Evaluation of THUNDERBEAT® in open liver resection- a single-center experience

verfasst von: Bibek Aryal, Teruo Komokata, Hiroto Yasumura, Daisaku Kamiimabeppu, Maki Inoue, Kota Yoshikawa, Mamoru Kaieda, Yutaka Imoto

Erschienen in: BMC Surgery | Ausgabe 1/2018

Abstract

Background

THUNDERBEAT® (TB) (Olympus Medical Systems Corp., Tokyo, Japan) is a dynamic energy system device that simultaneously delivers ultrasonically generated frictional heat energy and electrically generated bipolar energy. TB is being routinely used in various operative procedures, however, less is known about its utility in liver resection. We, herein, report our early experience of using TB in open liver resection particularly in patients with normal or near-normal liver parenchyma.

Methods

We retrospectively reviewed the clinical characteristics, and evaluated the perioperative outcome of twenty-eight patients who underwent liver resection with TB, and twenty-four patients who underwent liver resection with basic procedure in our institution. The resection type was stratified into: major hepatectomy; resection of 3 or more than 3 Couinauds segments, and minor hepatectomy; resection of less than 3 Couinauds segments.

Results

Liver resection time (mean ± SD) in TB group with major hepatectomy was significantly shorter: 16.7 ± 8.8 compared to 62.8 ± 39.4 min in basic procedure group (P < 0.0001). Accordingly, the liver resection time (mean ± SD) in TB group with minor hepatectomy was also significantly shorter, 8.3 ± 2.9 min compared to 45.2 ± 23.9 min in liver resection with basic procedure (P < 0.001). No significant difference was observed between the groups in terms of intraoperative blood transfusion ratio, postoperative complication and postoperative liver dysfunction.

Conclusion

TB as a new energy device can offer a safe, reliable and substantially rapid liver resection particularly in patients with normal or near-normal liver parenchyma.

Background

The steady progress achieved in the outcomes of liver resection has emerged as a result of the upgraded surgical techniques that carry the advantage of technological advancements. A majority of liver resections performed these days involves the use of some energy device for cutting, coagulation or sealing, such as Cavitron Ultrasonic Surgical Aspirator (CUSA, Tyco Healthcare, Mansfield, MA, USA), the Harmonic Scalpel (Ethicon Endo-Surgery, Cincinnati, OH, USA), the Ligasure Vessel Sealing System (Covidien, Mansfield, MA, USA) and several other dissecting sealers [18].
Use of energy devices in the liver parenchyma transection is popular for faster and safer hepatectomy [6, 7]. THUNDERBEAT® (TB) (Olympus Medical Systems Corp., Tokyo, Japan) is the first device to simultaneously integrate ultrasonically generated frictional heat energy and electrically generated bipolar energy [9]. The ultrasonic technology is used for rapid cutting and precise dissection while the bipolar technology performs reliable vessel sealing. TB is considered vibrant due to its ability to gain a rapid surge in temperature with minimal thermal spreading [9].
Despite plenty of reports on the utility of other devices, little is known about the relevance of TB in liver resection. In this retrospective cohort study, we used TB as a single device, assisted by Pringle maneuver with or without infra-hepatic inferior vena cave (IVC) clamping, in open liver resection, and compared the perioperative outcomes with the patients who underwent liver resection with basic procedure.

Methods

Between April 2016 and April 2018, patients who underwent open liver resection with TB were retrospectively enrolled in the study. Similarly, a control group of patients who underwent open liver resection with basic procedure between June 2013 and February 2018 was included.
All the patients underwent routine blood tests, tumor markers evaluation, triphasic liver dynamic computed tomography (CT), and double contrast magnetic resonance imaging (MRI). Liver function was estimated using technetium-99 m galactosyl serum albumin scintigraphy, and remnant liver volume was assessed by a three-dimensional volume analyzer (SYNAPSE VINCENT; FUJIFILM Medical Co., Tokyo, Japan) in major hepatectomy.
The fibrosis score was assessed based on a five-point scale: F0 = no fibrosis, F1 = portal fibrosis without septa, F2 = few septa, F3 = numerous septa without cirrhosis, F4 = cirrhosis. Liver parenchyma with F0 was defined as normal and F1 was defined as having near normal parenchyma.

