Evaluating miniscrew position using orthopantomograms compared to cone-beam computed tomography

verfasst von: R. Bennemann, Dr. M. Baxmann, L. Keilig, S. Reimann, B. Braumann, C. Bourauel

Erschienen in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Ausgabe 3/2012

Einloggen, um Zugang zu erhalten

Abstract

Objective

Our objective was to compare the diagnostic capabilities of the orthopantomogram (OPT) as a standard dental diagnostic procedure with those of cone-beam computed tomography (CBCT). We examined the extent to which the OPT allowed miniscrew position to be estimated, and whether probands from different backgrounds in dental medicine arrived at the same conclusions regarding the screw position using the same radiological images (OPTs).

Materials and methods

First, a tomas® pin 8 mm in length and 1.2 mm in diameter (Dentaurum, Ispringen, Germany) was inserted in all four quadrants of nine macerated skulls. These nine skulls were then imaged by OPT and CBCT. The OPTs were presented to probands whose task it was to estimate the miniscrew position in relation to the neighboring structures in an evaluation form. The probands surveyed came from three different backgrounds in dental medicine: orthodontists, oral and maxillofacial surgeons, and dental students. Comparison with the CBCT was made by measuring the CBCTs using the GALILEOS Viewer (Sirona, Bensheim, Germany), and these results were compared with those in the evaluation forms.

Results

In this study, it could be shown that none of the three proband groups estimated the screw’s position in the OPT significantly better or worse in comparison to the CBCT. The probands’ assessments of the miniscrew’s proximity to the dental root differed, and in many cases did not correspond exactly with the CBCT measurement. The majority of probands estimated the likelihood of an opposing corticalis perforation as small, as confirmed in the CBCTs. The proximity of the miniscrew to the maxillary sinus or inferior alveolar nerve was similarly assessed by most of the probands, as also determined in the CBCT. However, the OPT assessments of some probands of the proximity to the maxillary sinus and/or inferior alveolar nerve differed from the proximity measured in the CBCT.

