Evaluation of a novel normal database with matched SPECT systems and optimal pre-filter parameters for 3D-SSP

verfasst von: Hideo Onishi, Norikazu Matsutomo, Yuji Kai, Yoshiharu Kangai, Hizuru Amijima, Takeshi Yamaguchi

Erschienen in: Annals of Nuclear Medicine | Ausgabe 1/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

A normal perfusion database (NDB) is imperative for the statistical imaging of brain function. This study validates a novel NDB created under the same injection dose and acquisition conditions for three gamma camera systems and evaluates optimal pre-filter parameters for three-dimensional stereotactic surface projections (3D-SSPs).

Method

We compared a novel NDB that matched the databases in each of three vendor gamma camera systems with a conventionally constructed NDB (conventional NDB) and a NDB constructed in-house for 3D-SSP. We generated hypoperfused regions where pre-specified volumes were simulated for various areas in SPECT images. The properties of each NDB were evaluated based on the distribution of the standard deviation (SD). Abnormal accumulation regions were validated using Z, extent, and artifactual scores. Detection error was used to evaluate the optimal Butterworth pre-filter cutoff frequency with the perfusion defect rate (PDR) in 3D-SSP.

Results

The SD distribution was the same in the novel NDB and in the NDB constructed in-house, and the SD of the peak distribution was 0.08–0.07. The Z and extent scores of the novel DB and the NDB constructed in-house were similar, but increased along with the artifactual scores when using the conventional NDB. Many artifacts appeared in the Z score map when using the conventional NDB. The detection error deviated from the actual value by −1.3% at a cutoff frequency of 0.58 cycles/cm and a PDR of 30%, which was the lowest. The cutoff frequency became lower or higher, and the low-perfusion defect rate increased according to the increasing detection error. The optimal cutoff frequency was between 0.52 and 0.58 cycles/cm.

