CT-based attenuation correction and resolution compensation for I-123 IMP brain SPECT normal database: a multicenter phantom study

verfasst von: Yoshitaka Inui, Takashi Ichihara, Masaki Uno, Masanobu Ishiguro, Kengo Ito, Katsuhiko Kato, Hajime Sakuma, Hidehiko Okazawa, Hiroshi Toyama

Erschienen in: Annals of Nuclear Medicine | Ausgabe 5/2018

Einloggen, um Zugang zu erhalten

Abstract

Objective

Statistical image analysis of brain SPECT images has improved diagnostic accuracy for brain disorders. However, the results of statistical analysis vary depending on the institution even when they use a common normal database (NDB), due to different intrinsic spatial resolutions or correction methods. The present study aimed to evaluate the correction of spatial resolution differences between equipment and examine the differences in skull bone attenuation to construct a common NDB for use in multicenter settings.

Methods

The proposed acquisition and processing protocols were those routinely used at each participating center with additional triple energy window (TEW) scatter correction (SC) and computed tomography (CT) based attenuation correction (CTAC). A multicenter phantom study was conducted on six imaging systems in five centers, with either single photon emission computed tomography (SPECT) or SPECT/CT, and two brain phantoms. The gray/white matter I-123 activity ratio in the brain phantoms was 4, and they were enclosed in either an artificial adult male skull, 1300 Hounsfield units (HU), a female skull, 850 HU, or an acrylic cover. The cut-off frequency of the Butterworth filters was adjusted so that the spatial resolution was unified to a 17.9 mm full width at half maximum (FWHM), that of the lowest resolution system. The gray-to-white matter count ratios were measured from SPECT images and compared with the actual activity ratio. In addition, mean, standard deviation and coefficient of variation images were calculated after normalization and anatomical standardization to evaluate the variability of the NDB.

Results

The gray-to-white matter count ratio error without SC and attenuation correction (AC) was significantly larger for higher bone densities (p < 0.05). The count ratio error with TEW and CTAC was approximately 5% regardless of bone density. After adjustment of the spatial resolution in the SPECT images, the variability of the NDB decreased and was comparable to that of the NDB without correction.