Operative technique

Following exploratory laparotomy, with the aid of intraoperative ultrasonography, the extent of the disease along with its relationship to vascular structures was assessed in all resections. The description of the procedure in each resection group is listed below.

TB resection

The liver was mobilized, and the hepatic pedicles were tapped with or without the intra-hepatic IVC taping. 1–0 Vicryl® stay sutures were placed on the hepatic edges for symmetric traction (for major hepatectomy), and the transection of the liver parenchyma was started with TB (Fig. 1). During parenchymal transection, only the pedicles equal or larger than 3rd branches of glissonian sheath or main branches of major hepatic veins were ligated and divided, the other vessels were sealed with TB. Liver resection was performed with Pringle maneuver in cycles of clamp/unclamp time of 15/5 min.

Basic resection

Inflow occlusion was performed by hepatic pedicle clamping along with ligation of glissonian or selective ligation of the portal vasculature and hepatic artery supplying the portion of the liver being resected. Liver was mobilized and stay sutures were placed on the hepatic edges, and about 1 cm of the superficial liver parenchyma was transected with an energy device like harmonic scalpel. Parenchymal transection was performed either by the crush-clamp method or using the CUSA. CUSA was mostly used in association with Kelly clamp rather than a single device. The exposed vessels or ducts were divided using sutures ligation. Intermittent inflow occlusion with the Pringle maneuver was performed in a similar way to the TB resection.
In both techniques, after removal of the specimen, hemostasis was achieved with Z-shaped sutures (4–0/5–0 Prolene®) when required, and Tachosil® was applied to the raw surface. A closed suction drain was placed at the end of the procedure.

Perioperative assessment

Liver resection time indicates the duration recorded from the beginning of the parenchyma resection to the removal of the en-bloc specimen of liver.
Liver resections were defined according to the International Hepato-Pancreato-Biliary Association, Brisbane (IHPBA) 2000 nomenclature; major hepatectomy: resection of 3 or more than 3 Couinauds segments, minor hepatectomy; resection of less than 3 Couinauds segments. Clavien-Dindo classification was applied to define postoperative complication [10]. The postoperative liver dysfunction was assessed on the basis of Balzan 50–50 criteria [11].

Statistical analyses

Statistical analyses were conducted with SPSS software (version 25; SPSS, Inc., Chicago, IL) and Graph Pad Prism (version 6.0d for Mac OS X, USA, GraphPad Software, San Diego California, USA), and were based on nonparametric tests (Mann-Whitney’s U test). The Chi-square or Fisher’s exact test was used to evaluate frequencies between categorical variables.

Results

The clinical details and perioperative outcomes of the groups are compared in Table 1 . Most of the patients were diagnosed with HCC, followed by metastatic colorectal cancer (mCRC), perihilar or intra-hepatic cholangiocellular carcinoma (CCC), gallbladder cancer (GB Ca) and a few other lesions (hemangioma, IgG4-related cholangitis, cystic adenoma, hepatolithiasis and gastric cancer metastasis). On examining the fibrosis score, except HCC patients, all other patients had normal to near-normal liver parenchyma (F0 - F1). Two patients in TB group and 4 patients in basic procedure group, with mCRC, received chemotherapy after colon resection and prior to liver resection.
Table 1
Comparative analysis of patients’ characteristics and perioperative variables
Variables
Resection with TB
(N = 28)
Resection with basic procedure
(N = 24)
P-value
Age (Years)
 Median (Range)
73.5 (57–88)
71.0 (31–88)
0.61
Sex (M/F)
 Male, n
12
20
0.003
 Female, n
16
4
 
Diagnosis, n (%)
 HCC
12 (42.9)
8 (33.3)
0.687
 mCRC
6 (21.4)
5 (20.8)
 
 CCC
4 (14.2)
2 (8.3)
 
 GB Ca
3 (10.7)
3 (12.5)
 
 Others
3 (10.7)
6 (25.0)
 
Fibrosis score (HCC)
 F0 – F1, n
9
7
1.00
 F2 – F4, n
3
1
 
Chemotherapy prior liver resection, n
2
4
0.39
Transection time (Minutes)
 Major (mean ± SD)
16.7 ± 8.8
62.8 ± 39.4
< 0.001
 Range (Min- Max)
7.0–45.0
18.0–115.0
 
