Depth of sedation during drug induced sedation endoscopy monitored by BiSpectral Index® and Cerebral State Index®

verfasst von: Michael Herzog, Mathias Rudzki, Sebastian Plößl, Stefan Plontke, Patrick Kellner

Erschienen in: Sleep and Breathing | Ausgabe 2/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Drug induced sedation endoscopy (DISE) is performed to investigate patterns and sites of obstruction in patients with sleep-disordered breathing (SDB). During DISE the patients are sedated to obtain a muscular relaxation of the upper airway which mimics the relaxation during natural sleep. Different sleep stages are intended to be simulated by drug induced sedation, and it is helpful to measure the depth of sedation. The BiSpectral Index® (BIS) is often used for this procedure. Besides the BIS, other means of sedation depth monitoring exist in anaesthesiology but have not yet been investigated with respect to DISE. Monitoring of the Cerebral State Index® (CSI) is one of these methods. The aim of the study was to compare the BIS and CSI for sedation depth monitoring during DISE.

Methods

Sixty patients underwent DISE monitored by the BIS and CSI in parallel. The BIS and CSI values were compared using the Bland-Altman analysis.

Results

The BIS and CSI values differed during the course of sedation during DISE by a mean of – 6.07. At light sedation (BIS 60–80), lower values by 10 scale points of CSI compared with BIS were detectable. At deeper sedation levels (BIS 40–50), the CSI turned to present equal and even higher values compared with the BIS.

Conclusion

Sedation depth measurement during DISE can be performed by the BIS or CSI, but the differences should be interpreted carefully as comparable data for sleep stages in natural sleep are available only for BIS.
Literatur
1.
Zurück zum Zitat Pringle MB, Croft CB (1991) A comparison of sleep nasendoscopy and the Muller manoeuvre. Clin Otolaryngol Allied Sci 16(6):559–562CrossRef Pringle MB, Croft CB (1991) A comparison of sleep nasendoscopy and the Muller manoeuvre. Clin Otolaryngol Allied Sci 16(6):559–562CrossRef
2.
Zurück zum Zitat Pringle MB, Croft CB (1993) A grading system for patients with obstructive sleep apnoea--based on sleep nasendoscopy. Clin Otolaryngol Allied Sci 18(6):480–484CrossRef Pringle MB, Croft CB (1993) A grading system for patients with obstructive sleep apnoea--based on sleep nasendoscopy. Clin Otolaryngol Allied Sci 18(6):480–484CrossRef
3.
Zurück zum Zitat de Vito A, Carrasco Llatas M, Ravesloot MJ, Kotecha B, de Vries N, Hamans E, Maurer J, Bosi M, Blumen M, Heiser C, Herzog M, Montevecchi F, Corso RM, Braghiroli A, Gobbi R, Vroegop A, Vonk PE, Hohenhorst W, Piccin O, Sorrenti G, Vanderveken OM, Vicini C (2018) European position paper on drug-induced sleep endoscopy: 2017 Update. Clin Otolaryngol 43(6):1541–1552. https://doi.org/10.1111/coa.13213CrossRefPubMed de Vito A, Carrasco Llatas M, Ravesloot MJ, Kotecha B, de Vries N, Hamans E, Maurer J, Bosi M, Blumen M, Heiser C, Herzog M, Montevecchi F, Corso RM, Braghiroli A, Gobbi R, Vroegop A, Vonk PE, Hohenhorst W, Piccin O, Sorrenti G, Vanderveken OM, Vicini C (2018) European position paper on drug-induced sleep endoscopy: 2017 Update. Clin Otolaryngol 43(6):1541–1552. https://​doi.​org/​10.​1111/​coa.​13213CrossRefPubMed
9.
Zurück zum Zitat Nieuwenhuijs D, Coleman EL, Douglas NJ et al (2002) Bispectral index values and spectral edge frequency at different stages of physiologic sleep. Anesth Analg 94(1):125–129 table of contentsPubMed Nieuwenhuijs D, Coleman EL, Douglas NJ et al (2002) Bispectral index values and spectral edge frequency at different stages of physiologic sleep. Anesth Analg 94(1):125–129 table of contentsPubMed
10.
Zurück zum Zitat Sleigh JW, Andrzejowski J, Steyn-Ross A, Steyn-Ross M (1999) The bispectral index: a measure of depth of sleep? Anesth Analg 88(3):659–661CrossRef Sleigh JW, Andrzejowski J, Steyn-Ross A, Steyn-Ross M (1999) The bispectral index: a measure of depth of sleep? Anesth Analg 88(3):659–661CrossRef
11.
Zurück zum Zitat Dahaba AA, Xue JX, Xu GX, Liu QH, Metzler H (2011) Bilateral bispectral index (BIS)-vista as a measure of physiologic sleep in sleep-deprived anesthesiologists. Minerva Anestesiol 77(4):388–393PubMed Dahaba AA, Xue JX, Xu GX, Liu QH, Metzler H (2011) Bilateral bispectral index (BIS)-vista as a measure of physiologic sleep in sleep-deprived anesthesiologists. Minerva Anestesiol 77(4):388–393PubMed
16.
Zurück zum Zitat Jensen EW, Lindholm P, Henneberg SW (1996) Autoregressive modeling with exogenous input of middle-latency auditory-evoked potentials to measure rapid changes in depth of anesthesia. Methods Inf Med 35(3):256–260CrossRef Jensen EW, Lindholm P, Henneberg SW (1996) Autoregressive modeling with exogenous input of middle-latency auditory-evoked potentials to measure rapid changes in depth of anesthesia. Methods Inf Med 35(3):256–260CrossRef
22.
Zurück zum Zitat Pilge S, Kreuzer M, Kochs EF et al (2012) Monitors of the hypnotic component of anesthesia - correlation between bispectral index and cerebral state index. Minerva Anestesiol 78(6):636–645PubMed Pilge S, Kreuzer M, Kochs EF et al (2012) Monitors of the hypnotic component of anesthesia - correlation between bispectral index and cerebral state index. Minerva Anestesiol 78(6):636–645PubMed
24.
Zurück zum Zitat Yang S-h, Zhong T, Guo Q-l et al (2006) Bispectral index, cerebral state index and the predicted effect-site concentration at different clinical end-points during target-controlled infusion of propofol. Zhong Nan Da Xue Xue Bao Yi Xue Ban 31(6):929–933PubMed Yang S-h, Zhong T, Guo Q-l et al (2006) Bispectral index, cerebral state index and the predicted effect-site concentration at different clinical end-points during target-controlled infusion of propofol. Zhong Nan Da Xue Xue Bao Yi Xue Ban 31(6):929–933PubMed
Metadaten
Titel
Depth of sedation during drug induced sedation endoscopy monitored by BiSpectral Index® and Cerebral State Index®
verfasst von
Michael Herzog
Mathias Rudzki
Sebastian Plößl
Stefan Plontke
Patrick Kellner
Publikationsdatum
28.08.2020
Verlag
Springer International Publishing
Erschienen in
Sleep and Breathing / Ausgabe 2/2021
Print ISSN: 1520-9512
Elektronische ISSN: 1522-1709
DOI
https://doi.org/10.1007/s11325-020-02180-4

Weitere Artikel der Ausgabe 2/2021

Sleep and Breathing 2/2021 Zur Ausgabe

Sleep Breathing Physiology and Disorders • Original Article

Severity evaluation of obstructive sleep apnea based on speech features

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.