Effect of respiratory motion on quantitative myocardial gated SPECT: a simulation study

verfasst von: Ahmad Bitarafan-Rajabi, Hossein Rajabi, Feridoon Rastgou, Ali Akbar Sharafi

Erschienen in: Annals of Nuclear Medicine | Ausgabe 6/2009

Einloggen, um Zugang zu erhalten

Abstract

Objective

Respiratory motion is a potential cause of artefact and downgrading the quality of ECG-gated single photon emission computed tomography (SPECT) images that may result in clinical misinterpretation. We studied qualitatively the effects of respiratory motion on gated SPECT myocardial perfusion and function using Monte Carlo simulated data.

Methods

NCAT phantom was used to model a human torso. The cardiac and respiratory cycles of torso were 1 and 5 s, respectively. Eight realizations of the phantom, having diaphragmatic motion amplitudes of 0–7 cm were generated. SimSET Monte Carol simulator was used to image the phantom and generate gated studies of 16 frames per cardiac cycle.

Results

Our results demonstrated the underestimation of left ventricle end-diastolic, end-systolic, stroke volumes and ejection fraction and overestimation of wall motion and wall thickening (p < 0.01). In addition, the mean percentage of count in the basal-inferior, mid-inferior, apical-inferior, basal-septal and mid-septal segments were significantly lower due to respiratory motion when compared with control (p < 0.01). The changes in uptake were not significant in the apex, antroapical, apicoseptal, apicolateral, mid-anterior, basal-anterior, mid-lateral and basal-lateral segments.