Transection time (Minutes)
 Minor (mean ± SD)
8.3 ± 2.9
45.2 ± 23.9
< 0.0001
 Range (Min- Max)
5.5–16.0
9.0–95.0
 
Intraoperative blood transfusion, n (%)
8 (28.6)
11 (45.8)
0.19
Vascular clamping
 Pringle, n
17
23
0.003
 Pringle + IVC, n
11
1
 
Postoperative complication
 CD ≥ 3a, n (%)
3 (7.1)
2 (8.3)
1.00
Postoperative liver dysfunction, n
0
0
NA
M male, F female, N number, HCC hepatocellular carcinoma, mCRC metastatic colorectal carcinoma, CCC cholangiocellular carcinoma, GB Ca gall bladder carcinoma, SD standard deviation, IVC inferior vena cava, CD Clavien- Dindo NA not applicable
Liver resection time in TB group with major hepatectomy was significantly shorter than the patients who underwent liver resection with basic procedure (16.7 ± 8.83 vs. 62.8 ± 39.4 min; P < 0.0001). Similarly, the minor group who underwent liver resection with TB had also significantly shorter liver resection time than the group with basically (8.4 ± 2.9 vs. 45.2 ± 23.9 min; P < 0.001). There was no significant difference in terms of intraoperative blood transfusion ratio between the groups. There was no difference in terms of postoperative complication between the groups; CD3a was encountered in 2 patients in TB (1 bile leakage and 1 hepatico-jejunostomy failure) and 1 patient in basic procedure group (wound infection), and CD3b was encountered in 1 patient in TB (postoperative bleeding) and 1 patient in basic procedure group (wound dehiscence). Pringle along with IVC clamping was carried out in 11 patients in TB group, and only 1 patient in the basic procedure group. None of the patients in any group developed postoperative liver dysfunction. None of the cases encountered thermal injury or device-related complications in TB group. Tumor-free margin (R0) was achieved in all cases except for one case in TB group. No mortality was recorded in any group.

Discussion

Our data provide early evidence on the feasibility of TB as an advance device in open liver resection for selected cases. In this study, we reserved the use of TB in patients, particularly with normal to near-normal liver parenchyma, undergoing open liver resection. The commendable feature of TB observed in our study was an extensively rapid parenchyma transection time.
We have been using TB in open major or minor hepatectomy cases with HCC, perihilar CCC, T2 GB Ca, metastatic lesions and specifically in patients with normal or near-normal liver parenchyma. Parenchyma transection time was remarkably shorter even in major hepatectomies. In most of the cases, liver resection was completed with a single Pringle maneuver with or without infra-hepatic IVC clamping.
TB produces high heat compared to other energy devices (more than 200 °C compared to 100 °C for electrosurgery), thus, may not be suitable in cirrhotic, friable or steatotic liver parenchyma where the transection becomes riskier posed by the frequent distortion or displacement of intra-hepatic vessels by the tumor.
There is still inadequate evidence for one energy device to be considered superior to the other. TB has been frequently compared with other common energy devices [9, 1217]. TB was found superior to other energy devices in animal models [15]; however, none of the existing human studies have concluded its salient advantage over the other energy devices. In a randomized controlled trial in patients undergoing laparoscopic radical hysterectomy, TB was linked to shortened operation time and less postoperative pain [18]. Use of TB in laparoscopic colorectal surgery has also concluded the safety of the device [16, 17]. There is still a paucity of evidence on its utility in liver resection. In a recent retrospective study, Badawy et al. compared TB and ultrasonic Harmonic devices in 80 patients who underwent laparoscopic liver resection and concluded that TB is safe and effective, but not superior to ultrasonic harmonic devices [16, 17]. In this study, we compared the advantages of the TB assisted liver resection with basic procedure, our results indicate that compared to the basic procedure, TB enables exceptionally rapid parenchyma transection time with no substantial difference in complications or the transfusion ratio. In this study, the number of infra-hepatic IVC clamping was higher in the TB group that is attributed to the fact that the majority of patients in the basic resection underwent minor resection.
From the prospect of cost containment, TB retains the cost-effective benefit, in that it equips the surgeon with one instrument, without the aid of any other devices, capable of any tissue dissection and sealing efficiently. TB did not only make the dissection rapid but also offered additional benefits of reliable sealing without jeopardizing the safety and oncological clearance.
Each device used in liver resection has relative advantages and disadvantages; understanding the utility of the devices are essential in deciding which energy source better suits for a specific procedure [19]. Open liver resection with TB is our preliminary experience; we, therefore, do not stipulate a clear consensus for the procedure. Furthermore, this study has limitations including a relatively small number of patients, the retrospective design of the study and limited to the open procedure. Clamp-crushing or CUSA-based technique in liver resection is conventional, standard and still commonly used [6]; we also adopt the same technique during anatomical resections. In our cohort, we encountered a few postoperative complications including a case of bile leakage in TB group; this issue also requires careful consideration in future studies. In our experience, TB appears as an appealing device in liver transection for its rapidity, safety, and simplicity; however, the clinical benefits and the inclusion criteria for liver resection need to be better clarified from a larger randomized trial.

