Dynamic torsional misalignment of eyes during laser in-situ keratomileusis

verfasst von: Mehdi Shajari, Jens Bühren, Thomas Kohnen

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To determine the amount and characteristics of dynamic torsional misalignment of eyes during excimer ablation in laser in-situ keratomileusis (LASIK).

Methods

Retrospective trial for evaluation of dynamic intraoperative torsional misalignment of 179 eyes that underwent LASIK for correction of myopia and/or astigmatism. Patients were treated with the Keracor 217z excimer laser implementing 25 Hz dynamic eye tracker ACE 100 (both Technolas Perfect Vision, Munich, Germany). From dynamic torsional misalignments, temporal power spectra were obtained by Fourier analysis up to a frequency of 12.5 Hz and an amplitude of ±15° from initial torsional status (limited by the tracking system). The f90, f95, and f99 criteria were defined as the frequency below which 90 %, 95 %, and 99 % of misalignments occur. A Wilcoxon rank sum test was performed to detect differences of f90, f95, and f99 in groups’ gender, age, and eye (if both eyes underwent surgery at same day). Multiple regression analysis (MRA) was performed to evaluate possible preoperative predictors of f90, f95, and f99.

Results

Fourier analysis showed a dominance of high-frequency, low-power dynamic torsional misalignment. Mean f95 threshold of rotational movements was 4.89±2.12 Hz (median 4.54, ranging from 0.44 to 9.23 Hz). Wilcoxon rank sum test showed no differences in f90, f95, and f99 between groups' gender, age, and eye. MRA revealed age, gender, and optical zone as preoperative predictors on intraoperative f90, f95, and f99.

