they did not take axial length (AL) into consideration, in fact that axial length elongation is the primary reason for progressive myopia [13]. Wang et al. [14] suggested that there were significant differences between emmetropia, moderate myopia and high myopia in terms of deformation amplitude, first- and second-applanation time, and radius at highest concavity. However, Matalia et al. [15] showed that corneal deformation parameters were unaffected by myopia. These studies did not investigate the correlation between the degree of myopia and the corneal biomechanical properties in severe myopia, which have an increased risk of myopic complications and deserve more attention.
The purpose of our study was to investigate the relationship between corneal biomechanical properties and the degree of myopia using the Corvis ST in a large sample taking into consideration AL and anterior segment parameters.

Methods

Research was conducted in accordance with the Declaration of Helsinki and approved by the Ethics Committee of the Eye Hospital of Wenzhou Medical University. Informed consent was written from each subject before the examination. The study consisted of 265 eyes of 265 participants who presented to The Eye Hospital of Wenzhou Medical University seeking refractive surgery or for an optometric examination between January 2015 and December 2015. Subjects were excluded if they had a history of ocular surgery, any other ocular disease (e.g. corneal scaring, glaucoma and keratoconus) other than a refractive error. The exclusion criteria also included the presence of corticosteroid use, or systemic disease which could interfere with the eye such as diabetes mellitus and connective tissue diseases. Subjects who were pregnant or with astigmatism more than 3.00D were also excluded. Participants were required to stop wearing soft contact lenses at least 2 weeks and rigid lens at least 1 month before measurements were acquired. A clinician excluded glaucoma according IOP (intraocular pressure), optic nerve head assessment via fundus examination, mild keratoconus according to corneal topography. Based on spherical equivalent refraction (SE), subjects were divided into four groups: low myopia or control (SE from − 0.50 to − 3.00D), moderate myopia (SE from − 3.00 to − 6.00D), high myopia (SE from − 6.00 to − 10.00D) and severe myopia (SE greater than − 10.00D). In this study, all subjects had axial myopia.
All participants who met the inclusion criteria were selected randomly, with simple random sampling, and then received complete ophthalmic examinations including best visual acuity, subjective refraction, slit lamp microscope examination, fundus examination and measurements with the Pentacam HR (Oculus, Wetzlar, Germany), Lenstar LS900 (Haag-Streit AG, Koeniz, Switzerland), and Corvis ST (Oculus, Wetzlar, Germany). Only the right eye was measured to avoid the influence from high correlation between eyes. In cases were the right eye did not meet the inclusion criteria mentioned above, the left eye was included. All examinations were performed by an experienced ophthalmologist between 9:00 AM and 5:00 PM. The Corvis ST was the last device that was used, so as to eliminate the bias in corneal biomechanical variables caused by the air puff. All the instrument measurements were completed before pupil dilation. Three measurements with good quality were taken and the average value was used for statistical analysis. All corneal deformation images were reviewed immediately following each measurement and only images with an “OK” quality index were used in the analysis. If an “OK” was not achieved, the measurement acquisition was repeated until an “OK” was achieved. The Pentacam is a comprehensive anterior segment tomographer [16, 17] and the Lenstar LS900 is an optical biometer [18, 19]. The Corvis ST is a noncontact tonometer with an ultra-high-speed (4330 frames/s) Scheimpflug camera incorporated. It can dynamically record the deformation process of the cornea to an air impulse covering 8.0 mm. It automatically generates ten parameters: DA (the amount of corneal displacement at highest degree of concavity), A1-time and A2-time (time from start to the first and second applanation), A1-length and A2-length (length of flattened cornea at the first and second applanation), A1-velocity and A2-velocity (corneal velocity during the first and second applanation), HC-time (time from start to the highest concavity), PD (distance between the two peaks of the cornea at highest concavity), HC radius (radius of curvature at highest concavity). IOP values were also obtained with Corvis ST. Anterior segment parameters were measured with the Pentacam HR, including steepest keratometry (Ks), flattest keratometry (Kf), central corneal thickness (CCT), corneal volume (CV) and anterior chamber depth (ACD). AL was assessed with Lenstar LS900.

Statistical analysis

Data were analyzed using SPSS software version 17.0. Normal distribution was assessed with the Kolmogorov–Smirnov test. Stepwise multivariable linear regression analysis was calculated to study the relationship between corneal biomechanical properties and ocular characteristics. A one-way analysis of variance (ANOVA) and a Fisher’s least significant difference t-test was used to analyze the differences in corneal biomechanical properties among the four groups. A P-value of < 0.05 was considered statistically significant.

