Erschienen in: European Journal of Clinical Microbiology & Infectious Diseases 12/2021

Open Access 04.06.2021 | COVID-19 | Brief Report

Performance of a SARS CoV-2 antibody ELISA based on simultaneous measurement of antibodies against the viral nucleoprotein and receptor-binding domain

verfasst von: Nina Reiners, Carolin Schnurra, Henning Trawinski, Judith Kannenberg, Thomas Hermsdorf, Andrea Aebischer, Torsten Schöneberg, Sven Reiche, Christian Jassoy

Erschienen in: European Journal of Clinical Microbiology & Infectious Diseases | Ausgabe 12/2021

Abstract

SARS CoV-2 antibody assays measure antibodies against the viral nucleoprotein (NP) or spike protein. The study examined if testing of antibodies against both antigens increases the diagnostic sensitivity. Sera (N=98) from infected individuals were tested with ELISAs based on the NP, receptor-binding domain (RBD), or both proteins. The AUROCs were 0.958 (NP), 0.991 (RBD), and 0.992 (NP/RBD). The RBD- and NP/RBD-based ELISAs showed better performance than the NP-based assay. Simultaneous testing for antibodies against NP and RBD increased the number of true and false positives. If maximum diagnostic sensitivity is required, the NP/RBD-based ELISA is preferable. Otherwise, the RBD-based ELISA is sufficient.
Begleitmaterial
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s10096-021-04284-5.
Nina Reiners and Carolin Schnurra contributed equally to this work.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Infection with the SARS CoV-2 induces a virus-specific immune response [14]. IgG antibodies against the virus indicate previous infection and examination of the antibody response in a population provides information about the level of exposure to the virus [58].
Commercial SARS CoV-2 antibody tests are based on measuring antibodies against the viral nucleoprotein (NP); the spike glycoprotein regions S1, S2, or the receptor-binding domain (RBD); or, more rarely, a combination of two viral proteins [915]. Antibody tests based on either the NP or the spike protein have similar sensitivity but react with slightly different sets of sera [1619]. This suggests that tests that concurrently measure antibodies against the two proteins are more sensitive than tests for antibodies against only NP or the spike glycoprotein. Based on this assumption, the aim of the study was to examine if simultaneous testing for IgG against the NP and RBD recognizes more sera than testing for antibodies against NP or RBD alone.

Study design

We performed a diagnostic study to compare the diagnostic sensitivity and specificity of IgG antibody ELISAs based on SARS CoV-2 NP, RBD, or the two proteins simultaneously.

Serum samples

Serum (N=91) and plasma (N=89) collected before 2019 were tested as putative negative specimens. Plasma and 80 sera were from blood donors (Institute for Transfusion Medicine, University Clinics, University of Leipzig). Eleven sera were from laboratory personnel. Ninety-eight sera were from adults with a positive SARS CoV-2 RT-PCR and mild to moderate or no symptoms. Forty-eight of the sera were obtained 4–10 weeks after infection; 50 blood specimens were drawn 6 months after infection. Participants were infected in March or April 2020. The sera from 4–10 weeks have previously been used to compare the sensitivity of commercial antibody tests [18]. Most (46 of 50) of the sera were from the same individuals.

Viral antigens

The SARS CoV-2 nucleoprotein was produced in Escherichia coli as a maltose-binding domain fusion protein. The receptor-binding domain (RBD) was generated as a strep-tag fusion protein in Expi293 cells (see Suppl. Material S1).

Antibody ELISA

Antibody ELISAs were performed with antigen-coated microtiter plates according to a modified standard protocol (see Suppl. Material S2).

Calibrator and cut-off value

A SARS CoV-2 antibody–positive serum prediluted in a protein stabilizer (Antibody Stabilizer PBS, Candor Bioscience GmbH) was used as the calibrator sample. To establish the calibration factor, the calibrator and 91 control sera from before 2019 were tested in parallel on the same plate. The calibration factor was calculated as the ratio of the OD of the calibrator serum and the mean plus two standard deviations of the OD of the control sera. In subsequent assays with plasma and sera, the cut-off was calculated by dividing the OD of the calibrator sample by the calibration factor.