Conclusions

The specific use of TB in liver resection remains unclear, with little evidence available in literature.
Besides safety, the principal feature in using this device, and what endorses it peculiarly in liver resection, would be the combination of rapidity and simplicity it offers during the parenchyma transection.

Acknowledgements

We thank Miss Tomoko Matsueda, Department of Surgery, Kagoshima Medical Center, for her kind assistance.

Funding

Dr. Bibek Aryal is a fellow of Japan society for the promotion of science (JSPS) and his fellowship is supported by Grants-in-Aid for Scientific Research, JSPS KAKENHI Grant Number: 16F16420 and 18F16420.

Availability of data and materials

The authors declare that data supporting the findings of this study are available within the article. Additional data sets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
Ethical approval was not applicable because of the retrospective design of the study. Informed consent for surgery was obtained in a written document from all individual participants included in the study.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Bodzin AS, Leiby BE, Ramirez CG, Frank AM, Doria C. Liver resection using cavitron ultrasonic surgical aspirator (CUSA) versus harmonic scalpel: a retrospective cohort study. Int J Surg. 2014;12(5):500–3.CrossRef Bodzin AS, Leiby BE, Ramirez CG, Frank AM, Doria C. Liver resection using cavitron ultrasonic surgical aspirator (CUSA) versus harmonic scalpel: a retrospective cohort study. Int J Surg. 2014;12(5):500–3.CrossRef
2.
Zurück zum Zitat Campagnacci R, De Sanctis A, Baldarelli M, Di Emiddio M, Organetti L, Nisi M, et al. Hepatic resections by means of electrothermal bipolar vessel device (EBVS) LigaSure V: early experience. Surg Endosc. 2007;21(12):2280–4.CrossRef Campagnacci R, De Sanctis A, Baldarelli M, Di Emiddio M, Organetti L, Nisi M, et al. Hepatic resections by means of electrothermal bipolar vessel device (EBVS) LigaSure V: early experience. Surg Endosc. 2007;21(12):2280–4.CrossRef
3.
Zurück zum Zitat Gotohda N, Yamanaka T, Saiura A, Uesaka K, Hashimoto M, Konishi M, et al. Impact of energy devices during liver parenchymal transection: a multicenter randomized controlled trial. World J Surg. 2015;39(6):1543–9.CrossRef Gotohda N, Yamanaka T, Saiura A, Uesaka K, Hashimoto M, Konishi M, et al. Impact of energy devices during liver parenchymal transection: a multicenter randomized controlled trial. World J Surg. 2015;39(6):1543–9.CrossRef
4.
Zurück zum Zitat Ikeda M, Hasegawa K, Sano K, Imamura H, Beck Y, Sugawara Y, et al. The vessel sealing system (LigaSure) in hepatic resection: a randomized controlled trial. Ann Surg. 2009;250(2):199–203.CrossRef Ikeda M, Hasegawa K, Sano K, Imamura H, Beck Y, Sugawara Y, et al. The vessel sealing system (LigaSure) in hepatic resection: a randomized controlled trial. Ann Surg. 2009;250(2):199–203.CrossRef
5.
Zurück zum Zitat Nanashima A, Abo T, Arai J, Takagi K, Matsumoto H, Takeshita H, et al. Usefulness of vessel-sealing devices combined with crush clamping method for hepatectomy: a retrospective cohort study. Int J Surg. 2013;11(9):891–7.CrossRef Nanashima A, Abo T, Arai J, Takagi K, Matsumoto H, Takeshita H, et al. Usefulness of vessel-sealing devices combined with crush clamping method for hepatectomy: a retrospective cohort study. Int J Surg. 2013;11(9):891–7.CrossRef