Results

Participant demographics are displayed in Table 1. This study included 265 eyes of 265 Chinese participants. Mean age was 27.02 ± 7.32 years (range: 18 to 52 years). Mean SE was − 7.29 ± 4.31D (range: − 0.63 to − 25.75D). Mean AL was 26.31 ± 1.82 mm (range: 21.87 to 31.94 mm). There were significant differences among groups in terms of refractive error and AL (P < 0.001). No significant differences were found between groups for CCT, IOP and CV (P > 0.05). Though there was a significant difference in age among four groups (P < 0.05), the difference was small. The one-sample Kolmogorov–Smirnov test showed that all data were normally distributed.
Table 1
Participant demographics
Parameters
Low myopia
(n = 49)
Moderate myopia
(n = 61)
High myopia
(n = 91)
Severe myopia
(n = 64)
P-value
Gender (male: female)
24:25
21:40
54:37
30:34
< 0.05
Age (years)
29.2 ± 8.85
24.58 ± 4.97
25.85 ± 5.98
29.34 ± 8.56
< 0.05
SE (D)
−1.92 ± 0.72
−4.64 ± 0.85
−7.84 ± 1.04
− 13.14 ± 3.27
< 0.001
Kf (D)
42.87 ± 1.83
42.49 ± 1.16
42.87 ± 1.31
43.34 ± 1.41
0.007
Ks (D)
43.85 ± 1.70
43.46 ± 1.23
44.02 ± 1.44
44.56 ± 1.41
< 0.001
CCT (μm)
537.22 ± 30.06
536.72 ± 32.95
538.92 ± 32.47
531.89 ± 31.72
0.599
CV (mm3)
60.12 ± 3.07
59.89 ± 3.49
60.57 ± 3.36
59.55 ± 3.42
0.291
ACD (mm)
3.12 ± 0.38
3.37 ± 0.31
3.25 ± 0.30
3.19 ± 0.31
< 0.001
AL (mm)
24.61 ± 1.28
25.42 ± 1.09
26.36 ± 1.13
28.39 ± 1.52
< 0.001
IOP (mmHg)
12.91 ± 2.21
13.07 ± 1.86
13.40 ± 2.39
13.67 ± 2.43
0.262
SE spherical equivalent, AL axial length, Ks steepest keratometry, Kf flattest keratometry, CCT central corneal thickness, CV corneal volume, ACD anterior chamber depth, AL Axial length, IOP intraocular pressure
Six of ten parameters (DA, A2-time, A2-velocity, A2-length, HC-time and HC-radius), were significantly different among the four groups (Table 2). DA in the low, moderate, high and severe myopia groups were 1.14 ± 0.11, 1.12 ± 0.09, 1.13 ± 0.10 and 1.18 ± 0.12 mm, respectively. SE was negatively correlated with AL (r = − 0.776, P < 0.001). DA in the severe myopia group was significantly longer than in the low, moderate and high myopia group (P = 0.04, P = 0.001, P = 0.006, respectively). A2-time in the low myopia group was significantly longer than the severe and high myopia group (P = 0.001, P = 0.023, respectively). A2-velocity in severe myopia group was significantly greater than in the low, moderate and high myopia group (P < 0.001, P < 0.001, P = 0.008, respectively). A2-length in the high and severe myopia groups were significantly smaller than in the other two groups. HC-time in the low myopia group was significantly greater than in the moderate, high and severe myopia groups (P = 0.02, P = 0.005, P < 0.001, respectively). HC-radius in the severe myopia group was significantly smaller than in the high myopia group. Table 3 described the coefficient (β) and P value for multivariate regression analysis. We did not find any relationship between PD and ocular parameters. AL was positively associated with DA whereas negatively associated with A1-velocity and A2-length. SE was positively associated with A2-time, HC-time and A2-velocity, whereas negatively associated with DA. IOP was positively associated with four corneal biomechanical parameters and negatively associated with three parameters. Age was positively associated with DA, A1-velocity and A2-time. CCT was positively associated with A1-length, HC radius and A2-length. ACD was negatively associated with A1-time, A2-velocity and HC-time.
Table 2
All the parameters acquired by the Corvis ST
Parameters
Low myopia
(n = 49)
Moderate myopia
(n = 61)
High myopia
(n = 91)
Severe myopia (n = 64)
P-value
A1-time (m/s)
7.07 ± 0.21
7.07 ± 0.18
7.12 ± 0.24
7.14 ± 0.26
0.212
A1-length (mm)
1.76 ± 0.16
1.73 ± 0.12
1.73 ± 0.08
1.72 ± 0.09
0.293
A1-velocity (m/s)
0.15 ± 0.01
0.15 ± 0.01
0.15 ± 0.01
0.15 ± 0.01
0.602
A2-time (m/s)
21.93 ± 0.45
21.79 ± 0.36
21.76 ± 0.43
21.66 ± 0.42
0.011
A2-length (mm)
1.72 ± 0.28
1.71 ± 0.30
1.60 ± 0.25
1.55 ± 0.28
0.001
A2-velocity (m/s)
−0.38 ± 0.08
−0.40 ± 0.07
− 0.42 ± 0.07
− 0.46 ± 0.08
< 0.001
HC-time (m/s)
16.89 ± 0.38
16.69 ± 0.35
16.72 ± 0.32
16.59 ± 0.30
< 0.001
PD (mm)
4.16 ± 0.79
4.35 ± 0.66
4.23 ± 0.72
4.24 ± 0.79
0.619
HC radius (mm)
6.51 ± 0.79
6.63 ± 0.69
7.02 ± 0.86
6.52 ± 1.02
< 0.001
DA (mm)
1.14 ± 0.11
1.12 ± 0.09
1.13 ± 0.10
1.18 ± 0.12
0.007
A1-time and A2-time: time from start until the first and second applanation; A1-length and A2-length: length of flattened cornea at the first and second applanation; A1-velocity and A2-velocity: corneal velocity during the first and second applanation; HC-time: time from start until highest concavity is arrived; PD: distance between the two peaks of the cornea at highest concavity; HC radius: central curvature at highest concavity; DA: maximum deformation amplitude at highest concavity. One-way analysis of variance (ANOVA) and LSD-t test was used to analyze the differences of corneal biomechanical parameters between the four groups. P < 0.05 was considered statistically significant
Table 3
Stepwise multivariate regression model results
Parameters
Age (years)
SE (D)
Kf (D)
Ks(D)
CCT (μm)
CV (mm3)
ACD (mm)
AL (mm)
IOP (mmHg)
β P
β P
β P
β P
β P
β P
β P
β P
β P
A1-time (m/s)
      