Data analysis

Measurement units (U) were calculated by dividing the OD of the serum sample by the cut-off OD. As a start, sera with less than 1.0 U were negative and sera higher than 1.0 were positive. Receiver operating characteristic (ROC) analysis was performed and the cut-off values of the tests were adjusted such that the ELISAs had similar specificity. Sensitivity and specificity of the ELISAs were calculated with a contingency table and ROCs were calculated with a spreadsheet program. The area under the receiver operating characteristic curve (AUROC) was calculated with the MedCalc statistical software and the AUROCs were compared with VassarStats (http://​vassarstats.​net/​index.​html). The 95% confidence intervals were calculated using the MedCalc “Test for one proportion” and the Clopper-Pearson confidence interval. Proportions of unpaired serum samples were compared with the MedCalc “Comparison of proportions calculator” that uses the chi-squared test (https://​www.​medcalc.​org/​calc/​comparison_​of_​proportions.​php). Proportions of paired serum samples were compared with the exact binomial test (https://​scistatcalc.​blogspot.​com/​2013/​11/​mcnemars-test-calculator.​html).

Results

Area under the receiver operating characteristic curve (AUROC)

Receiver operating characteristic analysis was performed to determine the AUROC as a measure for the discrimination level of the tests. The AUROCs of the NP-, RBD-, and NP/RBD-based test were 0.958 (95% confidence interval (CI): 0.927–0.978), 0.991 (95% CI: 0.971–0.998), and 0.992 (95% CI: 0.974–0.999), respectively. The AUROC of the RBD- and NP/RBD-based assays was significantly larger than the AUROC of the NP-based ELISA (p-values 0.039 and 0.032, Table 1).
Table 1
Area under the receiver operating characteristic curve (AUROC) of the SARS CoV-2 antibody ELISAs
 
NP
RBD
NP/RBD
AUROC
0.958
CI: [0.927; 0.978]
0.991*
CI: [0.971; 0.998]
0.992*
CI: [0.974; 0.999]
*p-values of two-tailed comparisons of the AUROC of the antibody assays. NP- and RBD-based ELISA: 0.039, NP- versus NP/RBD-based ELISA: 0.032

Comparison of the sensitivity of the assays

To compare the diagnostic sensitivity of the tests, the cut-offs of the ELISAs were adjusted to a specificity of 98.33% by ROC analysis. The cut-off values were taken to determine the sensitivity of the tests with sera from SARS CoV-2-infected individuals. The NP-based ELISA found antibodies in 81 of the sera (82.7%, 95% confidence interval (CI): 73.7–89.6%). The RBD-based test showed 91 positive results (92.9%, 95% CI: 85.9–97.1%) and the NP/RBD-based assay recognized 94 (95.9%, 95% CI: 89.9–98.9%) of the sera. Subgroup analysis with the 48 sera obtained 4–10 weeks after infection showed positivity rates of 95.8% (95% CI: 85.7–99.5%), 97.9% (95% CI: 88.9–99.95%), and 100% (95% CI: 92.6–100%). The 50 sera from 6 months after infection gave 70.0% (95% CI: 55.4–82.1%), 88.0% (95% CI: 75.7–95.5%), and 92.0% (95% CI: 80.8–97.8%) positive results. The diagnostic sensitivity of the NP-based and NP/RBD-based tests was lower with sera from 6 months than from early after infection (Table 2).
Table 2
Sensitivity of the nucleoprotein (NP), RBD-based, and NP/RBD-based IgG ELISAs at a specificity of 98.3%
 
Positive sera (%)
Time post infection
No. of sera
NP
RBD
NP/RBD
> 4 weeks
98
82.7
CI [73.74; 89.60]
92.9
CI [85.89; 97.11]
95.9
CI [89.85; 98.87]
4–10 weeks
48
95.8
CIa [85.70; 99.48]
97.9
CI [88.9; 99.95]
100.0
CI [92.60; 100.00]
6 months
50
70.0
CI [55.39; 82.14]
88.0
CI [75.69; 95.47]
92.0
CI [80.77; 97.78]
p-levelb
 