Conclusion

The proposed protocol showed potential for constructing an appropriate common NDB from SPECT images with SC, AC and spatial resolution compensation.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat McArthur C, Jampana R, Patterson J, Hadley D. Applications of cerebral SPECT. Clin Radiol. 2011;66:651 – 61.CrossRefPubMed McArthur C, Jampana R, Patterson J, Hadley D. Applications of cerebral SPECT. Clin Radiol. 2011;66:651 – 61.CrossRefPubMed
2.
Zurück zum Zitat Ballard C, Gauthier S, Corbett A, Brayne C, Aarsland D, Jones E. Alzheimer’s disease. Lancet. 2011;377:1019–31.CrossRefPubMed Ballard C, Gauthier S, Corbett A, Brayne C, Aarsland D, Jones E. Alzheimer’s disease. Lancet. 2011;377:1019–31.CrossRefPubMed
3.
Zurück zum Zitat McKeith IG, Boeve BF, Dickson DW, Halliday G, Taylor JP, Weintraub D, et al. Diagnosis and management of dementia with Lewy bodies: Fourth consensus report of the DLB Consortium. Neurology. 2017;89:88–100.CrossRefPubMedPubMedCentral McKeith IG, Boeve BF, Dickson DW, Halliday G, Taylor JP, Weintraub D, et al. Diagnosis and management of dementia with Lewy bodies: Fourth consensus report of the DLB Consortium. Neurology. 2017;89:88–100.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Dubois B, Feldman HH, Jacova C, Dekosky ST, Barberger-Gateau P, Cummings J, et al. Research criteria for the diagnosis of Alzheimer’s disease: revising the NINCDS-ADRDA criteria. Lancet Neurol. 2007;6:734 – 46.CrossRefPubMed Dubois B, Feldman HH, Jacova C, Dekosky ST, Barberger-Gateau P, Cummings J, et al. Research criteria for the diagnosis of Alzheimer’s disease: revising the NINCDS-ADRDA criteria. Lancet Neurol. 2007;6:734 – 46.CrossRefPubMed
5.
Zurück zum Zitat Minoshima S, Frey KA, Koeppe RA, Foster NL, Kuhl DE. A diagnostic approach in Alzheimer’s disease using three-dimensional stereotactic surface projections of fluorine-18-FDG PET. J Nucl Med. 1995;36:1238–48.PubMed Minoshima S, Frey KA, Koeppe RA, Foster NL, Kuhl DE. A diagnostic approach in Alzheimer’s disease using three-dimensional stereotactic surface projections of fluorine-18-FDG PET. J Nucl Med. 1995;36:1238–48.PubMed
6.
Zurück zum Zitat Friston KJ. Analyzing brain images: Principles and overview. In: Frackowiak RSJ, Friston KJ, Frith CD et al, editors. Human Brain Function. San Diego, CA: Academic Press; 1997. pp. 25–41. Friston KJ. Analyzing brain images: Principles and overview. In: Frackowiak RSJ, Friston KJ, Frith CD et al, editors. Human Brain Function. San Diego, CA: Academic Press; 1997. pp. 25–41.
7.
Zurück zum Zitat Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shoji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39:181–9.PubMed Iida H, Narita Y, Kado H, Kashikura A, Sugawara S, Shoji Y, et al. Effects of scatter and attenuation correction on quantitative assessment of regional cerebral blood flow with SPECT. J Nucl Med. 1998;39:181–9.PubMed
8.
Zurück zum Zitat Shimada H, Otake H, Higuchi T, Arisaka Y, Oriuchi N, Endo K. Normal database (NDB) of 123I-IMP brain perfusion SPECT examination is affected by statistical image analysis in the presence or absence of scatter correction and attenuation correction. Kaku Igaku. 2012;49:341–9. (ArticleJapanese).PubMed Shimada H, Otake H, Higuchi T, Arisaka Y, Oriuchi N, Endo K. Normal database (NDB) of 123I-IMP brain perfusion SPECT examination is affected by statistical image analysis in the presence or absence of scatter correction and attenuation correction. Kaku Igaku. 2012;49:341–9. (ArticleJapanese).PubMed
9.
Zurück zum Zitat Onishi H, Matsutake Y, Matsumoto N, Kai Y, Amijima H. Effect of prefiltering cutoff frequency and scatter and attenuation corrections during normal database creation for statistical imaging analysis of the brain. J Nucl Med Technol. 2011;39:231–6.CrossRefPubMed Onishi H, Matsutake Y, Matsumoto N, Kai Y, Amijima H. Effect of prefiltering cutoff frequency and scatter and attenuation corrections during normal database creation for statistical imaging analysis of the brain. J Nucl Med Technol. 2011;39:231–6.CrossRefPubMed
10.
Zurück zum Zitat Chang LT. A method for attenuation correction in radionuclide computed tomography. IEEE Trans Nucl Sci. 1977;25:638–43.CrossRef Chang LT. A method for attenuation correction in radionuclide computed tomography. IEEE Trans Nucl Sci. 1977;25:638–43.CrossRef
11.
Zurück zum Zitat Stodilka RZ, Kemp BJ, Prato FS, Nicholson RL. Importance of bone attenuation in brain SPECT quantification. J Nucl Med. 1998;39:190–7.PubMed Stodilka RZ, Kemp BJ, Prato FS, Nicholson RL. Importance of bone attenuation in brain SPECT quantification. J Nucl Med. 1998;39:190–7.PubMed
12.
Zurück zum Zitat Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of methods of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.PubMed Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of methods of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.PubMed
13.
Zurück zum Zitat Farid K, Habert MO, Martineau A, Caillat-Vigneron N, Sibon I. CT nonuniform attenuation and TEW scatter corrections in brain Tc-99m ECD SPECT. Clin Nucl Med. 2011;36:665–8.CrossRefPubMed Farid K, Habert MO, Martineau A, Caillat-Vigneron N, Sibon I. CT nonuniform attenuation and TEW scatter corrections in brain Tc-99m ECD SPECT. Clin Nucl Med. 2011;36:665–8.CrossRefPubMed
14.
Zurück zum Zitat Ishii K, Hanaoka K, Okada M, Kumano S, Komeya Y, Tsuchiya N, Hosono M, Murakami T. Impact of CT attenuation correction by SPECT/CT in brain perfusion images. Ann Nucl Med. 2012;26:241–7.CrossRefPubMed Ishii K, Hanaoka K, Okada M, Kumano S, Komeya Y, Tsuchiya N, Hosono M, Murakami T. Impact of CT attenuation correction by SPECT/CT in brain perfusion images. Ann Nucl Med. 2012;26:241–7.CrossRefPubMed
15.
Zurück zum Zitat Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34:2216–21.PubMed Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT. J Nucl Med. 1993;34:2216–21.PubMed
16.
Zurück zum Zitat Meikle SR, Hutton BF, Bailey DL. A transmission dependent method for scatter correction in SPECT. J Nucl Med. 1994;35:360–7.PubMed Meikle SR, Hutton BF, Bailey DL. A transmission dependent method for scatter correction in SPECT. J Nucl Med. 1994;35:360–7.PubMed
17.
Zurück zum Zitat Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: a simulation study. Phys Med Biol. 1997;42:1619–32.CrossRefPubMed Beekman FJ, Kamphuis C, Frey EC. Scatter compensation methods in 3D iterative SPECT reconstruction: a simulation study. Phys Med Biol. 1997;42:1619–32.CrossRefPubMed
18.
Zurück zum Zitat Ljungberg M, King A, Hademensos G, Strand E. Comparison of four scatter correction methods using Monte Carlo simulated source distributions. J Nucl Med. 1994;35:143–51.PubMed Ljungberg M, King A, Hademensos G, Strand E. Comparison of four scatter correction methods using Monte Carlo simulated source distributions. J Nucl Med. 1994;35:143–51.PubMed
19.
Zurück zum Zitat Minoshima S, Koeppe RA, Frey KA, Kuhl DE. Anatomic standardization: linear scaling and nonlinear warping of functional brain images. J Nucl Med. 1994;35:1528–37.PubMed Minoshima S, Koeppe RA, Frey KA, Kuhl DE. Anatomic standardization: linear scaling and nonlinear warping of functional brain images. J Nucl Med. 1994;35:1528–37.PubMed
Metadaten
Titel
CT-based attenuation correction and resolution compensation for I-123 IMP brain SPECT normal database: a multicenter phantom study
verfasst von
Yoshitaka Inui
Takashi Ichihara
Masaki Uno
Masanobu Ishiguro
Kengo Ito
Katsuhiko Kato
Hajime Sakuma
Hidehiko Okazawa
Hiroshi Toyama
Publikationsdatum
19.03.2018
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 5/2018
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-018-1248-x

Weitere Artikel der Ausgabe 5/2018

Annals of Nuclear Medicine 5/2018 Zur Ausgabe