−0.020 0.025
 
0.989 < 0.001
A1-length (mm)
    
0.221 < 0.001
    
A1-velocity (m/s)
0.133 0.022
 
0.222 < 0.001
    
−0.156 0.004
−0.373 < 0.001
A2-time (m/s)
0.111 0.005
0.129 0.001
   
0.169 < 0.001
  
−0.789 < 0.001
A2-length (mm)
   
−0.136 0.017
0.289 < 0.001
  
−0.248 < 0.001
0.177 0.004
A2-velocity (m/s)
 
0.408 < 0.001
    
−0.119 0.008
 
0.597 < 0.001
HC-time (m/s)
 
0.289 < 0.001
0.143 0.016
   
−0.234 < 0.001
  
HC radius (mm)
    
0.207 0.001
   
0.355 < 0.001
DA (mm)
0.155 < 0.001
−0.119 0.039
 
0.084 0.026
   
0.151 0.008
−0.785 < 0.001
SE spherical equivalent, Ks steepest keratometry, Kf flattest keratometry, CCT central corneal thickness, CV corneal volume, ACD anterior chamber depth, AL Axial length, IOP intraocular pressure; A1-time and A2-time: time from start until the first and second applanation; A1-length and A2-length: length of flattened cornea at the first and second applanation; A1-velocity and A2-velocity: corneal velocity during the first and second applanation moment; HC-time: time from start until highest concavity is arrived; HC radius: central curvature at highest concavity; DA: maximum deformation amplitude at highest concavity

Discussion

The present study aimed to assess the corneal biomechanics in eyes with different degrees of myopia, especially eyes with severe myopia, using the Corvis ST with a large sample size. In the current prospective study, we found that the severe myopia group had a significantly longer DA than the low, moderate and high myopia group. Moreover, DA was negatively correlated with SE (β = − 0.119, P = 0.039) and positively correlated with AL (β = 0.151, P = 0.008). DA is the deformation amplitude at highest concavity and is regarded as a surrogate for corneal stiffness [10, 14, 20]. A higher DA value would represent a softer cornea due to a lower resistance to deformation [20, 21]. Our findings showed that the higher the degree of myopia or the longer the axial length, the softer of the cornea, which is consistent with previous studies [12, 14, 2124]. Miki et al. [24] have found similar trends but did not divide subjects into different groups according to SE or AL and had relatively few participants. In comparison, the present study included a large sample of subjects with severe myopia with rigorous screening criteria. Xiang et al. [25] suggested that eye elongates particularly fast before the onset of myopia. It is conceivable that changes in corneal and scleral biomechanics precede the occurrence of myopia. However, our data does not prove whether corneal biomechanical properties have changed with the onset of myopia. This would require a longitudinal study. In clinical practice, if a larger DA indicates a greater risk of axial elongation, preventive methods can be considered. Further, assessment of corneal biomechanics may also be crucial for surgical planning (and preoperative screening) to reduce the risk of corneal ectasia after refractive surgery.
A2-velocity in the severe myopia group was significantly greater than in the other three groups. A2-time in the high and moderate group was significantly shorter than in the low myopia group. The severe and high myopia groups had a smaller A2-length than other groups. These findings are not in line with those from Wang et al. [14], their study did not show the difference of A2-velocity between groups. Another study, consistent with our results, showed high myopia had greater A2-velocity compared to either low to moderate myopia or emmetropia using the Corvis ST [12]. A2-time and A2-velocity were positively correlated with SE, as per the ANOVA in the present study. A previous study found that A2-time may be an indicator of the total viscoelasticity of the cornea [26]. The higher the degree or the longer the AL of the myopic eye, the less time is needed to return to the origin after applanation, as its viscoelasticity increased. In a recent study [27], myopia had greater A2-velocity indicating that the cornea tends to be more deformable and softer than emmetropia and hyperopia in children. Therefore, A2-velocity may be another indicator of corneal elasticity, as Lee et al. [12], Miki et al. [24] and Wang et al. [28] suggested.
In accordance with a previous study by Wang et al. [14], we found that the HC radius was significantly smaller in eyes with severe myopia than high myopia. Frings et al. [29] demonstrated that in eyes following corneal refractive surgery, there was a tendency to have a smaller radius at highest concavity. That is, the softer the cornea, the easier it is to deform, resulting in a smaller HC-radius.
We found longer DA values with increasing age, in agreement with previous studies [14, 30]. However, others supported a view that the cornea becomes stiffer in elderly individuals because of more glycation-induced cross-linking [31, 32]. The relatively narrow range of age in this study may be a reason for the conflicting results. Further study is required to determine the influence of age on corneal biomechanics, controlling for other variables including SE. Our analysis showed a lower IOP was associated with less stiffness of the cornea with a larger DA (β = − 0.785, P < 0.001). This phenomenon is in accordance with previous studies [24]. Alonso et al. indicated that the variation in corneal deformation may depend on the IOP rather than simply affected by the corneal structure [33]. Tian et al. also found that DA was negatively correlated with IOP in both primary open-angle glaucoma (POAG) and normal eyes [34]. In other words, we should take the corneal biomechanical into consideration when it comes to the accuracy of IOP measurement.
It has been suggested that myopic eyes have a lower ocular rigidity than emmetropic eyes, [6, 7] and the diameter of collagen fiber bundles of sclera in highly myopic eyes is less and the sclera is significantly thinner [13, 35]. Less proteoglycan and glycosaminoglycan synthesis [3638] and a reduction in the extracellular matrix [39] may explain the remodeling of the sclera during the development of myopia. In this way, the scleral mechanical properties weaken [35] whilst the deformability increases [40]. Since the corneal stroma is the continuation of the sclera, [35] the expansion of the sclera may result in the reduction of corneal stiffness. If a change in corneal biomechanical properties potentially indicate a risk for myopia development, such measurements may become clinically useful in the management of myopia. On the other hand, various studies have demonstrated that corneal refractive surgery may change corneal biomechanics, [21, 29, 4143] which can lead to iatrogenic keratectasia [44]. Wang et al. [28] demonstrated that the postoperative refractive outcome of small-incision lenticule extraction (SMILE) can be predicted by using corneal topographic and biomechanical parameters (CTBPs). Therefore, a better understanding of the biomechanical characteristics of the cornea is of great importance in the field of corneal refractive surgery [45].
The main limitations of our study are as follows: first, we did not assess corneal deformation in emmetropia. Second, a longitudinal study is required to determine whether differences in corneal biomechanics is the consequence or the cause of myopia development.