0.0008*
0.0585
0.0465*
aCI: 95% confidence interval
bComparison of the sensitivity 4–10 weeks versus 6 months post infection by the chi-squared test. Significant p-values are labeled with an asterisk
The difference in sensitivity of the tests at 4-10 weeks after infection was statistically insignificant, but the RBD- and NP/RBD-based tests were more sensitive than the NP-based ELISA with the sera from 6 months after infection (p = 0.0413 and 0.0002). The NP/RBD-based ELISA showed a higher percentage of positive results compared with the RBD-based antibody assay, but this was not statistically different (Table 3). Five sera that were positive in the NP ELISA and negative in the RBD test were positive in the NP/RBD-based ELISA (Supplementary Table 1).
Table 3
Comparison of the sensitivity of the NP, RBD, and NP/RBD antibody ELISA
 
Time post infection
> 4 weeks
4–10 weeks
6 months
ELISA
RBDa
NP/RBD
RBD
NP/RBD
RBD
NP/RBD
NP
0.049b*
0.0009*
1
0.5
0.0413*
0.0002*
RBD
x
0.687
x
1
x
0.453
aRBD, receptor-binding domain; NP, nucleoprotein
bp-values. Sensitivity values from Table 1 were compared with the exact binomial test. An asterisk indicates statistical significance at a significance level of 0.05

Discussion

Most commercial SARS CoV-2 antibody tests measure the antibody response against either the viral NP or portions of the spike glycoprotein. Some tests measure antibodies against both NP and the spike protein [9, 13, 15, 2022]. We investigated if simultaneous testing with the NP and RBD improves the diagnostic sensitivity of SARS CoV-2 antibody testing compared with testing for antibodies against the NP and RBD alone.
The diagnostic sensitivity and specificity of the ELISAs were in the same range as commercial NP- and spike glycoprotein–based SARS CoV-2 ELISA tests [18, 2327]. The AUROC of the NP-based test was significantly smaller than those of the RBD- and NP/RBD-based ELISAs indicating that the NP-based ELISA has lower discriminatory potency. Among the other two assays, the NP/RBD-based test recognized more sera than the RBD-based ELISA. Several sera that were negative in the RBD-based ELISA were positive in both the NP-based and the NP/RBD-based ELISAs. However, the risk of false positive results also increased. With the samples examined, the AUROC and the diagnostic sensitivity of the combined antibody test were slightly higher than those of the RBD-based assay, but this was not statistically significant. The results are similar to those recently obtained with a Luminex-based SARS CoV-2 antibody test that simultaneously measured antibodies against NP, RBD, and other spike protein fractions [22].
We also evaluated the percentage of positive results with sera from different time points. The three ELISAs showed more positive results with sera obtained at 4–10 weeks after infection. In that time period, the NP-, RBD- and NP/RBD-based IgG ELISAs showed comparable proportions of positive sera. When tested with sera obtained 6 months after infection, the RBD- and NP/RBD-based tests detected more sera than the NP-based ELISA. The lower rate of positive results is most likely due to the slightly shorter half-life of NP-specific antibodies in the serum compared to spike glycoprotein–specific antibodies [2830].
The lower limits of detection of the NP- and RBD-based ELISAs were approximately 30 and 10 IU/ml (First WHO International Standard for anti-SARS-CoV-2 immunoglobulin, NIBSC code 20/136, data not shown). The analytical sensitivity remains unaffected by the antibody concentration in the sera. The loss of diagnostic sensitivity shown here is due to the decline of SARS CoV-2 antibodies after infection [2830].
The main technical limitation of the ELISAs is the occurrence of false positive sera and plasma. Sera that were false positive in the NP- or the RBD-based ELISA reacted with the NP/RBD-based ELISA. A similar observation was previously made with a Luminex-based antibody test [22]. Thus, although the sensitivity of the ELISA with the two proteins increased, the specificity declined concomitantly. Interestingly, the false positive sera were negative when tested in the nucleoprotein-based Roche Elecsys and the RBD-based Siemens Atellica IM COV2T antibody bridging assays (data not shown). This suggests that further improvement of the ELISA technique may reduce the frequency of false positive reactions.
Most vaccines against SARS CoV-2 induce immune responses against the spike glycoprotein but not the nucleoprotein. Therefore, tests that detect antibodies against the viral NP do not measure vaccine-induced antibodies. On the other hand, NP-specific antibodies are a sign of previous infection. Ideally, SARS CoV-2 antibody tests provide information to differentiate between previous infection and immunization.
In summary, the SARS CoV-2 RBD-specific antibody ELISA was diagnostically more sensitive than the NP-specific ELISA. Simultaneous measurement of NP- and RBD-specific antibodies further increased the percentage of positive sera, but this was not statistically significant due to small sample size. The ELISAs showed fewer positive results with sera from 6 months after infection compared with sera from early time points. The loss of diagnostic sensitivity was most pronounced with the nucleoprotein-specific assay. This suggests that assays using the RBD or NP/RBD are better for serosurveillance beyond 6 months after infection.

