Erschienen in:
Open Access
01.06.2020 | Short Commentary
Diagnostic accuracy of a smartphone bedside
test to assess the fixation suppression of the vestibulo-ocular reflex: when nothing else
matters
verfasst von:
Florin Gandor, Manfred Tesch, Hannelore Neuhauser, Doreen Gruber, Hans-Jochen Heinze, Georg Ebersbach, Thomas Lempert
Erschienen in:
Journal of Neurology
|
Ausgabe 7/2020
Einloggen, um Zugang zu erhalten
Abstract
Objective
Validation of a bedside test to objectify the fixation suppression of
the vestibulo-ocular reflex (FS-VOR) in patients with a cerebellar syndrome and
healthy controls.
Methods
The vestibulo-ocular reflex and its fixation suppression were assessed
by video-nystagmography (VNG) in 20 healthy subjects (mean age 56 ± 15) and 19
patients with a cerebellar syndrome (mean age 70 ± 11). The statistical cutoff
delineating normal from pathological FS-VOR was determined at the 2.5th percentile of
the normal distribution of the healthy cohort. VNG was then compared to a bedside
test, where eye movements were recorded with a smartphone while patients were rotated
on a swivel chair at a defined speed and amplitude. These videos were rated as normal
or pathological FS-VOR by six blinded raters, and results compared to VNG.
Results
VNG in healthy controls showed FS-VOR with a reduction of nystagmus
beats by 95.0% ± 7.2 (mean ± SD). The statistical cutoff was set at 80.6%. Cerebellar
patients reduced nystagmus beats by only 26.3% ± 25.1. Inter-rater agreement of the
smartphone video ratings was 85%. The sensitivity of the video ratings to detect an
impaired FS-VOR was 99%, its specificity 92%. Inter-test agreement was 91%.
Conclusion
The smartphone bedside test is an easily performed, reliable, sensitive,
specific, and inexpensive alternative for assessing FS-VOR.