Conclusions

In conclusion, the myopic eyes with a greater degree and longer AL exhibited longer DA and A2-velocity, and less A2-time under stress, suggesting that the cornea in higher myopia tends to be more elastic and deformable. Further research is warranted in this field as this may have clinical significance the management of high myopia and also in refractive surgery.

Acknowledgments

The authors thank all the funders for their support.
Research was carried out in accordance with the Declaration of Helsinki and approved by the Ethics Committee of the Eye Hospital of Wenzhou Medical University. Informed consent was written from each subject before the examination.
Not applicable.

Competing interests

The authors declare that they no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Rahi JS, Cumberland PM, Peckham CS. Myopia over the lifecourse: prevalence and early life influences in the 1958 British birth cohort. Ophthalmology. 2011;118(5):797–804.PubMed Rahi JS, Cumberland PM, Peckham CS. Myopia over the lifecourse: prevalence and early life influences in the 1958 British birth cohort. Ophthalmology. 2011;118(5):797–804.PubMed
2.
Zurück zum Zitat Wang TJ, Chiang TH, Wang TH, Lin LL, Shih YF. Changes of the ocular refraction among freshmen in National Taiwan University between 1988 and 2005. Eye (Lond). 2009;23(5):1168–9.. Wang TJ, Chiang TH, Wang TH, Lin LL, Shih YF. Changes of the ocular refraction among freshmen in National Taiwan University between 1988 and 2005. Eye (Lond). 2009;23(5):1168–9..
3.
Zurück zum Zitat Brien A, Holden TRF, Wilson DA. Global Prevalence of Myopia and High Myopia and Temporal Trends from 2000 through 2050. Ophthalmology. 2016;123:1036–42. Brien A, Holden TRF, Wilson DA. Global Prevalence of Myopia and High Myopia and Temporal Trends from 2000 through 2050. Ophthalmology. 2016;123:1036–42.
4.
Zurück zum Zitat Saw SM, Gazzard G, Shih-Yen EC, Chua WH. Myopia and associated pathological complications. Ophthalmic Physiol Opt. 2005;25(5):381–91.PubMed Saw SM, Gazzard G, Shih-Yen EC, Chua WH. Myopia and associated pathological complications. Ophthalmic Physiol Opt. 2005;25(5):381–91.PubMed
5.
Zurück zum Zitat Wong TY, Ferreira A, Hughes R, Carter G, Mitchell P. Epidemiology and disease burden of pathologic myopia and myopic choroidal neovascularization: an evidence-based systematic review. Am J Ophthalmol. 2014;157(1):9–25.PubMed Wong TY, Ferreira A, Hughes R, Carter G, Mitchell P. Epidemiology and disease burden of pathologic myopia and myopic choroidal neovascularization: an evidence-based systematic review. Am J Ophthalmol. 2014;157(1):9–25.PubMed
6.
Zurück zum Zitat Lam AKC, Wong S, Lam CSY, To CH. The effect of myopic axial elongation and posture on the pulsatile ocular blood flow in young normal subjects. Optom Vis Sci. 2002;79(5):300–5.PubMed Lam AKC, Wong S, Lam CSY, To CH. The effect of myopic axial elongation and posture on the pulsatile ocular blood flow in young normal subjects. Optom Vis Sci. 2002;79(5):300–5.PubMed
7.
Zurück zum Zitat Berisha F, Findl O, Fuchsaeger-Mayrl G, Schmetterer L. Dependence of ocular rigidity on eye length: a study comparing ocular pressure pulse and ocular fundus pulse in healthy subjects. Invest Ophthalmol Vis Sci. 2004;45:U972. Berisha F, Findl O, Fuchsaeger-Mayrl G, Schmetterer L. Dependence of ocular rigidity on eye length: a study comparing ocular pressure pulse and ocular fundus pulse in healthy subjects. Invest Ophthalmol Vis Sci. 2004;45:U972.
8.
Zurück zum Zitat Hayashi M, Ito Y, Takahashi A, Kawano K, Terasaki H. Scleral thickness in highly myopic eyes measured by enhanced depth imaging optical coherence tomography. Eye. 2013;27(3):410–7.PubMedPubMedCentral Hayashi M, Ito Y, Takahashi A, Kawano K, Terasaki H. Scleral thickness in highly myopic eyes measured by enhanced depth imaging optical coherence tomography. Eye. 2013;27(3):410–7.PubMedPubMedCentral
9.
Zurück zum Zitat Huang J, Wen D, Wang Q, McAlinden C, Flitcroft I, Chen H, Saw SM, Chen H, Bao F, Zhao Y, et al. Efficacy comparison of 16 interventions for myopia control in children: a network meta-analysis. Ophthalmology. 2016;123(4):697–708.PubMed Huang J, Wen D, Wang Q, McAlinden C, Flitcroft I, Chen H, Saw SM, Chen H, Bao F, Zhao Y, et al. Efficacy comparison of 16 interventions for myopia control in children: a network meta-analysis. Ophthalmology. 2016;123(4):697–708.PubMed
10.
Zurück zum Zitat Hon Y, Lam AK. Corneal deformation measurement using Scheimpflug noncontact tonometry. Optom Vis Sci. 2013;90(1):e1–8.PubMed Hon Y, Lam AK. Corneal deformation measurement using Scheimpflug noncontact tonometry. Optom Vis Sci. 2013;90(1):e1–8.PubMed