Conclusion

Overall our study demonstrates that the OPTs enabled a rough evaluation of the miniscrew position in relation to the surrounding structures. However, the probands’ assessments were imprecise. Major deviations appeared in the determination of the proximity of the screw to the dental root. For this reason, if there is any doubt, patients should take advantage of three-dimensional (3D) diagnostics, preferably before any surgical intervention. This is true, for example, if it is unclear whether a root will be or has been perforated, or indeed whether there is sufficient transverse bone tissue.
Literatur
1.
Zurück zum Zitat Ammar HH, Ngan P, Crout R J et al (2011) Three-dimensional modeling and finite element analysis in treatment planning for orthodontic tooth movement. Am J Orthod Dentofacial Orthop 139:e59–e71PubMedCrossRef Ammar HH, Ngan P, Crout R J et al (2011) Three-dimensional modeling and finite element analysis in treatment planning for orthodontic tooth movement. Am J Orthod Dentofacial Orthop 139:e59–e71PubMedCrossRef
2.
Zurück zum Zitat Chen Y, Kyung HM, Zhao WT, Yu WJ (2009) Critical factors for the success of orthodontic mini-implants: a systematic review. Am J Orthod Dentofacial Orthop 135:284–291PubMedCrossRef Chen Y, Kyung HM, Zhao WT, Yu WJ (2009) Critical factors for the success of orthodontic mini-implants: a systematic review. Am J Orthod Dentofacial Orthop 135:284–291PubMedCrossRef
3.
Zurück zum Zitat European Commission (2004) European guidelines on radiation protection in dental radiology. Office for Official Publications of the European Communities, Luxembourg European Commission (2004) European guidelines on radiation protection in dental radiology. Office for Official Publications of the European Communities, Luxembourg
4.
Zurück zum Zitat Fayed MMS, Pazera P, Katsaros C (2010) Optimal sites for orthodontic mini-implant placement assessed by cone-beam computed tomography. Angle Orthod 80:939–951PubMedCrossRef Fayed MMS, Pazera P, Katsaros C (2010) Optimal sites for orthodontic mini-implant placement assessed by cone-beam computed tomography. Angle Orthod 80:939–951PubMedCrossRef
5.
Zurück zum Zitat Kim SH, Kang SM, Choi YS et al (2010) Cone-beam computed tomography evaluation of mini-implants after placement: Is root proximity a major risk factor for failure? Am J Orthod Dentofacial Orthop 138:264–276PubMedCrossRef Kim SH, Kang SM, Choi YS et al (2010) Cone-beam computed tomography evaluation of mini-implants after placement: Is root proximity a major risk factor for failure? Am J Orthod Dentofacial Orthop 138:264–276PubMedCrossRef
6.
Zurück zum Zitat Kuroda S, Yamada K, Deguchi T (2007) Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop 131:68–73CrossRef Kuroda S, Yamada K, Deguchi T (2007) Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop 131:68–73CrossRef
7.
Zurück zum Zitat Lim JE, Lim WH, Chun YS (2008) Quantitative evaluation of cortical bone thickness and root proximity at maxillary interradicular sites for orthodontic mini-implant placement. Clin Anat 21:486–491PubMedCrossRef Lim JE, Lim WH, Chun YS (2008) Quantitative evaluation of cortical bone thickness and root proximity at maxillary interradicular sites for orthodontic mini-implant placement. Clin Anat 21:486–491PubMedCrossRef
8.
Zurück zum Zitat Liu H, Lv T, Wang N (2011) Drift characteristics of miniscrews and molars for anchorage under orthodontic force: 3-dimensional computed tomography registration evaluation. Am J Orthod Dentofacial Orthop 139:e83–e89PubMedCrossRef Liu H, Lv T, Wang N (2011) Drift characteristics of miniscrews and molars for anchorage under orthodontic force: 3-dimensional computed tomography registration evaluation. Am J Orthod Dentofacial Orthop 139:e83–e89PubMedCrossRef
9.
Zurück zum Zitat Ludwig B, Glasl B, Lietz T, Kopp S (2008) Radiological location monitoring in skeletal anchorage: Introduction of a positioning guide. J Orofac Orthop 69:59–65PubMedCrossRef Ludwig B, Glasl B, Lietz T, Kopp S (2008) Radiological location monitoring in skeletal anchorage: Introduction of a positioning guide. J Orofac Orthop 69:59–65PubMedCrossRef
10.
Zurück zum Zitat Martinelli FL, Luiz RR, Faria M, Nojima LI (2010) Anatomic variability in alveolar sites for skeletal anchorage. Am J Orthod Dentofacial Orthop 138:252.e1–252.e9PubMedCrossRef Martinelli FL, Luiz RR, Faria M, Nojima LI (2010) Anatomic variability in alveolar sites for skeletal anchorage. Am J Orthod Dentofacial Orthop 138:252.e1–252.e9PubMedCrossRef