Conclusions

We generated a novel NDB according to the individual devices and compared it with a conventional and a NDB constructed in-house. The Z and extent scores were essentially equal when using the novel DB and the NDB constructed in-house, but considerably differed when using the conventional NDB. The optimal cutoff frequency of the Butterworth filter evaluated from the detection error was in the range of 0.52–0.58 cycles/cm. The detection error increased the perfusion defect rate by <15% and this was undetectable in 3D-SSP. The next step will be to improve the accuracy of the extent of abnormal regions and the sensitivity of the Z score using a novel NDB constructed according to the individual devices.
Literatur
1.
Zurück zum Zitat Minoshima S, Frey KA, Koeppe RA, Foster NL, Kuhl DE. A diagnostic approach in Alzheimer’s disease using three-dimensional stereotactic surface projections of fluorine-18-FDG PET. J Nucl Med. 1995;36:1238–48.PubMed Minoshima S, Frey KA, Koeppe RA, Foster NL, Kuhl DE. A diagnostic approach in Alzheimer’s disease using three-dimensional stereotactic surface projections of fluorine-18-FDG PET. J Nucl Med. 1995;36:1238–48.PubMed
2.
Zurück zum Zitat Minoshima S, Giordani B, Berent S, Frey KA, Foster NL, Kuhl DE. Metabolic reduction in the posterior cingulate cortex in very early Alzheimer’s disease. Ann Neurol. 1997;42:85–94.PubMedCrossRef Minoshima S, Giordani B, Berent S, Frey KA, Foster NL, Kuhl DE. Metabolic reduction in the posterior cingulate cortex in very early Alzheimer’s disease. Ann Neurol. 1997;42:85–94.PubMedCrossRef
3.
Zurück zum Zitat Ishii K, Imamura T, Sasaki M, Yamaji S, Sakamoto S, Kitagaki H, et al. Regional cerebral glucose metabolism in dementia with Lewy bodies and Alzheimer’s disease. Neurology. 1998;51:125–30.PubMed Ishii K, Imamura T, Sasaki M, Yamaji S, Sakamoto S, Kitagaki H, et al. Regional cerebral glucose metabolism in dementia with Lewy bodies and Alzheimer’s disease. Neurology. 1998;51:125–30.PubMed
4.
Zurück zum Zitat Kogure D, Matsuda H, Ohnishi T, Asada T, Uno M, Kunihiro, et al. Longitudinal evaluation of early Alzheimer’s disease using brain perfusion SPECT. J Nucl Med. 2000;41:1155–62.PubMed Kogure D, Matsuda H, Ohnishi T, Asada T, Uno M, Kunihiro, et al. Longitudinal evaluation of early Alzheimer’s disease using brain perfusion SPECT. J Nucl Med. 2000;41:1155–62.PubMed
5.
Zurück zum Zitat Ishii K, Sakamoto S, Sasaki M, Kitagaki H, Yamaji S, Hashimoto M, et al. Cerebral glucose metabolism in patients with frontotemporal dementia. J Nucl Med. 1998;39:1875–8.PubMed Ishii K, Sakamoto S, Sasaki M, Kitagaki H, Yamaji S, Hashimoto M, et al. Cerebral glucose metabolism in patients with frontotemporal dementia. J Nucl Med. 1998;39:1875–8.PubMed
6.
Zurück zum Zitat Honda N, Machida K, Matsumoto T, Matsuda H, Imabayashi E, Hashimoto J, et al. Three-dimensional stereotactic surface projection of brain perfusion SPECT improves diagnosis of Alzheimer’s disease. Ann Nucl Med. 2003;17:641–8.PubMedCrossRef Honda N, Machida K, Matsumoto T, Matsuda H, Imabayashi E, Hashimoto J, et al. Three-dimensional stereotactic surface projection of brain perfusion SPECT improves diagnosis of Alzheimer’s disease. Ann Nucl Med. 2003;17:641–8.PubMedCrossRef
7.
Zurück zum Zitat Ishii K, Kono A, Sasaki H, Miyamoto N, Fukuda T, Sakamoto S, et al. Fully automatic diagnostic system for early- and late-onset mild Alzheimer’s disease using FDG PET and 3D-SSP. Eur J Nucl Med Mol Imaging. 2006;33:575–83.PubMedCrossRef Ishii K, Kono A, Sasaki H, Miyamoto N, Fukuda T, Sakamoto S, et al. Fully automatic diagnostic system for early- and late-onset mild Alzheimer’s disease using FDG PET and 3D-SSP. Eur J Nucl Med Mol Imaging. 2006;33:575–83.PubMedCrossRef
8.
Zurück zum Zitat Van Laere K, Koole M, Versijpt J, Vandenberghe S, Brans B, D’Asseler Y, et al. Transfer of normal 99mTc-ECD SPET database between different gamma cameras. Eur J Nucl Med. 2001;28:435–49.PubMedCrossRef Van Laere K, Koole M, Versijpt J, Vandenberghe S, Brans B, D’Asseler Y, et al. Transfer of normal 99mTc-ECD SPET database between different gamma cameras. Eur J Nucl Med. 2001;28:435–49.PubMedCrossRef
9.
Zurück zum Zitat Matsuda H, Mizumura S, Soma T, Takemura N. Conversion of brain SPECT images between different collimators and reconstruction processes for analysis using statistical parametric mapping. Nucl Med Commun. 2004;25:67–74.PubMedCrossRef Matsuda H, Mizumura S, Soma T, Takemura N. Conversion of brain SPECT images between different collimators and reconstruction processes for analysis using statistical parametric mapping. Nucl Med Commun. 2004;25:67–74.PubMedCrossRef
10.
Zurück zum Zitat Mizumura S, Kumita S. Stereotactic statistical imaging analysis of the brain using the easy Z-score imaging system for sharing a normal database. Radiat Med. 2006;24:545–52.PubMedCrossRef Mizumura S, Kumita S. Stereotactic statistical imaging analysis of the brain using the easy Z-score imaging system for sharing a normal database. Radiat Med. 2006;24:545–52.PubMedCrossRef
11.
Zurück zum Zitat Ogawa K, Harata Y, Ichihara T, Kubo A, Hashimoto A. A practical method for position-dependent Compton-scatter correction in single photon emission CT. IEEE Trans Med Imaging. 1991;10:408–12.PubMedCrossRef Ogawa K, Harata Y, Ichihara T, Kubo A, Hashimoto A. A practical method for position-dependent Compton-scatter correction in single photon emission CT. IEEE Trans Med Imaging. 1991;10:408–12.PubMedCrossRef
12.