11.
Zurück zum Zitat Nemeth G, Hassan Z, Csutak A, Szalai E. Repeatability of ocular biomechanical data measurements with a Scheimpfl ug-based noncontact device on Normal corneas. J Refract Surg. 2013;29(8):558–63.PubMed Nemeth G, Hassan Z, Csutak A, Szalai E. Repeatability of ocular biomechanical data measurements with a Scheimpfl ug-based noncontact device on Normal corneas. J Refract Surg. 2013;29(8):558–63.PubMed
12.
Zurück zum Zitat Lee R, Chang RT, Wong IY, Lai JS, Lee JW, Singh K. Assessment of corneal biomechanical parameters in myopes and emmetropes using the Corvis ST. Clin Exp Optom. 2016;99(2):157–62.PubMed Lee R, Chang RT, Wong IY, Lai JS, Lee JW, Singh K. Assessment of corneal biomechanical parameters in myopes and emmetropes using the Corvis ST. Clin Exp Optom. 2016;99(2):157–62.PubMed
13.
Zurück zum Zitat Harper AR, Summers JA. The dynamic sclera: extracellular matrix remodeling in normal ocular growth and myopia development. Exp Eye Res. 2015;133:100–11.PubMedPubMedCentral Harper AR, Summers JA. The dynamic sclera: extracellular matrix remodeling in normal ocular growth and myopia development. Exp Eye Res. 2015;133:100–11.PubMedPubMedCentral
14.
Zurück zum Zitat Wang J, Li Y, Jin Y, Yang X, Zhao C, Long Q. Corneal biomechanical properties in myopic eyes measured by a dynamic Scheimpflug analyzer. J Ophthalmol. 2015;2015:1–8. Wang J, Li Y, Jin Y, Yang X, Zhao C, Long Q. Corneal biomechanical properties in myopic eyes measured by a dynamic Scheimpflug analyzer. J Ophthalmol. 2015;2015:1–8.
15.
Zurück zum Zitat Matalia J, Francis M, Tejwani S, Dudeja G, Rajappa N, Sinha RA. Role of age and myopia in simultaneous assessment of corneal and Extraocular tissue stiffness by air-puff Applanation. J Refract Surg. 2016;32(7):486–93.PubMed Matalia J, Francis M, Tejwani S, Dudeja G, Rajappa N, Sinha RA. Role of age and myopia in simultaneous assessment of corneal and Extraocular tissue stiffness by air-puff Applanation. J Refract Surg. 2016;32(7):486–93.PubMed
16.
Zurück zum Zitat Guber I, McAlinden C, Majo F, Bergin C. Identifying more reliable parameters for the detection of change during the follow-up of mild to moderate keratoconus patients. Eye Vis (Lond). 2017;4:24. Guber I, McAlinden C, Majo F, Bergin C. Identifying more reliable parameters for the detection of change during the follow-up of mild to moderate keratoconus patients. Eye Vis (Lond). 2017;4:24.
17.
Zurück zum Zitat McAlinden C, Khadka J, Pesudovs K. A comprehensive evaluation of the precision (repeatability and reproducibility) of the oculus Pentacam HR. Invest Ophthalmol Vis Sci. 2011;52(10):7731–7.PubMed McAlinden C, Khadka J, Pesudovs K. A comprehensive evaluation of the precision (repeatability and reproducibility) of the oculus Pentacam HR. Invest Ophthalmol Vis Sci. 2011;52(10):7731–7.PubMed
18.
Zurück zum Zitat McAlinden C, Wang Q, Pesudovs K, Yang X, Bao F, Yu A, Lin S, Feng Y, Huang J. Axial length measurement failure rates with the IOLMaster and Lenstar LS 900 in eyes with cataract. PLoS One. 2015;10(6):e0128929.PubMedPubMedCentral McAlinden C, Wang Q, Pesudovs K, Yang X, Bao F, Yu A, Lin S, Feng Y, Huang J. Axial length measurement failure rates with the IOLMaster and Lenstar LS 900 in eyes with cataract. PLoS One. 2015;10(6):e0128929.PubMedPubMedCentral
19.
Zurück zum Zitat McAlinden C, Gao R, Yu A, Wang X, Yang J, Yu Y, Chen H, Wang Q, Huang J. Repeatability and agreement of ocular biometry measurements: Aladdin versus Lenstar. Br J Ophthalmol. 2017;101(9):1223–9.PubMed McAlinden C, Gao R, Yu A, Wang X, Yang J, Yu Y, Chen H, Wang Q, Huang J. Repeatability and agreement of ocular biometry measurements: Aladdin versus Lenstar. Br J Ophthalmol. 2017;101(9):1223–9.PubMed
20.
Zurück zum Zitat Ye C, Yu M, Lai G, Jhanji V. Variability of corneal deformation response in Normal and Keratoconic eyes. Optom Vis Sci. 2015;92(7):e149–53.PubMed Ye C, Yu M, Lai G, Jhanji V. Variability of corneal deformation response in Normal and Keratoconic eyes. Optom Vis Sci. 2015;92(7):e149–53.PubMed
21.
Zurück zum Zitat Shen Y, Chen Z, Knorz MC, Li M, Zhao J, Zhou X. Comparison of corneal deformation parameters after SMILE, LASEK, and femtosecond laser-assisted LASIK. J Refract Surg. 2014;30(5):310–8.PubMed Shen Y, Chen Z, Knorz MC, Li M, Zhao J, Zhou X. Comparison of corneal deformation parameters after SMILE, LASEK, and femtosecond laser-assisted LASIK. J Refract Surg. 2014;30(5):310–8.PubMed
22.
