Skip to main content
Erschienen in: International Orthopaedics 3/2008

01.06.2008 | Original Paper

Foot pumps without graduated compression stockings for prevention of deep-vein thrombosis in total joint replacement: efficacy, safety and patient compliance

A comparative, prospective clinical trial

verfasst von: Rocco P. Pitto, Simon Young

Erschienen in: International Orthopaedics | Ausgabe 3/2008

Einloggen, um Zugang zu erhalten

Abstract

Mechanical prophylaxis with foot pumps provides an interesting alternative to chemical agents in the prevention of thromboembolic disease following major orthopaedic surgical procedures. Recent studies have suggested that the simultaneous use of graduated compression stockings (GCS) may hinder the pneumatic compression effect of foot pumps. The hypothesis of this prospective study was that the use of foot pumps without GCS does not affect the efficacy of deep-vein thrombosis (DVT) prophylaxis and improves patient compliance. A total of 846 consecutive patients admitted at a single institution undergoing total hip (THR) or knee replacement (TKR) were included in the study. The A-V Impulse System foot-pump unit (Orthofix Vascular Novamedix, Andover, UK) was used in all patients. Of these 846 patients, 46 discontinued the use of foot pumps, leaving 400 patients who received foot pumps in combination with GCS and 400 patients with foot pumps alone. Eleven patients of the stocking group (2.7%) and nine patients of the no-stocking group (2.3%) developed postoperative symptomatic DVT (p = 0.07). DVT was more frequent in TKR (10/364; 2.7%) than in THR (10/436; 2.3%). Non-fatal pulmonary embolism occurred in four of the 20 patients with symptomatic DVT, two patients each of the stocking and no-stocking groups. The foot-pump discontinuation rate of patients treated with stockings was 7% versus 4% of the patients treated without stockings (p < 0.05). In conclusion, management of patients with foot pumps without GCS does not reduce the efficacy of DVT prophylaxis after THR and TKR and improves patient compliance.
Literatur
1.
Zurück zum Zitat Andrews BT, Sommerville K, Wilson N, Browse NL (1993) The effect of foot compression on the velocity and volume of blood flow in the deep veins. J Bone Jnt Surg Br 80:198–200 Andrews BT, Sommerville K, Wilson N, Browse NL (1993) The effect of foot compression on the velocity and volume of blood flow in the deep veins. J Bone Jnt Surg Br 80:198–200
2.
Zurück zum Zitat Borghi B, Casati A; Rizzoli Study Group on Orthopaedic Anaesthesia (2002) Thromboembolic complications after total hip replacement. Int Orthop 26:44–47PubMedCrossRef Borghi B, Casati A; Rizzoli Study Group on Orthopaedic Anaesthesia (2002) Thromboembolic complications after total hip replacement. Int Orthop 26:44–47PubMedCrossRef
3.
Zurück zum Zitat Charalambous C, Cleanthous S, Tryfonidis M, Goel A, Swindell R, Ellis D (2003) Foot pump prophylaxis for deep venous thrombosis-rate of effective usage following knee and hip arthroplasty. Int Orthop 27:208–210PubMedCrossRef Charalambous C, Cleanthous S, Tryfonidis M, Goel A, Swindell R, Ellis D (2003) Foot pump prophylaxis for deep venous thrombosis-rate of effective usage following knee and hip arthroplasty. Int Orthop 27:208–210PubMedCrossRef
4.
Zurück zum Zitat Fleming P, Fitzgerald P, Devitt A (2000) The effect of position of the limb on venous impulse foot pumps. J Bone Jnt Surg Br 82:433–435CrossRef Fleming P, Fitzgerald P, Devitt A (2000) The effect of position of the limb on venous impulse foot pumps. J Bone Jnt Surg Br 82:433–435CrossRef
5.
Zurück zum Zitat Fordyce MJF, Ling RSM (1992) A venous foot pump reduces thrombosis after total hip replacement. J Bone Jnt Surg Br 74:45–49 Fordyce MJF, Ling RSM (1992) A venous foot pump reduces thrombosis after total hip replacement. J Bone Jnt Surg Br 74:45–49
6.
Zurück zum Zitat International Consensus Statement Group (2001) Int Angiol 20:1–37 International Consensus Statement Group (2001) Int Angiol 20:1–37
7.
Zurück zum Zitat Kim YH, Choi IY, Park MR, Park TS, Cho JL (1998) Prophylaxis for deep vein thrombosis with aspirin or low molecular weight dextran in Korean patients undergoing total hip replacement. A randomized controlled trial. Int Orthop 22:6–10PubMedCrossRef Kim YH, Choi IY, Park MR, Park TS, Cho JL (1998) Prophylaxis for deep vein thrombosis with aspirin or low molecular weight dextran in Korean patients undergoing total hip replacement. A randomized controlled trial. Int Orthop 22:6–10PubMedCrossRef
8.
Zurück zum Zitat Ko PS, Chan WF, Siu TH, Cheng A, Lee OB, Lam JJ (2003) Duplex ultrasonography after total hip or knee arthroplasty. Int Orthop 27:168–171PubMed Ko PS, Chan WF, Siu TH, Cheng A, Lee OB, Lam JJ (2003) Duplex ultrasonography after total hip or knee arthroplasty. Int Orthop 27:168–171PubMed
9.