Conclusion

Respiratory motion has significant effect on the calculation of the left ventricular functional and regional myocardial perfusion in the GSPECT. The amount of deterioration and quality distortion of the images depends on the amplitude of the diaphragmatic motion.
Literatur
1.
Zurück zum Zitat Cooper JA, Neumann PH, McCandless BK. Effect of patient motion on tomographic myocardial perfusion imaging. J Nucl Med. 1992;33:1566–71.PubMed Cooper JA, Neumann PH, McCandless BK. Effect of patient motion on tomographic myocardial perfusion imaging. J Nucl Med. 1992;33:1566–71.PubMed
2.
Zurück zum Zitat Germano G, Erel J, Kiat H, Kavanagh PB, Berman DS. Quantitative LVEF and qualitative regional function from gated thallium-201 perfusion SPECT. J Nucl Med. 1997;38:749–54.PubMed Germano G, Erel J, Kiat H, Kavanagh PB, Berman DS. Quantitative LVEF and qualitative regional function from gated thallium-201 perfusion SPECT. J Nucl Med. 1997;38:749–54.PubMed
3.
Zurück zum Zitat Go V, Bhatt MR, Hendel RC. The diagnostic and prognostic value of ECG-gated SPECT myocardial perfusion imaging. J Nucl Med. 2004;45:912–21.PubMed Go V, Bhatt MR, Hendel RC. The diagnostic and prognostic value of ECG-gated SPECT myocardial perfusion imaging. J Nucl Med. 2004;45:912–21.PubMed
4.
Zurück zum Zitat Pretorius P, King MA, Tsui BM, LaCroix KJ, Xia W. A mathematical model of motion of the heart for use in generating source and attenuation maps for simulating emission imaging. Med Phys. 1999;26:2323–32.PubMedCrossRef Pretorius P, King MA, Tsui BM, LaCroix KJ, Xia W. A mathematical model of motion of the heart for use in generating source and attenuation maps for simulating emission imaging. Med Phys. 1999;26:2323–32.PubMedCrossRef
5.
Zurück zum Zitat Segars PW, Tsui BMW. Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT. IEEE Trans Nucl Sci. 2002;49:675–9.CrossRef Segars PW, Tsui BMW. Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT. IEEE Trans Nucl Sci. 2002;49:675–9.CrossRef
6.
Zurück zum Zitat Tsui BMW, Segars WP, Lalush DS. Effects of upward creep and respiratory motion in myocardial SPECT. IEEE Trans Nucl Sci. 2000;47:1192–5.CrossRef Tsui BMW, Segars WP, Lalush DS. Effects of upward creep and respiratory motion in myocardial SPECT. IEEE Trans Nucl Sci. 2000;47:1192–5.CrossRef
7.
Zurück zum Zitat Wang Y, Riederer SJ, Ehman RL. Respiratory motion of the heart: kinematics and the implications for the spatial resolution in coronary imaging. Magn Reson Med. 1995;33:713–9.PubMedCrossRef Wang Y, Riederer SJ, Ehman RL. Respiratory motion of the heart: kinematics and the implications for the spatial resolution in coronary imaging. Magn Reson Med. 1995;33:713–9.PubMedCrossRef
8.
Zurück zum Zitat Bruyant P, King MA, Pretorius PH. Correction of the respiratory motion of the heart by tracking of the center of mass of thresholded projections: a simulation study using the dynamic MCAT phantom. IEEE Trans Nucl Sci. 2002;49:2159–66.CrossRef Bruyant P, King MA, Pretorius PH. Correction of the respiratory motion of the heart by tracking of the center of mass of thresholded projections: a simulation study using the dynamic MCAT phantom. IEEE Trans Nucl Sci. 2002;49:2159–66.CrossRef
9.
Zurück zum Zitat Segars WP, Lalush DS, Tsui BMW. A realistic spline-based dynamic heart phantom. IEEE Trans Nucl Sci. 1999;46:503–6.CrossRef Segars WP, Lalush DS, Tsui BMW. A realistic spline-based dynamic heart phantom. IEEE Trans Nucl Sci. 1999;46:503–6.CrossRef
10.
Zurück zum Zitat Zubal IG, Harrell CR, Smith EO, Rattner Z, Gindi GR, Hoffer PB. Computerized three-dimensional segmented human anatomy. Med Phys. 1994;21(2):299–302.PubMedCrossRef Zubal IG, Harrell CR, Smith EO, Rattner Z, Gindi GR, Hoffer PB. Computerized three-dimensional segmented human anatomy. Med Phys. 1994;21(2):299–302.PubMedCrossRef
11.
Zurück zum Zitat Haynor D, Harrison R, Lewellen T. The use of importance sampling techniques to improve the efficiency of photon tracking in emission tomography. Med Phys. 1991;18:990–1001.PubMedCrossRef Haynor D, Harrison R, Lewellen T. The use of importance sampling techniques to improve the efficiency of photon tracking in emission tomography. Med Phys. 1991;18:990–1001.PubMedCrossRef
12.
Zurück zum Zitat He ZX, Iskandrian AS, Gupta NC, Verani MS. Assessing coronary artery disease with dipyridamole technetium-99m-tetrofosmin SPECT: a multicenter trial. J Nucl Med. 1997;38(1):44–8.PubMed He ZX, Iskandrian AS, Gupta NC, Verani MS. Assessing coronary artery disease with dipyridamole technetium-99m-tetrofosmin SPECT: a multicenter trial. J Nucl Med. 1997;38(1):44–8.PubMed
13.
Zurück zum Zitat Hassan N, Escanyé JM, Juillière Y, Marie PY, David N, Olivier P, et al. 201Tl SPECT abnormalities, documented at rest in dilated cardiomyopathy, are related to a lower than normal myocardial thickness but not to an excess in myocardial wall stress. J Nucl Med. 2002;43(4):451–7.PubMed Hassan N, Escanyé JM, Juillière Y, Marie PY, David N, Olivier P, et al. 201Tl SPECT abnormalities, documented at rest in dilated cardiomyopathy, are related to a lower than normal myocardial thickness but not to an excess in myocardial wall stress. J Nucl Med. 2002;43(4):451–7.PubMed
14.
Zurück zum Zitat Pitman AG, Kalff V, Every BV, Risa B, Barnden LR, Kelly MJ. Effect of mechanically simulated diaphragmatic respiratory motion on myocardial SPECT processed with and without attenuation correction. J Nucl Med. 2002;43:1259–67.PubMed Pitman AG, Kalff V, Every BV, Risa B, Barnden LR, Kelly MJ. Effect of mechanically simulated diaphragmatic respiratory motion on myocardial SPECT processed with and without attenuation correction. J Nucl Med. 2002;43:1259–67.PubMed