6.
Zurück zum Zitat Poon RT. Current techniques of liver transection. HPB (Oxford). 2007;9(3):166–73.CrossRef Poon RT. Current techniques of liver transection. HPB (Oxford). 2007;9(3):166–73.CrossRef
7.
Zurück zum Zitat Romano F, Garancini M, Caprotti R, Bovo G, Conti M, Perego E, et al. Hepatic resection using a bipolar vessel sealing device: technical and histological analysis. HPB (Oxford). 2007;9(5):339–44.CrossRef Romano F, Garancini M, Caprotti R, Bovo G, Conti M, Perego E, et al. Hepatic resection using a bipolar vessel sealing device: technical and histological analysis. HPB (Oxford). 2007;9(5):339–44.CrossRef
8.
Zurück zum Zitat Saiura A, Yamamoto J, Koga R, Sakamoto Y, Kokudo N, Seki M, et al. Usefulness of LigaSure for liver resection: analysis by randomized clinical trial. Am J Surg. 2006;192(1):41–5.CrossRef Saiura A, Yamamoto J, Koga R, Sakamoto Y, Kokudo N, Seki M, et al. Usefulness of LigaSure for liver resection: analysis by randomized clinical trial. Am J Surg. 2006;192(1):41–5.CrossRef
9.
Zurück zum Zitat Obonna GCMR. Differences between Thunderbeat, LigaSure and harmonic scalpel energy system in minimally invasive surgery. World J Lap Surg. 2014;7(1):41–4. Obonna GCMR. Differences between Thunderbeat, LigaSure and harmonic scalpel energy system in minimally invasive surgery. World J Lap Surg. 2014;7(1):41–4.
10.
Zurück zum Zitat Dindo D, Demartines N, Clavien PA. Classification of surgical complications: a new proposal with evaluation in a cohort of 6336 patients and results of a survey. Ann Surg. 2004;240(2):205–13.CrossRef Dindo D, Demartines N, Clavien PA. Classification of surgical complications: a new proposal with evaluation in a cohort of 6336 patients and results of a survey. Ann Surg. 2004;240(2):205–13.CrossRef
11.
Zurück zum Zitat Balzan S, Belghiti J, Farges O, Ogata S, Sauvanet A, Delefosse D, et al. The “50-50 criteria” on postoperative day 5: an accurate predictor of liver failure and death after hepatectomy. Ann Surg. 2005;242(6):824–8 discussion 8-9.CrossRef Balzan S, Belghiti J, Farges O, Ogata S, Sauvanet A, Delefosse D, et al. The “50-50 criteria” on postoperative day 5: an accurate predictor of liver failure and death after hepatectomy. Ann Surg. 2005;242(6):824–8 discussion 8-9.CrossRef
12.
Zurück zum Zitat Allaix ME, Arezzo A, Giraudo G, Arolfo S, Mistrangelo M, Morino M. The Thunderbeat and other energy devices in laparoscopic colorectal resections: analysis of outcomes and costs. J Laparoendosc Adv Surg Tech A. 2017;27(12):1225–9.CrossRef Allaix ME, Arezzo A, Giraudo G, Arolfo S, Mistrangelo M, Morino M. The Thunderbeat and other energy devices in laparoscopic colorectal resections: analysis of outcomes and costs. J Laparoendosc Adv Surg Tech A. 2017;27(12):1225–9.CrossRef
13.
Zurück zum Zitat Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Ma J, et al. Evaluation of the safety, efficacy, and versatility of a new surgical energy device (THUNDERBEAT) in comparison with harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. J Laparoendosc Adv Surg Tech A. 2012;22(4):378–86.CrossRef Milsom J, Trencheva K, Monette S, Pavoor R, Shukla P, Ma J, et al. Evaluation of the safety, efficacy, and versatility of a new surgical energy device (THUNDERBEAT) in comparison with harmonic ACE, LigaSure V, and EnSeal devices in a porcine model. J Laparoendosc Adv Surg Tech A. 2012;22(4):378–86.CrossRef