Zurück zum Zitat Pulmonary Embolism Prevention (PEP) study Group (2000) Prevention of pulmonary embolism and deep vein thrombosis with low dose aspirin trial. Lancet 355:1288–1289CrossRef Pulmonary Embolism Prevention (PEP) study Group (2000) Prevention of pulmonary embolism and deep vein thrombosis with low dose aspirin trial. Lancet 355:1288–1289CrossRef
10.
Zurück zum Zitat Pitto RP, Kuehle J, Hamer H, Radespiel-Troger M, Pietsch M (2001) Hemodynamics of the lower extremity with pneumatic foot compression. Effect on leg position. Biomed Tech 46:124–128CrossRef Pitto RP, Kuehle J, Hamer H, Radespiel-Troger M, Pietsch M (2001) Hemodynamics of the lower extremity with pneumatic foot compression. Effect on leg position. Biomed Tech 46:124–128CrossRef
11.
Zurück zum Zitat Pitto RP, Hamer H, Heiss-Dunlop W, Kuehle J (2004) Mechanical prophylaxis of deep-vein thrombosis after total hip replacement. A randomized clinical trial. J Bone Jnt Surg Br 86:639–642CrossRef Pitto RP, Hamer H, Heiss-Dunlop W, Kuehle J (2004) Mechanical prophylaxis of deep-vein thrombosis after total hip replacement. A randomized clinical trial. J Bone Jnt Surg Br 86:639–642CrossRef
12.
Zurück zum Zitat Robertson KA, Bertot AJ, Wolfe MW, Barrack RL (2000) Patient compliance and satisfaction with mechanical devices for preventing deep venous thrombosis after joint replacement. J South Orthop Assoc 9:182–186PubMed Robertson KA, Bertot AJ, Wolfe MW, Barrack RL (2000) Patient compliance and satisfaction with mechanical devices for preventing deep venous thrombosis after joint replacement. J South Orthop Assoc 9:182–186PubMed
13.
Zurück zum Zitat Santori FS, Vitullo A, Stopponi M, Santori N, Ghera S (1994) Prophylaxis against deep-vein thrombosis in total hip replacement. Comparison of heparin and foot impulse pump. J Bone Jnt Surg Br 76:579–583 Santori FS, Vitullo A, Stopponi M, Santori N, Ghera S (1994) Prophylaxis against deep-vein thrombosis in total hip replacement. Comparison of heparin and foot impulse pump. J Bone Jnt Surg Br 76:579–583
14.
Zurück zum Zitat Stone MH, Limb D, Campbell P, Stead D, Culleton G (1996) A comparison of intermittent calf compression and enoxaparin for thromboprophylaxis in total hip replacement. A pilot study. Int Orthop 20:367–369PubMedCrossRef Stone MH, Limb D, Campbell P, Stead D, Culleton G (1996) A comparison of intermittent calf compression and enoxaparin for thromboprophylaxis in total hip replacement. A pilot study. Int Orthop 20:367–369PubMedCrossRef
15.
Zurück zum Zitat Thomas DP (2000) Whither thromboprophylaxis after total hip replacement? J Bone Jnt Surg Br 82:469–472CrossRef Thomas DP (2000) Whither thromboprophylaxis after total hip replacement? J Bone Jnt Surg Br 82:469–472CrossRef
16.
Zurück zum Zitat Warwick D, Harrison J, Glew D (1998) Comparison of the use of a foot pump with the use of low-molecular-weight heparin for the prevention of deep-vein thrombosis after total hip replacement. A prospective, randomized trial. J Bone Jnt Surg Am 80:1158–1166 Warwick D, Harrison J, Glew D (1998) Comparison of the use of a foot pump with the use of low-molecular-weight heparin for the prevention of deep-vein thrombosis after total hip replacement. A prospective, randomized trial. J Bone Jnt Surg Am 80:1158–1166
17.
Zurück zum Zitat Warwick DJ, Pandit Hermant, Shewale S, Sulkin T (2002) Venous impulse foot pumps. Should graduated compression stockings be used? J Arthoplasty 17:446–448CrossRef Warwick DJ, Pandit Hermant, Shewale S, Sulkin T (2002) Venous impulse foot pumps. Should graduated compression stockings be used? J Arthoplasty 17:446–448CrossRef
18.
Zurück zum Zitat Westrich GH, Sculco TP (1996) Prophylaxis against deep venous thrombosis after total knee arthroplasty. Pneumatic plantar compression and aspirin compared with aspirin alone. J Bone Jnt Surg Am 78:826–834 Westrich GH, Sculco TP (1996) Prophylaxis against deep venous thrombosis after total knee arthroplasty. Pneumatic plantar compression and aspirin compared with aspirin alone. J Bone Jnt Surg Am 78:826–834
19.
Zurück zum Zitat Westrich GH, Haas SB, Mosca P, Peterson M (2000) Meta-analysis of thromboembolic prophylaxis after total knee arthroplasty. J Bone Jnt Surg Br 82:795–800CrossRef Westrich GH, Haas SB, Mosca P, Peterson M (2000) Meta-analysis of thromboembolic prophylaxis after total knee arthroplasty. J Bone Jnt Surg Br 82:795–800CrossRef
Metadaten
Titel
Foot pumps without graduated compression stockings for prevention of deep-vein thrombosis in total joint replacement: efficacy, safety and patient compliance
A comparative, prospective clinical trial
verfasst von
Rocco P. Pitto
Simon Young
Publikationsdatum
01.06.2008
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 3/2008
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-007-0326-9

Weitere Artikel der Ausgabe 3/2008

International Orthopaedics 3/2008 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.