15.
Zurück zum Zitat Kovalski G, Israel O, Keidar Z, Frenkel A, Sachs J, Azhari H. Correction of heart motion due to respiration in clinical myocardial perfusion SPECT scans using respiratory gating. J Nucl Med. 2007;48:630–6.PubMedCrossRef Kovalski G, Israel O, Keidar Z, Frenkel A, Sachs J, Azhari H. Correction of heart motion due to respiration in clinical myocardial perfusion SPECT scans using respiratory gating. J Nucl Med. 2007;48:630–6.PubMedCrossRef
16.
Zurück zum Zitat Murase K, Ishine M, Katoaka M, Itoh H, Mogami M, Iio A. Simulation and experimental study of respiratory motion effect on image quality of single photon emission computed tomography (SPECT). Eur J Nucl Med. 1987;13:244–9.PubMed Murase K, Ishine M, Katoaka M, Itoh H, Mogami M, Iio A. Simulation and experimental study of respiratory motion effect on image quality of single photon emission computed tomography (SPECT). Eur J Nucl Med. 1987;13:244–9.PubMed
17.
Zurück zum Zitat Zubal G, Bizais Y, Bennett GW, Brill AB. Dual gated nuclear cardiac images. IEEE Trans Nucl Sci. 1984;31:566–9.CrossRef Zubal G, Bizais Y, Bennett GW, Brill AB. Dual gated nuclear cardiac images. IEEE Trans Nucl Sci. 1984;31:566–9.CrossRef
18.
Zurück zum Zitat Segars WP, Lalush DS, Tsui BMW. Modelling respiratory mechanics in the MCAT and spline-based MCAT phantoms. IEEE Trans Nucl Sci. 2001;48(1):89–97.CrossRef Segars WP, Lalush DS, Tsui BMW. Modelling respiratory mechanics in the MCAT and spline-based MCAT phantoms. IEEE Trans Nucl Sci. 2001;48(1):89–97.CrossRef
19.
Zurück zum Zitat Cho K, Kumiata S, Okada S, Kumazaki T. Development of respiratory gated myocardial SPECT system. J Nucl Cardiol. 1999;6:20–8.PubMedCrossRef Cho K, Kumiata S, Okada S, Kumazaki T. Development of respiratory gated myocardial SPECT system. J Nucl Cardiol. 1999;6:20–8.PubMedCrossRef
20.
Zurück zum Zitat Beach RD, Pretorius PH, Boening G, Bruyant PP, Feng B, Fulton RR, et al. Feasibility of stereo-infrared tracking to monitor patient motion during cardiac SPECT imaging. IEEE Trans Nucl Sci. 2004;51:2693–8.PubMedCrossRef Beach RD, Pretorius PH, Boening G, Bruyant PP, Feng B, Fulton RR, et al. Feasibility of stereo-infrared tracking to monitor patient motion during cardiac SPECT imaging. IEEE Trans Nucl Sci. 2004;51:2693–8.PubMedCrossRef
21.
Zurück zum Zitat Hansen CL, Goldstein RA, Berman DS, Churchwell KB, Cooke CD, Corbett JR, et al. Myocardial perfusion and function single photon emission computed tomography. J Nucl Cardiol. 2006;13(6):e97–120.PubMedCrossRef Hansen CL, Goldstein RA, Berman DS, Churchwell KB, Cooke CD, Corbett JR, et al. Myocardial perfusion and function single photon emission computed tomography. J Nucl Cardiol. 2006;13(6):e97–120.PubMedCrossRef
22.
Zurück zum Zitat Pretorius PH, King MA. Spillover compensation in the presence of respiratory motion embedded in SPECT perfusion data. IEEE Trans Nucl Sci. 2008;55(1):537–42.CrossRef Pretorius PH, King MA. Spillover compensation in the presence of respiratory motion embedded in SPECT perfusion data. IEEE Trans Nucl Sci. 2008;55(1):537–42.CrossRef
23.
Zurück zum Zitat Dodge HT, Sandler H, Ballew DW, Loard JD. Use of biplane angiocardiography for the measurement of left ventricular volume in man. Am Heart J. 1960;60(5):762–76.PubMedCrossRef Dodge HT, Sandler H, Ballew DW, Loard JD. Use of biplane angiocardiography for the measurement of left ventricular volume in man. Am Heart J. 1960;60(5):762–76.PubMedCrossRef
24.
Zurück zum Zitat Germano G, Erel J, Lewin H, Kavanagh PB, Berman DS. Automatic quantitation of regional myocardial wall motion and thickening from gated technetium-99m sestamibi myocardial perfusion single-photon emission computed tomography. J Am Coll Cardiol. 1997;30:1360–7.PubMedCrossRef Germano G, Erel J, Lewin H, Kavanagh PB, Berman DS. Automatic quantitation of regional myocardial wall motion and thickening from gated technetium-99m sestamibi myocardial perfusion single-photon emission computed tomography. J Am Coll Cardiol. 1997;30:1360–7.PubMedCrossRef
25.
Zurück zum Zitat Galt JR, Garcia E, Robbins W. Effects of myocardial wall thickness on SPECT quantification. IEEE Trans Med Imaging. 1990;9(2):144–50.PubMedCrossRef Galt JR, Garcia E, Robbins W. Effects of myocardial wall thickness on SPECT quantification. IEEE Trans Med Imaging. 1990;9(2):144–50.PubMedCrossRef
26.
Zurück zum Zitat Sharir T, Berman DS, Waechter PB, Areeda J, Kavanagh PB, Gerlach J, et al. Quantitative analysis of regional motion and thickening by gated myocardial perfusion SPECT: normal heterogeneity and criteria for abnormality. J Nucl Med. 2001;42(11):1630–8.PubMed Sharir T, Berman DS, Waechter PB, Areeda J, Kavanagh PB, Gerlach J, et al. Quantitative analysis of regional motion and thickening by gated myocardial perfusion SPECT: normal heterogeneity and criteria for abnormality. J Nucl Med. 2001;42(11):1630–8.PubMed
Metadaten
Titel
Effect of respiratory motion on quantitative myocardial gated SPECT: a simulation study
verfasst von
Ahmad Bitarafan-Rajabi
Hossein Rajabi
Feridoon Rastgou
Ali Akbar Sharafi
Publikationsdatum
01.08.2009
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 6/2009
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-009-0277-x

Weitere Artikel der Ausgabe 6/2009

Annals of Nuclear Medicine 6/2009 Zur Ausgabe