Zurück zum Zitat Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of method of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.PubMed Hayashi M, Deguchi J, Utsunomiya K, Yamada M, Komori T, Takeuchi M, et al. Comparison of method of attenuation and scatter correction in brain perfusion SPECT. J Nucl Med Technol. 2005;33:224–9.PubMed
13.
Zurück zum Zitat Lancaster JL, Summerlin JL, Rainey L, Freitas CS, Fox PT. The Talairach Daemon, a database server for Talairach Atlas Labels. Neuroimage. 1997;5:S633. Lancaster JL, Summerlin JL, Rainey L, Freitas CS, Fox PT. The Talairach Daemon, a database server for Talairach Atlas Labels. Neuroimage. 1997;5:S633.
14.
Zurück zum Zitat Lancaster JL, Woldorff MG, Parsons LM, Liotti M, Freitas CS, Rainey L, et al. Automated Talairach Atlas labels for functional brain mapping. Hum Brain Mapp. 2000;10:120–31.PubMedCrossRef Lancaster JL, Woldorff MG, Parsons LM, Liotti M, Freitas CS, Rainey L, et al. Automated Talairach Atlas labels for functional brain mapping. Hum Brain Mapp. 2000;10:120–31.PubMedCrossRef
15.
Zurück zum Zitat Mizumura S, Kumita S, Cho K, Ishihara M, Nakajo H, Toba M, et al. Development of quantitative analysis method for stereotactic brain image: assessment of reduced accumulation in extent and severity using anatomical segmentation. Ann Nucl Med. 2003;17:289–95.PubMedCrossRef Mizumura S, Kumita S, Cho K, Ishihara M, Nakajo H, Toba M, et al. Development of quantitative analysis method for stereotactic brain image: assessment of reduced accumulation in extent and severity using anatomical segmentation. Ann Nucl Med. 2003;17:289–95.PubMedCrossRef
16.
Zurück zum Zitat Shimizu S, Hanyu H, Iwamoto T, Koizumi K, Abe K. SPECT follow-up study of cerebral blood flow changes during Donepezil therapy in patients with Alzheimer’s disease. J Neuroimaging. 2006;16:16–23.PubMedCrossRef Shimizu S, Hanyu H, Iwamoto T, Koizumi K, Abe K. SPECT follow-up study of cerebral blood flow changes during Donepezil therapy in patients with Alzheimer’s disease. J Neuroimaging. 2006;16:16–23.PubMedCrossRef
17.
Zurück zum Zitat Ishii K, Kanda T, Uemura T, Miyamoto N, Yoshikawa T, Shimada K, et al. Computer-assisted diagnostic system for neurodegenerative dementia using brain SPECT and 3D-SSP. Eur J Nucl Med Mol Imaging. 2009;36:831–40.PubMedCrossRef Ishii K, Kanda T, Uemura T, Miyamoto N, Yoshikawa T, Shimada K, et al. Computer-assisted diagnostic system for neurodegenerative dementia using brain SPECT and 3D-SSP. Eur J Nucl Med Mol Imaging. 2009;36:831–40.PubMedCrossRef
18.
Zurück zum Zitat Uemura T, Ishii K, Miyamoto M, Yoshikawa K. Computer-assisted system for diagnosis of alzheimer disease using data base—independent estimation and fluorodeoxyglucose—positron-emission tomography and 3D-stereotactic surface projection. AJNR Am J Neuroradiol. 2011;32:556–9.PubMedCrossRef Uemura T, Ishii K, Miyamoto M, Yoshikawa K. Computer-assisted system for diagnosis of alzheimer disease using data base—independent estimation and fluorodeoxyglucose—positron-emission tomography and 3D-stereotactic surface projection. AJNR Am J Neuroradiol. 2011;32:556–9.PubMedCrossRef
19.
Zurück zum Zitat Yanamoto T, Onishi H, Murakami T, Takahashi S, Odajima T, Uchida K. Research report: accuracy and evaluation of the stereotactic statistical imaging analysis of the brain. Nippon Hoshasen Gijutsu Gakkai Zashi. 2008;64:752–65. (in Japanese).CrossRef Yanamoto T, Onishi H, Murakami T, Takahashi S, Odajima T, Uchida K. Research report: accuracy and evaluation of the stereotactic statistical imaging analysis of the brain. Nippon Hoshasen Gijutsu Gakkai Zashi. 2008;64:752–65. (in Japanese).CrossRef
20.
Zurück zum Zitat Matsutake Y, Onishi H. Influence of smoothing process on statistical brain function image analysis. Nippon Hoshasen Gijutsu Gakkai Zashi. 2008;64:426–33. (in Japanese).CrossRef Matsutake Y, Onishi H. Influence of smoothing process on statistical brain function image analysis. Nippon Hoshasen Gijutsu Gakkai Zashi. 2008;64:426–33. (in Japanese).CrossRef
21.
Zurück zum Zitat Ohnishi H, Ota T, Takada M, Kida T, Noma K, Matsuo S, et al. Two optimal pre-filter cutoff frequencies needed for SPECT images of myocardial perfusion in a one-day protocol. J Nucl Med Technol. 1997;25:256–60.PubMed Ohnishi H, Ota T, Takada M, Kida T, Noma K, Matsuo S, et al. Two optimal pre-filter cutoff frequencies needed for SPECT images of myocardial perfusion in a one-day protocol. J Nucl Med Technol. 1997;25:256–60.PubMed
22.
Zurück zum Zitat Onishi H, Matsutake Y, Matsutomo N, Kai Y, Amijima H. Effect of prefiltering cutoff frequency and scatter and attenuation corrections during normal database creation for statistical imaging analysis of the brain. J Nucl Med Technol. 2011;36:1–6. (in press). Onishi H, Matsutake Y, Matsutomo N, Kai Y, Amijima H. Effect of prefiltering cutoff frequency and scatter and attenuation corrections during normal database creation for statistical imaging analysis of the brain. J Nucl Med Technol. 2011;36:1–6. (in press).
Metadaten
Titel
Evaluation of a novel normal database with matched SPECT systems and optimal pre-filter parameters for 3D-SSP
verfasst von
Hideo Onishi
Norikazu Matsutomo
Yuji Kai
Yoshiharu Kangai
Hizuru Amijima
Takeshi Yamaguchi
Publikationsdatum
01.01.2012
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 1/2012
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-011-0534-7

Weitere Artikel der Ausgabe 1/2012

Annals of Nuclear Medicine 1/2012 Zur Ausgabe