Zurück zum Zitat Chang PY, Chang SW, Wang JY. Assessment of corneal biomechanical properties and intraocular pressure with the ocular response analyzer in childhood myopia. Br J Ophthalmol. 2010;94(7):877–81.PubMed Chang PY, Chang SW, Wang JY. Assessment of corneal biomechanical properties and intraocular pressure with the ocular response analyzer in childhood myopia. Br J Ophthalmol. 2010;94(7):877–81.PubMed
23.
Zurück zum Zitat Cong Ye MY, Lai G. Variability of Corneal Deformation Response in Normal and Keratoconic Eyes. Optom Vis Sci. 2015;92(7):149–53. Cong Ye MY, Lai G. Variability of Corneal Deformation Response in Normal and Keratoconic Eyes. Optom Vis Sci. 2015;92(7):149–53.
24.
Zurück zum Zitat Miki A, Maeda N, Ikuno Y, Asai T, Hara C, Nishida K. Factors associated with corneal deformation responses measured with a dynamic Scheimpflug analyzer. Invest Ophthalmol Vis Sci. 2017;58(1):538–44.PubMed Miki A, Maeda N, Ikuno Y, Asai T, Hara C, Nishida K. Factors associated with corneal deformation responses measured with a dynamic Scheimpflug analyzer. Invest Ophthalmol Vis Sci. 2017;58(1):538–44.PubMed
25.
Zurück zum Zitat Xiang F, He M, Morgan IG. Annual changes in refractive errors and ocular components before and after the onset of myopia in Chinese children. Ophthalmol. 2012;119(7):1478–84. Xiang F, He M, Morgan IG. Annual changes in refractive errors and ocular components before and after the onset of myopia in Chinese children. Ophthalmol. 2012;119(7):1478–84.
26.
Zurück zum Zitat Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and Keratoconic corneas measured using a noncontact Scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55(6):3651–9.PubMed Ali NQ, Patel DV, McGhee CN. Biomechanical responses of healthy and Keratoconic corneas measured using a noncontact Scheimpflug-based tonometer. Invest Ophthalmol Vis Sci. 2014;55(6):3651–9.PubMed
27.
Zurück zum Zitat Long W, Zhao YM, Hu Y. Characteristics of corneal biomechanics in Chinese preschool children with different refractive status. Cornea. 2019;00(00):1–5. Long W, Zhao YM, Hu Y. Characteristics of corneal biomechanics in Chinese preschool children with different refractive status. Cornea. 2019;00(00):1–5.
28.
Zurück zum Zitat Wang M, Zhang Y, Wu W, Young JA, Hatch KM, Pineda R 2nd, Elze T, Wang Y. Predicting refractive outcome of small incision Lenticule extraction for myopia using corneal properties. Transl Vis Sci Technol. 2018;7(5):11.PubMedPubMedCentral Wang M, Zhang Y, Wu W, Young JA, Hatch KM, Pineda R 2nd, Elze T, Wang Y. Predicting refractive outcome of small incision Lenticule extraction for myopia using corneal properties. Transl Vis Sci Technol. 2018;7(5):11.PubMedPubMedCentral
29.
Zurück zum Zitat Frings A, Linke SJ, Bauer EL, Druchkiv V, Katz T, Steinberg J. Effects of laser in situ keratomileusis (LASIK) on corneal biomechanical measurements with the Corvis ST tonometer. Clin Ophthalmol. 2015;9:305–11.PubMedPubMedCentral Frings A, Linke SJ, Bauer EL, Druchkiv V, Katz T, Steinberg J. Effects of laser in situ keratomileusis (LASIK) on corneal biomechanical measurements with the Corvis ST tonometer. Clin Ophthalmol. 2015;9:305–11.PubMedPubMedCentral
30.
Zurück zum Zitat Vellara HR, Ali NQ, Gokul A, Turuwhenua J, Patel DV, McGhee CN. Quantitative analysis of corneal energy dissipation and corneal and orbital deformation in response to an air-pulse in healthy eyes. Invest Ophthalmol Vis Sci. 2015;56(11):6941–7.PubMed Vellara HR, Ali NQ, Gokul A, Turuwhenua J, Patel DV, McGhee CN. Quantitative analysis of corneal energy dissipation and corneal and orbital deformation in response to an air-pulse in healthy eyes. Invest Ophthalmol Vis Sci. 2015;56(11):6941–7.PubMed
31.
Zurück zum Zitat Daxer A, Misof K, Grabner B, Ettl A, Fratzl P. Collagen fibrils in the human corneal stroma: structure and aging. Invest Ophthalmol Vis Sci. 1998;39(3):644–8.PubMed Daxer A, Misof K, Grabner B, Ettl A, Fratzl P. Collagen fibrils in the human corneal stroma: structure and aging. Invest Ophthalmol Vis Sci. 1998;39(3):644–8.PubMed
32.
Zurück zum Zitat Elsheikh A, Geraghty B, Rama P, Campanelli M, Meek KM. Characterization of age-related variation in corneal biomechanical properties. J R Soc Interface. 2010;7(51):1475–85.PubMedPubMedCentral Elsheikh A, Geraghty B, Rama P, Campanelli M, Meek KM. Characterization of age-related variation in corneal biomechanical properties. J R Soc Interface. 2010;7(51):1475–85.PubMedPubMedCentral
33.