Conclusions

Dynamic intraoperative torsional misalignments of eyes undergoing LASIK are dominated by low-frequency (slow), high-power (large) movements, with 95 % being slower than 4.89Hz regarding the spectrum analyzed (0–12.5Hz, ±15°). Movements can be predicted preoperatively by eye treated, patients’ gender, and age in pre-LASIK diagnostics.
Literatur
1.
Zurück zum Zitat Mrochen M, Eldine MS, Kaemmerer M, Seiler T, Hütz W (2001) Improvement in photorefractive corneal laser surgery results using an active eye-tracking system. J Cataract Refract Surg 27:1000–1006CrossRefPubMed Mrochen M, Eldine MS, Kaemmerer M, Seiler T, Hütz W (2001) Improvement in photorefractive corneal laser surgery results using an active eye-tracking system. J Cataract Refract Surg 27:1000–1006CrossRefPubMed
3.
Zurück zum Zitat Ditchburn RW, Foley-Fisher JA (1967) Assembled data in eye movements. Opt Acta 14:113–118CrossRef Ditchburn RW, Foley-Fisher JA (1967) Assembled data in eye movements. Opt Acta 14:113–118CrossRef
4.
Zurück zum Zitat Park SH, Kim M, Joo C-K (2009) Measurement of pupil centroid shift and cyclotorsional displacement using iris registration. Ophthalmologica 223:166–171CrossRefPubMed Park SH, Kim M, Joo C-K (2009) Measurement of pupil centroid shift and cyclotorsional displacement using iris registration. Ophthalmologica 223:166–171CrossRefPubMed
5.
Zurück zum Zitat Chernyak DA (2004) Cyclotorsional eye motion occurring between wavefront measurement and refractive surgery. J Cataract Refract Surg 30:633–638CrossRefPubMed Chernyak DA (2004) Cyclotorsional eye motion occurring between wavefront measurement and refractive surgery. J Cataract Refract Surg 30:633–638CrossRefPubMed
6.
Zurück zum Zitat Ciccio AE, Durrie DS, Stahl JE, Schwendeman F (2005) Ocular cyclotorsion during customized laser ablation. J Refract Surg 21:S772–S774PubMed Ciccio AE, Durrie DS, Stahl JE, Schwendeman F (2005) Ocular cyclotorsion during customized laser ablation. J Refract Surg 21:S772–S774PubMed
7.
Zurück zum Zitat Swami AU, Steinert RF, Osborne WE, White AA (2002) Rotational malposition during laser in situ keratomileusis. Am J Ophthalmol 133:561–562CrossRefPubMed Swami AU, Steinert RF, Osborne WE, White AA (2002) Rotational malposition during laser in situ keratomileusis. Am J Ophthalmol 133:561–562CrossRefPubMed
8.
Zurück zum Zitat Kohnen T, Kühne C, Cichocki M, Strenger A (2007) Cyclorotation of the eye in wavefront-guided LASIK using a static eyetracker with iris recognition. Ophthalmologe 104:60–65CrossRefPubMed Kohnen T, Kühne C, Cichocki M, Strenger A (2007) Cyclorotation of the eye in wavefront-guided LASIK using a static eyetracker with iris recognition. Ophthalmologe 104:60–65CrossRefPubMed
9.
Zurück zum Zitat Chang J (2008) Cyclotorsion during laser in situ keratomileusis. J Cataract Refract Surg 34:1720–1726CrossRefPubMed Chang J (2008) Cyclotorsion during laser in situ keratomileusis. J Cataract Refract Surg 34:1720–1726CrossRefPubMed
10.
Zurück zum Zitat Febbraro J-L, Koch DD, Khan HN, Saad A, Gatinel D (2010) Detection of static cyclotorsion and compensation for dynamic cyclotorsion in laser in situ keratomileusis. J Cataract Refract Surg 36:1718–1723CrossRefPubMed Febbraro J-L, Koch DD, Khan HN, Saad A, Gatinel D (2010) Detection of static cyclotorsion and compensation for dynamic cyclotorsion in laser in situ keratomileusis. J Cataract Refract Surg 36:1718–1723CrossRefPubMed
11.
Zurück zum Zitat Waring GO (2009) One-kilohertz eye tracker and active intraoperative torsion detection in the NIDEK CXIII and Quest excimer lasers. J Refract Surg 25:S931–S933CrossRefPubMed Waring GO (2009) One-kilohertz eye tracker and active intraoperative torsion detection in the NIDEK CXIII and Quest excimer lasers. J Refract Surg 25:S931–S933CrossRefPubMed
12.
Zurück zum Zitat Porter J, Yoon G, MacRae S, Pan G, Twietmeyer T, Cox IG, Williams DR (2005) Surgeon offsets and dynamic eye movements in laser refractive surgery. J Cataract Refract Surg 31:2058–2066CrossRefPubMed Porter J, Yoon G, MacRae S, Pan G, Twietmeyer T, Cox IG, Williams DR (2005) Surgeon offsets and dynamic eye movements in laser refractive surgery. J Cataract Refract Surg 31:2058–2066CrossRefPubMed
13.