Acknowledgements

The study was approved by the Ethics Committee of the Medical Faculty of the University of Leipzig. The authors thank T. Kaiser, R. Biemann (Institute for Laboratory Medicine, University Hospitals, University of Leipzig), and H. Zaiß (Labor alphaomega, Leipzig) for re-testing of several sera with the Roche Elecsys anti-SARS-CoV-2 and Siemens Atellica IM COV2T tests.

Code availability

Not applicable.

Declarations

Ethics approval

The study was performed in line with the principles of the Declaration of Helsinki. It was approved by the Ethics Committee of the Medical Faculty of the University of Leipzig.
Oral and written informed consent was obtained from the participants.
The authors have seen the final version of the manuscript and approved submission for publication.

Conflict of interest

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Anhänge

Supplementary Information

Literatur
9.
Metadaten
Titel
Performance of a SARS CoV-2 antibody ELISA based on simultaneous measurement of antibodies against the viral nucleoprotein and receptor-binding domain
verfasst von
Nina Reiners
Carolin Schnurra
Henning Trawinski
Judith Kannenberg
Thomas Hermsdorf
Andrea Aebischer
Torsten Schöneberg
Sven Reiche
Christian Jassoy
Publikationsdatum
04.06.2021
Verlag
Springer Berlin Heidelberg
Schlagwort
COVID-19
Erschienen in
European Journal of Clinical Microbiology & Infectious Diseases / Ausgabe 12/2021
Print ISSN: 0934-9723
Elektronische ISSN: 1435-4373
DOI
https://doi.org/10.1007/s10096-021-04284-5

Weitere Artikel der Ausgabe 12/2021

European Journal of Clinical Microbiology & Infectious Diseases 12/2021 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

„Restriktion auf vier Wochen Therapie bei Schlaflosigkeit ist absurd!“

06.05.2024 Insomnie Nachrichten

Chronische Insomnie als eigenständiges Krankheitsbild ernst nehmen und adäquat nach dem aktuellen Forschungsstand behandeln: Das forderte der Schlafmediziner Dr. Dieter Kunz von der Berliner Charité beim Praxis Update.

GLP-1-Rezeptoragonisten und SGLT-2-Hemmer: zusammen besser

06.05.2024 Typ-2-Diabetes Nachrichten

Immer häufiger wird ein Typ-2-Diabetes sowohl mit einem GLP-1-Rezeptor-Agonisten als auch mit einem SGLT-2-Inhibitor behandelt. Wie sich das verglichen mit den Einzeltherapien auf kardiovaskuläre und renale Komplikationen auswirkt, wurde anhand von Praxisdaten aus Großbritannien untersucht.

Bessere Prognose mit links- statt rechtsseitigem Kolon-Ca.

06.05.2024 Kolonkarzinom Nachrichten

Menschen mit linksseitigem Kolonkarzinom leben im Mittel zweieinhalb Jahre länger als solche mit rechtsseitigem Tumor. Auch aktuell ist das Sterberisiko bei linksseitigen Tumoren US-Daten zufolge etwa um 11% geringer als bei rechtsseitigen.

Echinokokkose medikamentös behandeln oder operieren?

06.05.2024 DCK 2024 Kongressbericht

Die Therapie von Echinokokkosen sollte immer in spezialisierten Zentren erfolgen. Eine symptomlose Echinokokkose kann – egal ob von Hunde- oder Fuchsbandwurm ausgelöst – konservativ erfolgen. Wenn eine Op. nötig ist, kann es sinnvoll sein, vorher Zysten zu leeren und zu desinfizieren. 

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.