Zurück zum Zitat Alonso-Caneiro D, Karnowski K, Kaluzny BJ, Kowalczyk A, Wojtkowski M. Assessment of corneal dynamics with high-speed swept source optical coherence tomography combined with an air puff system. Opt Express. 2011;19(15):14188–99.PubMed Alonso-Caneiro D, Karnowski K, Kaluzny BJ, Kowalczyk A, Wojtkowski M. Assessment of corneal dynamics with high-speed swept source optical coherence tomography combined with an air puff system. Opt Express. 2011;19(15):14188–99.PubMed
34.
Zurück zum Zitat Tian L, Wang DJ, Wu Y, Meng XL, Chen B, Ge M, Huang YF. Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmol. 2016;94(5):e317–24.PubMed Tian L, Wang DJ, Wu Y, Meng XL, Chen B, Ge M, Huang YF. Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmol. 2016;94(5):e317–24.PubMed
35.
Zurück zum Zitat McBrien NA, Gentle A. Role of the sclera in the development and pathological complications of myopia. Prog Retin Eye Res. 2003;22(3):307–38.PubMed McBrien NA, Gentle A. Role of the sclera in the development and pathological complications of myopia. Prog Retin Eye Res. 2003;22(3):307–38.PubMed
36.
Zurück zum Zitat Rada JA, Nickla DL, Troilo D. Decreased proteoglycan synthesis associated with form deprivation myopia in mature primate eyes. Invest Ophthalmol Vis Sci. 2000;41(8):2050–8.PubMed Rada JA, Nickla DL, Troilo D. Decreased proteoglycan synthesis associated with form deprivation myopia in mature primate eyes. Invest Ophthalmol Vis Sci. 2000;41(8):2050–8.PubMed
37.
Zurück zum Zitat Rada JA, Shelton S, Norton TT. The sclera and myopia. Exp Eye Res. 2006;82(2):185–200.PubMed Rada JA, Shelton S, Norton TT. The sclera and myopia. Exp Eye Res. 2006;82(2):185–200.PubMed
38.
Zurück zum Zitat Guo L, Frost MR, Siegwart JT Jr, Norton TT. Scleral gene expression during recovery from myopia compared with expression during myopia development in tree shrew. Mol Vis. 2014;20:1643–59.PubMedPubMedCentral Guo L, Frost MR, Siegwart JT Jr, Norton TT. Scleral gene expression during recovery from myopia compared with expression during myopia development in tree shrew. Mol Vis. 2014;20:1643–59.PubMedPubMedCentral
39.
Zurück zum Zitat Seko Y, Shimokawa H, Tokoro T. Expression of bFGF and TGF-beta 2 in experimental myopia in chicks. Invest Ophthalmol Vis Sci. 1995;36(6):1183–7.PubMed Seko Y, Shimokawa H, Tokoro T. Expression of bFGF and TGF-beta 2 in experimental myopia in chicks. Invest Ophthalmol Vis Sci. 1995;36(6):1183–7.PubMed
40.
Zurück zum Zitat Jones BE, Thompson EW, Hodos W, Waldbillig RJ, Chader GJ. Scleral matrix metalloproteinases, serine proteinase activity and hydrational capacity are increased in myopia induced by retinal image degradation. Exp Eye Res. 1996;63(4):369–81.PubMed Jones BE, Thompson EW, Hodos W, Waldbillig RJ, Chader GJ. Scleral matrix metalloproteinases, serine proteinase activity and hydrational capacity are increased in myopia induced by retinal image degradation. Exp Eye Res. 1996;63(4):369–81.PubMed
41.
42.
Zurück zum Zitat Chen MC, Lee N, Bourla N, Hamilton DR. Corneal biomechanical measurements before and after laser in situ keratomileusis. J Cataract Refract Surg. 2008;34(11):1886–91.PubMed Chen MC, Lee N, Bourla N, Hamilton DR. Corneal biomechanical measurements before and after laser in situ keratomileusis. J Cataract Refract Surg. 2008;34(11):1886–91.PubMed
43.
Zurück zum Zitat Kamiya K, Shimizu K, Ohmoto F. Comparison of the changes in corneal biomechanical properties after photorefractive keratectomy and laser in situ Keratomileusis. Cornea. 2009;28(7):765–9.PubMed Kamiya K, Shimizu K, Ohmoto F. Comparison of the changes in corneal biomechanical properties after photorefractive keratectomy and laser in situ Keratomileusis. Cornea. 2009;28(7):765–9.PubMed
44.
Zurück zum Zitat Klein SR, Epstein RJ, Randleman JB, Stulting RD. Corneal ectasia after laser in situ keratomileusis in patients without apparent preoperative risk factors. Cornea. 2006;25(4):388–403.PubMed Klein SR, Epstein RJ, Randleman JB, Stulting RD. Corneal ectasia after laser in situ keratomileusis in patients without apparent preoperative risk factors. Cornea. 2006;25(4):388–403.PubMed
45.
Zurück zum Zitat McAlinden C. Corneal refractive surgery: past to present. Clin Exp Optom. 2012;95(4):386–98.PubMed McAlinden C. Corneal refractive surgery: past to present. Clin Exp Optom. 2012;95(4):386–98.PubMed
Metadaten
Titel
Corneal biomechanical properties in myopic eyes evaluated via Scheimpflug imaging
verfasst von
A-Yong Yu
Hui Shao
Anpeng Pan
Qinmei Wang
Zixu Huang
Benhao Song
Colm McAlinden
Jinhai Huang
Sisi Chen
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2020
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-020-01530-w