Zurück zum Zitat Wang L, Koch DD (2008) Residual higher-order aberrations caused by clinically measured cyclotorsional misalignment or decentration during wavefront-guided excimer laser corneal ablation. J Cataract Refract Surg 34:2057–2062CrossRefPubMed Wang L, Koch DD (2008) Residual higher-order aberrations caused by clinically measured cyclotorsional misalignment or decentration during wavefront-guided excimer laser corneal ablation. J Cataract Refract Surg 34:2057–2062CrossRefPubMed
14.
Zurück zum Zitat Tjon-Fo-Sang MJ, de Faber J-THN, Kingma C, Beekhuis WH (2002) Cyclotorsion: a possible cause of residual astigmatism in refractive surgery. J Cataract Refract Surg 28:599–602CrossRefPubMed Tjon-Fo-Sang MJ, de Faber J-THN, Kingma C, Beekhuis WH (2002) Cyclotorsion: a possible cause of residual astigmatism in refractive surgery. J Cataract Refract Surg 28:599–602CrossRefPubMed
15.
Zurück zum Zitat Arba-Mosquera S, Merayo-Lloves J, de Ortueta D (2008) Clinical effects of pure cyclotorsional errors during refractive surgery. Invest Ophthalmol Vis Sci 49:4828–4836CrossRefPubMed Arba-Mosquera S, Merayo-Lloves J, de Ortueta D (2008) Clinical effects of pure cyclotorsional errors during refractive surgery. Invest Ophthalmol Vis Sci 49:4828–4836CrossRefPubMed
16.
Zurück zum Zitat Ghosh S, Couper TA, Lamoureux E, Jhanji V, Taylor HR, Vajpayee RB (2008) Evaluation of iris recognition system for wavefront-guided laser in situ keratomileusis for myopic astigmatism. J Cataract Refract Surg 34:215–221CrossRefPubMed Ghosh S, Couper TA, Lamoureux E, Jhanji V, Taylor HR, Vajpayee RB (2008) Evaluation of iris recognition system for wavefront-guided laser in situ keratomileusis for myopic astigmatism. J Cataract Refract Surg 34:215–221CrossRefPubMed
17.
Zurück zum Zitat Kohnen T, Knorz MC, Neuhann T, German Ophthalmological Society, Professional Association of German Ophthalmologists (2007) Evaluation and quality assurance of refractive surgery procedures by the German Ophthalmological Society and the Professional Association of German Ophthalmologists. Ophthalmologe 104:719–726CrossRefPubMed Kohnen T, Knorz MC, Neuhann T, German Ophthalmological Society, Professional Association of German Ophthalmologists (2007) Evaluation and quality assurance of refractive surgery procedures by the German Ophthalmological Society and the Professional Association of German Ophthalmologists. Ophthalmologe 104:719–726CrossRefPubMed
18.
Zurück zum Zitat Shannon CE (1949) Communication in the presence of noise. Proc IRE 37:10–21CrossRef Shannon CE (1949) Communication in the presence of noise. Proc IRE 37:10–21CrossRef
19.
Zurück zum Zitat Kim H, Joo C-K (2008) Ocular cyclotorsion according to body position and flap creation before laser in situ keratomileusis. J Cataract Refract Surg 34:557–561CrossRefPubMed Kim H, Joo C-K (2008) Ocular cyclotorsion according to body position and flap creation before laser in situ keratomileusis. J Cataract Refract Surg 34:557–561CrossRefPubMed
20.
Zurück zum Zitat Bolger C, Bojanic S, Sheahan NF, Coakley D, Malone JF (1999) Dominant frequency content of ocular microtremor from normal subjects. Vision Res 39:1911–1915CrossRefPubMed Bolger C, Bojanic S, Sheahan NF, Coakley D, Malone JF (1999) Dominant frequency content of ocular microtremor from normal subjects. Vision Res 39:1911–1915CrossRefPubMed
21.
Zurück zum Zitat Steinman RM, Cunitz RJ, Timberlake GT, Herman M (1967) Voluntary control of microsaccades during maintained monocular fixation. Science 155:1577–1579CrossRefPubMed Steinman RM, Cunitz RJ, Timberlake GT, Herman M (1967) Voluntary control of microsaccades during maintained monocular fixation. Science 155:1577–1579CrossRefPubMed
Metadaten
Titel
Dynamic torsional misalignment of eyes during laser in-situ keratomileusis
verfasst von
Mehdi Shajari
Jens Bühren
Thomas Kohnen
Publikationsdatum
12.03.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 5/2016
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-016-3309-y

Weitere Artikel der Ausgabe 5/2016

Graefe's Archive for Clinical and Experimental Ophthalmology 5/2016 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Die überschießende Wundheilung in der filtrierenden Glaukomchirurgie ist ein zentraler Faktor für ein operatives Versagen. Nach der Einführung der Trabekulektomie in den 1960er-Jahren wurden viele Faktoren erkannt, die mit einer vermehrten …

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.