14.
Zurück zum Zitat Milsom JW, Trencheva K, Sonoda T, Nandakumar G, Shukla PJ, Lee S. A prospective trial evaluating the clinical performance of a novel surgical energy device in laparoscopic colon surgery. Surg Endosc. 2015;29(5):1161–6.CrossRef Milsom JW, Trencheva K, Sonoda T, Nandakumar G, Shukla PJ, Lee S. A prospective trial evaluating the clinical performance of a novel surgical energy device in laparoscopic colon surgery. Surg Endosc. 2015;29(5):1161–6.CrossRef
15.
Zurück zum Zitat Tanaka R, Gitelis M, Meiselman D, Abar B, Zapf M, Carbray J, et al. Evaluation of vessel sealing performance among ultrasonic devices in a porcine model. Surg Innov. 2015;22(4):338–43.CrossRef Tanaka R, Gitelis M, Meiselman D, Abar B, Zapf M, Carbray J, et al. Evaluation of vessel sealing performance among ultrasonic devices in a porcine model. Surg Innov. 2015;22(4):338–43.CrossRef
16.
Zurück zum Zitat Badawy A, Seo S, Toda R, Fuji H, Ishii T, Taura K, et al. Evaluation of a new energy device for parenchymal transection in laparoscopic liver resection. Asian J Endosc Surg. 2017;11(2):123–8.CrossRef Badawy A, Seo S, Toda R, Fuji H, Ishii T, Taura K, et al. Evaluation of a new energy device for parenchymal transection in laparoscopic liver resection. Asian J Endosc Surg. 2017;11(2):123–8.CrossRef
17.
Zurück zum Zitat Van Slycke S, Gillardin JP, Van Den Heede K, Minguet J, Vermeersch H, Brusselaers N. Comparison of the harmonic focus and the thunderbeat for open thyroidectomy. Langenbeck’s Arch Surg. 2016;401(6):851–9.CrossRef Van Slycke S, Gillardin JP, Van Den Heede K, Minguet J, Vermeersch H, Brusselaers N. Comparison of the harmonic focus and the thunderbeat for open thyroidectomy. Langenbeck’s Arch Surg. 2016;401(6):851–9.CrossRef
18.
Zurück zum Zitat Fagotti A, Vizzielli G, Fanfani F, Gallotta V, Rossitto C, Costantini B, et al. Randomized study comparing use of THUNDERBEAT technology vs standard electrosurgery during laparoscopic radical hysterectomy and pelvic lymphadenectomy for gynecologic cancer. J Minim Invasive Gynecol. 2014;21(3):447–53.CrossRef Fagotti A, Vizzielli G, Fanfani F, Gallotta V, Rossitto C, Costantini B, et al. Randomized study comparing use of THUNDERBEAT technology vs standard electrosurgery during laparoscopic radical hysterectomy and pelvic lymphadenectomy for gynecologic cancer. J Minim Invasive Gynecol. 2014;21(3):447–53.CrossRef
19.
Zurück zum Zitat Scatton O, Brustia R, Belli G, Pekolj J, Wakabayashi G, Gayet B. What kind of energy devices should be used for laparoscopic liver resection? Recommendations from a systematic review. J Hepatobiliary Pancreat Sci. 2015;22(5):327–34.CrossRef Scatton O, Brustia R, Belli G, Pekolj J, Wakabayashi G, Gayet B. What kind of energy devices should be used for laparoscopic liver resection? Recommendations from a systematic review. J Hepatobiliary Pancreat Sci. 2015;22(5):327–34.CrossRef
Metadaten
Titel
Evaluation of THUNDERBEAT® in open liver resection- a single-center experience
verfasst von
Bibek Aryal
Teruo Komokata
Hiroto Yasumura
Daisaku Kamiimabeppu
Maki Inoue
Kota Yoshikawa
Mamoru Kaieda
Yutaka Imoto
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
BMC Surgery / Ausgabe 1/2018
Elektronische ISSN: 1471-2482
DOI
https://doi.org/10.1186/s12893-018-0423-2

Weitere Artikel der Ausgabe 1/2018

BMC Surgery 1/2018 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.