Weitere Artikel der Ausgabe 1/2020

BMC Ophthalmology 1/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Ophthalmika in der Schwangerschaft

Die Verwendung von Ophthalmika in der Schwangerschaft und Stillzeit stellt immer eine Off-label-Anwendung dar. Ein Einsatz von Arzneimitteln muss daher besonders sorgfältig auf sein Risiko-Nutzen-Verhältnis bewertet werden. In der vorliegenden …

Operative Therapie und Keimnachweis bei endogener Endophthalmitis

Vitrektomie Originalie

Die endogene Endophthalmitis ist eine hämatogen fortgeleitete, bakterielle oder fungale Infektion, die über choroidale oder retinale Gefäße in den Augapfel eingeschwemmt wird [ 1 – 3 ]. Von dort infiltrieren die Keime in die Netzhaut, den …

Bakterielle endogene Endophthalmitis

Vitrektomie Leitthema

Eine endogene Endophthalmitis stellt einen ophthalmologischen Notfall dar, der umgehender Diagnostik und Therapie bedarf. Es sollte mit geeigneten Methoden, wie beispielsweise dem Freiburger Endophthalmitis-Set, ein Keimnachweis erfolgen. Bei der …

So erreichen Sie eine bestmögliche Wundheilung der Kornea

Die bestmögliche Wundheilung der Kornea, insbesondere ohne die Ausbildung von lichtstreuenden Narben, ist oberstes Gebot, um einer dauerhaften Schädigung der Hornhaut frühzeitig entgegenzuwirken und die Funktion des Auges zu erhalten.   

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.