11.
Zurück zum Zitat Miyawaki S, Koyama I, Inoue M et al (2003) Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop 124:373–378PubMedCrossRef Miyawaki S, Koyama I, Inoue M et al (2003) Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop 124:373–378PubMedCrossRef
12.
Zurück zum Zitat Motoyoshi M, Yoshida T, Ono A, Shimizu N (2007) Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants. Int J Oral Maxillofac Implants 22:779–784PubMed Motoyoshi M, Yoshida T, Ono A, Shimizu N (2007) Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants. Int J Oral Maxillofac Implants 22:779–784PubMed
13.
Zurück zum Zitat Müssig E, Wörtche R, Lux CJ (2005) Indications for digital volume tomography in orthodontics. J Orofac Orthop 66:241–249PubMedCrossRef Müssig E, Wörtche R, Lux CJ (2005) Indications for digital volume tomography in orthodontics. J Orofac Orthop 66:241–249PubMedCrossRef
14.
Zurück zum Zitat Präger TM, Mischkowski R, Laube N et al (2008) Remodeling along the bone-screw interface. J Orofac Orthop 69:337–348PubMedCrossRef Präger TM, Mischkowski R, Laube N et al (2008) Remodeling along the bone-screw interface. J Orofac Orthop 69:337–348PubMedCrossRef
15.
Zurück zum Zitat Reddy MS, Mayfield-Donahoo T, Vanderven F JJ, Jeffcoat M K (1994) A comparison of the diagnostic advantages of panoramic radiography and computed tomography scanning for placement of root form dental implants. Clin Oral Implants Res 5:229–238PubMedCrossRef Reddy MS, Mayfield-Donahoo T, Vanderven F JJ, Jeffcoat M K (1994) A comparison of the diagnostic advantages of panoramic radiography and computed tomography scanning for placement of root form dental implants. Clin Oral Implants Res 5:229–238PubMedCrossRef
16.
Zurück zum Zitat Scherer P, Neugebauer J, Ritter L et al (2007) Indikationen für die 3-dimensionale Bildgebung in der Zahnheilkunde. ZWR 116:219–230CrossRef Scherer P, Neugebauer J, Ritter L et al (2007) Indikationen für die 3-dimensionale Bildgebung in der Zahnheilkunde. ZWR 116:219–230CrossRef
17.
Zurück zum Zitat Schnelle C (2001) Vergleich der Strahlenexposition bei der Digitalen-Volumen-Tomographie, der Panoramaschichtaufnahme und der Computertomographie. Med Diss, Heinrich Heine Universität, Düsseldorf Schnelle C (2001) Vergleich der Strahlenexposition bei der Digitalen-Volumen-Tomographie, der Panoramaschichtaufnahme und der Computertomographie. Med Diss, Heinrich Heine Universität, Düsseldorf
18.
Zurück zum Zitat Wawrzinek C, Sommer T, Fischer-Brandies H (2008) Microdamage in cortical bone due to the overtightening of orthodontic microscrews. J Orofac Orthop 69:121–134PubMedCrossRef Wawrzinek C, Sommer T, Fischer-Brandies H (2008) Microdamage in cortical bone due to the overtightening of orthodontic microscrews. J Orofac Orthop 69:121–134PubMedCrossRef
19.
Zurück zum Zitat Wilmes B, Rademacher C, Olthoff G, Drescher D (2006) Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop 67:162–174PubMedCrossRef Wilmes B, Rademacher C, Olthoff G, Drescher D (2006) Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop 67:162–174PubMedCrossRef
20.
Zurück zum Zitat Wilmes B, Su YY, Sadigh L, Drescher D (2008) Pre-drilling force and insertion torques during orthodontic mini-implant insertion in relation to root contact. J Orofac Orthop 69:51–58PubMedCrossRef Wilmes B, Su YY, Sadigh L, Drescher D (2008) Pre-drilling force and insertion torques during orthodontic mini-implant insertion in relation to root contact. J Orofac Orthop 69:51–58PubMedCrossRef
21.
Zurück zum Zitat Wriedt S, Jaklin J, Al-Nawas B, Wehrbein H (2012) Impacted upper canines: examination and treatment proposal based on 3D versus 2D diagnosis. J Orofac Orthop 73:28–40PubMedCrossRef Wriedt S, Jaklin J, Al-Nawas B, Wehrbein H (2012) Impacted upper canines: examination and treatment proposal based on 3D versus 2D diagnosis. J Orofac Orthop 73:28–40PubMedCrossRef
Metadaten
Titel
Evaluating miniscrew position using orthopantomograms compared to cone-beam computed tomography
verfasst von
R. Bennemann
Dr. M. Baxmann
L. Keilig
S. Reimann
B. Braumann
C. Bourauel
Publikationsdatum
01.05.2012
Verlag
Springer-Verlag
Erschienen in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Ausgabe 3/2012
Print ISSN: 1434-5293
Elektronische ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-012-0079-y

Weitere Artikel der Ausgabe 3/2012

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 3/2012 Zur Ausgabe

Informationen

DGKFO-Seiten

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.